Kimal plc: URGENT FIELD SAFETY NOTICE Field Safety Corrective Action IMMEDIATE ACTION REQUIRED

Similar documents
Alaris / Asena GS, GH, CC, TIVA, PK, Enteral Syringe Pump

Urgent Field Safety Notice (Field Safety Corrective Action) for CORAIL AMT Neck Trials Surgical Instruments

Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts

URGENT FIELD SAFETY NOTICE: Medical Device Field Safety Corrective Action / Recall

Chester Chest Model 2400 User s Manual

Chester Chest Model 2400 User s Manual

SFHCHS132 Obtain venous blood samples

SFHCHS132 - SQA Code HD10 04 Obtain venous blood samples

DL00 04 (SFH CHS132) Obtain Venous Blood Samples

DIAGNOSTIC CARDIOLOGY

URGENT MEDICAL DEVICE RECALL NOTICE

INSTRUCTION MANUAL Elisa Baby-PICC

PRODUCT CATALOGUE OF

Venepuncture. Clinical Skills. Venepuncture. Dr Brian Jenkins (Clinical Skills Lead) Sian Williams (Clinical Skills Manager)

PEDIATRIC INJECTABLE TRAINING ARM SIMULATOR LF00958U INSTRUCTION MANUAL

Name: Oxygen Care Ltd Address: 2 Holfeld Business Park, Kilmacanogue, Co. Wicklow, Ireland URGENT FIELD SAFETY NOTICE - UPDATE

Via Federal Express Rockville, MD 20850

SEP-1 Additional Notes for Abstraction for Version 5.0b

limbsandthings.com Advanced Venepuncture Arm User Guide For more skills training products visit Limbs & Things Ltd.

AQUARIUS Continuous Renal Replacement Therapy Competency

in84034.txt at Page 1 of 9 2/25/00

Product Inspection Advisory Notice Immediate Action Required

NOTES PATIENT DETAILS

Advanced IV Hand LF01139U & LF01146U Instruction Manual

Xcela Hybrid PICC. Clinician Reference Tool

AED Program 2017 January 2017

INTERVENTIONAL CARDIOLOGY

LABELLING OF MEDICAL GAS CONTAINERS

SERVICE BULLETIN. Owner/Operators of Pacific Scientific Company Restraint Systems using Buckle Assembly Part Numbers XX and XX.

1.0 PURPOSE 2.0 SCOPE 3.0 DEFINITIONS. ANSI: American National Standards Institute. CCC: Chemical Control Centre

Infant IV Leg LF03636U Instruction Manual

Peripheral Venous Catheter Placement Simulator

Pitfalls & Practical Tips

18.1 Introduction Maintaining pedestrian-related Problems arising in the long term. infrastructure

AQUARIUS Continuous Renal Replacement Therapy with Regional Citrate Anticoagulation (RCA) Competency

BSR GPTC Z TR GM References and Reporting Page 1 of 8

Conditions of Entry - Campdraft Super Series 2018

EASYPUMP II THE FLEXIBLE SOLUTION FOR SHORT AND LONG TERM INFUSION THERAPIES

Overview. Users of this standard will need to ensure that practice reflects up to date information and policies.

Port - Body in a Box User s Manual

URGENT FIELD SAFETY NOTICE

System Flow Chart. ventilation patient connection

IVIG Educational Presentation

Aseptic Techniques. Techniques for Sterile Compounding. Pharmacy Technician Training Systems Passassured, LLC

CS CANNULATION. Step-by-Step Guide. This guide contains recommendations as provided by Dr. Seth Worley, MD POWERED BY

Patients Guide for Self Administering Intravenous Antibiotics General points

The NAMIC Dream Kit. Single Operator Control. Reduce Contrast Waste. Less Force to Inject. Compensator * Manifold. How the Compensator Manifold Works:

AQUADEX FLEXFLOW SYSTEM QUICK REFERENCE GUIDE

Varithena Delivery System Instructions for Use INSTRUCTIONS FOR USE Please read the Product Monograph before using the product.

Service Bulletin

Aylesbury & District Swimming Club 2016

ALL SPFL CLUBS SCOTTISH HIGHLAND FOOTBALL LEAGUE CLUBS SCOTTISH LOWLAND FOOTBALL LEAGUE CLUBS

CONTROL OF SUBSTANCES HAZARDOUS TO HEALTH (COSHH) POLICY

Troubleshooting Guide Aquarius

Nursing Guide Portsystems

Standard. Bag & Stand Advanced Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00442

4-(4-(3-(4-chloro-3-(trifluoromethyl)phenyl)ureido)-3-fluorophenoxy)-N-methylpicolinamide

Child CRiSis Injectable Training Arm LF03612U Instruction Manual

Varithena Delivery System Instructions for Use

HEALTH SERVICES CODE A.6 CATEGORY: RN - SNP

STANDARD OPERATING PROCEDURE ADVANCED INTRAVENOUS TRAINING ARM (S )

To assist you in preparing STAT Baby for student activities we offer the following set up suggestions:

CENTRAL IOWA HEALTHCARE Marshalltown, Iowa

Fingertip Pulse Oximeter

(2) Review the CHP at least annually and revise the document as necessary to reflect current regulatory practice.

Service Bulletin

Product reference: POA FORGAS. no relevant. list. ground anchor stakes. product. be found in these instructions. Lyon Equipment Ltd

RECONSTITUTION AND ADMINISTRATION INSTRUCTIONS FOR HEALTHCARE PROFESSIONALS

RECONSTITUTION AND ADMINISTRATION INSTRUCTIONS FOR HEALTHCARE PROFESSIONALS. Hereditary Angioedema

Unblocking the MicroSprayer Aerosolizer Model IA-1C With the High Pressure Reverse Cleaning Adapter. Instructions for Use

Sterile Visual Flow Indicator

EMERGENCY EQUIPMENT TROLLEY DAILY SIGNATURE SHEET. Please refer to the Emergency Equipment stock poster for detailed information

New. The high security connection

Critical Disposables. Non-Sterile Angiographic Accessories. for Medical Device Manufacturers & Kit Packers

Haemodialysis. Bloodlines, Tubing Systems, Accessories Product Range. Product name / Headline Subline

Spill response Procedure

ARTERIAL MONITORING. Final Approval: VP Operations, Chief Nursing Officer Distribution: ICU, PACU POLICY:

1. Connect the attached anesthesia mask to your anesthesia system using the included tubing and connectors.

RYA BERTH HOLDERS CHARTER GUIDANCE NOTES FOR. Marina Berth Holder Associations and Marina Operators

WE CARE ACCURATELY. ACCURA Medizintechnik GmbH Max-Planck-Straße Karben / Germany Tel:

NORTH DIVISION ADMINISTRATIVE INSTRUCTIONS Nominated League Contact

ADVANCED CORE DECOMPRESSION SYSTEM X-REAM. Percutaneous Expandable Reamer PRO-DENSE. Core Decompression Procedure Kit SURGICAL TECHNIQUE

Nita Newborn Model 1800 User Manual

Making Life Easier Easypump II

SAFETY DATA SHEET according to the (US) Hazard Communication Standard (29 CFR )

Kids First, Safety First

How to place the EQUIVET Ocular Lavage Kit

Rules of Participation (MiniRoos - Under 6-11)

COLLECTION OF BLOOD. Types of patients

Alfred ECMO Circuit Priming and Storage Guide

The Wincanton Supplier Trading Process. Detailed Guidance Notes PO Transactions

CHILD CRiSis INJECTABLE TRAINING ARM LF03612U INSTRUCTION MANUAL

Instructions/Procedure for Use of the Speaking Valve

SAFETY ADVISORY: HIGH PRESSURE ALUMINUM SEAMLESS AND ALUMINUM COMPOSITE HOOP-WRAPPED CYLINDERS

C O L O G Y. New. g u O N. l o t a C a. The high security connection

CHEMICAL HANDLING AND SPILL KITS GROUP WHS JAN 2017

PIMA MEDICAL INSTITUTE RADIOGRAPHY PROGRAM ARRT GENERAL PATIENT CARE COMPETENCIES CPR

PURPOSE: LEVEL: Interdependent Principal Investigator/designee Biological Specimen Repository Staff SUPPLIES: CATALOG NUMBER

12V Mini Air Compressor

Compact Ball Valves ASAHI AV VALVES. Contents. User s Manual. (Page) (1) Be sure to read the following warranty clauses of our product 1

Transcription:

(Customer Name) (Address Line 1) (Address Line 2) (Address Line 3) (Address Line 4) (Post Code) Kimal plc: URGENT FIELD SAFETY NOTICE Field Safety Corrective Action IMMEDIATE ACTION REQUIRED Adult Altius Central Venous Catheter Kit (K2CS / K2CV) - 18G Echogenic Introducer Needle Dear Customer, Kimal plc is initiating a Field Safety Notice in regards to certain lot numbers within our Altius Central Venous Catheter range of products. The affected products have been circulated as individual Catheter Kits and included within a Procedure pack as listed below. The 18G Echogenic Introducer Needle included in the kit is used to perform the venepuncture, aspiration of blood and guidewire introduction and is the component affected by this Field Safety Notice. Description of issue: Kimal plc has received customer complaints indicating that the 18G Echogenic Introducer Needle, included in these kits, is in a small number of cases, cracking during use. The reported failures have occurred during catheter insertion, when aspirating blood into the syringe to confirm correct needle placement or during infusion. During the aspiration process a crack in the hub has been observed and occasionally air bubbles have been seen in the syringe. During infusion leaks have been observed. The confirmed cases reported to Kimal plc have equated to 0.091% of stock distributed from the affected lots. Within these affected lots the same batch of introducer needles had been used. In-house testing has been carried out as part of this investigation to further evaluate the safety and performance of these needles. The reported failures could not be replicated during in-house testing as all needles met the specified test criteria. This Field Safety Notice is being circulated due to the risk that there may be a small number of needles within kits from the affected lots, where the crack is already present prior to use or could occur during use. This potential defect could cause delay to the procedure if not identified prior to use. We kindly ask you to review the instructions carefully within this Field Safety Notice, inspect your stock for the affected products and lot numbers and advise on the quantities in your inventory. Kimal is offering free of charge 18G introducer needles to be used alongside your current stock. Alternatively you can return affected stock and receive a credit or return affected stock and receive replacement stock depending on stock availability. Timeframe: Kimal plc have assigned a timeframe of 90 days to complete this Field Safety Notice.

Products affected: TABLE 1: Altius Central Venous Catheter Kit Codes: Product Code Description LOT Quantity (units) K2CS115701P ALTIUS HP 1 Lumen CVC Set 7FR 14GA 15CM 17001 472 K2CS115703P ALTIUS HP 3 Lumen CVC Set 7FR 15CM 16012 138 17002 120 K2CS115804P ALTIUS HP 4 Lumen CVC Set 8FR 15CM 17002 312 K2CS115854 ALTIUS 4 Lumen CVC Set 8.5FR 15CM 16012 280 K2CS115855 ALTIUS 5 Lumen CVC Set 8.5FR 15CM 17001 164 K2CS115904P ALTIUS HP 4 Lumen CVC Set 9FR 15CM 16012 120 17001 736 K2CS115905P ALTIUS HP 5 Lumen CVC Set 9FR 15CM 17002 250 K2CS120701P ALTIUS HP 1 Lumen CVC Set 7FR 14GA 20CM 17001 110 16012 345 K2CS120702P ALTIUS HP 2 Lumen CVC Set 7FR 14GA 20CM 17001 570 K2CS120703 ALTIUS 3 Lumen CVC Set 7FR 20CM 17002 300 K2CS120703P ALTIUS HP 3 Lumen CVC Set 7FR 20CM 16012 290 K2CS120802 ALTIUS 2 Lumen CVC Set 8FR 20CM 16012 160 K2CS120804P K2CS120853 ALTIUS HP 4 Lumen CVC Set 8FR 20CM ALTIUS 3 Lumen CVC Set 8.5FR 20CM 16012 111 17002 200 17001 180 17002 200 K2CS120854 ALTIUS 4 Lumen CVC Set 8.5FR 20CM 16012 640 K2CS130703 ALTIUS 3 Lumen CVC Set 7FR 30CM 16012 47 K2CS215703 ALTIUS Pro+ 3 Lumen CVC Set 7FR 15cm 17001 792 K2CS215854 ALTIUS Pro+ 4 Lumen CVC Set 8.5FR 15cm 17002 166 K2CS215904P ALTIUS PRO+ HP 4 LUMEN CVC SET 9FR 15CM 17002 111 K2CS215905P ALTIUS Pro+HP 5 Lumen CVC Set 9FR 15cm 16012 1146 17001 1664 17002 850 K2CS220702 ALTIUS Pro+ 2 Lumen CVC Set 7FR 20cm 17002 210 K2CS220703 ALTIUS Pro+ 3 Lumen CVC Set 7FR 20CM 17001 70 K2CS220703P ALTIUS Pro+HP 3 Lumen CVC Set 7FR 20cm 16012 349 17001 30 K2CS220854 ALTIUS Pro+ 4 Lumen CVC Set 8.5FR 20cm 17002 50 K2CS220855 ALTIUS Pro+ 5 Lumen CVC Set 8.5FR 20cm 16012 50 K2CS220905P ALTIUS PRO+ HP 5 LUMEN CVC SET 9FR 20CM 17001 187 K2CV115701 ALTIUS 1 Lumen CVC Set 7FR (14GA) 15CM 17002 10 K2CV115702 ALTIUS 2 Lumen CVC Set 7FR 15CM 16012 1000 K2CV115703 ALTIUS 3 Lumen CVC Set 7FR 15CM 17001 2000 17002 1000

TABLE 2 Altius Central Venous Catheter Kit Procedure Packs: Quantities of the above mentioned lots have also been circulated as part of a Procedure Pack. The procedure packs affected as listed below: Product Code Description LOT Quantity (units) 17E0384 3 UK-ALT1905P-51145/P1 ICU Altius 15cm 5L 9FR HP 17H0259 100 CVC Insertion VASPack 17H0642 80 UK-ALT2855-51244/P1 UK-ALT2905P-51226/P1 Altius HP 5L 15cm 9Fr Pr Critical Care Altius 5 Lumen 15cm P 17D0414 140 17E0545 30 17E0636 50 17K0133 103 17D0415 100 17E0546 150 Our records indicate that Kimal plc has shipped a number of affected product(s) to your facility; therefore we draw your attention to the following instructions: 1. Please review the content of this Field Safety Notice. 2. Communicate immediately to all those who need to be aware within your organisation or to any organisation where the potentially affected devices have been transferred. 3. Please check your stock for the affected product code(s) and lot number(s) listed above. 4. Please identify and quarantine all affected stock. 5. Please complete Appendix 1 and return to Kimal indicating quantities of stock with lot codes per quantity along with your preferred action on the tables shown. 6. Kimal plc will action your response The co-ordinating competent authority (MHRA) is aware of this action and other regulatory authorities concerned have also been alerted. We regret any inconvenience this action may have caused and would appreciate your understanding as we have taken this action in the interest of patient safety. If you have any questions, or would like further assistance with this Field Safety Notice, please contact the following designated person. Vigilance / Compliance Executive: Mr. Paul Beard vigilance@kimal.co.uk Reference: FSCA 18359 Yours Sincerely,.. Kimal Plc Attachments Appendix 1) Confirmation of Receipt of Field Safety Notice

Appendix 1 Confirmation of Receipt of Field Safety Notice Kimal plc: URGENT FIELD SAFETY NOTICE Adult Altius Central Venous Catheter Kit K2CS and K2CV Kimal plc: Urgent Field Safety Notice KS837/838PS Bloodline Set For Braun Dialog, KS837838HDF1 Braun Dialog Limited Online Set Type of Action: Field Safety Corrective Action Please complete this form and return a copy either by FAX or email to confirm that you have received this confirmation, once all information has been obtained. Fax: 0845 4379541 Email: vigilance@kimal.co.uk Customer Name and Address: (Please Print) Reply confirmation completed by: (Please Print Name) Title: (Please Print) Telephone Number : Email : We confirm: We have read the Field Safety Notice. We have communicated the information to staff and other services / departments / units / facilities who need to know. We have none of the affected stock We have distributed the affected products to a third party organisation and will provide Kimal plc with these details. We have the following product that requires the following action (Tick appropriate action):

For stock identified in Table 1 - Altius Central Venous Catheter Kit Codes please complete the table below: TABLE 1 Affected Stock Product Code: Lot Number: Quantity (pcs): ACTION Tick appropriate action and credit note 18G introducer Needle Replacement and receive alternative For stock identified in Table 2 - Altius Central Venous Catheter Kit Procedure Packs please complete the table(s) below depending on procedure pack received: TABLE 2 Affected Procedure pack Stock Product Code: UK-ALT2855-51244/P1 UK-ALT2905P-51226/P1 Lot Number: Quantity (pcs): ACTION Tick to confirm you are following the instructions below This procedure pack contains a two part introducer needle Needle two part Introducer 18G 7cm,. Only use this two part introducer needle and dispose of the introducer needle provided in the Altius Kit This procedure pack contains a two part introducer needle, Needle 18G x 6.35cm Introducer. Only use this two part introducer needle and dispose of the introducer needle provided in the Altius Kit TABLE 2 Affected Stock ACTION Tick appropriate action Product Code: Lot Number: Quantity (pcs): 18G introducer Needle Replacement and credit note and receive alternative UK-ALT1905P- 51145/P1