SOP for use of the SX19 Stopped-Flow Instrument

Similar documents
5890II GC Standard Operating Procedure 9/2/2005

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

SOP: Buck Scientific BLC-20P HPLC Operation

AKTA pure 25 New Owner s Intro

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Standard Operational Procedure for Capillary Electrophoresis

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

PURDUE UNIVERSITY TEFLON CORE PCB THRU-HOLE PREPARATION FOR BLACK HOLE APPLICATION RF LAB. Professor in charge: William Chappell MSEE 289

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Using the Akta Prime plus October 22, 2012

Armfield Distillation Column Operation Guidelines

Determination of Zn using Atomic Absorption with Multiple Standard Additions

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

Evaluation copy. Vapor Pressure of Liquids. computer OBJECTIVES MATERIALS

User Service Manual SX Series Spectrometers

Determination of Zn using Atomic Absorption with Multiple Standard Additions

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Physics Experiment 17 Ideal Gas Law Qualitative Study

Cover Page for Lab Report Group Portion. Flow Visualization in a Water Channel

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Vapor Pressure of Liquids

Microfluidic Demonstration

FOURIER TRANSFORM INFRARED SPECTROSCOPY

Vapor Pressure of Liquids

ShellPa. Standard Mechanical Cell Stretch System Model No: NNMS Serial #: User Manual

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

Stopped-flow Kinetics

The Gas Laws: Boyle's Law and Charles Law

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

QFM-4000 User s Manual Version 2.0

Purpose and Scope: Responsibilities: Warnings and precautions: Background Beers law

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

VEVO 2100 ULTRASOUND USER GUIDE

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

Pipetting and Determining Protein Concentration

Procedures for operation of the TA Instruments DSC

System Overview. TCD Detector temperature setpoint Regulator

NEULOG OXYGEN LOGGER SENSOR GUIDE

Gas Absorption Draft Standard Operating Procedure

XpressFill XF2500 Counter Pressure Filler. Operating Instructions

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

Exploring the Properties of Gases. Evaluation copy. 10 cm in diameter and 25 cm high)

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

Automated Liquid Handling Station

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

6800 Maintenance Instruction System Flush Procedure

Vapor Pressure of Liquids

STANDARD OPERATING PROCEDURE TEMPLATE

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

SPLIT HOPKINSON PRESSURE BAR OPERATIONS MANUAL 5/9/2014

MULTITUBE BUBBLE TUBE BY ROMPA (U)

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

FLUORESCENCE DETERMINATION OF OXYGEN

BASIC RULES OF USAGE, OLIS DSM-20 CD Management The instrument is currently managed by the Analytical Biophysics Core Facility.

Traceability and Evidential Breath Analyser

[USING THE VITROBOT April 9, Using the Vitrobot

Liquid Handling Systems

ELIMINATOR COMPETITION DRAG RACE Program Manual Firm Ver 4.11

BIL 151 Laboratory Enzymes: Practicing the Protocol

Operation of the mask aligner MJB-55

Prep LC Training Manual S.V.

RhinoPro Cartridge Gun Manual

Hydro user manual. Malvern Instruments Limited

SPUTTER STATION STANDARD OPERATING PROCEDURE

Exploring the Properties of Gases

Version BA D

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

uemis CONNECT: Synchronisation of the SDA with myuemis

SHIMADZU LC-10/20 PUMP

Quintic Automatic Putting Report

XC2 Client/Server Installation & Configuration

6900 Maintenance Instruction System Flush

XpressFill XF4500C Counter Pressure Can Filler. Operating Instructions

1)! DO NOT PROCEED BEYOND THIS MARK

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

XpressFill XF4500 Counter Pressure Filler. Operating Instructions

BSS 2 Bottle Sampling System. Instruction manual Version 1.8

Standard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation

Observations Possible Causes Recommended Actions. Turquoise color

50KN stress rig for use on SALSA

PerkinElmer Aanalyst100

Ideal gas law. Introduction

Instruction Manual Otoflash G171 UV-flash-device for Light Curing

FROG-4000 TM Chemical Analysis System

Pipettor. User Manual

ROMPA SUPA SPARKLE BUBBLE TUBE 20288

USER MANUAL OPERATION AND THE USE OF A CAR WITH. Diego G3 / NEVO SEQUENTIAL GAS INJECTION SYSTEM

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

Standard Operating Manual

Warnings: Notes: Revised: October 5, 2015

RayPette Plus Autoclavable Pipette. User Manual

Calibration of a GC955

ELE INTERNATIONAL COMPRESSION TESTER Standard Operating Procedure

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

Transcription:

SOP for use of the SX19 Stopped-Flow Instrument The main components of the stopped flow system are: Always book the system before use. Starting up and preparing the SX19 for use: 1. Turn on the compressed air supply. An 8 bar gas supply power the pneumatic rams and stop valve of the sample-handling unit. Open the large lever value on the wall and check the pressure is 8 bar. 2. Turn on the water bath. Turn on the water bath on the floor under the instrument. Check there is sufficient water covering the heater coil, and top up with mq water as necessary (there is a large plastic container by the sink in 0.14 for this purpose). (Let Katherine or Joe know if there is air in the sample handling unit, which the water bath fills and controls.) 3. Turn on the lamp. Power on the lamp power supply unit using the rocker switch at the rear. The green light on the arc lamp ignite switch should illuminate. Wait for at least 30 seconds before pressing the green button firmly to ignite the lamp. Ensure the shutter between the lamp housing and the monochromator is in the open position. You can check the lamp is on since light comes out of the slit in the lamp housing: 1

If the lamp fails to ignite, then wait for two minutes before pressing the green button again. 4. Turn on the electronics unit. Power on the electronics unit via the black rocker switch on the front panel. There are three light emitting diodes, one green and two red. The green diode will flash for a short time after the unit is powered on, but should then remain on constantly. Whenever the software is running the red diodes will flash continuously. It is recommended that if the instrument is to be used daily the electronics unit is left powered on. 5. Power on the computer and load the Pro-Data software. Power on the computer and log on. The default account is lab which has the password bcssxxxx (see the sticker on the side of the PC). Load Pro-Data SX using the desktop icon: 6. Load the Pro-Data Viewer software and set the file save location and seed names. Start the Pro-Data Viewer program using the Viewer icon on the Pro-Data Viewer toolbar: Navigate in the Pro-Data Viewer file browser to your preferred save directory and right click on it selecting Set Working Directory Here. New data files will now automatically be saved in this directory: Now open the Files Names dialog box by clicking on the icon on the toolbar: Enter a name for the new data files and baseline file (where relevant). Subsequent files of each type will be numbered starting from the values set in the window: 2

Preparing the sample handling unit: 1. Configure the flow lines. The following description applies to a single-mixing stopped flow experiment with a 1:1 mixing ratio. (See the manual for details of how to perform more complicated experiments.) The single-mixing experiment always utilises the left-hand drive ram and the reagents are always loaded in the F (left) and C (right) drive syringes: Check that the instrument is in the configuration above before proceeding. (The drive volume is pre-set to approximately 120 μl. Do not adjust this under any circumstances.) 2. Flush the system with water. The stopped-flow unit should have been left from previous use with mq water/ethanol in the flow circuit and cell. In any event it is best to ensure that mq water/buffer is in the flow circuit and cell prior to loading sample. To do this: a) Set the beige drive valves to the load position: 3

b) Empty any contents of the drive syringes into waste reservoir syringes by pushing the drive syringe piston upwards into reservoir syringes. Always be careful when handling the drive syringe pistons as they can break. c) Replace the reservoir syringes with ones containing clean mq water/buffer. d) Load each drive syringe with about 1 ml of water/buffer by pushing down on each reservoir syringe piston while at the same time pulling on the drive syringe piston. Take care not to introduce air bubbles into the flowcircuit: one of the most common causes of artifacts in stopped-flow data is the presence of air bubbles. Repeat the up and down motion a few times to develop your technique for eliminating air bubbles*. * In general, to minimise the introduction of bubbles and to facilitate their removal, with the drive valves in the load position, shunt liquids back and forth gently through the drive valves whilst keeping the liquid compressed from both ends of the flow (by applying downward pressure on the piston on the reservoir syringe and upward pressure on the drive syringe piston). When loading drive syringes avoid simply pulling the syringes down, to suck liquid from the reservoirs, as this will encourage cavitation and bubble formation. Fill them by both pushing liquid in from the reservoirs above, while pulling the drive syringe pistons down, to keep a positive pressure in the valve and syringe cavity. e) With the drive valves still in the load position, raise the drive ram until it is in contact with both drive syringe pistons. f) Now set the beige drive valves to the drive position: g) Select Drive on the Sample handling unit panel (using the Drive function rather than the Acquire function ensures that there will be no data collection and generation of a data file*). * It is however more efficient to run washes using Acquire: use < 10 data points and < 1 s time frame, tick the panel labelled repeats and enter 10 as the number of repeats. The files generated can easily be deleted afterwards. h) Do at least 5 repeat drives to ensure that the flow circuit and cell contents have been completely displaced by the clean water. 4

Before executing a drive always check that the drive valves are in the drive position (not the load position) and the drive ram is in contact with the syringe plungers. Failure to do so may result in damage to the flow circuit. Stopped-Flow Absorbance Kinetics 1. Select Absorbance mode. On the Signal panel of the Pro-Data main window select Absorbance from the drop down list. Ensure that AutoHV is ticked: 2. Select kinetics mode, set timebase and points. Ensure that Kinetics mode is selected on the Sequencer panel: Select External from the drop down list in the Trigger panel to activate stopped-flow operation: On the Timebase panel choose an appropriate acquisition time based on the expected timescale of the reaction. The Points should normally be set to a value between 400 and 1000 data points. 3. Select monochromator wavelength and slit width. On the Monochromator panel enter the selected wavelength: Using the dials on the front of the monochromator set the slit widths to give the required absorbance signal. 5

4. Zero the detector. Fill the optical cell with a suitable reference material (e.g. water or other solvent). Click the Reference button in the Signal panel to zero the absorbance detector. 5. Load the samples. Samples may now be loaded into the drive syringes. At least 1.3 ml is necessary for a complete series of approx. 25 drives. Do not to push/pull the drive syringe pistons while they are in the drive position after loading. Equalizing the piston levels should only be done in the load position, otherwise the sample from one syringe (i.e. enzyme) is easily pushed into the other syringe (i.e. substrate). The reaction will then occur prematurely in the drive syringes and you will have to discard the contaminated material, and then wash and load the syringes again using fresh samples. 6. Acquire data. Ensure that the drive valves are in the drive position and click the Acquire button to perform a drive and collect the data. The first few drives will consist of buffer and then diluted sample. By the fourth or fifth drive clean sample should be mixed with each drive and results should be consistent. Absorbance measurements will now be in units of absolute absorbance since water was used when measuring the reference. It is possible to measure the reference with something other than water in the cell, in which case the measurements will now be relative absorbance. This may be convenient when measuring small absorbance changes on significant backgrounds. Collecting a reference when only one reagent (often colourless) is pushed into the cell is one typical option as is reference collection to the chemical end point of the chemical reaction under investigation. When you have finished your experiments: 1. Quit the software and save your data to a clean USB drive. 2. Power off the computer. 3. Power off the lamp. 4. Power off the electronics unit. 5. Turn off the water bath. Cleaning the observation cell and flow circuit after use: 1. Flush the drive syringes. With the beige valves in the load position, push up the drive syringes to empty them as far as possible (taking care as they are fragile). Replace the reservoir syringes with ones containing mq water and flush the drive syringes with water. Replace the reservoir syringes with ones containing 2% Hellmanex (provided next to the instrument; a supply of neat Hellmanex is under the fume hood and mq water is in the plastic container by the sink in 0.14 if this solution runs out). Flush the drive syringes with at approx. 5 ml of this cleaning solution. 2. Wash the cell and stop syringe. The cleaning solution can be pushed through the cell, stop syringe and associated circuitry using repeated drives as described on pages 4-5 (put the valves in the load position and the drive ram in contact with both drive syringe pistons, then put the valves in the drive position and perform repeated single Drives, or Acquire with repeats). However, this is unlikely to fully replace the liquid in the stop syringe. To flush the stop syringe, cell and associated circuitry more completely it is necessary to perform manual drives. To do this, put the drive 6

valves in the load position and move the drive ram so it is in contact with the syringe plungers, then put the valves in the drive position. Next, turn off the pressure using the large valve handle on the wall, and then release residual pressure by depressing the in-line red button behind the sample-handling unit: Identify the stop valve, which will be in the inject position (facing forwards): Turn it by hand 180 clockwise ( waste position; it will only turn if there is no pressure left in the system) and empty the stop syringe manually by pushing it upwards before returning the stop valve to its normal position. While doing this, take great care not to (a) move the stop assembly, which is only weakly fixed to the sample handling unit, or (b) touch the copper trigger plate. Fresh mq water/buffer can then be introduced into the drive syringes ( load position) and flushed through the flow circuit manually (by raising the drive ram upwards against the drive syringe pistons in the drive position). The stop syringe will fill, and can be emptied again by turning the stop valve to waste. This procedure can be repeated (e.g. x5) until the user is confident that the circuit is clean and also free from air bubbles. Wash the stop syringe using the manual drive procedure described above. Repeat approx. 5 times until you are confident that the circuit is clean. Leave it filled with 80:20 water:ethanol when not in use. 3. Final wash. Finish by replacing the reservoir syringes with ones containing clean purified water and 20% ethanol. Flush the drive syringes with this solution and leave them filled with 80:20 water:ethanol when not in use. 4. Turn off the air supply. Turn off the compressed air supply on the wall, and release the residual pressure by depressing the red in-line button behind the sample-handling unit. 5. Tidy up! Please keep the area around the instrument clean and clear. Re-fill empty bottles and wipe up any spills immediately. Katherine Stott and Joe Maman, January 2019 7