Paradise Pump Series Operator s Manual

Similar documents
Med Aire Alternating Pressure Pump and Pad System

Operator s Manual. Dynamic Low-Air-Loss Therapy with Alternating Pressure

AltaDyne AutoFloat. Alternating Pressure Management Plus. Model # Model # OPERATORS MANUAL

SC505 Sof Care Inflator. Operating Instructions and Service Manual

Alternating Pressure / Low Air Loss Mattress System

APH166 Morpheus Air USER MANUAL. Apollo Healthcare Technologies Ltd. Holme Street, Liversedge, West Yorkshire, WF15 6JF.

Operation Manual. H & R Healthcare 1750 Oak Street Lakewood, NJ Phone: Fax:

Key Product Features. Primary Use: Aggressive Treatment of Pressure Ulcers

Instruction Manual for Configura Cushionair Portable Pump

USER MANUAL. AltaDyne XT Model A AltaDyne XD Model A Alternating Pressure / Low Air Loss Replacement Mattress System

RIK Fluid Overlay INSTRUCTIONS FOR USE

Comfortmax Thermax Hot/Cold Water Circulation System. Instructions for Use

SELECT SERIES LS200 ALTERNATING PRESSURE / LOW AIR LOSS MATTRESS SYSTEM

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual

Diamond 8 Plus. Alternating + Micro Low Air Loss Pressure Relief System. User Manual. Distributed by: Quart Healthcare Inc.

PressureGuard APM 2 and PressureGuard APM 2 Safety Supreme

Pegasus Aircare Mattress Overlay System Models 2200 and 2201

Air Plus User Manual

77-I0PN230 Rev. A. Model 1030 User Manual

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

Pressure Regulator. Instructions for Use

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS

User Manual. soft air simplex wds. Health. Security. Independence. Pressure ulcer prevention and treatment-system Assembly and Operation instructions

SELECT SERIES LS300 ALTERNATING PRESSURE / LOW AIR LOSS MATTRESS SYSTEM

Dialflow Regulator. Instructions for Use

DYNAMIC COMPRESSION THERAPY

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS

Panacea AIR PRO MATTRESS. Owner s Manual. Please keep and refer to this Owner s Manual.

Manual for HoverJack Devices

User Guide Antidecubitus systems EURO BASIC

PRESSUREGUARD APM OWNER S MANUAL. PressureGuard APM Bariatric. Span-America Medical Systems, Inc. Greenville, SC

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified

Instructions for Cleaning & Maintaining Air Compressor

LYMPHA PRESS MANUFACTURED BY MEGO AFEK IMPORTED BY LYMPHA PRESS USA, Ltd. USERS MANUAL LYMPHA PRESS MODEL 201-M

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

HZ-300 HAZER MACHINE USER MANUAL. Congratulations on the purchase of your new ANTARI HZ-300 Haze Machine.

User Manual Models: S36, S40, S50

Hi Flow Stroller H I F L O W S T R O L L E R

OPERATION MANUAL ASPEN MASSAGE TABLE OR MULTI-PURPOSE LIFT TABLE. (800) (760) Fax: (760)

Operating Instructions for BAIR22-6 AIR-POWERED CRIMPING TOOL

DuPont Personal Protection Instruction Manual for Universal Pressure Test Kit No

Pressure Regulators. Instructions for Use

MA60 Series. MA60 Alternating Pressure System. MA65 Alternating Pressure On-Demand Low Air Loss System. Owner s Operator and Maintenance Manual

Oxygen Dialflow Meter. Instructions for Use

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

400H HARDMOUNT AIR COMPRESSOR KIT PART NO H HARDMOUNT AIR COMPRESSOR KIT PART NO

770000S CLINICAL MAX LOW AIR LOSS WITH ALTERNATING PRESSURE SYSTEM USER MANUAL

12V AIR STATION ALL-IN-ONE

Lateral Turning System. User Guide.

MA90 Series Mattresses. MA95 Series Mattresses

User Manual. 10. Dimensions and Ordering Codes. Integrity Alternating Pressure Mattress System. Alternating Pressure Mattress Systems

User Manual. This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.

INFLATION STATION. Cordless OWNER S MANUAL MODEL NO: RCP-C65B

Stroller. Sprint S T R O L L E R S P R I N T

Owner s Guide. ElevateSUP.com

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank

POWERPRESS UNIT. User Manual. Gradient Pneumatic Sequential Compressor. Neomedic

MODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE

42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only)

450P AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

Model PSI Compressor with 3-Gallon Air Tank 12VDC

Cascade Patient Transfer Lift System Safe Operation & Daily Maintenance Instructions

Instruction Manual. Low air loss/alternating pressure mattress

USER MANUAL LTH-255A Hand Piece

INSTRUCTION MANUAL MANUAL INFLATION BLOOD PRESSURE MONITOR

Vapotherm Q50 Compressor. Instructions for Use

(DynaLaL III) Dynamic True Low Air Loss Mattress Replacement System

MODEL NUMBER: P-A AUTOMATIC PORTABLE COMPRESSOR

100C Air Compressor Kit

450P- RV AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

Contents. Introduction...1 About Breeze... 1 Breeze Overlay... 2 Breeze Mattress Replacement... 2 Breeze Pump... 4 Controls and Indicators...

EasyLife INSTRUCTIONS FOR USE. Respironics Inc Murry Ridge Lane Murrysville, PA USA JDH 3/25/10

MODEL NUMBER: P PORTABLE COMPRESSOR

200 PSI COMPRESSORS - MODEL NUMBERS

M5400 Operators Manual

Pressure Relief Valve Instruction Manual

[ACT 005] Manual E 5/25/05 9:24 PM Page 3 MULTI-FUNCTION

Oxygen Dialflow Meter. Instructions for Use

Portable oxygen concentrator

May 2016 Med Tips. Helpful Hints and FAQs. Replacing AirSep VisionAire Mixing Tank

Owner s Manual & Safety Instructions

Model 1700 Patient Operating Instructions. Model 1700

Cocoa Patio Pond with Lit Spillway

Operator s Manual. The Bullet Blender Gold BB24-AU, BB5E-AU

Leglifter. user instructions. Simple solutions for everyday independence

! Warning, refer to accompanying documents.

12 Volt Inflator. Operator s Manual. Model: ICC12F, ICC12FC, ICC12FCN, ICC12FCM, ICC12N

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

CART FOR CENTURION 1500+

Operations/Maintenance Name of Manual

User Manual DIRECTHEALTHCARESERVICES.CO.UK

HoverJack. Air Patient Lift. User Manual. From HoverTech International.

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

microair MA65 Series Alternating Pressure Low Air Loss Mattress User Manual MA65, MA65B42, MA65B48, MA65RSR

Everyday INFLATION STATION CORDLESS

200 PSI FAST-FILL AIR SOURCE KIT

200 PSI HIGH-FLOW AIR SOURCE KIT

Instruction Manual LIMITED 1 YEAR WARRANTY. Hydraulic Punch Driver Read this material before using this product.

420C AIR COMPRESSOR KIT PART NO C AIR COMPRESSOR KIT PART NO

250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO

Transcription:

Paradise Pump Series Operator s Manual P/N 12421-000 9/05

Paradise Pump Operator s Manual Important Before using the Paradise Pump Alternating Pressure Relief Systems, please read and understand this manual and all safety precautions prior to each application. If you have purchased a Paradise Pump through a local dealer and you have questions or are experiencing problems, please contact that local dealer for assistance. For additional assistance, contact our Customer Service department at: Gaymar Industries, Inc. 10 Centre Drive Orchard Park, NY 14127-2295 USA Telephone: 1 800 828-7341 (716) 662-2551 Fax: 1 800 993-7890 (716) 662-0748 Warranty The Paradise Pump System is warranted free of defects in material and workmanship for a period of one (1) year. A copy of the warranty is available upon request. Gaymar disclaims all implied warranties including, but not limited to, the implied warranties of merchantability and of fitness for a particular purpose. Symbols Attention, consult accompanying documents Class II, double insulated equipment Type BF equipment Dangerous voltage Protective earth ground Latex Free 1.0 Contraindications Air support treatment is not recommended when spinal stability is a concern. 2.0 Indications for Use This device is intended to assist in treating and preventing pressure ulcers. 2.1 Safety Precautions When using electrical products, especially when children are present, basic precautions should always be followed, including the following: DANGER Electric Shock Hazard. Do not remove pump cover. No user serviceable parts inside. To reduce the risk of electrocution: Do not place pump where it can fall or be pulled into a tub or sink containing water or other liquids. Do not reach for a pump that has fallen into water. Unplug immediately. 3

Paradise Pump Operator s Manual WARNING To reduce the risk of burns, electrocution, fire, or personal injury: Never operate this pump if it has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water. Return the pump to your Dealer for examination. Keep the cord away from heated surfaces. Do not place pump in bed, in contact with the patient, or under sheets or other coverings. Deflate pad/mattress before CPR or CPR will be ineffective. CAUTION Some medical conditions may not respond to treatment of this type. Patient s skin condition should be inspected regularly. Consult physician if any redness or skin breakdown occurs. Keep bedding dry. Use incontinence pads if necessary. Failure to keep the patient s skin clean and dry may lead to maceration. For grounding reliability, plug only into a properly grounded outlet. Make certain all mattress straps are secured to the bed frame to prevent mattress from sliding and causing patient injury. Use minimal layers of sheeting and incontinence pads. Too many layers between the patient s skin and the support surface will reduce the effectiveness of the system. Do not pull linens tightly over mattress. Tight sheets can cause hammocking and reduce effectiveness of treatment. Portable and RF communications equipment can affect medical electrical equipment. 3.0 Description Gaymar Paradise Pump Alternating Pressure Relief Systems are designed for the prevention and treatment of pressure ulcers. A system consists of a Paradise Pump air control unit (pump) and an alternating pressure air overlay or alternating pressure mattress. Compatible Gaymar overlays and mattresses contain two sets of air cells that are alternately inflated and deflated once every 5 to 6 minutes. To minimize movement, inflation of one set of air cells begins before the alternate set is fully deflated. The Gaymar Paradise Pump is available with friction connectors. Pump model number PAR30 F has friction connectors and is designed for use with Gaymar Alternating Pressure Overlay model numbers PAR01, PAR02, PAR03, E301, EFF301, EFF302 and P500. 4.0 Setting Up the Paradise Pump System 4.1 Setup with Overlays PAR01, PAR02, PAR03, E301, EFF301, EFF302 and P500: 1. Place pump on flat surface or suspend pump on end of bed using attached hooks. 2. Position overlay on existing mattress, smooth and flat, with hose connections at the foot end of the bed. 3. Connect the hose assembly between the overlay and the pump. 4. Plug the power cord into a properly grounded outlet. 5. Adjust pressure control knob to highest setting for quick inflation and turn pump on using green on/off switch. 6. When applicable, tuck end flaps of overlay under mattress. 7. Wait approximately 10 minutes for overlay to inflate. 8. Apply sheet loosely over the top of the overlay. Linens should be loose to prevent hammocking. 9. Position patient on overlay. Adjust pressure control knob for patient comfort. 4

Paradise Pump Operator s Manual 10. Perform Hand Check. A Hand Check must be performed every 8 hours to verify proper operation. To perform the Hand Check: With the patient on his or her back, slide hand, flat and palm up, between the overlay and the mattress. Hand should be directly under the air cell that is under the patient s buttocks (or other bony area). Wait for full inflation of the air cell directly above hand. If the patient s body is not in direct contact with hand, the system is operating correctly. If during full inflation of air cell, the patient s body is in direct contact with flat hand, the system is not operating properly. Adjust the pressure control to a higher setting. Wait 10 minutes and repeat the Hand Check. If the Hand Check fails, check that the hoses are not kinked or pinched. If repeated Hand Check fails and hoses are not kinked, contact Dealer or Gaymar for further instruction. 5.0 CPR (Cardio Pulmonary Resuscitation) WARNING Deflate the overlay or mattress before CPR or CPR will be ineffective. To deflate overlay or mattress for CPR: 1. Disconnect the hoses from the pump. The overlay or mattress will deflate in approximately 20 seconds. 2. Proceed with CPR procedures. 6.0 System Cleaning, Care and Storage 6.1 Cleaning WARNING Disconnect the power cord from the outlet before attempting to clean the pump. Do not heat or steam autoclave any component of the system. 1. To clean, use soap, water and a clean cloth to wipe down the pump, power cord, hoses and overlay or mattress. Do not use abrasive cleaners, alcohol or alcohol-based solutions. Wipe dry with a clean, dry cloth. Note: Blood and other body fluids must be thoroughly cleaned from all surfaces before applying disinfectants. 2. Apply any EPA approved disinfectant to the external surfaces of the pump, hoses and overlay or mattress. Allow solution contact time according to manufacturers instructions. The contact time is what makes disinfection effective. 3. Wipe down the overlay or mattress with a clean, dry cloth to remove any excess disinfectant. 5

Paradise Pump Operator s Manual 6.2 Care and Storage Pump Check the power cord and plug for abrasions or excessive wear. Coil power cord and place both cord and pump in plastic bag for storage. Overlay, Hoses Deflate the overlay by disconnecting hoses from pump. Check the hoses for kinks or breaks. Replace if necessary. Do not place sharp objects in contact with overlay. Holes in the overlay may compromise pressure relief therapy. Heavy Duty overlays (PAR03 and P500) are furnished with a Pinhole Repair Kit (P/N 78061-000) and include instructions for pad repair. Follow instructions in kit. Single patient use overlay should be discarded after therapy is discontinued. Do not discard hose assembly. Overlay should be rolled for storage, not folded. Roll the deflated overlay starting at the head end. Coil the hose assembly loosely. Place overlay and hoses in plastic bag for storage. 7.0 Troubleshooting Symptom: Overlay or Mattress does not inflate or is too soft. Corrective Action Make certain the air hoses are not kinked, cracked, or split. hose couplings are fully engaged to pump Verify the 6

8.0 Technical Specifications Paradise Pump Operator s Manual Catalog PAR30 F Input 108-126V, 60Hz, 1.0A Overcurrent Protection 1A Operating Temperature 60 90 F (16-32 C) Classification Class 1 grounded Type BF equipment UL60601-1, CSA C22.2 IPXO, enclosed equipment without protection against ingress of water. Not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen (O 2 ) or nitrous oxide (N 2 O). Continuous operation Electromagnetic Compatibility EN60601-1-2:2003 CISPR11 Classified as Class B, Group 1 ISM Equipment Ground functionality can be checked by measuring resistance between the function earth screw (accessed through small hole on back of unit adjacent to the ground symbol) and the grounding pin on the line cord attachment plug. The value shall be less than 5 ohms. Functional test requirements are available upon request. 7

10 Centre Drive Orchard Park, New York 14127-2295 USA www.gaymar.com Phone 1 800 828-7341 716 662-2551 Fax 1 800 993-7890 716 662-0748 In Canada contact: Waterloo Bedding Company, Ltd. 825 Trillium Drive Kitchener, Ontario, N2R 1J9 Phone 1 800 203-4293 519 742-4447 Fax 519 742-3699