Performance Verification of BioTek s Precision 2000 using Artel s Multichannel Volume Verification (MVV ) System

Similar documents
A Flexible, Multi-Purpose, Multi-Detector HTS Automated Platform

MVS Volume Verification Using Any Microtiter Plate or Small Volume Container, Such as a Tube or Vial

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor

New Validation and Optimization Techniques for the PerkinElmer JANUS platform using the ARTEL Multichannel Verification System (MVS )

Artel MVS as a Tool For Measuring Liquid Mixing Efficacy in Microplates

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Transferpette. Lightweight precise rugged reliable!

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Transferpette. Lightweight precise rugged reliable!

Single Channel Pipette Calibration and Operator Competency Assessment Using a Dual-Dye Ratiometric Photometry System

Air Displacement Pipetting Modules. Total Integrated Liquid Handling

Eliminating Sources of Pipetting Error in the Forensic Laboratory David M. Epstein Director of Scientific Operations. Abstract.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Application Note AN-107

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

25/04/56. LTC Carmen A. Bell Chief, Laboratory Operations USAMC-AFRIMS

Pipetting and Determining Protein Concentration

Rat myeloperoxidase-antineutrophil cytoplasmic antibody(mpo-anca) ELISA Kit

Air Displacement Pipetting Module. Total Integrated Liquid Handling

Improving Pipetting Techniques

HumiSys HF High Flow RH Generator

MULTICHANNEL. Labnet International, Inc.

The pipettes cover a volume range from 0.1µl to 10ml.

RayPette Plus Autoclavable Pipette. User Manual

Pipettor. User Manual

Qi Litre Ultrasonic Cleaner

ali-q TM Gravimetric Verification Procedure

Standard Operating Procedure. Air Displacement Pipette Calibration

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING

WHEATON. Peristaltic Pumps. Advancing the Science of Liquid Handling

Pressure calibration. 20

High-Throughput. Faster, Better 96 & 384-Well Pipetting Ultra Efficient and Extremely Versatile

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

Time Pressure Dispensing

Chapter 2. Pipetting techniques. Guide to Pipetting

Register your instrument! Eppendorf Research plus. User Adjustment

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014

Ensuring pipette & weighing device performance: Verification & Calibration

Adjust the Volume Display

Proline Plus Mechanical Pipette

Transferpette -8/-12

A N F G J I C D D L A B C D E F

Improve pipetting results in pharmaceutical formulation by using MICROMAN E

VITLAB micropipette. Standard Operating Procedure (SOP)

PERFORMANCE TEST INSTRUCTIONS FOR BIOHIT PROLINE PIPETTORS

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

ab IgG1 Human ELISA Kit

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs

AUTOMATIC HOSE TEST UNIT, TYPE SPU

Pipetting Small Volumes

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts

Automated Liquid Handling Station

Good Pipetting Practice

RayBio Human vwf ELISA Kit

Fisherbrand. Finnpipette INSTRUCTIONS FOR USE PARK LANE Pittsburgh PA

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

Register your instrument! Eppendorf Xplorer / Eppendorf Xplorer plus. Adjustment

Constant Pressure Inlet (CCN) Operator Manual

Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features

Technical Procedure for General Laboratory Equipment

Series Ruska Digital Pressure Indicator. GE Sensing & Inspection Technologies. Features

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

Convergent Technologys product overview

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS

biosensis Mouse CXCL10/IP-10 ELISA Kit Protocol

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

PIPETMAN F ENGLISH. Other languages available on our website:

INTERNATIONAL STANDARD

Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx

Using a micropipette to transfer exact liquid measurements in scientific experiments By: Nicole Hume

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

Selecting the right pressure sensor for your application

RayBio Human TNF-alpha ELISA Kit

Low gas volume & flow measurements made easier Gas Endeavour

Improve Process Reliability

Engineering Note. Algorithms. Overview. Detailed Algorithm Description. NeoFox Calibration and Measurement. Products Affected: NeoFox

Transferpette. Single and Multichannel Pipettes

SOP-111 Pipetting Small Volumes

PIPET TIPS. UniFit ZFR (Zero Fluid Retention) Tips Universal Fitting Pipet Tips

BUYER S GUIDE AQUAlogger 530WTD

Standard Procedure. for using ConfiRM with the SoloVPE. Slope

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

biosensis Human IGF-II, Insulin-like growth factor II, Somatomedin-A ELISA Kit Protocol

Pneumatic high-pressure controller Model CPC7000

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

Intelligent Pipette Automation made Simple

Deadweight Tester / Pressure Balance High pressure up to 5,000 bar / 70,000 psi CPB5000-HP. Applications. Special features. Description. Rel.

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

How to Measure R7.1. Reference. I. Linear dimensions

Ruskinn Hypoxia and Anoxia Workstations. innovation in incubation

RayBio Human IFN alpha/beta R2 ELISA Kit

Human ABCD1 ELISA KIT

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

DS/PI-14 DIAPHRAGM SEAL TEMPERATURE ERROR

Additel 761 Automated Pressure Calibrators

Prepared By: Lisa McAvoy 30/Nov/17 dd/mmm/yy. Approved By: Veronica Harris-McAllister 30/Nov/17 dd/mmm/yy

Pneumatic high-pressure controller Model CPC7000

Transcription:

Performance Verification of BioTek s Precision 2000 using Artel s Multichannel Volume Verification (MVV ) System Authors: Paul Held 1, John Bradshaw 2, Jason Greene 1, Alex Rogers 2, and Tanya Knaide 2 1 BioTek Instruments, Winooski, Vermont; 2 Artel Inc., Westbrook, Maine Abstract Biomolecular screening often entails repetitive pipetting of fluids into and out of 96- and 384-well microplates. The standardized matrix and spacing of the wells in microplates, 8 x 12 and 16 x 24, for 96- and 384-well microplates, respectively, has allowed the use automated systems to pick up and dispense liquids when these microplates are employed. The use of automation has allowed tremendous numbers of samples to be screened using microplates as the reaction vessel. Towards that end, BioTek Instruments (Winooski, VT) has developed the Precision 2000 Automated Pipetting System to eliminate the need for manual pipetting into these plate formats. Due to the absence of human intervention, the increased reliance on automation to perform repetitive pipetting tasks on large numbers of samples makes the need for pipetting accuracy even more important than it a;ready was. Artel has developed a system that photometrically measures liquid handler performance from 2-200 µl. This Multichannel Volume Verification (MVV ) System uses dimensionally characterized microplates and accuracy controlled dual-dye sample solutions to measure both the accuracy and precision of liquid dispensed into 96-well plates from multichannel liquid handler heads. Because of the small volumes associated with the wells of plates, both 96- and 384-well, the ability to dispense small volumes accurately and precisely is necessary to perform many assays. While the performance of the Precision 2000 Automated Pipetting System has been specified at relatively large volumes such as 100 µl, its typical performance at much lower volumes is quite remarkable. Here we describe the process of using the MVV system to validate the Precision 2000 in a matter of minutes, as well as demonstrate the performance of the Precision 2000 Automated Pipetting System. Introduction The use of automation has allowed tremendous numbers of samples to be screened using microplates as the reaction vessel. Towards that end, BioTek Instruments has developed the Precision 2000 Automated Pipetting System. The Precision 2000 has a completely configurable multi-station platform to hold the required pipette tips, reagent troughs, and microplates (96- and 384-well) for fluid transfer (Figure 1). The platform is removable, allowing for multi-user friendliness, easy cleaning, and setup of the instrument. The 8- or 12-channel pipette arm moves up and down as well as side to side, while the platform moves front to back to provide complete access to all locations on the work platform and complete configurability. The pipette arm uses a proprietary technology to reliably pick up and seal any standard tip with BioTek Instruments, Inc., P.O. Box 998, Highland Park, Winooski, Vermont 05404-0998 USA COPYRIGHT 2006 TEL: 888-451-5171 FAX: 802-655-7941 Outside the USA: 802-655-4740 E-mail: customercare@biotek.com www.biotek.com

individual, free-floating barrels that compensate for tips out of position (Figure 2). An optional bulk-dispense 8- or 12-channel manifold, which uses a precise bidirectional syringe pump to accurately and rapidly dispense fluids from a large unpressurized reservoir, is also available. Precision Power PC software accompanies the Precision 2000 Automated Pipettor to provide complete programming flexibility. In addition to the external PC software, the Precision 2000 has a built-in microprocessor that can control all movements, which allows for program downloading. Once downloaded, the PC can be disconnected and the pipetting programs can be run via a keypad. The flexible onboard software, which provides complete programming for the most complex fluid transfers, can store up to 80 programmed assays. For more complete automation robotics, interfaces can be developed using ActiveX software commands. The Precision 2000 s small size, with a 15 x 21-inch footprint and a height of 16 inches, allows it to be used almost anywhere including most biological safety cabinets or chemical fume hoods. Also available is an optional aerosol cover for use on the laboratory bench. Building upon their expertise in liquid delivery verification, Artel has developed a system (MVV ) that photometrically measures liquid handler performance from 2-200 µl. The MVV system uses dimensionally characterized microplates and accurately controlled concentrations of dual-dye sample solutions to measure both the accuracy and precision of liquid dispensed into 96-well plates from multichannel liquid handler heads. The MVV system uses two distinctly different dye solutions, which are very stable over time and traceable to national standards. The blue dye, with an absorbance peak at 730 nm, is used at the same concentration for all volumes to be tested. The red dye, with an absorbance peak of 520 nm, is used at different concentrations depending on the dispense volume to be tested. The microplates have truncated conical wells that are uniform and have been well characterized with regard to their geometry. Using the device to be tested, the test sample is pipetted into the microplate. Note that the sample contains both blue and red dyes. However, with very small test-volumes, additional diluent will be required in order to have a sufficient volume to adequately cover the bottom of the well. Because the diluent contains the same concentration of blue dye as is present in the test sample, the precise amount of diluent is not critical. Once the test sample and any diluent necessary have been pipetted, the microplate s absorbance at 520 nm and 730 nm is measured. The first calculation performed determines the liquid depth or pathlength of each well. This calculation is based on the ratio of the absorbance of each well at 730 nm to an absorbance per unit pathlength constant that has been determined by Artel (Figure 3A). Once the depth of the liquid has been calculated, the total volume of liquid can be calculated for each well using the volume formula for a truncated cone (Figure 3B). The test sample volume that has been pipetted into the well is then calculated using the calculated total volume, the ratio of absorbance at 520 nm to 730 nm, as well as the ratio of the absorbance per unit of pathlength constants of both dyes (Figure 3C). In order to calibrate microplate reader response (i.e., the absorbance values) to the test dyes of the MVV system, a calibrator plate that contains the same solutions as the sample and diluent solutions in sealed precision cuvettes is placed into the microplate reader and the absorbance at 520 nm and 730 nm determined. Each calibrator plate is bar-coded to specifically identify the factory-determined absorbance properties of the calibrator plate to the MVV software. This establishes a direct correlation between the factory reference spectrophotometer and the microplate reader. With the reader calibrated and the test plate pipetted, the absorbance of the test plate is measured and the performance of the instrument (e.g., Precision 2000) determined in a matter of minutes.

Materials and Methods Using the Multichannel Volume Verification (MVV ) system from Artel, the dispense accuracy and precision of the Precision 2000 Automated Pipetting System was tested. For each indicated volume, microplate wells were filled with the indicated volume of dye using a Precision 2000 Automated pipettor. The absorbance at 520 and 730 nm was determined using an ELx800NB Microplate Absorbance Reader (BioTek Instruments, Winooski, Vermont) according to the MVV instructions. Dispense volumes and precisions were calculated using MVV software (Artel Westbrook, Maine). The calculated accuracy and precision data were then exported to Microsoft Excel for the purposes of graphing. For single dye experiments, a concentrated blue dye solution was dispensed into clean dry plates and colorless TRIS buffer diluent was added using the Precision 2000 (BioTek Instruments, Inc.). The resulting absorbance was then read at 595 nm using a microplate absorbance reader. Volume data were then exported to Microsoft Excel for analysis. Results As demonstrated in Figure 4, the Precision 2000 is very accurate over the range of volumes tested using the MVV system from Artel. The specification for accuracy with the Precision 2000 is an error of less than 2.0% at 100 µl. In these experiments the percent error at all volumes tested was well below specification, and at volumes greater than 50 µl the error was less than 0.5%. In addition to being accurate, the Precision 2000 was found to be precise. Despite having 8 or 12 independent pipette channels that work in unison, there was very little channel-tochannel variability. Figure 5 demonstrates the dispense-precision of the Precision 2000. As an entire plate, the coefficient of variance (%CV) was found to be less than 2.0% at volumes down to 5 µl. When a direct comparison between calculated volume and expected volume is plotted, the resultant linear regression shows a very high degree of correlation, with a correlation coefficient (r 2 ) of 0.9999 (Figure 6). In addition, these data correlate well with precision data generated using a one-dye test (Figure 7). Conclusions Reliance on automation to perform repetitive pipetting tasks on large numbers of samples makes pipetting accuracy and precision even more important than they already were. These data indicate that the Precision 2000 can be used to dispense aqueous solutions precisely and accurately. Using either a one-dye test procedure, which tests precision, or the two-dye MVV system, which tests both accuracy and precision, the Precision 2000 performed admirably. Several factors influence the ability to dispense small volumes with the Precision 2000. The positive displacement syringes depend on the displacement of air to push the fluid out of the barrel of the tip. Unfortunately, gaseous materials are prone to compression when the gas volume is large relative to the fluid volume to be dispensed (e.g., 100 µl). This often precludes the ability to be accurate at distinctly smaller volumes. Because air is also prone to expansion and contraction as a result of temperature changes, having samples and reagents equilibrated to ambient temperature will result in more accurate and precise dispensing as well. For more accurate and precise dispenses, it is advisable to include an initial pre-pickup of a small volume. This will provide some compensation for air gas compression, as well as provide a humid environment inside the pipette tip. Additional steps can be taken to ensure accurate dispensing as well. If multiple dispenses of small volumes are needed, it is faster and more accurate to aspirate a large volume and perform multiple dispenses from the same pipette tip. However, in many instances the reuse of tips is precluded by cross-contamination concerns. In the case of dispensing into dry microplates, it is advisable to perform a bottom touch-off, which will assist in transferring the remaining droplet caused by the surface tension of the liquid from the pipette tip to the well. Contacting the liquid to the well bottom will caused the droplet to adhere to the well rather than remain with the tip.

Summary: The Precision 2000 is an accurate and precise automated multichannel pipettor. The MVV System is a fast and reliable method to test the calibration of multichannel pipetting devices. The MVV System can test both accuracy and precision at the same time in a matter of minutes. Figure 1. Precision 2000 Automated Pipettor with Rapid-Dispense Manifold and Aerosol Cover. Figure 2. Closeup of Pipette head of an 8-channel Precision 2000 Automated Pipettor.

Calculate depth of liquid in each well Based on the absorbance at 730 nm Independent of the ratio of sample to diluent a b = absorbance per unit pathlength of blue dye in both solutions l l = A 730 a b Figure 3A. Liquid Depth Calculation. Volume calc ulation is based on a truncated cone Total volume c alculated from liquid depth and known dimensions l γ D D 2 V l tan( ) tan( ) T 4 Dl 2 γ 2 l 3 γ = π + π + π 3 Figure 3B. Liquid Volume Calculation.

Calc ulate sample volume based on total volume and measured absorbance ratios a r = absorbance per unit pathlength of red dye in sample solution V S a b a r A 520 ( )( ) V S = V T A 730 Figure 3C. Sample Dispense Volume Calculation. Precision 2000 Accuracy Using MVV Method 2.00% 1.50% % Error 1.00% 0.50% 0.00% -0.50% 5 10 15 38 50 100 Dispense Volume (µl) Figure 4. Accuracy of the Precision 2000 Automated Pipettor. Using the MVV System, the volume of liquid dispensed by the Precision 2000 was determined for several different volumes. The percent error of the calculated mean at each volume was then plotted.

Precision 2000 Precision Using the MVV Method 2.00% 1.60% % CV 1.20% 0.80% 0.40% 0.00% 5 10 15 38 50 100 Dispense Volume (µl) Figure 5. Dispense precision of the Precision 2000 Automated Pipettor. Using the data depicted in Figure 4, the coefficient of variance (%CV) of the data at each volume was calculated and plotted. Determined vs. Expected Dispense Volumes Determined Volumes (µl) 120 100 80 60 40 20 0 R 2 = 0.9999 0 20 40 60 80 100 120 Expected Volumes (µl) Figure 6. Comparison between expected and determined dispense-volumes. Using the MVV System (Artel, Westbrook, ME) the volume of liquid dispensed by the Precision 2000 (BioTek Instruments, Winooski, VT) was determined for several different programmed volumes. The calculated volume was then plotted against the expected volume and a linear regression analysis performed. The resultant correlation coefficient of the regression is indicated.

70.0 Dispense Precision Precision 2000 Pipette %CV 60.0 50.0 40.0 30.0 20.0 10.0 0.0 1 2 3 4 5 6 7 8 9 10 15 20 25 Dispense Volume per Well (µl) Figure 7. Dispense precision into dry 96-well plates using the Precision 2000 pipette at various dispense volumes. The indicated volume of TRIS buffer containing blue food coloring was dispensed to all the wells of a 96-well plate using the pipette. After dispensing the colored dye, 100 µl of TRIS buffer was dispensed into each well using the dispense manifold at a speed setting of 5. The absorbance at 595 nm for each well of every plate was determined using a microplate absorbance reader. Note that each data bar represents the %CV of an entire plate. Rev. 10/29/03