Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

Similar documents
Autocalibration Systems For In Situ Oxygen Analyzers

Autocalibration Systems for In Situ Oxygen Analyzers

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

Technical Information

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

TECHNICAL INFORMATION

Oxygen Flue Gas Analyzer BA 2000

Installation, Operation, and Maintenance Manual Parker Balston Model N2-04 Nitrogen Generator

Model Series 62 Constant Differential Relay

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

Installation, Operation, and Maintenance Manual

Model EXAxt AV550G. Averaging Oxygen Analyzer

Bermad Pressure Reducing. Model: 42T

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Instruction Manual O-Ring Ejector Chlorine & Sulfur Dioxide to 500 PPD (10 kg/h)

User and Installation Manual

Best Practice Guide, Servomex 2700

Zirconia Oxygen Sensors. Model 1230 Series

OXYGEN MONITORING SYSTEM OMS

TECHNICAL INFORMATION

Generating Calibration Gas Standards

CSA Sample Draw Aspirator Adapter Operator s Manual

NGP-250/500 Nitrogen Generator Quick Start Guide

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

ACCUMULATOR OPERATING & MAINTENANCE INSTRUCTIONS

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

with O 2 Controller Instruction Manual

RK-IR Sample Draw Aspirator Adapter Operator s Manual

MEDICAL EQUIPMENT CATALOG

Installation, Operation, and Maintenance Manual

Inert Air (N2) Systems Manual

Oxygen Dialflow Meter. Instructions for Use

Operating Instructions

Installing a Probe. Typical Installation

Instruction Manual - Diaframless TM Ejector Chlorine, Sulfur Dioxide and Ammonia

ACV-10 Automatic Control Valve

RAM Operation Manual

Pressure Independent Control Series

5 Channel Calibrator

Dissolved Oxygen Sensor

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

GILMONT ACCUCAL FLOWMETERS

Anderson Greenwood Series 93 Positive Pressure POSRV Installation and Maintenance Instructions

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION SPECIFICATIONS. CE (EMC) Compliant

GAS SUPPLY APPLICATION GUIDE

SERIES 30. Spring Operated Tank Blanketing Valve PROTECTOSEAL.

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

DS05C,D,G Dial Set Pressure Regulating Valves

RD(H)20/25 Pressure-Reducing Regulator User Manual

Installation, Operation, and Maintenance Manual

INSTRUCTIONS FOR MODELS UPH & UPHS HIGH FLOW, HIGH PURITY GAS REGULATORS

SERVICE MANUAL No. I-0043

RDK-408D2 Cold Head. Technical Manual. SHI-APD Cryogenics Inc Vultee Street Allentown, PA U.S.A. Revision A: September 2005

SERIES 10. Pilot Operated Tank Blanketing Valve, 2" PROTECTOSEAL.

Manual Solvent Purification System. Innovative Technology, Inc.

Nitrogen Generation Systems

RHPS Series RD(H)F40 User Manual. Read the complete manual before installing and using the regulator.

Endura AZ25 series probe Combustion oxygen monitor

Description Advanced Features Typical Applications

1805 Series Relief Valves

ABB MEASUREMENT & ANALYTICS DATA SHEET. Endura AZ40 Oxygen and carbon monoxide equivalent (COe) analyzer

CAPITAL CONTROLS SERIES NXT3000

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &

64 Series Pressure Reducing Regulators

ECS Protector Nitrogen Inerting Vent (PAV-WN)

PERFORM Operating Document

6890 GC Site Preparation

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

IMPORTANT SAFETY INSTRUCTIONS

XMO2. Panametrics Smart Oxygen Analyzer. Applications. Features. bhge.com. An oxygen transmitter for use in:

Micro Dial-Flowmeter. Instructions for Use

Model 4000 Pressure Controller

MEGR-1912 Instruction Manual

Type S301 & S302 Gas Regulators INTRODUCTION INSTALLATION. Scope of Manual. Description. Specifications. Type S301 and S302. Instruction Manual

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI

MODEL GENERA PURGE TURBINE INSTR DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.

2 Sentry MCL Installation, Operation & Maintenance

250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO

250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO

SPECIFICATIONS. Approximate Weight: 10 oz. Surface Finish: Ra micro inch or less MATERIALS OF CONSTRUCTION

8.1 Calibration of gas sensors

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

Type EZH Relief or Backpressure Regulator

User Manual for the Mars Calibration Bench

LRS(H)4 Pressure-Reducing Regulator User Manual

Purge Star & Ring Purge Systems

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

PART 1 GENERAL..2 Components of the gas chlorination system include, but are not limited to the following:

RAM 4021-DPX Operation Manual

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

Types ACE95 and ACE95Sr Tank Blanketing Valves

SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER

ATS430 turbidity sensor Retractable insertion assembly

GAS FUEL VALVE FORM AGV5 OM 8-03

Maintenance Manual. Liquid Level Transmitter CT801-SI, CT801-TA

HP/UHP Spring-Loaded Regulator

Transcription:

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes Instruction Bulletin Addendum IB PB1000 Part no. Serial no. Order no. Oxymitter 4000

Applicability: Used with pressure-balancing option, p/n 3D39811 (50 psig maximum process pressure). This product option may be utilized with Rosemount Analytical s Oxymitter 4000/5000 O 2 analyzer or with Rosemount Analytical s CENELEC World Class O 2 Analyzer. This may also be used with 3081/4081FG high temperature systems (remote mounted only) and may be used in conjunction with optional Isolation Valving System Oxy-Iso 1000 General Rosemount Analytical s line of in situ oxygen analyzers utilize zirconium oxide sensing technology, which is sensitive to pressure variations in the process. The ZrO 2 sensing cell develops and EMF O 2 relative to content based upon the Nernst Equation: where: EMF = KT log10 (P1/P2) + C P2 is the partial pressure of oxygen on the flue gas side P1 is the partial pressure of oxygen on the reference side T is the absolute temperature K is a constant C is the cell constant. Rosemount Analytical s Oxymitter and World Class probes place the sensing cell at the end of an in situ probe that inserts directly into the process gas stream. This arrangement provides for fast response and is also very resistant to pluggage from particulate material, or the acids that frequently condense out of normal flue gas. The ZrO 2 sensing technology is sensitive to pressure changes of the process, and one can expect an output change of approximately 1% of reading (not 1 % FS, or 1% O 2 ) for every 4 inches of H 2 O pressure or vacuum in the process. The pressure errors of a few psi can be calibrated by calibrating the probe in the installed, pressurized condition. However, pressure fluctuations from the calibrated state will induce errors. A better accommodation for pressurized processes is to balance the process pressure with reference air (common instrument air) inside of the oxygen probe (see following page). The sensing cell will have the same pressure on both sides, eliminating any pressureinduced error. Pressure fluctuations in the process will be balanced, maintaining the most accurate O 2 reading throughout normal operating conditions. Rosemount Analytical s Oxymitter 4000/5000 probes (both hazardous area and general purpose styles) and the CENELEC World Class probes are suitable for pressure balancing. Process pressure is routed to a balancing relay via a nitrogen purge system or via the probe s internal calibration gas line. The balancing relay duplicates the process pressure with instrument air and this is routed to the reference side of the probe (inside of the probe). This pressure balancing system is integrally mounted to the probe and gauges are provided to confirm proper balancing operation.

Preferred Arrangement Utilizing N 2 Purge into the Process and Representing Process Pressure Ambient Proces Probe Tube Sensing Cell Check Valve (if Autocal system is Diaphrag N2 Purge into Diffusion Element, and Optional Flame Arrestor Balancing Relay Pressure Regulator, set to above max. process Instrument Air 1 Installation notes for N2 Purge Arrangement Install T for N2 tap as close to process as possible. Calibration Gas Fitting- capped tight between calibrations If autocal system is utilized, or if cal gas lines are permanently installed, use check valve kit, p/n 7307A56G02 Reference Gas Inlet- Piped to pressure/balancing system Reference Gas Exhaust- Capped, ½ turn open from tight Operation Note- Set N2 pressure regulator to 10 PSI above maximum process pressure

Integral Pressure Balancing Arrangement- Manual Calibration Ambient Process Probe Sensing Calibration Gas Line/Process Gages Diaphragm Selector Valve CG Cal Gas (manually Diffusion Element, and Optional Flame Balancing Relay Pressure set to above process Instrument Air 1 If N2 purge is unavailable, the probe s calibration gas line may be utilized as the process pressure tap for the Balancing Relay. Ensure that all parts wetted by process gases are well insulated to prevent condensation.

Integral Pressure Balancing Arrangement- Auto Ambient Proces Probe Tube Sensing Cell Calibration Gas Line/Process Tap Gage Diaphrag Check Valve CG Flowmeter Cal Gas (from Autocal system) Diffusion Element, and Optional Flame Arrestor Reference Air at Process Pressure Balancing Relay Pressure Regulator, set to above max. process Instrument Air 1

An optional isolation valving system facilitates the insertion and withdrawal of the probe for service while the process is under pressure. This valving system is recommended if the process cannot be shut down for probe maintenance. Unpacking Probe will be shipped with the pressure-balancing valving option mounted and tested. Installation of Pressure Balancing System Reference Air Follow standard instructions in the instruction bulletin for the mechanical installation of the probe system. Instrument air is required and must be piped into the pressure regulator provided on the pressure balancing assembly. Set this pressure regulator to 50 psi. Reference air exhaust port should be capped loosely, ½ turn open from tight. Calibration Gases If the probe system will be manually calibrated, ensure that the blocking valve is shut except during calibration. When calibrating, connect bottles to the bottom part of the calibration gas flowmeter and set bottle pressure regulators to 50 psi. If an automatic calibration system will be used, mount the SPS, IMPS or MPS per standard instruction bulletin and pipe calibration gas outlet line to the check valve provided on the probe (per page 3 of the appendix). If an automatic calibration system will be used, again set the pressure regulator on the calibration gas bottles to 50 psi. Any time the calibration gases will be permanently connected, ensure that a check valve is provided to prevent the migration of process flue gases back down the calibration gas line. To prevent corrosion from condensation of process gases insulate all process wetted parts; including the probe flange, the calibration gas/process line and balancing relay. If this is an outdoor installation where rain and snow will be present, cover the entire probe electronics (Oxymitter) or probe junction box (CENELEC World Class) to inhibit the cooling effect of precipitation. Electrical Follow the probe electrical installation instructions in the Instruction Manual. Note that electrical penetrations into the CENELEC World Class probe will be under pressure, and must be sealed. The normal procedures for installing probes into hazardous areas are usually sufficient to maintain pressure inside a probe. The Oxymitter utilizes an internal bulkhead which isolates the terminations side of the probe from the pressure balancing system. The termination side cover can be removed and the wires disconnected while the system is under pressure. Do not remove the electronics side cover while the process is under pressure, and the pressure balancing system is active.

Note: The pressure balancing system must be connected and operational at all times. If an optional isolation valving system is utilized for inserting or withdrawing the probe while the process is under pressure, the pressure balancing system must be active at all times until the gate valve is closed. This may require an installation whereby the reference instrument air lines and electrical cables can ride back and forth with the probe. See Instruction Bulletin Addendum for Isolation Valving System, IB ISO1000 for more information. Calibration gas lines may be removed for probe installation or removal. Start-up and Operation Pressure Balancing system works automatically, and requires no adjustment. Two pressure gages are provided with the system, and should always read very close to the same pressure. Reference gas side gage may read 1-2 PSI lower than the process side gage. Consult the troubleshooting table whenever these gages do not read the same. For most accurate operation, calibrate the probe after the process is up, and at operating temperature and pressure. Caution: The inside of the probe is pressurized. If using the CENELEC World Class product do not unscrew the junction box cover or the electrical penetrations until reference gas pressure is bled off. If using the Oxymitter, do not unscrew the electronics side cover while pressurized. Oxymitter terminations side cover may be opened while pressurized. Operational Note: Some catalyst regeneration applications (notably the UOP CCR process) use nitrogen in the process. During shutdown, the O 2 is completely displaced with N 2 for some period of time. Residual CO and other combustible components can build up on the diffuser and cell of the probe. When coming back on line, the probe will read zero percent O 2 until these components are burned completely off the heated probe. We recommend that nitrogen be bled down the probe s calibration gas line at 5 SCFH during any periods where the probe s O 2 reading approaches zero. This flow may be removed and the cal gas line capped tightly when the unit comes back on line. This will insure that the O 2 reading will respond normally as air is reintroduced into the process.

Pressure Balancing System Troubleshooting The pressure balancing system will operate without periodic adjustment. Whenever the probe is taken out of service, inspect the pressure balancing system for condensation or particulate pluggage in the lines. Most surfaces that are in contact with the process gases are stainless steel, but condensate can still corrode stainless steel. If condensation is found, insulate the process lines to keep them warm. Leaks are the most likely problems in the pressure-balancing system. Symptom Diagnosis Cause Correction Watch the two pressure Diaphram on Replace balancing gages on the pressure balancing relay or relay or process tap. balancing system. If needles process tap has a are not stable and do not hole in it. follow each other, there is a hole in the balancing relay diaphram or in the process tap/cal gas line feeding the balancing relay. Disassemble and bench test the balancing relay to isolate the leak. O 2 ocillates up and down repeatedly and at a frequency greater than the normal process O 2 changes, (up and down every few seconds). O 2 reading increases above normal readings. Disassemble probe and look for leaks in the calibration gas line. Process tap/calibration gas line has a hole in it or the elastomer tube has popped off. Process gas is mixing with reference air inside the probe. For disassembly of pressure balancing system, see page 9.

Spare Parts Lists Balancing Relay Block Valve Check Valve Calibration Gas Flowmeter Cal Gas/Process Dial Gage Reference Air Dial Gage Reference Air Pressure Regulator Mounting Bar Bracket 1A99408H01 3526B30H01 7307A56G02 771B635H01 1A99410H03 1A99410H01 1A99094H01 4851B62H01 4851B61H01

Pressure Balancing System Detail Optional Isolation Valve (probe uninstalled) Oxygen probe Reference Air Press. Gage Process Gage Block Valve Cal Gas Flowmeter Mounting Bar Booster Relay Bracket Reference Air Pressure Regulator Fig, 8 Disassembly of Pressure Balancing System 1) Disconnect instrument air supply line at the pressure regulator. 2) Disconnect the instrument air reference line out of the balancing relay and into the probe. 3) Disconnect the calibration gas line into the flowmeter. 4) Disconnect the calibration gas line from probe into balancing relay. 5) Remove the entire pressure balancing assembly from the mounting bar that is welded to the probe flange. 6) Unscrew calibration gas assembly feeding into the balancing relay, including the CG flowmeter, valve (or check valve if an autocal system is utilized), and tee fitting. 7) Remove balancing relay from mounting bracket. 8) Unscrew reference gas assembly from balancing relay, including the instrument air reference gas pressure regulator, and gage. Assembly procedure is reverse of disassembly procedure.