Gas exchange measurement in module M- COVX

Similar documents
Understanding the Issues Affecting the Accuracy of Measuring Oxygen Consumption Utilizing Metabolic Carts

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR)

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR)

HESTA Issues affecting the accuracy of Metabolic Carts. Danny Rutar

Review of gas frac9ons in atmospheric air 2/2/16. Metabolic Calculations of Indirect Calorimetry. Expired gas analysis.

iworx Sample Lab Experiment HE-4: Respiratory Exchange Ratio (RER)

GE Healthcare. CARESCAPE R860 Nutritional Assessment Tools Appliguide

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

UNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

Medical Instruments in the Developing World

Automatic Transport Ventilator

Application Note 268 Metabolic Analysis Systems and Propane-based Verification

CARDIOVIT AT-104 ergospirometry

New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters.

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805)

Content Display. - Introduction. Laboratory Unit : Lab 2 - Measurement of Oxygen Consumption. KINE xxxx Exercise Physiology

V8600 Ventilator. Integrated Invasive & Noninvasive Ventilation

Panther 5 Acute Care Ventilator

Using Common Ventilator Graphics to Provide Optimal Ventilation

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use

ALVEOLAR - BLOOD GAS EXCHANGE 1

Stikstofuitwas en Heliumdilutie. Eric Derom Universitair Ziekenhuis Gent

Technical Data and Specifications

Oxylog 3000 plus Emergency & Transport Ventilation

Mechanical Ventilation

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

Blease900 Ventilator PATIENT CENTERED VENTILATION

Unit 15 Manual Resuscitators

Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities

VENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE

RESPIRATORY PHYSIOLOGY, PHYSICS AND

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine

PART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV )

Innovating with you Shaping exceptional care

O 6 = 6 CO H 2 O H 2 = 16 CO 2. consumed during the oxidation of fatty acids, a ratio of 0.7.

EVALUATION OF A CANOPY SYSTEM AND A SIMPLE CALORIMETER FOR RESTING METABOLISM USING A RESPIRATORY SIMULATOR

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

Update to RS-232 commands. Changing patient from NIV to INVASIVE Vent Type. SNDF command

Recording established using the VO2 and RER graph template file (h19.gtl)

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

Quality assurance of Modular Metabolic Systems. Danny Rutar, Redback Biotek, Ireland/Australia Jeffrey Swen(Intern), University of Ulster, Belfast.

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Cornell Institute for. Biology Teachers. Respirometry Part I: Lung Volumes and Capacities. Lab issue/rev. date: 12/12/96. Title:

INTENSIVE CARE VENTILATORS

Let s talk about Capnography

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers

Pressure Controlled Modes of Mechanical Ventilation

SLE5000 Infant Ventilator with HFO

AUTOVENT 4000 VENTILATOR

Respiration Lab Instructions

Indications for Mechanical Ventilation. Mechanical Ventilation. Indications for Mechanical Ventilation. Modes. Modes: Volume cycled

Guide to Understand Mechanical Ventilation Waveforms

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

Virginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

The centerpiece of a comprehensive anaesthesia workstation

WILAmed INTENSA. Intensive Care Ventilator System

BREATH-BY-BREATH METHOD

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Carestation 650. Physical Specifications Product Description Carestation 620 A1 ANESTESI CARESTATION 650

Inspire rpap REVOLUTION FROM THE FIRST BREATH

Selecting and Connecting Breathing Systems

Collin County Community College. Lung Physiology

Hypoxia Following Rapid Decompression to 18,288 m (60,000 ft) Attributable to Alveolar Hypoventilation

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

VENTILATORS PURPOSE OBJECTIVES

IMPACT. Uni-Vent. Eagle. Model 754 Portable Ventilator

Introduction to Conventional Ventilation

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

Physiology of the Respiratory System

Ventilator Training Module

IMPC phenotyping SOPs in JMC

Oxylog VE300 Emergency & Transport Ventilation

A few seconds after turning on, the screen changes to Set Up Screen Before adjusting the pressure, calibrate the O 2 sensors

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

The Time Constant of The Respiratory System

Inspiration 7i Ventilator

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams)

Selecting the Ventilator and the Mode. Chapter 6

LUNG CLEARANCE INDEX. COR-MAN IN Issue A, Rev INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark

Initiation and Management of Airway Pressure Release Ventilation (APRV)

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Your patients, our primary focus. Siesta i Whispa anesthesia machine

PROCEDURE (TASK): ROUTINE VENTILATOR CHECK. 5. Verifies current ventilator Insures correspondence between physician's

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation

PROBLEM SET 8. SOLUTIONS April 15, 2004

Draeger Fabius GS Pre-Use Check

Overview. Front Panel: Keypad and Display

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

Monnal T60. Touch and Breathe.

Automatic Transport Ventilators. ICU Quality Ventilation on the Street.

Transcription:

Gas exchange measurement in module M- COVX This document explains how the gas exchange measurement in the gas module M-COVX works. 1. Basic equations To obtain the oxygen consumption VO2 of a patient, one must measure the amount of oxygen that is inhaled and subtract from it the amount exhaled. These amounts can be obtained by multiplying each measured volume piece dv by the corresponding gas concentration f O2: CO2 production VCO2 is obtained similarly Using inspiratory and expiratory minute volumes MVi and MVe and volume-averaged inspiratory and expiratory concentrations fi and fe, these equations can be written VO2 = fio2 MVi - feo2 MVe VCO2 = feco2 MVe - fico2 MVi To make the results less sensitive to errors in volume measurements, the well-known Haldane transformation can be applied. This means taking advantage of the fact that the patient is not consuming nor producing nitrogen, so that the amount of nitrogen inhaled is equal to the amount exhaled: fin2 MVi = fen2 MVe. VO2 and VCO2 can then be written using either inspiratory minute volume: VO2 = (fio2 - f Hald feo2) MVi VCO2 = (f Hald feco2 - fico2 ) MVi or expiratory minute volume: VO2 = (fio2/f Hald - feo2) MVe VCO2 = (feco2 - fico2/f Hald) MVe where f Hald = fin2 / fen2 = (1 - fio2 - fico2 )/ (1 - feo2 - feco2 ).

M-COVX measures inspiratory volumes more accurately than expiratory volumes, and therefore the former equations with MVi are applied. At fio2 levels higher than 65%, however, VCO2 is approximated by VCO2 = (feco2 - fico2) MVe. Deltatrac measures only expiratory minute volume and uses the MVe-based formulation. 2. Synchronization of gas and volume data The measurements of gas concentration and airway flow/volume are not performed simultaneously. When a gas sample is passing through the Dlite, the flow at that moment is obtained almost instantly as the pressure difference travels via the spirometry tubings with the speed of sound. In contrast, it takes about 1.1 s for the gas sample to travel through the sampling system to the CO2 sensor, and about 0.6 s more to reach the O2 sensor. These delay times are continuously measured and compensated for in order to obtain the volume-averaged inspiratory and expiratory gas concentrations. Figures 1-2 describe the measurement schematically. A sampling connector (1) for a gas analyzer and a spirometer is located between the Y-piece (2) of the ventilatory circuit and the intubation tube (3). A gas sample is continuously sucked via the sampling tube (4) from the connector to the gas analyzer unit (5) of the measuring device. Pressures on both sides of a flow restriction (6) are communicated via the spirometer tubes (7) to the spirometer unit (8) of the device.

Upper part of Figure 2 is a schematical picture of the gas analyzing unit (5). CO2 concentration is measured by an infrared absorption sensor (9), and O2 concentration is measured by a paramagnetic resonance sensor (10). The rise times of these measurements are compensated for by deconvolution procedures. When the gas sample enters the device, moisture is removed in the water separator (11). The flow is divided into two parts, one of which goes through the gas sensors (9) and (10) and to the bypass flow (12). The reference gas (13) required for the paramagnetic measurement is taken from room air. When the zero valve (14) is closed, input flow is taken from room air for a readjustment of the zero levels of the gas concentration measurements. If an occlusion takes place in the sampling tubes, the occlusion valve (15) is closed and sample flow will completely bypass the analyzers. The power of the pump (16) is then effectively employed in an effort to clear the occlusion. Sample flow vs is measured by the pressure transducer (17) which senses the pressure difference?ps over a flow restrictor between the CO2 and O2 sensors. Lower part of Figure 2 shows the spirometer unit (8) of the device. Pressures communicated via the spirometry tubes are converted to an airway pressure signal paw (19) and an airway pressure difference signal?paw (20). Airway flow vaw is computed from?paw.

2.1. Corrected sample flow measurement 2.2. Compensation of pressure variations 2.3. Reducing sample flow variations 2.4. Compensation of variations in sample line volume Vs 2.4.1. Measurement of Vs 2.4.2 Adjustment of Vs 2.5. Integration of gas concentration samples 3. Deformation of gas concentration curves The gas curves as we see them on the monitor screen do not look the same as they would had they been measured at Dlite by an imaginary infinitely fast device. First, some mixing of gas which softens the front regions between inspiratory and expiratory gases has taken place during the transport of the gas via the sampling system. Second, the CO2 and O2 sensors are not infinitely fast but have a finite measurement rise time, largely due to the fact that they measure a finite volume of gas. A set of transformations is performed that

reduces the measured gas signal back to the concentration pattern at the sampling point of the Dlite at the time instant when the sample is taken. 3.1. Rise time 3.2. Mixing 4. Accuracy of VO2 and VCO2 in the M-COVX module According to the laboratory measurements and clinical tests, the method gives VO2 and VCO2 within +/- 10% accuracy up to fio2 < 65% and within +/- 15% accuracy for 65% < fio2 < 85%. Some further restrictions for applying the method have been found: respiration rate RR < 35/min inspiratory time T i > 1 s if pressure swings are very high (p peak-p peep > 25 cmh 2O) stable (mixed) fio2 level less accurate in BIPAP mode It is concluded that the method can be applied with sufficient accuracy in the majority of clinical situations; cases which meet the restrictions above are relatively rare. 5. Comparison to Deltatrac performance The overall accuracy of the M-COVX method is less good than that of Deltatrac, which gives VO2 and VCO2 within +/- 6% accuracy for fio2 < 65%. For Deltatrac high respiration rates or very short inspiratory times are not difficult, whereas Deltatrac also encounters problems for unstable fio2 levels and in the BIPAP mode. The main factor in favor of M-COVX is its supreme ease of use. However, in certain situations M-COVX also possesses technical advantages over Deltatrac: 1. M-COVX adjusts to changes in gas concentration levels much faster. This was demonstrated in the laboratory in the following way: Both VO2 and VCO2 were kept constant, while the fio2 level was switched from room air level to 40%. It took 13 minutes before Deltatrac again showed values with less than +/- 10% error; for M-COVX this took only 2 minutes which corresponds to the time that is needed for the respiratory system to reach a new dynamic equilibrium state. The reason for slow recovery of Deltatrac is that it takes time to clear the mixing chamber from old gas: some of it is left in the corners of the chamber affecting the measurement for a longer time than the absolute volume/flow ratio would suggest. 2. For low respiration rates Deltatrac shows higher standard deviations of VO2 and VCO2 values than M-COVX does. At respiration rate of 6/min, successive minute values may in a stable state differ by 20%. The reason is that Deltatrac has no means to find out when a breath starts or ends. Certain minute value may in the worst case consist of, say, seven inspirations but only six expirations, while the next minute then includes six inspirations and seven expirations. M-COVX shows balanced values of complete breaths, so low respiration rates are not a problem. 3. At low diluted CO2 levels Deltatrac is unable to measure accurately. If the metabolic rate of a patient is very low, CO2 concentration in the mixing chamber of Deltatrac may be too low for accurate measurements. The problem can be solved by switching Deltatrac to the child mode. This is, however, only possible if the minute volume is below the maximum limit specified for this mode. These situations do not pose any problem for M- COVX as no dilution is performed.