ACCURACY OF ESTIMATES OF STEP FREQUENCY FROM

Similar documents
Treadmill and daily life

Ambulatory monitoring of gait quality with wearable inertial sensors

APPLICATION OF THREE DIMENSIONAL ACCELEROMETRY TO HUMAN MOTION ANALYSIS

SCHEINWORKS Measuring and Analysis Systems by

Effects of fatigue on center-of-mass acceleration during a prolonged overground run

INTERACTION OF STEP LENGTH AND STEP RATE DURING SPRINT RUNNING

An investigation of kinematic and kinetic variables for the description of prosthetic gait using the ENOCH system

THE INFLUENCE OF SLOW RECOVERY INSOLE ON PLANTAR PRESSURE AND CONTACT AREA DURING WALKING

Gait Analyser. Description of Walking Performance

REPLACING REDUNDANT STABILOMETRY PARAMETERS WITH RATIO AND MAXIMUM DEVIATION PARAMETERS

Towards determining absolute velocity of freestyle swimming using 3-axis accelerometers

APPROACH RUN VELOCITIES OF FEMALE POLE VAULTERS

Impact Points and Their Effect on Trajectory in Soccer

Foot side detection from lower lumbar spine acceleration

Use of Accelerators for Biomechanical Analysis of Walking Motion Aided by Wheeled Walking Frames

HHS Public Access Author manuscript Int J Cardiol. Author manuscript; available in PMC 2016 April 15.

NIH Public Access Author Manuscript Gait Posture. Author manuscript; available in PMC 2009 May 1.

Sensitivity of toe clearance to leg joint angles during extensive practice of obstacle crossing: Effects of vision and task goal

Inertial compensation for belt acceleration in an instrumented treadmill

Analysis of Backward Falls Caused by Accelerated Floor Movements Using a Dummy

Accelerometers: An Underutilized Resource in Sports Monitoring

Mutual and asynchronous anticipation and action in sports as globally competitive

The Effect of a Seven Week Exercise Program on Golf Swing Performance and Musculoskeletal Screening Scores

Mobility Lab provides sensitive, valid and reliable outcome measures.

A Description of Variability of Pacing in Marathon Distance Running

Fall Prevention Midterm Report. Akram Alsamarae Lindsay Petku 03/09/2014 Dr. Mansoor Nasir

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

Neurorehabil Neural Repair Oct 23. [Epub ahead of print]

Keywords. Electronic supplementary material

Effect of the Grip Angle on Off-Spin Bowling Performance Parameters, Analysed with a Smart Cricket Ball

Motion Control of a Bipedal Walking Robot

Smita Rao PT PhD. Judith F. Baumhauer MD Josh Tome MS Deborah A. Nawoczenski PT PhD

Monitoring Walking and Cycling of Middle-Aged to Older Community Dwellers Using Wireless Wearable Accelerometers*

The Optimal Downhill Slope for Acute Overspeed Running

Relationship between Ground Reaction Force and Stability Level of the Lower Extremity in Runners Background: Objective: Design and Setting:

SPRINTING CHARACTERISTICS OF WOMEN S 100 METER FINALS AT THE IAAF WORLD CHAMPIONSHOPS DAEGU 2011

Validated assessment of gait sub-phase durations in older adults using an accelerometer-based ambulatory system

ETB Pegasus equipment was used in the following paper

Jogging and Walking Analysis Using Wearable Sensors *

Interface Pressure Mapping (IPM) Clinical Use of the Literature

A Development of Mechanical Arm for Weight Scale Calibration

PUV Wave Directional Spectra How PUV Wave Analysis Works

Artifacts Due to Filtering Mismatch in Drop Landing Moment Data

Technical Report. Evaluation of a Body-Worn Sensor System to Measure Physical Activity in Older People With Impaired Function

Knee Kinematic Improvement after Total Knee Replacement Using a Simplified Quantitative Gait Analysis Method

Gait Analysis System. User Manual and Outcome parameters Patent WO2012/ A1

Megan E. Krause, BSBSE, Young Hui Chang, Ph.D. Comparative Neuromechanics Laboratory. Georgia Institute of Technology

The Influence of Load Carrying Modes on Gait variables of Healthy Indian Women

Simulation-based design to reduce metabolic cost

BAUKJE DIJKSTRA, WIEBREN ZIJLSTRA, ERIK SCHERDER, YVO KAMSMA. Abstract. Introduction

GAIT MEASUREMENTS AND MOTOR RECOVERY AFTER STROKE. Plamen S. Mateev, Ina M. Tarkka, Ekaterina B. Titianova

Analysis of Foot Pressure Variation with Change in Stride Length

Friction properties of the face of a hand-held tennis racket

An investigation of lower-extremity functional asymmetry for nonpreferred able-bodied walking speeds

BODY FORM INFLUENCES ON THE DRAG EXPERIENCED BY JUNIOR SWIMMERS. Australia, Perth, Australia

2) Jensen, R. Comparison of ground-reaction forces while kicking a stationary and non-stationary soccer ball

Video Based Accurate Step Counting for Treadmills

A Study of Human Gaits across Different Speeds

The three steps for biomechanical assessment are the following: > Periodically verify the results and the efficacy of treatment

Geometric moments for gait description

Normative data of postural sway by using sway meter among young healthy adults

Scanning Laser Vibrometry Assessment of Sports Equipment

Image-Based Measurement of Alveoli Volume Expansion in an Animal Model of a Diseased Lung

Validation of Measurements from a ZephIR Lidar

Continuous sweep versus discrete step protocols for studying effects of wearable robot assistance magnitude

Analysis of stroke technique using acceleration sensor IC in freestyle swimming

G-EOL. Discover the simplicity of gait therapy intended for daily use

SIMULTANEOUS RECORDINGS OF VELOCITY AND VIDEO DURING SWIMMING

ROSE-HULMAN INSTITUTE OF TECHNOLOGY Department of Mechanical Engineering. Mini-project 3 Tennis ball launcher

AN ACTIVITY MONITORING SYSTEM FOR ELDERLY

Gait Analysis by High School Students

CARDIOVIT AT-104 ergospirometry

Step Detection Algorithm For Accurate Distance Estimation Using Dynamic Step Length

NEUROLOGICAL INSIGHTS FOR TEACHING GOLF TO TODAY S FITNESS CHALLENGED

Walking speemtmmkubjects and amputees: aspects of validity of gait analysis

Supplementary materials

What is the optimal design of a rocker shoe

LocoMorph Deliverable 4.4 part 2.4

Available online at ScienceDirect. Procedia Engineering 112 (2015 )

Gait Analysis An Introduction Michael W Whittle

QUALITY OF TREADMILL EMBEDDED FORCE PLATES FOR GAIT ANALYSIS

(12) Patent Application Publication (10) Pub. No.: US 2009/ A1

Bicycling Simulator Calibration: Proposed Framework

Analysis of Variance. Copyright 2014 Pearson Education, Inc.

Biomechanical analysis of the medalists in the 10,000 metres at the 2007 World Championships in Athletics

Calibration and Validation of the Shell Fatigue Model Using AC10 and AC14 Dense Graded Hot Mix Asphalt Fatigue Laboratory Data

New Monitoring Approaches for Improving Athletes Performance Vesa Linnamo

TEMPORAL STRUCTURE OF A LEFT HAND TOSS VS. A RIGHT-HAND TOSS OF THE VOLLEYBALL JUMP SERVE

A COMPARISON OF SELECTED BIOMECHANICAL PARAMETERS OF FRONT ROW SPIKE BETWEEN SHORT SET AND HIGH SET BALL

Developing an Effective Quantification Method of Tongue Deviation Angle to Assess Stroke Patients

siot-shoe: A Smart IoT-shoe for Gait Assistance (Miami University)

E STIMATING KILN SCHEDULES FOR TROPICAL AND TEMPERATE HARDWOODS USING SPECIFIC GRAVITY

MRI-2: Integrated Simulation and Safety

Inertial sensor, 3D and 2D assessment of stroke phases in freestyle swimming

ABSTRACT AUTHOR. Kinematic Analysis of the Women's 400m Hurdles. by Kenny Guex. he women's 400m hurdles is a relatively

KINEMATIC ANALYSIS OF SHOT PUT IN ELITE ATHLETES A CASE STUDY

Figure 2: Principle of GPVS and ILIDS.

PURPOSE. METHODS Design

Naval Postgraduate School, Operational Oceanography and Meteorology. Since inputs from UDAS are continuously used in projects at the Naval

Assessment of spatio-temporal gait parameters from trunk accelerations during human walking Zijlstra, Wiebren; Hof, At L.

Transcription:

ACCURACY OF ESTIMATES OF STEP FREQUENCY FROM A WEARABLE GAIT MONITOR M Punt, MSc 1,HWittink, PhD 1,Fvan der Bent, Ing 1,Jhvan Dieën, PhD 2,3 1 Research group Lifestyle and Health, Utrecht University of Applied Sciences, Utrecht, The Netherlands; 2 Move Research Institute Amsterdam, Faculty of Human Movement Sciences, VU University Amsterdam, Amsterdam, the Netherlands; 3 King Abdulaziz University, Jeddah, Saudi Arabia Corresponding Author: Michiel.punt@hu.nl Background: Assessment of gait activity by accelerometry requires data analysis. Currently several methods are used to estimate step frequency. At present the relation between step frequency estimation, gait speed and minimal required time window length remains unknown. Aims: The purpose of the study was to assess the accuracy of estimates of step frequency (SF) from trunk acceleration data analyzed with commonly used algorithms and time window lengths, at a wide range of gait speeds. Method: Twenty healthy young subjects performed an incremental treadmill protocol from 1 km/h up to 6 km/h, with steps of 1 km/h. Each speed condition was maintained for two minutes. A waist worn accelerometer recorded trunk accelerations, while video analysis provided the correct number of steps taken during each gait speed condition. Accuracy of two commonly used signal analysis methods (autocorrelation, fast Fourier transformation) was examined with time windows of two, four and eight seconds. Results: Our main finding was that accuracy of SF estimates with fast Fourier transformation and autocorrelation improved with increasing time window size, only at the lower gait speeds. Accuracy of SF estimation was lower at low gait speeds independent of the algorithm and time window used. Conclusion: We recommend a minimum TW length of 4 seconds when using AC and PSD algorithms and when using the PSD algorithm to use spectral averaging, as this leads to better results at short TW and low gait speeds. Journal MTM 4:1:27, 2015 doi:10.7309/jmtm.4.1.2 www.journalmtm.com Introduction Quantitative assessment of gait patterns usually involves laboratory methods, such as force plates and optical motion analysis systems 1. The ecological validity of such laboratory-based assessment can be questioned, i.e. does the movement performed in the gait lab represent the normal movement of a subject during daily life? 2. Tri-axial accelerometers fixed on the human body allow quantitative analysis of gait patterns and offer the advantage that they are not limited to a laboratory setting 3. Accelerometry-based assessment has become more and more accessible, with examples of new applications being: real-time gait parameter #JOURNAL OF MOBILE TECHNOLOGY IN MEDICINE VOL. 4 ISSUE 1 JANUARY 2015 2

recognition 4, continuous activity monitoring 5 and telerehabilitation 6. A basic parameter to be estimated in gait analysis is the number of steps taken per unit of time, i.e. the step frequency (SF). Algorithms based on Autocorrelation (AC) 7,8 and Power Spectral Density (PSD) 9 have been used to this end. These methods require analysis of data collected over a time window (TW) of a given size. The choice of TW size for data analysis depends on many factors. Accuracy of the estimation increases with TW size. In addition, smaller TWs reduce the frequency resolution and increase the chance of spectral leakage. However, step frequency (SF) within one person may vary over time and the length of the TW limits resolution in determining such variations. In addition, a longer TW produces a time delay due to the longer data collection and calculation times. This time delay might be unacceptable in real-time and continuous monitoring applications, especially because compact wearable systems are restricted in computational capacity 10. Finally, gait speed is often reduced in specific populations, such as patients post stroke 11 and with Parkinson s disease 12. Consequently, SF tends to be lower than in normal populations 11. Low SF may require a longer TW as the events occur less frequently. To the best of our knowledge, it is at present unknown which TW size provides the most accurate SF determination across the range of gait speeds that humans produce. The main objective of this study was to determine the effect of TW size on accuracy of SF determination over a range of gait speeds, for two most common used estimation methods (based on AC and PSD). For the PSD based estimates, we compared estimates of SF as the frequency at the peak PSD to a weighted average of the frequency at peak PSD and the nearest neighboring frequencies. We speculated that the latter would improve the accuracy of SF estimation, particularly for small time windows as spectral averaging might overcome effects of spectral leakage. In addition, it has been assumed that combining acceleration signals from different directions might improve the accuracy of gait parameter detection 10,4, at the cost of requiring more computational capacity. Therefore our second aim was to determine whether or not including acceleration signals from different directions improves the accuracy of measuring SF. Method and data analysis 2.1 Participants Twenty subjects (7 males, 13 females; age 28,6 9 11,2yr; height 172,6 9 8 cm; weight 69,4 9 9,7 kg; BMI 23,2 9 2,5; mean 9SD) voluntarily participated in the study. This study was approved by the local Ethics Committee and written informed consent was obtained from each participant. Treatment of the participants was according to the Helsinki declaration 13. 2.2 Protocol All subjects performed a twelve-minutes walk on a calibrated treadmill (En Mill treadmill, Enraf Nonius, the Netherlands). The first speed was set at one km/h and speed was increased by one km/h after every two minutes. Walking speed ranged from one km/h up to six km/h. The treadmill speed was manually adjusted by a research assistant. Only the last 88 seconds of each two-minutes walking bout were analyzed in order to avoid acceleration effects. The actual number of steps taken was derived from video observation. 2.3 Materials One tri-axial, piezo-capacitive accelerometer was worn around the waist (70X80X25mm, 150 grams, range 9 2.5g, output is in mv, a change of 1mV corresponded to a change of 0.08 m/s 2 (resolution)). Acceleration signals were digitally stored on a memory card with a sampling rate of 25 samples/s. A camera was placed behind the treadmill (Panasonic type HC-V70, 50 samples/s). 2.4 Data analysis Signal processing was performed using MATLAB (Matlab 7.10.0, The MathWorks, USA). Based on sensor alignment, acceleration signals were identified as anterior-posterior (AP), medio-lateral (ML) and vertical (VT). A low-pass second-order Butterworth filter with a cut-off frequency of 10 Hz was used. We compared two data analysis methods: AC and PSD (with and without spectral averaging with TW of two, four and eight seconds. Furthermore, accuracy of SF estimates derived from the AP acceleration signal as well as from combined AP and VT signals with of AC and PSD was determined. #JOURNAL OF MOBILE TECHNOLOGY IN MEDICINE VOL. 4 ISSUE 1 JANUARY 2015 3

2.5 Autocorrelation (AC) We used the unbiased AC sequence of the acceleration signals 8 where in x(i) (1,2..N) represents the time serie used, N represents the amount of samples used in the equation and m represents the timelagged phase shift in samples from the same time serie. 1 adðmþ ¼ N jmj X N jmj i¼1 xðiþxði þ mþ (1) We used an algorithm for automatic detection of the first dominant period (ad1) and second dominant period (ad2) 14. ad1 was subtracted from ad2, resulting in the number of samples between two contralateral steps (delta). Delta was divided by sample frequency and subsequently multiplied by the non-overlapping TW length. This procedure was implemented for all time windows in the 88 seconds walking bouts and results were summed, to estimate the total number of steps in the trial. 2.6 Power Spectral Density (PSD) Fast Fourier Transformation (FFT) was used to estimate the PSD of the acceleration signals. A custom-made algorithm searched for the peak in the PSD. We assumed that the highest peak in the power spectrum was the SF. For a more detailed description of the calculations, See 9. 2.7 Spectral averaging As an extension of the above method, we calculated a weighted average over the frequency at the highest power density and its nearest neighboring frequencies. Herein is Power(i) the amount of presents at a certain frequency (Df) in the time serie. estimated frequency ¼ P jþ1 i¼j 1 ðpowerðiþðidfþþ P jþ1 i¼j 1 PowerðiÞ (2) 2.8 Combining AP and VT signals Similar calculations for AC and PSD were made; however, input was the vertical (VT) acceleration signal. AC and PSD results from AP and VT signals were averaged. This resulted in AC and PSD derived estimates from both AP and VT signals. Further analyses were similar to the description above. 2.9 Statistics The difference in estimated number of steps from acceleration data and number of steps counted from video observation was expressed as mean absolute percentage error. A smaller mean absolute percentage error reflects a higher accuracy. Normality of the data was confirmed by the Kolmogorov-Smirnov test. We used a repeated measure ANOVA, to test for effects of gait speed, TW and algorithm type and their interactions. Tukey HSD tests were used for post-hoc analyses. To compare the different algorithms in relation with TW and gait speed we used a three-way factorial ANOVA repeated measures to test for any interaction effects between: gait speed * TW * and type of algorithm used. Results 3.1 Power Spectral Density and Autocorrelation Figure 1 illustrates the accuracy of the autocorrelation (AC) and Power spectral density (PSD) based Figure 1: Autocorrelation and Power Spectral Density results for estimating step frequency at a wide range of gait speeds (1 6km/h), at different TW lengths * 2 second window length, h 4 second window and o an 8 second window. Data is presented as the absolute percentage differences against camera observation. Please notice the difference in vertical scale between left and right panel. #JOURNAL OF MOBILE TECHNOLOGY IN MEDICINE VOL. 4 ISSUE 1 JANUARY 2015 4

estimates of SF, at all TW lengths and gait speeds. A three-way ANOVA repeated measures for type of algorithm (PSD and AC), TW length and gait speed demonstrated a main effect of algorithm (F 69.11, P B0.001), a main effect of TW length (F 97.11, P B0.001) a main effect on gait speed (F 8.26 P B0.001) and a significant interaction effect between these three factors (F 4.33, P B 0.001). Post-hoc analyses demonstrated a higher accuracy for PSD at 2 seconds TW at all gait speed conditions compared to AC at a 2 seconds TW. Comparing AC with PSD at the 4-seconds TW revealed a higher accuracy for AC for the first two gait speed conditions. No differences were found at the 8-seconds TW. 3.2 Autocorrelation Evaluation of the AC algorithm revealed a significant effect of TW size (F 106.3, P B0.0001), gait speed (F 6.02, P B0.0001) and an interaction effect of TW and gait speed (F 3.52, P B0.001). Post-hoc analyses revealed that SF estimates were less accurate for 2-seconds TW compared to 4- and 8-seconds TW for all gait speeds. Accuracy for 4- seconds TW was higher compared to 8-seconds TW at slow and fast speeds (1, 2 and 5, 6 km/h). Posthoc analyses within the TW size conditions revealed a significant increase in accuracy with increasing gait speed for TW of 2 and 8 seconds. 3.3 Power Spectral Density (PSD) PSD based estimates were compared between 2-, 4- and 8-seconds TW and all gait speed conditions. A significant effect of TW (F 16.3, P B0.0001) was found indicating higher accuracy with longer time windows. Furthermore a significant effect of gait speed (F 6.8, P B0.0001) indicating that accuracy improved with increasing speed, moreover no significant interaction effect between TW and gait speed (F 1.7, P 0.07). 3.4 Spectral averaging Figure 2 illustrates the effect of spectral averaging on the PSD based estimates. Significant main effects were found for gait speed (F 8.58, PB 0.001) for TW length (F 28.79, P B0.001) and type of algorithm used (PSD and PSD with spectral averaging) (F 18.21, P B0.001). A significant interaction effect was found between gait speed, TW length and algorithm (F 2.15, P 0.022). Posthoc analyses demonstrated that spectral averaging improved accuracy for 2-seconds TW across all gait speed conditions in comparison to PSD. In addition, spectral averaging for a TW of 4 seconds significantly improved accuracy in comparison to PSD at a TW length of 4 seconds at gait speeds of 1 and 2 km/h. No differences were found between PSD and spectral averaging at 8-seconds TW. However, accuracy improved for both methods at gait speeds of 3 km/h and faster. 3.5 Combining AP and VT accelerations Figures 3 and 4 illustrates the effect on accuracy of combining the AP and VT accelerations for the AC algorithm and PSD algorithms. Addition of the VT signals did yield similar and in some cases less accurate results for both algorithms. Discussion Motivated by the possibility of developing wearable gait systems for real-time gait parameter recognition, real-time gait parameter feedback and gait activity monitoring, our first aim was to examine the accuracy of SF estimates over a wide range of gait speeds for AC and PSD algorithms at different time window (TW) lengths. We found more accurate SF estimates when both gait speed and TW were Figure 2: Results for the spectral averaging in comparison with PSD algorithm at different time windows and at a wide range of gait speeds. h spectral averaging and * represents PSD. A 2 seconds TW, B 4 seconds TW and C 8 seconds TW. Data is presented as the mean absolute percentage error against camera observation. #JOURNAL OF MOBILE TECHNOLOGY IN MEDICINE VOL. 4 ISSUE 1 JANUARY 2015 5

Figure 3: Results for the superimposed autocorrelation in comparison to autocorrelation at different time windows and at a wide range of gait speeds. Superimposed autocorrelation h and regular autocorrelation *. A 2 seconds TW, B 4 seconds TW and C 8 seconds TW. Data is presented as the mean absolute percentage error against camera observation. Figure 4: Results for the superimposed PSD in comparison to PSD at different time windows and at different gait speeds. Superimposed PSD h and regular PSD *. Respectively A 2 seconds TW, B 4 seconds TW and C 8 seconds TW. Data is presented as the mean absolute percentage error against camera observation. higher. Furthermore, we found an interaction effect between TW length and gait speed for AC. These results support the idea to enlarge the TW length when the population of interest consists of slow walkers. Differences between the AC and PSD based estimates were small and not consistent across TW and speeds. Spectral averaging did improve accuracy for short TW lengths i.e. 2 or 4 seconds and at slow gait speeds. Combining estimates from AP accelerations with estimates form VT accelerations did not yield more accurate SF estimates. Finally, independent of the algorithm and TW used, SF estimation accuracy is always less at slow gait speeds compared to higher gait speeds. To the best of our knowledge, only a limited number of studies have presented guidance on minimum numbers of meters or steps for valid and reliable gait parameter recognition. For example, Moe- Nilssen et al. 8 used five strides, while Auvinet et al. 15 recommend 40 m. However, these studies did not explore effects of TW length and changes in gait speed systematically. Yang et al. 4 recently pointed out that the use of time windows may be necessary in view of variations in gait speed over time. Therefore future research in the field of activity monitoring should focus on the determination of the optimization of window length with respect to accuracy, reliability and sensitivity to variation in gait speed. Study limitations Over ground walking is different compared to treadmill walking 16 ; in treadmill walking gait variability is reduced compared to over ground walking 17. However, dominant frequencies measured around the waist in over ground walking reflect SF as well 3. Therefore, we expect no major differences in effects of the methodological choices investigated when this experimental protocol had been carried out in an over ground condition. Most advanced wearable gait systems are and will be developed for abnormal or pathological gaits such as in stroke and Parkinson s disease 5,18. Our conclusions and recommendations can be useful in developing algorithms for pathological gait, but have to be interpreted with caution as our study used young, healthy subjects only. 4.1 Conclusions We examined the accuracy of estimating step frequency over a wide range of gait speeds derived from accelerometer signals in relation to signal analysis. When developing specific algorithms for the detection of step frequency, the optimal TW #JOURNAL OF MOBILE TECHNOLOGY IN MEDICINE VOL. 4 ISSUE 1 JANUARY 2015 6

length depends on gait speed for both AC and PSD based estimates. We recommend a minimum TW length of 4 seconds when using AC and PSD algorithms and when using the PSD algorithm to use spectral averaging, as this leads to better results at short TW and low gait speeds. Combining AP with VT acceleration data did not improve estimates of step frequency. Conflict of interest statement The authors state that there was no conflict of interests with any financial or personal relationships or organizations that could influence the research results. Acknowledgements We would like to thank our research assistants Femmy Troost and Desiree Pultrum for their contribution in data collection and data analysis. We also would like to thank our financial funder SIA RAAK, 2010-2-024 INT. References 1. Hreljac A, Marshall RN. Algorithms to determine event timing during normal walking using kinematic data. J Biomech 2000;33:7836. 2. Mulder T, Nienhuis B, Pauwels J. Clinical gait analysis in a rehabilitation context: some controversial issues. Clin Rehabil 1998;12:99106. 3. Moe-Nilssen R. A new method for evaluating motor control in gait under real-life environmental conditions. Part 2: Gait analysis. Clin Biomech 1998;13:32835. 4. Yang C-C, Hsu Y-L, Shih K-S, Lu J-M. Real-time gait cycle parameter recognition using a wearable accelerometry system. Sensors 2011;11:731426. 5. Wagenaar RC, Sapir I, Zhang Y, Markovic S, Vaina LM, Little TDC. Continuous monitoring of functional activities using wearable, wireless gyroscope and accelerometer technology. Conf Proc IEEE Eng Med Biol Soc 2011; 48447. 6. Vitacca M, Bianchi L, Guerra a, Fracchia C, Spanevello a, Balbi B, et al. Tele-assistance in chronic respiratory failure patients: a randomised clinical trial. Eur Respir J Off J Eur Soc Clin Respir Physiol 2009;33:4118. 7. Henriksen M, Lund H, Moe-Nilssen R, Bliddal H, Danneskiod-Samsøe B. Test-retest reliability of trunk accelerometric gait analysis. Gait Posture 2004;19:28897. 8. Moe-Nilssen R, Helbostad JL. Estimation of gait cycle characteristics by trunk accelerometry. J Biomech 2004;37:1216. 9. Ichinoseki-Sekine N, Kuwae Y, Higashi Y, Fujimoto T, Sekine M, Tamura T. Improving the accuracy of pedometer used by the elderly with the FFT algorithm. Med Sci Sports Exerc 2006;38:167481. 10. Jovanov E, Wang E, Verhagen L, Fredrickson M, Fratangelo R. defoga real time system for detection and unfreezing of gait of Parkinson s patients. Conf Proc IEEE Eng Med Biol Soc 2009;51514. 11. Goldie PA, Matyas TA, Evans OM. Deficit and Change in Gait Velocity During Rehabilitation After Stroke. Arch Phys Med Rehabil 1996;77:107482. 12. Egerton T, Williams DR, Iansek R. Comparison of gait in progressive supranuclear palsy, Parkinson s disease and healthy older adults. BMC Neurol 2012;12:116. 13. Subjects IH. World Medical Association Declaration of Helsinki. Ethical principles for medical research involving human subjects. Nurs Ethics 2002;9:1059. 14. Tura A, Rocchi L, Raggi M, Cutti AG, Chiari L. Recommended number of strides for automatic assessment of gait symmetry and regularity in above-knee amputees by means of accelerometry and autocorrelation analysis. J Neuroeng Rehabil 2012;9:11. 15. Auvinet B, Chaleil D, Barrey E. Accelerometric gait analysis for use in hospital outpatients. Rev Rhum Engl Ed 1999;66:38997. 16. Alton F, Baldey L, Caplan S, Morrissey MC. A kinematic comparison of overground and treadmill walking. Clin Biomech (Bristol, Avon) 1998;13:434 40. 17. Dingwell JB, Cusumano JP, Cavanagh PR, Sternad D. Local Dynamic Stability Versus Kinematic Variability of Continuous Overground and Treadmill Walking. J Biomech Eng 2001;123:27. 18. Yang C-C, Hsu Y-L, Shih K-S, Lu J-M. Real-time gait cycle parameter recognition using a wearable accelerometry system. Sensors 2011;11:731426. #JOURNAL OF MOBILE TECHNOLOGY IN MEDICINE VOL. 4 ISSUE 1 JANUARY 2015 7