TMA 2940 CE. Operator s Manual. Thermomechanical Analyzer

Similar documents
Liquid Nitrogen Cooling Accessory CE

TGA 2950 CE. TA Instruments. Operator s Manual. Thermogravimetric Analyzer. PN Rev. D (Text and Binder) PN Rev.

TGA Operator s Manual. Thermogravimetric Analyzer. PN Rev. D (Text and Binder) PN Rev. D (Text Only)

DSC Differential Scanning Calorimeter

! Warning, refer to accompanying documents.

ACV-10 Automatic Control Valve

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

User s Guide Temperature Sensor Converter TSC-599

Apex Evaporator Manual

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Procedures for operation of the TA Instruments DSC

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

RPS900W Redundant Power Supply. Installation Guide.

USER MANUAL LTH-255A Hand Piece

Preferred Instruments Danbury, CT USA

FOURIER TRANSFORM INFRARED SPECTROSCOPY

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

Generating Calibration Gas Standards

THE BP-690 SERIES. Operating and Service Manual. Series includes all variants of BP-LF/MF-690/691

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Best Practice Guide, Servomex 2700

2-1- Shipping Receiving Lifting Location on Site Installation Stages 6

Preparation and Installation of the Sanitary BDI-FLX Sensor and Connection to the BDI-FLX Interface Cable

LOCKOUT/TAGOUT PROGRAM

Installing ASI Umbilical Delivery Systems

BOTTOM HINGED FLOOD BARRIER

Lockout/Tagout Training Overview. Safety Fest 2013

PERFORM Operating Document

Approved by Principal Investigator Date: Approved by Super User: Date:

Purpose and Scope: Responsibilities: Warnings and precautions: Background Beers law

Spiratec ST14, ST16 and ST17 Sensor Chambers and sensors

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #:

Discovery HP-TGA 75/750. Site Preparation Guide

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

Installation Instructions

RAM 4021-DPX Operation Manual

Installation Instructions

INSTRUCTIONS FOR MODEL SG6500 GAS HEATERS

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

Compressed Gas Cylinders - Guideline

BGA244 Binary Gas Analyzer

INSTRUCTION MANUAL PARTICLE SENSOR KS-40AF Higashimotomachi, Kokubunji, Tokyo , Japan

Installation, Operation and Maintenance Manual for Back Pressure Regulator

RH800 & 2000 BASIC OPERATIONS

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Notice. VeraCode Vortex Incubator User Guide

Operation Instruction Manual

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

RAM Operation Manual

GRI-1300/1350Li. User Manual

Armfield Distillation Column Operation Guidelines

Gas Switching Accessory

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Pressure Measurement Single-range transmitters for general applications

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

CP10 Sensor Installation and Maintenance Instructions

Information Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100

GALISO SOAP OPTION FOR PCT-ADW-D

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

Propane Conversion Kit Instructions

Vapor Sorption Analyzer

All Purpose AP-Series Operation Manual

Approved by BNC Managing Director Paul Lum Date: Approved by PI (s) /Super User (s): _Peter Hosemann/Andy Minor_Date:

PERFORM Operating Document

LOCKOUT/TAGOUT PROGRAM

Warnings: Notes: Revised: January 8,

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage

PQI Model 301 & 300Test Block Procedure

Agilent 1220 Infinity II LC Mobile Upgrade Kit

PROCESS ANALYSERS. SERVOPRO PureGas. Gas Purifier. Operator Manual. Part Number: A Revision: 0 Language: UK English

5890II GC Standard Operating Procedure 9/2/2005

Inert Air (N2) Systems Manual

i) Filtration machine, measuring equipment and other related machinery

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

PURA. Pure Gas Dewpoint Transmitter. Users Guide

MANUFACTURING TECHNICAL INSTRUCTIONS - SAFETY FALL HAZARD CONTROL REQUIREMENTS SERIES & NO. SMI - 157

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

Stirring Water Bath. Instruction Manual

BT6HC Hygienic Sanitary Balanced Pressure Steam Trap for High Capacity and CIP/SIP Applications

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

QUICK START GUIDE TO DRINKING WATER MONITORING

Standard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation

ShellPa. Standard Mechanical Cell Stretch System Model No: NNMS Serial #: User Manual

Transcription:

109 Lukens Drive New Castle, DE 19720 Thermal Analysis & Rheology A SUBSIDIARY OF WATERS CORPORATION TMA 2940 CE Thermomechanical Analyzer Operator s Manual PN 825615.001 Rev. C (Text and Binder) PN 825615.002 Rev. C (Text Only) Issued November 1999 I

II

Table of Contents III

Table of Contents IV

Table of Contents V

Table of Contents VI

Table of Contents VII

Table of Contents VIII

Table of Contents IX

Table of Contents X

Notes, Cautions, and Warnings NOTE: CAUTION:!WARNING Highlights important information about equipment or procedures. Emphasizes a procedure that may damage equipment or cause loss of data if not followed correctly. Marks a procedure that may be hazardous to the operator or to the environment if not followed correctly. XI

Helplines To TA Instruments XII

Safety CE Compliance XIII

Safety Instrument Symbols XIV

Safety Electrical Safety!WARNING!WARNING Because of the high voltages in this instrument, untrained personnel must not remove the cabinet cover unless specifically directed to do so in the manual. Maintenance and repair of internal parts must be performed by TA Instruments or other qualified service personnel only. After transport or storage in humid conditions, this equipment could fail to meet all the safety requirements of the safety standards indicated. Refer to the NOTE on page 2-10 for the method of drying out the equipment before use. Handling Liquid Nitrogen XV

Safety!WARNING WARNING: Liquid nitrogen evaporates rapidly at room temperature. Be certain that areas where liquid nitrogen is used are well ventilated to prevent depletion of oxygen in the air. XVI

Safety Lifting the Instrument!WARNING Use two people to lift and/or carry the instrument. The instrument is too heavy for one person to handle safely. XVII

Using This Manual XVIII

!WARNING Potential Asphyxiant Liquid nitrogen can cause rapid suffocation without warning. Store and use in an area with adequate ventilation. Do not enter confined spaces where nitrogen gas may be present unless the area is well ventilated. The warning shown above applies to the use of liquid nitrogen. Oxygen depletion sensors are sometimes utilized where liquid nitrogen is in use. XIX

XX

CHAPTER 1: Introducing the TMA 2940 CE 1 1

Introduction 1 2

Introduction Introduction Components 1 3

Introduction Furnace Assembly Display Keypad Cabinet Balance Enclosure 1 4

The 2940 CE Instrument The 2940 CE Instrument 1 5

Introduction Standby DimChg 23.25 C 0.135µm TMA 2940 Thermomechanical Analyzer 1 6

The 2940 CE Instrument 2940 CE Display 2940 CE Keypad HEATERPOWER NOTE: Experiment information and instrument constants are entered from the controller keyboard, not the instrument keypad. 1 7

Introduction SCROLL ZERO LENGTH MEASURE LENGTH PROBE 1 8

The 2940 CE Instrument Probe PROBE FURNACE START Probe 1 9

Introduction STOP REJECT SCROLL STOP STOP SCROLL-STOP 1 10

NOTE: The 2940 CE Instrument SCROLL operates normally (scrolls the text) until the STOP key is pressed. Then the display returns to the signal displayed on the screen before SCROLL was pressed. SCROLL-STOP STOP HEATER Switch HEATER NOTE: POWER Switch The light in the HEATER switch will glow whenever the power control circuits are enabled. This occurs when either a method is running or an end of method function (return to temp range) is selected and active. (See Heater Indicator Light in Chapter 6) for more information. POWER 1 11

Introduction Accessories Standard Accessories 1 12

Accessories Optional Accessories 1 13

Introduction Specifications 1 14

Specifications 1 15

Introduction 1 16

CHAPTER 2: Installing the TMA 2940 CE 2 1

Installation 2 2

Unpacking the TMA Unpacking/Repacking the 2940 CE NOTE: These instructions are also found as separate unpacking instructions in the shipping box. Unpacking the 2940 CE!WARNING Have an assistant help you unpack this unit. Do not attempt to do this alone. 2 3

Installation 2 4

Unpacking the TMA 2 5

Installation Removing the Shipping Material CAUTION: Steps 1 through 9 may be completed by the customer. Only qualified service personnel should proceed with steps 10 through 19. 2 6

Unpacking the TMA!WARNING Whenever you remove the balance enclosure, make sure that the power is off and the instrument is unplugged. CAUTION: Only qualified service personnel should proceed beyond this point. 2 7

Installation Repacking the 2940 CE 2 8

Installing the Instrument Installing the Instrument CAUTION: To avoid mistakes, read this entire chapter before you begin installation. 2 9

Installation Inspecting the System NOTE: Drying out the instrument may be needed, if it has been exposed to humid conditions. Certain ceramic materials used in this equipment may absorb moisure, causing leakage currents to exceed those specified in the applicable standards (see page xi) until the moisture is eliminated. It is important to be certain that the instrument ground is adequately connected to the facilities ground for safe operation. Run the following method to dry out the instrument (refer to Running Experiments for further information). 1 Ramp at 10 C/min to 400 C 2 Isothermal for 30 min. 2 10

Installing the Instrument Choosing a Location 2 11

Installation Installing the Weight Tray Door 2 12

Installing the Instrument Installing the Stage 2 13

Installation 2 14

Installing the Instrument SPRING CLIP 2 15

Installation Stage Shield Thermocouple 2 16

Installing the Instrument Installing the Expansion/ Penetration Probes NOTE: To learn how to install the film/fiber probe, turn to Chapter 4, "Using Your Options." Installing a Probe 2 17

Installation Removing a Probe 2 18

Installing the Instrument Connecting Cables and Gas Lines NOTE: NOTE: CAUTION:!WARNING * All directional descriptions for this section are written on the assumption that you are facing the back of the instrument. Connect all cables before connecting the power cords to outlets. Tighten the thumbscrews on all computer cables. When plugging or unplugging power cords, always handle them by the plugs, not by the cords. Protect power and communications cable paths. Do not create tripping hazards by laying them across accessways. GPIB Cable 2 19

Installation Power Cord Serial Connector Ready Light Reset Button Address Dial GPIB Connector NOTE: If you have a multiple instruments system, each instrument must have a different address. 2 20

Installing the Instrument NOTE: The instrument s GPIB address is displayed during startup and can also be viewed on the instrument s status display. 2 21

Installation Purge Line!WARNING CAUTION: Use of an explosive gas as a purge gas is dangerous and is not recommended for the TMA instrument. For a list of the gases that can be used in the TMA, see Chapter 3. Use of corrosive gases will shorten the life of the instrument. 2 22

Installing the Instrument 2 23

Installation Air Cool Line 2 24

Installing the Instrument Power Cable NOTE: Connect all other cables and gas lines before connecting the power cable to a wall outlet. POWER TMA 2940 Thermomechanical Analyzer CAUTION: Before plugging the TMA power cable into the wall outlet, make sure the instrument is compatible with the line voltage. Check the label on the back of the unit to verify the voltage. Make sure that you use the stepdown transformer (PN 994560.901) with the instrument to provide the correct line voltage and ensure safety. 2 25

Installation Starting the 2940 CE NOTE: Allow the TMA to warm up for at least 30 minutes before performing an experiment. POWER 2 26

Starting the 2940 CE Standby DimChg 23.25 C 0.135µm TMA 2940 Thermomechanical Analyzer 2 27

Installation Shutting Down the 2940 CE POWERHEATER 2 28

CHAPTER 3: Running Experiments 3 1

Running Experiments 3 2

Overview Overview 3 3

Running Experiments Before You Begin 3 4

Calibrating the TMA Calibrating the TMA NOTE: Please note that this chapter provides guidelines on when to perform these calibrations. For details on how to perform each calibration procedure, refer to the instrument control software online documentation. Force Calibration When to Calibrate 3 5

Running Experiments NOTE: A force calibration performs the functions of a probe calibration (see the next section). Therefore, it is not necessary to perform a probe calibration immediately following a force calibration. Probe Calibration Temperature Calibration 3 6

Calibrating the TMA When to Calibrate: NOTE: Regardless of the size or shape of the sample that you are running on the TMA, position the tip of the thermocouple so that it bends at a 90 angle and lies flat against the platform. It should be close to, but not touching the sample. 3 7

Running Experiments Cell Constant 3 8

Calibrating the TMA 3 9

Running Experiments Running a TMA Experiment Experimental Procedure 3 10

Running a TMA Experiment Selecting a Probe NOTE: Refer to Chapter 4 for details regarding the optional probes. 3 11

Running Experiments Zeroing the Auto Measure System ZERO LENGTH 3 12

Running a TMA Experiment Sample Preparation and Loading Sample Preparation Sample Loading ZERO LENGTH PROBE 3 13

Running Experiments NOTE: It is recommended that you place a quartz wafer or a piece of thin aluminum foil between the stage and any thermoplastic samples to prevent damage to the stage. 3 14

Running a TMA Experiment NOTE: Low force levels may accentuate vibrationinduced signal noise. MEASURE LENGTH Thermocouple Positioning Measuring the Sample Length MEASURE LENGTH 3 15

Running Experiments Setting Up an Experiment 3 16

Running a TMA Experiment Setting Up Accessories Using Air Cool 3 17

Running Experiments Using a Purge Gas Suggestions 3 18

Running a TMA Experiment Starting an Experiment START 3 19

Running Experiments Stopping an Experiment CAUTION: Removing Samples STOP SCROLL-STOP The REJECT function discards all experiment data. 3 20

Running a TMA Experiment CAUTION: Use extreme caution when working with acid solutions. 3 21

Running Experiments Subambient Experiments NOTE: If the Return to temperature range methodend condition is activated, press the STOP key to deactivate it before trying to cool the TMA. Performing a Cooling Experiment 3 22

Subambient Experiments CAUTION: If you are using liquid nitrogen as a coolant, be sure to follow all of the safety precautions listed in the front of this manual for handling liquid nitrogen. NOTE: Adjusting the coolant during the course of the run may cause nonlinear heating rates, which will be reflected in noisy or discontinuous curves. Techniques for Improved Subambient Operation 3 23

Running Experiments CAUTION: Do not program furnace temperatures greater than 600 o C when using the aluminum mass. 3 24

CHAPTER 4: Using Your Options 4 1

Using Your Options 4 2

Overview Overview 4 3

Using Your Options Film/Fiber Accessory Description 4 4

Film/Fiber Accessory Operating Procedures NOTE: When comparing run data, make sure that all comparison scans are made with the thermocouple in the same position and at the same heating rate. Accessory Installation 4 5

Using Your Options Removing the Existing Probe Removing the Stage 4 6

Film/Fiber Accessory Changing the Instrument Mode CAUTION: The instrument must be placed into the film/ fiber mode before the film/fiber accessory is installed, in order to prevent damage to the instrument or to the accessory. Installing the Film/ Fiber Probe NOTE: The special configuration of the film/fiber accessory requires the installation of the probe before the stage. 4 7

Using Your Options Installing the Film/Fiber Stage 4 8

Film/Fiber Accessory NOTE: After the thermocouple is installed, its tip may interfere with the zero length operation. Temporarily move the thermocouple when performing this operation. Calculating the Offset 4 9

Using Your Options 4 10

Film/Fiber Accessory Calibration 4 11

Using Your Options Sample Preparation Using the Aluminum Balls Sample Length 12 to 25 mm (0.5 to 1.0 inch) 4 12

Film/Fiber Accessory 4 13

Using Your Options Using the Stainless Steel Clamps 4 14

Film/Fiber Accessory 4 15

Using Your Options 4 16

Flexure Accessory Flexure Accessory Description 4 17

Using Your Options Accessory Installation Calibration 4 18

Flexure Accessory Sample Preparation ZERO LENGTH 4 19

Using Your Options 4 20

Dilatometer Accessory Dilatometer Accessory Description β β µ µ 4 21

Using Your Options β µ β β α α µ 4 22

Dilatometer Accessory Operating Procedures Considerations NOTE: If you compare data runs, make sure that all comparison scans are made with the vial filled to the same level with filling medium, with the thermocouple in the same position, and heated at the same rate. Accessory Installation 4 23

Using Your Options Calibration Baseline Calibration NOTE: The dilatometer filling medium must be kept dry. Store it in a desiccator when it is not being used. If the sample has become contaminated through long exposure to humidity, dry it in a vacuum oven at 200 o C for one hour. NOTE: 4 24 Marking the vial at the fill level and filling future sample runs to this level will insure that the baseline effect due to the fill medium are minimized from run to run.

Dilatometer Accessory Displacement Calibration 4 25

Using Your Options Dilatometer Accessory Sample Preparation NOTE: The dilatometer filling medium must be kept dry. Store it in a desiccator when it is not being used. If the sample has become contaminated through long exposure to humidity, dry it in a vacuum oven at 200 o C for one hour. 4 26

4 27

Using Your Options ZERO LENGTH NOTE: To avoid false thermocouple readings, make sure that the thermocouple tip does not touch the sample vial. 4 28

Parallel Plate Rheometer Accessory Parallel Plate Rheometer Accessory Description Accessory Installation 4 29

Using Your Options NOTE: Calibration To avoid false thermocouple readings, make sure that the thermocouple tip does not touch the alignment cages or plates. Exposed portions of the two thermocouple leads must not touch each other. 4 30

Parallel Plate Rheometer Accessory Sample Preparation Sample Encapsulating Press 4 31

Using Your Options Removing the Existing Die Assembly!WARNING The plunger in the upper hermetic die assembly is springloaded and should be removed with care. 4 32

Parallel Plate Rheometer Accessory NOTE: The upper crimping die is installed over, and locked to, the upper hermetic die. NOTE: The upper hermetic die is now held in position by either the lower hermetic or crimping die. NOTE: As the lower die-holder adjustment screw is being removed, spring tension on the hermetic die plunger is relieved. 4 33

Using Your Options Installing the Pellet Press Die Set 4 34

Parallel Plate Rheometer Accessory Preparing Moldable Powders 4 35

Using Your Options 4 36

Parallel Plate Rheometer Accessory 4 37

Using Your Options Preparing Resinous Materials!WARNING This procedure involves heating up the sample and mold assembly. Handle with tongs to prevent burns. 4 38

Parallel Plate Rheometer Accessory 4 39

Using Your Options Preparing Sheet Materials 4 40

Parallel Plate Rheometer Accessory Sample Loading NOTE: To avoid false thermocouple readings, make sure that the thermocouple tip does not touch the alignment cage, sample, or plates. Exposed portions of the two thermocouple leads must not touch each other. Zero Length 4 41

Using Your Options Cleanup CAUTION: These procedures involve the use of heat or chemicals; appropriate safety measures should be observed at all times. Thermoplastic Material 4 42

Parallel Plate Rheometer Accessory Thermosetting Material Calculations Correction Calculations for Thermal Expansion of Parallel Plates 4 43

Using Your Options α α α α x x x 1/ 4 44

Parallel Plate Rheometer Accessory α 4 45

Using Your Options Calculations for Viscosity and Wall Shear Rate Data η γ η π x x γ x η γ 4 46

Hemispherical Probe Hemispherical Probe 4 47

Using Your Options 4 48

CHAPTER 5: Technical Reference 5 1

Technical Reference 5 2

Description of the TMA 2940 CE Description of the TMA 2940 CE 5 3

Technical Reference Uses of Thermomechanical Analysis 5 4

Description of the TMA 2940 CE Calculating the Coefficient of Expansion α α α 5 5

Technical Reference Method Considerations 5 6

Status Codes Status Codes 5 7

Technical Reference START 5 8

Status Codes 5 9

Technical Reference START 5 10

Status Codes 5 11

Technical Reference 5 12

CHAPTER 6: Maintenance and Diagnostics 6 1

Maintenance and Diagnostics 6 2

Overview Overview 6 3

Maintenance and Diagnostics Routine Maintenance!WARNING Because there are high voltages in this instrument, untrained personnel must not attempt to test or repair any electrical circuits. Inspection 6 4

Routine Maintenance Cleaning Cleaning the Keypad Cleaning the Probe Assembly 6 5

Maintenance and Diagnostics Cleaning the Stage For small amounts of residue: For thorough cleaning: 6 6

Routine Maintenance Cleaning the Thermocouple 6 7

Maintenance and Diagnostics Replacements Replacing the Thermocouple 6 8

Replacements 6 9

Maintenance and Diagnostics 6 10

Diagnosing Power Problems Diagnosing Power Problems Fuses!WARNING Fuse 1 Fuse 2 Always unplug the instrument before you examine or replace the fuses. 6 11

Maintenance and Diagnostics Furnace Power Check POWER HEATER HEATER NOTE: The light in the HEATER switch will glow only after an experiment is initiated. The HEATER switch will continue to glow, even if it is switched to the OFF (0) position, until the "Stand By" status code is displayed. 6 12

Diagnosting Power Problems Heater Indicator Light HEATER STOP 6 13

Maintenance and Diagnostics Power Failures 6 14

TMA 2940 CE Test Functions TMA 2940 CE Test Functions The TMA Confidence Test 6 15

Maintenance and Diagnostics 2940 TMA V1.0 Confidence Test 00 TMA 2940 Thermomechanical Analyzer 6 16

TMA 2940 CE Test Functions 6 17

Maintenance and Diagnostics Replacement Parts 6 18

Replacement Parts 6 19

Maintenance and Diagnostics 6 20

Appendix A: Ordering Information A 1

Appendix A A 2

Index Index A B C I 1

Index D E F α I 2

Index G K H I I 3 L M

Index O P I 4

I 5 Index R S

Index I 6 T V

Index W Z I 7

Index I 8