Title: Haematoxylin & Eosin (H&E) Staining & Subsequent Mounting of Slides. Effective from: 01/03/2016 Reviewed annually

Similar documents
Title: Heat-induced Epitope Retrieval (HIER) Using a Pressure Cooker. Effective from: 01/03/2016 Reviewed annually

Tissue-Tek Prisma Plus & Tissue-Tek Glas TM g2

Imagine... full flexibility linked to high capacity. Tissue-Tek Prisma & Film. Automated Slide Stainer & Coverslipper

Imagine... full flexibility linked to high capacity

University of Saskatchewan Department of Mechanical Engineering Standard Operating Procedure # Mat0002

Histology Protocols. Safety and Good Laboratory Practice (GLP)

Standard Operating Procedure (SOP)

EXPERIMENT. Introduction: Microbiology- Rules for Success

PRODUCT SHEET. Order probe only as RXPROBE02

Before you start sampling, be sure to read

Laboratory Safety Training. Environmental Health and Safety

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Standard. Standard. Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00332

SOP Standard Operating Procedure. Etching All Materials

6900 Maintenance Instruction System Flush

PURDUE UNIVERSITY TEFLON CORE PCB THRU-HOLE PREPARATION FOR BLACK HOLE APPLICATION RF LAB. Professor in charge: William Chappell MSEE 289

Patients Guide for Self Administering Intravenous Antibiotics General points

VOLUMETRIC TECHNIQUES

Section 3. Classical Anatomy

STANDARD OPERATING PROCEDURE VORTEX-GENIE 2

Exeter Clinical and Health Research

Tissue-Plus. User Manual. O.C.T. Compound. Optimal Cutting Temperature Embedding Medium for Frozen Tissue Specimens

Exeter Clinical and Health Research

v /07/17 Page 1 of 18

Hydrofluoric Acid SOP

CHM 317H1S Winter Section P Procedures and Tables

Douglas Call 4/11/2008 1/5. Ammonia Treatment

Best Practices for Infinium Assays

ACID FAST BACTERIA MICROBIOLOGY STAINER / CYTOCENTRIFUGE WESCOR, INC

Laboratory Self Inspection Checklist LAB01

Toolkit for safe Use of Microbiological Safety Cabinets(MSC s)

SOP for use of the SX19 Stopped-Flow Instrument

Working Safely with Hydrofluoric Acid

Guidelines on Collection of Water Samples for Phytoplankton Analysis

CHEMISTRY FACULTY LABORATORY SAFETY CONTRACT CENTRAL CAMPUS

INVESTIGATING HEART FAILURE

ST 18 Aluminium Gas Model

Pipetting Small Volumes

ST 18 Aluminium Gas Model

1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique.

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

Standard. Bag & Stand Advanced Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00442

Electron Microscopy Centre

K-STATE GRAIN QUALITY RESEARCH PROGRAM STANDARD OPERATING PROCEDURES

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

Elcometer 138. Bresle Kit and Patches. Operating Instructions

PERFORM Operating Document

6800 Maintenance Instruction System Flush Procedure

Elcometer 138. Bresle Kit and Patches. Operating Instructions

UNH Department of Forensic Sciences

Spa Start Up & Maintenance

Standard Operating Procedure. Air Displacement Pipette Calibration

Lab-Specific Standard Operating Procedure (LSOP)- Pyrophoric Chemicals

Digital Melting Point Apparatus

Wolfson Campus Natural Science Department. Biology Laboratory. Student Safety Contract

PRE LABORATORY ASSIGNMENT: Lab Section Score: /10 READ THE LAB TEXT BEFORE ATTEMPTING THESE PROBLEMS!

Transfer Procedures. By the end of this chapter, you will be able to: The Transfer Process. Chapter 12

[USING THE VITROBOT April 9, Using the Vitrobot

Installation guide Scale Control. Before installing read this guide first

GV Standard X-Vent. Setup, Commissioning & Installation Guide

PREPARATION AND PROPERTIES OF ATMOSPHERIC GASES I: Carbon Dioxide and Oxygen

Quenching of Self-Heating Substances and Mixtures

Salt Lowers the Freezing Point of Water

Risk Assessment for: Roaches Pyrotec dyeing machines

INSTRUCTION MANUAL FOR USE OF BABY WALKER FLOWER

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

Pool Owners Guide. As Recommended by Barrier Reef Pools LLC.

Ammonia Treatment Instructions Page 1 of 5

MOLEBIO LAB #1: Microquantity Measurement

Safety Procedures on Vacating Premises

The ultimate in water softening

Instruction Manual Updated 7/26/2011 Ver. 2.2

Version BA D

Department of Chemistry

Appendix A: Induction check list

Urine Sediment Preparation

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.

Armfield Distillation Column Operation Guidelines

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

TJF-Q180V Cleaning and Disinfection Checklist

Varipette Operating manual

PERFORM Operating Document

AN EXCITING NEW CONCEPT MICROBIOLOGY SLIDE STAINER/ CYTOCENTRIFUGE BIOMEDICAL PRODUCTS WESCOR, INC

Chapter 4: Implementing Effective Exposure Controls

CHEM254 #4 The Synthesis of a Tertiary Alcohol Using a Pre-Made Grignard Reagent 1

IMPROVING QUALITY THROUGH INNOVATION: A FULLY AUTOMATED SOLUBLE SALT METER

BIOSAFETY SELF AUDIT PRINCIPAL INVESTIGATOR/PERMIT HOLDER: CONTAINMENT LEVEL: 1 2 (Shaded) RISK GROUP: 1 2 AUDIT TO BE COMPLETED BY (DATE):

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

Lab: Safety Due: Lab # 0

STANDARD OPERATING PROCEDURE TEMPLATE

SECTION 2 SMART PAYOUT MANUAL SET FIELD SERVICE MANUAL

Pipettor. User Manual

Student Information & Laboratory Safety Contract

Waste Management System.

Ultima Salt System Overview

Instruction Sheet. Calibration of Multiple 1720 and FT660 Turbidimeters

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

limbsandthings.com Advanced Venepuncture Arm User Guide For more skills training products visit Limbs & Things Ltd.

Katy Independent School District Science Safety Rules

TJF-Q180V Cleaning and Disinfection Checklist

Transcription:

Document Number: SuRF-HIS-052.05 Title: Haematoxylin & Eosin (H&E) Staining & Subsequent Mounting of Slides Version 5.0 Author: Melanie McMillan Effective from: 01/03/2016 Reviewed annually SOP History Number: Date: Reasons for Change: 01 01/06/2013 Original 02 17/03/2014 Title change & change to COSHH hazard listing 03 01/12/2014 Title change & change to procedure 04 01/11/2015 Change to purpose, scope, responsibilities, procedure, related documents / references and appendices 05 01/03/2016 Change to procedure and related documents / references 1.0 Purpose: The purpose of this Standard Operating Procedure (SOP) is to describe the current procedure for the haematoxylin and eosin (H&E) staining of paraffin embedded tissues and subsequent mounting of slides within the Histology, Immunodetection & Aquila-HistoPlex sections of the Shared University Research Facilities (SuRF), hereafter collectively referred to as SuRF Histology. H&E staining is a popular method in histology, used widely in medical diagnosis. The staining method involves application of hemalum, which is a complex formed from aluminium ions and oxidized haematoxylin. This colours nuclei of cells (and a few other objects, such as keratohyalin granules) blue. The nuclear staining is followed by counterstaining with an aqueous or alcoholic solution of eosin Y, which colours other eosinophilic structures in various shades of red, pink and orange. Page 1 of 10

2.0 Scope: This SOP applies to all staff, including students, visitors and any other supervised / trained individuals involved in this procedure within SuRF Histology, based in the Queen s Medical Research Institute (QMRI), Edinburgh. 3.0 Responsibilities: This document is a guide only on site training is essential before use. 3.1 All staff are responsible for ensuring that methods are followed in accordance with this SOP after suitable training, and where relevant, update their SOP Training Record (Standard document QA008) accordingly. 3.2 All staff involved in this procedure must be familiar with the location of any manufacturers manuals, instructions or guidance pertaining to equipment or methodology, and are strongly advised to read and understand this material before performing this procedure. 3.3 All staff must have read any corresponding relevant risk assessment / COSHH documents before performing this procedure. 4.0 Procedure: Sections to be stained should have been baked in a 55 C oven overnight. Control slides must always be used to check staining efficiency. Practically every tissue has an internal control so no other control is needed, if a control is desired, lung aorta strips will provide good material. Validation of the Shandon Varistain Gemini ES Slide Stainer is performed every time the machine is used prior to the first run of the day with a control slide of lung aorta strips. This must be dated and signed. Tissue that has been dehydrated, embedded in wax and then sectioned must be dewaxed and rehydrated before staining with specific stains or dyes. The reagents for dewaxing and rehydrating by hand (similarly those for dehydrating and clearing) slides are laid out in the laboratory within the fume extraction cabinet or on downflow benching. When not in use replace the lids on the reagent dishes to minimise the release of fumes. 4.1 Preparation of solutions: 95% ethanol (GP grade): ethanol (GP grade) deionised water 95ml 5ml Page 2 of 10

80% ethanol (GP grade): ethanol (GP grade) deionised water 70% ethanol (GP grade): ethanol (GP grade) deionised water 80ml 20ml 70ml 30ml 1% acid alcohol: hydrochloric acid, concentrated (sp.gr.1.19) 1ml 70% ethanol 99ml Scott s tap water: Commercially available sodium bicarbonate 2g magnesium sulphate 20g deionised water 1000ml Dissolve the salts in the water. Store stock solutions at room temperature. Haematoxylin (Harris ): Commercially available haematoxylin 2.5g ethanol 25ml potassium alum 50g deionised water 500ml mercuric oxide 1.25g or sodium iodate 0.5g acetic acid (glacial) 20ml The haematoxylin is dissolved in ethanol, and is then added to the alum, which has previously been dissolved in the warm deionised water in a 2-litre flask. The mixture is rapidly brought to the boil and the mercuric oxide or sodium iodate is then slowly and carefully added. Plunging the flask into cold water or into a sink containing chipped ice rapidly cools the stain. When the solution is cold, the acetic acid is added, and the stain is ready for immediate use. The glacial acetic acid is optional but its inclusion gives more precise and selective staining of nuclei. Eosin Y: Commercially available Note historically there are 2 working solutions of Eosin Y in use in Histology. (1) An aqueous solution from Thermo Fisher Scientific, used in E1.24 and (2) a mix of Eosin Y 515, alcoholic and Eosin Y, 1% aqueous both from Leica Biosystems, used in E1.27 (1 part alcoholic to 3 parts aqueous). Both achieve the same result. 4.2 Dewax and rehydrate by hand: Note: Allow any excess fluid to drain from the slide rack before proceeding to the next solution. Page 3 of 10

Xylene (1) Fume extraction unit 5 minutes Xylene (2) Fume extraction unit 5 minutes Xylene (3) Fume extraction unit 5 minutes Ethanol (1) Fume extraction unit 20 seconds Ethanol (2) Fume extraction unit 20 seconds Ethanol (3) Fume extraction unit 20 seconds 95% ethanol (GP grade) Fume extraction unit 20 seconds 80% ethanol (GP grade) Fume extraction unit 20 seconds 70% ethanol (GP grade) Fume extraction unit 20 seconds Wash in running water Sink 2 minutes 4.3 Staining and expected results by hand: The Haematoxylin used is commercially available ready-made Harris Haematoxylin. This should be filtered before use and replaced every two weeks. Although slides should always be quality controlled by eye and solution changed sooner if there is a problem. Place the slides into haematoxylin for 5 minutes. Remove and wash in running water for 20 seconds. Differentiate for 5 seconds maximum in 1% acid alcohol. Transfer to a dish of Scott s tap water substitute for 2 minutes until the tissue sections turn blue. Place slides in the eosin solution and stain for 2 minutes. Remove and wash in running water for 20 seconds. Control slides must always be used to check staining efficiency. Proceed to section 4.4 dehydrate, differentiate in ethanol and clear in xylene. Then proceed to section 4.6 and mount either by hand or by using Shandon ClearVue Coverslipper. 4.4 Dehydrate and clear by hand: Note: Allow any excess fluid to drain from the slide rack before proceeding to the next solution. 70% ethanol (GP grade) Fume extraction unit 20 seconds 80% ethanol (GP grade) Fume extraction unit 20 seconds 95% ethanol (GP grade) Fume extraction unit 20 seconds Ethanol (1) Fume extraction unit 20 seconds Ethanol (2) Fume extraction unit 20 seconds Ethanol (3) Fume extraction unit 20 seconds Xylene (1) Fume extraction unit 5 minutes Xylene (2) Fume extraction unit 5 minutes Xylene (3) Fume extraction unit 5 minutes Results: nuclei blue background pink to red Page 4 of 10

4.5 Using Shandon Varistain Gemini ES Slide Stainer, Protocol 2 see appendix 1 The Shandon Varistain Gemini ES Slide Stainer is designed for high throughput, precision, safety, durability and ultimate flexibility in histology and cytology staining applications. This Shandon stainer features load-ondemand multiple baskets and programs, power outage backup and built in fume control to make staining easier, faster and safer. Training and guidance will be provided prior to initial usage. NOTE: this machine is pre-programmed so there is NO need to change any of the settings. An access code is now in place and no program can be changed. In summary: Check the levels of all the reagents in the baskets and, if low, top these up. Turn the cold water tap on (until the black lines on the tap align). Switch on the electrical supply on the front of the control box (located underneath the staining machine). Wait until the machine completes its test cycle with the validation control slide of spleen. Place the microscope slides into the special slide racks with the label end next to the arrow on the rack and sections facing in the direction of this arrow. Lock the slides in place by moving the locking bar on the rack away from the arrow. Push the TOP clear window to the right (the monitor display will read H&E other and place the rack of slides into the LOAD basket. Close the window and press the URGENT START button on the control box. Press the OVERIDE button if the machine alarms. The machine will start automatically. Up to 5 racks may be loaded at any one time. Switch off in reverse order. Proceed to section 4.6 and mount either by hand or by using Shandon ClearVue Coverslipper. If an error message occurs then refer to the key instructions in the manufacturers manual. Cleaning and maintenance should be carried out every 15 cycles as per manufacturer s instruction manual. Results: nuclei blue background pink to red 4.6 Mounting slides: 4.6.1 by hand: Slide mounting should always be performed under the fume extraction unit within the main laboratory. Page 5 of 10

Unless otherwise stated slides should be mounted directly from xylene using Pertex. Nitrile gloves are available for use when mounting sections. Place appropriately sized coverslips onto the blotting paper under the fume extraction unit. Using a pastette, place a drop of mountant onto each coverslip. Remove a slide from the xylene and align the long edge of the slide with the coverslip, ensuring that the section is facing towards the coverslip. Tilt the slide towards the coverslip until it touches the mountant. Gently release the slide allowing the mountant to spread between the coverslip and the slide. Turn the slide over so that the coverslip is now on top of the slide. If necessary, center the coverslip over the tissue section and remove any air bubbles by pressing down gently with a finger. Wipe away any excess mountant using a clean cloth. If unsuccessful, place the slide and coverslip into the xylene and slide the coverslip away from the section, then remount using a new coverslip. (Do not pull the coverslip off.) Continue until all slides are mounted. Once slides are mounted they can be left on cardboard slide trays in the fume hood to dry. 4.6.2 using Shandon ClearVue Coverslipper: The Shandon ClearVue Coverslipper provides a fast and accurate automated method for coverslipping glass slides. Training and guidance will be provided prior to initial usage. In summary: Switch on the electrical supply at the mains plug & on the side of the machine. Wait until the machine completes its test cycle. Check the mountant level in the glass bottle at the front of the machine and, if low, top up by unscrewing the plastic lid and add more mountant to the level marked on the bottle. Place the microscope slides into the special slide racks with the label end next to the arrow on the rack and sections facing in the direction of this arrow. Lock the slides in place by moving the locking bar on the rack away from the arrow. Place the rack into the coverslipper (open the lid on the lower left hand of the coverslipper and hook the rack on the metal ridge with the sections facing upwards) Close the lid and the machine will start automatically. Once complete, the mounted slides will be dispensed into the top left hand chamber. These can be left there to dry or removed by opening the lid and lifting out the rack of slides. Up to 5 racks may be loaded on the machine at any one time. Switch off in reverse order. If an error message occurs then refer to the key instructions in the manufacturers manual. Page 6 of 10

Cleaning and maintenance should be carried out as per manufacturer s instruction manual once a week. 4.7 Changing reagents: All working reagents are changed weekly but individual reagents may be changed more often if required. Stock reagents are changed every 2 months or as according to expiry date on bottle. A record of this is kept on the wall next to the reagent dishes and should be filled in whenever reagents are changed. Discard the reagents into a properly labelled waste container. Clean work area with mild detergent solution. Clean the set of staining dishes and fill with fresh reagents. Place clean lids on the dishes. Complete and sign the reagent change record. 5.0 Related documents / references: COSHH: SuRF-COSHH-052: Haematoxylin and Eosin (H&E) Staining and Subsequent Mounting of Slides QA008 SOP Training Record QA013 Shandon Varistain Gemini ES Reagent Change Form Manufacturer s manuals / instructions The University of Edinburgh Health & Safety Policy / Codes of Practice (available on University s Health and Safety Department website) College of Medicine and Veterinary Medicine Health and Safety Manual (available on University s Health and Safety Department website) Bancroft J.D., and Gamble M. Theory and Practice of Histological Techniques 6 th Edition (2008) Churchill Livingstone, Elsevier Limited Page 7 of 10

6.0 Appendices Appendix 1 Shandon Varistain Gemini ES Protocol 2 Step Reagent Conc. (%) Uses Time Limit Agitate 1 Dry Storage 0 00:00 No Maximum Initial 2 Xylene 20 05:00 No Maximum Initial 3 Xylene 20 05:00 No Maximum Initial 4 Xylene 20 05:00 No Maximum Initial 5 Xylene 20 05:00 No Maximum Initial 6 Ethanol 100 20 02:00 No Maximum Initial 7 Ethanol 100 20 02:00 No Maximum Initial 8 Ethanol 95 20 02:00 No Maximum Initial 9 Running Water Wash 20 01:00 No Maximum Initial 10 Haematoxylin 20 05:00 Standard Initial 11 Running Water Wash 20 01:00 Standard Initial 12 Acid Alcohol 20 00:05 Critical Initial 13 Running Water Wash 20 01:00 No Maximum Initial 14 Scotts Tap Water Sub. 20 01:00 Standard Initial 15 Running Water Wash 20 01:00 No Maximum Initial 16 Aqueous Eosin Y 20 01:00 Standard Initial 17 Running Water Wash 20 01:00 Standard Initial 18 Ethanol 95 20 00:30 No Maximum Initial 19 Ethanol 100 20 00:30 No Maximum Initial 20 Ethanol 100 20 00:45 No Maximum Initial 21 Xylene 20 01:00 No Maximum Initial 22 Xylene 20 01:00 No Maximum Initial 23 Xylene 20 01:00 No Maximum Initial 24 Xylene 20 01:00 No Maximum Initial Page 8 of 10

7.0 Approval and sign off: Author: Name: Position: Signature: Melanie McMillan Principal Investigator Date: Management Approval by: Name: Mike Millar Position: Facility Manager Signature: Date: QA Release by: Name: Position: Signature: Robin Sellar QA Manager Date: Page 9 of 10

8.0 Verification by users: Sign below to indicate you have read and understood the activity outlined above and the risk control measures that you must implement, use and / or wear. Ensure you have received sufficient information, instruction and training to enable you to conduct this activity with the minimum of risk to yourself and others. Signature: Date: Page 10 of 10