USER MANUAL AUTOMATIC AND MANUALLY TRIGGERED RESUSCITATOR. O-TWO MEDICAL TECHNOLOGIES INC. Innovation in Resuscitation

Similar documents
2 PRE USE FUNCTION CHECK 3 OPERATING PROCEDURE 4 SERVICING. 5 CAREvent ALS ACCESSORIES 11

USER MANUAL (For G2005 Model)

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

Equinox II NITROUS OXIDE / OXYGEN ADMINISTRATION EQUIPMENT 01EQ7000E

O-Two Self-Study Guide. Equinox Relieve

O-Two Equinox Relieve Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System

45A, Armthorpe Road, Brampton, Ontario, Canada, L6T 5M4. Made in Canada by O-Two Medical Technologies Inc.

AUTOMATIC TRANSPORT VENTILATORS

AUTOVENT 4000 VENTILATOR

Equinox Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System USER MANUAL

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR

CPR MEDICAL DEVICES INC. OPERATING MANUAL

Operation and Maintenance of the EPV200 Portable Ventilator

O-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes

CONTROL MODULE FOR RESUSCITATOR

OXY-VIVA 3 RESUSCITATOR OPERATING INSTRUCTIONS

Virginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Hospital and Transport for Controlled Breathing

Operation and Maintenance of the MCV100 and MCV100-B Portable Ventilator

Rev. B. Operating Instructions

C O R P O R AT E B R O C H U R E

OXY-life TM. Oxygen Resuscitator OPERATING MANUAL

MACS Mask CPAP System

Emergency Transport and Ventilation

Breathing Devices. Chapter 8 KNOWLEDGE OBJECTIVES SKILL OBJECTIVES. 6. List four precautions to take when using oxygen.

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

Unit 15 Manual Resuscitators

Quick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT

Chapter 9 Airway Respirations Metabolism Oxygen Requirements Respiratory Anatomy Respiratory Anatomy Respiratory Anatomy Diaphragm

Airways and Resuscitators. CRC 330 Cardiorespiratory Care University of South Alabama

SERIES-O REGULATOR. Instructions for use

Mirage Vista Nasal Mask

Blendox AII5000F Inhalation Antalgic System. User s Manual :SP-Y Series No. Version :A1. Date of Review:

E C C. American Heart Association. Basic Life Support for Healthcare Providers. Written Examinations. March 2011

E C C. American Heart Association. BLS for Healthcare Providers. Written Exams. January 12, 2012

2017 North America Catalogue

How to use the SAVe II Ventilator

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

NeoNatalie Resuscitator Oxygen kit

MEDUMAT Standard a MEDUMAT Standard. The Ventilation Standard for Professionals

e700 Automatic Transport Ventilator User Manual e700 User Manual (15PL Rev.14 28_8_2017)

Dialflow Regulator. Instructions for Use

Oxygen Demand Valve. Instructions for Use

INTERNATIONAL CATALOGUE

Yoke Block Instruction Manual

INTERNATIONAL CATALOGUE

COALINGA STATE HOSPITAL. Effective Date: August 31, 2006

Breathing Process: Inhalation

Oxygen Dialflow Meter. Instructions for Use

Recommendations on Checking Anaesthesia Delivery Systems

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

Data Sheet... RUDOLPH, inc. HANS RUDOLPH,

Complex Care Hub Manual: Pressure Ventilation on the LTV 1200

NeoNatalie Resuscitator Oxygen kit

R S A B C CPR. Basic Life Support Flow Chart Check for danger. Check Response. Send for Help. Check Airway. Check for Breathing.

Automatic Transport Ventilator

User Operation Manual and Safety Instructions

Training. Testor Training Manual

OPERATION MANUAL NTF-60 Plus

Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators

Flow meter. bellow vaporizer. APL valve. Scavenging system

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL

Definition An uninterrupted path between the atmosphere and the alveoli

Technical Data and Specifications

Pressure Regulators. Instructions for Use

NORTH AMERICA CATALOGUE

NORTH AMERICA CATALOGUE

ASSISTANCE. Cardinal Health Pulmonetic Systems Medina Rd., Suite 100 Minneapolis, Minnesota

Provide Comfort to your Patients is our commitment PRODUCT CATALOGUE HOSPTIAL SUPPLIES. Copyright ReadMed Company Ltd

OPERATION MANUAL NTF-15

AIR HAMMERS MODEL NO: CAT138/CAT139 OPERATING & MAINTENANCE INSTRUCTIONS PART NO: / GC064

User Instruction Manual

SCK01 AIR COMPRESSOR KIT WITH QUICK FITTINGS - SPRAY/BLOW/OIL GUN, TYRE INFLATOR & HOSE

CHAPTER 3: TAKING ACTION AND CARING FOR BREATHING EMERGENCIES Multiple Choice

Simplified Automated Ventilator Training Slides

Yoke Block Instruction Manual

e600 Automatic Transport Ventilator User Manual e600 User Manual (15PL Rev.7 28_8_2017)

The Univentor 1250 Anaesthesia Unit

Invacare Twilight II Nasal Mask

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

How to Perform CPR. The CABs of CPR are Circulation, Airway and Breathing. C - Circulation (Chest Compressions)

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

SERVICE MANUAL MODEL BA020ANM-S1 BREATHING AIR PANEL

MODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE

C-1 C-7. C-6 Short air tubing Upper Velcro straps. C-a C-5 C-2 C-4 C-3 B-1 B-9 B-10 B-2 B-7 B-8 B-3 B-6 D-1 D-2 D-13 D-3 D-12 D-10 D-11 D-4 D-5 D-8

SERVICE MANUAL MODEL BA020AMST BREATHING AIR PANEL

BASIC LIFE SUPPORT - PAEDIATRIC

Basic Life Support & Automated External Defibrillation Course. OBJECTIVES

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank

Electrical Shock Survival

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

Analgesic Demand Valve. Instructions for Use

Transcription:

MANUFACTURERS AND DISTRIBUTORS OF: Administration Systems Burn Relief Burn Kits and Dressings CAREvent Range of Automatic Transport Ventilators CPR Bio-Barrier Face Shields Demand Valve Resuscitators EasyGrip Bag-Valve-Mask Resuscitators Equinox 50% Nitrous Oxide / 50% Oxygen Genesis II IDLH Rescue Resuscitators Oxygen Demand & BVM Oxygen Supply Refill Valves Rescue Breather CPR Pocket Ventilator Respiratory Disposable products SMART BAG Bag-Valve-Mask Resuscitators SMART BAG MO Bag-Valve-Mask Resuscitators SQ B Aluminum Free Unibody Regulators SQ Unibody Regulators Universal Pocket Ventilators Vent R ALS / BLS AUTOMATIC AND MANUALLY TRIGGERED RESUSCITATOR Your Representative is: EU Representative: Marcel Houben Rue Vinâve, 32 4030 LIEGE Belgique O-TWO MEDICAL TECHNOLOGIES INC. Innovation in Resuscitation 7575 Kimbel Street Mississauga, Ontario Canada L5S 1C8 Tel. (905) 677-9410 Fax (905) 677-2035 Website: www.otwo.com E-mail: resuscitation@otwo.com For your nearest Authorized O-Two Distributor In North America call Toll Free 1-800-387-3405 USER MANUAL CAREvent CAREvent CAREvent CAREvent ALS Standard ALS G05 BLS Standard BLS G05 Made in Canada by O-Two Medical Technologies Inc. Part Number 17MP9656 Rev K Sept. 2008

CAREvent BLS Handheld Resuscitator Accessories 17MP9039-MOH 01PV1121 01CV8001-Cs 01FG6500 01FG6501 01FG6502 01CV8010-Cs Reusable patient valve swivel housing (without anti-entrainment ring). Reusable diaphragm (duckbill style) Single use patient valve with bio-filter (patient valve dead space 8 ml) Mask Retaining Insert Mask Insert Extraction Tool Head Harness System Single Use Transport Ventilation Circuit (patient valve dead space 8 ml) 12

THE RESUSCITATOR MUST BE THOROUGHLY CLEANED AFTER EACH PATIENT USE. 1. Operate CAREvent ALS / CAREvent BLS Handheld Resuscitator to blow out any contaminant from the patient valve. 2. Ensure CAREvent ALS / CAREvent BLS Handheld Resuscitator is disconnected from the gas supply source. 3. Remove the patient valve swivel housing (a) from the body of the resuscitator (c), being careful to ensure that the diaphram (b) is retained (fig 3). 4. Remove the facemask from the resuscitator (after removing the mask retaining insert (if supplied) using the extraction tool). 5. Shake out any foreign material. 6. Wash all components thoroughly in a mild soap solution and disinfect as required. 7. The resuscitator can be wiped over with a soft cloth and mild soap solution. 8. Dry all components thoroughly. 9. Reassemble unit, connect to an air or oxygen supply to check operation prior to packaging for emergency use. NOTE: If the single use valve/mask combination is being used, safely dispose of these items. CHAPTER 5 CAREvent ALS Handheld Resuscitator Accessories 17MP9039 Reusable patient valve swivel housing 01PV1121 Reusable diaphragm (duckbill style) 01CV8005-Cs Single use patient valve with bio-filter and PEEP Port 01FG6500 Mask Retaining Insert 01FG6501 Mask Insert Extraction Tool 01FG6502 Head Harness System 01CV8015-Cs Deluxe Single Use Transport Ventilation Circuit with PEEP Port (patient valve dead space 8 ml) 01CV8020 Reusable Transport Ventilation Circuit with PEEP Port (patient valve dead space 8 ml) 17MP7010 Single Use PEEP Valve 11 TABLE OF CONTENTS CHAPTER TITLE PAGE 1 THE CAREvent ALS / CAREvent BLS HANDHELD AUTOMATIC AND MANUALLY TRIGGERED RESUSCITATOR 1.1 Introduction 1 1.2 Warranty Information 1 1.3 The CAREvent ALS CAREvent BLS Features 2 1.4 Performance Specifications 3 1.5 Safety Precautions 4 2 PRE USE FUNCTION CHECK 2.1 Set Up 5 2.2 Testing of the Individual Features of the Resuscitator 5 3 OPERATING PROCEDURE 4 SERVICING 3.1 Connecting the Supply Hose and Patient Circuit 6 3.2 Manual Ventilation and Cardiac Compressions 6 3.3 Automatic Ventilation 7 3.4 Action to be Taken if Patient Vomits During Resuscitation 8 3.5 Demand Breathing and Automatic Circuit Shut Off (CAREvent ALS Only) 9 4.1 Routine Maintenance 9 4.2 Cleaning and Sterilization 10 5 CAREvent ALS / CAREvent BLS ACCESSORIES 11

1.1 Introduction CHAPTER 1 The CAREvent ALS / CAREvent BLS Handheld Resuscitator provides trained individuals with a safe and effective means of providing artificial ventilation during respiratory and/or cardiac arrest. Regulator working pressure, suction (if equipped), and ventilator limiting pressures should be checked at least every six months, and more frequently in high use applications. Units with test pressures outside of the ranges listed in the product specifications should not be used. The product is not designed for field disassembly or service outside that indicated in this manual. Any malfunctioning units should be returned to the manufacturer or an Authorised Dealer. Unauthorised repairs will nullify the product warranty. The CAREvent ALS / CAREvent BLS Handheld Resuscitator is lightweight, portable, and extremely durable. Designed for the demands of the emergency medical and rescue environment, they can be operated anywhere an oxygen cylinder or piped wall outlet or breathing air supply is present. NOTE: Units with test parameters outside of their ranges listed in the product specifications, should not be used. Any units not meeting performance criteria should be returned to the Manufacturer or an authorized repair centre. NOTE: An Automatic and Manually Triggered Resuscitator is considered a critical device, and its components considered critical components. Only those individuals trained in Cardio-Pulmonary Resuscitation and the operation of oxygen-powered ventilators should use this equipment. Thoroughly review this instruction manual before use. NOTE: This resuscitator is intended for first response to a breathing emergency only and that patients must be transferred to a transport and emergency ventilator as equipment becomes available. WARNING: Resuscitator used in contaminated environments can be hazardous unless entrainment is prevented or appropriate filtration is provided. 4.2 Cleaning the CAREvent ALS / CAREvent BLS Handheld Resuscitator and Accessories Routine cleaning of the equipment should be undertaken to maintain the equipment in a clean condition. Reusable patient valve swivel housing and diaphram can be cleaned using a mild soap solution and disinfected using a legally marketed commercially available disinfectant, suitable for the application. Single use patient valves and masks should be discarded after each patient use and replaced with a new unit. All other components should be wiped clean with a mild soap solution. Under no circumstances should the complete unit be allowed to be soaked or immersed in cleaning solutions. 1.2 Warranty This equipment is manufactured from the finest quality materials. Each individual part is subject to strict quality control tests to ensure exceptionally high standards. The manufacturer warrants to the purchaser of the CAREvent ALS / CAREvent BLS Handheld Resuscitator that its component parts are free from defects in material and workmanship for a period of two years from the date of purchase. The manufacturer will replace and/or repair all parts of the resuscitator at its option for two years from the date of purchase at no cost to the purchaser, upon the notification of the defects, in writing by the purchaser. All shipping costs shall be borne by the purchaser. The manufacturer shall be liable under this warranty only if the resuscitator and its parts have been used and serviced in the normal manner described in the instruction manual. There are no other expressed or implied warranties. This warranty gives no specific legal rights. You may also have other rights that may vary according to local regulations. (g) Resuscitator Body (h) Gas Supply Inlet Disassembly of the resuscitator (fig 3.) (Shown with re-usable patient valve) 1 10 (f) Pressure Relief System (e) Viewing Window (d) Manual Button (c) Setting Selector (b) Diaphragm (a) Patient Valve Swivel-Housing

4. Rotate the selector to the appropriate position and either operate the manual button to blow out any contaminant or cycle automatically for a few breaths 5. Re-assemble diaphragm, patient valve and face mask and operate the manual button or automatically cycle the resuscitator for a few breaths to ensure correct function. 6. Restart resuscitation as previously indicated. NOTE: When using the single use patient valve, vomitus may be forced past the diaphragm and contaminate the bio- filter. This may necessitate the use of a new single use valve. 3.5 Demand Breathing and Automatic Circuit Shut Off (CAREvent ALS Only) Should the patient commence spotaneous breathing at a flowrate of greater than 30 lpm for more than 1 second the CAREvent ALS will sense the patient s inspiratory effort and will stop cycling automatically allowing the patient to Demand Breathe at their own rate and volume on 100% oxygen (if connected to an oxygen supply). If they cease spontaneous breathing the ventilator will recommense automatic cycling after a delay of 5-8 seconds (depending on the depth of the patients previous respiration) without intervention by the rescuer. NOTE: The use of a Transport Ventilator Circuit with the CAREvent ALS will increase Demand Valve Triggering Pressure due to the secondary patient valve. The use of the Transprt ventilator Circuit does not increase the patient valve deadspace. 4.1 Routine Maintenance WARNING: CHAPTER 4 SERVICING The CAREvent ALS / CAREvent BLS Handheld Resuscitator is designed to provide respiratory support in all emergency situations. Failure to follow the maintenance and inspection routines properly could result in incorrect operation of the resuscitator. To ensure proper operation of the resuscitator regular inspection and checking of the resuscitator and accessories for correct function should be undertaken by a responsible member of staff on a routine basis. This check is to ensure that all of the accessories and resuscitator components are present, the air or oxygen cylinder is full and that the resuscitator is in working order. 1.3 Features The CAREvent ALS / CAREvent BLS Handheld Resuscitator is a pneumatically powered, time/volume cycled ventilatory resuscitator with the added feature of a Manually Actuated, Automatic Ventilation Override Button (Manual Button) to allow the operator to control the ventilations manually at a rate and volume they desire. In addition the CAREvent ALS allows the breathing patient to Demand Breathe on 100% oxygen while their inspiratory efforts causes the automatic cycling to cease. Should they stop breathing, the ventilator will automatically restart cycling in the setting selected. The pneumatic logic circuit can be run on either approved, compressed, breathing air or medical oxygen. The unit is self contained and only requires its attachment to a regulated oxygen or air supply (as specified) for immediate use. The CAREvent ALS / CAREvent BLS Handheld Resuscitator:. Delivers 100% oxygen during resuscitation (when attached to an oxygen source).. Meets American Heart Association/J.A.M.A. recommendations for C.P.R.. Provides physiologically normal respiratory rates and volumes.. Has an Audible Airway Pressure Limiting System set in accordance with the recommendations of the American Heart Association/ JAMA.. Is lightweight and extremely durable.. Is designed for emergency resuscitation and pre-hospital care / patient transport and for resuscitation and inter-departmental transport in the hospital environment where the potential patient use is with children and adults.. Has a Manually Actuated, Automatic Ventilation Override Button (Manual Button).. Has a seven position selector providing six, preset, automatic settings for a range of patient sizes from children to adults with tidal volumes and frequencies of ventilation in line with established guidelines for the range of patient sizes indicated and an OFF position to allow the ventilator to be left in a standby mode with the oxygen supply turned on.. CAREvent ALS Provides Demand Breathing with automatic cycling shut off and re-start. Guidelines 2005 compliance. The G05 model complies with the requirements of the ERC and AHA for 30:2 compression:ventilation ratio. 9 2

1.4 Performance Specifications (All specifications are subject to a tolerance of +/- 10% except the I:E Ratio which is subject to a tolerance of +/- 20% and Maximum Airway Pressure +0/-10%) TIDAL VOLUME: 150-600 ml (G05) 200-1100 ml (Standard) BREATHS PER MINUTE: 20-10 (G05) 20-12 (Standard) I:E RATIO: 1 : 2 AUTOMATIC FLOW RATE: MANUAL FLOW RATE: MANUAL OVERIDE DELAY TIME: DEMAND BREATHING FLOWRATE (CAREvent ALS Only): 9.0-18.0 L/min (G05) 12.0-39.6 L/min (Standard) As per automatic setting 17 ~ 23 Sec (G05) 5 ~ 8 Sec (Standard) 0-120 L/min DEMAND BREATHING TRIGGERING PRESSURE (CAREvent ALS Only): -2.0 cmh 2 0-2.0 mbar INSPIRATORY RESISTANCE: -6.0 cmh 2 0-5.9 mbar EXPIRATORY RESISTANCE: 6.0 cmh 2 0 5.9 mbar AUTO SHUT OFF DELAY TIME: 5-8 Sec (CAREvent ALS Only) INPUT PRESSURE: 45-70 PSI (3.0-5.0 Bar) MAXIMUM AIRWAY PRESSURE: 60 cmh 2 0 (58,8 mbar) OPERATING TEMPERATURE: -18 to + 50 o C (0 to+ 122 o F) tidal volume. This is accomplished by moving the control toward the child setting if the blow off valve operates or towards the adult setting if chest rise is insufficient. This can be done by simply using your thumb to slide the control without removing your hand from the mask. WARNING: Automatic ventilation of the patient who is intubated or whose mask is held in place with the optional head harness system, does not mean that the patient is safe to be left unattended and constant observation of the patient s pulse and chest movement must be continued. WARNING: The use of gas pressure regulators that do not maintain a minimum output pressure and flowrate in line with the requirements of the specification may cause the device to fail resulting in the patient not being ventilated. Control Position 1 2 3 4 5 6 Tidal Volume Vt (ml) 150 200 300 400 500 600 * 200 300 400 600 800 1100 Frequency (BPM) 20 20 10 10 10 10 * 20 15 15 12 12 12 Minute Volume Vm 3.0 4.0 3.0 4.0 5.0 6.0 (ltrs) * 4.0 4.5 6.0 7.2 9.6 13.2 Automatic flowrate 9.0 12.0 9.0 12.0 15.0 18.0 (L/min) * 12.0 13.5 18.0 21.6 28.8 39.6 Body Weight (kg) min. 21 28 42 57 71 85 (6-7 ml/kg) max. 25 34 50 67 84 100 Automatic Adjustable Setting Selections (fig 2) 3.4 Action to be taken if patient vomits during resuscitation Should the patient vomit into the mask during resuscitation the following steps should be followed to clear the foreign material: STORAGE TEMPERATURE: - 40 to + 60 o C (- 40 to + 140 o F) INPUT CONNECTION: PATIENT CONNECTION: WEIGHT(approx) 9/16 DISS 15 / 22 mm 16 OZ (0.45 Kg)(ALS) 14 OZ (0.40 Kg)(BLS) 1. Remove the mask from the patient s face and clear any foreign material from the patients airway. Depress the manual button or allow the resuscitator to cycle automatically for a few breaths to clear the mask and valve of foreign material. 2. If depressing the manual button repeatedly or automatically cycling the resuscitator does not clear the foreign material from the patient valve,either disconnect the resuscitator from the gas supply or turn the selector to the OFF position, remove the facemask and unscrew patient valve swivel housing (a) from the resuscitator body being careful to ensure that the diaphragm (b) is retained (Fig.3). SIZE: 140 x 63 x 73 mm 5.5 x 2.5 x 2.9 CYLINDER DURATION: (a) CAREvent ALS / CAREvent BLS: 31 ~ 103 minutes (b) CAREvent ALS / CAREvent BLS G05: 69 ~ 138 minutes (Based on an Aluminum D size cylinder containing 415 Litres of oxygen.) 3 8 3. Shake out any foreign material from the resuscitator, diaphragm, face mask and patient valve swivel housing.

cardiac output without the danger of gastric distension.) The flowrate provided is equivalent to the setting on the automatic setting selector providing flowrates equivalent to the size of patient being ventilated, thereby reducing airway pressures and the risk of aspiration of stomach contents still further. 1. If no respiratory effort is observed, position yourself above the patient s head and apply the face mask over the patients mouth and nose. The thumb and index fingers are used to hold the mask to the face while the remaining three fingers of each hand are placed along the angle of the jaw. A tilt action is used to hyperextend the neck and move the jaw forward. This helps displace the tongue away from the back of the throat and maintains an open airway. 2. Select the tidal volume/frequency of ventilation for the size of patient being resuscitated. Depress the manual button and observe the rise of the patient s chest. Release the button when chest rise is adequate. 3. If the patient s chest does not rise or gas escapes around the mask or the blow off valve (f) (Fig.3) operates, reposition the patients head and adjust your hand position to obtain an effective mask seal and an open airway. 4. Monitor the patient s skin, nailbed and lip colour. 5. If mask indicates signs of vomitus, remove immediately and clear the airway. Restart ventilation immediately after clearing airway. 6. Continue ventilation at an appropriate rate until relieved or until spontaneous breathing returns. 3.3 Automatic Ventilation. 1. If you have been ventilating manually simply release the manual button and after a short pause (5-8 seconds) (17-23 seconds in the G05 model), the ventilator will commence automatic cycling at the rate and volume selected. If you are commencing automatic ventilation immediately, rotate the setting selector to the setting appropriate for the size of patient being ventilated and the ventilator will commence automatic cycling (fig 1). 1.5 Safety Precautions The CAREvent ALS / CAREvent BLS Handheld Resuscitator is designed to provide emergency ventilatory support to patients suffering from respiratory and/or cardiac arrest. The CAREvent ALS / CAREvent BLS Handheld Resuscitator is intended for use by suitably trained and qualified personnel. The following precautions should always be observed: 1. DO NOT LEAVE THE PATIENT UNATTENDED. 2. WHEN NOT IN USE, ALWAYS TURN OFF THE CYLINDER. 3. NEVER ALLOW OIL OR GREASE TO COME INTO CONTACT WITH ANY PART OF THE CYLINDER, REGULATOR OR RESUSCITATOR. 4. DO NOT DISASSEMBLE ANY PART OF THE RESUSCITATOR EXCEPT WHERE DESCRIBED IN THIS MANUAL, AS ANY UNAUTHORIZED DISASSEMBLY WILL INVALIDATE THE WARRANTY. 5. AFTER USE, ALWAYS ENSURE THAT ALL COMPONENTS ARE CLEANED IN ACCORDANCE WITH THE INSTRUCTIONS PROVIDED IN THIS MANUAL. 6. ENSURE THAT ALL COMPONENTS ARE REASSEMBLED CORRECTLY AND THAT ALL ITEMS ARE REPLACED IN THE CARRYING CASE. 7. AFTER USE, ALWAYS ENSURE THAT A FULL AIR OR OXYGEN CYLINDER IS ATTACHED BEFORE RETURNING THE UNIT TO ITS NORMAL STORAGE POSITION. 8. ENSURE THAT A NEW SEALING WASHER IS USED EVERY TIME YOU ATTACH THE REGULATOR TO THE CYLINDER. 9. IT IS RECOMMENDED THAT AN ALTERNATIVE MEANS OF VENTILATING THE PATIENT BE AVAILABLE IN CASE OF GAS SUPPLY FAILURE. 2. Closely observe the patient s chest movements. If there is any leak from around the mask or any obstruction in the patients airway (blow off valve will operate) reposition patients head and adjust mask and hand position to ensure a good airway and mask to face seal. 3. Should repositioning the mask and adjusting hand/neck position not resolve the situation adjust the automatic selector control to establish the correct 7 4

2.1 Set Up Along with the contents of the shipping cartons you will require the following items to enable you to undertake the pre-use functional check: [1] Full oxygen cylinder CHAPTER 2 PRE USE FUNCTION CHECK CHAPTER 3 OPERATING PROCEDURE 3.1 Connecting the Supply Hose. The supply hose provided is attached to the oxygen inlet on the rear of the resuscitator and is tightened finger tight (fig 1). [2] Oxygen regulator with a 60 PSI 9/16 DISS outlet. The regulator must be able to output a minimum of 120 L/min at no less than 45 PSI ( 3.1 Bar). WARNING: The use of excessive force in tightening the supply hose may damage the seal and /or thread. Having connected the supply hose to the regulator, ensure that the Setting Selector is in the OFF position and turn on the oxygen supply. Using a mild soap solution, spray the input connection to the resuscitator to check for leaks. If any leak is present, tighten the connection and re-test. Once no leaks are found, connect the Test Lung to the 15/22mm patient connector on the resuscitator. Using the Sliding Setting Selector, select the first automatic setting. The facemask is attached to the patient connection port by simply pushing the mask onto the 22mm taper. Manual Button 2.2 Testing of the Individual Features of the Ventilator. The following features can be individually tested during the pre-use Functional Check: [1] Maximum Airway Pressure and Audible Alarm. [2] Frequency/Tidal Volume Adjustment Attach supply hose here Setting Selector [3] Manual Ventilation [4] Demand Valve Function and Automatic Circuit Shut Off (CAREvent ALS Only) Connecting the supply hose (fig 1) 3.2 Manual Ventilation and Cardiac Compressions. The CAREvent ALS / CAREvent BLS Handheld Resuscitator has a Manually Actuated, Automatic Ventilation Override Button (Manual Button) to assist in the timing of ventilations in conjunction with external cardiac compressions. By using the Manual Button, the operation of the ventilator can be easily timed with the chest compressions so as to avoid the potential problem of the aspiration of stomach contents due to gastric distension which may occur if overlap of chest compression and inflation occurs. (It has been shown in some studies that, in patients that are intubated, this overlap of compression and inflation may increase 5 6