Firesafe Flowmeter. Instructions for Use

Similar documents
Micro Dial-Flowmeter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use

Pressure Regulator. Instructions for Use

Dialflow Regulator. Instructions for Use

Pressure Regulators. Instructions for Use

Oxygen Demand Valve. Instructions for Use

Analgesic Demand Valve. Instructions for Use

Domiciliary Oxygen Switch. Instructions for Use

Analgesic and Oxygen Demand Valve. Service Manual

SERIES-O REGULATOR. Instructions for use

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

Rev. B. Operating Instructions

BOC: Living healthcare. Manual. LIV IQ BOC Integrated Valve with digital display portable delivery system for Medical Oxygen. BOC: Living healthcare

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

AUTOMATIC GAS MANIFOLDS

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

MANUAL GAS MANIFOLDS

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

User manual. ZX trolley.

Home Oxygen Products. > Innovation > Quality > Safety

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A

Medical oxygen. LIV IQ Digital therapy device.

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS

AUTOVENT 4000 VENTILATOR

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

Peninsula Dental Social Enterprise (PDSE)

Service Manual. Regulators for Medical Gases

MEDAP-MOBILE GAS SUPPLY

BRITISH COMPRESSED GASES ASSOCIATION Registered office: 4a Mallard Way, Pride Park, Derby, UK. DE24 8GX Company Number: 71798, England

SEMEM3-73 Maintaining medical gas pipeline systems and equipment

AND OPERATION INSTRUCTIONS

Instructions for use Evacuation system compressed air injector SL 20 / SL 20 Max SL 30 / SL 30 Max

Instructions for use Compressed air injector with water trap

PRESSURE REGULATORS AND FLOWMETERS, ALL MEDICAL GASES

SAFETY PRINCIPLES FOR PRESSURE REGULATORS FOR MEDICAL OXYGEN CYLINDERS

PRESSURE REDUCTION FLOWMETERING, QUICK COUPLING PROBES AND ACCESSORIES

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

Operating Manual. R280 Pressure regulator made of brass. 1. Intended Use

Service Manual. Dial Flow Regulators

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

AREA VALVE SERVICE UNITS

Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions

Presence PR, PRC & PRE

- GAS REGULATOR - COMPACT STYLE

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

Instruction Manual for Configura Cushionair Portable Pump

45A, Armthorpe Road, Brampton, Ontario, Canada, L6T 5M4. Made in Canada by O-Two Medical Technologies Inc.

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

Budget Range Operators Handbook

Before using or working with the Grab n Go Heliox system, please read and understand the usage and safety instructions provided in this booklet.

Flow meter. bellow vaporizer. APL valve. Scavenging system

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

Recommendations on Checking Anaesthesia Delivery Systems

ISO INTERNATIONAL STANDARD. Flow-metering devices for connection to terminal units of medical gas pipeline systems

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS

CPR MEDICAL DEVICES INC. OPERATING MANUAL

Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

O-Two Self-Study Guide. Equinox Relieve

Training. Testor Training Manual

Working Environment Safety Assessment

FLOWMETER UNITS FLOWMETER DEVICES, BY-PASS SYSTEMS

Instructions for Use

Manifolds & pressure reducing sets

Allspeeds Ltd. Royal Works, Atlas St Clayton le Moors Accrington Lancashire England BB5 5LW. Tel +44 (0)

I. Subject: Compressed gas cylinders, regulators, and portable liquid systems.

Operating Instructions Part No

Testing anaesthetic machines, ventilators and related equipment

Rejuvenation Instructions

Pre-filter for Brushless Blower Instruction Manual

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI

PRODUCT INFORMATION AND INSTRUCTIONS OM-810

WW-720. Pressure Reducing Control Valve

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS

U S E R M A N U A L. (CGA 870 Connection)

OXYGEN THERAPY FLOWMETERS, REUSABLE AND SINGLE-PATIENT HUMIDIFIERS, PRESSURE REGULATORS, FLOW SELECTORS

Pressure Regulators. Operating Instructions. Instrumentation

BS ml Bottle Sampling Unit

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters

REL-46 WARNING NOTICE 15 TON SINGLE ACTING REMOTE HYDRAULIC CRIMPING HEAD. Compatible with RELIABLE R15 and P Style dies. REL-46 Manual

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

Welcome to the Specialized Medical Services Respiratory training webinar series!

LUBRICATOR ASSEMBLY AND OPERATING INSTRUCTIONS

USER MANUAL. AIR-OXYGEN BLENDER (DISS and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS

Installation & Maintenance Manual. C2 & C3 Manifolds

FILTER REGULATORS MODEL NO: CAT155 & CAT156 FITTING & MAINTENANCE INSTRUCTIONS PART NO: & ORIGINAL INSTRUCTIONS

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified

Instruction manual. for. high purity gas pigtails 200 bar / 300 bar

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

Transcription:

Firesafe Flowmeter Instructions for Use 702-0049.16 May 2014

Contents 1. Symbols... 2 2. Warnings and Cautions... 2 2.1. Warnings!... 2 2.2. Cautions!... 3 3. Functional Description... 4 3.1. Intended Use... 4 3.2. Technical Description... 4 4. Operating Instructions... 6 4.1. Connecting to Gas Supply... 6 4.2. Adjusting the Flow... 6 4.3. Reading the Flowmeter... 6 4.4. Connecting to Patient... 6 4.5. During Use... 7 4.6. After Use... 7 5. Cleaning and Disinfection... 7 5.1. Cleaning... 7 5.2. Suspected Contamination... 7 6. Maintenance... 7 6.1. Interim Inspection... 7 6.2. Service and Repair... 7 7. Specification... 8 8. Troubleshooting... 9 9. Parts and Spares... 10 10. Distributor Details... 11

1. Symbols Warning! Caution! Note Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user or others Indicates a potentially hazardous situation which, if not avoided, could result in damage to the equipment or property Highlights points that might allow more convenient or efficient operation of the equipment Attention, consult accompanying documents Use no oil 0120 CE Marked to Medical Device Directive 93/42/EEC as amended by directive 2007/47/CE No smoking Flow should be read level with the top of the float 1304X12345 YYMM Date of manufacture identification 2. Warnings and Cautions 2.1. Warnings! Read through this entire instruction manual before using or showing others how to use this flowmeter. As with all medical equipment, attempting to use this device without a thorough understanding of its operation may result in patient or user injury. Medical gases are, or should be considered a drug and should only be used for medical purposes as prescribed by a physician or authorised clinician and in accordance with their instructions. Oxygen therapy may be a critical treatment. Additional risk control measures such as SpO 2 monitoring with an alarm and a backup oxygen supply with delivery device should be provided for patients who might suffer distress or injury if their oxygen supply is not available or is stopped unexpectedly. The flowmeter is not MRI compatible. Ensure that the medical gas supply is sufficient for the proposed therapy and is supplied within the pressure range given in the Device Specification. If the supply is a gas cylinder, check the cylinder contents gauge regularly. This flowmeter is only for use with the medical gas indicated on the device label. Check the cylinder or supply is the correct medical grade gas before use. Always check that the flowmeter is connected to the correct gas specific terminal unit. 2

Gas specific connectors are fitted to the flowmeter. Do not attempt to modify the fittings to suit other gases or fitting systems. Oxygen is not flammable; however the presence of it will drastically increase the rate and severity of combustion. Oil and/or grease in the presence of an oxygen enriched atmosphere will become highly combustible. Oxygen must never be allowed to come into contact with oil, grease or other petroleum-based substances. Do not use oil or grease on this flowmeter. Do not smoke around oxygen equipment. Only appropriately trained personnel working in controlled conditions may disassemble or assemble this flowmeter. Do not disassemble the flowmeter whilst under pressure. Secure cylinders to a wall, stand or cart in accordance with local safety regulations. Do not stand in front of a flowmeter or regulator outlet when opening a cylinder valve. Do not allow any fluid to enter this device. The accuracy of the flowmeter can be affected significantly if the input pressure is other than stated in the technical specification and the device label. Do not directly connect the flowmeter output to another pressurised gas source as this may lead to corruption of one or other of the gas supplies (e.g. connecting an oxygen flowmeter to an air flowmeter via a Y-piece in an attempt to blend their outputs). Only use this flowmeter with the flow tube in a vertical position. Failure to do so may result in incorrect flow rates being displayed. Arrange the oxygen tubing and gas hose carefully to avoid damage to the hose and the potential for causing a trip hazard. Never pull or apply excessive force to the gas hose. A leaking hose may result in high local oxygen and nitrous oxide concentrations and an increased risk of fire. Gas flow may be prevented if the supply tubing is kinked or occluded. Care should be taken to ensure that the flow control valve is free from obstruction and unintended pressure. These scenarios may result in temporary changes in flow. 2.2. Cautions! The performance of the flowmeter may be affected if it is stored or transported in temperatures outside of the range -20 C to 60 C (-4 F to 140 F). The flowmeter is not suitable for autoclaving. Where a threaded fastener is used, do not use the device to tighten or lock connection. The performance of the flowmeter may be affected if the flow control valve is over-tightened when switching off flow. Apply the minimum force required to switch off gas flow. 3

3. Functional Description Scale Flow Tube Precision Rotating Float Gas Specific Inlet Connector Flow Control Valve Firesafe Nozzle 3.1. Intended Use The Firesafe Flowmeter is used to precisely control the flow of medical oxygen or medical air (model dependant) from a low pressure supply. Inlet pressure ranges can be found in the Technical Specification. Medical gases are, or should be considered, drugs and should only be used for medical purposes on the authority of a physician or authorised clinician and then strictly in accordance with their instructions. This product is designed for use in homecare applications, in hospitals, other managed clinical environments and by the emergency services. 3.2. Technical Description The Firesafe enables accurate control of the flow of either medical oxygen or medical air (model dependent) in the range of 0.5 to 15 l/min. A precision rotating float rises inside a transparent graduated flow tube. The flow tube has a scale printed upon it that indicates the rate of gas flow in litres per minute (l/min). The reading should be taken at the top of the float. The flow of gas through the tube is controlled by a valve downstream of the flow tube. BPR Medical Ltd products enhance patient and user safety and as such the Firesafe Flowmeter includes many unique features to maximise safety and minimise exposure to risk. These features include: 4

3.2.1. Firesafe Nozzle Every Firesafe Flowmeter is fitted with a Firesafe Nozzle, minimising the known risks of fire associated with oxygen therapy. The Firesafe Nozzle is discernible from a standard tubing connector as it features a bright orange band. The Firesafe Nozzle can limit the impact of an oxygen fire by automatically arresting the oxygen flow. Acting like a fuse, it automatically isolates the oxygen supply as a fire tracks back along the tubing towards the oxygen source. Activation of the Firesafe Nozzle during a fire requires is highly reliable at oxygen flows above 1 l/min. The Firesafe Nozzle is not suitable for use as a fire risk control measure outside this range. 3.2.2. Flowsafe Valve The Flowsafe Valve is another feature unique to the Firesafe Flowmeter that further lowers the risk of fire associated with oxygen use. In the event of the flow tube being inadvertently sheared from the flowmeter body, it automatically isolates the oxygen flow, preventing a free flow of oxygen into the hospital ward. The pipeline pressure is therefore maintained, negating the need to isolate the ward oxygen supply, upon which other patients may be dependent. 3.2.3. Precision Rotating Float Unlike conventional ball and glass style flowmeters, the Firesafe Flowmeter incorporates a precision float with a fluted rim, which continuously rotates in the gas stream to provide visual confirmation that gas is flowing and the device is providing an accurate reading. Historically, incidents have occurred with conventional designs whereby static charge within the flow tube has caused the ball to stick to the surface and provide a false reading. Unlike conventional ball style flowmeters, where there is often doubt as to where a reading should be taken, the Precision Rotating Float has a red band indicating the position at which the flow reading should be taken. 3.2.4. Over Pressure Valve The over-pressure (pressure relief) valve further enhances patient and user safety in the event of the source pressure rising to a dangerous level e.g. source equipment malfunction, such as a creeping pressure regulator. 3.2.5. Static Control (Triboelectric Charge Dissipation) The Firesafe Flowmeter incorporates two separate static control measures in order to rapidly dissipate any charge generated during use. These unique measures control the recognised phenomena of static charge, which is known to cause the ball in conventional flowmeters to stick to the flow tube inner surface, leading to either an inaccurate reading or indicating a flow exists when the valve is closed. 3.2.6. Flush Flow Mode The flowmeter has a Flush Flow Mode. This mode may be used to deliver approximately 25-30 l/min to supply a resuscitation bag. 3.2.7. Inlet Filter The flowmeter is fitted with a 25 µm filter to ensure that large particles of debris that might be resident in the gas supply do not contaminate the flowmeter or pass through to the patient. 3.2.8. Gas Specific Connector The gas specific quick connection probe has characteristics to prevent connections between different gas services. 5

4. Operating Instructions 4.1. Connecting to Gas Supply Before use, visually check both the flowmeter and hose (if fitted) for any damage or contamination. Do not connect or use the device if there are any doubts about its condition. The flowmeter is supplied with a gas specific connector that is designed to connect to a mating gas specific outlet. Gas specific outlets might be a terminal unit in a medical gas pipeline system or part of a pressure regulator outlet on a gas cylinder. If you are using a cylinder supply, ensure that the cylinder contents are adequate for planned therapy and turn on the supply at the cylinder. Connect the gas specific quick connect to the appropriate gas specific outlet. For indirect connection flowmeters the Medirail bracket should be fastened to a suitable medical rail system (compliant with BS EN ISO 19054 Rail systems for supporting medical equipment) using the thumb screw in the Medirail bracket. Warning! Where the gas specific connector is dependent on a threaded fastener (e.g. DISS CGA V5 1240, AS 2896/SIS handwheel) offer device to outlet and connect a few turns. Align device to the final vertical position and fully hand tighten the connection before turning on the supply pressure. Do not use the device to tighten or lock the connection. For quick connector probes (e.g. BS 5682, DIN, AFNOR), ensure that the connection is correctly made by gently pulling the flowmeter body or hose (as applicable) before turning on the supply pressure. Connect a sufficient length of patient tubing (not supplied) to the outlet barb. Note: Removal of the Firesafe Nozzle when fitting a humidifier will temporarily disable the oxygen arresting capability of the flowmeter in the event of a cannula fire. Always remember to refit the Firesafe Nozzle (indicated by a distinctive orange band) when the humidifier is no longer needed. 4.2. Adjusting the Flow Turn the flow control valve anti-clockwise to turn on and increase the flow of gas. If the flowmeter does not operate correctly, remove it from use and refer to the trouble shooting guide in Section 8 of this booklet. 4.3. Reading the Flowmeter The diagram opposite shows the correct reading level in relation to the float position for both float type and ball type variable area flowmeters. Reading Level The right hand illustration indicates the correct reading level for this Firesafe Flowmeter i.e. at the top of the float. 4.4. Connecting to Patient Connect the other end of the oxygen tubing to the patient or patient enclosure using the appropriate connector (not supplied). Float Ball 6

4.5. During Use Continue to monitor the level of gas flow during the procedure or therapy and the contents level of the cylinder (if applicable) and be aware that the supply tubing and hose (if applicable) may be a trip hazard. 4.6. After Use If the supply is from a gas cylinder, close the cylinder valve and disconnect the flowmeter when therapy is no longer required. 5. Cleaning and Disinfection Ensure the flowmeter is disconnected from the gas supply before attempting to clean it. Caution! The flowmeter is not suitable for autoclaving. 5.1. Cleaning Wipe over the outside of the flowmeter and the gas supply hose with an alcohol or disinfecting wipe. 5.2. Suspected Contamination If you suspect that the flowmeter is contaminated, remove it from use and refer the device to the appropriate department. Never immerse the flowmeter in any fluid or attempt to clean internal parts. 6. Maintenance 6.1. Interim Inspection The flowmeter should be regularly cleaned, inspected for damage and checked for performance. The frequency of inspection and performance checks depends upon usage. As a guideline, if the flowmeter is used daily this may need to be performed every six months; if used infrequently an annual check may suffice. 6.1.1. Leak Test Attach the flowmeter to a low pressure medical gas supply with a nominal supply pressure as specified on the rating label of the device and close the flow control valve. Connect supply tubing to the outlet barb and immerse the other end of the tube in water, the presence of gas bubbles in the water indicates a leak. Remove the tube and apply an oxygen compatible leak test solution to all outlets and fittings and check for leaks. A flowmeter that fails these tests should be removed from use. 6.1.2. Flow Test Verify flow rates at all flow settings against specification i.e. within ±10% at 1 l/min and above. 6.2. Service and Repair Direct connection flowmeters do not require routine servicing but can be repaired. The device can be returned to a recognised BPR Medical Service Centre, details of which can be found at www.bprmedical.com. Indirect connection flowmeters should have the hose assembly replaced after four years use or five years from the date of manufacture. The manufacture date can be found on the device label. Device life can be calculated by adding 7 years to the date of manufacture. 7

Disassembly, repair or reassembly must be carried out by a suitably qualified person working in a controlled environment. 7. Specification Specification Flow Scale Inlet Pressure Range Supply Pressure Range Accuracy Over Pressure Valve Set Point Gas Compatibility Maximum Flow (Flush Flow Mode) Firesafe Nozzle Flow Range Value 0 15 l/min Model specific between 370 550 kpa Note: The flowmeter is able to operate safely within the pressure range 370 kpa to 550 kpa, but the flowmeter will only deliver accurately when connected to an input pressure as shown on the device. Any input pressure other than shown on the device will result in flow rates different from the flow settings Minimum 370 kpa Maximum 550 kpa Nominal Pressure: ±10 % of indicated reading between 1 15 l/min Valve Settling: < 2 % from initial setting within 3 hours Temperature: 3 % flow increase/decrease per 5 C temperature increase/decrease Inlet Pressure: 4 % increase/decrease per 10 kpa increase/decrease from nominal pressure Outlet Flow Resistance: Negligible up to 230 kpa gauge back pressure Nominally 700 kpa (550 kpa minimum to 800 kpa maximum) Medical Oxygen or Medical Air (model dependant) 25 30 l/min 1 25 l/min Inlet Filtration 25 µm Environmental Regulatory Transport and Storage Temperature: -20 C to 60 C (-4 F to 140 F) Operating Temperature: 5 C to 40 C (41 F to 104 F) Humidity: 0-100% RH non-condensing CE: Medical Device Directive 93/42/EEC Active Medical Device Class IIa FDA: Class I Applied Standards BS EN ISO 5359 BS EN ISO 14971 BS EN ISO 15001 BS EN ISO 15002 BS EN ISO 15223-1 BS EN ISO 19054 BS 5682 Low pressure hose assemblies for use with medical gases Medical devices. Application of risk management to medical devices Anaesthetic and respiratory equipment. Compatibility with oxygen Flow-metering devices for connection to terminal units of medical gas pipeline systems Medical devices. Symbols to be used with medical device labels, labelling and information to be supplied. General requirements Rail systems for supporting medical equipment Probes (quick connectors) for use with medical gas pipeline systems 8

DIN 13260-2 AS 2902 Supply systems for medical gases. Part 2: Dimensions and allocation of probes and gas specific connection points for terminal units for compressed medical gases and vacuum Medical gas systems. Low pressure flexible hose assemblies 8. Troubleshooting Fault Possible Cause Solution No gas flow Flowmeter is not connected properly. Gas cylinder empty. Medical gas terminal unit on a pipeline system is isolated. Inlet filter blocked. Flowsafe Valve has activated. Firesafe Nozzle has activated due to fire or mechanical damage. Check gas supply. Check that the gas specific probe is correctly connected. Replace gas cylinder. Seek advice from someone authorised to operate the medical gas pipeline system isolation valves. Repair required. Flowsafe Valve is designed to activate when an exceptionally high flow of gas passes through the device. Disconnect from the gas supply and check the flow meter for damage or rupture. Replace Firesafe Nozzle. Audible gas leak Seal is worn or damaged. Repair required. Over Pressure Valve leaking due to input pressure being too high. Flow tube cracked or damaged. Probe nose worn or damaged. Medical gas hose damaged. Check gas supply pressure is within specification. Repair required. Repair required. Replacement required. Constant gas flow Valve or valve seat damaged. Repair required. 9

9. Parts and Spares Part Number* Description Calibration Pressure 829-0101 Firesafe Flowmeter - O 2 Single Direct 0 15 l/min - DIN 13260 connector 500 kpa 829-0301 Firesafe Flowmeter - O 2 Single Direct 0 15 l/min - BS 5682 connector 400 kpa 829-0302 Firesafe Flowmeter - O 2 Dual Direct 0 15 l/min - BS 5682 connector 400 kpa 829-0303 829-0304 Firesafe Flowmeter - O 2 Single Indirect (1.5m) 0 15 l/min - BS 5682 connector Firesafe Flowmeter - O 2 Dual Indirect (1.5m) 0 15 l/min - BS 5682 connector 400 kpa 400 kpa 829-0351 Firesafe Flowmeter - Air Single Direct 0 15 l/min - BS 5682 connector 400 kpa 829-0352 Firesafe Flowmeter - Air Dual Direct 0 15 l/min - BS 5682 connector 400 kpa 829-0353 Firesafe Flowmeter - O 2 Dual Indirect (3m) 0 15 l/min - BS 5682 connector 400 kpa 829-0901 Firesafe Flowmeter - O 2 Single Direct 0 15 l/min - SIS connector (AS 2896) 400 kpa 829-0902 Firesafe Flowmeter - O 2 Dual Direct 0 15 l/min - SIS connector (AS 2896) 400 kpa 829-0951 Firesafe Flowmeter - Air Single Direct 0 15 l/min - SIS connector (AS 2896) 400 kpa 829-0952 Firesafe Flowmeter - Air Dual Direct 0 15 l/min - SIS connector (AS 2896) 400 kpa 829-1001 829-1002 829-1051 829-1052 Firesafe Flowmeter - O 2 Single Direct 0 15 l/min - BS 5682 connector (New Zealand) Firesafe Flowmeter - O 2 Dual Direct 0 15 l/min - BS 5682 connector (New Zealand) Firesafe Flowmeter - Air Single Direct 0 15 l/min - BS 5682 connector (New Zealand) Firesafe Flowmeter - Air Dual Direct 0 15 l/min - BS 5682 connector (New Zealand) 400 kpa 400 kpa 400 kpa 400 kpa *All listed part numbers are standard configurations; please contact your BPR Medical distributor for others Spare Parts 827-0031 Firesafe Nozzle (10 pack) 9/16 UNF (DISS) Female 10

10. Distributor Details United Kingdom BOC Healthcare Customer Service Centre Priestly Road Worsley Manchester M28 2UT T: +44 (0)8456 447 764 F: +44 (0)8450 870 062 www.bochealthcare.co.uk Australia BOC Healthcare Customer Service Centre 10 Julius Avenue North Ryde Sydney NSW, 2113 T: +61 (0)2 8874 4400 F: +61 (0)2 9886 9000 www.boc.com.au New Zealand BOC Healthcare Customer Service Centre 970-992 Great South Road Penrose 1061 Auckland New Zealand T: +64 (0)9 525 5600 www.boc.co.nz 11

12

13

BPR Medical Limited T: +44 (0)1623 628 281 22 Hamilton Way, Mansfield F: +44 (0)1623 628 289 Nottinghamshire NG18 5BU, United Kingdom cs@bprmedical.com www.bprmedical.com