KOBOLD KSM FLOWMETER. User Instructions

Similar documents
GILMONT ACCUCAL FLOWMETERS

Variable Area Flow Meters. VS Series

KDV HIGH ACCURACY GLASS TUBE ROTAMETER

KDV HIGH ACCURACY GLASS TUBE ROTAMETER

! WARNING. Model PFC-1-G (direct acting) PFC-1-GR (reverse acting) Modulating Pneumatic Liquid Level Controls INSTRUCTION MANUAL MM-110B

HAYWARD FLOW CONTROL FLV SERIES, PVC/CPVC FILTER VESSEL INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

GM Series Dual-Block Multi-Function Gas Control Valves

Installation, Operation, and Maintenance Manual

Installation Instructions and valve Maintenance Rev. 0 of 21/09/15

HAYWARD FLOW CONTROL Series PBV Back Pressure Valve and Series RPV Pressure Relief Valve INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS

Variable Area Flow Meters

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI

NC MAGNETIC FLOAT SWITCHES FOR LIQUIDS

Tri-Tech Medical Inc.

Installation and Operating Instructions Bypass Flow Meter DST

V46 Series 2-Way Pressure-Actuated Water-Regulating Valves With Union Fittings

RC 195 Receiver-Controller

Installation Operating Instructions for Duplex Manual Manifolds PX-TMD Series

HAYWARD FLOW CONTROL TFS SERIES FLOW METER INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

Engineering Data Sheet

NGP-250/500 Nitrogen Generator Quick Start Guide

FlashGARD HP Reverse Osmosis Filtration System Model Number FSTMO75 Part Number

INSTALLATION, OPERATION AND MAINTENANCE GUIDE

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Operating Procedures for GripTight 15.5 SDR & IPS Test Plugs

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

Instrument Operating Manual

OFFICINE OROBICHE S.p.A. 1/9

Magnetic level switch type MR783 Instruction Manual

Operating instructions for variable area flow meter with needle valve. Model: UVR/UTR

Types 749B and R130 Changeover Manifolds

Float Operated Level Controllers

V146 Series 2-Way Pressure-Actuated Water-Regulating Valves

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

V48 Series 3-Way Pressure-Actuated Water-Regulating Valves With Union Fittings

NC MAGNETIC FLOAT SWITCHES FOR LIQUIDS

KTM OM-2 SPLIT BODY FLOATING BALL VALVES INSTALLATION AND MAINTENANCE INSTRUCTIONS

MODEL 200 KNIFE GATE VALVES INSTALLATION & MAINTENANCE MANUAL

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual

Automatic Non-freeze Valve NF6

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May

WATER HEATER THERMAL EXPANSION TANKS Owner s Manual. Safety Instructions Installation Maintenance Warranty. Models: 2-5 Gallon Capacity

Flowmeter. Original operating manual DFM

Model Series 62 Constant Differential Relay

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

Series SFD. Clean Air Filter. Variations. Made to Order. Type. Cartridge (replaceable element) Disposable (non-replaceable element) Up to 100

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Welker Magnetic Indicator (With Gravity Mixer) Model CP2GM-HP

Type 1367 High-Pressure Instrument Supply System with Overpressure Protection

Instruction Manual - Diaframless TM Ejector Chlorine, Sulfur Dioxide and Ammonia

2.0 INSTALLATION & SERVICE

RHPS Series RD(H)F40 User Manual. Read the complete manual before installing and using the regulator.

OPERATING AND MAINTENANCE MANUAL

Filter-Regulator-Lubricator

THE BP-690 SERIES. Operating and Service Manual. Series includes all variants of BP-LF/MF-690/691

TECHNICAL DATA CAUTION

Operating Instructions

PENBERTHY SUBMERSIBLE AUTOMATIC SUMP DRAINER INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

MODEL 154XST INSTALLATION AND SERVICE INSTRUCTIONS FOR MODEL 154XST STROBE LIGHT FOR USE IN HARSH ENVIRONMENTS/ HAZARDOUS LOCATIONS

Instruction Manual Pressure Relief Valve and Rupture Disc Chlorine or Sulfur Dioxide, Series 869

Booster Pump PB4-60 Replacement Kits

TESCOM 50-4X Series Safety, Installation & Start-Up Procedures

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.

Installation, Operation, and Maintenance Manual

Instruction Sheet. SUPERSEDES: New EFFECTIVE: August 1, 2013

INSTALLATION & MAINTENANCE INSTRUCTION

Unit 24: Applications of Pneumatics and Hydraulics

Pilot Check Valve: Metal Body Type

KTM 50 (DN ) Installation, maintenance and operating instructions

1200B2 Series Service Regulators. Instruction Manual

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER CONSTANT PRESSURE CYLINDER

Installation, Operation, and Maintenance Manual

FVA Trogflux variable area flowmeter. Operating Instructions. MECON GmbH 05/2013 OI_FVA_Trogflux_EN

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

200 PSI COMPRESSORS - MODEL NUMBERS

MUELLER. A Wall Type. Indicator Post. Reliable Connections. General Information 2. Technical Data/ Dimensions 3. Installation 4-5.

TECHNICAL DATA. Page 1 of 12

Sterile Visual Flow Indicator

Bermad Pressure Reducing. Model: 42T

Model No Product Name Handpiece Only Complete System

Installation and commissioning instructions 255 series and 256 series

TOP VALVE. Pat. #5,857,486 & 5,944,050. Mid-Range Pressure PSIG Back Pressure and Pressure Relief Valves. Instruction Manual

Manual Actuated Boiler Blowdown Valves

Measurement of Low GAS & Liquid Flows

SPECIFICATIONS CAPACITY

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

G92 Series BASOTROL Automatic Pilot Gas Valve

RD(H)20/25 Pressure-Reducing Regulator User Manual

TECHNICAL DATA. Pressure Regulation 531a. April 24, 2009

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

Discontinued. Powers Controls. Technical Instructions Document No P25 RV Rev. 1, May, RV 201 Pressure Reducing Valves.

100C Air Compressor Kit

1800C and 1800C-HC Series Service Regulators

DPC-30 DPC-100. Reference Manual

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters

Powers Controls RC 195 Single Input Receiver-Controller

Glass tube flowmeters Series 2000

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

Transcription:

KOBOLD KSM FLOWMETER User Instructions USA 1801 Parkway View Drive Pittsburgh, PA 15205 PH 412-788-2830 Canada 9A Aviation Point Claire, QC H9R 4Z2 PH 514-428-8090 www.koboldusa.com KSM_manual_08-10

Table of Contents 1.0 General... 1 2.0 Specifications... 1 3.0 Installation instructions... 4 3.1 Mounting... 4 4.0 Operation... 5 4.1 Reading the Flow Rate... 5 4.2 Adjusting the Setpoints; KSM -5000/6000... 5 4.3 Operating the Setpoints; KSM -5000/6000... 5 5.0 Maintenance... 6 6.0 Arrival of Damaged Equipment... 6 7.0 Need Help With Your KSM... 6 List of Diagrams Diagram 2.3 Switch Wiring Schematic... 2 Diagram 2.4 Dimensions & Pressure Loss... 3 List of Tables Table 2.1 Technical Data... 1 Table 2.2 Material construction... 2 Table 2.5 Reed Contacts - Technical Data... 4

1 KSM User Instructions KOBOLD KSM FLOWMETER CAUTION: For safety reasons, please read the cautionary information located at the end of the manual, before attempting installation. 1.0 General The KOBOLD KSM flowmeter is a high volume flow measuring device, intended for applications in which the corrosion resistance of synthetic materials (plastics) is required. The KSM operates on the principle of the variable area flowmeter (float in a conical tube or rotameter). The KSM may be outfitted with a setpoint switch (reed type - 1 or 2 pc.) to allow control of external electronics by triggering upon flow rate changes. 2.0 Specifications Table 2.1; Technical Data Operating Principle: Variable area Dimensions: See Diagram 2.4 Display: Direct reading, calibrated for Water or Air at standard conditions (20 C, 0 PSIG Outlet) Range: See Diagram 2.4 Operating Temperature: Maximum Internal Pressure: 32 F to 140 F 145 PSIG

KSM 2 Table 2.2; Materials of Construction Body: Float: KSM- X001 to KSM-X300 KSM-X600 Float Stops: O-Rings: End Fittings: Trogamid-T or Polysulfone PVDF PVC PVDF EPDM PVC Diagram 2.3: Switch Wiring Schematic

3 KSM Diagram 2.4; Dimensions & Pressure Loss

KSM 4 Table 2.5; Reed Contact - Technical Data Maximum Voltage: Maximum Current: Maximum Power: Switch Hysteresis: 230 VAC 0.5 A 10 W /12 VA 3-12 mm Maximum Ambient Temperature: 32 F to 130 F 3.0 Installation Instructions CAUTION: For safety reasons, please read the cautionary information located at the end of the manual, before attempting installation. NOTE: Before installation, remove any protective netting surrounding the float within the metering tube and discard. This netting is for shipping purposes only. 3.1 - Mounting The KSM is supplied with PVC end connectors and PVC cap nuts as standard. To install the instrument, simply: 1. Remove cap nuts and end fittings from your KSM. 2. Place cap nut over pipe to which you wish to connect. 3. Per manufacturer's instructions, use a PVC adhesive/solvent to glue fittings in place and let cure. 4. Fasten connectors to KSM with PVC nuts. Take care not to overtighten the nuts. If necessary, use a strap wrench on the nuts. Do NOT use a pipe wrench, pliers, or any other sharp instrument on the fittings, as this may compromise the strength, and eventual safety, of the instrument. Alternately, the KSM can be ordered with an optional NPT threaded fittings; alter the installation procedure accordingly. Use an appropriate thread sealant. NOTE: The KSM must be installed vertically (long axis), with the inlet at the bottom.

5 KSM 4.0 Operation We recommend that, if possible, the medium be introduced to the instrument in gradually increasing amounts. This procedure serves two basic functions: 1. It prevents oscillation of the float upon sudden introduction of media (particularlygases). 2. It prevents possible damage to the instrument caused by pressure surges (as described in the cautionary section). 4.1 - Reading the Flow Rate Flow rates measured by float type devices are typically read at the largest float diameter. In the case of the KSM, this is the top edge of the float. 4.2 - Adjusting the Setpoint; KSM...-5000/6000 series To adjust a setpoint: 1. Loosen the holding screw found on the setpoint indicator body, near the V-groove. 2. Slide the switch to the desired setpoint value. 3. Tighten the holding screw, taking care not to overtighten. 4. Verify desired operation with media flowing and adjust if necessary. 4.3 - Operating the Setpoints; KSM...-5000/6000 series Typical reed switch operation is as follows: The setpoint switch is actuated by the motion of the magnet-containing float past the switch position. The behavior of the switch is bistable - that is, the switch toggles from one state to the other whenever the float passes, and remains in that state until the float again passes back in the reverse direction. Since a reed switch is a switch in the purest sense, external electronics may be driven directly through it, provided that the electrical current/power loads are not demanding or excessive. (See Table 2.4 for reed switch performance data.) Should switching of currents or voltages in excess of the abilities of the reed switch be required, a suitable isolation relay must be installed.

KSM 6 5.0 Maintenance The major enemy of float-type flowmeters is dirt. We recommend that clean or filtered media only, be passed through the device. If using the setpoint switch, it is particularly important to guard from contamination by ferritic (metal) materials. To protect from these, we suggest the installation of a magnetic filter, such as a KOBOLD MFR.., or equivalent. 6.0 Arrival of Damaged Equipment Your instrument was fully inspected prior to shipment and found to be defect-free. If damage is visible on the unit, we advise that you carefully inspect the packing in which it was delivered. If damage is visible, notify your local carrier at once, since the carrier is liable for a replacement under these circumstances. If your claim is refused, please contact KOBOLD Instruments for further advisement. 7.0 Need help with your KSM? Call one of our friendly engineers at 412-788-2830.

7 KSM Caution PLEASE READ THE FOLLOWING GENERAL FLOW METER/ MONITOR WARNINGS BEFORE ATTEMPTING INSTALLATION OF YOUR NEW DEVICE. FAILURE TO HEED THE INFORMATION HEREIN MAY RESULT IN EQUIPMENT FAILURE AND POSSIBLE SUBSEQUENT PERSONAL INJURY.

KSM 8 Inspect instrument for damage upon arrival. Cracked, fractured, bent or otherwise damaged instruments must not be put into use, since the device is weakened to an unknown extent. (The operations and installation guide will explain how to make a claim on damaged instruments.) Under NO circumstances must the maximum tolerances (temperature and pressure) be exceeded. The maximum tolerances of the device have been determined using water, air and/or oil. If using other media, especially corrosive ones, it is critically important that the user determine chemical compatibility with our instruments. A list, detailing material composition of our instruments, is available from KOBOLD Instruments Inc. upon request. KOBOLD Instruments Inc. cannot accept responsibility for failure and consequences resulting from use of media other than water, mineral oil, air, and nitrogen. Install the device in a fully supported position within your flow system. This avoids excessive stresses which may damage the instrument. In particular: a. Ensure that the plumbing leading to and from the instrument is fully supported and that the instrument does not perform the physical function of a joint. b. When calculating stress on the device caused by plumbing, the weight of the medium in the pipes must be considered as well. c. Misaligned runs of rigid piping can cause large stresses when connected to the instrument. Do not connect in such a fashion. During installation, avoid stresses on the instrument by following guidelines given below: a. When connecting fittings, hold the instrument fittings rigid with a correctly sized wrench. Do not install by twisting the instrument into the pipe fittings. b. Do NOT install by holding the device housing to provide counter-torque to the pipe fitting. c. Use an appropriate amount of PTFE tape on male threads of fitting. This reduces the twisting stresses produced by tightening the fittings into each other. d. Do not use pliers or wrenches on the housing, as this may damage it. e. Do not overtighten, as this may fracture the fittings.

9 KSM During operation, there are a number of situations to avoid: a. The sudden cessation of fluid flow causes what is typically referred to as "water hammer". Most people are familiar with this phenomenon from their home experience - it is the cause behind the loud clank of water pipes which occurs when faucets are turned off too suddenly. The cause behind this "water hammer" is quite easy to visualize. Water is fairly massive. The amount of water in long runs of pipe is quite substantial. When the faucets are turned off suddenly, especially from a full on condition, the water has considerable momentum and does not want to stop flowing. The situation is similar to stopping a car by running into a wall, rather than by applying brakes. Both are sudden rather than gradual. The damage to the wall can be substantial (not to mention the car). The "water hammer" causes surges in fluid pressure which could cause the measurement instrument's pressure limit to be exceeded, resulting in failure and possible personal injury. b. Fluid surges, as well as the water hammer, can be particularly damaging to empty flowmeters since there is no back pressure in the device. The damage is caused, once again, by momentary excess pressure. To avoid these surges, fluid lines should remain full (if possible) and water flow should be introduced to the device slowly. c. If the instrument is isolated with inlet and outlet valves, the flowmeter must be completely drained when said valves are both closed. Failure to do so could result in damage to the device caused by thermal expansion of fluid. d. Freezing of water in the instrument must be avoided since the resultant expansion will damage the flowmeter and make it unsafe for use.