Warnings: Notes: Revised: October 5, 2015

Similar documents
Warnings: Notes: Revised: January 8,

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

Karl Suss MA6 Mask Aligner SOP

Karl Suss MJB4 Mask Aligner

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Standard Operating Manual

University of Minnesota Nano Fabrication Center

SOP for Karl Suss MJB3 #1 Mask Aligner

University of Minnesota Nano Fabrication Center

Standard Operating Manual

Approved by Principal Investigator Date: Approved by Super User: Date:

MJB-3 Mask Aligner Property of TAU MNCF August 2012

Quintel Mask Aligner

ABM MASK ALIGNERS. NanoFab 26 March 2009 A Micro Machining & Nanofabrication Facility

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure

Karl Suss Contact Aligner Operation

OAI Model 200 Tabletop Mask Aligner Portland State University

Approved by Principal Investigator Date: Approved by Super User: Date:

March CS-1701F Reactive Ion Etcher

COBILT CA-800 Mask Aligner Equipment Operation

Operation of the mask aligner MJB-55

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera

Photolithography. Operating Instructions

Revised: June 7, 2017

OPERATION OF THE DIMPLER

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

Angstrom Dielectric Sputterer Operation Manual

University of MN, Minnesota Nano Center Standard Operating Procedure

Cryo-Evaporator Operation

Operating Procedures for Metal Evaporator I

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

Standard Operating Procedure. For. PVD E-Beam

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

Standard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

Xactix XeF2 OPERATION MANUAL

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

Approved by Principal Investigator Date: Approved by Super User: Date:

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

Standard Operating Manual

Usage Policies Notebook for Parylene Coating System

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

Scanning Electron Microscope JEOL JSM F

Underwater Housing for Panasonic Lumix DMC-ZS60, DMC-TZ80

Standard Operating Manual

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

ELE INTERNATIONAL COMPRESSION TESTER Standard Operating Procedure

EE 432 Lab 3 PMOS source/drain lithography and diffusion

Columbus Instruments

Standard Operating Manual

STS ICP-RIE. Scott Munro (2-4826,

RH800 & 2000 BASIC OPERATIONS

Cover Page for Lab Report Group Portion. Compressible Flow in a Converging-Diverging Nozzle

Standard Operating Manual

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

Model 130M Pneumatic Controller

Oerlikon Sputtering Evaporator SOP

ACV-10 Automatic Control Valve

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer

Mounting and Operating Instructions EB EN. Type Supply Pressure Regulator. with increased air capacity

BASIC RULES OF USAGE, OLIS DSM-20 CD Management The instrument is currently managed by the Analytical Biophysics Core Facility.

[needs to be discussed with faculty if any new material to be done]

Product Information News

User Manual GRI- 1500Li

Plasma Etching: Safety Operating Procedures

[USING THE VITROBOT April 9, Using the Vitrobot

OPERATING INSTRUCTIONS FOR THE MODEL 85B

Cover Page for Lab Report Group Portion. Compressible Flow in a Converging-Diverging Nozzle

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

Plasma Asher: March PX-500 User guide (May-30, 2017)

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Underwater Housing for Canon EOS M

Maintenance Manual. Note: This document is broken up into the following 6 sections of prescribed maintenance.

BAUSCH & LOMB ENGINEERING ROCHESTER. Title: Membrane Press Procedure Page 1 of 14

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

LCR-6-WT-EPP CYLINDER ROLLER EXPLOSION PROOF SERIES INSTRUCTION MANUAL MANUAL NUMBER Rev A Issued November 2011

Glove Box Installation Manual

Pressure Balance Trim

MyControl Bathing System Signature Series

ASSEMBLY & OPERATIONS MANUAL. Straight & Clean Torch Cuts. Every Time.

SPUTTER STATION STANDARD OPERATING PROCEDURE

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions

Underwater Housing for Canon G7X Mark II

Standard Operating Procedure - First Draft for Review Only- StarWeld Ultra DX

Nordiko Metal Sputtering System Standard Operating Procedure

VACUUM REGULATORS CONTENTS

Birck Nanotechnology Center

SOP for use of the SX19 Stopped-Flow Instrument

Sterisil Ac+ Installation & Operating Manual

Transcription:

Karl Suss MA6 Mask Aligner Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any campus phone.) (617)627-3246 (From off-campus or from a cell phone) Tufts Emergency Medical Services are at x66911. Revised: October 5, 2015 Warnings: 1. Whenever the shutter is open you may be exposed to UV radiation. Never look directly at the lamp. Do not put your hands under the lamp while the shutter is open. 2. The lamp contains mercury and is pressurized. If the lamp explodes, everyone should evacuate the cleanroom immediately (take your suit off in the hall outside). Call the safety office (number is at the top of this page) and Prof. White. Notes: 1. The lamp is calibrated to 20 mw/cm 2 intensity at 365 nm (i-line). However there is additional energy at 405 nm (h-line) and 436 nm (g-line). The total broadband UV intensity from the lamp is approximately 60 mw/cm 2 although this is not calibrated. 2. In order to see your alignment marks at the lowest magnification (5x) you should size and place your alignment marks appropriately. Here is a suggestion: 30 mm 30 mm Place alignment marks along the wafer centerline, 30 mm from the center. Make them symmetrically positioned about the wafer center!

Try using this alignment mark. Dimensions are in microns. If you make the crosses and squares transparent on your mask, make the crosses and squares slightly smaller (18 and 8 microns wide crosses, something like 48 and 28 micron square squares, but positioned at the same centerpoints!!) on the substrate. If you make the crosses and squares opaque on the mask, then do the reverse. I m sure this isn t optimal. It s just a suggestion. 1.0 Material Requirements: 2.0 Procedure: 1.1 Equipment: A 5 by 5 by 0.09 mask plate (usually we use Chrome on Soda Lime glass) 1.2 Personal Protective Equipment: Nitrile gloves and safety glasses (standard cleanroom gear) Facilities Facilities should already be hooked up and turned on. However, if there is any question: 1. The air and vacuum valves at the wall should be open (handle in line with the body of the valve) 2. The two air regulators should be set to 75 psi (5 bar) for the line going to the compressed air inlet, and 15 psi (1 bar) for the line going to the nitrogen inlet. 3. Vacuum should read < -0.6 bar on the dial on the front of the MA6. Turning on the Lamp: When you first come in to use the aligner, the lamp may be powered down. To check if the lamp is powered down, look at the lamp controller under the aligner on the right. If the power switch is off then the lamp is off. If the lamp is off: 1. Hold down the Up Arrow and CI1 buttons on the front of the controller and simultaneously turn on the power, continue to hold down for 2-3 seconds after the controller powers up. This sets the controller to be in Constant Intensity mode based on CI1 (365 nm).

2. The controller should turn on and briefly flash C1 to show it is in constant intensity mode. It is very difficult to see the CI flash on the controller it goes by very fast - try to pay careful attention to make sure it worked and you are in constant intensity mode! 3. It will then go through a series of checks and say rdy. 4. Push start the lamp will attempt to ignite. It will make a nasty buzzing sound and flash the word FIRE. This is normal. It will attempt to ignite up to 9 times, and, if not successful, will give up. If this happens, notify lab staff and put a note on the tool telling people the tool is down. 5. If the lamp starts successfully, it will stop buzzing/beeping and the display will flass COLD (this is normal). Wait 15 minutes for the lamp to warm up; it should stop flashing COLD. Turning on the aligner: 1. Once the lamp has been started, turn on the aligner by gently turning the green switch on the electronics front panel to on and letting go. 2. The electronics should immediately start up. 3. Watch the screen as it goes through startup and note any error messages. 4. The screen should say Rdy for start push load button. 5. Push Load This will complete the startup process. You are not actually loading a wafer yet. 6. The screen should read Ready for load 7. Make sure the X and Y stage micrometers are set to 10 (this is the middle of travel). 8. Make sure the Theta micrometer is adjusted so that the white cross marker on the front of the stage is centered. (Steps 7 and 8 make sure you start out with everything centered). Loading the Mask: 1. Handle the mask by the edges, it is bad practice to touch the mask s exposure area. Also a fresh pair of gloves before handling the mask is a good idea 2. Inspect your mask under the microscope and make sure it is clean. 3. Push Change mask 4. Slide out the mask frame, flip it over and put it on the wing-table to the left of the aligner. 5. Pull the small metal clip gently a few millimeters towards you by putting your fingernail into the notch in the black plastic. It should snap into place. 6. Load the mask against the alignment pins with the metal facing up (this is the side that will touch the wafer after you flip the frame over to put it back in the tool). 7. Push enter to engage mask vacuum. The mask should now be held to the frame. 8. Push down on the metal part of the clip with your fingernail the clip should pop back into place, securing the mask further. 9. Slide the mask frame back into the tool. 10. Push Change mask to complete the process. 11. If you are using a mylar mask, load the glass blank following the directions above. The mylar mask can be secured to the blank using bits of vacuum type tape. Be careful not to place any tape in the exposure area, or where the mask and wafer may contact. Alternatively, the mylar mask can be trimmed to match your substrate size and carefully placed directly onto the substrate when you load the substrate later. Setting up the Exposure Parameters: (Do this before loading the wafer once the wafer is loaded you cannot change the contact type) 1. Push edit parameters 2. Use the left and right arrows to select the exposure parameter to change. 3. Use the up and down arrows to change the value. 4. Choose either proximity, soft contact, hard contact, or vacuum contact - whichever you want to use. If you don t know what these are, look them up. 5. Change the exposure time in seconds based on the total exposure dose you want. Remember, the source is calibrated to 20 mw/cm 2 at i-line (365 nm) and has approximately 60 mw/cm 2 broadband UV intensity (not calibrated). It is highly recommended to check the log to see the parameters other users have chosen and the results.

6. Change the alignment separation distance here too, if desired. A typical separation distance would be about 30 microns. 7. Do not change the following settigns Pre Vac [s] : 4 ; Vac Purge [s] : 4 ; Full Vac [s] : 4 WEC type : Cont ; WEC offset: 0 8. Once everything is all set, push Edit parameters again to lock those parameters in. Loading the substrate: 1. Push Load 2. Pull out the wafer slide. 3. Load your wafer, pushing it up against the alignment pins. Make sure the wafer is not ON TOP of the pins or it may break! 4. Push enter to engage the wafer vacuum. The wafer should be held down. A hissing indicates misalignment, or the back of the wafer is contaminated and cannot make a good vacuum seal with the holder. 5. Slide the frame back in. 6. Push enter again to indicate that the wafer is in. 7. The tool will automatically perform WEC (wedge error compensation) at this point, parallelizing the wafer and mask, and will then go into separation. Wait for the process to complete. Lithography results depend on perfectly parallel surfaces, thus the wafer and mask must be flat. Any contamination (particles, old resist) that interferes with surface flatness will degrade your results (and possibly harm the aligner), see image below Alignment Align/focus microscopes on mask targets 1. Adjust overall illumination using the TSA illumination knob on the main panel. 2. Adjust individual illumination to the two sides by opening or closing the apertures (using the small silver levers located under each microscope camera). 3. Move the microscopes over the mask by using the X and Y arrow buttons on the main control panel. You can toggle fast on and off as needed. 4. Set the coarse and then the fine focus - a. Set fine individual microscope focus left/right TSA adjustments - to midpoint of travel (3 turns total, so 1.5 turns from clockwise stop) b. Adjust coarse microscope focus together using the large aluminum knob on the top of the tool. c. Adjust individual microscope fine focus using the left and right TSA adjustments d. If you run out of fine focus travel, reset to midpoint, readjust coarse focus and try again 5. Rotate microscope theta using the small knob near the spring above the microscopes. 6. Move the individual microscope objectives left and right in the X direction using the two medium silver knobs on the left and right sides of the microscope.

Alignment Align substrate to mask Note: The Contact light on the aligner dashboard. If it is lit, the mask and substrate are in contact, no alignment adjustments can be made in this condition 1. Move the wafer in X, Y, and Theta using the large labelled knobs under the stage. Coarse and fine adjustment is available. 2. You can rotate the microscope objective turrets to change magnification if needed. Be careful not to crash an objective into the mask!! 3. Once alignment has been achieved, you can toggle contact by pushing the Align/Cont button the panel. When the wafer and mask are in contact the yellow contact light will be on do not attempt to move the wafer when the wafer and mask are in contact!! 4. To change the separation between the mask and the wafer, push Edit parameters and change the alignment distance parameters (use left and right arrows to find it and up and down arrows to change it). Next time you go in and out of contact it will go to the new separation distance. This is useful if the substrate targets are outside the depth of field of your microscope objectives. 5. If you need to move the microscope out of the way to get a look at your wafer, push F1 and enter. Repeat F1 and enter to bring them back down. Exposure 1. If you change your mind and don t want to expose, push Unload 2. When everything is aligned and all parameters are set push Exposure Exposure will occur for the specified length of time! Do not look at the UV light, turn away. If there are others in the lab with you tell them you are exposing so they know to look away as well. Removing your substrate 1. The wafer will automatically be ready to be removed after exposure 2. Substrate removal is the reverse of loading described above. 3. Pull out the substrate slide, toggle the wafer vacuum off by pushing enter (if needed see instructions on the screen). 4. Remove the wafer. 5. Slide the wafer slide back into position. Removing your mask: 1. Mask removal is the reverse of loading described above. 2. Push change mask 3. Pull the mask frame out and flip it over. 4. Push enter to release the mask vacuum. 5. Pull the clip forward as before. 6. Remove the mask taking care not to touch the exposure area. 7. Slide the mask frame back into position. 8. Push change mask to lock the frame back in. Shutting down the system: 1. Reset the X and Y micrometers back to the center (value = 10), Set the Theta micrometer back to center (white cross alignment marker on the front of the stage centered). 2. Once exposure is complete, if another user will do lithography soon, just leave the tool on with no mask and no wafer in it, but with the slide and mask frame loaded. 3. If no other user will use the tool today: a. Turn off the lamp controller (just turn the power switch off on the lamp controller, it is that easy, cooling air is still flowing to the lamp, don t worry) b. Turn off the green key switch on the electronics to shut the system down.

Terminology WEC Wedge Error Compensation. Procedure employed to parallelize the mask and substrate, performed automatically. There are two basic options: Contact WEC utilizes global planarity of mask and wafer to bring both into planarity. Wafer and mask will be brought into contact to complete procedure. Proximity WEC utilizes a 3 point contact procedure to bring mask and wafer into planarity. TSA Top Side Alignment, use of upper optics to align substrate to mask BSA Back Side Alignment, use of bottom side optics to align substrate and mask. This feature is not available on our aligner Soft Contact one possible lithography mode, the mask and wafer are brought into contact at a preset pressure. Recommended for feature sizes in the 10 s or 100 s of microns scale. Hard Contact one possible lithography mode, the mask and wafer are brought into contact at the same preset pressure with additional pressure provided by N 2 gas. Recommended for features sizes in the 10 s of microns or less. Vacuum Contact one possible lithography mode, the mask and wafer are brought into contact and the vacuum seal inflates to seal the volume which is then evacuated. The parameter PreVac Time allows the vacuum to proceed slowly, preventing any alignment shift. Recommended for absolute minimum feature sizes. I have been told sub micron feature sizes are achievable with this aligner. Proximity one possible lithography mode, the mask and wafer are separated by a user-defined exposure distance. The mask and wafer never make contact. Recommended for particularly sticky and viscous resists and very large feature sizes If at any time you feel a situation is dangerous, do not hesitate to call the safety office (x73246, Peter Nowak) the faculty supervisor/lab manager (x72210, Robert White), or Tufts Emergency Services (Police/Fire/Ambulance at x66911). Report all accidents (injuries, major spills, fires) to the safety office at x73246 (Peter Nowak) and Prof. White at x72210. For emergencies, call Tufts Emergency Services at x66911.