This GC has two columns (I think new columns can be ordered from Restek)

Similar documents
5890II GC Standard Operating Procedure 9/2/2005

Multiple Gas#5 GC configuration Jan 2016

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

New improved robot for gas kinetics

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

SRI Multi Gas Analyzer 2016

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

SRI 8610C Gas Chromatograph Multiple Gas #3 GC configuration

Quantitative Analysis of Hydrocarbons by Gas Chromatography

System Overview. TCD Detector temperature setpoint Regulator

Agilent 490 Micro GC Natural Gas Analyzer

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Gas Leak Measurement of TRT Barrel Module

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

GCMSD-Headspace Analysis SOP

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Rapid and Reliable Detection of Dissolved Gases in Water

AN-03 Rotary valves applications

A New GC/MS System Designed for Helium Carrier Gas Conservation

Zebron Gas Management

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Automated Determination of Dissolved Gases in Water Anne Jurek

AKTA pure 25 New Owner s Intro

Operation of the Perkin Elmer TGA-GC/MS

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Heat Pump Connections and Interior Piping

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Calibration of a GC955

Carrier Gases in Capillary GC

During today's webinar we're going to look at the following topics:

Using the Akta Prime plus October 22, 2012

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

FROG-4000 TM Chemical Analysis System

Leaks, contamination and closed loop control how RGAs make coating processes more profitable

Mass Spec will not Autotune

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Chromat Calibration Updated October 27th, 2017

Bernhard Thomas September EnCal 3000 prochain Energy in Gas with ROI

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

Retention Time Locking: Concepts and Applications. Application

Weiyi Zheng, Xiaoxia Dong, Evan Rogers, Jimmie C. Oxley, and James L. Smith*

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

6890 GC Not Ready Messages (A16010)

Your local gas generation partner. Precision series Modular gas generation solution for GC.

FROG-4000 TM Chemical Analysis System

Manual for continuous distillation

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Hazards associated with the gas system and how to mitigate them

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

SOP: Buck Scientific BLC-20P HPLC Operation

Background Statement for SEMI Draft document 4657B NEW STANDARD: SPECIFICATIONS FOR TUNGSTEN HEXAFLUORIDE (WF 6 )

6890 GC Site Preparation

28 Site Preparation. Supplying valve actuator air

Svensk läkemedelsstandard

Carrier Gases in Capillary GC

no peaks missing peaks too small peaks increasing retention decreasing retention

Svensk läkemedelsstandard

REFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY

CP-3800 GC Operator s Manual

Experiment GC : Analysis of BTEX by GC-FID

Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results

Ultimate Performance in Gas Chromatography

Svensk läkemedelsstandard

Standard Operating Manual

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

3 Flow and Pressure Control

3-Way Decoking Valve Option - Installation and Operation Instructions Accessory G Background

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin

User manual. TC-20 Sample conditioning and dry-purge rig. QUI-0003 Version 6.3 September

Continuous gas analysis ULTRAMAT 23 Supplement 2 for Operating Instructions C79000-G52xx- C216-03

Prep LC Training Manual

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector

Buck Scientific, Inc.

The. «Modular Alliance» Gas Generators. ALLIANCE, the partnership that keeps the gas flowing. One step ahead!

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating

MicroJet Cryo-Trap MODEL MJT-1030Ex OPERATION MANUAL

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

MMS Mercury Monitoring System for Natural Gas

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

Figure 1: Plug in ion inlet of mass spec

CONNECTIONS. The Thermal Conductivity Detector

Space product assurance

Transcription:

SRI 8610C GC Ann s Notes 11-17-2014 Introduction This is the GC as it came from the manufacturer This GC has two columns (I think new columns can be ordered from Restek) 1. Hayesep D: Column 1 a. Separating CO, CO 2 and other gaseous carbon compounds ex. acetylene 2. Molecular sieve: Column 2 a. Separating extremely small gases b. IMPORTANT: CO 2 and H 2O cannot enter this column because the pore size is not big enough to allow them to pass through. They must be burned off the column at 200 C (limit of stability for molecular sieve) if they enter. Because CO 2 cannot enter into the molecular sieve column, there is a switching mechanism between the columns to only allow certain gases to pass through the molecular sieve column. This switching mechanism is designated G on or G off. When G is off the gas flows through only the hayesep D column. When G is on the sample flows through both columns. In addition there is a solenoid valve that stop flow through the hayesep D column to allow the separation to occur on the molecular sieve column. When the solenoid valve is open, it allows the

carrier gas to pressurize both the inlet in the outlet of column A stopping the flow through this column. When A is off the solenoid is closed and gas flows through the hayesep D column. When A is on there is no flow through the hayesep D column. The GC was intended to be operated in the mode where you constantly have sample flowing in and out. This does not work for us since we don t have enough sample and we want to use manual injection. The manual injection in deposited straight into the carrier gas. However, even when using manual injection you have to flow carrier gas through the sample loop to flush the remaining sample out. This had been verified now twice by Joerg that you get double peaks if you switch the G valve while injecting sample manually because there is still sample in the loop that then re-elutes on the next run. We didn t want to run carrier gas through the sample loop since it is expensive helium (Mass spec grade) so we rewired the GC. We rewired it to not have the sample loop in the flow. I relabeled the diagram on the GC below to detail how Joerg and I rewired the GC. The rewiring involved 1. Hardwiring the carrier gas to the inlet of column 1.

2. Hardwiring connection 4 to connection 9 a. In G off position column 2 is connected to itself so nothing will flow b. In G on position the sample loop is connected to itself Based on this rewiring we have created two programs to measure gases. Ann s program is intended to measure CH 4, CO 2 and CO. Joreg s method is to measure CH 4 and H 2. We edit the method every time because the software sucks. To edit the method go to Edit/Channel. Then click on the appropriate box to edit what you want. It should be noted you should change the name where the file will save by clicking the postrun box. You need to make sure the postrun box for channel 1 and 2 have different names or one will overwrite the other when saving. Also, the change the carrier flow you must manually do this by changing the value for the carrier 1 panel on the front of the GC. Random GC comments When switching carrier gas make sure the run the new carrier gas through both columns and make sure all of the previous carrier gas has been purged before making measurements. When integrating make sure to ingrate from the baseline and draw peaks as designated below. This involves manual integration because the GC is not programed to integrate this way. The gas flows through the TCD detector followed by the FID since the sample is destructively measured in the FID. Therefore the peak for the TCD is 0.1 min earlier than the FID peak Joerg s method Carrier Gas: N 2 Carrier Flow: 13 Temperature: 40 C Run Time: 3 minutes FID sensitivity: medium Event File: G off entire time Elution Time Compound Elution Time TCD (min) H 2 0.516 CH 4 1.130 CO 2 2.553

1% CH 4/1% H 2 Combined Standard Curve Ann s method Carrier Gas: He Carrier Flow: 13 Temperature: 40 C Run Time: 10 minutes FID sensitivity: medium Detection: TCD works for high concentrations (in the percent range). Event File Time (min) Event Flow 0 G on Flow through both columns 0 A off 3.4 A on Stop flow Hayesep D, flow through molecular sieve 6.7 G off Flow through Hayesep D only 6.75 A off Resume flow through Hayesep D

10% CO Standard in He Compound Elution Time TCD CO 4.1 CH 4 5.1 CO 2 8.1 100% CH 4 20% CO 2 in N 2 More information in the manual