Operating Guidelines for PermSelect Modules Liquid Contacting

Similar documents
Technical Service Bulletin March 2015 TSB142.01

OPERATING MANUAL LO-CROSS-FLOW CF M

Design & Operating Guidelines for Liqui-Cel Extra-Flow Membrane Contactors

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

TECHNICAL INFORMATION

Integrity Testing LifeASSURE BNA045 and BNA065 Series Filter Cartridges

Installation, Operation, and Maintenance Manual Parker Balston Model N2-04 Nitrogen Generator

POWER FLUSH KIT INSTRUCTIONS

ACCU-PULSE Installation and Operation Instructions

A Perspective on Filter Cartridge Issues

COMBINATION AIR RELEASE DEGASSING (CARD) VALVES INSTALLATION AND MAINTENANCE MANUAL

Technical Information

Application Worksheet

Technical Service Bulletin October 2013 TSB135.03

Operating Instructions

Heat Pump Connections and Interior Piping

FIK In-Tank Filters FIK. Max Flow: 170 gpm (643 lpm) Working Pressures to: Rated Static Burst to: Flow Range to:

Best Practices for Installation of Gaseous Sample Pumps

Winery Trials on Gas Control Using 3M Liqui-Cel. 3M Separation & Purification Sciences Erwin Ona, Senior Applications Engr

Alkylphosphines. Storage and Handling Recommendations

3M Liqui-Cel EXF-10x28 Series Membrane Contactor

Type C-CUT. Quickstart. English Deutsch Français

CHAPTER 3 : AIR COMPRESSOR

AIR-OPERATED DOUBLE DIAPHRAGM PUMP USER S MANUAL

SEPAREL. Hollow Fiber Membrane Module INSTRUCTION MANUAL (For IJ) EF External Flow Small Size Series. EF-MICRO EF-G2 Series EF-G3 Series EF-G5 Series

3M Liqui-Cel EXF-10x28 Series Membrane Contactor with ANSI or JIS Connections with Integrated End Caps

The most common terms rating air flow capacity are ICFM, FAD, ANR, SCFM or nl/min

SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

WHP Operations and Methods { July William M. Smethie, Jr. Lamont Doherty Geological Observatory Palisades, New York U.S.A.

Concentration and diafiltration of viral antigens with SmartFlow TFF

DIAPHRAGM VACUUM PUMPS AND COMPRESSORS

INSTRUCTION MANUAL FOR POLYKEG DISPOSABLE KEG

CS Controlled Dissolved Gas assembly with optional CS Air Aspirator assembly. Installation and Operation information

STANFORD UNIVERSITY W.W. HANSEN EXPERIMENTAL PHYSICS LABORATORY GRAVITY PROBE B, RELATIVITY GYROSCOPE EXPERIMENT STANFORD, CALIFORNIA

ULTRA PURE SYSTEMS 625 GPD RODI INSTALLATION AND OPERATION MANUAL

ENSURING AN ACCURATE RESULT IN AN ANALYTICAL INSTRUMENTATION SYSTEM PART 1: UNDERSTANDING AND MEASURING TIME DELAY

SERIES 30. Spring Operated Tank Blanketing Valve PROTECTOSEAL.

OPERATION MANUAL NTF-15

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

1020 Industrial Drive, Orlinda, TN fax

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

OPERATION MANUAL NTF-60 Plus

Inert Air (N2) Systems Manual

API Piping Plan 62: A Reliable Quench System

Series AMJ. Drain Separator for Vacuum

COMPRESSOR REPLACEMENT PROCEDURE (R-134A)

Goulds 3296 EZMAG. Chemical Process Pump

Nitrogen Supply System

JANIS OPERATING INSTRUCTIONS FOR SUPERCONDUCTING MAGNET CRYOSTATS

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist

Pilot Check Valve: Metal Body Type

VACUUM TESTING PRECAST CONCRETE MANHOLES

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION

Stainless Steel 142 mm Filter Holder YY

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

OPERATING INSTRUCTIONS

hydro-pac, inc. Low-Pressure Gas Compressors 1500 to 6000 PSI

CASH VALVE TYPE KP BACK PRESSURE VALVES

SERIES 10. Pilot Operated Tank Blanketing Valve, 2" PROTECTOSEAL.

Another convenient term is gauge pressure, which is a pressure measured above barometric pressure.

Table of Contents - EverLUX Filters

TECHNICAL INFORMATION

Reduce Turnaround Duration by Eliminating H 2 S from Flare Gas Utilizing VaporLock Scrubber Technology

HAYWARD FLOW CONTROL TFS SERIES FLOW METER INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

NGP-250/500 Nitrogen Generator Quick Start Guide

Fast. Efficient. Precise control. 3M Liqui-Cel Membrane Contactors. Dissolved gas control solutions for the beverage industry

Smart Water Application Technologies (SWAT) TM

After filling the pool with clean water in the summer, two basic factors can take place due to the exposure of water to the sun and the air.

Hydraulic Reliability & Preventive Maintenance Report

maintaining storage tank dissolved oxygen levels utilizing gas transfer membranes

4400 TwinHybrid Gas Seal

DIAPHRAGM PUMPS WITH KNF STABILIZATION SYSTEM

3M Liqui-Cel EXF-14x28 and 14x40 Series Membrane Contactor

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

THE INNER WORKINGS OF A SIPHON Jacques Chaurette p. eng. January 2003

Installation & Instruction Manual For Portable Series R.O. units:

D R A F T. Operating Procedures for the NPDGamma Liquid Hydrogen Target at the BL 13. Version 1.00

ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE

HAYWARD FLOW CONTROL Series PBV Back Pressure Valve and Series RPV Pressure Relief Valve INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS

DIAPHRAGM GAS SAMPLING PUMPS

2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters. Kent-Taylor

Cash Valve TYPE KP PILOT OPERATED BACK PRESSURE VALVE. ISSUED - DECEMBER 2000 CAVMC-0518-US-0208 ISO 9001 Certified

VACUVENT Valve functions - benefits - flowpath - VG series - Sewage

Air Eliminators and Combination Air Eliminators Strainers

NITROGEN GENERATOR Tradition Trail Plano, Texas 75093

Type C-CUT. Operating Instructions. Bedienungsanleitung Manuel d utilisation

ULTRA PURE SYSTEMS 400 GPD RODI INSTALLATION AND OPERATION MANUAL

Lesson 6: Flow Control Valves

Table of Contents - ALpHA Filters

School of Chemistry SOP For Operation Of Glove Boxes

Membrane Air Dryer. IDG1-C06 Series IDG1-C06-P Series

Installation Operating Instructions for Duplex Manual Manifolds PX-TMD Series

TEST BENCH FOR SAFETY VALVES

Dean Pump Self-Priming Chemical Process Pumps

The Shand & Jurs Model Vapor Guard Tank Blanketing Valve

FLUID POWER FLUID POWER EQUIPMENT TUTORIAL ACCUMULATORS. This work covers part of outcome 2 of the Edexcel standard module:

PROCESS GAS SAMPLING SYSTEMS

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

Transcription:

Operating Guidelines for PermSelect Modules Liquid Contacting PermSelect PDMSXA-2500 (0.25m 2 ) PermSelect PDMSXA-7500 (0.75m 2 ) PermSelect PDMSXA-1.0 (1m 2 ) PermSelect PDMSXA-2.1 (2.1m 2 ) PermSelect membrane modules can be used in a variety of liquid contacting or gas separation applications. The module design allows operational flexibility to achieve targeted objectives utilizing dense PDMS hollow fiber membranes. Depending on the application of interest, please follow outlined start up and operating guidelines. Please note that these operating guidelines apply only when using with the following fluids: water, air, and the components of air (O 2, N 2, CO 2 ). Every customer s application is different and unique. It would be very difficult to predict the performance of the membrane modules for each intended application with different fluids and operating conditions. Thus this document should only be used as an initial guideline, in combination with the chemical compatibility charts, to help determine if a PermSelect membrane module is suitable for a specific application. It is up to each customer to determine the suitability of a PermSelect membrane module in their own application by their own testing and evaluation. WARNING: While we may provide information to help our customers determine if our membrane modules are suitable for their intended application, MedArray does not represent or warrant that our products are fit or safe for a particular application. PermSelect membrane modules are only to be used by persons thoroughly familiar with its use in its intended application. It is the responsibility of the user to determine the suitability of membrane modules in its specific application. Membrane modules can fail, permitting fluid discharge into the environment and mixing of shell and tube side fluids. Users must take all precautions to ensure safety to people and property in case of module failure. Purchaser assumes all responsibility for the suitability and fitness for use as well as for the protection of the environment and for health and safety involving this product. Please read our Terms of Sale for further information. Selecting Operating Mode For liquid contacting applications, there are three modes of operation to choose from: sweep, vacuum and combination of the two. Sweep gas mode of operation is commonly used to transfer gases into the liquid such as O 2, N 2 or CO 2. Vacuum mode is most commonly used to remove dissolved gases and bubbles from the liquid. Combination mode (sweep with vacuum) is recommended to enhance degassing performance by increasing the gas transfer driving force. Modes of operation are shown below: MedArray, Inc. 3915 Research Park Dr., Suite A-4 Ann Arbor, MI 48108, USA Page 1

1a. SWEEP 1b. VACUUM 1c. SWEEP w/ VACUUM Figure 1. Liquid Contacting Operating Modes Liquid in the shell side The most efficient mode of operation of a PermSelect membrane module for liquid contacting is with the liquid on the shell side (outside the hollow fibers) as shown in the diagrams above. However, this mode of operation typically has a lower fluid pressure limitation compared to operating with the liquid on the lumen side. The trans-membrane pressure (TMP) defined as the difference in pressure between both sides of the membrane, should not exceed 15 psi from shell to the lumen side, and 45 psi from lumen to shell at ambient conditions. The reason for the lower pressure limit from shell to lumen is because the hollow fibers may collapse and be damaged if the pressure from the outside of the hollow fiber to the lumen side of the hollow fiber exceeds 15 psi. On the other hand, if the pressure from the lumen side to the outside of the hollow fiber exceeds 45 psi, the hollow fibers may over-distend and potentially be damaged. Gas will transfer across the membrane in either mode of operation, but more efficient transfer (per unit area) is achieved with liquid flowing in the shell side. Thus it is important to know the fluid operating conditions before selecting the mode of operation, and depending on the operating conditions of the system, the maximum allowable trans-membrane pressure (TMP) may limit the way a module should be used. If the liquid supply pressure is fixed and known, then subtract the sweep gas (or vacuum) pressure from this value to determine the TPM. For example suppose water is supplied from a centrifugal pump at a maximum of 30 psig. Assume the atmospheric pressure is 14.7 psia, thus the water pressure is 30 psig + 14.7 psia = 44.7 psia. In this example water can flow in the shell side provided the gas (lumen) pressure is no less than 44.7 psia 15 psia = 29.7 psia, which is above atmospheric pressure, so this configuration would not be suitable for degassing applications. On the other hand liquid can flow in the lumen side provided the shell side pressure is no less than 44.7 psia 45 psia = -0.3 psia which is essentially full vacuum (0 psia). Note that the maximum TMP for each mode of operation applies whether gas or liquid is used on either side of the membrane. However in liquid contacting, it is always preferable to maintain the liquid side pressure higher that the gas pressure to prevent gas bubbles from eventually forming in the liquid. MedArray, Inc. 3915 Research Park Dr., Suite A-4 Ann Arbor, MI 48108, USA Page 2

Thus in general, with water in the lumens of the hollow fibers, full vacuum can be used with as high as 30 psig water pressure. With water flowing in the shell side and full vacuum was applied to the lumens, the maximum water pressure should not exceed 0.3 psig. This is why liquid on the shell side is limited by maximum TMP. Figure 1 above illustrates the module installation / operation when liquid is flowing on the shell side. Figure 2 below illustrates proper installation / operation with liquid flowing in the lumens. Table 1 below provides a fluid pressure guide for vacuum degassing applications. Table 1a. Vacuum Pressure Guide with Liquid on the Shell Vacuum Pressure (Gauge) Maximum Membrane TMP Maximum Liquid Pressure (Gauge) inhg mmhg (Torr) psi psi psi Pa atm mbar -29.9-760 -14.7 15.0 0.3 2100 0.02 21-29.5-750 -14.5 15.0 0.5 3425 0.03 34-26.6-675 -13.1 15.0 1.9 13425 0.13 134-23.6-600 -11.6 15.0 3.4 23424 0.23 234-20.7-525 -10.2 15.0 4.8 33424 0.33 334-17.7-450 -8.7 15.0 6.3 43423 0.43 434-14.8-375 -7.3 15.0 7.7 53423 0.53 534-11.8-300 -5.8 15.0 9.2 63423 0.63 634-8.9-225 -4.4 15.0 10.6 73422 0.72 734-5.9-150 -2.9 15.0 12.1 83422 0.82 834-3.0-75 -1.5 15.0 13.5 93422 0.92 934 0.0 0 0.0 15.0 15.0 103421 1.02 1034 Table 1b. Vacuum Pressure Guide with Liquid in the Lumens Vacuum Pressure (Gauge) Maximum Membrane TMP Maximum Liquid Pressure (Gauge) inhg mmhg (Torr) psi psi psi Pa atm mbar -29.9-760 -14.7 45.0 30.3 208943 2.06 2089-29.5-750 -14.5 45.0 30.5 210268 2.08 2103-26.6-675 -13.1 45.0 31.9 220267 2.17 2203-23.6-600 -11.6 45.0 33.4 230267 2.27 2303-20.7-525 -10.2 45.0 34.8 240267 2.37 2403-17.7-450 -8.7 45.0 36.3 250266 2.47 2503-14.8-375 -7.3 45.0 37.7 260266 2.57 2603-11.8-300 -5.8 45.0 39.2 270265 2.67 2703-8.9-225 -4.4 45.0 40.6 280265 2.77 2803-5.9-150 -2.9 45.0 42.1 290265 2.86 2903-3.0-75 -1.5 45.0 43.5 300264 2.96 3003 0.0 0 0.0 45.0 45.0 310264 3.06 3103 2a. SWEEP 2b. VACUUM 2c. SWEEP w/ VACUUM General Start-Up Instructions Figure 2. Liquid Contacting Operating Modes Liquid in the Lumens NOTE: Maximum TMP applies at startup also. Some users fail to ensure that the maximum TMP is not exceeded at startup when only one fluid is flowing, or if valves, flowmeters, and pumps are not set appropriately. Use pressure gages wherever possible and start slowly!! 1. Connect all the lines to the module according to the chosen mode of operation as shown in Figures 1 and 2. MedArray, Inc. 3915 Research Park Dr., Suite A-4 Ann Arbor, MI 48108, USA Page 3

2. Slowly introduce liquid into the system ensuring that the liquid pressure and flow rate do not exceed the maximum operating limits shown in Table 2. 3. Adjust the liquid inlet flow rate and pressure to targeted set points by adjusting the valves in the system. 4. Choose desired mode of operation: a. Sweep: i. Set the sweep gas pressure and flow rate entering the module to the desired levels. A typical sweep gas pressure is <15psig with flow rates from 0.1 to 1scfm / l/min of liquid flow rate. These are just recommendations and optimal levels should be determined experimentally. b. Vacuum only: i. Start vacuum pump per the manufacturer s instructions. ii. Open the valve connecting the module to the vacuum pump. Depending on the pressures and resulting flow configuration, the vacuum can be pulled through one or two ports. If single port operation is desirable, block the top port and use the bottom port for the vacuum line. iii. Adjust vacuum pressure as necessary. c. Sweep with vacuum: i. Set the sweep gas pressure at the module inlet to 1psig or less. ii. Set the sweep flow rate to the desired levels. A typical sweep flow rate used with vacuum is usually 5 to 10 smaller than outlined in the section a when only sweep gas is used. iii. Apply vacuum as described above in section b. Table 2. PermSelect Module Operating Limits Membrane Area Module m 2 (ft 2 ) 0.25 (2.69) PDMSXA-2500 0.75 (8.07) PDMSXA-7500 1.0 (10.76) PDMSXA-1.0 2.1 (22.6) PDMSXA-2.1 Typical Liquid Flow L/min (gpm) 0.2 1.9 (0.05 0.5) 0.5 6 (0.125 1.6) 1 7.5 (0.25-2) 2.75 19 (0.75-5) Maximum Lumen to Shell TMP bar (psi) 3 (45) @ 25 C (77 F) Maximum Shell to Lumen TMP bar (psi) 1 (15) @ 25 C (77 F) Maximum Lumen Pressure bar (psia) 3 (45) 2 (30) Maximum Shell Pressure bar (psia) 3 (45) 2 (30) Sweep Flow Rate (no Vacuum) L/min (scfm) 0.5 6 (0.02-0.2) 3 15 (0.1 0.5) 3 23 (0.1 0.75) 5 30 (0.2-1) Sweep Flow with Vacuum m 3 /hr (scfm) 0.1 1.2 (0.004 0.04) 0.6 3 (0.02 0.1) 0.6 4.5 (0.02 0.15) 1-6 (0.04 0.2) Other Notes o It is highly recommended to use 10µm or smaller prefilter on the incoming liquid. o Prior to use, flush and rinse all lines to and from the module before introducing the liquid inside the PermSelect Module. o Follow cleaning recommendations as necessary. o If using compressed air for sweep, be sure it is oil free. A 3µm prefilter is recommended for any gas used inside the module. MedArray, Inc. 3915 Research Park Dr., Suite A-4 Ann Arbor, MI 48108, USA Page 4

o In rare instances, a reduction in performance may be caused by condensation build up (droplets may be visible in such a case). In such an instance, remove PermSelect module and dry it with sweeping air at 50 o C (122 o F) on both sides of the fiber, especially on the side where sweep gas is used or vacuum applied. Technical Support Questions related to PermSelect Module use and operation? Call technical support at 734-769-1066. Cleaning Following CIP procedure is a general membrane cleaning protocol and depending on the application may not be optimal (for example if chemicals are not allowed in the system). If you are having problems restoring membrane performance with cleaning or would like to evaluate a different cleaning protocol, please contact technical support. For the best results DI water or softened water should be used for CIP, otherwise use tap water. 1. Initial Rinse flush liquid side of the membrane with water at ambient temperature for 5 minutes in once through mode water goes to the drain after exiting module. 2. Alkaline Cycle recirculate 2-5% w/w caustic (NaOH) at 122 o F (50 o C) for 30 minutes. 3. Rinse Cycle flush with water at ambient temperature in once through mode until neutral ph is reached (<5minutes). 4. Acid cycle recirculate 2-5% w/w citric acid at 122 o F (50 o C) for 30 minutes. 5. Final Rinse - flush with water at ambient temperature in once through mode until neutral ph is reached (<5minutes). 6. Purge membrane with gas (air, N2 or CO 2 ) at high flow rate to remove remaining water droplets. 7. Dry the membrane using warm gas such as air or nitrogen at 40-80 o C (104-176 o F) for minimum of 1-2 hrs before returning to operation. Higher drying temperature will result in shorter drying times. Handling Handle the module with care to avoid hits and sudden shocks from shaking or rapid movement, which can potentially cause internal damage. Careful handling will ensure optimal performance. Storage It is recommended to store modules in a dry, heat-sealed plastic bag at temperatures above 2 o C (35 o F) and below 50 o C (122 o F). Storing a module inside the original opaque box is important to prevent UV damage. If storing a module in the transparent plastic sealed bag outside of an opaque box, it must be away from direct sunlight. In order to avoid possible contamination, it is highly recommended for the modules to remain inside the plastic bag until use. Shelf life Silicone is known for its stability and if properly stored it will retain its physical properties such as tensile strength and modulus of elasticity for many years. MedArray, Inc. 3915 Research Park Dr., Suite A-4 Ann Arbor, MI 48108, USA Page 5