Series Ruska Digital Pressure Indicator. GE Sensing & Inspection Technologies. Features

Similar documents
Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications

Additel 780 Series Pressure Controller

Additel 761 Automated Pressure Calibrators

Additel 780 Series Pressure Controller

Additel 761 Automated Pressure Calibrators Selection Guide

Additel 761 Automated Pressure Calibrators

Pressure calibration. 20

Additel 761 Automated Pressure Calibrators Selection Guide

Additel 761 Automated Pressure Calibrators

Additel 780 Series Pressure Controller

PPC4E Pressure Controller/Calibrator Superior rangeability and reliability

DPI 610/615 Series. Druck Portable. Pressure Calibrators. GE Sensing. Features. Setting the Standard for Portable. Pressure Calibrators

ADT 761 Automated Pressure Calibrator

Pneumatic high-pressure controller Model CPC7000

Pneumatic High-Speed Pressure Controller Model CPC3000

Pneumatic high-pressure controller Model CPC7000

The PACE6000. Modular Pressure Controller/Indicator. GE Measurement & Control Solutions. Features

PACE1000 Precision Pressure Indicator

PACE Modular Pressure Controller

P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features

Industrial pressure controller Model CPC4000

628 Differential Pressure Decay Leak Tester

High-speed pressure sensor Model CPT6140

Industrial pressure controller Model CPC4000

Additel. Overvie. vacuum. Standalone. to 200 bar. products for. to use. The establishes. the future. Calibration Experts EN ISO / IEC

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

Digital deadweight tester Model CPD8500

Precision digital pressure gauge Model CPG1000

Hand-held pressure calibrator with integrated pump Model CPH6600

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

ADT. Series. oller. Overvie. vacuum Removable. Standalone. Easy-to-use. products for. to use. The establishes

P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features

Modular pressure controller Model CPC6050

Precision pressure sensor Basic version Model CPT6020

FPG8601 Force Balanced Piston Gauge


HumiSys HF High Flow RH Generator

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

DO NOT PRINT THIS PAGE

Pressure & Temperature Calibration Equipment

Digital dead-weight tester Model CPD8500

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. Type: ETd 04. class 0.05

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

Pressure Calibration Solutions

PPC2+ BG0002. Operation and Maintenance Manual (Use with PPC2+ Operation and Maintenance Manual) 2001 DH Instruments, Inc.

Automatic Isokinetic Sampler. ST5 Evo

Pressure Calibrators. Series PC..., PM...

DPC-30 DPC-100. Reference Manual

Pressure Technical Information

F&J SPECIALTY PRODUCTS The nucleus of quality air monitoring programs

The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions

Automatic Isokinetic Sampler. ST5 Evo

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

E-DWT-H Electronic Deadweight Tester

Digital pressure gauges

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

WTSeries C R Y S T A L. Digital Pressure Gauge. Operation Manual. engineering corporation

PPC4 Pressure Controller/Calibrator

Cecomp. Electronics. Test. Measure. Monitor. Transmit. Alarm. Cecomp High Perfor. formance Digital Pressur

Datasheet TM40. HUBER INSTRUMENTE AG, Grellingerstrasse 23, CH-4208 Nunningen. All rights reserved Paige 1 / 5 Datasheet_TM40_

Pressure Calibration Equipment

PTG100 Precision Test Gauge

UNIK 5600/5700 Marine Certified Pressure Sensing Platform

陕西创威科技有限公司. Shaanxi Creat Wit Technology Co., Ltd. Precision pressure calibrator-cwy300 Operating Manual

Digital Pressure Gauge Main features: Data Sheet: BIT02B.417.R6.EN COVER on request

2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters. Kent-Taylor

PPC4 Pressure Controller/Calibrator Operation and Maintenance Manual

DFX II Series. Digital Force Gauge

Design Features. General Description

Pneumatic service kit Model CPG-KITP

PG9607 Piston Gauge. Automated primary pressure reference for absolute and gauge pressures to 500 kpa

Pressure Calibration. Technical Data. Pressure Calibration with Fluke Calibrators

GP1 & GP2. Electropneumatic Regulators FOR PRESSURE CONTROL TO 1,000 PSI

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

Fluke 700 Series pressure modules

PPC2 AF. Operation and Maintenance Manual

Pressure transmitter PASCAL Ci4

86BSD Backside Digital Output

High Precision Gauge Pressure Indicators

Definition - Pressure

BAPI Pressure Line of Products - FAQs

Pressure & Vacuum Measurement. Series 370. Solutions. Stabil-Ion Vacuum gauge and Controller

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

Selecting the right pressure sensor for your application

molbox1+ Flow Terminal ± % of reading lowest uncertainty for gas flow calibration

RTX 1000H Series. Versatile Transmitters for a World of Pressure. GE Measurement & Control. Features

SPD Pressure Sensor Families

Pressure Automated Calibration Equipment

Pressure Calibration Equipment

SX150A. Features. Typical Applications. Description

Electronic Measuring and Calibration Instruments

AIR FLOW CALIBRATORS Single Sensor Designs

6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

EPP4 Pressure Regulator Basic G 1/4"

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

Transcription:

GE Sensing & Inspection Technologies Series 7050 Ruska Digital Pressure Indicator Features Pressure ranges from 0 to 10 inh2o and 0 to 1500 psi (0 to 25 mbar and 0 to 100 bar) Choose from three models: 7050i, 7050 and 7050LP Model 7050i provides advanced precision of 0.005% of reading Model 7050 provides 0.003% of full scale precision The Series 7050 digital pressure indicator expands upon the popular line of Series 7000 products such as the Series 7250 digital pressure controller and Series 7252 dual output pressure controller. The 7250 and 7252 provide both pressure measurement and control, and the 7050 provides an instrument for applications that only require pressure measurement. Stability: 0.0075% of reading per year Active matrix color screen with enhanced navigation menus

Three models are available to meet a wide range of applications: Model 7050i with advanced percent of reading precision Model 7050 with a high performance-to-price ratio Model 7050LP for low pressure measurements The Model 7050i, 7050 and 7050LP all use GE s unique quartz sensor, the most accurate pressure sensing technology available. Each quartz sensor is manufactured and tested to provide the ultimate performance required by a GE pressure calibrator, ensuring that every customer receives quality, precision and long term stability in their instrument. Advanced Precision The Model 7050i offers advanced percent of reading precision for increased capability with a single instrument, reducing the investment required to measure a wide pressure range. This model provides 0.005% of reading precision from 25% to 100% of the instrument s range. For pressures below the lower threshold of 25%, the 7050i precision is 0.005% of the lower threshold value. For example, a 100 psi (7 bar) Model 7050i provides 0.005% of reading from 25 to 100 psi (1.7 to 7 bar); the precision for pressures from 0 to 25 psi (0 to 1.7 bar) is 0.005% of 25 psi (1.7 bar). This unmatched precision is achieved through unique quartz pressure sensing technology and multiple quartz sensors in a single instrument. Various full scale pressure ranges from 0 to 5 to 0 to 1500 psi (0 to 400 mbar to 0 to 100 bar) are available. For absolute mode operation, select either the barometric reference option, or the vacuum reference option, which requires an external vacuum pump connected to the reference port. The latter features an on-board vacuum sensor that allows automatic zeroing in absolute mode. Permanent absolute ranges to 50 psia (4 bar) are also available. Series 7050 features GE s unique fused-quartz sensor. This rugged transducer offers unequalled precision and a stability of 0.0075% of reading per year. Model 7050 i Precision Stability (One year) Calibration Standard Environmental: Temperature (included in precision) Head Pressure Uncertainty (2 sigma) 0.005% of reading 0.0075% of reading 0.0010% of reading 0.000% of reading 0.001% of reading Expanded Uncertainty (2 sigma) 0.009% of reading Standard Precision For applications that do not require the level of performance provided in the Model 7050i, the Model 7050 offers an economical approach to high-accuracy pressure measurement, with a precision of 0.003% of full scale. Various ranges from 0 to 5 to 0 to 1500 psi (0 to 400 mbar to 0 to 100 bar) are available. For absolute mode operation, select either the barometric reference option, or the vacuum reference option, which requires an external vacuum pump connected to the reference port. The latter features an on-board vacuum sensor that allows automatic zeroing in absolute mode. Permanent absolute ranges to 50 psia (4 bar) are also available.

Automating Pressure Test and Calibration The 7050i, 7050 and 7050LP are provided with both an RS232 and IEEE-488 interface, and Series 7050 syntax follows SCPI protocol for easy programming. As a standard feature, software written for GE s previous generation Series 7215, 7010, 7000 and 6000 instruments is fully supported by the Series 7050. Intecal, an off-the-shelf software package is available. The Series 7050 features multi-lingual menus and displays. Low Pressure 7050LP The Model 7050LP is a specially configured instrument for low pressure applications and is available in three different range combinations with two ranges in a single instrument: 10 inh2o and 30 inh 2 O (25 and 75 mbar) 20 inh2o and 60 inh 2 O (50 and 150 mbar) 35 inh2o and 100 inh 2 O (87 and 250 mbar) The 7050LP provides a precision of 0.005% of reading from 10% to 100% of the maximum range. For example, the 10/30 in H 2 O 7050LP provides a precision of 0.005% of reading when measuring pressure from 30 to 3 inh 2 O (75 to 7.5 mbar). Changing ranges is automatic both upscale and downscale and does not require operator intervention. Since the 7050LP utilizes a true differential sensor, the reference port of the 7050LP can be connected to the test system eliminating unwanted pressure disturbances due to room drafts caused by HVAC systems. Long Term Stability All three models not only provide unequalled precision, but also excellent long term stability of 0.0075% of reading per year due to the inherent properties of quartz. A free download of the LabVIEW driver is available at www.gesensing.com. Firmware updates can be performed over the RS232 interface (updates can be downloaded from www. gesensing.com) A MET/CAL driver is also available as an option. Versatility The Series 7050 is versatile enough to handle almost any type of pneumatic pressure measurement application. Wide pressure range available in a variety of standard or custom full scale pressure ranges from 10 inh2o to 1500 psi (25 mbar to 100 bar) Pressure units/scales select from over twelve standard units of measure, including inhg at 0 C and 60 F, kpa, bar, psi, inh2o at 4 C, 20 C and 60 F, kg/cm2, mmhg at 0 C, cmhg at 0 C, and cmh2o at 4 C, and two user-defined units Head pressure automatic correction for head pressure differences. Absolute mode The 7050i and 7050 offer three different methods to make absolute pressure measurements. The barometric reference option provides the most convenient method and is available on ranges 15 psi (1.0 bar) and higher. Alternatively, the vacuum reference option allows the connection of an external vacuum pump to the reference port of the instrument. An onboard vacuum sensor monitors the reference vacuum and allows for automatic zeroing in absolute mode. This option provides the lowest overall uncertainty since it does not include the additional uncertainty of a secondary barometric reference sensor. For pressures to 50 psia (4 bar), permanent absolute models are also available.

Pressure limits Set upper and lower pressure limits to sound an audible alarm. Automatic zeroing All models feature automatic zeroing, including units with the vacuum reference option, for automated absolute mode zeroing through the front panel or over the PC interfaces. Options The following options are available for the Series 7050: Vacuum (negative gauge) mode for bidirectional measurements Barometric reference for absolute mode operation with the 7050i and 7050; ranges 15 psi (1.0 bar) and higher Vacuum reference for absolute mode operation via an external vacuum pump connected to the reference port for 7050i and 7050 Permanent absolute ranges to 50 psia (4 bar) full scale, which include a tare feature for simulated gauge mode operation NVLAP-accredited calibration report Intecal software for Windows The Series 7050 features an easy to navigate menu structure with full text descriptions for menus and commands. The large color display allows the pressure value to be displayed even when viewing a submenu selection such as the units selection screen shown above. The Series 7050 digital pressure indicators provide high performance pressure measurement with a wide variety of pressure ranges and options. All are easy to use, easy to maintain, and have the reliability, performance and features that you want.

Series 7050 Specifications Pressure Ranges Standard Pressure Ranges Model 7050i Select any full scale from 5 to 1500 psig (400 mbar to 100 barg). Or, permanent absolute ranges are available from 15 to 50 psia (1 to 4 bara). Model 7050 Select any Full Ccale pressure range from 5 to 1500 psig (400 mbar to 100 barg). Or, permanent absolute ranges are available from 15 to 50 psia (1 to 4 bara). Model 7050LP 10/30 inh 2 O (25/75 mbar) 20/60 inh 2 O (50/150 mbar) 35/100 inh 2 O (87/250 mbar) Optional Modes Absolute using barometric reference sensor for ranges from 15 to 1500 psig (1.0 to 100 barg) Absolute using vacuum reference option for ranges 5 to 1500 psig (400 mbar to 100 barg) Negative gauge Performance Precision Model 7050i From 25% to 100% full scale (FS): 0.005% of reading Below 25% FS: 0.005% of 25% FS Model 7050 0.003% of FS Model 7050LP From 10% to 100% of maximum FS, 0.005% reading, Below 10% maximum FS: 0.005% of 10% maximum FS Stability All models: 0.0075% of reading per year Display Resolution User selectable to 1:1,000,000 Negative Gauge Precision (optional) Model 7050i 0.005% of 25% FS or 0.0005 psi (0.034 mbar), whichever is greater Model 7050 0.003% of FS Model 7050LP 0.005% of reading from 10% to 100% maximum FS Below 10% maximum FS 0.005% of 10 maximum FS Barometric Reference (optional) 0.002 psi (0.1379 mbar) maximum error per year Vacuum Reference (optional) 0.0002 psi (0.014 mbar) maximum error per year Calibration A calibration report with traceability to National Institute of Standards and Technology (NIST) is provided. An ISO-17025 accredited calibration is available.

Total Uncertainty The maximum deviation from the true value of pressure including precision, stability, temperature effects and the calibration standard is: Model 7050i 25% to 100% FS 90 day: 0.006% reading 1 year: 0.009% reading Model 7050 Ranges to 1500 psi (100 bar) 90 day RSS: 0.003% FS + 0.002% of reading 1 year RSS: 0.003% FS + 0.0075% of reading Model 7050LP 10/30 inh2o (25/75 mbar) RSS: 0.005% and 0.0076% of reading per year down to 10% of maximum FS, Below 10% RSS: 0.00015 inh 2 O (0.000375 mbar) and 0.0076% of reading per year 20/60 inh2o (50/150) RSS: 0.005% and 0.0076% of reading per year down to 10% of maximum FS, Below 10% RS: 0.003 inh 2 O (0.00075 mbar) and 0.0076% of reading per year 35/100 inh 2 O (87/250 mbar) RSS: 0.005% and 0.0076% of reading per year down to 10% of maximum FS, Below 10% RSS: 0.0005 inh 2 O (0.00125 mbar) and 0.0076% of reading per year Communications RS232 and IEEE-488, SCPI syntax. Series 7250, 7215, Model 7000 and Series 6000 emulation are standard. MET/CAL drive optional LabView drive download from www. gesensing.com Firmware updates are performed via RS232 interface download from www.gesensing.com Languages The Series 7050 is capable of displaying menus and functions in: English, French, Chinese, German, Japanese, Spanish and Italian Options Barometric reference (absolute and negative gauge) Vacuum reference* (absolute) Negative gauge only MET/CAL driver Intecal software NVLAP-accredited calibration Rack mount kit Liquid trap assembly *Requires external vacuum pump General Display TFT, VGA, active matrix, 6.4 in (162.5 mm) 640 x 480 resolution, 65,000 colors Temperature Operating: 64 F to 96 F (18 C to 36 C) Storage: -4 F to 158 F (-20 C to 70 C) Humidity 5% to 95% relative humidity, non-condensing Dimensions (h x w x d) All versions: 7 in x 16.5 in x 19 in (178 mm x 419 mm x 483 mm) Weight All models: 16 lb (7.5 kg) Power 90 to 260 VAC, 50/60 Hz, 150W Test Port and Reference Connection 1/4 in NPT female

Warm Up Time 2 to 3 hours; may be left on indefinitely Pressure Medium Nitrogen or clean dry air Precision is defined as the combined effects of linearity, repeatability and hysteresis throughout the operating temperature range. Expression of total uncertainty conforms with the recommendations of the ISO Guide to the Expression of Uncertainty in Measurement. Other Products and Services In addition to a wide range of digital pressure controllers and indicators from 1 to 40,000 psi (25 mbar to 2750 bar), GE manufactures primary standard piston gauges from 0.2 to 72,5000 psi (14 mbar to 5000 bar).

NVLAP LAB CODE 200491-0 www.gesensinginspection.com 920-144B 2009 General Electric Company. All Rights Reserved. Specifications are subject to change without notice. GE is a registered trademark of General Electric Company. Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective companies, which are not affiliated with GE.