Operating Manual. MELAseal 100+ Bar sealer. Dear Dr.

Similar documents
Operating Instructions. Ball valve fitting according to ZB For pressure transmitter VEGABAR 82. Document ID: 50027

Digital Melting Point Apparatus

Type 4708 Supply Pressure Regulator

Theatre Control Panel. Instructions for Use. January 2017 ASF-29 (1)

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25

Instructions for Use. Plug connection for medical gas and vacuum acc. to DIN / 2004 DIN EN ISO

CJMJ.DPNQBDU. Instruction for use. Neonatal phototherapy device. BILI-COMPACT WY 1816/ Year of manufacture D2017_00

Gas plug-in hose, straight-through form. Instructions for Use. suitable for gas socket model G2019T, G2016T. G2023 from 01/1990.

Product instruction manual Easymount Sign Laminators EM-S1400C, EM-S1600C, EM-S1400H, EM-S1600H

KGW - ISOTHERM. Operating Instructions. Level controlling device for liquid nitrogen LEVEL CONTROL LN2

A17W49_MASEM_F_EN MANUAL

Safety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14

Type 4708 Supply Pressure Regulator

Siphon vessel / Priming aid

DOLOMITE Legacy. Rollator User Manual

MODELS (2 foot Version) (4 foot Version)

Model 130M Pneumatic Controller

User Manual PTW-DensiX. D / Di/Zi

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

KERN EG/EW Version /02

Operating instructions - Translation of the original -

The Univentor 1250 Anaesthesia Unit

KTM OM-2 SPLIT BODY FLOATING BALL VALVES INSTALLATION AND MAINTENANCE INSTRUCTIONS

INSTRUCTION MANUAL. Slavia 630/631

SG36KTL-M Quick Installation Guide. 1 Unpacking and Inspection

INSTRUCTION MANUAL CZ 630/631

Installation and Maintenance Instruction Manual

Mounting and Operating Instructions EB EN. Type Supply Pressure Regulator. with increased air capacity

OPERATORS MANUAL MODEL 365 BACK LAPPING MACHINE WARNING

icreasepro Creaser Operators Manual

BMR 2100XL Operating Instructions Page 1 of 13

Operating Instructions

Cleaning rod: spring steel, stainless steel or carbon fibre cleaning rod - only use a one-piece rod. Avoid using snakes.

Instruction Manual Contact Pressure Vacuum Gauge

Poollift Delphin. Version: February, 2015 MANUAL AND OPERATING INSTRUCTIONS

Technical Data Sheet TI-F50 Locking Units series KFH

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

LIGHTWEIGHT TRIWALKER OWNER S HANDBOOK

3 Post Pressure Fit System Owner s Manual

Squeegee Unit for MRS Multi-Recovery System

Type Operating Instructions. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies.

RayPette Plus Autoclavable Pipette. User Manual

FPD20 Series. Gear wheel flow sensor

Mounting and Operating Instructions EB 8546 EN. Supply Pressure Regulator Type Fig. 1 Supply pressure regulators

INSTALLATION INSTRUCTIONS PUSH BAR PULLMAN PANIC BOLT DEVICE TO EN 1125 : 2008

Overspeed governors (113 OG Моment 250 MR, 114 OG Moment 250 A3, 115 OG Moment 250 MRL, 119 OG Moment 250 MRL/1)

Supratec Gesellschaft für Umweltund Verfahrenstechnik mbh OXYFLEX - M T / M T M E M B R A N E D I S C D I F F U S ER AERATION TECHNOLOGY

contents introduction

Auto-Rewind Hose Reels INSTRUCTION MANUAL FOR OXY-LPG MODEL

Type BBS-03, BBS-05, BBS-06, BBS-25

Supplementary instructions. Rod and cable components. for VEGAFLEX series 80. Document ID: 44968

PPS PORTABLE POWER SUPPLY

REMS Swing INSTRUCTION MANUAL SINGLE-HAND TUBE BENDER SWING ANCRA

BR 27g. Operating, assembly and maintenance instructions for discontinuous sampling valve. 1. Design, operation and dimensions:

e.do Gripper Safety requirements, technical features and assembly and integration instructions

Type Pressure Regulator. for increased air capacity. Fig. 1: Type Pressure Regulator. Mounting and Operating Instructions EB EN

Mounting and Operating Instructions EB 2555 EN. Self-operated Pressure Regulators. Pressure Reducing Valves Type 50 ES Type 50 EM.

Inverted Bucket Steam Trap IB 16A-7

Pectoral Machine. User manual E S S E N T I A L S T R E N G T H

Instruction Manual AC350 GUIDED-TYPE FALL ARRESTER ON RIGID ANCHORAGE LINE EN 353-1:2002

The pipettes cover a volume range from 0.1µl to 10ml.

Instructions for Installation, Use and Maintenance FSI Safety Tank Shower Systems

8MAY15 US RACK, Inc Falcon Drive, Madera, CA

A18W35_MASE 130_F_EN MANUAL

Propane Conversion Kit Instructions

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer

Instruction manual. for. high purity gas pigtails 200 bar / 300 bar

Data Distribution Cabinets. Exclusive CD edition March 2006

Actuating elements for ComEx

Operation Manual High Speed Knife Mill HOM , Hagavish st. Israel Tel: , Fax: MRC.10.

12V MINI AIR COMPRESSOR

testo Leakage detector for refrigerants Instruction manual

Technical Information Conducal CLY421

ROLLATOR (R6 and R8) OWNER S HANDBOOK

Contents. Operating instructions. Pressure switch, heavy-duty version Model PSM-520

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar

Operating Instructions. ATMOS Chair Comfort. English GA1GB Index 01

Operation Manual. PEGAS NovoTap, PEGAS NovoTap+

Float Operated Level Controllers

SG33KTL-M Quick Installation Guide. 1 Unpacking and Inspection

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

HAI UPB Wall Switch and Auxiliary Switch Installation and Operating Instructions

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

EXTERNALLY ROUTED DROPPER POST owner S MANUAL

OPERATING INSTRUCTIONS

MANUAL SEALLESS STEEL STRAPPING TOOL MODEL A

User Manual. Quantos Automated Dosing Liquid Module

Discussion and guidance on the definition and qualification of porous loads

Agilent 1220 Infinity II LC Mobile Upgrade Kit

SAUTER FK Version /2016 GB

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

Operating Instructions VETTER Industrial Lifting Bags 8 bar - Type IKV

Model 405 &407 Small Animal Ventilator. User Manual v3.5

NovelBright LED Illuminator

OPERATION MANUAL PNEUMATIC VALVES

RO LLATO R (R6 and R8) O W NER S HANDBO O K

Instructions for High Altitude Conversion

PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100

Transcription:

Operating Manual MELAseal 100+ Bar sealer Dear Dr. We should like to extend our thanks for the expression of trust in our company which you have displayed through the purchase of this MELAG product. As an owner-run and operated family concern founded in 1951, we have a long history of successful specialization in hygiene products for practice-based use. Our focus on innovation, quality and the highest standards of operational reliability has established MELAG as the world s leading manufacturer in the instrument treatment and hygiene field. You, our customer are justified in your demand for the best products, quality and reliability. Providing "competence in hygiene" and "Quality made in Germany", we guarantee that these demands will be met. Our certified quality management systems is subject to close monitoring: one instrument to this end is our annual multi-day audit conducted in accordance with ISO 13485 and ISO 9001 conducted by the certification agency BSI. This guarantees that all MELAG products are manufactured and tested in accordance with strict quality criteria. The MELAG management and team.

General Guidelines Please read these operating instructions carefully before commissioning the sealing device. The instructions include important safety information. The functionality and value-retention of this autoclave depend on the care accorded to it. Make sure to keep the Operating Manual near to the device. It represents a component of the product. User Group This manual is targeted at doctors, medical assistants and service. Symbols used Symbol Explanation Indicates a dangerous situation which if not avoided could entail slight to life-threatening injuries. Indicates a dangerous situation which if not avoided could entail damage to the instruments, the practice equipment or the sealing device. Draws your attention to important information. Formatting rules Symbol OPEN see Chapter 2 Explanation Words or phrases appearing on the display of the sealing device are marked as display text. Reference to another text section within this manual. Fig. 1/(3) Reference to a detail in a figure in the example, to part no. 3 in figure 1.

Contents Contents Chapter 1 Performance Specifications... 5 Proper use... 5 Chapter 2 Device Description... 6 Views of the device... 6 Control lamp status display and acoustic signals... 7 Chapter 3 Initial Commissioning... 8 Transport and storage... 8 Conditions of storage and installation... 8 Requirements of the set-up location... 8 Wall mounting... 8 Connecting the sealing device... 9 Commissioning... 10 Simple sealing procedure... 10 Sealing procedure for rollstock... 11 Seal seam width... 13 Sealing temperature... 13 Chapter 4 Optional Accessories... 14 Reel dispenser standard... 14 Reel dispenser Comfort and Deluxe... 14 Wall-mounted reel dispenser... 15 Chapter 5 Maintenance... 16 Cleaning and regular controls... 16 Replacing the blade... 16 Manufacturer's Recommendation for Routine Operation... 18 DIN Specifications... 20 Terms... 20 General information regarding the packaging and sealing procedure... 20 Seal seam width... 20 Strength of seal seam... 20 Accessories and Replacement Parts... 21 Technical Data... 21

Chapter 1 Performance Specifications Safety Instructions When operating the sealing device, please ensure that you observe the following safety instructions as well as those contained in following chapters. Power cable and mains socket Only the power cable included in the scope of delivery may be connected to the sealing device. Liquids may not be permitted to reach the interior of the sealing device. This could result in an electrical shock or short circuiting. Repair Never open the housing of the sealing device. Incorrect opening and repair can compromise electrical safety and pose a danger to the user. The guarantee and warranty are forfeited as soon as the sealing device is opened by anyone other than a member of a MELAG-authorized technical customer service. 4

Chapter 1 Performance Specifications Chapter 1 Performance Specifications Proper use This sealing device is designed for application in a medical context, e.g. clinics and medical and dental practices. It was developed for the heat sealing of instruments in sterilization packaging. Suitable materials It is suitable for the heat sealing of transparent sterilization packaging according to DIN EN 868-5 such as e.g. MELAfol. Should you wish to use packaging materials other than those specified above, please consult your stockist first or contact MELAG directly. Unsuitable foils and materials Tubular film (double-sided film) is not suitable for use in this sealing device as these types tend to become glued together on the sealing rail, thus restricting the functionality of the sealing device. Polyethylene film Soft PVC film Hard PVC film Polyamide film Polypropylene film PLEASE NOTE The use of unsuitable packaging materials carries the risk of damage to or malfunction of the device. Please observe the manufacturer information regarding the respective packaging materials as well as the recommended sealing temperatures. 5

Chapter 2 Device Description Chapter 2 Device Description Scope of delivery Please check scope of delivery before connecting the sealing device. Standard scope of delivery MELAseal 100+ sealing device Power cable Operating Manual Certificate of Conformity Warranty certificate Lever Torx key for housing screws Views of the device Fig. 1: Views of the device (1) Knife handle (2) Control lamp (3) Rotary knob for temperature adjustment (4) Square hole for lever (double-sided) (5.1/5.2) Reel dispenser comfort (6) Power cable socket (7) Lever (8) Power switch (ON/OFF) 6

Chapter 2 Device Description Control lamp status display and acoustic signals Table 1: Explanation of the warning lights and the acoustic signals Warning lights/acoustic signals illuminated yellow Flashes red, warning signal Illuminated green Flashes green illuminates red, warning signal (malfunction) Possible cause The sealing device is currently warming up or cooling down The lever is depressed during the heating phase, The sealing temperature has not yet been reached. The sealing device has reached the pre-set sealing temperature and is ready for operation. The pre-set sealing time (3 sec) has been reached and the sealing procedure has been ended. The sealing procedure is running (3 sec) and the lever has been depressed. The lever has been raised early, despite the required sealing time not having been reached. The lever has not been raised, despite the required sealing time having been completed. The heating phase is taking too long (> 5 min.); the sealing device cannot reach the pre-set sealing temperature. Risk management Please wait until the pre-set sealing temperature has been reached. Please wait until the LED is constantly illuminated green. Raise the lever and remove the packaging. Please wait until the LED is constantly illuminated green. Please keep the lever depressed until the green LED is constantly illuminated. Lift the lever as soon as the sealing time has been reached so that the film is not burnt. 7

Chapter 3 Initial Commissioning Chapter 3 Initial Commissioning Transport and storage WARNING The use of unsuitable transport packaging can result in damage to the housing and the device interior. The sealing device should only be transported in its original or otherwise suitable packaging. Storage The device should be stored in such a way to protect against humidity. Conditions of storage and installation In accordance with current VDE specifications, the sealing device is unsuitable for operation in areas exposed to the danger of explosion. The sealing device is conceived for use outside patient surroundings. The device should be located at least 1.5 m away (radius) from the treatment area. Observe all the information contained in this chapter during commissioning. DANGER Damages to the sealing device and/or injuries can occur in case the installation conditions are not observed. Requirements of the set-up location Observe the following requirements: The site of setup is dry and dust-protected. Sufficient clearance to the surrounding surfaces is available in order to ensure sufficient ventilation. The sealing device is located away from direct sunshine and outside the range of other sources of heat The sealing device is protected against blows or vibration. Wall mounting If the device is not to be placed on a table, it can be mounted on a wall. If this is the case, we recommend using the wall-mounted roll bracket optionally available. Proceed as follows: 1. Remove the perforated metal wall-mounting metal panels from the base of the sealing device. 2. Drill two Ø 6 mm boreholes in the wall with a clearance of 16.5 mm at the desired mounting height. 8

Chapter 3 Initial Commissioning 16.5 cm 3. Insert two rawl plugs (6mm) with round-head screws (Ø3.5 x 45 mm) in the bore holes. 4. Hang the sealing device on the screws. Connecting the sealing device Connecting the sealing device Check the following before connecting the sealing device: The device power switch is off. The power cable included in the scope of delivery is used for connecting. 1. Connect the one end of the power cable with the port on the rear of the sealing device and the other end with a mains socket with an approved voltage supply (in accordance with the name plate on the rear of the sealing device). 2. Insert the lever in the square hole on the left or right-hand side of the device as required (Fig. 1/(4)). 9

Chapter 3 Initial Commissioning Commissioning Switching on the sealing device Activate the sealing device at the power switch. The LED at the front of the sealing device (Fig. 1/(2)) will illuminate yellow after switching on. As soon as the control lamp is continuously illuminated green, the pre-set sealing temperature has been reached and the sealing device is ready to operate. Simple sealing procedure DANGER The sealing bar is heated continuously after the sealing device has been switched on. The surrounding metal components are hot! Touching the hot metal surfaces can cause burns. Never touch the metal surfaces on the cutting bar and on the rear and fore paper guide directly. In order to perform a simple sealing procedure, proceed as follows: 1. Introduce the packaging in the paper guide with the foil side facing upwards and push the packaging through the pressure bar and sealing bar. Ensure maintenance of the correct clearance between instrument and seal seam (consult the relevant section of the DIN Specifications). WARNING The film side of the packaging must always face upwards. Should the packaging be inserted incorrectly, this could result in the deposit of film residue on the sealing rail. 10

Chapter 3 Initial Commissioning 2. Press the lever forwards to its fullest extent until it clicks. 3. Leave the lever depressed as long as the control lamp flashes green in short intervals (for 3 seconds). 4. When the control lamp is continually illuminated green again, return the leaver to its starting position. 5. The sealing procedure has been completed successfully. 6. After every successful sealing process carry out a visual inspection of the film. Sealing procedure for rollstock DANGER The sealing bar is heated continuously after the sealing device has been switched on. The surrounding metal components are hot! Touching the hot metal surfaces can cause burns. Never touch the metal surface on the cutting bar and on the rear and fore paper guide directly. Cutting the film If the instruments are packaged from the roll in transparent sterilization packaging bags, proceed as follows: 1. Introduce the film in the paper guide (lower slot) from the rear with the paper side facing downwards and push the packaging forward through the pressure bar and sealing bar until reaching the desired length. 2. Press the lever forwards to its fullest extent until it clicks. 11

Chapter 3 Initial Commissioning 3. Leave the lever depressed as long as the control lamp flashes green in short intervals (for 3 seconds). 4. To cut the film during the sealing process, move the knife handle to the other end of the sealing device quickly. The lever should remain depressed. Do not return the knife handle. Sealing film bags In order to seal the cut to length film bag, proceed as follows: 1. Place the instruments to be sterilized in the bag thus produced. Ensure maintenance of the correct clearance between instrument and seal seam (consult the relevant section of the (DIN Specifications). 2. Slide the open side of the bag from the front into the paper guide. WARNING The film side of the packaging must always face upwards. Should the packaging be inserted incorrectly, this could result in the deposit of film residue on the sealing rail. 12

Chapter 3 Initial Commissioning 3. Press the lever forwards to its fullest extent until it clicks. 4. Leave the lever depressed as long as the control lamp flashes green in short intervals (for approx. 3 seconds). 5. When the control lamp is continually illuminated green again, return the leaver to its starting position. 6. The sealing procedure has been completed successfully. PLEASE NOTE When sealing MELAfol sealing bags with lateral creases, please observe and comply with the special information contained in the sealing bag operating manual regarding sealing cassettes. Seal seam width The sealing device produces homogenous 10 mm wide seal seams with every sealing procedure. The recommended nominal size for the width of the seal seam in the German standard DIN 58953 part 7 is only 6 mm. Section 4.3.2 of DIN EN 868-5 requires a min. total seal width of 6 mm. The sum of the individual seams should amount to min. 6 mm. Sealing temperature Continuously variable temperature regulation is performed using the rotary knob (Fig. 1/(3)) on the lefthand side of the sealing device. The sealing temperature is determined by the type of sterilization packaging. When using MELAfol transparent sterilization packaging as provided by MELAG, the rotary knob should point to the middle area marked with MELAFOL (corresponds to 180 C). To reduce the sealing temperature, turn the knob leftwards in an anti-clockwise direction. To increase the sealing temperature, turn the rotary knob rightwards in a clockwise direction. 13

Chapter 4 Optional Accessories Chapter 4 Optional Accessories Reel dispenser standard The reel dispenser "standard is placed directly behind the sealing device. The rolls of film are placed into the cavity and held in position using additional spacers located to the left and right. This prevents them from slipping. The reel dispenser standard is hooked onto the brackets on the rear of the sealing device. Reel dispenser "standard" (Rear detail view) Reel dispenser Comfort and Deluxe The reel dispenser Comfort enables space-saving storage of the film rolls over the sealing device. The rolls are slid onto the rod laterally and held in position via additional spacers to the left and right. This prevents the roll from slipping and turning as well. In the roll dispenser "deluxe", the film rolls are stored above the sealing device, thus saving space. The integrated feeding device enables easy feeding of the film via a hand wheel. 14

Chapter 4 Optional Accessories The reel dispensers Comfort and Deluxe are hooked onto the brackets on the rear of the sealing device. Reel dispenser "Comfort"/ Deluxe (Rear detail view) Wall-mounted reel dispenser Mounted on the wall directly over the sealing device, the wallmounted roll dispenser saves space. 15

Chapter 5 Maintenance Chapter 5 Maintenance Cleaning and regular controls The outside of the sealing device should be cleaned at least twice a year. Only use a lint-free damp cloth stainless steel-cleaning agent suitable for use with medical products. It should not deposit any oily residue. Observe the following instructions when cleaning: Before every cleaning switch off the sealing device at the power switch and remove the power plug. The cleaning cloth may never be allowed to become entirely wet in order to prevent water from entering the interior of the sealing device. Replacing the blade DANGER Failure to observe this provision can result in a short circuit and/or an electric shock. Switch off the sealing device at the mains and remove the cable. DANGER Metal surfaces of the sealing device become hot during operation. Touching the hot metal surfaces can cause burns. Allow the sealing device to cool before opening. Proceed as follows to replace the blade: 1. Remove the 2 TORX screws (Torx 20) from the upper side of the sealing device on the cutting bar using the TORX screwdriver included in the scope of delivery. 2. Remove the upper rear housing cover rearwards. 3. Loosen the screws on the blade. 4. Turn the blade 180 in order to use the other side of the blade or replace it entirely. 16

Chapter 5 Maintenance 5. Return the screws. 6. Replace the unit housing and secure with the Torx screws. Press the fore housing cover against the rear housing cover in order to return the Torx screws. PLEASE NOTE When assembling the sealing device, ensure that the blade is able to move freely from side to side. 17

Manufacturer's Recommendation for Routine Operation Manufacturer's Recommendation for Routine Operation Frequency Check/action Criteria Daily before Visual check: operation Making and checking a sample seal seam The seal seam should be distinct and even and not contain any creases or defects. The sealing seam should be almost identical with the works sealing seam, i.e. an industrial seal with an identical edge in terms of colour. If the sealing seam is too light, the sealing temperature and/or the contact pressure (sealing force) is too low; If the sealing seam is brown, the sealing temperature is too high or the sealing duration too long Mechanical check Tear-resistance check: the strength required to open the seam must be exactly equal to that required to open the factory-produced seam Peel test 1) When peeled off, the paper should free itself from the film, leaving almost no residue Following every sterilization Batch-related check of the seam Control the entire packaging for dryness and integrity. Clearance after sterilization, documentation within the scope of process clearance Check for dryness and integrity once again, before using the instruments. Weekly Annually Checking the seal seam with aids Check the seal seam for tearresistance in accordance with DIN EN 868-5, Appendix D. Perform a visual control using MELAG seal check and document the results - for criteria and further information, see the MELAG seal check operation manual See the supplementary instructions of the MELAG seal seam strength test After 50,000 cycles or 2 years (MELAseal 200, MELAseal Pro) Upon malfunction messages or obviously faulty seals Maintenance Remedying the fault Perform the maintenance in accordance with the MEALG maintenance instructions, including the replacement of wear parts Do not operate a defective sealing device. Notify the authorized customer services 1) Performing the peel test: 1. Seal the sterilization packaging in the sealing device. 2. Place a sealed sterilization packaging in a sterilization cycle. 3. Slowly pull apart (by hand) the sealing seam in the direction of peeling. The visual check is to verify whether the sealing seam extends along the whole width and length. No paper residue greater than 10 mm is permitted on the seal seam. Document the results. 18

Manufacturer's Recommendation for Routine Operation MELAG seal seam strength test For validating your sealing processes MELAG provides a seal seam stability test for a cost price of 125.00 plus V.A.T. (state 01/2015) After the film test strips have been tested, MELAG will issue verification of testing and if successful, a certificate. This certificate confirms conformity of the seal seam with the standard DIN EN 868-5, Appendix D. Use the MELAG seal seam strength test application form (http://www.melag.de/en/service/downloadcenter/recommendations). 19

DIN Specifications DIN Specifications Terms Sterile barrier system DIN EN ISO 11607-2:2006 replaces the terms "packaging" "final packaging" and immediate packaging" with the single term "sterile barrier system." A sterile barrier system is the minimum level of packaging facilitating successful sterilization, serving as a micro-biological barrier and permitting aseptic provision. This includes transparent sterilization packaging, a sterilization bags, reusable containers etc. Protective packaging The protective packaging is designed to provide the sterile barrier system with protection up until its final application, Packaging system The sterile barrier system and protective packaging combine to form the packaging system. Peel test A procedure to determine the peeling characteristics of paper/plastic composite material in accordance with DIN EN 868-5, Appendix E. General information regarding the packaging and sealing procedure Please observe the following during packaging and sealing: Do not select undersized packaging. Packaging made of porous materials and plastic composite film should be filled with a max. of up to 3/4 of its volume (DIN 58953-7:2010). When using transparent sterilization packaging from a roll, the removal side must have an overlap of min. 1 cm between the cutting edge and the seal seam, enabling an aseptic removal (DIN 58953-7:2010). Pressing together should remove all air before sealing. Seal seam width Maintain the seal seam width stipulated in the standard: The width of the seal seam for sterilization packaging e.g. transparent sterilization packaging should amount to a min. 6 mm. With grooved seal seams, the sum of the individual seams should amount to min. 6 mm (DIN EN 868-5:2009). Strength of seal seam When using MELAfol transparent sterilization packaging, the sealing device guarantees a seal seam strength in accordance with EN 868-5. 20

Accessories and Replacement Parts Accessories and Replacement Parts Article Article no. Accessories Reel dispenser standard 00117 Reel dispenser Comfort 00111 Reel dispenser Deluxe 00108 Wall-mounted reel dispenser 00106 Distance plate (1 pcs) for reel dispenser "Comfort" or wallmounted reel dispenser 13330 Distance plate (1 pcs) for reel dispenser "Deluxe" 88110 Distance plate for reel dispenser "standard" (2 pcs.) 72330 MELAG seal check 01079 Replacement parts Knife blade 17781 Knife holder black (without blade) 45610 knife handle black 45600 Technical Data Model name MELAseal 100+ Device dimensions (WxDxH) Weight 41.5 x 24 x 15 cm 5.4 kg Electrical connection 220-240 V 50/60 Hz 100 110 V 50/60 Hz Power input 300 W Seal temperature range 160-200 C Seal seam width Seal seam length Sealing time Sealing force Overheating protection 10 mm max. 27.5 cm 3.0 sec. (factory settings, fix) ~100 N (factory settings, fix) >240 C 21

MELAG Medizintechnik ohg Geneststraße 6-10 10829 Berlin Germany E-Mail: info@melag.de Web: www.melag.de Responsible for content: Technical Office We reserve the right to technical alterations Your stockist: BA_EN_seal100+_v4.docx Rev.: 4 14/1917