YSI Environmental Monitoring Systems. Water Quality Training Seminar

Similar documents
QUICK START GUIDE TO DRINKING WATER MONITORING

Optical Dissolved Oxygen Meter

Optical Dissolved Oxygen Meter

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

STERILIZABLE DISSOLVED OXYGEN SENSORS

PRODUCT SHEET. Order probe only as RXPROBE02

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

6-SERIES ENVIRONMENTAL MONITORING SYSTEMS. Service Manual

Heavy Duty Dissolved Oxygen Meter

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

D-OptoLogger Dissolved Oxygen Logger Operation Manual

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Dissolved Oxygen Guide

LogIT Dissolved Oxygen Probe

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

6-Series Multiparameter Water Quality Sondes. User Manual. 6-Series: 6600 V2 6600EDS V V V2 600 OMS V2 600XL 600XLM 600LS 600R 600QS

Best Practice for Calibrating LTH Conductivity Instruments

Columbus Instruments

Dissolved Oxygen and measurement possibilities. Berno Lüpkes, 15 th March 2017

6-SERIES 6600 Sonde 6600 EDS Sonde 6920 Sonde 6820 Sonde 600XLM Sonde 600 OMS Sonde 600XL Sonde 600QS Sonde 600R Sonde 650 MDS Display/Logger

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

Gas Pressure Probe. Information Sheet. Total Dissolved Gas

BI-680 Online Dissolved Oxygen Controller Instruction Manual

User Manual DOCUMENT # Pro20 Pro20i USER MANUAL. English

Environmental Monitoring Systems 6-SERIES

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

PRO20 USER MANUAL. English. Français. Español. Deutsch

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

Radnoti Liver Perfusion System

Unisense In Situ UniAmp Connector System

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

Yeo-Kal Electronics Pty Ltd 18/26 Wattle Road, Brookvale NSW Australia Telephone Fax USER MANUAL

Measuring the Conductivity of Pure Water Using USP<645>

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

FLUORESCENCE DETERMINATION OF OXYGEN

CCS240 and CCS241. Technical Information. Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

Pool and Spa. Water Testing

Oxygen Measurement Systems High Reliability and Wide Application Coverage

Improve Process Reliability

SS32L Dissolved O 2 Probe

Oxygen Sensor with Electrode

Introduction The Smart Q Oxygen Adaptor enables an Oxygen electrode to be connected to an EasySense unit.

D-Opto Dissolved Oxygen Sensor Operation Manual

Dissolved Oxygen/Temp Waterproof Tester

Oxygen Measurement in Natural Gas

MODEL GT820 OXYGEN SENSOR

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

D-Opto. Dissolved Oxygen Sensor. Operation Manual (Software Version 1.0)

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

CB-150 Data Buoy Quick Start Guide

Aerobic Respiration. Evaluation copy

Refillable ORP/Redox Probe: Model MTC30101 or MTC30103

Best Practice Guide, Servomex 2700

HOBO U20 Water Level Logger (U x and U x-Ti) Manual

HI 2314 HI 2315 HI 23151

DISSOLVED OXYGEN SENSOR BT34i

New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters.

Pneumatic high-pressure controller Model CPC7000

Onset Computer Corporation Inside this package: HOBO U20 Water Level Logger Calibration Certificate

EC214 EC215 - EC215R Bench Conductivity Meters

Calcimeter Instruction Manual

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

Oxymax W COS51D. Technical Information. Digital sensor for measuring dissolved oxygen Membrane-covered amperometric sensor

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

NEULOG OXYGEN LOGGER SENSOR GUIDE

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications

Oxygen Dipping Probe PSt3 / PSt6 / PSt9

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

OxyProbe II Instruction Manual

Instruction Manual HI Portable Dissolved Oxygen Meter. DO meter HI 8043 STB C OFF O 2 ZERO SLOPE C. ins rum n s

Dissolved Oxygen Sensor

_ pressure transducers. User Manual

HOBO U20L Water Level Logger (U20L-0x) Manual

Water Samplers, Water Sampling with Internal Recording, Cabling, and Deployment

Salinity and Dissolved Oxygen (D.O.) Lecture 2

How to specify a product. Process Sensors and Mechanical Instruments

Blood Parameter Monitoring System 550

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

Overview. Front Panel: Keypad and Display

OUS31. Technical Information

Calibration and Maintenance Procedures: YSI ProDSS rev 8/31/15

Test Plans & Test Results

Three Lakes Association Hydrolab for Dummies. Table of Contents:

Application Note AN-107

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa)

T02.0.3X.6C-06 Page 1 of 6. Operating Manual for Dissolved Oxygen-Measuring transducer OXY 3610 MP

High-performance submersible pressure transmitter For level measurement Model LH-10

Instruction Manual DO 6. Economy Hand-held Dissolved Oxygen Meter. 68X Rev 2 Apr 08. Technology Made Easy... Part of Thermo Fisher Scientific

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

MMS Mercury Monitoring System for Natural Gas

STARLOG. Capacitive Water Depth Probe

RDO Optical Dissolved Oxygen Sensor

Triton DO8 Dissolved Oxygen Sensors ELECTRO-CHEMICAL DEVICES

Transcription:

YSI Environmental Monitoring Systems Water Quality Training Seminar

6-Series Products 600QS 6820/6920 600R 6820V2/6920V2 600LS 6600/6600EDS 600OMS 6600V2 600XL/600XLM

600 QS (Quick Sample)

600 QS, 600R

600LS

600XL, 600XLM, 600 OMS

6820V2-1, 6920V2-1

6820 V2-2, 6920 V2-2

6600V2-2

6600EDSV2-2

6600 V2-4

6600 V2-4 w/ PAR Photosynthetically Active Radiation (PAR)

Telemetry Platforms

Telemetry Platforms

Autonomous Underwater Vehicles

Autonomous Underwater Vehicles

ProPlus

ProODO

556 & 85

EcoSense

Sensor Technology

Sensor Technology Chlorophyll Temperature Rhodamine Conductivity Blue-Green Algae Dissolved Oxygen Chloride ph/orp Ammonia Water Depth/Level Nitrate Turbidity

Temperature High precision thermistor 2252 ohms at 25C (+/- 1%) Resistance changes with temperature - uses thermistor algorithm to convert to temperature Thin wall titanium sleeve Fast response Resists corrosion

Conductivity AC voltage applied to nickel electrodes Conductivity electrodes placed in sample to be measured Electrical current flows through the electrodes and the sample Electrical current level has direct relationship with conductivity of the solution 4 pure nickel electrodes Cell constant = 5.00 Temperature probe Fresh, brackish and sea water capability

Conductivity Conductivity will vary with temperature changes Specific Conductance is commonly expressed in reference to a specific temperature Specific Conductance = Conductivity corrected to 25 C.

Differences between ROX and Rapid Pulse Sensors ROX No electrolyte Self-cleaning sensor Not affected by H 2 S Zero-cal option Rapid Pulse Electrolyte required Not self-cleaning (EDS sonde required for membrane cleaning) Affected by H 2 S No zero-cal option

Dissolved Oxygen Two main types of Polarographic Oxygen Sensors Steady State Pulsed

Dissolved Oxygen Oxygen diffuses through Teflon membrane Oxygen reduced at gold cathode Electrons flow = electrical signal Oxygen concentration proportional to signal level e - O 2 O 2 e - Gold Cathode H 2 O e - H 2 O e - O 2 Silver Anode O 2 KCl O 2 H 2 O H2 O O 2 O 2 O 2 H 2 O H 2 O H 2 O H 2 O H 2 O H 2 O O 2 O 2 O 2 Teflon membrane H 2 O H 2 O H 2 O

Rapid Pulse Dissolved Oxygen On Rapid-Pulse refined the technology Pulses on / off during measurement sequence Incorporates new electrode design which utilizes a separate silver reference electrode and anode instead of combining the two. This configuration is required to polarize / de-polarize the electrode consistently for a very short period of time (40mS) Utilizes digital electronics Eliminates need for stirring 40ms Off 3960ms 4 sec

Rapid Pulse Dissolved Oxygen Long Term Stability Less susceptible to fouling accurate long term data Accurate results in low flow conditions no stirring device required, <2% stirring dependence Ultra low power requirements long battery life Reliable long term deployments Diagnostics Serviceable

Rapid Pulse Dissolved Oxygen Removed Stirring 100 Percent Saturation Rapid Pulse Response Steady State Response

Optical DO Step 1: The sensor emits blue light, which causes the sensing element to luminesce red light. In the presence of oxygen, the sensor measures a change in phase-shift of luminescent emissions as a function of oxygen concentration. The luminescence signal from the blue excitation is compared to that of the red excitation (next slide) and a stable DO concentration is calculated.

Optical DO Step 2: The sensor emits red light, which is reflected by the sensing element. The reflected red light is measured by the sensor and serves as a reference for the lifetime luminescence calculations

ph Glass sensing bulb is filled with solution of stable ph (ph 7) so inside of glass surface experiences constant binding of H+ ions Outside of bulb is exposed to sample where H+ varies Differential of H+ creates a potential which is read versus the stable potential of the reference electrode AgCl-coated Ag wire H + H + H + H + H + H + E membrane KCl Gel Reference Junction

ph Plot - Nernst Equation 200 100 mv 0-100 4 7 10 ph -200 ph 4.0 +180 slope ph 7.0 0 +/- 50mV ph 10.0-180 slope

Sensor - Glass imbedded platinum ring - Solution potential read versus ph reference electrode Oxidation reduction potential (ORP) of the sample is determined by measuring the potential of a chemically-inert (platinum) electrode which is immersed in the solution The sensing electrode potential is read relative to the reference electrode of the ph probe and the value is presented in millivolts Value represents the overall oxidizing (positive values) and reducing (negative values) ability of the sample ORP

ISE Works exactly like ph except the sensor is a PVC membrane selective for the analyte rather than a glass bulb selective for H+ ions Sensor module contains a static concentration of the analyte which binds to the inner membrane Potential is related to the analyte concentration using the Nernst Equation

ISE ph Reference Electrode + - ISE Module AgCl-coated Ag wire NH 4+ solution PVC membrane with ionpore NH 4 + NH + NH + 4 4 NH + 4 NH 4 + NH 4 + E membrane

Optical Sensor Design Integrated Improved Replaceable Non-corroding Easy to wiper replace, o-ring optical wiper with titanium seals durable fiber shaft new seal switch wiper membrane for shaft controlled longer for with long-life a sensor usable parking life Hsystem 2 of S and one low year oxygen environments

Optical Probes Wiper Motor Wiper Sponge Wiper

Turbidity Techniques Nephelometric Angle of Reflection 90º ISO Recommended method

Turbidity Method Probe contains an LED with a wavelength of 860 NM (Emitter) Probe contains a photodiode (detector) Optical fibers connected to emitter and detector intersect probe face at opposing 45 degree angles - net angle is 90 degrees Light from emitter enters sample and scatters off the particles - light scattered at 90 degrees enter detector fiber and is measured by the photodiode Circuit board in the sonde quantifies sample signal relative to standards and generates a value in NTU Based on ISO 7027 Method

Turbidity Light Source (infrared) Photodetector Fiber optic cables molded at a 45 degree angle 90º Intersect

Chlorophyll Chlorophyll fluoresces when irradiated with light of a particular wavelength (435-470 nm), emits light of a higher wavelength (630-700 nm.) The ability of chlorophyll to fluoresce is the basis of commercial in-situ fluorometers. These fluorometers induce chlorophyll to fluoresce by shining a beam of light of the proper wavelength into the water and then measuring the higher wavelength light which is emitted. Measurement is made in the field without disrupting cells.

Chlorophyll

YSI 6025 Chlorophyll Probe Light Source 470 nm Photodetector Optical Filter Optical Fiber

YSI 6025 Chlorophyll Probe

Blue-Green Algae (Cyanobacteria) Theory of detection same as Chlorophyll sensors (in vivo fluorescence of pigments in living cells) Chlorophyll a not available for fluorescence detection in cyanobacteria We look for accessory pigments Phycocyanin: Freshwater Species Phycoerythrin: Saltwater Species Recommended to use chlorophyll and BGA sensors in concert If a significant BGA population is present, relying solely on the chlorophyll sensor will underestimate total algal biomass

Blue-Green Algae (Cyanobacteria)

Blue-Green Algae (Cyanobacteria) Early Warning Indicator

Rhodamine Probe contains an LED with a wavelength of 540 nm (Emitter) Probe contains a photodiode (detector) Optical fibers connected to emitter and detector Light from emitter enters sample and the Rhodamine WT dye in the sample fluoresces. Probe picks up the high wavelength emission, and uses filter to restrict other source of emission.

Rhodamine Light Source Photodetector Optical Filter Fiber optic cables

Depth Sensor Sensor: Stainless Steel Strain Gauge in four different sensitivities shallow vented, shallow, medium and deep. Shallow Vented 0-3m ± 0.003/3-9.1m ± 0.02m Shallow 0-9.1m ± 0.02m Medium 0-60.5m ± 0.12m Deep 0-200m ± 0.3m (YSI 6600 only) Voltage from Strain Gauge is proportional to hydrostatic pressure.

Non-Vented Depth Measurement Pressure Due to Atmosphere Pressure Due to Water Column P Total = P Water + P Atmosphere

Vented Depth Measurement Pressure Due to Atmosphere Pressure Due to Water Column

Care & Maintenance

Probe Installation Remove the port plug. Lightly grease the o-rings of the probe use only enough grease to make the o-ring shiny. More grease is not better it may actually hinder the seal. Align the probe in its corresponding port. Once the probe key falls into place, press the probe firmly in the bulkhead to completely seat the probe. Hold the probe with one hand to prevent twisting it while finger threading the slip nut.

Probe Installation If you meet resistance while threading, DO NOT continue threading. Back the slip nut out and realign the threads. Finger tighten the probe, then use the provided tool to snug the probe by turning the nut an additional ¼ turn. DO NOT over tighten. For optical probes, finger tighten, then use the provided tool to tighten the probe completely. Be sure to point the probes towards the floor while pulling from the port to minimize port contamination.

Temperature No maintenance required.

Conductivity Be sure the conductivity conduits are free from contamination/blockage. Use the small test tube brushes and a mild soap solution to periodically clean the conduits.

Dissolved Oxygen Membrane changes should be done as necessary (according to application) between 2-6 weeks. If the electrodes look tarnished, recondition electrode with 2400 grit sand paper (provided in 6035 kit) How do I know when it s time to recondition?

Dissolved Oxygen

Reconditioning DO Probe Remove membrane, rinse probe in clean water and dry off probe surface Use the YSI 6035 kit to service electrodes (note that the sanding disc must be struck parallel with the electrodes). Rinse probe surface area with clean water. Re-membrane probe and check DO Charge (after 5 minutes of probe burn-in). If reconditioning DO Probe does not reduce / increase charge repeat conditioning. If repeat conditioning fails, replace probe.

Membrane Replacement Large KCl meniscus

Membrane Replacement Hold the membrane with both hands and come straight down on the DO probe. Be sure not to stretch the membrane laterally, just hold it taut, and stretch it as you pull it down over the probe.

Be sure to adequately stretch the membrane over the probe. Membrane Replacement

Membrane Replacement Hold the membrane taut while placing the o-ring on to reduce bubbles and wrinkling. Be careful not to leave fingerprints on the membrane while installing the o-ring.

Trim the membrane about 2 to 3 mm below the o-ring. Membrane Replacement

The final inspection of the membrane should reveal no bubbles and no wrinkles. Membrane Replacement

Membranes Membranes have specific response coefficients. Membrane response coefficients are stored in the sensor and in the.glp file. Replacement membranes require new coefficients to be entered in sonde software. Membranes must remain hydrated.

Membrane Replacement Remove the new sensor membrane assembly from its hydrated container and dry the underside completely. Ensure both sides of the membrane cavity are dry. Topside Underside

Membrane Replacement 1. After installing the ROX sensor in the sonde bulkhead and replacing the membrane, connect the sonde to your computer or 650 MDS logger. 2. Locate the alphanumeric string on the sticker which contains the coded calibration information for this particular sensor membrane. 3. Access the Main Menu and select Calibrate, then Optical Dissolved Oxy. 4. Select option 3-Enter cal sheet and enter the corresponding calibration coefficients.

ph Clean ph Probe Dirty ph/orp Probe

ph Remove ph probe (if applicable) Soak in mild soap solution for 15 minutes. Rinse, then soak in tap water for 5 minutes. Soak in 1 M HCl for 20-30 minutes. Rinse, then soak in tap water for 10 minutes. Soak in a 1:1 solution of bleach and water for 45 minutes. Rinse, then soak in tap water for 15 minutes.

Same basic care & maintenance as ph probe. ORP

Optical Probes Replace wiper on a regular schedule maintenance, e.g. every 30 days / before each long term deployment Wiper calibration is automatic in sondes with version 2.17 or higher. For earlier software versions, remove all power from the sonde, or if on a PC, ESC to # and type RESET <enter> Wiper calibration should be observed to confirm wiper operation Ensure wiper is not covering emitter and detector. Do not force the wiper as this will damage the motor beyond repair

Clean Optical Wiper

Dirty Optical Wiper

Optical Wiper

Depth Use the syringe provided with your maintenance kit to flush out the depth set. Do not stick anything in the holes on the depth set--you will damage the pressure transducer.

Calibration

Temperature Factory Set - No calibration required.

Conductivity Calibrate using SpCond (Specific Conductivity). Place into clean conductivity standard e.g.: 1000 µs/cm (most standards are labeled in µs/cm, BUT the calibration value has to be entered as a ms/cm). Allow one minute to stabilize before accepting the calibration, or if reading has not changed for 30 Sec. Check Cell Constant in Advanced Menu. Readings should be 5.0 ± 0.45 If Cell Constant is out of range clean and recalibrate. Never accept an Out of Range

ph In Report / Setup Parameters, enable ph mv Place into a ph 7.0 buffer and allow to stabilize for at least one minute. Record mv Offset value 0 +/- 50. Input value of ph 4.0 buffer into Sonde, then rinse. Place into a ph 4.0 buffer, the reading should be close to 4 within a few seconds. Allow to stabilize and record mv slope. Should be 180 mv different than the value recorded for ph buffer 7, +/- 50 mv. Repeat a third time using ph 10.0 buffer. mv should have a slope of -180 mv different than the value recorded for ph buffer 7, +/- 50 mv.. If mv readings/slope are out of spec, or if there is slow response, perform cleaning procedure. If the mv is out of range, or response does not improve then soak probe for 24-48 hours in 2M KCl or ph 4 Buffer.

ORP Allow temperature to stabilize. Use temperature compensation chart that came with ORP Standard to verify value.

Dissolved Oxygen The following tips are how to get the most accurate results for a SAMPLING APPLICATION: Allow probe to warm-up for 4-5 minutes before accepting calibration (RS232 auto sleep disabled) Make sure probe is in a saturated environment Enter current barometer reading (mmhg) Record DO Charge for QA/QC purposes Record DO Gain for QA/QC purposes (after calibration) Confirm DO Warm-up Pattern Never accept an Out of Range

Dissolved Oxygen The following tips are how to get the most accurate results for an UNATTENDED APPLICATION: Probe will self calibrate after the selected warm-up time from advanced menu. (RS232 auto sleep enabled) Make sure probe is in a saturated environment Enter current barometer reading (mmhg) Record DO Charge for QA/QC purposes Record DO Gain for QA/QC purposes (after calibration) Confirm DO Warm-up Pattern Never accept an Out of Range

Calibration Values for Various Atmospheric Pressures & Altitudes In Hg mmhg mbars feet meters %sat 30.23 768 1023-276 -84 101 29.92 760 1013 0 0 100 29.61 752 1003 278 85 99 29.33 745 993 558 170 98 29.02 737 983 841 256 97 28.74 730 973 1126 343 96 28.43 722 963 1413 431 95 28.11 714 952 1703 519 94 27.83 707 942 1995 608 93

Checking DO Warm-up Stop Discrete and Unattended sampling. Confirm that auto-sleep RS-232 is enabled Be sure DO% is set in the report menu. Turn off Sonde for 1-2 minutes. Power up Sonde. Start discrete sampling at 4 second intervals. DO% must display a positive number and decrease with each 4 second sample, eventually stabilizing to the calibration value in approximately 60-120 seconds. Disregard the first two samples they can be affected by electronics warm-up.

DO Warm-up Accept/Reject Criteria The DO% must start at a positive number and decrease during the warm-up. Example: 123, 117, 114, 113, 110, 107, 104, 102, 101, 100, 100 Should the display read a negative number or start at a low number and climb up to the calibration point, the probe should be rejected and must not be deployed.

Turbidity Choose 1 or 2-point calibration. Zero must be the first calibration point. Be sure your zero/di water is clear--no particulates. When using the cal cup, only engage one thread. If using the 6136 probe you must have a black bottom cal cup and probe guard. Never accept an Out of Range

Chlorophyll Do 1 point calibration to zero the probe and use for screening and/or monitoring trends Check for drift and slightly enhance accuracy using dye solution to simulate chlorophyll (2 point cal) Use extractive analysis results from field samples to post calibrate all readings Never accept an Out of Range

Post-calibration of In Vivo Chl Readings

Rhodamine Do a 2-point calibration initially. A 1 point calibration can be used to zero the probe and check for screening and/or monitoring trends Never accept an Out of Range

Depth Make sure flow through holes are cleaned with syringe (proved in Maintenance Kit) Leaving the Sonde in air, zero the depth sensor Using advanced menu check the Pressure Offset readings. Readings should be -14.7 ± 6.0 Never accept an Out of Range

Tips & Tricks

Biofouling Prevention

Biofouling Prevention

Biofouling Prevention

Biofouling Prevention

Questions? Barrett Gaylord Northeast Representative 914-260-6085 bgaylord@ysi.com