Evaluation of the relevance of the water test for the control of sterilization containers

Similar documents
REQUIREMENTS FOR PACKAGING. Wim Renders

User Guide. Pall Laboratory Manifold. For laboratory use. Not for use in a manner other than indicated. Introduction. Regulatory References

Discussion and guidance on the definition and qualification of porous loads

Detection of Micro-organisms in Compressed Gases

THE RIGHT LEVEL OF PROTECTION FOR YOUR NEEDS.

Microbial Impaction Sampler for Clean Room Gas Monitoring TEST AIR

SterileBarrierPackaging: CommonCausesofFailures. PackagingSectionLeader NelsonLaboratories

CTG Certification Protocol -- Version 3.0 SUMMARY

Steam penetration in steam sterilization processes

HPI-4N1 NEGATIVE PRESSURE COMPOUNDING ASEPTIC ISOLATOR PERFORMANCE EVALUATION

Glove Management. PDA Ireland Chapter Dublin 04 th Oct Dr. Aidan Harrington DPS Group

What is the D-ifference in D-value?

Sole reliance for sterility or other quality aspects must not be placed on finished product test.

1 5 2 Definition of Isolator Technology Isolator Types and Applications 3 Critical Design Considerations 4 Biodecontamination Conclusion and Summary 2

NuAire Model NU PR797 Positive Pressure Recirculating Compounding Aseptic Isolator Performance Evaluation Compliance to USP797

Product Test Report F.A. Lennon Drive December 2015 Solon, Ohio U.S.A. Page 1 of 7. Valve Series Description Test Pressure Test Qty.

PERFORMANCE QUALIFICATION AND OCCUPATIONAL EXPOSURE LIMIT (OEL) FOR ISOLATOR

ISO 7218 INTERNATIONAL STANDARD

Instructions for use Compressed air injector with water trap

USP <797> LABELING ESTABLISHING BEYOND-USE DATES AND IN-USE TIMES

H 2 O 2 Bio-Decontamination and Aseptic Transfer of Heat Sensitive Materials Sergio Mauri, Director, Global Marketing and Business Intelligence

Aseptic Vial and Syringe Filling Narlin B. Beaty, Ph.D. Adelaide Nye

Hygea Ultrasonic Validation Kit

Webinar Expert Module 1: Garment Logistics and Gowning. Dupont September GOP-Innovations

ECOBULK FOODCERT + ASEPTIC The highest packaging safety for your food products

Ad Hoc GMP Inspections Services Group EC GUIDE TO GOOD MANUFACTURING PRACTICE REVISION TO ANNEX 1

Annex 6 Good manufacturing practices for sterile pharmaceutical products

Filtration FILTRATION PRODUCT TIPS

Regulatory Basics for Facility Design (WHO GMP): Current GMP Requirements

TEMPLE UNIVERSITY ENVIRONMENTAL HEALTH AND RADIATION SAFETY

Important considerations for use of Cembrane

LEAK TESTING UNDER WATER

OEL LIMITS: Products are classified according to hazards levels that correspond to these five OEL levels as shown in the chart below :

R2S Standard Sampling Assembly OPERATION PROCEDURE

CORESTA RECOMMENDED METHOD N 6

RESPIRATOR PERFORMANCE TERMINOLOGY

PRODUCT NUMBER: Rd40 Aluminium Canister HgP3

Index. 1. Introduction. 2. Available models. 3. Sterilization Programs. 4. Mechanical features. 5. Electrical features

NuAire Model NU NTE797 Negative Pressure Total Exhaust Compounding Aseptic Containment Isolator Performance Evaluation Compliance to USP797

Instructions for use Evacuation system compressed air injector SL 20 / SL 20 Max SL 30 / SL 30 Max

Pseudomonas aeruginosa Disinfection in Swimming Pool Water - Laboratory Efficacy Testing.

Standard Pneumatic Test Procedure Requirements for Piping Systems

Corning HYPERStack Cell Culture Vessel Closed System

Air Removal from the Sterilization Chamber

AMBLER SURGICAL INSTRUCTIONS FOR USE - OPHTHALMIC DUET-STYLE SCISSORS

NuAire Model NU NR797 Negative Pressure Recirculating Compounding Aseptic Containment Isolator Performance Evaluation Compliance to USP797

Manual Chapter 1. 1 st Generation 2 nd Generation

A Perspective on Filter Cartridge Issues

Gerard Sheridan Inspector

Transfer Procedures. By the end of this chapter, you will be able to: The Transfer Process. Chapter 12

1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique.

HYGIENIC DESIGN, TESTING AND CERTIFICATION

VALITEQ DMS v2.o Document Management Software Tutorial

Development Report. Walker Barrier 10 Ergoclean Gloveport Interface. And. Single Piece Gauntlet Performance Testing

PVP Copyright 2009 by ASME

Texas Department of Insurance. Respiratory Protection. Provided by. Workers Health & Safety HS01-027B (1-06)

Int r oduct ion t o Int egr it y Tes t ing Theor y

Test lungs from IMT Analytics simulate a range of realistic ventilation situations during training.

Steritech: A World Leader in Sterilisation and Decontamination

Aseptic Processing Achieving Sterility by Design

Shukla Medical. TITLE: Manual Surgical Orthopedic Instruments Recommendations for Care, Cleaning, Maintenance and Sterilization

This program is registered with the AIA/CES for continuing professional education. As such, it does not include content that may be deemed or

DESIGN AND INSTALLATION CONSIDERATIONS FOR TSI AEROTRAK VHP-RESISTANT REMOTE PARTICLE COUNTERS

LIGHTHOUSE. Stopper Pop-Up and the Effects on. Container Closure of Sterile Vials. Application Note 101 LIGHTHOUSE

1. The laboratory supervisor must enforce the institutional policies that control access to the laboratory.

Beta Study Report. Results of Nosocomial Infection Study In Whirlpool Bathing

FlexAct BT (Bag Tester) Validation November, 2015

EMT EnvironmentalMonitoringTECHNOLOGIES

Suction Regulators. Recommendations for Cleaning, Disinfection and Reprocessing: Boehringer 3800 &3900 Series Suction Regulators


Aseptic Techniques. Techniques for Sterile Compounding. Pharmacy Technician Training Systems Passassured, LLC

_ pressure transducers. User Manual

Pneumatic switch module 3/2-way valve (On/Off) Types P Types FP Function tested for use of the float in Ex-zone 0

HYDROSTATIC LEAK TEST PROCEDURE

E PB 127/2013 TEST CERTIFICATE. Reusability of sealing cushions (valve) coated with swelling material

STANDARDS FOR COMPRESSED GAS TESTING

CONTAINMENT OF CO2 REFRIGERATION INSTALLATIONS

Manufacturer s Information

Pressure Equipment Directive PED 2014/68/EU Commission's Working Group "Pressure"

Combination Air Valve Model

amchydroflow HP Validation Guide

Fire Escape Hood OPERATION AND INSTRUCTIONS

Instruction Manual. CG16K Barometrically Compensated Capsule Dial Gauge. CG16K Capsule Dial Gauge, 0 to 25 mbar

Instructions For Use USD3156. Kleenpak Capsule with Pegasus Protect Prefiltration Membrane

Combination Air Valve

New Technological Guidelines Required for Pipette Management (Proposal of a Pipette Management Method)

á1207ñ PACKAGE INTEGRITY EVALUATION STERILE PRODUCTS

MEASURING INSTRUMENTS Pressure Gauge T-Piece...77 Pressure Gauge TC 25.0, connection below...78 Pressure Gauge TC 25.0, connection behind...

EuroBioConcept s.a.s Hospital Pharmacy Isolator Systems

Determination of density by volumetric displacement Skeleton density by gas pycnometry

PRODUCT NUMBER: Rd40 Aluminium Canister K2

Section 3-5: Principle 3: Determine Critical Limits

RHIC CRYOGENIC SYSTEMS MAINTENANCE PROGRAM

AES-28/50/75/110/150 RFG Sterilization Autoclaves with counter pressure and fast cooling

INFORMATION NOTE No 353: MICROBIOLOGICAL TESTING

Gas Pressure Regulator HON 300

Can Temperature And Pressure Datalogging Replace The Bowie and Dick Test? Dr. Brian Kirk

MANUFACTURING RISK ASSESSMENT STUDY FOR STERILE DRY POWDER INJECTION OF CEFTRIAXONE SODIUM

ISO INTERNATIONAL STANDARD. Breathing system filters for anaesthetic and respiratory use Part 2: Non-filtration aspects

Generating Calibration Gas Standards

Transcription:

Decarout L., Lambert C. Unité de Stérilisation CH Métropole Savoie, Chambéry. France Evaluation of the relevance of the water test for the control of sterilization containers

Concept of the SBS Container = reusable Sterile Barrier System (EN 11607-1) ISO 11607-1 : after forming the sterile barrier system, the closures must constitute a barrier against micro-organisms. SBS integrity = preservation of the sterile state 2

Container checks Mandatory functional tests in France Before sterilization, each reusable container is visually inspected, and its suitability for operation is verified according to the manufacturer's recommendations. (1) (1) Bonnes pratiques de pharmacie Hospitalière. Arrêté du 22 juin 2001 EN ISO 11607-1 requirements Acceptance criteria must be established for the inspection prior to each reuse. 3

Container performance controls Ultrasound test Vacuum cloche Visual checks EN 868-1 (1997) : vacuum test Water test 4

The water test Objective: evaluation of the watertightness of the container/lid junction 1 leak: test + no leak: test - (1) FDS 98-053 : Test protocol to be performed on reusable sterile barrier systems (containers) to assess the watertightness of the container/lid closure (2014) 5

Relevance of the water test The water Test is not recognized by container manufacturers National survey: containers with positive leak test France : 29% (3) Switzerland : 17% (4) (3) Validation des emballages en stérilisation centrale hospitalière Atelier des Journées nationales d études sur la stérilsation. Marseille 2013 B. Valence et J. Molina 4) Cavin F. Analyse de risque suite aux contrôles d étanchéité des jointes des conteneurs dans plusieurs établissements suisses. 16èmes congrès WFHSS; 2015; Lille 6

Does the water Test have a predictive character for an integrity breach of the SBS? 7

Permeability of the microbial barrier Gas exchange through the SBS: Gas contraction during cooling Pressure variations (atmospheric pressure, overpressure, elevation variation ) Preliminary tests Volume changes during cooling (exit autoclave) Impact of pressure changes 8

Relevance of the water test for the control of sterilization containers 9

Material & methods 15 containers 30x30 x11 cm 5 with a positive leak test (F+) 5 with a negative leak test (F-) 5 Positive control without filter (T+) PP trays/tsa agar/bottom of the containers Sterilisation at 125 C/ 20 min. Cooling in unloading zone (EN 14664-1:ISO 8) Containers placed in aerosol chamber 10

Material & methods (2) Self-contained ultrasonic nebulizer Suspension of Microccus luteus (5.10 8 UFC) 3 pressures (25, 50, 75 mbar) 3 containers per series: 1 container F +, 1 container F -, 1 container T + Exposure to overpressure for 30 min. in sealed chamber 15 series of 3 containers/pressure Protective packaging before transport to the bacteriology laboratory 11

Materiel & methods (3) 15 sets of 3 containers: F + = water test (+) and filter F-= water test (-) and filter T += without filter Nebulization of 5.10 8 CFU M. Luteus during 20 min 3 pressure: 25, 50 and 75 mbar overpressure 30 min 12

Material & methods (4) Aseptic opening of containers Packaging of trays under laminar flow Positivecontrol : M. luteus colonies Negative bcontrol : without colonies Incubation at 37 C for 5 days 13

Results % d'ufc 100 90 80 70 60 50 40 30 20 10 0 73 27 T+ 25 mbar 33 33 54 13 T+ 50 mbar 67 T+ 75 mbar 7 20 27 73 73 F- 25 mbar F- 50 mbar 40 60 F- 75 mbar 27 20 27 26 F+ 25 mbar 53 47 F+ 50 mbar 6 27 67 F+ 75 mbar T+ : positive control F+ : positive leak test F- : negative leak test <1 [1-9] [10-99] >100 Distribution of CFU/containers/pressure 14

Results : determination of log reduction value Determination of log reduction value (LRV): LRV = log CFU 0 - log CFU 1 Sterility Assurance Level = LRV 6 Using the same containers LRV mean Type conteneur Pression (mbar) LRV moyenne T1+ T+ 25 3,39299769 F1+ F+ 25 3,89573862 F1- F- 25 5,81830839 T2+ T+ 25 3,39952207 F2+ F+ 25 4,28467999 F2- F- 25 4,87054405 T3+ T+ 25 3,52134567 F3+ F+ 25 5,34152134 F3- F- 25 6,7934848 T4+ T+ 25 3,39233701 F4+ F+ 25 3,75675486 F4- F- 25 6,29979445 T5+ T+ 25 3,37396797 F5+ F+ 25 4,37900185 F5- F- 25 5,81830839 T1+ T+ 50 4,10323955 F1+ F+ 50 5,74166619 F1- F- 50 5,88630564 T2+ T+ 50 3,88242412 F2+ F+ 50 5,57898955 F2- F- 50 5,73192096 T3+ T+ 50 4,15804139 F3+ F+ 50 6,14487132 F3- F- 50 6,25812373 T4+ T+ 50 4,28447045 F4+ F+ 50 5,57683265 F4- F- 50 6,59050818 T5+ T+ 50 4,45160687 F5+ F+ 50 6,75682986 F5- F- 50 7,10902212 T1+ T+ 75 4,32170488 F1+ F+ 75 5,90768056 F1- F- 75 5,90768056 T2+ T+ 75 4,20448772 F2+ F+ 75 5,74500392 F2- F- 75 6,26529125 T3+ T+ 75 4,19387396 F3+ F+ 75 6,60866252 F3- F- 75 7,2502129 T4+ T+ 75 4,2301899 F4+ F+ 75 6,14143961 F4- F- 75 6,75652254 T5+ T+ 75 4,94608948 F5+ F+ 75 7,44135817 F5- F- 75 6,75652254 15

Results : group comparison Statistically significant difference between negative water test containers (f-) and positive (f +) containers (P-value = 0.002). 16

Results : group comparison Statistically significant difference between T + positive control and F + groups (P-value = 0.002) as well as T + positive control and F-group 17

Discussion Relevance of maintenance Reduction in leakage rate from 58% to 15% after maintenance (5) Results in % : insufficient for demonstate filtering barrier 99,9999% of barrier = 100 micro-organisms if bioburden is 10 6 Bioburden of 10 8 Micrococcus luteus too high? Exposure of 30 min. vs. storage for 6 months Containers vs. wrapped trays: 87% contaminated vs. 0% (6) (5) C. Lambert. Contrôle de performance des conteneurs : Intérêt du test à l eau. JIFS 2014. (6) Harry L. Shaffer. Sterility maintenance study : dynamic evaluation of sterilized containers and wrapped instrument trays to prevent bacterial ingress. AJIC 18

Discussion Are visual controls sufficient? 76% containers that passed the visual inspection leak (2) Depends on operator Using the leak test Not feasible before each assembly of a tray Periodic inspection necessary (periodicity to be defined according to QMS sterilization unit) Provider: after maintenance In case of a positive test? Removal of container from the reprocessing Curative maintenance Use fold inside container (SBS) 19

Conclusions Overpressure promotes the penetration of microorganisms: impact for facilities located at different altitudes Particle control is required in the unloading area of sterilizers (ISO 8 recommended) Transport of sterile sets after cooling protects from recontamination Visual controls: necessary but insufficient to evaluate container performance Water Test: highlights the lack of sealing of the microbial barrier! 20

Take Home Message The choice and use of an SBS are only a matter of Strong Common Sense (Solide Bon Sens in French). 21

Act or play ostrich? 22

Thank you for your kindly attention christophe.lambert@ch-metropole-savoie.fr 23