DeFelsko PosiTector Dew Point Meter DPM & DPMD Probes

Similar documents
DeFelsko PosiTector CMM IS Probe

Calibration Procedure

Calibration Procedure

Calibration Procedure

Calibration Procedure

PosiTector UTG, UTG-STD & UTG-ME

Calibration Procedure. PosiTest DFT Combo

Calibration Procedure

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

BI-680 Online Dissolved Oxygen Controller Instruction Manual

ali-q TM Gravimetric Verification Procedure

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Relative Humidity Calibration Kit

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Model of a Standard Operating Procedure (SOP) for SO 2 measurements with a UV fluorescence analyzer

2014 Programme of Training Courses in Measurement & Calibration

FAULT CODE TROUBLESHOOTING INDEX

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

CALIBRATING THE GMA PRESSURE SENSORS AND THE ECU ENGINEERING UNIT

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

TECHNICAL MANUAL CALIBRATION INSTRUCTIONS FUEL PRESSURIZING & DUMP VALVE TEST STAND 1TL

PRODUCT SHEET. Order probe only as RXPROBE02

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

Calibration certificates

Owner s Manual Humiport 10/20

Optical Dissolved Oxygen Meter

Soil Water Characteristic Curve (SWCC) Using the WP4C

DO NOT PRINT THIS PAGE

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

! Warning, refer to accompanying documents.

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

Technical Procedure for General Laboratory Equipment

Digital Vacuum Regulator

MARATHON SENSORS INC. TROUBLESHOOTING A CARBON PROBE

OPTIMA 7 - BIOGAS Calibrating the gas sensors SENSOR ADJUSTMENT. Diamond Scientific 625 Peachtree St., Cocoa, Florida /Tel:

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

Digital Vacuum Regulator

8.1 Calibration of gas sensors

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

DBI version 002 User Manual US version Free Balloon Flight Instrument

SL G3. Warning

Standard Operating Procedure

LM LUMEN MAINTENANCE TEST REPORT

Installation Instructions

Installation Instructions

Moisture Monitor Series 3. Service Manual

SL G3 - Fuel Conversion to Natural Gas (LP to NG) P-Kit 301

AMS 2710 PCB pressure sensor module with V output

operating Instructions

Heavy Duty Dissolved Oxygen Meter

Annual Inspection and Calibration Checklist VAF Oilcon MK-6(M) MEPC.108(49) Page 1 of 5

SL G3 - Fuel Conversion to Propane (NG to LP) P-Kit 302

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

WBEA Standard Operating Procedure

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

Pneumatic high-pressure controller Model CPC7000

QUICK START GUIDE TO DRINKING WATER MONITORING

DQM Annual Hopper QA Checks

Additel 761 Automated Pressure Calibrators

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1

Operation Manual. Pro 4 Warn Analyzer TM Rev 08.17

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

USER'S GUIDE. Vaisala CARBOCAP Carbon Dioxide Calibrator for the GMP220 series probes GM K220 U421EN-1.1

Location of CC Number if not located with the identification:

LOW PRESSURE VOLTAGE BREAKDOWN TEST PROCEDURE FOR GSS FSU FLIGHT SPARE HVA BOARD October 28, P0938 Rev

CCPaSEC. Procedure for the Quality Control Yearly Equipment Check. July 2018 CCPaSEC Quality Control Team

Pneumatic high-pressure controller Model CPC7000

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

WeatherStation Monitoring for Industrial marine Applications

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

D-Opto. Dissolved Oxygen Sensor. Operation Manual (Software Version 1.0)

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Operation Manual. Pro CO O 2 Alarm. Carbon Monoxide and Oxygen Analyzer. Rev

Installation Instructions

D-OptoLogger Dissolved Oxygen Logger Operation Manual

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING

YSI Model 550a Dissolved Oxygen (DO) Meter

OC508 Programmable mv Calibrator

Transmitter CS 21 Operation Manual

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR PRESSURE GAUGE. This publication replaces T.O. 33K dated 30 January 2012.

DISSOLVED OXYGEN SENSOR BT34i

Optical Dissolved Oxygen Meter

DURRIDGE STAINLESS STEEL SOIL GAS PROBE

Simrad GD10P Extended On-site Maintenance

DigiTRON Electrical Flying Leads, Jumpers and Harness Assemblies Site Received Test Manual

ASTM D86 - Procedural Outline

BGA244 Binary Gas Analyzer

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

Digital Melting Point Apparatus

In Vivo Scientific, LLC INSTRUCTION MANUAL

Operating Manual. BACVis. Manual BACVis. for. Sensors and MilliGascounter. Rev

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

Additel 761 Automated Pressure Calibrators Selection Guide

Coda cx1 Thermal Cycle Test Procedure

Dissolved Oxygen/Temp Waterproof Tester

Transmitter CS 21 Operation Manual

Transcription:

Management Procedure 2581 Revision: D Date Issued: November 9, 2007 Date Revised: April 19, 2017 Calibration Procedure DeFelsko PosiTector Dew Point Meter DPM & DPMD Probes Table of Contents 1 Introduction and UUC Performance Requirements... 2 Table 1-1 Measurement Ranges... 2 2 Measurement Standards and Support Equipment Performance Requirements... 2 Table 2-1 UUC Accuracy Requirements and Description... 2 Table 2-2 Minimum Use Specifications... 2 Table 2-3 Actual Air Temperature & Relative Humidity Equipment Specifications... 3 Table 2-4 Actual Surface Temperature Test Equipment Specifications... 3 Table 2-5 Calibration Environment and Warm-Up Requirements... 4 3 Preliminary Operations... 4 4 Humidity and Air Temperature Calibration Process... 4 5 Surface Temperature Calibration (DPM only)... 5 6 Performance Requirements... 5 Table 6-1 Requirements and Calibration Data for DeFelsko PosiTector DPM & DPMD Probes... 5 Management Procedure Change Notice... 6 Management Procedure MP 2581 Rev D Page 1 of 6

1 Introduction and UUC Performance Requirements 1.1 This procedure describes the calibration of the DeFelsko PosiTector Dew Point Meter (DPM) Built-in and (DPMD) cabled probes with the following specifications: Table 1-1 Measurement Ranges Function Measurement Range Resolution Air Temperature -40 to 80 C (-40 to 175 F) 0.1 C (0.1 F) Relative Humidity Surface Temperature (DPM only) 0 to 100% RH -40 to 190 C (-40 to 375 F) 0.1% 0.1 C (0.1 F) 1.2 The unit being calibrated will be referred to as the UUC (Unit-Under-Calibration). 2 Measurement Standards and Support Equipment Performance Requirements 2.1 The UUC accuracy requirements are based upon the published UUC performance specifications. 2.2 The test uncertainty ratio applied in this calibration procedure is 4:1 unless otherwise stated. The surface temperature uncertainty ratio for the range -40 to 80 C is 2.9:1. 2.3 Minimum-Use-Specifications are the minimum test equipment specifications required to meet all the UUC accuracy requirements and the test uncertainty ratio applied. Table 2-1 UUC Accuracy Requirements and Description UUC Function Range Accuracy Test Method Air Temperature Relative Humidity -40 to 80 C (-40 to 175 F) 0 to 100% RH + 0.5 C (1.0 F) + 3.0 % Humidity Generator Surface Temperature -40 to 80 C (-40 to 175 F) 80 to 190 C (175 to 375 F) + 0.5 C (1.0 F) + 1.5 C (3.0 F) Temperature Calibrator Table 2-2 Minimum Use Specifications Function Range Accuracy Air temperature -40 to 80 C (-40 to 175 F) + 0.125 C (0.25 F) Relative Humidity 0 to 100% RH + 0.75% Surface Temperature -40 to 80 C (-40 to 175 F) + 0.125 C (0.25 F) 80 to 190 C (175 to 375 F) + 0.375 C (0.75 F) Management Procedure MP 2581 Rev D Page 2 of 6

Table 2-3 Actual Air Temperature & Relative Humidity Equipment Specifications Actual Equipment Specifications Manufacturer/Model # s Applicable Range Accuracy Thunder Scientific Model 1200 Humidity Generator 10 to 60 C (60 to 140 F) + 0.05 C (+ 0.09 F) 10 to 95% RH + 0.5% Table 2-4 Actual Surface Temperature Test Equipment Specifications Equipment Name Actual Equipment Specifications Manufacturer/Model # s Applicable Range Accuracy 1. Type K Thermocouple -200 1250 C 0.138 C (-328 2282 F) 2. Distilled Water Ice Bath 3. Keithley 2000 Multimeter 4. EDL STS-SC2 Calibrator N/A Up to 100 mv 40 C 100 C 0.05 C 0.090 C 0.016 C 0.25 C Keithley 2000 Multimeter calculation Use ITS-90 (International Temperature Standard) Table for Type K Thermocouple Temperatures at 40 C (1.612 mv) and 100 C (4.096mV) Scale Range 100 mv with 1 year accuracy = 50 ppm of reading + 35 ppm of range Accuracy @ 40 C = (50 ppm x 1.612 mv) + (35 ppm x 100 mv) = 3.580 uv (40 C /1.612mv) * 0.003580mv =.089 C Accuracy @ 100 C = (50 ppm x 4.096 mv) + (35 ppm x 100 mv) = 3.704 uv (100 C /4.096mv) * 0.003704mv = 0.090 C Surface Temperature Combined Accuracy @ 40 o C = (Thermocouple 2 + Bath 2 + Keithley 2 + Surface plate 2 ) 0.5 = (0.138 2 + 0.05 2 + 0.090 2 + 0.016 2 ) 0.5 = 0.17 C Surface Temperature Combined Accuracy @ 100 o C = (Thermocouple 2 + Bath 2 + Keithley 2 + Surface plate 2 ) 0.5 = (0.138 2 + 0.05 2 + 0.090 2 + 0.25 2 ) 0.5 = 0.30 C Surface Temperature Uncertainty Ratio (-40 to 80 C): = 2.9:1 (0.5 C / 0.17 C) Caution: The instructions in this Calibration Procedure relate specifically to the equipment and conditions listed in this section. If other equipment is substituted, the information and instructions must be interpreted accordingly. Management Procedure MP 2581 Rev D Page 3 of 6

Table 2-5 Calibration Environment and Warm-Up Requirements Temperature: 23 ± 5 C. Relative Humidity: Less than 95% Barometric Pressure 30 ± 1.5 in Hg (1016 ± 50mbar) Measurement Standards & Support Equipment Environmental Requirements: Measurement Standards & Support Equipment Warm-up and Stabilization Requirements: 3 Preliminary Operations Thunder Scientific Humidity Generator: 60 minutes EDL STS-SC2 surface temperature calibrator: 15 minutes Note: Review the entire document before starting the calibration process. 3.1 Visual Inspection 3.1.1 Damage or excess wear must be repaired prior to beginning the calibration process. 3.1.2 Visually inspect the UUC for: Wear or damage to the probe body or tip Missing parts Proper identification 3.2 If the probe has a protective cap, remove it before proceeding. 4 Humidity and Air Temperature Calibration Process Refer to UUC and equipment instruction manual(s) for menu navigation instructions, details on features and operating instructions. Note: Whenever a test requirement is not met as indicated in table 6-1, verify the results of the test and take corrective action before proceeding. 4.1 Attach the probe(s) to the adapter cable connectors in the humidity chamber. Multiple probes can be calibrated in the chamber simultaneously. Make note of what probe is attached to which body. Record the probe serial number(s) and the Barometric pressure (mbars). Note: If adapter cables are not available, the entire Gage/Probe combination can be placed in the chamber. The gage would need to be set in data logging mode to obtain the readings. Refer to the gage user manual on how to set data logging mode. 4.2 Adjust the set point of the humidity chamber to 35 %RH and allow to stabilize for at least 60 minutes. After the stabilization time turn on the gage(s), wait about 1 minute to verify Management Procedure MP 2581 Rev D Page 4 of 6

the reading is stable, and record the gage and the chamber relative humidity and the gage and chamber air temperature readings. Note: When turning on the gage do not perform a full reset. 4.3 Adjust the set point of the humidity chamber to 65 %RH and allow to stabilize for at least 60 minutes. After the stabilization time turn on the gage(s), wait about 1 minute to verify the reading is stable, and record the gage and the chamber relative humidity readings. Turn off the gage and disconnect the probe(s) from the connectors in the chamber. 5 Surface Temperature Calibration (DPM only) 5.1 Prepare an ice bath using distilled water per ASTM E 563-11. Connect the thermocouple to TC1 of the surface temperature calibrator, the ice bath and the Keithley 2000 Multimeter. 5.2 Set the surface temperature calibrator to 40 C and allow to stabilize for at least 15 minutes. 5.3 Connect the probe to a body and turn on the gage. Place the DPM Probe on the center of the top surface of the surface temperature calibrator for 15 seconds and record the highest readings for the Keithley 2000 Multimeter* and the gage. *Note: The multimeter readings must be adjusted by the surface temperature offsets as indicated on the surface temperature calibrator calibration certificate. 5.4 Repeat steps 5.2 & 5.3 for 100 o C. 6 Performance Requirements Note: The technician will collect the data needed to complete columns D and E. The technician shall then calculate the values for Column F and record all information as shown in table 6-1. Do not write in this procedure. Table 6-1 Requirements and Calibration Data for DeFelsko PosiTector DPM & DPMD Probes Reference (A) Units (B) Set Point (C) Test Equipment Reading (D) Gage Reading (E) Probe Measurement Accuracy (F) Allowable Tolerance (G) Relative Humidity %RH 35 + 3.0 Ambient Temperature C N/A + 0.5 Relative Humidity %RH 65 + 3.0 Surface Temperature Low C 40 + 0.5 Surface Temperature High C 100 + 1.5 Note: To convert from C to F T F = 1.8 * T C + 32 Management Procedure MP 2581 Rev D Page 5 of 6

Management Procedure Change Notice Procedure Number: MP 2581 Revision Level: D Date of Change: April 19, 2017 Title: Calibration Procedure for DeFelsko PosiTector Dew Point Meter DPM & DPMD Probes Reason for Change: Update for new version probe Description of Change: Added DPMD code throughout procedure Table 1-1 added DPM only to Surface Temperature Added wait period to 4.2 and 4.3 Moved surface temperature calibration to separate section Removed Ts delta calculation I confirm I have read and understand the procedure and the change described above. Printed Name Signature Date Management Form 0010.02-05/1998 Management Procedure MP 2581 Rev D Page 6 of 6