Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

Similar documents
Acclaim RSLC. (Rapid Separation Liquid Chromatography)

HPLC systems. Overview

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

Troubleshooting. Select the Table of Contents here Rheodyne All rights reserved

AKTA pure 25 New Owner s Intro

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

HPLC troubleshooting guide

no peaks missing peaks too small peaks increasing retention decreasing retention

HPLC. Solvent Delivery Module Systems

The instrument should have a mass range from 2 to 2000 m/z or better Interface

DOS in Food Science & Nutrition, Manasagangotri, Mysore

GCMSD-Headspace Analysis SOP

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

BETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Setting a New Standard in Flash Chromatography Performance

Automated Determination of Dissolved Gases in Water Anne Jurek

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Indian Institute of Technology Kanpur Environmental Engineering and Management Department of Civil Engineering

Figure 1: Plug in ion inlet of mass spec

Agilent 1260 Infinity Purification Solution

Valves FLUIDICS VALVE OVERVIEW & FUNCTIONS ROTARY SHEAR VALVES STAND ALONE VALVE PRODUCTS MANUAL VALVES SPARE PARTS VALVE ACCESSORIES

For that reason VICI Jour developed:

UP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule

DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178

ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information

Prep LC Training Manual S.V.

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

SOP: Buck Scientific BLC-20P HPLC Operation

Autosampler problems!

Running HPLC Methods on an ACQUITY UPLC System

Gradients in ion chromatography. Optimal separation with shorter analysis time

The instrument will process the samples one by one till all the samples have been cleaned up.

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

Generating Calibration Gas Standards

OPERATORS MANUAL G U A R A N T E E D T R Y C H E M APPROVED

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Dionex IonPac AS19-4µm

STANDARD OPERATING PROCEDURES

FROG-4000 TM Chemical Analysis System

LC, RSLC, Bio RS and Dual Gradient Systems

Extended Application Note

Prep LC Training Manual

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

Best Practice for Identifying Leaks in GC and GC/MS Systems

Multiple Gas#5 GC configuration Jan 2016

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Video Notes. LC Troubleshooting Series Retention Time Shifts

Fiber Optic SMA Z-Flow Cell Manual Version 2.1

Rev. CO Packed Columns for Normal Phase Chromatography. TSKgel NH m INSTRUCTION MANUAL

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Peak Biotech - Company & Product Presentation

Rapid and Reliable Detection of Dissolved Gases in Water

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

FROG-4000 TM Chemical Analysis System

NMR Service. How to Prepare Samples for NMR

Operating Instructions Multi-Channel Vacuum Degasser

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

MV5: Automated concentration system

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Zebron Gas Management

HICHROM. HPLC Accessories

Buck Scientific, Inc.

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions

Advances in Capillary Column Backflushing

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

Columbus Instruments

Organic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results

System Overview. TCD Detector temperature setpoint Regulator

Svensk läkemedelsstandard

Preparative Valve-Based Fraction Collector

Standard Operational Procedure for Capillary Electrophoresis

Oxygen Contamination

POLYMER TEST KIT (CLAPPER METHOD) OFI PART No

INSTALLATION AND OPERATING INSTRUCTIONS

Agilent 1290 Infinity Pump Head Maintenance

Transcription:

HPLC Hints, Tips, Guidelines and Important Considerations Presenters: Wilhad Reuter / Tom Breslin

Important HPLC Solvent Characteristics: ALL injected compounds/diluents must be soluble High purity and degassed Noncorrosive Low viscosity UV transparency Solvents should not alter component s activity in sample Commercial bottled solvents should not be re-filtered Add 1% MeOH or 0.5% ACN to water. Not recommended using pure water > 2 days old Do not use halogenated acids; e.g., HCl Typically excludes IPA For UV detection, solvents must not absorb close to wavelength(s) of interest; above UV cutoff Especially for pharmaceuticals and biomolecules Select such that components elute between 2 to 20 column volumes ( k )

Solvent Degassing Mobile phase degassing is important for accurate pump blending and to avoid bubble formation in the detector flowcell. Techniques: BOILING Hmm!, not advised VACUUM Useful for only 2-42 4 hours; Must combine w/ sonication HELIUM (at least High Purity) SPARGE CONTINUOUS SPARGE SPARGE AND PRESSURIZE ON-LINE VACUUM { Uses too much helium Best choice, when available Sparge for 5 min; then let pressurize Good choice when helium unavailable; 80% as effective as helium degassing

Effect of (Not) Degassing on Retention Time Pump mixed mobile phase at 30 O C; Injections every 6 min and, after Helium stopped, every 4 hours 2.0 1.8 Solvent A = 1% HOAc in MeOH Solvent B = 1% HOAc in H 2 O Mobile Phase = 34% A 0.22% CCA Chlorthalidone 1.6 Retention Time (min) 1.4 1.2 3.9% (RSD) Not Degassed Not Pressurized Degassed and Pressurized 1.5% Stop Helium Pressurization 1.0 0.24% 0.8 3.5% 1.3% 0.6 48 min 16 Hours 2 4 6 8 10 12 14 16 18 20 22 24 26 28 Injection Number

Effect of temperature Bottom line: Avoid ambient temperatures 2.4 2.2 Premixed solvent; starting at 30 Degrees 3.4% CCA Chlorthalidone Retention Time (min) 2.0 1.8 1.6 1.4 1.2 0.31% (RSD) o 30 C w/ Oven Pressurized and Degassed Room Temperature o 22-25 C Pressurized and Degassed 2.5% 0.27% o 30 C w/ Oven Pressurized and Degassed 1.0 0.24% 0.30% 0.8 24 Hours 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 Injection Number

Common Mobile Phase Modifiers Buffers To stabilize phs of mobile phase under reverse phase (RPC) or ionexchange (IEC) chromatography (phosphate, acetate, citrate); ~5-20 mm Triethylamine (TEA) To reduce tailing of basic analytes RPC; ~0.1% Phosphoric, acetic acid or formic acids Ion suppressors To suppress ionization of acidic analytes under RPC; ~0.1% Ion-pairing reagents For separation of ionic compounds using RPC column (C18) Anions: Quaternary ammonium salts (e.g., Tetrabutylammonium hydroxide); ~2-20 mm Cations: Alkyl sulfonates (e.g., Hexane sulfonate); ~2-20 mm

Series 200 Analytical Pump: Low Pressure Mixing Solvent toward column Solvents Solvent into metering chamber 100 µl 107 µl Solvent mixing delay volume: ~800-1000uL, depending on pressure Advantages: Great solvent mixing and built-in solvent compressibility factor

Low-Pressure Mixing Pump Blending Performance

Series 200 Micro-Pump: Binary High-Pressure Mixing To Injector Purge Valve Purge Valve Tee Pump A Pump B Solvent mixing delay volume: ~10-400uL with static mixer, depending on flow rate Advantage: Very low solvent mixing delay; good for high speed gradients and LC/MS

Method Transferability Concerns Low-Pressure Mixing: 1.0mL Delay Volume = High-Pressure Mixing: 0.1mL Delay Volume = 2 ml/min

Using Injection Delay Time (for Method Transferability) Low-Pressure Mixing: 1.0mL Delay Volume = High-Pressure Mixing: 0.1mL Delay Volume = Low-Pressure mixing with 0.45 min Injection Delay Time 2 ml/min

HPLC: Preferred Materials for Wettable Parts STAINLESS STEEL Industry Standard SS 316; May rarely effect biological activity or allow adverse chealation to occur TITANIUM (alternative to SS) PEEK (Poly-Ether Ether-Ether-Ketone) TEFLON / VESPEL / TEFZEL POLYAMIDE / POLYIMIDE QUARTZ-SILICA GLASS Good biocompatibility. Somewhat more brittle than SS Good biocompatibility and often used for basic compounds. Becomes brittle w/ >20% THF or MeCl 2 Used for rotor seals and low- pressure tubing. Note: : Use PEEK for basic compounds Do not use with MeCl 2 Commonly used for flow cells RUBY ` Used for balls in pump check valves SAPPHIRE Used for pump pistons

Pressure Problems PRESSURE TOO HIGH Dirty column Blockages e.g., system filter, guard column or column inlet frit, tubing, injector needle or valve, detector flow cell. >> Check by isolating pressure point(s); like a plumber Check flow rate Different solvent mixture? Are solvent lines switched?

Pressure Problems PRESSURE TOO LOW Leaks Pump malfunction e.g., piston seal, connections, rotor seal e.g., vapor lock, sticking check valves or solenoid valve, broken piston Plugged solvent filter Check flow rate Different solvent mixture? Lines switched?

Autosampler: Basic Injection Cycle Based on Rheodyne injector valve Poor peak shape may result if either the connections to injector ports 2 and 3 are reversed or one fills sample loop volume over 50% (in partial-loop mode). Sample from Syringe

Precision Problems Possibly Related to Autosampler Overfilling the sample vial Recommended: No more than 2/3 full Tightly sealed vials not vented Turn venting option on Excessive sampling speed Especially for viscous and low volume samples (<5µL) L): : Reduce speed to slow Bubbles in the autosampler sampling lines or syringe Degas flush solvent Worn rotor seal in the injection valve Results in poor RSD: Replace seal

Detectors for Any Application PerkinElmer Series 200 Detectors UV/Vis PDA UV/VIS considerations: Can detect predominant compounds being analyzed by HPLC Useful for any compounds that absorb between 190 and 700 nm Very good sensitivity (ppb). Typically, better than RI but not as good as Fluor. Exceptional baseline stability Wavelength programmable during run PDA considerations: Approaches UV/VIS sensitivity (ppb) Dual lamps (Deuterium and Tungsten) for full UV/Vis wavelength range Spectral data useful for confirmation and identification of components Capability of peak purity assessment Network-Direct connection for direct transfer of data and spectra to data directories on the network

Detectors for Any Application PerkinElmer Series 200 Detectors FL RI Fluorescence considerations: Very selective (must fluoresce) Exceptional sensitivity (ppt) May have to derivatize if compounds do not fluoresce in native state. Refractive Index considerations: Universal detection Good sensitivity (ppm) Sensitive to temperature changes (RI must have built-in Peltier temp. control; 35 C usually good for most labs) Can not be used for gradient method

Detectors: What to Consider Lamp Energy: Lamp Hours: Wavelength Accuracy: Reference Wavelength (PDA): Flow Cell Condition: RI Reference Purge Valve: Column Connections: Check periodically Lamp lifetime: typically ~1500 hours for UV/Vis, ~1000 for PDA and 500 for FL detectors Check periodically Choose a wavelength transparent to both sample and solvent On PDA, remove flow cell periodically; hold it toward light and check clarity Shine light into UV/Vis flow cell Check position Check that column is connected to Inlet tubing on detector

Detector Pointers Baseline problems related to detectors may be: Mechanical Leaky flow cell, trapped bubble Optical Aging lamp, dirty flow cell Bad mobile phase Not detector Electrical Call vendor Don't wash columns while they are connected to the detector. When possible, the use of a 50psi back pressure device is recommended after the detector (suppresses outgassing)

Suggested Periodic Performance Testing LC PUMPS Flow, composition calibration and pressure Leak test AUTOSAMPLER Precision, carryover, linearity testing DETECTORS Wavelength check Noise and drift testing Linearity testing

Suggested Spare Parts and Replacement Period LC PUMPS High/low pressure piston seals and backup rings kit (~ once a year); Part# N291-0383 Might want to replace high pressure piston every other year AUTOSAMPLER Injector rotor seal (~ once a year) Injector needle (if properly maintained, should hardly ever need replacing) Might want to replace flush and/or sampling syringe periodically (inspect first) DETECTORS Lamp (order new one only after 1000 hours on current UV/Vis or 500 hours on PDA - they have a shelf life)

Chromatographic Diagnostics NO PEAKS: Lamp off/bad, check wavelength Injector problem Check solvent selection / strength No sample in vial CARRYOVER: Materials incompatibility eg., Teflon vs PEEK Insufficient injector flushes or flush speed Column overloaded Some columns exhibit component carryover For that application, switch to another column vendor

Chromatographic Diagnostics Normal noise Expanded; always a good practice before starting a run

Chromatographic Diagnostics 5 min Air bubbles in solvent Degas solvent with in-line degasser or 5min helium sparge then pressurize Aging/bad detector lamp Replace lamp (typically, order new lamp after 500 hours on current one) Damaged pulse dampener on pump Replace

Chromatographic Diagnostics 5 min Contamination in system (solvent, degasser, pump, AS, column, detector) Remove column(s)/guard column and flush any buffers out of system; Rinse all solvent lines with HPLC water (may require co-solvent flush); Place all solvent lines into common bottle of 10% Ammonium Hydroxide; Flush for 20 min at 3mL/min. then switch to HPLC water for 2 minutes; Replace solvent bottle with 10% Nitric Acid and do another 20 min flush; Replace solvent bottle with HPLC water and do another 10 min flush; May also want to perform 20 min./1ml/min Isopropanol flush (usually not necessary); otherwise, move back to original solvents with clean bottles

Chromatographic Diagnostics Noise from detector, surroundings or poor analog cable grounding Monitor lines and recommend adding line conditioner Aging/bad detector lamp Replace lamp Air bubbles in solvent or dirty flow cell Degas solvent / rinse flow cell with strong co-solvent; The latter may require base/acid rinse (previous slide) May arise from fines from column > flush thoroughly without detector or change vendor Especially observed on RI and electrochemical detectors

Chromatographic Diagnostics A B Void volume; solvent front Unretained peak volume/time Elution volume k = 1 (capacity factor) Refractive index effect (typically from flow cell) Replace flow cell or retire detector

Chromatographic Diagnostics Column voids/channeling Bad column: Discard

Chromatographic Diagnostics Sample overload Inject less sample, dilute sample or change diluent to match mobile phase Forgot to add anti-tailing modifier to mobile phase Add Triethylamine (TEA) or similar Deteriorating column Change column Components may be too basic or acidic If using ion exchange column, switch to C18 and ion pairing mode

Example of Fronting Fronting Peak Caused by sample overload or component miscibility problem Inject less sample, dilute sample or change diluent to match mobile phase Check sample miscibility

Chromatographic Diagnostics All could be caused by sample overload Inject less sample, less concentrated sample or change diluent to match mobile phase Change to another column with higher sample capacity

Chromatographic Diagnostics Proportionally increasing retention times Low flow rate If same peak height, solvent leak before injector Column temperature may be lower; use column oven Possible mobile phase change (modifier, ph, molarity); however, not always proportional Possible column deterioration; however, usually not proportional

Chromatographic Diagnostics Disproportional retention time change Mobile phase change (modifier, ph, molarity); Column deterioration; replace column Change in column temperature; use column oven

Chromatographic Diagnostics Proportional drop in peak height Air in autosampler flush line Check lamp energy: lamp energy may be getting significantly lower. Time to change lamp Check wavelength (not always proportional) Sample absorbing to column or other material; change all teflon contact points in autosampler to PEEK material (rotor seal, etc.) or change to other vendor s column Autosampler: leaky rotor seal or plugged needle Volatile components: use capped vials (not always proportional)

Thank you for attending!

Chromatographic Diagnostics Component interconversion or isomerization Check ph, especially relative to pka(s) of components Check Temperature