Keeping anaesthetic pollution to minimum in theatre environment

Similar documents
HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

Recommendations on Checking Anaesthesia Delivery Systems

Checking a Breathing System

Selecting and Connecting Breathing Systems

Anes Machine Use and Pre-use Check - CA1 focus

Flow meter. bellow vaporizer. APL valve. Scavenging system

Anesthesia Machine Intro

Proceedings of the World Small Animal Veterinary Association Sydney, Australia 2007

Breathing Systems. Professor Khalid Bashir

International Journal of Science, Environment and Technology, Vol. 5, No 5, 2016,

Anesthesia. Midmark Animal Health Products

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

Breathing Circuits. Product training

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Anesthesiology Research and Scholarship. Follow this and additional works at:

M5400 Operators Manual

DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE

Operators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500

Figure 3. Miniscav scavenging unit

Section 14: Equipment failure

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

Hallowell EMC Vacuum / Scavenging Interface

Isoflurane: Safe Use Guidelines. (2-chloro-2-(difluoromethoxy)-1,1,1- trifluoro-ethane) is a halogenated ether used for inhalational anesthesia.

Theatre Control Panel. Instructions for Use. January 2017 ASF-29 (1)

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

Working version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee

Medical Instruments in the Developing World

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Anesthesia Breathing Circuit

Effective 2/1/17. SC1 Service Call PER Visit $76.00

Oxygen Dialflow Meter. Instructions for Use

Anaesthetic infrastructure and supplies

BGC HOLDINGS LTD RESPIRATORY PROTECTIVE EQUIPMENT (RPE) POLICY

We Warm Animals. User Manual For. Darvall Heated Smooth-Wall Circuits. Darvall Heated ZDS Qubes

SERVICE KITS AND CALIBRATION NOTES

FLOWMETER UNITS FLOWMETER DEVICES, BY-PASS SYSTEMS

KOALA. Disposable Adult Anaesthetic Circuits for Low Flow Anaesthesia KO.S5.22MM. Adult Anaesthetic Circuits. Medical Pty Ltd

Oxygen Dialflow Meter. Instructions for Use

NARKOVET ANAESTHETIC UNIT

Draeger Fabius GS Pre-Use Check

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

Hartle et al. Checking anaesthetic equipment 2012 Anaesthesia 2012, 67, The pre-use check to ensure the correct functioning of anaesthetic equ

INTEGRUSPSV ANAESTHESIA WORKSTATION USER MANUAL

Hallowell EMC Anesthesia WorkStation Operating Manual

CONTROL OF SUBSTANCES HAZARDOUS TO HEALTH PROCEDURE

COSHH Hazardous Substances Health & Safety Management Standard Issue 2 (September, 2010)

Prima SP2/3 Anaesthetic Machine Range Service Manual Partnership for Life

POLICY FOR THE CONTROL OF SUBSTANCES HAZARDOUS TO HEALTH (COSHH)

Landmark. Veterinary Anesthesia Systems. Operator s Manual P/N

The centerpiece of a comprehensive anaesthesia workstation

Figure 5. Nasal mask in place

BleaseFocus Anesthesia System VERSATILE ANESTHESIA

Pressure Regulator. Instructions for Use

Breathing systems. Non-rebreathing system using a unidirectional valve pair. Non-rebreathing systems

Recommendations - USA

Table Top and Convertible Anesthesia Machine Models: M3000 and M3100 (SHOWN WITH OPTIONAL EQUIPMENT)

Dialflow Regulator. Instructions for Use

Respiratory Protection for Producers

SterlingOSB Material Safety Data Sheet

Narkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

Recommended Minimum Facilities for Safe Anaesthetic Practice in Operating Suites

VIMA. (Volatile Induction and Maintenance Anesthesia) How and Why. James H. Philip M.E.(E), M.D.

o An experienced and trained assistant is available to help you with induction.

Pulmonex II Training Presentation

Respiratory Protective Equipment

ACUC ۰ ROUTINE PROCEDURE

PURPOSE OF THE POLICY

የIትዮዽያ የደረጃዎች ኤጀንሲ ETHIOPIAN STANDARDS AGENCY ES DRAFT UPLOAD REQUEST FORM FOR PUBLIC COMMENTS

Nitrous oxide Quiz 2017 Answers

Standard Specification for Anesthetic Gas Scavenging Systems Transfer and Receiving Systems 1

The right atmosphere for health. A system for medical gas supply

Operate Breathing Apparatus

GSA POLICY ON LOCAL EXHAUST AND OTHER VENTILATION (FUME CUPBOARDS)

Aespire 7100 Essential performance Compact design

Unit 15 Manual Resuscitators

Anesthesia Delivery Systems What Do I Need to Know About My Machine? A Workshop James H. Philip, MEE, MD, CCE

Minimum equipment list for cardiopulmonary resuscitation Primary dental care

SIGNET 615 ANAESTHESIA WORKSTATION USER MANUAL

Micro Dial-Flowmeter. Instructions for Use

MATERIAL SAFETY DATA SHEET

Scottish House Builders Health and Safety Forum Meeting February 2013

CONTRAfluran. Contrafluran TM -Anaesthetic Gas Filter with Standard Mount. Instruction for Use

Forced or Mechanical Ventilation - which uses mechanical supply and/or extraction to provide fresh air and is controllable.

Cambridge Refrigeration Technology

PORTER INSTRUMENT COMPANY, INC.

Homecare Pneumology NEONATOLOGY ANAESTHESIA INTENSIVE CARE VENTILATION Step beyond the future

A200SP Circle Absorber Service Manual

Recommendations for Standards. Monitoring. Cardiopulmonary Bypass

Safety Data Sheet. Bromotrifluoromethane (Halon 1301, R13B1) Version : 1

E3628 THE RESPIRATORY PROTECTION PROGRAM: EMPLOYEE TRAINING. Leader s Guide ERI Safety Videos EMPLOYEE TRAINING

General Health & Safety Information Sheet

VAPOMATIC ANESTHETIC VAPORIZER

INSTALLATION AND OPERATION OF THE HYPERBARIC CHAMBER

Update in Anaesthesia 31

THE GE AISYS ANAESTHETIC MACHINE

Toolkit for safe Use of Microbiological Safety Cabinets(MSC s)

Welcome to Maxstar Medical

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS

Transcription:

Vet Times The website for the veterinary profession https://www.vettimes.co.uk Keeping anaesthetic pollution to minimum in theatre environment Author : Gianluca Bini Categories : General, Vets Date : September 5, 2016 Atmospheric pollution in anaesthetic practice is of paramount importance and cannot be stressed enough. Figure 1. Male (blue) and female (green) 30mm connectors of a collecting system. Anaesthetic gases and vapours are a potential pollutant in veterinary practices, with a prolonged lifetime of up to 120 years in the atmosphere 1. They impact not only on ozone depletion and, consequently, on global warming, but also on the health of operating theatre staff. In 2014, 2 of the 30 human adverse events reported involving veterinary surgeons and veterinary nurses in the UK were due to inhaled anaesthetic exposure 2. This article aims not to debate the effects of inhalational agents, but describe how to avoid pollution in practice. It is important to remember UK employers are legally required, under the Control of Substances Hazardous to Health (COSHH) regulations 3, to control medical pollution and avoid passing the workplace exposure limits to inhalational agents set by the Health and Safety Executive 4. Scavenging system 1 / 10

Figure 2. A female 22mm connector (white) and female 30mm connector (green). The COSHH code of practice recommends putting in place the most effective and reliable control options, to keep substances hazardous to health in the working environment to the smallest possible amount. This can be achieved with the installation of a scavenging system. Moreover, the RCVS Practice Standards Scheme sets the presence of a scavenging system and regular pollution control as a core requirement for veterinary practices 5. A scavenging system collects waste gases from the breathing system and releases them outside the building at a safe distance. Two types of scavenging systems essentially exist passive and active. Both are composed of a collecting, transfer, receiving and disposal system. The waste gases from the exhaust port of the anaesthetic machine, ventilator or adjustable pressure-limiting valve of a breathing system move through the collecting system, which is usually composed of 30mm male and female conical connectors (Figure 1). In addition, certain collecting systems include a pressure relief valve. This valve is usually set to open if the pressure is equal to, or greater than, 10cmH 2 O. The connector s diameter is deliberately different to the conventional 15mm or 22mm tubing used for breathing systems (Figure 2) to avoid wrongful connection. 2 / 10

Figure 3. An open receiver cylinder with filtered openings. 3 / 10

Figure 4. The filtered opening of a receiver. Dust can be noted on the filters. 4 / 10

Figure 5. An open receiver showing a non-fluctuating (a) and fluctuating (b) purple disc. Figure 6. The switch of an active scavenging system. 5 / 10

Figure 7. Ventile. 6 / 10

Figure 8. Charcoal adsorber. In human medicine, the use of a collecting system with the same diameter of a breathing circuit has been reported to potentially cause serious accidents6. Ultimately it is important the collecting system does not result in any resistance to expiration, especially in veterinary medicine where we often deal with very small animals. A tube from the collecting system to the receiving system makes up the transfer system. Both connectors of this tube have a 30mm diameter for the same aforementioned reason. Moreover, this tube should be as short as possible to prevent any risk of kinking 6. The pressure increase in the system caused by a kink in the tube can lead to barotrauma and potentially fatal consequences, 7 / 10

particularly in small patients. The receiving system is basically a safety device, protecting the patient from excessive negative or positive pressure. Two types of receiving systems exist an open system is used if an active scavenging system is in place, while the receiving system is a closed type if a passive scavenging system is used. The most important part of the receiving system is the reservoir. In the open system, this is a cylinder of rigid material (Figure 3), with one or more openings acting as pressure breaker between the patient breathing system and the final stage of the scavenging system. Some models have a filter over the openings to stop debris entering. Dust can accumulate on these filters (Figure 4), so they should be cleaned monthly. Some reservoirs have a coloured disc that fluctuates or lifts when the scavenging system is active (Figure 5). In the closed system, this part is usually a reservoir bag and two valves, one for negative and the other for positive pressure relief. Complications may occur if a receiving device is not in use due to several factors, including negative pressure from either the fan of the active scavenging system or the wind on the exterior tubing of the passive system, or positive pressure if the tubing is kinked. Disposal Figure 9. A leak test on a rebreathing system. Close the adjustable pressure-limiting valve (1) and press the oxygen flush button while occluding the opening of the breathing system with your thumb (1), until the reservoir bag is completely inflated and distended (3), and wait 30 seconds. If the bag is inflated and distended as before, the circuit does not have a leak. Once finished, open the valve (4). The disposal system can be active or passive. In both cases, the end part of the disposal system is positioned outside the building, in a rarely attended location. An active disposal system requires a fan or exhaust unit creating negative pressure and a vacuum regulating valve controls pressure fluctuations in the system. The BS EN ISO 7396-2:2007 8 / 10

standard defines a range of 50L/min to 80L/min for high flow and 25L/min to 50mL/min for low flow active scavenging systems. The switch (Figure 6) is usually placed inside the practice, in an easy to reach position, although an active disposal system may be working at all times. A passive disposal system is made of a wide-bore tube passing through a wall or the roof. This can either terminate outside or in a ventile (Figure 7). A ventile uses wind to move the exhaust gases from the system. A short, wide tube is preferred to minimise resistance. It is important to choose the best position for the external opening of the tube. A non-windy position should be chosen; if not, suction can be created inside the tube. The opening should be covered with a filter to prevent insects, small animals or debris entering. Passive scavenging systems are less expensive to set up and maintain and commonly used in small premises, while active scavenging systems may be more convenient in large practices. Adsorber Another option commonly employed in veterinary practices to control operating theatre pollution is using an adsorber (Figure 8). These are canisters containing about 1kg of activated charcoal, but their weight increases as they adsorb vapour. Such adsorbers cannot deal with nitrous oxide. Their capacity can vary between brands and individual canisters 7. They are disposed of by incineration and should be stored safely to prevent leaks. They should be weighed regularly and disposed of when the canister reaches the weight indicated on the label. Other pollution factors Figure 10. The Association of Veterinary Anaesthetists safety checklist. 9 / 10

Powered by TCPDF (www.tcpdf.org) Other factors impacting on pollution in clinical areas are filling vaporisers, patients recovering from anaesthesia, mask or box inductions and capnography outlets (when a line to reintroduce the gases and vapours into the system is not in use). For the correct management of these, vaporisers with a key-fill system and adequate ventilation are important. Between 15 and 25 air changes per hour are suggested, which can be addressed using a correctly maintained air conditioning system, keeping in mind recirculating or partially recirculating air conditioning systems may affect their extracting efficacy. Checking the anaesthetic machine and breathing system, including a leak test (Figure 9), should be performed at least every day. Anaesthetic safety checklists, such as those produced by the Association of Veterinary Anaesthetists (Figure 10), are freely available 8. Even if a scavenging system is in place, anaesthetic gases can leak into the room. Ultimately, it is important to remember the use of rebreathing systems, such as the circle system combined with low-flow anaesthesia, will reduce waste gas production thereby reducing the economic and environmental cost of anaesthesia. References 1. World Meteorological Organization/United Nations Environment Programme (1998). Scientific Assessment of Ozone Depletion: 1998. 2. VMD (2016). Veterinary Medicines Pharmacovigilance Annual Review 2014. 3. Health and Safety Executive (2013). Control of Substances Hazardous to Health: Approved Code of Practice and Guidance (6th edn), HSE Books, London. 4. Health and Safety Executive (2011). EH40/2005 Workplace Exposure Limits (2nd edn), HSE Books, London. 5. RCVS (2015). Practice Standards Scheme Modules and Awards: Small Animal, http://bit.ly/2bfpb6y 6. Farquhar-Thomson DR and Goddard JM (1996). The hazards of anaesthetic gas scavenging systems, Anaesthesia 51(9): 860-862. 7. Smith JC and Bolon B (2003). Comparison of three commercially available activated charcoal canisters for passive scavenging of waste isoflurane during conventional rodent anesthesia, Contemporary Topics in Laboratory Animal Science 42(2): 10-15. 8. Association of Veterinary Anaesthetists safety checklist: www.ava.eu.com/resources/checklists 10 / 10