Chapter 2. Pipetting techniques. Guide to Pipetting

Similar documents
Adjust the Volume Display

Improve pipetting results in pharmaceutical formulation by using MICROMAN E

Improving Pipetting Techniques

PIPETMAN F ENGLISH. Other languages available on our website:

UNCC Biotechnology and Bioinformatics Camp. Dr. Jennifer Weller Summer 2010

Pipettor. User Manual

MICROMAN ENGLISH. Other languages can be consulted and downloaded on our website,

Pipetting Small Volumes

RayPette Plus Autoclavable Pipette. User Manual

The pipettes cover a volume range from 0.1µl to 10ml.

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

Other languages available on our website:

The pipette of Choice for Problem Liquids!

Lab 1. Instrumentation Familiarity: Using Micropipetters and Serological Pipettes

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014

MULTICHANNEL. Labnet International, Inc.

Chapter 1. Selecting the best pipette for your application

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING

A N F G J I C D D L A B C D E F

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

25/04/56. LTC Carmen A. Bell Chief, Laboratory Operations USAMC-AFRIMS

User s Manual. Autoclavable & UV resistant. Multi-channel digital micro pipette. In Vitro Medical Diagnostic Devices (98/79/EC)

Standard Operating Procedure. Air Displacement Pipette Calibration

Ergonomic Pipetting for Problem Liquids!

Operating Instructions. Rainin Classic Manual Pipette

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

A N B P F G J I C D D L A B C D E F

User s Manual. Digital micro pipette for liquid handling CERTIFIED ISO9001

User s Guide VARIABLE VOLUME MODELS. Variable volume single and multichannel models and fixed volume models

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

Pipetting and Determining Protein Concentration

WITOPET PREMIUM PIPETTE USER S GUIDE

Varipette Operating manual

User s Manual. Autoclavable & UV r e s i s t a n t. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC)

NEO. User s Guide. Do not lubricate this pipette. The use of lubricant cancels the warranty of this pipette. NOTICE

Prepared By: Lisa McAvoy 30/Nov/17 dd/mmm/yy. Approved By: Veronica Harris-McAllister 30/Nov/17 dd/mmm/yy

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Comfortable and Secure by Nature!

Using a micropipette to transfer exact liquid measurements in scientific experiments By: Nicole Hume

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

Gel Loading Practice MiniLab

Transferpette -8/-12

MOLEBIO LAB #1: Microquantity Measurement

VWR HIGH PERFORMANCE PIPETTOR

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

SOP-111 Pipetting Small Volumes

A N B P G J I C D L D E

SoCal BioMed, LLC 4695 MacArthur Court, 11th Floor Newport Beach, CA Phone: +1 (424) Fax: +1 (323)

Inquiry Module 1: Checking the calibration of a micropipette

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

CHM250 Calibration and Measurement Lab. Balance Calibration

Eliminating Sources of Pipetting Error in the Forensic Laboratory David M. Epstein Director of Scientific Operations. Abstract.

Liquidator 96 Manual benchtop pipetting system

Fisherbrand. Finnpipette INSTRUCTIONS FOR USE PARK LANE Pittsburgh PA

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

Nichipet EX Plus II. User s Manual. Autoclavable & UV resistant. Digital micro pipette for liquid handling

Transferpette. Lightweight precise rugged reliable!

Transferpette. Lightweight precise rugged reliable!

APPENDIX A CORRECT USE OF DIGITAL PIPETTES

Air Displacement Pipetting Modules. Total Integrated Liquid Handling

Register your instrument! Eppendorf Research plus. User Adjustment

Register your instrument! Eppendorf Xplorer / Eppendorf Xplorer plus. Adjustment

User s Manual. Autoclavable & UV resistant Nichipet EX II. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC)

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

CALIBRATION AND ADJUSTMENT

Qi Litre Ultrasonic Cleaner

Contents Page Warranty

Air Displacement Pipetting Module. Total Integrated Liquid Handling

Biochemistry Laboratory I CHEM 4401 Pipetting & Scales

Experiment 1 Use of Micropipettes: measuring volume accurately

Decontamination Procedures for Gilson Pipettes

Eppendorf Research fix variable multi Instruction Manual Bedienungsanleitung Mode d emploi Instruzioni d impiego Manual de Instrucciones

High-Throughput. Faster, Better 96 & 384-Well Pipetting Ultra Efficient and Extremely Versatile

CHEM 321 Experiment 1

VOLUMETRIC TECHNIQUES

Real World Metric System

DISCOVERY PRO - features

Technical Procedure for General Laboratory Equipment

How to Measure R7.1. Reference. I. Linear dimensions

Calibration of Volumetric Glassware

Userts Manual. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC) Annex III self-declared ISO 8655 STANDARD

Lab Equipment ANALYTICAL BALANCE

SoftGrip Single Channel Pipettes User Manual New Photo To Come

Performance Verification of BioTek s Precision 2000 using Artel s Multichannel Volume Verification (MVV ) System

VITLAB micropipette. Standard Operating Procedure (SOP)

TMArrayer Users Guide

Good Pipetting Practice

Pipette Operation Guide

Corning Step-R. Repeating Pipettor. Instruction Manual

Ensuring pipette & weighing device performance: Verification & Calibration

VITLAB Volumetric instruments

[USING THE VITROBOT April 9, Using the Vitrobot

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS

BIOSAFETY SELF AUDIT PRINCIPAL INVESTIGATOR/PERMIT HOLDER: CONTAINMENT LEVEL: 1 2 (Shaded) RISK GROUP: 1 2 AUDIT TO BE COMPLETED BY (DATE):

Transcription:

Chapter 2 Pipetting techniques 16 Guide to Pipetting

2.1 Adjust the volume display CHAPTER 2 Reading and adjusting the volume The volume is shown on the volumeter Hold the body of the micropipette in one hand and use the other hand to rotate the thumbwheel or the push-button. With the push-button, the volume can be easily adjusted with one hand. Push-button volume adjustment is available on all MICROMAN pipettes and on PIPETMAN pipettes (except PIPETMAN L) manufactured after April 1995. Push-button Thumbwheel Volume indicator A helpful hint for improving reproducibility and accuracy Always finish setting clockwise for best reproducibility. This is how to obtain a clockwise volume setting: When decreasing the volume setting, turn the thumbwheel slowly until the desired setting is displayed. Do not overshoot the required position. When increasing the volume setting, rotate the thumbwheel approximately 1/3 of a turn above the desired setting, and then slowly turn back to decrease the volume until you reach the desired setting. 1 0 0 Incorrect alignment: error 1 0 0 Correct alignment: accurate reading To avoid parallax, hold the pipette in a horizontal position. Adjust the volume until the indicator is lined up with the desired volume. IMPORTANT To avoid internal damage to your pipette, never attempt to force the volume setting beyond the limits. More information available at www.gilson.com 17

2.2 Forward or reverse mode pipetting Air-displacement / Forward mode The forward mode is the standard way of pipetting with an air-displacement pipette like PIPETMAN. Preparation Aspiration Dispense Purge Hold the instrument in a nearly vertical position. Depress the plunger smoothly to the first stop position. Immerse the pipette tip in the liquid*. Allow the plunger to move up smoothly to the rest position. Wait one second so that all the liquid has time to move up into the tip. Place the pipette tip at an angle (10 to 45 ) against the inside wall of the receiving vessel. Depress the plunger smoothly to the first stop position. Wait one second, then depress the plunger to the second stop position. This purge stroke removes any remaining sample from the tip. Remove pipette tip end from sidewall by sliding it up the wall. Rest position First stop Second stop or purge 18 Guide to Pipetting

CHAPTER 2 In general, precision in forward mode depends on precise draining by air pressure (air-displacement pipettes) or internal wiping of the pipette barrel (positive-displacement pipettes). Home Allow the plunger to move up to the rest position. Volume Immersion depth (mm) 0.1-1 µl 1 1-100 µl 2-3 101-1000 µl 2-4 1001 µl - 10 ml 3-6 Pre-rinsing To obtain greater uniformity and precision of dispensing, it is better to provide identical contact surfaces for all aliquots. This is done by pre-rinsing with the same liquid as the one dispensed. For pre-rinsing, aspirate with the tip, and then dispense back into the original reservoir or to waste. Pre-rinse again when adjusting the volume. * The immersion depth of your tip can have a significant effect on your results (for depth per model, see table above). If the tip is immersed too deeply, droplets will form on the outside of the tip and they will be deposited along with your sample. If the tip is not immersed deeply enough, vortexing will occur and your pipette will not aspirate the selected volume. More information available at www.gilson.com 19

Forward or reverse mode pipetting Air-displacement / Reverse mode The reverse mode is only possible with air-displacement pipettes. It is sometimes used to pipette slightly viscous liquids. Preparation Aspiration Dispense Hold the instrument in a nearly vertical position. Depress the plunger smoothly to the second stop position. Immerse the pipette tip in the liquid. Allow the plunger to move up smoothly to the rest position. Wait one second so that all the liquid has time to move up into the tip. Place the pipette tip at an angle (10 to 45 ) against the inside wall of the receiving vessel. Depress the plunger smoothly to the first stop position. Wait one second. Rest position First stop Second stop or purge 20 Guide to Pipetting

Re-aspiration Complete purge If the pipette tip is to be reused for the same sample, maintain the plunger in the intermediate position for subsequent immersion for the next pipetting cycle and restart operation 2. Wait one second and purge. If the pipette tip is not to be re-used, depress the plunger to purge position over an appropriate waste container and then eject the tip. or More information available at www.gilson.com CHAPTER 2 In reverse mode pipetting, the purge stroke is used during preparation. During aspiration, an amount of liquid equal to the amount of purged air is added. This amount compensates for the liquid that remains inside the tip during dispensing. 21

Forward or reverse mode pipetting Positive-displacement / Forward mode Preparation Press the plunger button to the first stop. The piston moves to the appropriate position. Aspiration Immerse the capillary/piston in the liquid*. Release the plunger, letting it move up to the home position. The piston moves up and the ambient pressure forces the desired volume of liquid through the orifice into the capillary. Dispense Press the plunger button to the first stop. The piston moves down and expels the liquid out of the capillary. Rest position First stop Ejection 22 Guide to Pipetting

Ejection Press the plunger all the way down to the second and last stop. Capillary and piston are ejected without hand contact. Wiping To avoid distorted results linked to the volume pipetted, ensure that no liquid is on the outside part of the tip. If necessary (viscous liquids such as cream), wipe the outside of the tip or the capillary with a clean medical wipe. Do not touch the orifice. Choose a tissue which is resistant, lint-free, and inert to acids and solvents. Dispose of the tissue in a safe, hygienic manner. IMPORTANT When working with high risk specimens, do not wipe the disposable part. Make sure fluid depth penetration does not exceed the recommended immersion depth. Pre-rinsing Warning! Greater uniformity and precision of dispensing are usually obtained by providing identical contact surfaces for all aliquots. This is achieved by pre-rinsing with the same liquid that is being dispensed. To pre-rinse, aspirate with the tip and then dispense back into the original reservoir or to waste. * The immersion depth of your tip can have a significant effect on your results (for depth per model, see table page 19). If the tip is immersed too deeply, droplets will form on the outside of the tip and they will be deposited along with your sample. If the tip is not immersed deeply enough, vortexing will occur and your pipette will not aspirate the selected volume. More information available at www.gilson.com 23

2.3 Tips for mistake-free pipetting How to avoid typical pipetting mistakes Many factors may impact pipetting accuracy Influencing parameters and effects Leaky/poorly seated pipette tips may affect accuracy by 0.5 to 50 % Reuse of pipette tips may affect accuracy by up to 4 % The straightness of pipette tips may affect accuracy by up to 10 % The difference in vapor pressure of the liquid to be pipetted versus that of the water used for adjustment may affect accuracy by up to 2 % Failure to wipe pipette tip on the vessel wall can affect accuracy by up to 3 % Pipette tip immersion depth and handling angle during pipetting may affect accuracy by up to 1 % Irregular rhythm and timing during pipetting can affect accuracy by up to 1.5 % A leaky piston system can affect accuracy by 1 to 50 % Uneven piston movement can affect accuracy by up to 0.5 % Corrective measures Using original or recommended pipette tips Using pipette tips only once Using appropriate tips only Sufficient pre-wetting of pipette tips Wiping of the pipette tip on the vessel wall (wiping distance 8 to 10 mm)* Holding pipette in a vertical position while pipetting Applying a consistent pipetting technique Regularly checking the pipette and the volume aspirated Smooth operation of piston Information extracted from ISO 8655-2 - Annex B * Gilson recommends touching the tip to the vessel wall at an angle of 10 to 45. 2.4 Pipetting ergonomics Take a few minutes to get organized and ensure you have: 1 - An appropriate posture 2 - The right material Gilson offers various pipettes with forces adapted to user preferences. The forces of PIPETMAN L are some of the lowest. 3 - The appropriate gestures 4 - A good work organization A good test is to see if you can rest your elbow comfortably on the work surface. If not, your receptacle may be too low or too high, find the right height. Download the Gilson Ergonomics Poster for increased working comfort in your lab www.gilson.com/resources/ergonomics.pdf 24 Guide to Pipetting

CHAPTER 2 Take time to relax 1 - If possible, try to switch periodically between different types of work. Use a pipette holder Protect your pipette and always store it vertically on a pipette holder. Pipettes left on a work bench or stored in a drawer can easily come into contact with samples and become contaminated. 2 - Keep an appropriate, unrushed working speed. Let go of the pipette from time to time and give your fingers/ hand a (micro) break. 3 - Take frequent short breaks. Change your sitting position. Lean back and relax your shoulders and arms. Special attention should be paid to smooth pipetting 1 - To facilitate uniform timing and motion, keep all necessary items within arm's reach. 2 - Place the most frequently used objects in front of you. The more rarely used items can be placed a little further away from you. 3 - The opening of the receptacle for used tips should be at the same height as the end of your pipette. More information available at www.gilson.com 25