Alternating Pressure / Low Air Loss Mattress System

Similar documents
USER MANUAL. AltaDyne XT Model A AltaDyne XD Model A Alternating Pressure / Low Air Loss Replacement Mattress System

Med Aire Alternating Pressure Pump and Pad System

Paradise Pump Series Operator s Manual

770000S CLINICAL MAX LOW AIR LOSS WITH ALTERNATING PRESSURE SYSTEM USER MANUAL

SC505 Sof Care Inflator. Operating Instructions and Service Manual

Diamond 8 Plus. Alternating + Micro Low Air Loss Pressure Relief System. User Manual. Distributed by: Quart Healthcare Inc.

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual

SELECT SERIES LS200 ALTERNATING PRESSURE / LOW AIR LOSS MATTRESS SYSTEM

Instruction Manual for Configura Cushionair Portable Pump

Operation Manual. H & R Healthcare 1750 Oak Street Lakewood, NJ Phone: Fax:

Instruction Manual. Low air loss/alternating pressure mattress

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS

Instructions for Cleaning & Maintaining Air Compressor

Air Plus User Manual

Pegasus Aircare Mattress Overlay System Models 2200 and 2201

APH166 Morpheus Air USER MANUAL. Apollo Healthcare Technologies Ltd. Holme Street, Liversedge, West Yorkshire, WF15 6JF.

Key Product Features. Primary Use: Aggressive Treatment of Pressure Ulcers

SELECT SERIES LS300 ALTERNATING PRESSURE / LOW AIR LOSS MATTRESS SYSTEM

AltaDyne AutoFloat. Alternating Pressure Management Plus. Model # Model # OPERATORS MANUAL

Oxygen Concentrator Instruction

Operator s Manual. Dynamic Low-Air-Loss Therapy with Alternating Pressure

MA90 Series Mattresses. MA95 Series Mattresses

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

Hi Flow Stroller H I F L O W S T R O L L E R

VERTICAL AIR COMPRESSORS

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

VERTICAL AIR COMPRESSORS

Stroller. Sprint S T R O L L E R S P R I N T

Model 1700 Patient Operating Instructions. Model 1700

MA60 Series. MA60 Alternating Pressure System. MA65 Alternating Pressure On-Demand Low Air Loss System. Owner s Operator and Maintenance Manual

ORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual

Companion 5 TM. Oxygen Concentrator. User Manual

PRESSUREGUARD APM OWNER S MANUAL. PressureGuard APM Bariatric. Span-America Medical Systems, Inc. Greenville, SC

Portable oxygen concentrator

PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100

77-I0PN230 Rev. A. Model 1030 User Manual

Operator s Manual. The Bullet Blender Gold BB24-AU, BB5E-AU

User Guide Antidecubitus systems EURO BASIC

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

PressureGuard APM 2 and PressureGuard APM 2 Safety Supreme

User Manual DIRECTHEALTHCARESERVICES.CO.UK

(DynaLaL III) Dynamic True Low Air Loss Mattress Replacement System

Oxygen Concentrator. User Manual

CAUTION: YOU MUST READ THE FOLLOWING BEFORE OPERATING THE SNOWMAN.

! Warning, refer to accompanying documents.

AKRON CHILDS TILT TABLE

User Manual. This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.

12 Volt Inflator. Operator s Manual. Model: ICC12F, ICC12FC, ICC12FCN, ICC12FCM, ICC12N

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

Our technical information is based on Tip: calls attention to special information.

COMPACT AIR COMPRESSOR MODEL NO OWNER S MANUAL

Welcome to the World of In-Home Gardening! Simple Steps to Get Your AeroGarden Up and Growing (No Tools Required)

LYMPHA PRESS MANUFACTURED BY MEGO AFEK IMPORTED BY LYMPHA PRESS USA, Ltd. USERS MANUAL LYMPHA PRESS MODEL 201-M

Regalia Oxygen Concentrator

Contents. Introduction...1 About Breeze... 1 Breeze Overlay... 2 Breeze Mattress Replacement... 2 Breeze Pump... 4 Controls and Indicators...

User Manual. soft air simplex wds. Health. Security. Independence. Pressure ulcer prevention and treatment-system Assembly and Operation instructions

COMPRESSOR AND INFLATOR WITH TORCH

DOLOMITE Legacy. Rollator User Manual

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

In Vivo Scientific, LLC INSTRUCTION MANUAL

12V AIR STATION ALL-IN-ONE

O P E R A T O R S M A N U A O P E R A T O R S M A N U A L L

User Manual Models: S36, S40, S50

OPERATION AND MAINTENANCE INSTRUCTION MANUAL

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank

Welcome to the World of In-Home Gardening! Simple Steps to Get Your AeroGarden Up and Growing (No Tools Required)

450P- RV AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

STAGE, STADIUM & X-TREME SHOT.

LUBRICATOR ASSEMBLY AND OPERATING INSTRUCTIONS

MODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE

USER MANUAL. AirPal Air Supply. (AS-1100 North America) 1488 Limeport Pike, Coopersburg, PA airpal.com fax

microair MA65 Series Alternating Pressure Low Air Loss Mattress User Manual MA65, MA65B42, MA65B48, MA65RSR

AIR COMPRESSOR 12 V DIRECT DRIVE MULTI-FUNCTION INSTRUCTION MANUAL. Model no

User Manual DIRECTHEALTHCARESERVICES.CO.UK

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

Panacea AIR PRO MATTRESS. Owner s Manual. Please keep and refer to this Owner s Manual.

User Manual. 10. Dimensions and Ordering Codes. Integrity Alternating Pressure Mattress System. Alternating Pressure Mattress Systems

DYNAMIC COMPRESSION THERAPY

Operation Manual Airbrush Compressor AS196A ARTICLE 34207

TrekMill USER'S MANUAL MM5050

INFLATION STATION. Cordless OWNER S MANUAL MODEL NO: RCP-C65B

AIR COMPRESSOR. Failure to follow all instructions as listed below may result in electrical shock, fire, and/or serious personal injury.

EC DECLARATION OF CONFORMITY

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

450P AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

[ACT 005] Manual E 5/25/05 9:24 PM Page 3 MULTI-FUNCTION

200 PSI HIGH-FLOW AIR SOURCE KIT

200 PSI FAST-FILL AIR SOURCE KIT

VENTI-O 2 plus. O 2 switching valve for: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2

R E D I C O N T R O L S

OC4000 USER'S GUIDE I O OXYGEN CONCENTRATOR FOR VETERINARY USE ONLY. by Supera Anesthesia Innovations GLOSSARY OF SYMBOLS. : ON (power switched on)

SILENTAIRE TECHNOLOGY

Operating Instructions Part No

IMMERSION SUIT MAINTENANCE GUIDE IMMERSION SUIT MAINTENANCE GUIDE 1409 SERIES. SCMG003A Maintenance Guide - Immersion Suits USA ENGLISH.

Comfortmax Thermax Hot/Cold Water Circulation System. Instructions for Use

Arterial Puncture Wrist

PS Plasma 375 Plasma Cutter Kit Assembly & Operating Instructions

Transcription:

750000 Alternating Pressure / Low Air Loss Mattress System User Manual Important: Do not operate the Mattress System without first reading and understanding this manual! Save this manual for future use. 750000-INS-LAB-RevA08

Contents Introduction... 3 Intended use of this device... 3 Important safety precautions... 3 Danger... 4 Warning... 4 Caution... 5 Equipment Symbols... 6 Product description... 7 Pump and mattress system... 7 Pump: front... 7 Pump: rear... 8 Pump: top panel... 8 Installation... 9 Unpacking... 9 PUMP and MATTRESS INSTALLATION... 9 Operation... 11 Pressure adjustment... 11 Quick disconnect... 12 Transport cap... 12 CPR function... 12 Maintenance... 12 Cleaning... 12 GENERAL Maintenance... 13 Fuse replacement... 13 Air filter replacement... 14 Storage... 14 Specifications... 14 Warranty... 15 GF, Graham-Field, Lumex and AltaDyne are registered trademarks of GF Health Products, Inc. GF Health Products, Inc. is not responsible for typographical errors. Packaging, warranties, specifications and products are subject to change without notice. 750000-INS-LAB-RevA08

Introduction Use this manual for set-up and operation of the Alternating Pressure / Low Air Loss Mattress System. Read all instructions before using the mattress system. Save this manual for future reference. Intended use of this device The intended use of the Alternating Pressure / Low Air Loss Mattress System is: The treatment of pressure ulcers while optimizing user comfort. Pain management as prescribed by a physician. As described above, in either a homecare or long-term care setting. Important safety precautions The safety statements presented in this chapter refer to the basic safety information that should be observed by those using the Alternating Pressure / Low Air Loss Mattress System. There are additional safety statements in other chapters or sections, which may be the same as or similar to the following, or specific to the operations. DANGER: Indicates an imminent hazard situation that, if not avoided, will result in death or serious injury. WARNING: Indicates a potential hazard situation or unsafe practice that, if not avoided, could result in death or serious injury. s Caution: Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage. Note: Provides application recommendations or other useful information to ensure that you get the most from your product. 750000-INS-LAB-RevA08 3

Danger To reduce the risk of electrocution: DANGER: Always unplug this product immediately after using. DANGER: Do not use this product while bathing. DANGER: Do not place or store this product where it can fall or be pulled into a tub or sink. DANGER: Do not place this product in or drop into water or other liquid. DANGER: Do not reach for a product that has fallen into water. Unplug it immediately. Warning To reduce the risk of burns, electrocution, fire, or personal injury: WARNING: Do not leave this product unattended when plugged in. WARNING: Always use close supervision when this product is used by, on, or near children or those who require close supervision. WARNING: Use this product only as intended and described in this manual. Do not use attachments or accessories not recommended by Graham-Field. WARNING: Never operate this product if a) it has a damaged cord or plug b) it is not working properly c) it has been dropped or damaged d) it has been dropped into water. Return the product to your Graham-Field equipment provider. 4 750000-INS-LAB-RevA08

WARNING: Keep the cord away from heated surfaces. WARNING: Never block the air openings of this product or place it on a soft surface, such as a bed or couch, where the openings may be blocked. Keep the air opening free of lint, hair, and other similar particles. WARNING: Never drop or insert any object into any opening or hose. WARNING: Do not use this product in the presence of flammable anesthesia, or where oxygen or nitrous oxide is being administered. WARNING: Connect this product only to a properly grounded power outlet. WARNING: Maximum patient weight capacity for this product is 350 lb (159 kg). Caution s Caution: The pump can be used only with the accompanying mattress. Do not use it for any other purpose. 750000-INS-LAB-RevA08 5

Equipment Symbols Attention! Consult accompanying documents (this manual) Indicates that this product is according to the degree of protection against electric shock for type BF equipment Functional earth (for UL only) Class II (for UL only) Disposal of Electrical & Electronic Equipment (WEEE): Do not treat this product as household waste. For more detailed information with regard to returning and recycling this product, please consult your local city office, household waste disposal service, or Graham-Field Health Products equipment provider. For U.S. and Canada only E228589 53DG Medical Equipment: Air Pump with respect to electrical shock, fire and mechanical hazards only in accordance with UL60601-1 and CAN/CSA C22.2 No. 601.1 6 750000-INS-LAB-RevA08

Product description Pump and mattress system foot end symbol CPR valves mattress pump mattress-pump connectors pump and mattress system Pump: front quick connector 750000-INS-LAB-RevA08 7

Pump: rear Pump: top panel static function switch main power switch low pressure indicator normal pressure indicator pressure adjustment knob Pressure adjust knob: The pressure adjust knob controls the air pressure output. Turn it clockwise to increase output pressure (firmer); turn it counter-clockwise to decrease air output pressure (softer). Higher pressure output will better support the heavier-weight user. The setting is measured in kilograms (kg). Please consult your physician for an appropriate setting. Normal pressure indicator: The green normal pressure LED illuminates when the pressure inside the air mattress has reached the selected pressure setting. Low pressure indicator: The yellow low pressure LED illuminates when the pressure inside the air mattress is below the selected setting. 8 750000-INS-LAB-RevA08

Static function switch: Press the static function switch to suspend alternating function, if needed; the pressure inside the air tubes will then be adjusted to identical pressure. Press the static function switch again to change back to alternating mode. The static function switch toggles between static and alternating modes. Main power switch: The main power switch toggles between ON and OFF. When power is ON, either the normal or low pressure indicator illuminates. Installation Unpacking 1. Before unpacking, check for obvious damage to the carton or its contents. If damage is evident, please notify the carrier and your Graham-Field equipment provider. 2. Remove all loose packing from the carton. 3. Carefully remove all components from the carton. 4. Inspect all components. If damage is evident, please notify the carrier and your Graham-Field equipment provider. PUMP and MATTRESS INSTALLATION 1. Remove all sheets, blankets, and pillows from the bed. 2. Place the mattress on top of the bed frame; ensure that the foot end symbol faces up at foot end as shown at right. 3. Fold open the wire hangers on the back of the pump as shown at right. 4. Either hang the pump on the bed's foot end as shown at right, or 750000-INS-LAB-RevA08 9

place the pump on a flat surface easily accessible to the user, caregiver, and/or doctor. s Caution: Place the device in a position where the user, caregiver, and/or doctor can access it easily. 5. Connect the mattress-pump air hose quick connector; ensure that the connector has clicked securely into place before continuing. s Caution: Ensure that the air hoses are not kinked or tucked under the mattress. WARNING: Locate all cords so that they will not be stepped on, tripped over, or otherwise subjected to damage or stress. WARNING: Do not use a household extension cord if the electrical cord does not reach the power outlet. Use of an improper extension cord could result in fire and electric shock. If an extension cord must be used, use a three conductor cord with ground, properly wired, in good electrical condition, and keep it as short as possible. WARNING: Ensure that the local power voltage is appropriate for the pump unit. 5. Plug the power cord into a properly grounded electrical outlet. 10 750000-INS-LAB-RevA08

Operation 1. Turn the power switch to the ON position. 2. Adjust the pressure adjust knob to the maximum position. Note: Set the pressure adjust knob to the maximum position for the quickest inflation at set-up; the mattress can then be adjusted to the desired pressure. 3. The pump will begin delivering air into the mattress. While the mattress is filling, the low pressure indicator (yellow LED) will illuminate. 4. When the selected pressure level is reached, the low pressure indicator will de-illuminate, and the normal pressure indicator (green LED) will automatically illuminate. 5. Replace all sheets, blankets, and pillows on the bed. Pressure adjustment Generally, a lighter patient will need a lower (softer) setting while a heavier patient will need a higher (firmer) setting, but pressure adjustment must ultimately be based on the patient's weight distribution. A short, stocky person may be the same weight as a tall, thin person, but their settings will not be the same. The mattress should be firm enough to support the patient, yet soft enough so the patient will sink down to eliminate pressure points.test the pressure by placing four fingers below the mattress under 750000-INS-LAB-RevA08 11

the patient's buttock area; your fingers should just fit. If the fiit is too loose, decrease the setting. If your fingers do not fit well, increase the setting two notches firmer. Wait ten minutes and then check again. You should only have to adjust this setting one time per patient. Please consult your physician for an appropriate setting. Quick disconnect Pull the power plug from the wall connector to disconnect the device quickly. Transport cap In case of power failure or transport: Disconnect the quick connector and cap the mattress with the transport cap to slow deflation. CPR function If CPR must be performed on the patient while the mattress is in use, to immediately deflate the mattress, either: 1. Pull the CPR valves from the mattress in the location of the arrow shown at right, or 2. Disconnect the quick connector from the pump in the location of the arrow shown at right. Maintenance quick connector transport cap Cleaning Perform the following cleaning procedures before use. Pump: Wipe the pump with a clean, damp cloth and mild detergent, and keep it away from dust. Air dry. 12 750000-INS-LAB-RevA08

s Caution: Do not use corrosive or powdered cleansers to clean the pump. s Caution: Do not immerse or soak the pump. s Caution: Do not clean with phenolic products. Mattress: Wipe the mattress unit with a clean, damp cloth and mild detergent. The mattress may also be cleaned using a 10% solution of sodium hypochlorite (bleach) diluted in water. Air dry all parts thoroughly before use. s Caution: Do not clean with phenolic products. s Caution: After cleaning, air dry the mattress without direct exposure to sunlight. GENERAL Maintenance 1. Check main power cord and plug for abrasion or excessive wear. 2. Check mattress cover for signs of wear or damage. Ensure mattress cover and tubes are connected correctly. 3. Check airflow from the air hose connector. The airflow should alternate between each connector every halfcycle time if it s in alternating mode. 4. Check the air hoses for any kink or break. For replacement, please contact your GF equipment provider. Fuse replacement tool needed: small screwdriver 1. If you suspect a blown fuse, disconnect the plug from the wall outlet immediately. 2. Use a small screwdriver to remove the cover of the fuse holder (fuse locations are shown in picture at right). 3. Insert a new fuse of the correct rating (T1A/250V, VDE approved). 4. Replace the fuse holder cover. 750000-INS-LAB-RevA08 13

Air filter replacement 1. Remove the air filter plate located at rear of pump (filter location shown in picture at right). 2. Remove the filter. The filter is reusable, if not torn, and can be washed gently with a mild detergent and water. Dry the filter before use. 3. Replace the air filter regularly if it is gray, torn, or the environment is dirty. Storage 1. Lay the mattress out flat and upside down. 2. Roll from the head end toward the foot end. 3. Stretch the foot-end strap around the rolled mattress to prevent unrolling. 4. Store mattress and pump in a dust-free environment with no exposure to direct sunlight. s Caution: Do not fold, crease or stack mattress. Specifications Pump Power Supply Fuse Rating Cycle time Air input Dimensions (L x W x H) Weight Specification 120V System: AC 120V 60 Hz Note: see rating label on product T1A, 250V 8 minutes >4 LPM 9.8" x 5.3" x 3.7" (25 x 13.5 x 9.5 cm) 4 lb (1.8 kg) 14 750000-INS-LAB-RevA08

Specifications continued Pump Specification Environment Temperature Operation: 50 F to 104 F (10 C to 40 C) Storage: 5 F to 122 F ( -15 C to 50 C) Shipping: 5 F to 158 F ( -15 C to 70 C) Humidity Operation: 10% to 90% non-condensing Storage: 10% to 90% non-condensing Shipping: 10% to 90% non-condensing Classification Class II, Type BF, IPX0 Applied Part: Air Mattress Not suitable for use in the presence of a flammable anesthetic mixture (No AP or APG protection) Mattress Model Dimensions (L x W x H) Weight Arrangement Specification 8" mattress 78.7" x 35.4" x 8" (200 x 90 x 20.3 cm) 22.1 lb (10.0 kg) 20 x 8" bladders Material Cover Nylon/PU Cells Base Nylon/PU PVC Tarpaulin Warranty GF Health Products, Inc. warrants the Lumex AltaDyne 750000 Alternating Pressure / Low Air Loss Mattress System for a period of twelve months for defects in workmanship and materials. During the warranty period, defective items will be repaired or replaced at GF Health Products' option at no charge. 750000-INS-LAB-RevA08 15

USA, Corporate Headquarters: Graham-Field Health Products 2935 Northeast Parkway Atlanta, Georgia 30360 telephone: 800-347-5678, 770-447-1609 fax: 800-726-0601, 678-291-3232 www.grahamfield.com 750000-INS-LAB-RevA08 2008 GF Health Products, Inc.