Competency Record for Sonic Cleaning

Similar documents
SUGGESTED POLICY FOR THE MONITORING OF THE CLEANING EFFICIENCY OF AN ULTRASONIC CLEANERS( SONIC & TOSI) WEEKLY

Hygea Ultrasonic Validation Kit

Shukla Medical. TITLE: Manual Surgical Orthopedic Instruments Recommendations for Care, Cleaning, Maintenance and Sterilization

Suction Regulators. Recommendations for Cleaning, Disinfection and Reprocessing: Boehringer 3800 &3900 Series Suction Regulators

Ultrasonic baths. XUB range of digital ultrasonic baths. XUBA range of analogue ultrasonic baths. Digitally controlled range with a choice of 5 baths

TJF-Q180V Cleaning and Disinfection Checklist

TJF-Q180V Cleaning and Disinfection Checklist

QUALITY ASSURANCE HATCHERY

AMBLER SURGICAL INSTRUCTIONS FOR USE - OPHTHALMIC DUET-STYLE SCISSORS

Urgent Field Safety Notice (Field Safety Corrective Action) for CORAIL AMT Neck Trials Surgical Instruments

INFORMATION NOTE No 353: MICROBIOLOGICAL TESTING

HEALTH TECHNICAL MEMORANDUM 2030 Washer disinfectors: Validation and verification STATUS IN WALES ARCHIVED. This document was superseded by

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

Exeter Clinical and Health Research

For Establishing and Maintaining Your Spa using PROtech Spa Chemicals

Prestige Endoscopic Graspers Instructions for Use

1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique.

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual

Title: INSTRUMENT AND EQUIPMENT DOCUMENTATION AND RECORDS

Maintenance handbook

An Important Change to Pinery s Water in March

Oxygen Dialflow Meter. Instructions for Use

POOL CARE GUIDE. The E-Z Clor System To Perfect Pool Water.

Infection is caused by breathing in small droplets of water contaminated by the bacteria. The disease cannot be passed from one person to another.

TS1. Ultrasonic Tank Sender. Installation and Operating Instructions. For TS1 Firmware v3.8. Page 1 INST-TS1-V13 18/11/10

Product SDS. Purpose of product: To provide a pass/fail detection of cavitation energy within an ultrasonic bath.

General Warnings! A warning statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately could lead

Gas Injection Systems

Hercules 3 Universal Stabilizer Arm. Catalog Number

TITLE: DNA minitube - Red

Maintenance and Calibration

SKAGIT COUNTY AUTOMATED EXTERNAL DEFIBRILLATOR [AED] POLICY

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

MONTGOMERY COUNTY PUBLIC SCHOOLS Office of the Chief Operating Officer Automated External Defibrillator (AED) Guidelines for High Schools

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

Other terms If any term in this contract cannot be enforced or agreed to, then the Daizy s Castles and Cakes Hire equipment will not be supplied.

ANNEX 4 ALTERNATIVE TEXT FOR OPERATIONAL GUIDELINES FOR VERIFICATION OF DAMAGE STABILITY REQUIREMENTS FOR TANKERS

UNIVERSITY OF ROCHESTER MEDICAL CENTER

Effect of noise in the performance of the transducers in an ultrasonic flow meter of natural gas

PRESSURE SYSTEMS POLICY

Siphon vessel / Priming aid

Eurosonic 4D - 3D - Micro ENGLISH

4. Instruction in telemetry requires IACUC review if live vertebrates are used in the training session. a. True b. False

Laboratory and Satellite Rodent Housing

Exeter Clinical and Health Research

Guidelines. Milan Special School District. Automated External Defibrillator (AED) Program

LEGIONELLA MANAGEMENT PROCEDURE

Frequently Asked Questions about Mechanical Qualification. Ken Boda Applications Engineer Agilent Technologies

Safe management of industrial steam and hot water boilers A guide for owners, managers and supervisors of boilers, boiler houses and boiler plant

BIOSAFETY SELF AUDIT PRINCIPAL INVESTIGATOR/PERMIT HOLDER: CONTAINMENT LEVEL: 1 2 (Shaded) RISK GROUP: 1 2 AUDIT TO BE COMPLETED BY (DATE):

INFUSION SYSTEM (110V) (220V) USER S MANUAL

Service and Repair Manual

RMM 1204 First Aid Program Page 2 June 2016

HAZARDOUS INCIDENT SITE SAFETY PLAN

Introduction. Legal Duties. Risk Assessment

CLUB RISK ASSESSMENT TEMPLATE

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

RESPIRATORY PROTECTION PROGRAM

Recommendations on Checking Anaesthesia Delivery Systems

3M Steri-Vac Sterilizer/Aerator GSX Series. For Life Science Applications. Product. Profile

Significant Change to Dairy Heat Treatment Equipment and Systems

Ritter AquaBottleSystem

Essential Chemical Safety & Hygiene Training 2015

Sooriyan Pool and Spa Ionizer. by Hilltop Solar

Recommendations - USA

Our technical information is based on Tip: calls attention to special information.

Parallel Session 1A Sterile Manufacture. Denise Coakley. GMP Conference 12 th November 2014

Hazard and risk analysis in pharmaceutical products 1

Qualification Specification. Pool Water Treatment Y T. Version 17.1

OPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP

HEAT EXCHANGE AND TRANSFER, INC. 500 Superior Street Carnegie, PA (412) Fax: (412)

UNIT TITLE: CLEAN PUBLIC AREAS, FACILITIES AND EQUIPMENT NOMINAL HOURS: 60 hours

Brian Ballew SA-ALCRIESM 450 Quentin Roosevelt Road Kelly AFB, TX

Reliability engineering is the study of the causes, distribution and prediction of failure.

Kintrol Instructions for Use Product Number: VS4

Best Practice in the Breeder House

QUICK START GUIDE TO DRINKING WATER MONITORING

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

The University of Georgia Automated External Defibrillator (AED) Program Guidelines

Anaesthetic infrastructure and supplies

User Guide. Pall Laboratory Manifold. For laboratory use. Not for use in a manner other than indicated. Introduction. Regulatory References

EVALUATING KEGGING OPERATIONS: SEEING INTO THE BLACK BOX. Eric Johann Samp, Ph.D. Chris Nimptsch

Pipeline Pigging and Hydrostatic Test

BODY OF KNOWLEDGE TANK ENTRY SUPERVISOR CERTIFICATION PROGRAM Revision 2, October 2013

Ultrasonic level transmitter and indicator

BODY OF KNOWLEDGE API TES TANK ENTRY SUPERVISOR

Hi-Temp Spa Care Guide

MSE CONTROLLED DOCUMENT. Radiological HS&E Precautions for NORM Contaminated Tanks, Vessels & Separators

Instruction Manual for Configura Cushionair Portable Pump

SF 6 Product Guide. Gas Analysis Instruments and Accessories

Najo Backboards. by Ferno. Users Manual. February 2006 GLO Pub. No Najo RediBoard. Najo Lite. Najo. Najo RediHold. RediWide.

Vertebrate Animal Biosafety Level 1 Criteria

1.0 PURPOSE 2.0 SCOPE 3.0 DEFINITIONS. ANSI: American National Standards Institute. CCC: Chemical Control Centre

Environmental Health & Safety Policy & Procedure #35

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

Assembling & Installing your Pool Cleaner

889P & 889P/B Backboard

The routine maintenance and inspection of local exhaust ventilation (LEV)

Tools to Enhance Laboratory Biosafety and Biosecurity Performing Risk Assessments

Transcription:

Competency Record for Sonic Cleaning Name: Competency Achieved: (Date) Evaluator: Learner: Comments: Background information on sonic testing ST 79; 2009A section 7.5.3.3 states this on weekly testing Mechanical cleaning equipment should be tested upon installation, weekly (preferably daily) during routine use, and after major repairs. A major repair is a repair that is outside the scope of routine preventive maintenance and that significantly affects the performance of the equipment. Examples include replacement of the water pump(s), detergent delivery system, heating system, water delivery system, water treatment system, or computer control or an upgrade to software Sonic cleaners are considered mechanical cleaning equipment by AAMI. "Cleaning, not sterilization (or disinfection) is the first and most important step in any instrument processing protocol. Without first subjecting the instrument to a thorough, validated and standardized (and ideally automated) cleaning process, the likelihood that any disinfection or sterilization process will be effective is significantly reduced". The ultrasonic cleaning process cleans dirty surgical instruments so they can be handled safely, repackaged, and sterilized for a future surgery. The danger of handling instruments contaminated with blood is obvious in this age of hepatitis, CJD and HIV. The procedures for sterilizing instruments are based on years of scientific testing of cleaning instruments. If surgical instruments are not clean, the procedures are ineffective. Dried blood on instruments is hazardous to the employees of the hospital and to the next surgical patient upon which the instruments are used. 1

An ultrasonic cleaner enables thorough cleaning of equipment by ultrasonic agitation that dislodges soil from instruments. Ultrasonic cleaners do not disinfect instruments. They are used to assist with the cleaning of instruments that cannot be adequately cleaned manually such as spiral wound instruments like biopsy forceps. Ultrasonic vibration at the frequency used for cleaning does not kill microorganisms and infective aerosols may be produced. It is for this reason that the lid of the tank must be tightly closed during operation. Ultrasonic cleaners work by subjecting instruments to high frequency, highenergy sound waves. This causes the soil to be dislodged from instruments and drop to the bottom of the tank, or be sufficiently loosened that it will be removed during the rinsing process. The detergent used in the ultrasonic tank must be carefully selected in accordance with advice from the tank's manufacturer. Optimally it will be a neutral, low-foaming product and enzymatic cleaners will have enhanced benefits in this process. Degassing of cleaning solutions is extremely important in achieving satisfactory cleaning results. Fresh solutions or solutions which have cooled must be degassed before proceeding with cleaning. Degassing is done after the chemical is added and is accomplished by operating the ultrasonic energy and raising the solution temperature. The time required for degassing varies considerably, based on tank capacity and solution temperature, and may range from several minutes for a small tank to an hour or more for a large tank. An unheated tank may require several hours to degas. Degassing is complete when small bubbles of gas cannot be seen rising to the surface of the liquid and a pattern of ripples can be seen. Routine cleaning Cleaning the ultrasonic cleaner and replacement of the cleaning solution is necessary at least daily or more frequently if solution soiled. Performance testing The efficacy of the ultrasonic cleaner should be tested at a minimum weekly and daily if possible. The results of the testing shall be documented as part of the proof of process. Sonic cleaners fail for many reasons. Tests should provide a means of monitoring the variables that influence the effectiveness of the ultrasonic cleaning process. Some of these variables are water, time, detergent, enzyme, 2

temperature, high ph, agitation, speed, initial heat, drying, obstructions, and insufficient amount of chemicals and equipment faliure. Proper cleaning is critical. The SonoCheck test kit provides an independent objective test of clean and allows the Sterile Processing professional to monitor and ensure proper cleaning in the sonic process AAMI does list a sonic cleaner as a piece of medical equipment in section 7.5.3.3; there are other references that support the minimum of weekly testing of all type of sonic equipment with or without retro flow pulse adapter. Section 10.2 and ANNEX D states Health care personnel may perform verification tests as part of the overall quality assurance program. This verification may include the use of test devices that monitor the functionality of the cleaning equipment in cleaning surfaces and that ensure adequate fluid flow in equipment that has adaptors for lumened devices Section 7.5.3.3 states Mechanical cleaning equipment should be tested upon installation, weekly (preferably daily) during routine use, and after major repairs. A major repair is a repair that is outside the scope of routine preventive maintenance and that significantly affects the performance of the equipment. Examples include replacement of the water pump(s), detergent delivery system, heating system, water delivery system, water treatment system, or computer control or an upgrade to software.. The 2011 AORN RP for Cleaning and Care of Surgical Instruments and Powered Equipment Recommendation XXII Quality section is now supporting the testing of mechanical instrument washers before initial use, weekly during service, and after major maintenance. JCAHO in standard E.C.6.20 it states that medical equipment is maintained, tested and inspected So in accordance with ANSI/AAMI ST79-2009; ANSI/AAMI TIR12-2005; JCAHO EC6.20-2005, and AORN, Healthmark recommends at least weekly monitoring of the cleaning efficacy of any sonic cleaning unit with the following verification tests: Verification of sufficient cavitation energy with the Sonocheck monitoring vials*. Designed to change color (blue to yellow) when the ultrasonic cleaner is supplying sufficient energy and conditions are correct (degassed water, temperature, etc.) the Sonocheck is an easy to use and interpret method for monitoring cavitation energy. Failure to change color indicates that either the sonic bath conditions were not correct, or a failure of one or more of the ultrasonic transducers. 3

JCAHO and AAMI both recommend that Sterile Processing department have process performance in place. Using the sonic check according to the manufacture's guidelines helps ensure adherence to both JCAHO and AAMI standards and having a properly functioning cleaning process. 4

Competency Statement: Complies with policy and procedure for testing the Key 1 = Performs independently and consistently. Ask for assistance in new situations. 2 = Performs with minimal guidance and direction. Asks for assistance when necessary. 3 = Performs with maximal guidance and direction. Preceptor dependent. Consistently needs assistance. Critical Behaviors 1 2 3 Read manufactures manual for your specific sonic cleaner Completed / Attended the Self Study program supplied by the Vendor Describes the purpose of sonic cleaning at your facility Selects and wears the appropriate personal protective equipment Gather appropriate supplies to perform test on sonic cleaner Inspect sonic cleaner according to manufacture recommendations & hospital policy (document results) 1. Prepare a bath of cleaning solution (water and detergent) in compliance with instructions for use by the sonic manufacturer and the detergent manufacturer. 2. De-gas the bath in accordance with ultrasonic manufacturer s instructions. Insure that the bath is within the proper temperature range as provided by the detergent manufacturer Make sure that the ultrasonic cleaner has been degassed prior to running the test. 1. Select the appropriate number of SonoCheck TM vials and as shown below based on tank volume and choose the layout that matches the size of the equipment to be tested. 2. Place the SonoChecks as indicated below in an empty ultrasonic basket and place the basket in the ultrasonic cleaner that has been de-gassed. 3. Run the equipment as directed by the ultrasonic manufacturer and record the test results on the Log Sheet located on hmark.com. 4. All SonoChecks should change from blue/green to 5

yellow within specified time. The time needed for the color change will indicate the level of energy and degree of cavitation provided by the ultrasonic cleaner. A negative result will indicate a blind spot of ultrasonic energy. In case of an unsatisfactory result, refer to the SonoCheck TM troubleshooting guide. Small tank= Up to 5 ltrs Medium tank=5 to 20 ltrs Large=Above 20 ltrs Small Tank Medium Tank Large Tank x x x x x x Understand that a color change from blue/green to yellow indicates presence of cavitation energy. Failure for color change to yellow indicates a failure to achieve sufficient cavitation energy to clean. Ultrasonic energy is localized and failure to achieve color change may indicate one or more sonic transducers are failing. Record all results in a log Report any unsatisfactory results to the proper management for corrective action according to the policy of your facility. 6