How to evaluate pressure measurement specifications

Similar documents
PPC4E Pressure Controller/Calibrator Superior rangeability and reliability

PG7000 high line differential mode for defining differential pressure at line pressures above atmospheric pressure (PG7000 V2.

P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features

P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features

How to calibrate reference pressure transducers with COMPASS for Pressure Software Application Note

E-DWT-H Electronic Deadweight Tester

molbox1+ Flow Terminal ± % of reading lowest uncertainty for gas flow calibration

CALIBRATION SYSTEM REQUIREMENTS. ESCC Basic Specification No

Gerald D. Anderson. Education Technical Specialist

Pneumatic high-pressure controller Model CPC7000

P3000 Deadweight Tester Setup, Part 1: Limited Partial Correction Method

THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST

Four common PRESSURE CALIBRATION PAIN POINTS

Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features

Pressure Calibration. Technical Data. Pressure Calibration with Fluke Calibrators

P5523. Users Manual. Liquid to Gas Separator. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

I N N O V A T I O N C O N T I N U E S... International Patent Filed in 45 Leading Countries

Fluke 700 Series pressure modules

FPG8601 Force Balanced Piston Gauge

Calibrating pressure transmitters and switches in potentially explosive atmospheres

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

PG9607 Piston Gauge. Automated primary pressure reference for absolute and gauge pressures to 500 kpa

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

VT650. Gas Flow Analyzer. Technical data. Key benefits and features:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

BAPI Pressure Line of Products - FAQs

VT900. Gas Flow Analyzer. Technical data. Key benefits and features:

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Interlaboratory comparison in the pressure range from 0 to 2 MPa for accredited calibration laboratories

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Pressure calibration. 20

Pressure Calibrators. Series PC..., PM...

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

If a fair coin is tossed 10 times, what will we see? 24.61% 20.51% 20.51% 11.72% 11.72% 4.39% 4.39% 0.98% 0.98% 0.098% 0.098%

Guide for the uncertainty analysis in pressure when using P3000 Series Deadweight Testers

Improve Process Reliability

Precision pressure sensor Basic version Model CPT6020

PM105 PRESSURE METER USER S MANUAL

CALIBRATION AND TRACEABILITY: ITS IMPORTANCE IN THE GAS INDUSTRY

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. Type: ETd 04. class 0.05

CALIBRATION CERTIFICATE # 500

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z CALIBRATION

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications

Series Ruska Digital Pressure Indicator. GE Sensing & Inspection Technologies. Features

Pneumatic high-pressure controller Model CPC7000

Certified Accuracy. Ring Force Gauge. Models and Capacities Available. Design and Principle of Operation

Monitoring transformers with infrared cameras

PRESSURE TRANSDUCERS WITH UNIFIED (4-20) ma OR (10-50) ma OUTPUT SIGNAL

Ron Gibson, Senior Engineer Gary McCargar, Senior Engineer ONEOK Partners

Advanced Test Equipment Rentals ATEC (2832)

Next Generation Quartz Pressure Gauges

Liquid level measurement using hydrostatic pressure and buoyancy

ONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction

BaseCal 3, calculating the performance of Diaphragm Seals

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

Datasheet May, 2017 HPSD 4000 Pressure Transducer 1/5. Applications

PTG100 Precision Test Gauge

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

The Gold Standard in Calibration Mixtures

How to specify a product. Process Sensors and Mechanical Instruments

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

2014 Programme of Training Courses in Measurement & Calibration

INTRODUCTION TO GAS METERING

Selecting the right pressure sensor for your application

Sales Guide. 700HPPK Pneumatic Test Pump Kit

Hand-held pressure calibrator with integrated pump Model CPH6600

PMI ADVANCED AUTOMATED HYDRAULIC PRESSURE LEAK TESTER APCS-H-60K. Not just Products... Solutions!

Royal Process Gauges, ±0.5% Accuracy 316 SS Tube and Socket Dial Size

ITTC Recommended Procedures and Guidelines

Hydronic Systems Balance

ISO INTERNATIONAL STANDARD. Natural gas Measurement of properties Volumetric properties: density, pressure, temperature and compression factor

Calibration Procedure. PosiTest DFT Combo

Tools and techniques for pressure calibration

Design Features. General Description

Tutorial for the. Total Vertical Uncertainty Analysis Tool in NaviModel3

Engineering: Measurement Technology Pressure/Level (SCQF level 6)

Low Pressure AutoChoke Console Precise wellbore pressure control for your underbalanced and managed-pressure drilling operations

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

How we calibrate the Wave Height and Period Measurements from the Gravitational Acceleration Wave Buoys in RMIC/AP

Exercise 8. Closed-Loop Pressure Control, Proportional-Plus-Integral Mode EXERCISE OBJECTIVE

159 mm 154 mm. Subject to changes 10/2012

The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions

Hydraulic hand test pump Models CPP700-H and CPP1000-H

WaterSense Specification for Spray Sprinkler Bodies. Version 1.0

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Pneumatic service kit Model CPG-KITP

Ultima. X Series Gas Monitor

Traceable calibration of automatic weighing instruments in dynamic operation

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

Electronic Volume Correctors

Pressure is our Business

Chapter 7. SCIIB Pressure Sensor Performance Evaluations: Experiments, Results and Discussions

High-speed pressure sensor Model CPT6140

Green 0-3% Acceptable Orange 3.1-5% Alert Red 5.1 and up Fail. Calibrated By:

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

INTERNATIONAL HYDROGRAPHIC SURVEY STANDARDS

Transcription:

How to evaluate pressure measurement specifications Application Note There are thousands of different pressure measuring devices out there. How do you decide which instruments to invest in to calibrate them all? The secret to calibrating the most workload with the least investment lies in understanding the term rangeability. To do that, you ll need to know how to evaluate instrument specifications. This application note covers: Types and examples of percentage specifications Comparing specifications Test accuracy ratios (TARs) Determining rangeability Types of percentage specifications Pressure measurement devices, like gauges, transmitters, and transducers, have specifications that describe how well they measure pressure. When you calibrate or validate one of these devices, you want to confirm that it is performing within its specifications. During a calibration, you typically compare your device s measurements to the measurements of a reference standard that has the same types of functions but better accuracy. To ensure that your reference standard is accurate enough to do the calibration, it s common to compare the specifications of the device under test (DUT) to the reference standard s specifications using a test accuracy ratio. From the Fluke Calibration Digital Library @ www.flukecal.com/library

A pressure device s measurement specification is often expressed as a percentage. But there are different ways of expressing a percentage, and each one means something different: Percent of full scale This type of specification is a constant value throughout the range of the device. Accuracy (maximum allowable error) Pressure Reading (RDG) FS ± 0.25 % FS Upper range limit = 1,000 psi Accuracy specification= ± 2.5 % FS FS= Full scale = upper range limit = 1,000 psi ± 0.25 % of 1,000 psi = ± 0.25 psi The accuracy of the gauge is ± 2.5 psi at all pressures throughout the range. Percent of full scale specification. Percent of reading This type of specification is a function of the indicated pressure, so it changes as the pressure reading changes. Accuracy (maximum allowable error) Pressure Reading (RDG) FS ± 0.25 % RDG Upper range limit = 1,000 psi Accuracy specification= ± 0.25 % reading Reading is the indicated pressure: 0 psi, 500 psi, and 1,000 psi Find the maximum allowable error at the indicated pressures. Indicated pressure Accuracy specification 0 psi ± 0.25 % of 0 psi = ± 0 psi 500 psi ± 0.25 % of 500 psi = ± 1.25 psi 1,000 psi ± 0.25 % of 1,000 psi = ± 2.5 psi Note: In this case, the upper range limit does not affect the accuracy specification. Percent of reading specification. 2 Fluke Calibration How to evaluate pressure measurement specifications

Combined percent of reading and percent of full scale A maximum allowable error of ± 0 units as in the example above is probably unrealistic for any instrument. To avoid this problem, the manufacturer either limits the instrument s range or includes an error floor with the percent of reading specification. The error floor may be given as a constant, a percentage of full scale, or sometimes with a reference to the instrument s resolution. Accuracy (maximum allowable error) ± X % RDG + Error Floor X % FS FS Pressure Reading (RDG) Full scale = 1,000 psi Accuracy specification= ± 0.25 % reading plus 0.125 % full scale Indicated pressure Accuracy specification 0 psi ± ( 0.25 % of 0 psi + 0.125 % of 1,000 psi) = ± 1.25 psi 500 psi ± (0.25 % of 500 psi + 0.125 % of 1,000 psi) = ± 2.50 psi 1,000 psi ± (0.25 % of 1,000 psi + 0.125 % of 1,000 psi) = ± 3.75 psi Combined percent of reading and percent of full scale specification. Percent of span The specification is a constant value throughout the range of the device, but is based on the difference between the upper range limit and the lower range limit. Accuracy (maximum allowable error) ± X % Span Lower range limit Span Upper range limit Accuracy = ± 0.25 % of span Upper range limit = 100 psi Lower range limit = -10 psi Span = Upper range limit lower range limit = 110 psi ± 0.25 % of span = ± 0.25 % of 110 psi = ± 0.30 psi The accuracy is ± 0.30 psi at all pressures throughout the range. Percent of span specification. 3 Fluke Calibration How to evaluate pressure measurement specifications

Comparing specifications for three reference pressure gauges To ensure that you are properly comparing the DUT and the standard (master gauge), you need to convert their specifications into like quantities. Ideally, this means you ll convert all specifications into actual pressure values at each test pressure. Table 1 shows specifications for three digital pressure gauges, which can be used as reference standards in a calibration. Two of the specifications are expressed as percent of full scale, and the third is expressed as percent of full scale plus percent of reading. In this comparison, each gauge is the same full scale (1,000 psi) and we are showing the specification in 10 % increments. Test accuracy ratios A test accuracy ratio (TAR) is the ratio between the stated accuracy (calibration tolerance) of the DUT and the accuracy of the calibrator. For example, if the stated TAR is 4:1 (read: four to one) that means that the master gauge or calibrator is four times more accurate than the device under test. We use test accuracy ratios because calibration quality needs to be ensured, but conventional uncertainty analysis is complex, time consuming and may not be warranted for instrument performance verification. Over time, different industries have developed different standards for what a good TAR looks like. The most common is 4:1. By keeping the TAR to a ratio such 4:1, you are less likely to falsely accept or reject a pressure gauge based on the outcome of your calibration, especially when the error is near the established tolerance. Test accuracy ratios may depend on: DUT accuracy specification DUT full scale Master gauge accuracy specification Master gauge full scale Pressure gauge #1 Pressure gauge #2 Pressure gauge #3 0.04 % reading + 0.01 % FS Pressure 0.05 % FS 0.02 % FS 0 0.5 0.2 0.1 100 0.5 0.2 0.14 200 0.5 0.2 0.18 300 0.5 0.2 0.22 400 0.5 0.2 0.26 500 0.5 0.2 0.3 600 0.5 0.2 0.34 700 0.5 0.2 0.38 800 0.5 0.2 0.42 900 0.5 0.2 0.46 1000 0.5 0.2 0.5 Table 1. Comparing specifications for three digital pressure gauges. Maximum allowable error (psi) 0.6 0.5 0.4 0.3 0.2 0.1 0 Comparing three reference gauges with same full scale (1,000 psi) 0 200 400 600 800 1000 1200 Pressure reading (psi) 0.005 % FS 0.02 % FS 0.04 % reading + 0.01 % FS Figure 1. Comparing specifications for three Fluke Calibration digital pressure gauges. 4 Fluke Calibration How to evaluate pressure measurement specifications

Comparing reference pressure gauges and a DUT, where the DUT has specification of 0.1 % full scale Understanding specifications help you compare the capabilities of pressure standards. Consider the scenario shown in Figure 2 where the same DUT has a full scale of 1,000 psi, and an accuracy specification of ± 0.1 % full scale. In this scenario, the pressure standard with a percent of reading specification provides a sufficient test accuracy ratio at lower pressures, but at full scale it is only 2:1. That is problematic because, generally, the upper 80 % of the reference gauge s scale is the most used and the most important. Determining rangeability An instrument s rangeability is the ratio of the maximum to the minimum specified measured value at which the instrument has an acceptable performance. As we have seen, acceptable performance depends on a test accuracy ratio with the DUT, and the accuracy of the DUT usually depends on its own pressure range. Rangeability depends on the DUT specification, the specification of the standard, and full scale of both devices. On the other hand, a ± 0.02 % FS master gauge can meet a 4:1 ratio when calibrating ± 0.1 % FS devices with full scales from 800 psi to 1,000 psi, ± 0.25 % FS devices with full scales between 320 and 1,000 psi, and ± 0.5 % FS devices with full scales between 160 psi and 1,000 psi. The bottom line is that you can calibrate twice as many DUTs with one master gauge if it has twice the rangeability. Test Accuracy Ratio 12 : 1 10 : 1 8 : 1 6 : 1 4 : 1 2 : 1 Comparing three pressure gauges and a 0.1 % device under test with 1,000 psi FS 0 : 1 0 200 400 600 800 1000 1200 0.05 % FS 0.02 % FS 0.04 % reading + 0.01 % FS Figure 2. Comparing three pressure gauges and a device under test with 1,000 psi but a specification of 0.1 % full scale. 1,000 psi master guage accuracy 0.05 % FS 0.02 % FS Master gauge rangeability 1,000 psi 800 psi 600 psi 400 psi DUT Full Scale When a 1,000 psi master gauge is required to be four times more accurate than the device under test, a ± 0.05 % FS master gauge cannot calibrate a ± 0.1 % FS device. However, it can calibrate ± 0.25 % FS devices with full scales between 800 psi and 1,000 psi and ± 0.5 % FS devices with full scales between 400 and 1,000 psi. Figure 3. Rangeability of two 1,000 psi master gauges. Rangeability The ratio of the maximum to the minimum specified measured value at which the instrument has an acceptable performance. Rangeability depends on the DUT specification, the specification of the standard, and full scale of both devices. 5 Fluke Calibration How to evaluate pressure measurement specifications

Comparing reference pressure gauges and a DUT, where the standards and DUT have the same full scale Figure 4 uses the same three gauges as Figure 1 and shows the test accuracy ratio (TAR) with an example device under test (DUT). Our DUT has a specification of 0.25 % full scale and a full scale of 1,000 psi. All gauges in this scenario meet a 4:1 TAR requirement. Comparing reference pressure gauges and a DUT, where the standards and DUT have different full scale In the previous example, the standards and the DUT had the same full scale (1,000 psi). Now let s consider a scenario where the DUT has a full scale that is only 500 psi, half that of our standards. See Figure 5. Although the 500 psi gauge had the same percentage accuracy specification as the 1,000 psi gauge, the error limits were much tighter for the 500 psi gauge, because they were calculated as ± 0.25 % of 500 psi rather than ± 0.25 % of 1,000 psi. This caused the test accuracy ratios with the 500 psi pressure gauge to be smaller. In fact, they were so small that the test accuracy ratio for the ± 0.05 % FS gauge dropped below 4:1. This means the ± 0.05 % FS 1,000 psi pressure gauge probably would not be adequate to calibrate a ± 0.025 % FS device with a 500 psi upper range limit. This example further demonstrates the concept of rangeability, because a more accurate test standard can calibrate a wider range of pressure gauges. All things being equal, a master gauge with twice the accuracy as another will be able to calibrate twice the range of instruments. Conclusion When you are choosing a pressure standard to calibrate a device under test, it s not sufficient to look just at the standard s percent specification. You also need to account for the different types of percent specifications (full scale, reading, span and combined), and also any difference in full scales of the DUT and the pressure standard. Test Accuracy Ratio 30 : 1 25 : 1 20 : 1 15 : 1 10 : 1 5 : 1 Comparing three pressure gauges and a 0.25 % device under test with 1,000 psi FS 0 : 1 0 200 400 600 800 1000 1200 0.05 % FS 0.02 % FS 0.04 % reading + 0.01 % FS Figure 4. Comparing three pressure gauges and a device under test. Test Accuracy Ratio 14 : 1 12 : 1 10 : 1 8 : 1 6 : 1 4 : 1 2 : 1 Comparing three pressure gauges and a 0.25 % device under test with 500 psi FS 0 : 1 0 100 200 300 400 500 600 Figure 5. Comparing three pressure gauges and a device under test with 500 psi full scale. Fluke Calibration. Precision, performance, confidence. Fluke Calibration PO Box 9090, Everett, WA 98206 U.S.A. Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands Web access: http://www.flukecal.eu For more information call: In the U.S.A. (877) 355-3225 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.flukecal.com 2014 Fluke Calibration. Specifications subject to change without notice. Printed in U.S.A. 7/2014 6002525b-en Pub-ID 13157-eng Modification of this document is not permitted without written permission from Fluke Calibration. 6 Fluke Calibration How to evaluate pressure measurement specifications