GA24 Technical Datasheet

Similar documents
VA40 - VA45 Technical Datasheet

GA24. GA24 Handbook. Variable-area flowmeter. KROHNE 02/ MA GA24 R01 en

Liquid level instruments BM 24 /BM 51

KDV HIGH ACCURACY GLASS TUBE ROTAMETER

Liquid level switch BM 34

KDV HIGH ACCURACY GLASS TUBE ROTAMETER

Miniature flowmeters DK 46, DKR 46 DK 47, DK 48, DK 800

VersaFlow Vortex 100 Vortex Flow Meter Technical Datasheet. Specification. The all-in-one solution. Highlights. Industries.

Level transmitter LT 100

Glass tube flowmeters Series 2000

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure

DK48 - DK800. Quick Start. Variable area flowmeter

SITRANS P measuring instruments for pressure

DK46 - DK800 Supplementary instructions

7000 Series OPTIMASS Mass Flowmeters. Titanium Hastelloy Stainless Steel

Glass tube flowmeters Series 2000 Variable area flowmeter for low flows of liquids and gases

Technical Data Sheet. OPTIMASS 7000 Custody Transfer Mass Flowmeter

Monitoring device NTS 30

Electropneumatic Positioner and Pneumatic Positioner Type 3760 JIS

OPTIBAR P 1010 C Technical Datasheet

Level switches Series LD Vibrating fork level switch for liquids and solids

Pressure Measurement. Transmitters for basic requirements SITRANS P Z for gauge pressure. 2/4 Siemens FI Overview

Level switches Series LD Vibrating fork level switch for liquids and solids

1 Overview. Pressure Measurement Transmitters for basic requirements. 1/16 Siemens FI SITRANS P220 for gauge pressure

Electropneumatic Positioner Type 4763 Pneumatic Positioner Type 4765

DK32 - DK34 - DK37 Technical Datasheet

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/16 Siemens FI US Edition

Pressure Measurement Single-range transmitters for general applications

Measurement of Low GAS & Liquid Flows

3000 Series OPTIMASS Mass Flowmeter. Stainless Steel Hastelloy

Variable area flowmeter F VA 250

Type N-10, N-11 Hazardous Area Non-incendive Transmitters

Pressure Measurement Single-range transmitters for general applications

SITRANS. Pressure transmitter SITRANS P, Z series for gauge and absolute pressure. Introduction. Safety instructions 2.

L21S/SR/ST Glass Level Gauges

Electropneumatic Positioner and Pneumatic Positioner Type 3760 JIS

Pressure Switches Application Guideline

Pressure Reducing Valve DMV 755

High-performance submersible pressure transmitter For level measurement Model LH-20

SITRANS F I Gardex. Flowmeter. SITRANS F I Intra/Prima. Flap Flowmeter. Contents. English. General 1. SITRANS F I Gardex 2. SITRANS F I Intra/Prima 3

C-Flow Coriolis Mass Flow Meters

DK 37 M8. All-metal miniature flowmeters. For hazardous-duty devices, please refer to Supplementary Installation and Operating Instructions:

High-performance submersible pressure transmitter For level measurement Model LH-20

High-performance submersible pressure transmitter For level measurement in hazardous areas Model LH-20

High-performance submersible pressure transmitter For level measurement Model LH-20

Differential pressure

Technical Information. Overview Rotameter. TI 1R1A0-E-H 4th edition

TRZ 03-K Volumeter. Proven Technology. Superior Performance. TRZ 03-K Volumeter is a Turbine Gas Flowmeter for Secondary Gas Metering.

Installation and Operating Instructions Bypass Flow Meter DST

Hydrostatic. Continuous level measurement. Overview Page 188 VEGAWELL 52 Page 190 VEGABAR 86, 87 Page 192 Accessory Page 196 AL-EN

DMK 458. Pressure Transmitter for Marine and Offshore. Ceramic Sensor. accuracy according to IEC 60770: standard: 0.25 % FSO option: 0.

Technical Data. General specifications Switching element function Rated operating distance s n 5 mm

Pressure Reducing Valve DMV 755 set range: 1,0-9,0 bar

FVA Tubux M30 Variable Area Flowmeter. Operating Instructions. MECON GmbH 07/2013 OI_FVA_Tubux_M30_EN

Serie 06-M6. Swing wafer check valve. made in. Application fields. Check valves E U R O P E WATER CONDITIONING INDUSTRY

FVA Trogflux variable area flowmeter. Operating Instructions. MECON GmbH 05/2013 OI_FVA_Trogflux_EN

x act i Precision Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor accuracy according to IEC 60770: 0.

x act i Precision- Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor

Target disk flowmeter Series DP. Instructions manual DP65 DP500. R-MI-DP Rev.: 4 English version

High-performance submersible pressure transmitter For level measurement Model LH-10

Technical Data. General specifications. Rated operating distance s n 2 mm

Features. Up to 6 switch points Minimum degree of protection IP65 Optional temperature sensor PT1000 or thermostat Potted electrical contacts

Mechanical Pressure Measurement

Flow Measurement Equipment Armored Purge Meters

High-performance submersible pressure transmitter For level measurement Model LH-10

9700 Transmitter Submersible hydrostatic level transmitter

Colima TOR Magnetic Level Switches

Technical Data. Dimensions

RMG Meßtechnik GmbH. Publication No E

Mass Flow Meter (MFM) for Gases

Ball Float Steam Trap UNA 43 PN 16/CL 125/JIS 10K UNA 46 PN 40/CL 150/CL 300/JIS 10K/JIS 20K DN 80, 100, 150, 3", 4", 6"

JUMO dtrans p30 Pressure Transmitter

Gas density monitor With integrated transmitter Model GDM-100-TI

Operating instructions for variable area flow meter with needle valve. Model: UVR/UTR

Norval Pressure regulator

Flow controller Type ED/CCB311 Instruction manual

ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE

LMP 307. Stainless Steel Probe. Stainless Steel Sensor. accuracy according to IEC 60770: standard: 0.35 % FSO option: 0.25 % / 0.

Differential Pressure Transmiter

Technical Data. General specifications. Rated operating distance s n 5 mm

Technical Data. Dimensions

Safety shut-off valve S 50. Product information

Pressure Measurement Single-range transmitters for general applications

Variable Area Flow Meters. VS Series

Purge Device KPTD/KPTK/KPTH

Safety shut-off valve S 100. Product information

ATM/GR. Electronic pressure switches CUSTOMER BENEFITS

Expert Hydrostatic Level Transmitters

Electronic Turbine Meter TERZ 94

Vortex flowmeters. Product family introduction Principle of operation Product review Applications Key product features

Pressure Reducing Valve DMV 750 set range: bar

Compact differential pressure switch Flameproof enclosure Ex d Models DE, DEC

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

2/2- and 3/2-way Rocker Solenoid Valve - bistable (impulse) version

H250 M9 Supplementary instructions

Bourdon tube pressure gauge for diaphragm seals

Magnetic level switch type MR783 Instruction Manual

Technical Manual. Liquid Level Transmitter CT801-LB/S

Instruction Manual Differential Pressure Transmitter

Pressure switch with bourdon tube

Transcription:

GA24 Technical Datasheet Variable-area flowmeter Sturdy construction for several applications Local indication without auxiliary power Replaceable mounting parts KROHNE

CONTENTS GA24 1 Product features 3 1.1 GA24 Variable area flowmeter... 3 1.2 Operating principle... 4 2 Technical data 5 2.1 Technical data... 5 2.2 Dimensions and weights... 8 2.3 Measuring ranges... 9 3 Installation 11 3.1 Intended use... 11 3.2 Installation condtitions...11 4 Electrical connections 12 4.1 Limit switches GA24... 12 4.1.1 MS14 limit switch connection... 13 4.1.2 TG21 limit switch connection... 13 5 Order form 14 6 Notes 15 2

GA24 PRODUCT FEATURES 1 1.1 GA24 Variable area flowmeter The GA24 flowmeter is designed for measuring the volume or mass flow rate of liquids and gases and for measuring gases at low operating pressures. Highlights Universal installation length concept Low pressure loss in gas applications Low maintenance Optional limit switch Quick and easy to install and operate Easy feeding of suspended solid particle (can also be installed at a later date) Fragment protection Industries Can be used in all industrial sectors, for example Chemicals Heating, Ventilation & Air Conditioning (HVAC) Iron, Steel & Metals Oil and Gas Power plants Machinery Paper and Pulp Water Furnace construction Applications Gas measurements Industrial burner controlling Compressor monitoring Water circuits 3

1 PRODUCT FEATURES GA24 1.2 Operating principle The flowmeter operates on the float measuring principle The measuring unit consists of a glass cone in which a float can move freely up and down. The flow goes from bottom to top. The float changes position so that the lifting force acting on it F1 is in equilibrium with the form drag F2 and its weight F3: F3 = F1 + F2 The height of the float is read on the scale of the measuring glass and indicates the flow rate. The top edge of the float marks the reading line for flow values. 4

GA24 TECHNICAL DATA 2 2.1 Technical data The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local representative. Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website (Download Center). Measuring system Application range Flow measurement of liquids, gases and vapors Operating method / measuring principle Variable area measuring principle Measured value Primary measured value Float position Secondary measured value Operating and standard volumetric flow Measuring accuracy Directive VDI / VDE 3513, sheet 2 (q G =50%) Measuring accuracy 1.o% Operating conditions Temperature Max. operating temperature TS Pressure Max. permitted operating gauge pressure PS at TS = 100 C Test pressure PT DN15, DN25 10 bar 1 DN40 9 bar 1 DN50 7 bar 1-40...+120 C / -40...+248 F Pressure equipment directive 97/23/EC Pressure equipment directive 97/23/EC or AD 2000-HP30 Installation conditions Inlet condition, run Outlet condition, run 1 other pressures upon request 5 x DN 3 x DN 5

2 TECHNICAL DATA GA24 Materials Flange connection GA24/R Stainless steel 1.4404 (316 L) Flange connection GA24/PTFE Housing Measuring cone Suspended solid particle Suspended solid particle and insert Seals Stainless steel 1.14404 (316 L) with a liner made of PTFE Steel plate housing (zinc-plated with epoxy/polyester coating) Borosilicate glass Stainless steel 1.4571 (316 Ti) or Hastelloy C2000 PTFE/insert or TFM (PTFE) Aluminum Polypropylen (PP) PVDF (conforms to FDA) Neoprene PTFE Scherring PTFE sleeve Temperatures For devices to be used in hazardous areas, special temperature ranges apply. These can be found in the separate instructions. Max. measuring temperature T m -40...+120 C 1-40...+248 F Max. ambient temperature T amb. -20...+100 C -4...+212 F 1 higher temperatures on request Types of limit switches Type Switching function Connection Shape Note MS 14/A bistable 2-wire, floating Reed contact Float with magnet required TG 21 bistable 2-wire NAMUR Slotted proximity switch Float with magnet required 6

GA24 TECHNICAL DATA 2 Limit switch use Size Cone no. Limit switch Size Cone no. Limit switch DN15 N 18.07 MS14/A DN25 N 21.09 MS14/A TG21 N 18.09 MS14/A N 21.13 MS14/A TG21 N 18.13 MS14/A N 21.18 MS14/A TG21 N 19.09 MS14/A N 21.25 MS14/A TG21 N 19.13 MS14/A DN40 N 41.09 MS14/A TG21 N 19.19 MS14/A N 41.13 MS14/A TG21 N 19.26 MS14/A N 41.19 MS14/A TG21 DN50 N 51.10 MS14/A TG21 N 51.15 MS14/A TG21 N 51.21 MS14/A TG21 Technical Data Limit Switches Technical data MS14 Contact type Switching reproducibility Contact rating max. turn-on voltage Normally open or normally closed, can be reconnected < 2% of full-scale range 12VA 30VDC Max. switching current 0,5A Ambient temperature -20...+85 C / -4...+185 F Protection category acc. to IEC 60529 / EN IP44 529 Technical data TG21 Rated voltage 8VDC Current consumption, active surface open 3mA Current consumption, active surface 1mA covered Ambient temperature -25...+100 C / -13...+212 F Protection category acc. to IEC 60529 / EN IP 67 (NEMA 6) 529 7

2 TECHNICAL DATA GA24 2.2 Dimensions and weights Dimensions Nominal size a b c d DN Inch mm Inch mm Inch mm Inch mm Inch 15 1/2" 500 19.7 300 11.8 84 3.31 82 3.23 25 1" 500 19.7 300 11.8 105 4.13 102 4.02 40 1 1/2" 500 19.7 300 11.8 125 4.92 122 4.80 50 2" 500 19.7 300 11.8 165 6.50 147 5.74 Weights approx. kg approx. Ib DN15 6 13 DN25 10 22 DN40 13 29 DN50 18 40 Process connection Connection dimensions according to Connection Pressure class Flanged connections EN 1092 DN15, DN25 PN25 EN 1092 DN40 DN50 PN10 ASME B16.5 1/2"...2" Class 150 lb / RF ASME B16.5 1/2"...2" Class 300 lb / RF 8

GA24 TECHNICAL DATA 2 2.3 Measuring ranges Measuring span 10 : 1 Flow values 100% Reference condition: Water 20 C / Air 20 C - 1.013 bar abs. Float materials: 1 Stainless Steel or Hastelloy - 2 PTFE - 3 TFM - 4 Aluminium - 5 Polypropylen (PP) Materials 1 2 3 1 3 4 5 1 2 3 4 5 Cone Flow water [l/h] Flow air [m 3 /h] Max. pressure loss [mbar] N 18.07 DN15 40 25 13 1.5 0.6 0.8 0.5 9 6 2 3 1 N 18.09 63 40 22 2.2 0.95 1.2 0.7 9 7 3 3 2 N 18.13 100 63 35 3 1.5 1.8 1.2 9 8 3 4 2 N 19.09 160 100 55 5 2.2 2.8 1.8 13 9 4 5 2 N 19.13 250 160 85 8 3.3 4.5 2.8 16 11 4 5 2 N 19.19 400 250 140 - - - - 21 14 5 - - N 19.26 630 400 230 - - - - 27 17 6 - - N 21.09 DN25 630 400 230 18 1 9 11 7 22 14 6 8 3 N 21.13 1000 630 350 28 1 14 18 12 23 17 6 8 4 N 21.18 1600 1000 600 49 1-28 1 17 1 26 25 7 10 6 N 21.25 2500 1600 950 70 1-42 1 26 1 33 40 8 12 9 N 41.09 DN40 1600 1000 600 45 1 22 28 18 32 18 9 11 5 N 41.13 2500 1600 900 70 1 36 45 1 28 1 34 20 10 12 5 N 41.19 4000 2500 1500 128 1-76 1 46 1 38 24 11 15 8 N 51.10 DN50 4000 2500 1500 120 1 56 70 45 43 25 12 15 7 N 51.15 6300 4000 2400 190 1 90 110 1 70 1 47 30 13 16 7 N 51.21 10000 6300 3500 310 1-170 1 118 1 55 42 14 20 10 1 only possible with guided float The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Conversion to other products or operating data is performed by KROHNE using the calculation method stipulated by VDI/VDE Directive 3513. 9

2 TECHNICAL DATA GA24 Measuring span 10 : 1 Flow values 100% Reference condition: Water 68 F / Air 68 F - 14.7 psi Float materials: 1 Stainless Steel or Hastelloy - 2 PTFE - 3 TFM - 4 Aluminium - 5 Polypropylen (PP) Materials 1 2 3 1 3 4 5 1 2 3 4 5 Cone Flow, water [gph] Flow, air [scfm] Max. pressure loss [psi] N 18.07 DN15 10.6 6.6 3.43 0.93 0.37 0.5 0.31 0.1 0.1 0.03 0.04 0.02 N 18.09 16.6 10.6 5.81 1.36 0.59 1.2 0.43 0.1 0.1 0.04 0.04 0.03 N 18.13 26.4 16.6 9.25 1.86 0.93 0.74 0.74 0.1 0.1 0.04 0.06 0.03 N 19.09 42.3 26.4 14.5 3.1 1.36 1.7 1.1 0.19 0.13 0.06 0.07 0.03 N 19.13 66.0 42.3 22.5 4.96 2.05 2.8 1.7 0.2 0.16 0.06 0.07 0.03 N 19.19 105 66.0 37 - - - - 0.3 0.2 0.07 - - N 19.26 166 106 60.8 - - - - 0.4 0.25 0.09 - - N 21.09 DN25 166 106 60.8 11.2 1 5.58 6.8 4.3 0.3 0.2 0.09 0.1 0.06 N 21.13 264 166 92.5 17.4 1 8.68 11 7.4 0.3 0.25 0.09 0.1 0.06 N 21.18 423 264 158 30.4 1-17 1 10.5 1 0.3 0.3 0.1 0.15 0.9 N 21.25 660 423 251 43.4 1-26 1 16 1 0.48 0.58 0.1 0.17 0.13 N 41.09 DN40 423 264 158 27.9 1 13.6 17 11 0.48 0.26 0.1 0.16 0.07 N 41.13 660 423 238 43.4 1 22.3 28 1 17.4 1 0.49 0.29 0.15 0.17 0.07 N 41.19 1057 660 396 79.4 1-47 1 28.5 1 0.55 0.35 0.16 0.22 0.1 N 51.10 DN50 1057 660 396 74.4 1 34.7 43.4 27.9 0.62 0.36 0.17 0.22 0.1 N 51.15 1664 1057 634 118 1 55.8 68 1 43.4 1 0.68 0.44 0.19 0.23 0.1 N 51.21 2642 1664 925 192 1-105 1 73 1 0.8 0.61 0.2 0.29 0.15 1 only possible with guided float The operating pressure should be at least double the pressure loss for liquids and five times for gases. The specified pressure losses are valid for water and air at maximum flow rate. Conversion to other products or operating data is performed by KROHNE using the calculation method stipulated in VDI/VDE Directive 3513. 10

GA24 INSTALLATION 3 3.1 Intended use The variable area flowmeters are suitable for measuring gases and liquids. The devices are particularly suitable for the measurement of: Liquids Hydrocarbons Water Chemicals with low corrosiveness Industrial gases For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation. Responsibility for the use of the measurement devices with regard to suitability, intended use and corrosion resistance of the used materials against the measured fluid lies solely with the operator. The manufacturer is not liable for any damage resulting from improper use or use for other than the intended purpose. Do not use any abrasive or highly viscous media. 3.2 Installation condtitions When installing the device in the piping, the following points must be observed: The variable area flowmeter must be installed vertically (measuring principle). Flow direction from bottom to top. For installation recommendations please refer also to VDI/VDE 3513 Sheet 3. Before connecting, blow or flush out the pipes leading to the device. Pipes for gas flow need to be dried before the device is installed. Use connectors suitable for the particular device version. Align the pipes centrically with the connection bores on the measuring device so they are free of stresses. If necessary, support the pipeline to reduce vibrations being transmitted to the measuring device. Do not lay signal cables directly next to cables for the power supply. 11

4 ELECTRICAL CONNECTIONS GA24 4.1 Limit switches GA24 The flowmeters GA24 can be equipped with a maximum of two bistable limit switches. Bistable function: Stable switching when passing through the switching point Use, selection and function - see Technical Data Figure 4-1: Types of limit switch 1 MS 14/I - floating, bistable reed contact 2 TG21 - bistable, with integrated switching lug and slotted proximity switch 12

GA24 ELECTRICAL CONNECTIONS 4 4.1.1 MS14 limit switch connection Figure 4-2: Connection limit switch type MS14 1 Connection Reed-switch 2 Connection Receiver device 3 Terminal connection (potential-free) The switching function requires a float with integrated magnet. 4.1.2 TG21 limit switch connection Figure 4-3: Connection limit switch type TG21 1 Terminal - 2 Terminal + The TG 21 is adapted for switch amplifiers with an intrinsically safe circuit as per DIN EN 50227 NAMUR. The TG 21 includes an inductive slotted proximity switch with bistable switching characteristics. The slotted proximity switch is activated by the immersion of an aluminum lug. The magnet of the immersion lug is moved by the magnet in the float. The switching function requires a float with integrated magnet. 13

5 ORDER FORM GA24 Please provide us with the missing information so that we can be of help to you as quickly as possible. Then please fax this page to the appropriate sales associate. We will then contact you as soon as possible. Device data Connection type: EN 1092 ASME B16.5 Pressure rating: Limit switch: K1 1 K2 2 Approval: Disable ATEX 1 1 limit switch 2 2 limit switches Rating data Product: Operating pressure: Rated pressure: Operating temperature: Rated temperature: Density: Viscosity: Measuring range: Comments: Absolute pressure Standard density Overpressure Operating density Contact data Company: Contact person: Telephone number: Fax number: E-mail: 14