Experiment GC : Analysis of BTEX by GC-FID

Similar documents
Quantitative Analysis of Hydrocarbons by Gas Chromatography

Retention Time Locking: Concepts and Applications. Application

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

no peaks missing peaks too small peaks increasing retention decreasing retention

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Carrier Gases in Capillary GC

Rapid and Reliable Detection of Dissolved Gases in Water

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Carrier Gases in Capillary GC

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

GCMSD-Headspace Analysis SOP

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

Multiple Gas#5 GC configuration Jan 2016

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

ISO 7941 First edition

5890II GC Standard Operating Procedure 9/2/2005

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

CHEMTRACE Fremont Blvd. Fremont, CA 94538, Tel: (510) Fax: (510)

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Automated Determination of Dissolved Gases in Water Anne Jurek

Applications Note: Use of "pentane equivalent" calibration gas mixtures

Zebron Gas Management

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

Environmental. Gas Canister sampling and analysis. EQLF, Cambridge, July 20 th Right Solutions Right Partner

3 Flow and Pressure Control

Buck Scientific, Inc.

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Weiyi Zheng, Xiaoxia Dong, Evan Rogers, Jimmie C. Oxley, and James L. Smith*

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

During today's webinar we're going to look at the following topics:

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

Traceability of greenhouse gases and volatile organic compounds for air monitoring in South Africa

System Overview. TCD Detector temperature setpoint Regulator

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

Calibration of a GC955

ISO INTERNATIONAL STANDARD

FROG-4000 TM Chemical Analysis System

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

FROG-5000TM Chemical Analysis System User s Manual

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

APPLICATION NOTE. GC Integrated Permeation Device

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Manual for continuous distillation

May 1, By John Heim,Doug Staples

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

FROG-4000 TM Chemical Analysis System

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

The. «Modular Alliance» Gas Generators. ALLIANCE, the partnership that keeps the gas flowing. One step ahead!

Boyle s Law: Pressure-Volume Relationship in Gases. PRELAB QUESTIONS (Answer on your own notebook paper)

Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector

Agilent Dimension Software for ELSD User Manual

instrumentation Pulsed Discharge Detectors Non-Radioactive, Multiple Mode Electron Capture / Helium Photoionization

6850 Series Control Module User Information

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

METHOD: 1024, Issue 2 EVALUATION: FULL Issue 1: 15 August 1987 Issue 2: 15 August 1994

Standard Test Method for Residual Acrylonitrile Monomer Styrene-Acrylonitrile Copolymers and Nitrile Rubber by Headspace Gas Chromatography 1

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Intermolecular Forces

Agilent 490 Micro GC Natural Gas Analyzer

AKTA pure 25 New Owner s Intro

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Svensk läkemedelsstandard

Gases and welding distributors continue

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

HiQ laboratory gas generators.

AN-03 Rotary valves applications

STANDARD OPERATING PROCEDURES

A New GC/MS System Designed for Helium Carrier Gas Conservation

The Gas Laws: Boyle's Law and Charles Law

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

1. Study the performance of a binary distillation column operated in batch mode.

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

REFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

ANALYSIS OF GASES PRODUCED BY THREE UNDERWATER CUTTING DEVICES

Transcription:

46 Experiment GC : Analysis of BTEX by GC-FID Learning Goals: Familiarity with gas chromatography Gain experience in temperature programming and method development Correctly use an internal standard, and compare results with and without Learn proper injection technique Reading: Before coming to the lab, please do some reading on gas chromatography from a reliable analytical source, like Skoog et al, Instrumental Analysis; Harris, D., Quantitative Analysis, 6 th ed.;, and/or Harvey, Modern Analytical Chemistry. Prelab: 1. What are the boiling points of benzene, toluene, ethylbenzene, o-xylene and m- xylene? Based on your answers, decide on suitable temperatures for the injector and detector of the GC, bearing the following in mind: the injector temperature should be at least 20 C-40 C higher than the highest boiling point in the mixture, and the detector should be at least 220 C, not higher than 280 C, and 20 C-40 C higher than the injector. 2. The Varian 3400 GC is equipped with the choice of split or splitless injection. Given the conditions of this experiment, which mode is more appropriate? Why? 3. The Varian 3400 GC has an FID detector. What is an FID detector? What are the advantages and disadvantages of this detector? Would you say it is a good choice for the analysis of BTEX? Why or why not? Introduction: Gas Chromatography (GC): Chromatography involves separating a mixture of analytes according to their partitioning between a stationary phase and a mobile phase. In gas chromatography, the mobile phase is an inert carrier gas, and the stationary phase is found in the column. The instrument being used in this lab is the Varian 3400 Gas Chromatograph. It is equipped with a capillary (or open tubular) column, which is 30 m in length. These long narrow columns provide much better resolution than the alternative, packed columns, which now find their best use in preparative GC applications. The volatile liquid or gaseous analyte mixture is injected onto the GC through a septum into a heated injection port. The temperature of the injector is selected so as to vapourize the sample upon injection. The sample vapour is then carried through the column by the carrier gas. As

they interact with the stationary phase to varying degrees, they are separated. With nonpolar stationary phases, the principal determining factor as to relative retention times on the column is the volatility of the analytes. Therefore one would expect to see an elution order that followed, or nearly followed, the order of the boiling points of the analytes. The detector temperature is chosen to be at least 20 C higher than the highest boiling point, in order to ensure all analytes are detected as gases. To get a better and more efficient separation of analytes, temperature programming is often employed (see Figure 1). The temperature of the column is raised during the course of the analysis. If there are analyte peaks at the beginning of a chromatogram that are overlapping, you might choose to start the analysis at a lower temperature, and raise the temperature back up after a few minutes. As well, there could be some late eluting peaks that can be forced to elute faster if the temperature is raised during the analysis. Raising column temperature decreases retention time for anlaytes. Since analyte peaks broaden as time on the column increases, so decreasing retention time will improve resolution. BTEX: BTEX are an important class of volatile environmental contaminants, and are frequently analyzed in environmental and drinking waters. Current CRD regulations, for example, require that all waste entering the municipal sewer system contain no more than 1 mg/l BTEX; the other organic contaminants which must be monitored are chlorinated phenols, polycyclic aromatic hydrocarbons (PAH), phenols and petroleum hydrocarbons. In this experiment, % levels of BTEX are quantified using the internal standard technique as described below. Pentane is used as an internal standard, as it fulfills the necessary requirements for an internal standard. Hexane is used as a solvent. Sources of BTEX include: Benzene: solvents in printing, paints, dry cleaning solutions, etc. Ethyl benzene: solvent for coatings; used to make styrene, and is a component in rubber and plastic wrap. Toluene: used to make benzene and urethane Xylenes: gasoline, solvents; used to make plasticizers, polyester film, film. 47 Procedure: Week 1: Varian 3400 GC: The Varian 3400 GC is equipped with an FID detector. From your prelab, you will know a hydrogen flame is used in detection. As most people are aware, hydrogen is an explosive gas, and must be handled properly. 1. You will need to turn on all of the necessary gases at the cylinders. Helium is the inert carrier gas, medical air is used as the oxidant, and hydrogen for the fuel. Open all of these cylinders.

48 The regulators should be set at 80 psi for each already, so there is no need to adjust these. Next, open the helium tap on the benchtop. Make sure this is the tap that is connected to the 3400 GC. Leave the other taps closed for now. 2. Next you will have to build a method. Compare your selected temperatures for the injector and detector with your partner. Decide together on temperatures that you agree upon, and check that your selected injector and detector temperatures are ok with your TA. Select a column temperature. Remember the column temperature is now bound between room temperature, and at least 20 C below your injector temperature. You can run your first injection isothermally if you like, or you can try a program. Take careful notes of your conditions, so that you can track your changes as you run trials. 3. To build the method, press BUILD/MODIFY, then press METHOD 1. Always use method 1 for this lab. Next you will be asked for an initial column temperature. Enter the temperature you have chosen, and press enter. The screen will keep leading you through a series of prompts: -initial column hold time: enter the amount of time (in minutes) you want the column to remain fixed at the initial temperature. -add column program? enter yes or no. Yes means you want to add a temperature program, so that the temperature will change over the course of the run. If you select yes, you will be prompted to enter a final column temperature, a hold time for the final temperature, and a rate of temperature increase, in C/minute. Column Temperature ( C) 150 50 5.00 Time (min) 15.00 21.00 Figure 1. Schematic of a possible GC temperature program In the example shown in Figure 1, a column is held at an initial column temperature of 50 C for 5.00 minutes, then the temperature is ramped at 10 C/min to a final temperature of 150 C. The final temperature is then held for an additional 6 minutes. 4. After entering this information, you will be asked if you want to add an additional column program. This is used if you want to ramp the temperature at two different rates, at different

times in the run, or if you want to hold the column at a particular temperature partway through, then ramp again afterwards. For our purposes, this probably won t be necessary, but is an option. 5. Next, you will be asked to enter an injector temperature. Enter the approved temperature, then press enter. 6. The last part you will have to build is the detector program. You will be prompted to enter a detector temperature, attenuation and range. The attenuation should be 8, and the range should be 9. You can adjust these later if needed. 7. There will be further prompts, but at this point, you can just press ACTIVATE then press METHOD 1. A red NOT READY light will remain on until the temperatures you have set have been reached. To see where the temperatures are at any moment, push status. 8. When the detector has reached about 200 C, you can light the detector. Please get your TA to supervise this. Turn on the medical air on the benchtop tap. Do the same for the hydrogen. Then, on the front of the GC, push IGNITE and hold for about 4 seconds. You might hear a faint pop as the flame lights. After releasing the button, check for ignition by holding the glass slide over the exhaust. If it fogs up immediately, the flame is lit. If not, try pressing ignite for another 4 seconds. If there is still no luck, turn off the hydrogen and air taps, and try again in a couple of minutes. If there is still no success, get further assistance from the SLI. Once the detector has been lit, you are ready to proceed with the analysis. 9. Injection technique is important for this experiment. You should inject one or two microlitres. Choose an injection volume in this range, note it, and stick with it. You can adjust the attenuation if your peaks are too small or too large. You and your partner can both take turns injecting. 10. You can use the QC for the method development portion of the experiment. This part is done in pairs, so either of you (or both) can use your highest standards. This is not a quantitative exercise. You are looking for a good separation of the analyte peaks, as well as the sharpest possible peaks, and fastest possible method. You should only budget one hour for this part of the experiment, so get the best possible method you can in this time frame. If you need any guidance, ask your TA. 11. After you have a suitable method, you and your partner can start your standard preparation. Only one set of standards needs to be prepared between the two of you. You will want to calibrate from 0-10% each of benzene, toluene, ethylbenzene, o-xylene, and m-xylene. The easiest way to do this is to prepare several (at least 5) multistandards, that contain a mixture of the analytes, in varying concentrations. As well, in this experiment we will use an internal standard, pentane. Each prepared standard solution should therefore contain a constant amount (5%) of pentane. The solvent for the preparation is hexanes. 49

50 Week 2: 12. Follow the same procedure to start the instrument. Re-enter your method parameters, as many other people have been using different programs, and yours is probably no longer on the instrument. You and your partner will inject all of your standards, a QC, and one sample each. As many times as you feel necessary. 13. After all of the data has been collected, activate method 4, turn off the hydrogen at the cylinder, then on the benchtop. Then turn off the medical air at the cylinder and benchtop. After you have finished cleaning up your things and are ready to leave, turn off the helium at the benchtop, then at the cylinder. Report A full report is required for the experiment. Answers to the points below are to be included in the discussion. 1. Calculate the ratio of the peak area component/peak area of pentane internal standard for each of the 5 components of the BTEX standards and unknown. Plot analytical calibration curves for all 5 BTEX components. From these curves, determine and report the concentration of each component of BTEX in the unknown sample as %; remember that the unknown contains 5 % pentane internal standard. NOTE: the integrator is linear over several orders of magnitude; it s a good idea to examine the areas of off-scale peaks, and decide if they also represent accurate/adequate data. 2. Hand in your labeled chromatograms as per formatting guidelines. 3. Calculate the mean retention time (RT) ± standard deviation for all components. 4. Calculate the relative retention time (RRT) for each component. The RRT is calculated by dividing the retention time of the compound of interest by the retention time of the pentane internal standard. Calculate the mean RRT ± standard deviation. Compare with the results from 3 above and discuss. 5. Discuss results for two components (you can pick any two) obtained from both internal standard and external standard calibration, with respect to this experiment. This means you must prepare two calibration curves for each of the two components that you have chosen, one using the internal standard results, and one using the external standard results. Then calculate the concentration of unknown, QC and spike samples from each of the cal. curves. Compare accuracy and precision between internal/external calculations.