Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Similar documents
Osmometers. Product Selection Guide. Wissenschaftliche Gerätebau Dr. Ing. Herbert Knauer GmbH Hegauer Weg Berlin

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Using the Akta Prime plus October 22, 2012

SOP: Buck Scientific BLC-20P HPLC Operation

Video Notes. LC Troubleshooting Series Retention Time Shifts

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Rapid and Reliable Detection of Dissolved Gases in Water

AKTA pure 25 New Owner s Intro

GCMSD-Headspace Analysis SOP

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Standard Operational Procedure for Capillary Electrophoresis

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Automated Liquid Handling Station

Prep LC Training Manual S.V.

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit

Operation of the Perkin Elmer TGA-GC/MS

Agilent 1100 Series Analytical and Preparative Scale Fraction Collectors. User s Guide

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

SOP for use of the SX19 Stopped-Flow Instrument

Increasing the productivity of ICP-OES with the Thermo Scientific Qtegra ISDS Software

Prep LC Training Manual

Tensiometer For the measurement of surface and interfacial tension

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

Tensiometer For the measurement of surface and interfacial tension

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Setting a New Standard in Flash Chromatography Performance

FROG-4000 TM Chemical Analysis System

Multiple Gas#5 GC configuration Jan 2016

Gradients in ion chromatography. Optimal separation with shorter analysis time

The instrument will process the samples one by one till all the samples have been cleaned up.

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

Automated Determination of Dissolved Gases in Water Anne Jurek

LC, RSLC, Bio RS and Dual Gradient Systems

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Agilent 7697A Headspace Sampler

Peak Biotech - Company & Product Presentation

Panther 5 Acute Care Ventilator

Versatility. Accuracy. Compact. Versa. Automated Headspace Vial Sampler

The instrument should have a mass range from 2 to 2000 m/z or better Interface

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Running HPLC Methods on an ACQUITY UPLC System

no peaks missing peaks too small peaks increasing retention decreasing retention

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

CORESTA RECOMMENDED METHOD Nº 67

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

Degasser DG 2.1S User Manual V6880 HPLC

FROG-5000TM Chemical Analysis System User s Manual

FROG-4000 TM Chemical Analysis System

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

Quantitative Analysis of Hydrocarbons by Gas Chromatography

HPLC. Solvent Delivery Module Systems

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Agilent 1260 Infinity Purification Solution

Autosampler problems!

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

LC-SPE. Organizer 3.0 User Manual. Innovation with Integrity. Version 001 NMR

Figure 1: Plug in ion inlet of mass spec

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

Valve INV-907. Instructions. 1 Introduction. 2 Installation. Safety. Unpacking. Installing the valve

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178

SHIMADZU LC-10/20 PUMP

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Quick Installation Guide Doc nr.: C Date:

Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

Instrumental operation: ICS 3000

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

5890II GC Standard Operating Procedure 9/2/2005

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

CP-8400 AutoSampler and CP-8410 AutoInjector

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

Quick Start Guide for SA-9600 Series Surface Area Analyzer

Method Development Solutions

Laboratory Mortar Mixer (Testing)

UltiMate 3000 Viper Capillary Kits for Intelligent LC (LCi) System Solutions

Agilent Dimension Software for ELSD User Manual

Liquid Handling Application Designer V2.0. Software Manual P/N M5200R0

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

INTERNATIONAL STANDARD

Operating Manual. SUPREMA Calibration. Software for Fire and Gas Warning Units. Order No.: /01. MSAsafety.com

Transcription:

Application Note Rapid performance verification of AZURA systems with refractive index detector Method Keywords ID HPLC Quality control, system verification, refractive index, AZURA Analytical HPLC Plus system, OpenLAB CDS VSP0012N Summary Introduction This application note is a guideline describing step-by-step a compact performance verification (PV) of AZURA Analytical HPLC systems including a refractive index detector, using a reversed phase (RP) column and a standard test solution. This procedure is recommended for testing newly installed AZURA Analytical HPLC systems as well as for regular monitoring of system performance. We strongly recommend every HPLC user to check the reliability of their system before running an analysis. The accuracy and reproducibility of a HPLC separation method is strongly dependent on stable flow rates, the precision of detectors, injection valves and autosampler, respectively. This guideline provides a detailed instruction for testing the HPLC system performance using a standard test solution which is separated on a RP column. Retention time and area of two peaks from the standard test solution are used as parameters for the system performance and provide information about the status of the HPLC system. Five replicates are used for statistical analysis. The verification methods are adapted to the requirements of different AZURA Analytical HPLC systems. Note: This performance verification procedure is exclusively developed for refractive index detection mode. Performance Verification method KNAUER provides all method parameters as part of the performance verification document for your particular AZURA Analytical HPLC system. Note: The PV procedure is developed for standard system configurations which contain a 10 or 20 µl sample loop in the autosampler. Deviations from this configuration may require adaptions to the procedure. Note: The pictures exemplarily shown in this document are based on an AZURA Analytical HPLC Plus system and OpenLAB CDS.

Equipment Chemicals Double deionized water Uridin (CAS 58-96-8) Desoxyuridin (CAS 951-78-0) The data shown in this application note were produced with the KNAUER PV test mix containing uridin and desoxyuridin. The concentration of uridin and desoxyuridin is 1.0 mg/ml. For the verification procedure it is necessary to separate both substances and to evaluate the data regarding retention time and area of these two peaks. Note: We recommend producing your own test solution. 100 mg of uridin and 100 mg of desoxyuridin were dissolved in 100 ml water and sonicated for 10 minutes. Aliquots of the test solution can be stored for up to 24 months at -20 C. Column Software Eurospher II 100-5 C18A, 50 x 4 mm with precolumn (order no. 05WE184E2J) OpenLAB CDS Accessories 10 µl sample loop for manual injection valves (order no. A05645) 10 µl sample loop for autosampler 3950 (order no. A50078) 20 µl sample loop for manual injection valves (order no. A05646) 20 µl sample loop for autosampler 3950 (order no. A05646) After installing OpenLAB CDS and KNAUER drivers, ensure that all power saving functions of the operating system are switched off. Then open the software, create a project and configure the HPLC system. Note: It is strongly recommended to avoid the usage of space characters within any project names as well as data files or methods with OpenLAB CDS. Important note for OpenLAB After installing OpenLAB and KNAUER drivers, ensure that all power saving functions of the operating system are switched off. Then launch OpenLAB CDS and create a project. Next configure the HPLC system. Note: It is strongly recommended to avoid the usage of space characters within any project names as well as data files or methods with OpenLAB. VSP0012N www.knauer.net Page 2 of 8

Method parameters Injection The threshold values for the PV test were determined with the injection parameter Full loop. Deactivate the options Air segment and Headspace pressure. Set the Syringe speed to low and use the Needle wash option. Fig.1: AS 3950 Note: In the mode of Partial loopfill you can inject up to 50 % of the full loop volume, whereas in the mode of Full loop the volume of the entire loop is injected. Pump Fig.2: P 6.1L Applications based on refractive index detection are usually isocratic due to the principle of measurement. Nevertheless, also a high pressure gradient (HPG) or low pressure gradient (LPG) pump can be used. It is important to set the Maximum Pressure depending on the column specifications and the Compressibility factor (CF) of the solvent. VSP0012N www.knauer.net Page 3 of 8

Detector Fig.3: RID 2.1L The detection is performed with a time constant of 0.05 sec and 20 Hz sampling rate. Activate the options Autozero and set the Target temperature to 35 C. Additional settings Auxiliary traces It is always useful to record the pressure and the temperature of the detector to acquire further data in case of separation problems. Trigger If the HPLC system includes an autosampler the trigger type is always external. Integration events It is recommended to use the default settings for the integration parameters width and threshold. Sometimes it is necessary to define integration off to ensure correct baseline behavior. Peaks/groups After a test run the retention time should be inserted into the peaks/groups table to enable automated peak recognition and peak integration via report function. VSP0012N www.knauer.net Page 4 of 8

Verification procedure First steps This section describes step-by-step the procedure for validating an AZURA Analytical HPLC system including a refractive index detector. First of all enable the detector to warm up the lamp for at least 30 minutes to guarantee best performance. Use the warming-up time to connect channel A with a bottle of fresh double deionized water and to mount the column (Eurospher II 100-5 C18A, 50 x 4mm). Launch the HPLC system via OpenLAB CDS, open the verification method and check again the method configuration for the AZURA Analytical HPLC system. Note: Check the pressure limits of all HPLC components to prevent damage to your system. Open Control Instrument status to check whether all devices are ready. This is indicated by a green dot in the bottom left corner of the window. Purge the pump Always purge the channels you use before starting a method. Open the purge valve and choose the purge function to flush any air bubbles out of the system. Note: After installation of the HPLC system, maintenance or exchange of capillaries, we recommend flushing the system without column and detector. Then download the method (Control Download Method) close the purge valve and equilibrate the column for at least 10 minutes. Flush the detector cell Note: We recommend flushing the detector cell several times after the column is equilibrated with the solvent. VSP0012N www.knauer.net Page 5 of 8

Sequence Create a new sequence (File New Sequence Blank.sec) and enter all required parameters like method, injection volume as well as the corresponding vial position for your test solution. The Run Type should be configured as indicated in the figure below (Summary Begin, Summary Run and Summary End). Report The report template is defined on the right side of the window if you choose the Begin Summary option. You can choose your preconfigured advanced report template (see below). Start the sequence Note: If you activate the report option Sequence you will find the report as a pdf file in the Result path folder. VSP0012N www.knauer.net Page 6 of 8

Threshold values The given threshold values are specific for KNAUER analytical HPLC systems including a refractive index detector such as AZURA RID 2.1L. For a standard system configuration as described above, we recommend to accept values of % RSD of < 0.5 % regarding retention time and < 1.0 % regarding peak area to pass the PV test. Note: KNAUER provides specific limiting values together with the PV method. They can vary depending on the system configuration. How to create an Advanced Report with OpenLAB? This short description will help you to create an Advanced Report in OpenLAB for calculation of verification parameters. The arrows indicate the obligate selections. File New Report Blank.tpl. Right Click Table Wizard Sequence Summary Table. Follow the wizard and select the parameters retention time and area for verification. In the next selection windows choose: Named Peaks that were detected (deactivate the other options) Data Filename down yes Finish the wizard and save the report template. Troubleshooting If the system does not pass the PV test, check the following points to locate the cause of the issue. Compare the pressure trace with former successful PV tests to ensure that the pump is working properly. Check the RID 2.1L temperature trace (if available) to see whether the temperature is stable or out of limits. Compare also the peak shape with former runs. Deviations can indicate that capillary connections are not suitable for a successful PV test. For more information please visit our homepage, consider KNAUER HPLC troubleshooting guide and FAQs on www.knauer.net or contact us via email or phone. Note: Keep in mind that the methods are optimized for new columns and certain parameters e.g. retention time may vary after a certain amount of injections or the usage of alternative eluents. VSP0012N www.knauer.net Page 7 of 8

Conclusion With this procedure we can offer a robust and fast method to validate the reproducibility of the performance of a KNAUER AZURA Analytical HPLC system including refractive index detector. We recommend running a PV test at least once a year or whenever a part of the HPLC system is exchanged or varied. Author Hagen Schlicke, LC LAB, Project Manager / Application Specialist Contact information KNAUER Tel: +49 30 809727-0 Wissenschaftliche Geräte GmbH Fax: +49 30 8015010 Hegauer Weg 38 E-Mail: info@knauer.net 14163 Berlin, Germany Internet: www.knauer.net VSP0012N www.knauer.net Page 8 of 8