Application of PEEP using the i-gel during volume-controlled ventilation in anesthetized, paralyzed patients

Similar documents
6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge

Effect of Continuous Cuff Pressure Regulator in General Anaesthesia with Laryngeal Mask Airway

Why we should care (I)

LMA Airway portfolio. 3,000 published references. 300 million patient uses. Every 3 seconds an LMA airway is used somewhere in the world.

A simple method of partial inflation of the LMA cuff before insertion in children to allow cuff pressure without adjustment after insertion

ROUTINE PREOXYGENATION

QED-100 Clinical Brief

Description: Percentage of cases with median tidal volumes less than or equal to 8 ml/kg.

Mechanical Ventilation

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA

VENTILATORS PURPOSE OBJECTIVES

Effect of Fresh Gas Flow on Isoflurane Concentrations during Low-flow Anaesthesia

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

ONLINE DATA SUPPLEMENT. First 24 hours: All patients with ARDS criteria were ventilated during 24 hours with low V T (6-8 ml/kg

Lung recruitment maneuvers

27/10/58. Background and Rationale. Background and Rationale. Background and Rationale. Objective. Background and Rationale

Pressure -Volume curves in ARDS. G. Servillo

Javier García Fernández. MD. Ph.D. MBA. Chairman of Anaesthesia, Critical Care and Pain Service Puerta de Hierro University Hospital Associate

Javier García Fernández. MD. Ph.D. MBA. Chairman of Anaesthesia, Critical Care and Pain Service Puerta de Hierro University Hospital Associate

EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES

Mechanical power and opening pressure. Fellowship training program Intensive Care Radboudumc, Nijmegen

Stratégie ventilatoire pendant la RCP Pr Jean-Christophe M Richard

Disclosures. The Pediatric Challenge. Topics for Discussion. Traditional Anesthesia Machine. Tidal Volume = mls/kg 2/13/14

VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL

APNOEA AND PRE-OXYGENATION

Prof. Javier García Fernández MD, Ph.D, MBA.

VIMA. (Volatile Induction and Maintenance Anesthesia) How and Why. James H. Philip M.E.(E), M.D.

RESPIRATORY CARE POLICY AND PROCEDURE MANUAL. a) Persistent hypoxemia despite improved ventilatory pattern and elevated Fl02

Using the laryngeal mask airway to manage the difficult airway

SUPPLEMENTARY APPENDIX. Ary Serpa Neto MD MSc, Fabienne D Simonis MD, Carmen SV Barbas MD PhD, Michelle Biehl MD, Rogier M Determann MD PhD, Jonathan

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo

Clinical Benefits of Low- and Minimal- Flow Anesthesia

Let s talk about Capnography

The effect of tracheal tube size on air leak around the cuffs

Pulmonary gas exchange is well preserved during robot assisted surgery in steep Trendelenburg position

Recommendations for the minimal monitoring of the patient during anaesthesia.

Product information. The new LTS-D. The 2nd generation supraglottic airway device ideal for clinical use and pre-hospital environment

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

EQUIPMENT. The LMA ProSeal a laryngeal mask with an oesophageal vent

Update on Thoracic Anesthesia: Perioperative Lung Protective Strategies in One- and Two-Lung Ventilation

Volume Diffusion Respiration (VDR)

APRV: Moving beyond ARDSnet

Key words: intrahospital transport; manual ventilation; patient-triggered ventilation; respiratory failure

Unit 15 Manual Resuscitators

TESTCHEST RESPIRATORY FLIGHT SIMULATOR SIMULATION CENTER MAINZ

excellence in care Procedure Management of patients with difficult oxygenation. For Review Aug 2015

Monitoring, Ventilation & Capnography

PROBLEM SET 9. SOLUTIONS April 23, 2004

RESPIRATORY PHYSIOLOGY. Anaesthesiology Block 18 (GNK 586) Prof Pierre Fourie

Ventilating the Sick Lung Mike Dougherty RRT-NPS

Invasive mechanical ventilation:

Invasive Ventilation: State of the Art

Mechanical Ventilation of the Patient with ARDS

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

Comparison Of The Efficacy Of Two Low Fresh Gas Flow Techniques In Low Flow Anaesthesia

Difficult Oxygenation Distribution: Sydney X Illawarra X Orange X

Laryngeal Mask Airway (LMA) Indications and Use for the NH EMT-Intermediate and Paramedic

Flight Medical presents the F60

New Frontiers in Anesthesia Ventilation. Brent Dunworth, MSN, CRNA. Anesthesia Ventilation. New Frontiers in. The amount of gas delivered can be

RESPIRATORY PHYSIOLOGY, PHYSICS AND

Operation and Maintenance of the EPV200 Portable Ventilator

J.A. BAIN AND W.E. SPOEREL~

Chapter 3: Invasive mechanical ventilation Stephen Lo

Respiration (revised 2006) Pulmonary Mechanics

Model Based Approach to Estimate dfrc in the ICU Using Measured Lung Dynamics

Mechanical Ventilation

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

CEEA 2015, Kosice Luciano Gattinoni, MD, FRCP Università di Milano Fondazione IRCCS Ca Granda Ospedale Maggiore Policlinico Milan, Italy

Linear Infl ation Technology Non-invasive Blood Pressure Measurement. inibp Case Report Experiences in OR

Learning objectives. First Response Learning Module 2 Based on ILCOR and ANZCOR 2016

Neonatal tidal volume targeted ventilation

Javier García Fernández. MD. Ph.D. MBA. Chairman of Anaesthesia and Critical Care Service Puerta de Hierro University Hospital Associate Professor.

First Response Learning Module 2 Based on ILCOR and ANZCOR Victorian Newborn Resuscitation Project Updated March 2018

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine

VENTILATION AND PERFUSION IN HEALTH AND DISEASE. Dr.HARIPRASAD VS

The NuMask is as effective as the Face Mask in

Operating & troubleshooting a self inflating bag. Victorian Newborn Resuscitation Project Updated February 2012

RESPIRATORY PHYSIOLOGY RELEVANT TO HFOV

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

Initiation and Management of Airway Pressure Release Ventilation (APRV)

Bunnell LifePulse HFV Quick Reference Guide # Bunnell Incorporated

Organis TestChest. Flight Simulator for Intensive Care Clinicians

The Internet Journal of Anesthesiology 2010 : Volume 22 Number 2

How does HFOV work? John F Mills MBBS, FRACP, M Med Sc, PhD Neonatologist Royal Children s Hospital. Synopsis

Neonatal Assisted Ventilation. Haresh Modi, M.D. Aspirus Wausau Hospital, Wausau, WI.

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use

Physiological Basis of Mechanical Ventilation

Functional residual capacity in beagle dogs with and without acute respiratory distress syndrome


Section 2.9 Decannulation

Objectives. Respiratory Failure : Challenging Cases in Mechanical Ventilation. EM Knows Respiratory Failure!

Driving Pressure. What is it, and why should you care?

Collin County Community College. Lung Physiology

PROCEDURE (TASK): ROUTINE VENTILATOR CHECK. 5. Verifies current ventilator Insures correspondence between physician's

Selecting the Ventilator and the Mode. Chapter 6

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

Guidelines on Monitoring in Anaesthesia

Transcription:

J Anesth (2013) 27:827 831 DOI 10.1007/s00540-013-1628-2 ORIGINAL ARTICLE Application of PEEP using the i-gel during volume-controlled ventilation in anesthetized, paralyzed patients Yong Beom Kim Young Jin Chang Wol Seon Jung Sang Ho Byen Youn Yi Jo Received: 21 March 2013 / Accepted: 24 April 2013 / Published online: 7 May 2013 Ó The Author(s) 2013. This article is published with open access at Springerlink.com Abstract Purpose This prospective, randomized trial was designed to assess whether the i-gel supraglottic airway device is suitable for volume-controlled ventilation while applying positive end-expiratory pressure (PEEP) of 5 cmh 2 O under general anesthesia. It was believed that this device might improve arterial oxygenation. Methods Forty adult patients (aged 20 60 years) scheduled for elective orthopedic surgery were enrolled in this study. Twenty patients were ventilated without external PEEP [zero positive end-expiratory pressure (ZEEP) group], and the other 20 were ventilated with PEEP 5 cmh 2 O (PEEP group) after placing an i-gel device. Volume-controlled ventilation at a tidal volume (TV) of 8 ml/kg of ideal body weight, leak volume, and arterial blood gas analysis were investigated. Results The incidences of a significant leak were similar in the ZEEP and PEEP groups (3/20 and 1/20, respectively; P = 0.605), as were leak volumes. No significant PaO 2 difference was observed between the two groups at 1 h after satisfactory i-gel insertion (215 ± 38 vs. 222 ± 54; P = 0.502). Conclusions The use of an i-gel during PEEP application at 5 cmh 2 O did not increase the incidence of a significant air leak, and a PEEP of 5 cmh 2 O failed to improve arterial oxygenation during controlled ventilation in healthy adult patients. Y. B. Kim Y. J. Chang W. S. Jung S. H. Byen Y. Y. Jo (&) Department of Anesthesiology and Pain Medicine, Gachon University Gil Medical Center, 1198 Guwol-dong, Namdong-gu, Incheon 405-760, Korea e-mail: endless37@gilhospital.com Keywords Introduction i-gel PEEP Controlled ventilation The i-gel (Intersurgical, Wokingham, Berkshire, UK) was introduced into clinical practice in 2007. It is made of a thermoplastic elastomer, a soft gel-like substance [1]. This disposable device is designed to fit the perilaryngeal and hypopharyngeal structures without the use of an inflatable cuff, in contrast to other supraglottic airway devices [2], and has the advantages of easier insertion, minimal tissue compression, and fewer positional changes after cuff inflation. The gold standard for airway management remains endotracheal intubation, but in view of the fact that minimizing interruptions to chest compressions, to maximize coronary and cerebral perfusion pressure, is the most important aspect of resuscitation, supraglottic airway devices could be a good substitute for airway management during cardiopulmonary resuscitation. Furthermore, i-gel insertion has been reported to be faster than ProSeal laryngeal mask airway (LMA), tracheal tube, and classic LMA during resuscitation [3]. Because most supraglottic airway devices, including i-gel, have a low airway leak pressure, moderate tidal volumes of 6 8 ml/kg have been recommended during positive pressure ventilation [4]. Although there were no differences in the amount of atelectasis in the patients undergoing general anesthesia without lung injury between a tidal volume of 10 and 6 ml/kg [5], in patients with acute lung injury, changes in tidal volume from 10 to 6 ml/kg increase the alveolar collapse, which can reversed by positive end-expiratory pressure (PEEP) application [6]. Applying PEEP in addition to controlled ventilation has been suggested to increase functional residual capacity and

828 J Anesth (2013) 27:827 831 alveolar recruitment and to improve oxygenation and ventilation/perfusion mismatching under endotracheal intubation or when a ProSeal LMA is used [7 9]. However, because of the absence of an inflatable cuff, theoretically the i-gel might be more likely to have gas leaks during positive pressure ventilation than other supraglottic airway devices [10]. Recent studies support the use of i-gel during anesthesia for spontaneous breathing or controlled ventilation [10 12]. However, no study has yet shown that i-gel provides a good seal during PEEP application. Accordingly, the present study was designed to assess whether the i-gel is suitable for volume-controlled ventilation during PEEP application at 5 cmh 2 O under general anesthesia, and whether this device improves arterial oxygenation. Methods This study was approved by our institutional review board, and written informed consent was obtained from all eligible participants. Subjects Forty adult patients (aged 20 60 years) of American Society of Anesthesiologists (ASA) physical status I or II scheduled for elective orthopedic surgeries were enrolled in this study. The following exclusion criteria were applied: a morbidly obese status (body mass index [30 kg/m 2 ), a history of cerebrovascular or respiratory disease, neck or upper respiratory tract pathology, potentially difficult intubation, an increased risk of aspiration, and pregnancy. Patients were randomly allocated to one of two groups before the induction of anesthesia using computer-generated random numbers. The members of the ZEEP group were ventilated with no external PEEP application (n = 20), whereas members of the PEEP group were ventilated at a pressure of 5 cmh 2 O(n = 20) after placing the i-gel. Anesthesia All 40 patients received intramuscular midazolam (0.05 mg/kg) and glycopyrrolate (0.2 mg) as premedication 1 h before anesthesia induction. On arrival at the operating room, standard monitors, including an electrocardiogram, pulse oximeter, and a noninvasive arterial pressure, were applied to all patients. i-gel sizes were selected using patient weight, as follows: size #3 for patients\50 kg, size #4 for patients between 50 and 70 kg, and size #5 for patients [70 kg. Anesthesia was induced with propofol 1.5 2.0 mg/kg, rocuronium 0.3 mg/kg, and alfentanil 10 lg/kg. After i-gel insertion, anesthesia was maintained with desflurane to maintain a bispectral index (BIS) between 40 and 50. Adequate placement was determined by observing the end-tidal carbon dioxide (ETCO 2 ) waveform and chest movements, as previously reported [10]. If ventilation was inadequate, the device was gently pushed or the patient s head and neck were repositioned. A failed attempt was defined as removal of the device from the mouth for reinsertion and was excluded from the statistic analysis. All patients were ventilated using an S/5 Avance anesthetic machine (GE Healthcare, Madison, WI, USA). After anesthesia induction, volume-controlled ventilation at a constant flow and an I:E ratio of 1:2 was performed at a tidal volume (TV) of 8 ml/kg of ideal body weight. Values were estimated using the following equation: 50? 0.91 9 (height in cm - 152.4) for men and 45.5? 0.91 9 (height in cm - 152.4) for women. A respiratory rate of 8 16 breaths/min was adjusted to maintain an ETCO 2 of 30 35 mmhg at 40 % inspired oxygen in air using a fresh flow gas rate of 3 l/min. Evaluating leak volume and leak fraction Hemodynamic variables, ETCO 2, and ventilatory parameters were monitored and recorded at 5 min (T 1,baseline values), 30 min (T 2 ), and 1 h (T 3 ) after satisfactory i-gel insertion. An arterial blood gas sample was obtained by a single sterile puncture from a radial artery at 1 h after i-gel insertion. Leak volume was defined as the difference between inspired tidal volume and expired tidal volume, and leak fraction (LF) was defined as leak volume/inspired tidal volume. The primary outcome variable was LF. The sample size was calculated on the basis of a preliminary study of 10 patients. The mean LFs were 0.05 and 0.08 before and after application of PEEP 5 cmh 2 O, and to detect a mean difference ± SD in actual LF of 0.03 ± 0.03 with an a-error of 0.05 and power of 80 % between the two groups in terms of applying PEEP, 16 patients were required in each group. Assuming a dropout rate of *20 %, we calculated that 20 patients would be required per group. Statistical analysis The statistical analysis was performed using SPSS ver. 12.0 (SPSS, Chicago, IL, USA). Results are expressed as mean ± SD or as numbers of patients. Differences between the two groups were analyzed using the t test. Statistical significance was accepted for P values \0.05. Results All 40 patients enrolled and had their measures completed without any events. Patient characteristics and perioperative

J Anesth (2013) 27:827 831 829 Table 1 Patient characteristics and perioperative data data were similar in the ZEEP and PEEP groups (Table 1). Selected i-gel sizes were nonsignificantly different in the two groups (P = 0.071). All i-gel devices were inserted at first attempt and an acceptable airway was achieved in all patients. The incidence of significant leaks (defined as an LF [ 0.2) was 3/20 in the ZEEP group and 1/20 in the PEEP group, which was not a significant difference (P = 0.605). Actual leak fractions were similar between groups at T 1 (0.08 ± 0.1 vs. 0.06 ± 0.07; P = 0.408), T 2 (0.09 ± 0.09 vs. 0.05 ± 0.04; P = 0.135), and T 3 (0.09 ± 0.08 vs. 0.05 ± 0.03; P = 0.208). Changes in respiratory parameters and arterial blood gas analysis results are presented in Table 2. Leak volumes were similar in the two groups. However, respiratory rate was significantly higher in the ZEEP group (at T 1, T 2, and T 3 ; P = 0.05, 0.005, and 0.012, respectively). No significant intergroup difference was found for peak airway pressure, but mean airway pressure was significantly higher in the PEEP group at T 1 (7 ± 1 vs. 9 ± 1, P \ 0.001), T 2 (7 ± 1 vs. 9 ± 1, P \ 0.001), and T 3 (7 ± 1 vs. 9 ± 1, P = 0.002). Air entry into the stomach was not detected by auscultation over the epigastric area in any patient. Mean PaO 2 values in the ZEEP and PEEP groups at T 3 were similar (215 ± 38 vs. 222 ± 54; P = 0.502). No patient experienced desaturation or CO 2 retention gastric insufflation, regurgitation, or aspiration during surgery. Finally, no blood was visible on i-gels after removal, and no patient complained of a severe sore throat or soft tissue or tooth injury. Discussion ZEEP (n = 20) PEEP (n = 20) Age (years) 44 ± 13 41 ± 14 Males/females 12/8 10/10 Weight (kg) 65 ± 11 65 ± 11 Height (cm) 168 ± 9 166 ± 8 Anesthesia time (min) 81 ± 40 85 ± 26 Operation time (min) 50 ± 41 57 ± 24 i-gel number #3/4/5 3/16/1 9/9/2 Data are presented as mean ± SD or numbers of patients ZEEP zero positive end-expiratory pressure PEEP 5 cmh 2 O positive end-expiratory pressure We found that the use of an i-gel and the application of PEEP at 5 cmh 2 O did not increase significant air leak incidence or leak volume. Furthermore, the i-gel device provided acceptable airways without any desaturation or Table 2 Perioperative respiratory and blood gas parameters ZEEP (n = 20) PEEP (n = 20) P value Leak volume (ml/kg) T 1 39 ± 47 30 ± 37 0.515 T 2 45 ± 41 30 ± 19 0.226 T 3 47 ± 42 27 ± 19 0.063 Leak fraction T 1 0.08 ± 0.1 0.06 ± 0.07 0.408 T 2 0.09 ± 0.09 0.05 ± 0.04 0.135 T 3 0.09 ± 0.08 0.05 ± 0.03 0.208 Peak airway pressure (cmh 2 O) T 1 14 ± 3 16 ± 3 0.086 T 2 16 ± 4 17 ± 3 0.412 T 3 16 ± 4 18 ± 3 0.209 Mean airway pressure (cmh 2 O) T 1 7 ± 1 9 ± 1* \0.001 T 2 7 ± 1 9 ± 1* \0.001 T 3 7 ± 1 9 ± 1* 0.002 Expiratory tidal volume (ml) T 1 479 ± 92 504 ± 92 0.877 T 2 480 ± 97 466 ± 141 0.579 T 3 479 ± 92 504 ± 92 0.412 Respiratory rate (breaths/min) T 1 13 ± 2 12 ± 2* 0.015 T 2 13 ± 2 11 ± 2* 0.005 T 3 13 ± 2 11 ± 2* 0.012 ETCO 2 (mmhg) T 1 34 ± 2 33 ± 3 0.476 T 2 33 ± 2 32 ± 2 0.213 T 3 33 ± 2 32 ± 2 0.563 ph 7.47 ± 0.02 7.46 ± 0.05 0.246 PaO 2 (mmhg) 215 ± 38 222 ± 54 0.631 PaCO 2 (mmhg) 37 ± 2 36 ± 4 0.228 5 min (T 1 ), 30 min (T 2 ), and 1 h (T 3 ) after proper insertion of the i-gel device. Data are presented as mean ± SD ZEEP zero positive end-expiratory pressure PEEP 5 cmh 2 O of positive end-expiratory pressure * P \ 0.05 vs. ZEEP group at each time point CO 2 retention event regardless of PEEP. In addition, PEEP at 5 cmh 2 O failed to improve arterial oxygenation during controlled ventilation using an i-gel. I-gel is a novel supraglottic airway device with a shorter insertion time, higher mean leak pressure, and a better fiberoptic view score than the standard disposable laryngeal mask airway, and has been reported to reduce the incidence of sore throat, dysphagia, and neck pain [13]. In another study that compared the LMA Supreme and the i-gel in paralyzed, ventilated patients undergoing gynecological laparoscopic procedures [11], similar satisfactory results were obtained for ease of insertion, success rate on

830 J Anesth (2013) 27:827 831 first attempt, time to insertion, and oropharyngeal leak pressure. Gatward et al. [3] reported that time taken for i-gel insertion was *50 % that of other airway devices, such as the tracheal tube classic LMA and ProSeal LMA during chest compression for cardiopulmonary resuscitation. We believe that the use of i-gel with proper PEEP could contribute to adequate ventilation and prevent atelectasis during general anesthesia or resuscitation. Because, theoretically, the absence of an inflatable cuff might increase the risk of gas leaks during the application of PEEP, the objective of this study was to determined whether the application of PEEP with an i-gel fitted could maintain acceptable ventilation without clinically significant air leakage. Uppal et al. [10] reported that the i-gel achieved a median airway leak pressure of 28 cmh 2 O, which is higher than that of conventional LMA [22 ± 6 (8 40) cmh 2 O] and similar to that of the ProSeal LMA [27 ± 7 (10 40) cmh 2 O] [14]. Although peak airway pressure was slightly higher in the PEEP group (not significant), it might not exceed the leak pressure. Considering that peak airway pressure was 10 23 cmh 2 O and mean airway pressure was 6 11 cmh 2 O in both our study groups, it was not unexpected that we did not encounter any clinically eventful oropharyngeal air leakage. The use of high fraction inspired oxygen during anesthetic induction causes atelectasis to develop within minutes. However, the application of adequate PEEP can prevent atelectasis and provide a higher level of oxygenation [15]. Goldmann et al. [8] demonstrated that the application of PEEP 5 cmh 2 O under general anesthesia improved gas exchange in children when the ProSeal LMA was used with an inflatable cuff. However, improvements in functional residual capacity do not always correlate with major changes in oxygenation. For example, Futier et al. [16] reported, in nonobese patients, that PEEP at 5 cmh 2 O improved end-expiratory lung volume 15 % but not oxygenation. Although gas exchange is often used to assess lung function during controlled ventilation, it has been reported that oxygenation does not provide a specific measure of nonaerating lung tissue [17, 18]. In a study of anatomical shunt compartment by whole lung computed tomography and of functional shunt using blood gas analysis at PEEP values of 5 and 15 cmh 2 O, functional shunt was found to be poorly correlated with the anatomical shunt compartment (r 2 = 0.174), which was attributed to a large variability in apparent perfusion ratio [18]. It was concluded that gas exchange variations could not be used with sufficient confidence to assess anatomical lung recruitment [18]. In the present study, we did not detect improvements in arterial oxygenation, and thus we suggest that further study using computed tomography (CT) or imaging devices be conducted to determine the effect of PEEP on atelectasis and alveolar recruitment. The present study has several limitations. First, leak pressure was not measured using a manometer; instead, we checked for the absence of any audible throat noise and the proper achievement of tidal volume because we wanted the study to reflect clinical usability. Also, we did not confirm the proper positioning using fiberoptic bronchoscopy. However, we ascertained that ventilation was not compromised in any patient. Second, morbidly obese patients and those with a respiratory problem, an increased risk of aspiration, or pregnant status were excluded, and thus our results cannot be directly applied to patients with reduced respiratory compliance. We conclude that the use of an i-gel and the application of PEEP at 5 cmh 2 O did not increase the incidence of a significant air leak. However, a PEEP of 5 cmh 2 O failed to improve arterial oxygenation during controlled ventilation with the i-gel device in healthy adult patients. Conflict of interest declare. The authors have no competing interests to Open Access This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited. References 1. Singh I, Gupta M, Tandon M. Comparison of clinical performance of i-gel with LMA-Proseal in elective surgeries. Indian J Anaesth. 2009;53:302 5. 2. Levitan RM, Kinkle WC. Initial anatomic investigations of the I-gel airway: a novel supraglottic airway without inflatable cuff. Anaesthesia. 2005;60:1022 6. 3. Gatward JJ, Thomas MJ, Nolan JP, Cook TM. Effect of chest compressions on the time taken to insertion airway devices in a manikin. Br J Anaesth. 2008;100:351 6. 4. Keller C, Brimacombe J. Spontaneous versus controlled respiration with the laryngeal mask: a review. Anaesthesist. 2001;50: 187 91. 5. Cai H, Gong H, Zhang L, Wang Y, Tian Y. Effect of low tidal volume ventilation on atelectasis in patients during general anesthesia: a computed tomographic scan. J Clin Anesth. 2007;19:125 9. 6. Richard JC, Maggiore SM, Jonson B, Mancebo J, Lemaire F, Brochard L. Influence of tidal volume on alveolar recruitment. Respective role of PEEP and a recruitment maneuver. Am J Respir Crit Care Med. 2001;163:1609 13. 7. Hedenstierna G. Airway closure and atelectasis and gas exchange during anesthesia. Minerva Anesthesiol. 2002;68:332 6. 8. Goldmann K, Roettger C, Wulf H. Use of the ProSeal laryngeal mask airway for pressure-controlled ventilation with and without positive end-expiratory pressure in paediatric patients: a randomized, controlled study. Br J Anaesth. 2005;95:831 4. 9. Lesur O, Remillard MA, St.-Pierre C, Falardeau S. Prophylactic positive end-expiratory pressure and postintubation hemodynamics: an interventional, randomized study. Can Respir J. 2010;17:e45 50.

J Anesth (2013) 27:827 831 831 10. Uppal V, Fletcher G, Kinsella J. Comparison of the i-gel with the cuffed tracheal tube during pressure-controlled ventilation. Br J Anaesth. 2009;102:264 8. 11. Teoh WHL, Lee KM, Suhitharan T, Yahaya Z, Teo MM, Sia ATH. Comparison of the LMA supreme vs. the i-gel TM in paralysed patients undergoing gynaecological laparoscopic surgery with controlled ventilation. Anaesthesia. 2010;65:1173 9. 12. Gatward JJ, Cook TM, Seller C, Handel J, Simpson T, Vanek V, Kelly F. Evaluation of the size 4 i-gel trade mark airway in one hundred non-paralysed patients. Anaesthesia. 2008;63:1124 30. 13. Keijzer C, Buitelaar DR, Efthymiou KM, Srámek M, ten Cate J, Ronday M, Stoppa T, Huitink JM, Schutte PF. A comparison of postoperative throat and neck complaints after the use of the i-gel and the La Premiere disposable laryngeal mask: a double-blinded, randomized, controlled trial. Anesth Analg. 2009;109:1092 5. 14. Brimacombe J, Keller C, Fullekrug B, Agrl F, Rosenblatt W, Dierdorf SF, de Garcia Lucas E, Capdevilla X, Brimacombe N. A multicenter study comparing the ProSeal and classic laryngeal mask airway in anesthetized, nonparalyzed patients. Anesthesiology. 2002;96:289 95. 15. Rusca M, Proietti S, Schnyder P, Frascarolo P, Hedenstierna G, Spahn DR, Magnusson L. Prevention of atelectasis formation during induction of general anesthesia. Anesth Analg. 2003;97: 1835 9. 16. Futier E, Constantin JM, Petit A, Jung B, Kwiatkowski F, Duclos M, Jaber S, Bazin JE. Positive end-expiratory pressure improves end-expiratory lung volume but not oxygenation after induction of anaesthesia. Eur J Anaesthesiol. 2010;27:508 13. 17. Wyche MQ Jr, Teichner RL, Kallos T, Marshall BE, Smith TC. Effects of continuous positive-pressure breathing on functional residual capacity and arterial oxygenation during intra-abdominal operations: studies in man during nitrous oxide and d-tubocurarine anesthesia. Anesthesiology. 1973;38:68 74. 18. Cressoni M, Caironi P, Polli F, Carlesso E, Chiumello D, Cadringher P, Quintel M, Ranieri VM, Bugedo G, Gattinoni L. Anatomical and functional intrapulmonary shunt in acute respiratory distress syndrome. Crit Care Med. 2008;36:669 75.