DOC-0087 Rev B (DOC-0086 CCN Operators Manual)

Similar documents
Constant Pressure Inlet (CCN) Operator Manual

Cloud Condensation Nuclei (CCN) Counter

Armfield Distillation Column Operation Guidelines

ACV-10 Automatic Control Valve

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Race Screen: Figure 2: Race Screen. Figure 3: Race Screen with Top Bulb Lock

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

BUBBLER CONTROL SYSTEM

BUBBLER CONTROL SYSTEM

KEM Scientific, Inc. Instruments for Science from Scientists

DeeterFlow Automatic Liquid Dispenser

PRODUCT MANUAL. Diver-Mobile for Android

APA software instruction manual

TR Electronic Pressure Regulator. User s Manual

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

Overview. Front Panel: Keypad and Display

HumiSys HF High Flow RH Generator

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

AX5000 Operational Manual

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003

Instruction Manual. BZ7002 Calibration Software BE

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

REASONS FOR THE DEVELOPMENT

HyperSecureLink V6.0x User Guide

PART 5 - OPTIONS CONTENTS 5.1 SYSTEM EXPANSION 5-3

XC2 Client/Server Installation & Configuration

Hydro-Control V User Guide

Built-in Purge Control Functions

( ) ( ) *( A ) APPLICATION DATA. Procidia Control Solutions Coarse/Fine Control. Split-Range Control. AD Rev 2 April 2012

Specifications and information are subject to change without notice. Up-to-date address information is available on our website.

FOOD INTAKE MONITOR. Hardware and Software User s Manual DOC-047 Rev 3.0. Copyright 2009 All Rights Reserved

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

Instruction Manual. Biogas Analyzer InCa Bio 04

Sequence of Operations

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

RM-80 respiration monitor

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

EZ-Boom 2010 Automated Application Control System Quick Reference Card INTRODUCTION

Relative Humidity Calibration Kit

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

SAPCON. User Manual. Capacitance Continuous Level Indicator. . Introduction. . General Description. . Principle of Operation. .

SHIMADZU LC-10/20 PUMP

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

WL16 WATER LEVEL LOGGERS Submersible pressure transducer and USB datalogger combination

Pneumatic high-pressure controller Model CPC7000

WF STEUERUNGSTECHNIK GMBH. INFORMATION and TECNICAL DESCRIPTION

DRILL MONITOR DM-100 OPERATIONS MANUAL

This addendum describes the back pressure control implementation if FuelCell Version 4.1a and later.

Touch Screen Guide. OG-1500 and OG Part # T011

Section 1: Overview... 6 Features... 6 Supported Software Part 1: Getting Started 7

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

Quick Installation Guide Doc nr.: C Date:

STX 165 Operation. Data Knob Enter Pushbutton. Power/Mode Control. Select Knob Ident Pushbutton. VFR Pushbutton

Gas Analyzer for TC Sensors (CO2, H2, N2) Series 62101, 64101, Operating Manual

Safety Manual. Process pressure transmitter IPT-1* 4 20 ma/hart. Process pressure transmitter IPT-1*

Technology. In the My Files [My Files] submenu you can store all the programs that you have made on the NXT or downloaded from your computer.

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

TYPE DOSAODOR-D SOFTWARE FOR CONFIGURATION OF TYPE DOSAODOR-D ODORANT INJECTION SYSTEM

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

Exercise 8. Closed-Loop Pressure Control, Proportional-Plus-Integral Mode EXERCISE OBJECTIVE

COMPRESSOR PACK SUCTION CONTROLLER TYPE: LP41x

Valve Communication Solutions Axiom

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

Pneumatic high-pressure controller Model CPC7000

CONSOLE-320 ENGLISH. 230A: CONSOLE-320 with cable data output Item 230B: CONSOLE-320 with cable + wireless radio data output

MP15 Jockey Pump Controller

Boyle s Law: Pressure-Volume Relationship in Gases

Cover Page for Lab Report Group Portion. Head Losses in Pipes

PC Configuration software for Discovery MkVI v 1.03 User guide

SCIENTIFIC DATA SYSTEMS, INC. Depth Tension Line Speed Panel. DTLS Manual

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

Using Modbus Protocol with the ALTUS Net Oil Computer. Instruction Manual

EZ Boom 2010 System for the EZ Guide 500 Lightbar Triangle Ag Services Users Guide

Standard Chiller 1/4 compressors Manual version: /10/96

Manual Weighingblock VB2 series and Uniscale

ECL Comfort 110, application 131 (valid as of software version 2.00)

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

AMS 6915 Board mount pressure sensor with digital output (I²C)

Cover Page for Lab Report Group Portion. Pump Performance

WATER CONTROL SYSTEM QUICK START

User Manual. FlowAnalyser

16 Series. Mass and Volumetric Flow Controllers. Precision Gas Flow Controller

BekkTech s Procedures For Performing In Plane Membrane Conductivity Testing

Instruction and Maintenance Manual

D-OptoLogger Dissolved Oxygen Logger Operation Manual

FTC130 Transmitter. Operating Manual

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

Wickets Administrator

STARLOG. Capacitive Water Depth Probe

STAND alone & p.c. VERSION

KIV-SACS Safe automatic calibration system:

Operating Manual. BACVis. Manual BACVis. for. Sensors and MilliGascounter. Rev

MODEL 6600 Rev. B4 USER'S MANUAL LAVERSAB INC., 505 GILLINGHAM LANE. SUGAR LAND TX (281) FAX: (281)

WELCOME TO THE REVOLUTION

Nothing's as cool, as a freshwater pool. easychem 770 ORP and ph Controller for Pools and Spas with Timer Output for Pool Pump

Transcription:

DOC-0087 Rev B (DOC-0086 CCN Operators Manual) 5710 Flatiron Parkway, Unit B Boulder, CO 80301 USA

5710 FLATIRON PARKWAY, SUITE B BOULDER, COLORADO, USA 80301 TEL: +1 (303) 440-5576 FAX: +1 (303) 440-1965 WWW.DROPLETMEASUREMENT.COM All rights reserved. No part of this document shall be reproduced, stored in a retrieval system, or transmitted by any means, electronic, mechanical, photocopying, recording, or otherwise, without written permission from. Although every precaution has been taken in the preparation of this document, assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use of the information contained herein. Information in this document is subject to change without prior notice in order to improve accuracy, design, and function and does not represent a commitment on the part of the manufacturer. Information furnished in this manual is believed to be accurate and reliable. However, no responsibility is assumed for its use, nor any infringements of patents or other rights of third parties, which may result from its use. All. product names and the Droplet Measurement Technologies, Inc logo are trademarks of. All other brands and product names are trademarks or registered trademarks of their respective owners. DOC-0087 Rev B

! " " #!! $ % &"' &"' ()" )" ( * *% # +"!,, $ "! - &". # $! "#$%# & '#(! FIGURE 1. SUPERSATURATION SETTINGS TAB DISPLAY...6 FIGURE 2. PROGRAM CONTROL TAB...7 FIGURE 3. TEMPERATURE TAB SCREEN SHOWING THE TEMPERATURE SET POINTS AND READOUTS FOR THE CCN TEMPERATURES....8 FIGURE 4. FLOW CONTROL SCREEN....11 FIGURE 5. OPTICAL PARTICLE COUNTER...13 FIGURE 6. SCREEN FOR THE CHART TIME HISTORY LENGTH....14 FIGURE 7. ALARM MONITOR...15 FIGURE 8. TABLE FOR SETTING THE SUOPERSATURATION PROGRAM PARAMETER...18 FIGURE 9. SCREEN FOR SETTING CALIBRATION COEFFICIENTS...20 FIGURE 10. CCN DATA PLAYBACK...22 FIGURE 11. CCN DATA PLAYBACK...23 DOC-0087 Rev B

1.1 Introduction The computer system for the Cloud Condensation Nuclei Counter (known here in as CCN) is a PC-104 stack unit, utilizing a Celeron 650 MHz processor, 256 Mb of memory, a Diamond Systems MM-16 AT multifunction interface card. A 20 Gb hard drive is provided for program and data storage. The software for the CCN counter is provided as a LabView executable. The software revision level is shown in the status bar of the CCN counter program. This manual is intended for versions 3.0.0 and above. Four programs are provided: SINGLE CCN 3.0.1.exe -main program for operating the CCN unit. SS settings.exe -sets up the supersaturation tables CCN Calibration.exe -sets the instrument specific coefficients for the flow and pressure calibrations. CCN Playback.exe -displays and replays saved data files

1.1.1 SINGLE CCN 3.0.1.exe The CCN control program provides a completely automated system for running the CCN instrument under laboratory or flight conditions. The program is organized into a tab structure for control and monitoring of the system. The program is set to load and run automatically after the operating system boots. CCN, serial numbers 1-10, used a peristaltic pump for the liquid and drain flows and require a different control program. CCN, serial numbers 11 and up and those CCN s units(1-10) that have been retrofitted with the solenoid pumps. Use a control solenoid. This software manual should be used in conjunction with the CCN operator s manual DOC-0086.

1.1.1.1 Supersaturation Settings Tab (SS) Figure 1. Supersaturation settings tab display The control program will start with the supersaturation tab setting active (see Figure 1). After starting the program, the user has 20 seconds to select the desired supersaturation table from the SS Table pull down menu. If no selection is made, the last SS table used is selected by default. After the 20 second timer expires, the supersaturation values, durations, and OPC bin settings are loaded and cannot be altered. Current SS# displays the current supersaturation selection (1, 2, 3, 4, or 5) Current SS displays the current supersaturation. If the instrument adjusts the supersaturation to maintain proper temperature ranges, the supersaturation may differ from the current selection s value. (1.1.1.3 Temperature Tab) Time until next SS change counts down the remaining time in seconds for the current supersaturation selection. After the 5 th selection completes, the instrument continues and loops back to the 1 st selection. Current SS# and Time until next SS change are not visible when Auto SS Temp Adjust mode is turned off. Details on setting up the supersaturation tables are given in section 1.1.2.

1.1.1.2 Program Control Tab (Prog Control) Figure 2. Program Control Tab The Program Control tab contains controls for file saving and data streaming. (Figure 2) Saving File controls whether the data is saved or discarded. This button is on by default. (1.1.18 File Saving) Start new file forces a new file to be started Start New File Every (min) controls the length of data files. Default value is 60 minutes and the allowed range is 1-180 minutes. Shutdown stops the execution of the program WARNING: It is important to follow the shutdown procedure when shutting down the instrument; this will insure that the pumps are shut off in order. Serial Stream Out turns on/off the stream of CCN operational parameters to the RS-232 serial port. Default value is. (1.1.1.7 Serial Output Stream) Watchdog alternates ON and OFF every second when the computer is cycling properly. The watchdog light on the instrument should blink at the same time as this indicator. If the indicator light is steady OFF or ON, reboot the computer.

1.1.1.3 Temperature Tab (Temps) Figure 3. Temperature tab screen showing the temperature set points and readouts for the CCN temperatures. The Temps tab contains heater and temperature indicators and controls. (Figure 3) The following details the temperature designations: T1 T2 T3 T OPC T Inlet T Nafion T Sample Top of column temperature Middle of column temperature Bottom of column temperature Optical Particle Counter temperature Temperature of the particle inlet at the head of the column Nafion humidifier temperature Air sample temperature at the CCN manifold The four temperature differentials are read from THE CONFIG file at startup and can be changed when Auto SS Temp Adjust is OFF. Recommended Value Range T1-T Sample 2.0 C 0 10 T Nafion-T1-1.0 C -5 5 T Inlet-T1 1.0 C 0 5 T OPC- T3 2.0 C 1 10

Column Temps Stabilized indicates if T1, T2, and T3 are less than 0.3 from their set values. Even when the light is on, additional time may be needed before the entire unit is stabilized and data is consistent. The light will turn off during SS% changes. Typically for small SS% changes the light will be off for less than one minute. If the Column Temps Stabilized light never comes on refer to Trouble Shooting Section 5.0 of the Operators Manual. Auto SS Temp Adjust allows for automatic selection of Delta T(C) from Current SS or for manual input of Delta T(C). Default value is the last setting when the program was previously running. When Auto SS Temp Adjust is switched Tec Relay switches the 28 VDC power buss in the instrument. This supplies power to the heaters, the TEC units, and the sample pump. Default value is on. Delta T(C) is the temperature differential used to calculate values for T1 Set, T2 Set, T3 Set, Inlet Set, Nafion Set, and OPC Set. In Auto SS Temp Adjust mode, Delta T(C) is an indicator only where: Delta T(C) = Current SS * SS to TG + SS Offset In Manual SS Temp Adjust mode, Delta T(C) becomes a control and can be set through the serial port. (1.1.1.8 Serial Stream) The program will readjust the temperature set values if any of three following conditions are met: 1. Delta T(C) changes 2. T Sample +2*(T1-TSample) < T1 Set 3. T Sample > T1 Set When temperatures are adjusted, the new temperature set values are calculated using: T1 Set new = T Sample + (T1-TSample) T2 Set new = T1 Set new + ½ Delta T(C) T3 Set new = T1 Set new + Delta T(C) * (1 (%TG)/100) Nafion Set new = T1 Set new + (T Nafion-T1)

Inlet Set new = T1 Set new + (T Inlet-T1) OPC Set new = T3 Set new + (T OPC- T3) Then in order to protect the OPC if: 1. OPC Set new > 55 Delta T(C) is reduced using: Delta T(C) new = 50 + Delta T(C) - OPC Set new All temperatures are then readjusted using Delta T(C) new. This continues until OPC Set < 55 It is important that the OPC operate at a higher temperature than the bottom of the column, T3. This temperature differential insures that the OPC optics will not fog. Studies at DMT have shown that the OPC temperature can be as high as 5 deg above T3 with no loss of concentration. At an OPC temperature of 7 deg above T3 there was a 20% loss of droplets. If the supersaturation scans involve changing from a high supersaturation, where the OPC temperature is high to a low supersaturation, with a low OPC temperature adequate time must be allowed for the OPC to cool to 5-7 deg above T3 before taking data. Otherwise there will be undercounting of the activated particles. The temperature at the top of the column needs to be slightly above the temperature of the sample. If the top of the column temperature is not above the sample temperature, the air comes into the top of the column from the nafion humidifier, which is at 100 % RH, can have condensation and cause fog to form at the top of the column. If the sample temperature rises by more than the T1-T Sample value, the system will reset all of the temperatures up to maintain this temperature difference. If the sample temperature drops by twice the difference of the T1-T Sample value the T1 value will be reset lower. It is important that the operating temperature of the CCN be higher than the dew point of the air being sampled. If the CCN is operated in a room with substantial air conditioning, and the temperature of the air in the room is less than the dew point of the sampled air, aerosol changes can occur.

1.1.1.4 Flows Tab (Flows) Figure 4. Flow Control Screen. The Flows tab contains indicators and controls associated with the system flows. (Figure 4) Flow Set (Vccm) displays the current total flow rate set value. This value is read from the SS_Settings.ini file when the SS Table is selected, but can be changed at any time. The recommended flow range is from 200-1000 volume CC/min (Vccm). V a V Valve Set (V) displays the voltage sent to the proportional valve that regulates the total airflow through the CCN. Every proportional valve will use a different Valve Set (V) to maintain a given flow. The limits on Valve Set (V) are uniquely calibrated by the factory. (1.3 XXXX Calibrate) Manual Override switches the manual valve set mode ON/OFF. In this mode the Valve Set (V) will be ignored and a constant voltage, Valve Set M (V), will be used in its place. Valve Set M (V) controls the voltage supplied to the proportional control valve. Sample Flow (Vccm) displays the measured sample flow Sheath Flow (Vccm) displays the measured sheath flow Total Flow (Vccm) displays Sheath Flow (Vccm) + Sample Flow (Vccm). Flow Ratio displays Sheath Flow (Vccm) / Sample Flow (Vccm). The recommended flow range is from 200-1000 volume CC/min (Vccm). A

manual valve inside the instrument controls the split between the sample and sheath flows. The recommend flow ratio is 10. Liquid Supply Pump switches the supply pump ON/OFF. pumps run continuously. The two drain Sample Press (mb) displays sample pressure. Air Pump switches the air pump ON/OFF. To prevent fogging of the OPC, if T OPC < T3 Read the air pump switch will be overridden and the air pump will be turned off. Liquid Flow Set controls the frequency of the solenoid pumps used in the water supply and drains: Low, Medium, and High. Low is good for normal operation. This supplies sufficient water to the column for ambient CCN measurement when the sample temperatures are below 33. Medium is required for environments where the temperature is above 33 C, or the sample air to the CCN is dried with a diffusion drier. This supplies the extra water used by the Nafion humidifier and the column at higher temperatures. High is useful if water supply has run dry or if it is necessary to wet the column quickly. Liquid Supply Pump switches the solenoid pumps ON/OFF

1.1.1.5 Optical Particle Counter Tab (OPC) Figure 5. Optical Particle Counter The OPC tab contains controls and indicators for the operation of the OPC. The OPC measures the sizes of particles at the exit of the column. The particles are sized into 20 bins, and the top of the bin sizes at 0.75, 1.0, 1.5, 2.0, 2.5. in 0.5 micron increments to the largest bin at 10 microns. Bin Set #1-5 displays bin number settings for each SS #. These values are loaded when the SS Table is selected and cannot be changed. Particle counts in bins below the Bin Set value will not be included in CCN Number Conc. calculations. Particle Size displays the particle size that corresponds to the Bin Set value. Manual Control activates the slider bar that allows setting a manual Bin #. CCN Number Conc. displays the number concentration using the counts from particle Bins above the Particle Size and any overflow counts. Overflow displays the number of particles counted that are larger than 10 microns. Baseline Mon displays a voltage used by DMT for OPC diagnostic purposes. Laser Curr displays the laser current (ma) used in the OPC. 1 st Stage Mon displays 10 X the voltage off the first stage amplifier of the OPC. The normal operating range for this voltage is 0.25-0.30 volts. If the OPC is fogged, this voltage may be as high as 5 volts. If the OPC is fogged, turn up the temperature on the OPC to 5 deg above T3, and watch for a drop in the 1 st Stage Mon level. If the value drops as the OPC heater

cycles on, it is an indication that the OPC is not heavily fogged and it usually will recover in an hour or two. To speed the drying of the OPC, turn the cycling of the OPC drying pump to high speed. 1.1.1.6 Chart Properties (Chart) Figure 6. Screen for the chart time history length. The Charts tab contains controls for configuring the Flows, Temperature, and CCN charts. Chart History Length controls the how far back the charts extend. The blue slider selects the history length and the green indicator shows the length of data available. The default value is four hours. Show Legends controls the display of chart legends on top of the chart.

1.1.1.7 Alarm Monitor (Monitor) Figure 7. Alarm monitor The Monitor tab has indicators for monitoring and logging the status of systems. The status of the instrument is categorized as Normal, Alert, or Alarm. An Alert is triggered by a possible problem and takes no actions in the instrument. An Alarm is triggered by a probable problem and will put the CCN in safe mode. Safe mode shuts off the TEC relay and Liquid Supply Pump. The timers count up whenever the specified alert/alarm conditions are being met and reset to 00:00 if the conditions are not met. If any timer exceeds its set threshold, an alert/alarm is triggered and the indicator light turns red. Each alert is associated with a unique (2^n, where n=0,1,2 14) alert code. The sum of all triggered alerts is saved in the last column of the data file. This sum is then made negative if any alarm(s) have been triggered. Laser Current Status alerts if the laser current ever drops below 50ma. No alarm. (Alert Code = 1) -> (1) 1st Stage Monitor Voltage Status counts when the 1 st Stage monitor is >4.7V. Alerts after 5 minutes. No alarm. (2) Flow Status counts when total flow deviates >20% from flow set or when flow ratio is not within 5-15. Alerts after 30 minutes. No alarm (4) Temp Control Status counts when at least one of T1, T2, T3, T Inlet, T Nafion, and T OPC is more than 10 degrees from its set point. Alerts after 10 minutes. Alarms after 20 minutes. (8) Sample Temp Status counts when the sample temperature beyond the range of 10-50 degrees. Alerts after 1 minute. Alarms after 5 minutes (16)

OPC Status counts when the 1 st Stage Monitor > 4.7V and CCN counts are less than the configured threshold (factory default 10 counts/sec). No Alert. Alarms after 60 minutes. (32) CCN Concentration Status counts when CCN counts are less than the configured threshold (factory default 10 counts/sec). Alerts after 20 minutes. Alarms after 120 minutes. (64) Column Temps Stabilized Status counts when T1, T2, or T3 is >0.3 degrees from its set point. Alerts after 20 minutes. (128) e.g. Alarm code = 183 = 128 + 64 + 1 =Column Temp Stabilized + CCN Concentration + Laser Current Status Safe Mode flashes when an alarm has been triggered and the unit is in safe mode. Reset Status clears all alerts/alarms and resets their timers to zero. This button will not turn the TEC relays back on. Next to the program tabs are the Status window and indicator. The status window displays any alarms or alerts concerning the operation of the software or the instrument. During startup the window shows the time available to select a supersaturation table. Normal instrument operation displays a green status indicator. The indicator turns yellow when an alert has been triggered. The indicator blinks red when an alarm has been triggered and the instrument goes into safe mode. (1.1.1.7 Alarm Monitor). 1.1.1.8 Serial Stream A serial output stream of data is available for sending to a host computer or other device. A serial null modem cable must be used and is provided with the start up kit. The control for this the Serial Output button found on the Program Control tab (1.1.1.2 ). The data appears on the DB-9 connector labeled Data Out and uses the RS232 configurations with 9600 baud and 8-N-1. Connecting this port to another computer requires the use of a null modem between the computers so that pins 2 and 3 are swapped. If the Serial Output button is turned off, no data will be sent to the serial port. If the Serial Stream Out button is on, the following variables will be sent out on two lines at 1-second intervals, in a comma-delimited format. H, Time, SS setting, Temp Stable?, T1 read, T2 read, T3 read, Sample Temp, T inlet, T OPC, T Nafion, Sample flow, Sheath flow, Abs Press, Laser current, 1 ST Stage Monitor voltage, Temperature Gradient, proportional valve Voltage, Alarm Code C, Bin setting, CCN Number Conc., ADC overflow, bin 1, bin 2, bin 3, bin 4,

bin 5, bin 6, bin 7, bin 8, bin 9, bin 10, bin 11, bin 12, bin 13, bin 14, bin 15, bin 16, bin 17, bin 18, bin 19, bin 20 Note: The user can also change the temperature via the serial port (labeled DATA OUT ). To change the value, five characters must be inputted (example: 10.00 or 10.0 and a carriage return) To see what SS %, the temperature gradient equals to, press the Sup. Sat. Setting tab and look in the lower right hand corner for the display box labeled Current SS. 1.1.1.9 File Saving The CCN instrument provides on board file storage that can be retrieved by Remote Desktop or over an Internet connection. On the Program Control tab there is a setting New File Every (min) (see Figure 2) that controls when a new file is created. If the setting is 20, the program will save a new file every 20 minutes using the file convention described below. The files are named CCN data yymmddhhmmss.csv and are created in the C:\CCN data directory. The CC data directory is not automatically created by the software and must be created to store the data. yy represents the two digits for year MM the two digits for month dd the two digits for day hh for hours mm for minutes, and ss for seconds Example: May 6, 2004 at 4:18:27 would be 040506041827. This convention will list files created sequentially from start time and provides an easy way to locate specific files. The file contains the following variables: Time, SS Setting., TempStab?, Delta T Set, T1 Set, T1 Read, T2 Set, T2 Read, T3 Set, T3 Read, Nafion Set, T Nafion, Inlet Set, T Inlet, OPC set, T OPC, Sample Temp, Sample flow, Sheath flow, Abs. Press, Laser Current (ma), ADC overflow, Baseline Monitor, 1 st Stage Monitor, Bin Setting, Bin 1, Bin 2, Bin 3, Bin 4, Bin 5, Bin 6, Bin 7, Bin 8, Bin 9, Bin 10, Bin 11, Bin 12, Bin 13, Bin 14, Bin 15, Bin 16, Bin 17, Bin 18, Bin 19, Bin 20, CCN Number Conc., Alarm Code

1.1.2 SS settings.exe Figure 8. Table for setting the suopersaturation program parameter The program SS settings.exe allows the user to setup up to four predefined tables for operating the CCN in a scanning mode with variable supersaturation and time periods. When configuring supersaturation tables, it is recommended that the progression is from lower supersaturations to higher ones. The CCN instrument performs best with gradual increases in column temperature followed by a major cooling back to a low supersaturation. Tables 1 through 4, allow preconfigured settings to be loaded into the operating program. These setting include the supersaturation, time for the supersaturation, the bin point to be used in the CCN summation, sample flow set, and the temperature differentials to be used for the components outside the column. (1.1.1.5 OPC Tab)Once the table has been configured, click on SAVE to save this table, and move on to the next desired configuration. The tables have the option for four supersaturation settings. Following the configuration of all the desired tables, EXIT the program. 1.2 Network Connection (Remote Desktop) The CCN instrument can be connected to over the Ethernet port located with the interface connections on the instrument. The instrument can be connected to a single computer with a network crossover cable or to a local Internet.

Remote desktop is a program that can be run on a host machine that will connect to the CCN instrument over the Internet. Remote Desktop must first be installed on the host machine (available online). In order to connect to the CCN instrument run Remote Desktop from the host machine. In the computer selection type CCNStack or the IP address of the CCN instrument, then click Connect. The password for the CCN computer is password. Once Remote Desktop connects to the instrument, the user will be able to view and have complete access to the active desktop on the CCN computer from the host machine. The program controlling the CCN instrument will then be accessible from the host machine. Files can be transferred from the CCN onboard computer to the Host machine using the right click copy and paste features. Navigate to the CCN data folder and select the files to transfer and then copy. Then minimize the remote desktop function and navigate to the desired file location and paste the files. After right clicking to paste the files it may take up to a minute for the program to respond. Wait until the right click context box pops up in order to paste the files. In order to change flow, pressure, and temperature calibrations open the file press cal.exe. The program will allow the user to set the slope and intercept points for the sample flow, sheath flow, and the pressure. VNC is not supported by this version of windows XP at this time. It is the responsibility of the user to set up the Remote Desktop of the CCN. The following book is recommended for users needing assistance with networking to the CCN: Microsoft Windows XP Networking Inside Out: by Curt Simmons, published by Microsoft Press. )* # + )/,# *# +,

1.3 Press CAL.exe The flow measurements on the CCN for the sample and sheaths flows utilize a laminar flow element and a differential pressure transducer for each flow. In addition an absolute pressure transducer is used for the sample pressure measurement. These systems are calibrated before the unit leaves Droplet Measurement Technologies, but it will be necessary to recalibrate these sensors periodically. The calibration coefficients for each of these measurements are stored in a table that is accessed by the program press cal.exe (see Figure 12). To calibrate the transducers, set the slope value to 1 for each of these parameters and the intercept to 0. In this way the values on the flow tab screen will readout in volts. A new set of calibration coefficients can then be derived and entered into the program. Figure 9. Screen for setting calibration coefficients. #"%#*(* #" " (*" )-)**".%# / %$* #"%#*(* #" " (*" )-)**".%# / %$* *# %001*#2#*% #3(*" ###2

" #*##"#2# ' # "( *#2&1# *#!4 # 2# )*% #3(*" "" % # 5&4* *# * # 3!4! " #$ % "&'!! ()"!! #"%#* #2# #*.* 001 ) #2# 6100 ((7*# #"%#* # 8) #2 #$# / " ##"%#* #2# #*.* 001) #2# 6100 ((7 * ##"%#* #8) #2#$# / $% & & %"% #***".#"( %"9 8 "# ' $% * 2(#:(( (%) * " " ( #" ;#". " ;"#2 )*.#" " 2.# #"$&!. #*" " (%) '* * " ( #" ;#". " ;"#2 )*.#" " 2.# #"$!!. #*"

1.4 CCN Playback.exe The CCN Playback.EXE program allows you to playback data files from the CCN. It displays them graphically very similar to how they are displayed when the CCN is running. Figure 10. CCN Data Playback Read a File opens a file explorer for selecting the data file to be read Playback starts/stops data playback from current cursor position. Playback Speed is controlled with a knob under the Program Tab Selectable Chart selects data for bottom plot w/ default to temperature values.

Figure 11. CCN Data Playback

"# "# 2(#"$ " # ##("# #"#. "#" ##("+) #0#%"3, 0" 2* ## "#$<((=<(("2.$* 7 *##"#*( ##("#%""2* "#$. (" "" *>?:?''* 2?:?''0* #""###23 2# 7 (2#(" "#$%#2# *#"#%"*# * "