SRC-1 Soil Respiration System. Operators Manual for CIRAS-2. Version 4.01

Similar documents
SRC-1 / CPY-2 Closed System Chambers. For Use With All EGM s (1/2/3/4) And CIRAS-1. Operator s Manual. Version 3.31

GasSense NDIR User Manual

Rigel 601 CHECKBOX. Instruction Manual. 348A551 Issue 2.0. April Seaward Electronic Ltd. Issue 2.0

Sensoric 4-20 ma Transmitter Board Operation Manual

Mass Flow Controller (MFC) for Gases

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Instruction Manual. BZ7002 Calibration Software BE

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

RAM 4021 Operation Manual

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Technical Data Sheet MF010-O-LC

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Overview. Front Panel: Keypad and Display

Mass Flow Controller (MFC) for Gases

CSA Sample Draw Aspirator Adapter Operator s Manual

T i m i n g S y s t e m s. RACEAMERICA, Inc. P.O. Box 3469 Santa Clara, CA (408)

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

CIRAS-2 Portable Photosynthesis System Operator s Manual. Version 2.04

USER'S GUIDE. Vaisala CARBOCAP Carbon Dioxide Calibrator for the GMP220 series probes GM K220 U421EN-1.1

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100

RAM 4021-DPX Operation Manual

Model PDT Dewpoint Transmitter

A4s Operation Manual

FTC130 Transmitter. Operating Manual

Instruction Manual No. 742, 8/98

RAM Operation Manual

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

BI-680 Online Dissolved Oxygen Controller Instruction Manual

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

AMT-Ex Dewpoint Transmitter

with O 2 Controller Instruction Manual

Instructions for Use

Cover Page for Lab Report Group Portion. Pump Performance

Gas Sampling Fitting with MK2.1 and L-Probe

Columbus Instruments

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

HI 2314 HI 2315 HI 23151

Transmitter CS 21 Operation Manual

A4 Operation Manual. Fig.1-1 Controller Socket Diagram

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Mass Flow Meter (MFM) for Gases

RGC-IR Remote Gas Calibrator for IR400

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

Met One E-BAM Particulate Monitor

WHEATLEY Series 500 Swing Check Valve

AutoChanger Installation & User Guide Issue 2

The University of Hong Kong Department of Physics Experimental Physics Laboratory

Mass Flow Controller (MFC) for Gases

KERN EG/EW Version /02

VOLUME BOOSTER RELAYS YT-300 / 305 SERIES

MP15 Jockey Pump Controller

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

29 Pressure, Temperature relationship of a gas

Instruction Manual. Model DSP-Ex Portable Dewpoint Meter. Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England

The Univentor 1250 Anaesthesia Unit

Instruction Manual 742 5/1/2009. Eclipse Ratio Regulators ES Series Version 1

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

Instruction Manual. CG16K Barometrically Compensated Capsule Dial Gauge. CG16K Capsule Dial Gauge, 0 to 25 mbar

The Discussion of this exercise covers the following points:

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

FAULT CODE TROUBLESHOOTING INDEX

EC214 EC215 - EC215R Bench Conductivity Meters

SPECIFICATIONS ATTENTION

RK-IR Sample Draw Aspirator Adapter Operator s Manual

GILMONT ACCUCAL FLOWMETERS

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

ECL Comfort 110, application 131 (valid as of software version 2.00)

CENTER PIVOT EVALUATION AND DESIGN

RM-80 respiration monitor

Operating Instructions Temperature Measuring System MF420-5T-100

Instruction Manual Dräger MSI P7 and MSI P7 plus

PH01 Perfusion Cannula Manual

OC Panel High Limit Aquastat Kit, Manual Reset p/n

STARLOG. Capacitive Water Depth Probe

Constant Pressure Inlet (CCN) Operator Manual

Sweeper Dual Control. Automatic Sweeper Switching System SW SERIES SWEEPERS SW-4815 SW-4820 SW-4825 MAGUIRE PRODUCTS INC.

Installation, Operation, and Maintenance Manual

O2 Portable. User Manual

Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

DMM600 DUFF MOISTURE METER INSTRUCTION MANUAL

Operating instructions for. For visualization of your measurements

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

ACV-10 Automatic Control Valve

O3 3E 1 F Gas Sensor Module

combustion efficiency monitoring General Purpose Applications Flue gas oxygen analyzers Combustion & Environmental Monitoring An Company

Mass Flow Meter (MFM) for gases

Roller AC Servo System

Manual of SF6 Comprehensive Tester

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

Instruction Manual Contact Pressure Vacuum Gauge

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic

Transcription:

SRC-1 Soil Respiration System Operators Manual for CIRAS-2 Version 4.01 2002 PP Systems. All Rights Reserved 9th October 2002 Europe North America PP Systems PP Systems Inc. Unit 2, Glovers Court, Bury Mead Road. 241 Winter Street Hitchin, Hertfordshire, SG5 1RT, U.K. Haverhill, MA 01830 U.S.A. Tel: +44 1 462 453 411 Tel: +1 978 374-1064 Fax: +44 1 462 431 090 Fax: +1 978 374-0972 Email: support@ppsystems.com Web Site: http://www.ppsystems.com

This page is left intentionally blank.

Table of Contents Table of Contents Table of Contents... 1 Index of Figures... 2 Preface... 3 Document Conventions... 3 Other Documentation... 3 User Registration... 3 Notice... 4 Service... 5 Technical Specifications of the SRC... 6 Fan... 6 Housing... 6 Dimensions... 6 Weight... 6 Installation... 7 CIRAS-2 Connections... 7 Ciras Pumps... 8 General Advice... 8 Using a Water Vapour Equilibrator... 8 Removal of the SRC-1/STP-1... 8 Operation... 9 Warning... 9 System Parameters... 9 Volume of the System... 9 Area of soil exposed... 9 Maximum Time... 9 Maximum CO 2... 9 Chamber Temperature... 9 Note... 9 Windows Software... 10 Windows software... 10 Results... 11 Calculations... 12 Theory... 12 Measurement Units... 13 Water Vapour, Temperature and Pressure Corrections... 14 Maintenance... 15 Stirring Fan... 15 Appendix... 17 References... 17 SRC-1 Interface Connections... 18 User Notes... 19-1 -

Table of Contents Index of Figures Figure 1 SRC and STP Electrical and Gas Connections to CIRAS-2... 7 Figure 2 SRC connected to Water Vapour Equilibrator... 8 Figure 3 Simulated Measurements... 12 Figure 4 Replacing SRC Fan... 15 Figure 5 Remove Fan retaining Screw... 15 Figure 6 Remove and Replace Fan... 16-2 -

Preface Preface Document Conventions If viewed electronically, text marked blue acts as Hyperlinks. Some links refer to links to external files, which may cause the viewer to prompt whether the action should be performed. Other Documentation Refer to the CIRAS Operators Manual (Ciras2Operation.pdf) for Analyser Maintenance and operation of the Windows Software. Soil Respiration Measurements are also covered in the Closed System Help Files. (See Help\ClosedSys.hlp). User Registration It is very important that ALL new customers register themselves with us to ensure that our user s list is kept up to date. If you are a PP Systems user, please register yourself electronically on our web site at: http://www.ppsystems.com/register.html Only REGISTERED users will be allowed access to our protected Users section of our web site. This section will contain important product information including hardware/software updates, application notes, newsletters, etc. Thank you in advance for your co-operation. Please visit our web site for periodic updates on Technical Documentation and Software. http://www.ppsystems.com - 3 -

Notice Notice This instrument must not be used in situations where its failure could result in injury or death. For applications where failure of this instrument to function correctly would lead to consequential damage, the analyser must be checked for correct operation and calibration at intervals appropriate to the criticality of the situation. This manual is provided to help you install and operate the equipment. Every effort has been made to ensure that the information contained in this manual is accurate and complete. PP Systems does not accept any liability for losses or damages resulting from the use of this information. PP Systems' equipment warranty is limited to replacement of defective components, and does not cover injury to persons or property or other consequential damage. This manual, and the information contained in it, is copyright to PP Systems. No part of the manual may be copied, stored, transmitted or reproduced in any way or by any means including, but not limited to, photocopying, photography, magnetic or other mechanical or electrical means, without the prior written consent of PP Systems. - 4 -

Service Service Most equipment is covered under warranty for one complete year, parts and labor included. This, of course, is provided that the equipment is properly installed, operated and maintained in accordance with written instructions (i.e. Operator's Manual). The warranty excludes all defects in equipment caused by incorrect installation, operation or maintenance, misuse, alteration, and/or accident. If for some reason, a fault is covered under warranty, it is the responsibility of the customer to return the goods to PP Systems or an authorised agent for repair or replacement of the defective part(s). For service in Europe and U.S.A., please contact us at: In Europe In North America PP Systems PP Systems Inc. Unit 2, Glovers Court, Bury Mead Road. 241 Winter Street Hitchin, Hertfordshire, SG5 1RT, U.K. Haverhill, MA 01830 U.S.A. Tel: +44 1 462 453 411 Tel: +1 978 374-1064 Fax: +44 1 462 431 090 Fax: +1 978 374-0972 Email: support@ppsystems.com Web Site: http://www.ppsystems.com - 5 -

Technical Specifications of the SRC Technical Specifications of the SRC Fan 12v DC. Housing PVC and stainless steel. Dimensions 150 mm Height x 100 mm Diameter (excluding handle). Weight 900 g. PP Systems is continuously updating its products and reserves the right to amend its specifications without notice. - 6 -

Installation Installation CIRAS-2 Connections Electrical connections should be made before switching on CIRAS-2. The SRC-1 will be supplied with an adapter to connect the 15 pin D connector on the SRC to the probe input on CIRAS. If a soil temperature probe is being used, then a Y connector is required to double up the connection.to the adapter. The pipe, labelled R on the SRC should be connected to the REF inlet on CIRAS-2 and the second pipe should be connected to the REF outlet. After switching on, CIRAS-2 will automatically go through the start up procedure if the instrument is not already warmed up. Other SRC Pipe to REF IN SRC Pipe labelled R to REF OUT Connect (Via Adapter) to Probe Socket. Adapter and Y Piece If using STP (Shown) connect SRC and STP via Y Piece to adapter Figure 1 SRC and STP Electrical and Gas Connections to CIRAS-2-7 -

Installation Ciras Pumps The SRC-1 works as a closed system, using the reference side of the Ciras analyser connected in-line. The Ciras reference pump is not sealed as standard (not necessary in an open system). However, when Ciras is used as a closed system then the pump must be sealed. If an SRC is supplied with a new Ciras then a sealed pump is fitted as standard, leaving the original unsealed pump as a spare. If an SRC is supplied separately, it is supplied with a sealed pump which the user must fit. General Advice The CIRAS-2 must never be used in the horizontal position as there is the possibility of air tracking over the top of the Zero column(s) rather than passing through the absorber. This will give rise to an incorrect zero and an apparent loss in the analyser sensitivity. Do not calibrate with the SRC-1 connected. Using a Water Vapour Equilibrator To protect the analyser from excessive water vapour, an equilibrator may be in-lined with the SRC as shown. Unmarked SRC Pipe. Connect to Equilibrator Water Vapour Equilibrator SRC Pipe marked R Treat as usual Pipe from Water Vapour Equilibrator. Treat as unmarked SRC Pipe Figure 2 SRC connected to Water Vapour Equilibrator Removal of the SRC-1/STP-1 Switch off the analyser then carefully remove the electrical and gas connections. When the instrument is switched on again it will function as described in the Operator's Manual. - 8 -

Operation Operation Warning It is very important that the analyser should not be cooler than its surroundings when the respiration measurements are being made, otherwise, there is a risk of condensation particularly if the soil is wet. Keep a check on the pipelines to be sure that condensate is not coalescing and entering the analyser. If working on very wet soils then it is recommended that a PP Systems Water Vapour equilibrator is placed in the REF line before CIRAS-2. The water vapour equilibrator which is constructed from a material that is very permeable to water vapour but impervious to CO 2. If this is put in the pipe to the analyser then the water vapour concentration of the analysis air will be brought close to ambient. The problem may also be avoided by putting a suitable drier in the airline to the analyser. Drierite (anhydrous Calcium Sulphate) is suitable. Do NOT use Silica Gel, which also absorbs CO 2 System Parameters For the Closed System PC software the following parameters are user changeable. (selection of the SRC will automatically set default values shown). Volume of the System 1170 cm 3. Area of soil exposed 78 cm 2. Maximum Time Time for which the change in the chamber CO 2 is monitored (120 seconds). Maximum CO 2 The maximum change allowed in the CO 2 concentration from the start of measurement (60ppm). Chamber Temperature Approximate chamber temperature should be entered and the units that are to be used for the results. (µmol.co 2.m -2. s -1 or g CO 2 m -2..h -1 ) Note Readings are taken every 3.2 seconds by the PC and are fitted to a quadratic equation. A minimum of three readings is required to get the first fit. Try first with the default settings. If the record is automatically taken within the first 24 seconds, then Maximum Time or Maximum CO 2 is too low, and the results will be meaningless. - 9 -

Windows Software Windows Software Windows software Start up the CIRAS-2 program as usual and from the main display select SETTINGS then from the SETTINGS EDITOR select the PARAMETERS button. This will now show the PARAMETERS EDITOR. Select under PROBETYPE on the drop down menu, CLOSED SYSTEM CHAMBER then check or select SRC from CHAMBER CUVETTE TYPE box. Press OK, then a Dialog box appears when you should select to save the changes. When the SETTINGS EDITOR is again displayed you should press YES. Then the SELECT USER box is displayed so again press YES. The next display is the CLOSED SYSTEM display. From this, SETUP allows setting of the System parameters listed earlier as well as setting up the recording option and record file name. The message at the box at the top of the display is :- REMOVE FROM SOIL. PRESS START TO BEGIN. After pressing START then the message is:- HOLD CHAMBER IN AIR TO FLUSH (+ display of the elapsed time) Now, the fan motor in the respiration chamber will run at maximum speed. During this operation, the chamber should be held well clear of the soil, open to the air, so that it is thoroughly flushed out.. On completion the message is :- PLACE ON SOIL READY to continue. The area of soil for analysis should be level, without any obviously projecting stones. Push the chamber firmly down onto the soil so that the stainless steel perimeter ring is partially embedded to give a seal. Next, press the READY button. There is then a delay of approximately 20 seconds after pressing before any measurements begin. This is to allow equilibration of the system. This is followed by a further delay to allow 3 measurements to be collected and the first quadratic fit applied. During this period the message is :- EQUILIBRATING and the display of the elapsed time. The graph shows the progress of the measurement. Only 3 values are plotted, the CO 2 reference the change in CO 2 (dc) and the soil respiration (Sr). The current values are only plotted in the measurement cycle. At other times the values are set to 0 and the display will be a straight line. By clicking the <SETUP> button and clicking on the <Graphing Options> you may change the limits for the values. On completion of the measurement the Results Box appears with the option to record the data.. If at this point you elect to save the data then all subsequent measurements will be recorded to the selected file. - 10 -

Windows Software To take more measurements with the current settings then begin the Measurement cycle again or click<setup> to modify the parameters. A measurement may be aborted by clicking the <QUIT> button. To start a new set of measurements to a new file then click <SETUP> and then<new MEASUREMENT>. A different treatment may be selected by clicking<t> next to the <SETUP> button. Results Both dc and Sr are determined from the quadratic fitting. The first 3 sets of readings will show zero for these parameters, as a quadratic fit requires at least 3 data sets. The rate of increase in CO 2 should be linear, though any leakage to the outside air will cause it to decline with time. A quadratic equation is fitted to the relationship between the increasing CO 2 concentration and elapsed time. The coefficients of this are used to determine the rate of increase at time 0. A warning is given if the relationship is excessively non-linear. Measurements terminate when either the chosen elapsed time or the CO 2 concentration changes are exceeded. The PC records have headers that will be displayed when the file is imported into Microsoft Excel. Further details on Closed System Measurements are to be found in the Operators Manual and the Software Help Files. - 11 -

Calculations - Theory Calculations Theory The soil respiration is measured by placing a closed chamber on the soil and measuring the rate of increase of the CO 2. concentration inside the chamber. Then, assuming a well mixed sealed system :- ( Cn Co) V R * Tn A = (1) Where:- R is the soil respiration rate (Flux of CO 2 /unit area/unit time) Co is the CO 2 concentration at T=0 Cn is the concentration at a time Tn later, A is the area of soil exposed and V the total system volume. It has been suggested that to make accurate measurements of the respiration it is essential to start with a CO 2 concentration in the cuvette below ambient and measure until the concentration is above ambient, presumably with the intention of getting some compensation for leakage. However, this leakage can only take place at ground level, where the CO 2 concentration is unknown and most certainly will not be what we would consider as ambient. Over the short period of measurement and with the relatively small CO 2 concentrations in the cuvette compared with the soil concentrations, we would expect the soil respiration to be a constant flux, giving a constant rate of change in the cuvette CO 2 concentration. Any leakage should be a function of the concentration difference between the cuvette and the exchange air. Due to leakage, the apparent soil respiration rate decreases with time. Cn Cuvette Concentration Co No Leakage With Leakage T Time T Figure 3 Simulated Measurements With CIRAS, a quadratic relationship (y = a + bx + cx 2 ) is fitted between the cuvette concentration (C=y) and time (T=x) from the start of measurement. N.B. There is a delay after the chamber is first placed on the soil to allow for the establishment of stable gradients etc, before the measurements start. Then C= a + bt + ct 2 The true soil respiration rate will be calculated from dc/dt at T=0. - 12 -

Calculations - Measurement Units Now dc/dt = b + ct so at T=0 dc/dt = b (2) A comparison of b and ct gives an indication of the leakage. With the PP Systems program ct must be less than 20% of b or the non-linear error message is given. This is believed to be a better approach than lowering the CO 2 value at the start of the measurement. Measurement Units From (1) and(2):- R = b * V/A To give the flux in mass/unit area/unit time then b must be measured in mass/unit volume. The CO 2 analyser measures the volume/volume ratio (= ppm by volume = micro-litre/litre = micro-bar/bar = micro-mol/mol). Now one kg mol of gas (44.01kg of CO2) at STP occupies 22.41 m 3. Thus :- P 273 40.01 V R b * * * * 1000 273 + Ta 22.41 A = (3) Where :- R is the CO 2 flux in kg.m -2.s -1 V is the system volume (largely the soil respiration chamber volume) in m 3 A is the soil area exposed in the chamber in m 2, P is the atmospheric pressure in mb Tc is the temperature of the system volume (Soil Respiration Chamber) in ºC. We measure DC/DT = b as ppm/second, V in cm 3 and A in cm 2. R is expressed as g.m -2.hr -1. To convert (g.m -2.hr -1 ) to (micro-mol.m -2.s -1 ) multiply by 6.312. See References. - 13 -

Water Vapour, Temperature and Pressure Corrections Water Vapour, Temperature and Pressure Corrections CIRAS readings are pressure compensated and the analyser is thermostatted. Therefore a standard gas should read the correct ppm value independent of atmospheric pressure and temperature changes. The CO 2 concentration change is corrected for both the cross sensitivity of the CO 2 IRGA to water vapour and also for the dilution effects of any increase in the water vapour concentration. Since the ambient temperature is not measured, this is entered. - 14 -

Maintenance Maintenance Wipe around the rim of the chamber after each measurement. On completion of the day s measurements, leave the analyser running on ambient air for a period to ensure that there is no condensation in the unit. Please refer CIRAS Operator's Manual for more details regarding analyser maintenance. Stirring Fan To replace the fan, remove the 3 screws around the top of the chamber and remove the upper section. The fan is now accessible for replacement. New fans are available from PP Systems, or an authorised agent, or your local Micronel agent (Type V369L-0129K-1). When fitting a new fan, ensure that it blows vertically onto the soil surface. Remove screws (3 off) Figure 4 Replacing SRC Fan Remove Fan retaining screw. Note Slot below Screw. When replacing, ensure that the screw is not over tightened to avoid distorting the body of the fan Figure 5 Remove Fan retaining Screw - 15 -

Maintenance Push old Fan out of the cylinder. Fan electrical connector block. Ensure that Fan is rewired Push new Fan into the cylinder, observing Fan direction. Do not push to hard to avoid damaging fan. Figure 6 Remove and Replace Fan - 16 -

Appendix - References Appendix References Parkinson K.J. (1981). An improved method for measuring soil respiration in the field. Journal of Applied Ecology, 18, 221-228. - 17 -

Appendix - SRC-1 Interface Connections SRC-1 Interface Connections 15 Pin D connector with the following pin outs: Pin Number Function 1-6 9k 1 ohm resistor = SRC-1 5 Soil temperature output 6 +5v Supply (Interface circuit) 7 Alarm input used to switch relay selecting fan high & low speeds 8 Supply ground 12 +12v Supply (Fan & probe) 15 Signal ground - 18 -

User Notes User Notes - 19 -

User Notes - 20 -