USE, CARE & MAINTENANCE DIAGNOSTIX TM 972 MERCURIAL BLOOD PRESSURE INSTRUMENT

Similar documents
DIAGNOSTIX TM 922 MERCURIAL BLOOD PRESSURE INSTRUMENT. Use, Care, & Maintenance

Aneroid Sphygmomanometer. Use, Care, & Maintenance

Diagnostix TM Aneroids

ADC. Sphygmomanometer Accessories. Directions for Use, Care, & Maintenance AMERICAN DIAGNOSTIC CORPORATION

General Warnings! A warning statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately could lead

INSTRUCTION MANUAL MANUAL INFLATION BLOOD PRESSURE MONITOR

INSTRUCTIONS. Manual Home Blood Pressure Monitor MODEL Stethoscope Chestpiece. Cuff. Bulb

Wallmax Professional Aneroid Sphygmomanometer Model 222 User Manual

HOME BLOOD PRESSURE KIT INSTRUCTIONS TRILINGUAL INSTRUCTION GUIDE MODEL UA-100

Semi-Automatic Blood Pressure Monitor with Memory

INSTRUCTION MANUAL MANUAL INFLATION BLOOD PRESSURE MONITOR

STANDARD OPERATING PROCEDURE ARTSANA MERCURY SPHYGMOMANOMETER

Model ASSEMBLY and OPERATING INSTRUCTIONS

Introduction. Table of Contents. Automatic Wrist Blood Pressure Monitor With Voice-Guided Operation. Model No.: BP5K

J Air and Water Kit Instructions Part# 02584

Model AUTOMATIC UPPER ARM Blood Pressure Monitor

5 series Blood Pressure Monitor

Manual. FRESH AERO OneOP Model PC (Portable Compressors) Tire Inflator. Where Imagination Meets Innovation

FRESH AERO OneOP Model RC (Reservoir/tank) Tire Inflating System

Noninvasive Blood Pressure Measurement. D. J. McMahon rev cewood

Instruction Manual. Wrist Blood Pressure Monitor with Advanced Positioning Sensor (APS ) Model HEM-650 ESPAÑOL ENGLISH

3.1.3 Weight. Frequency. Weight is obtained at the baseline examination and annually. Equipment

D Series Air and Water Kit Part# 02550

RADAR. User Manual EASYCLIK MODEL Vascular. Compression. Device.

TECHNICAL CHARACTERISTIC CT 071

BLOOD PRESSURE SENSOR 0377i

CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE 3-4 BATTERY INSTALLATION

G7S Hand Pump Owner s Manual

Portable Compressed CO 2 Regulator

4-in-1 Professional Inflator Kit

Auto Arm Blood Pressure Monitor

VER: A00. Model No:BA2010

Digital Ear Thermometer

Instruction Manual LIMITED 1 YEAR WARRANTY. Hydraulic Punch Driver Read this material before using this product.

INSTRUCTION MANUAL. OMRON IntelliSense Blood Pressure Monitor. ModelHEM-907

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

Windy City II. User instruction manual WARNING: Fill pressure: 200 bar (approximately 2,900 psi)

92831 TEL: (714) FAX:

Blood Pressure Training System S415

Operating Instruction Manual for Liquid Nitrogen Containers. CRY/CRYC Series. Hagavish st. Israel Tel: , Fax:

BLOOD PRESSURE SENSOR BT17i USER S GUIDE

Sports Wrist Digital Blood Pressure Monitor Item #

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

Digital Blood Pressure Monitor. Model UA-705. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni 1WMPD B

AIR INLINE METAL SHEAR

TactAir PCP Fill Kit/Adapter. Operation Manual

Tactical Sniper air rifle & carbine

CSA Sample Draw Aspirator Adapter Operator s Manual

Please Do Not Return This Product To The Store!

Oxygen Concentrator. Installation and Operation Manual. ClearWater Tech, LLC. Integrated Ozone Systems

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

3 Scotch-Weld Polyurethane Reactive Adhesive Applicator Replacement Part Installation Guide

Vevazz. Operation Manual. Vevazz, LLC Rancho California Road. Suite Temecula, CA Phone: (773) Fax: (888)

92831 TEL: (714) FAX:

Research Grade Blood Pressure Transducer User's Manual

MEDICARE 60. Semi-automatic upper arm blood pressure monitor. Instruction manual

The RICH LIFE Project: Blood Pressure Training Certification Checklists

Blood Pressure Monitor

Monster Windy City Leopard

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

8675/8677/8678 OWNER S MANUAL

Vessel Installation and Operating Manual

INSTRUCTION MANUAL. Automatic Blood Pressure Monitor with Fit Cuff. =Fit Cuff=!"#$% IA1B. Model

6 digital caliper with case

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

Contents. Safety Notice. Wrist Type

Model UB-328. Wrist Digital Blood Pressure Monitor. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

BLOOD PRESSURE SIMULATOR LF01095U INSTRUCTION MANUAL

Parts and Owners Manual

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR

premium wrist blood pressure monitor

MODEL NUMBER: P PORTABLE COMPRESSOR

450P- RV AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only)

trigger magazine release

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

MODEL 100 NITROGEN INFLATION CART

Talking Wrist Style Blood Pressure Monitor Model INS-LAB-RevB08

User Instruction Manual

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

60 farm jack. Distributed exclusively by Harbor Freight Tools Mission Oaks Blvd., Camarillo, CA 93011

DDS CERVICAL TRACTION COLLAR

Digital Blood Pressure Monitor for the Upper Arm INSTRUCTION MANUAL

Owner s Manual & Safety Instructions

Installation Instructions Air Compressor Part # 2780, 2781

MODEL Beretta 92FS. Caliber.177 (4.5 mm) Pellet CO 2 Air Pistol. Operating instructions 3-15

DRAG HARROW SAFETY & OPERATING INSTRUCTIONS. Original Language. Serial No. Order No.

SC505 Sof Care Inflator. Operating Instructions and Service Manual

Instruction Manual. Auto Inflate Blood Pressure Monitor

Non-Invasive Blood Pressure (NIBP)

Blood Pressure Monitor

450P AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters

RULES FOR SAFE FIRING:

IMPACT WRENCH ASSEMBLY AND OPERATING INSTRUCTIONS

MODEL NUMBER: P PORTABLE COMPRESSOR

OIML R 16-1 RECOMMENDATION. Edition 2002 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Non-invasive mechanical sphygmomanometers

Transcription:

USE, CARE & MAINTENANCE DIAGNOSTIX TM 972 MERCURIAL BLOOD PRESSURE INSTRUMENT 1

A Special Thank You... Thank you for choosing an ADC blood pressure instrument. We re proud of the care and quality that goes into the manufacture of each and every sphygmomanometer that bears our name. Only the finest materials are used to assure you of a timeless instrument designed for optimum performance. You ll quickly appreciate the results for you now own the finest sphygmomanometer that money can buy. With proper care and maintenance your ADC blood pressure instrument is sure to provide you with many years of dependable service. Please read the following instructions and general information which will prove helpful in allowing you to enjoy your ADC product. Thank you for your patronage. It is indeed our pleasure to serve you. 2017 AMERICAN DIAGNOSTIC CORP. Sincerely, American Diagnostic Corp. Device Description and Intended Use ADC mercury sphygmomanometers are used by professional healthcare providers and individuals trained in auscultatory blood pressure technique to determine systolic and diastolic blood pressure in humans and animals. 2

Symbol Symbol Definition Important Warning/Caution Symbol Definition Manufacturer s Information Not made with naturalrubber latex Circumference Size F Temperature Limit Humidity Limitation Warnings and Precautions Warning - A warning statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately could lead to patient injury, illness, or death. Caution - Do not allow a blood pressure cuff to remain on patient for more than 10 minutes when inflated above 10 mmhg. This may cause patient distress, disturb blood circulation, and contribute to the injury of peripheral nerves. Warning - If luer lock connectors are used in the construction of tubing, there is a possibility that they might be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a blood vessel. Warning - Safety and effectiveness with neonate cuff sizes 1 through 5 is not established. Warning - Do not twist connectors when joining the blood pressure cuff to the tubing - this may cause the connector to stick and make changing BP cuffs difficult. Warning - When using with an infant or child cuff, extra care must be taken to prevent over-inflation. With smaller cuffs (infant or child) the cuff can inflate to over 300mmHg with just two full compressions of the bulb. To prevent discomfort or injury to the patient and damage to the instrument, bulb should only be partially squeezed, so that each stroke inflates the cuff in 40mmHg to 60mmHg increments until inflated to the desired level. Warning - This product may contain a chemical known to the state of California to cause cancer, birth defects, or other reproductive harm. Caution - Federal law restricts this device to sale by or on the order of a physician or licensed healthcare practitioner. Parts and Assembly This booklet contains operating and maintenance information for the DIAGNOS- TIX 972 series mobile mercurial blood pressure instrument. Please read and retain. 3

Your DIAGNOSTIX 972 consists of main unit with 5mm calibrated unbreakable plastic cartridge tube, inflation system (which includes the ADCUFF nylon cuff with Size Guide marking system, latex-free inflation bladder, bulb, and the AD- FLOW valve), 8 ft. length of coiled tubing, luer connectors, assembly tools, and operating instructions. To assemble inflation system to main unit, connect free end of coiled tubing to air inlet nipple at rear of mercury column above integral cuff storage compartment. Insert male luer adaptor on opposite end of coiled tubing into female luer connector on bladder. Store folded inflation system in cuff storage compartment behind mercury column of main unit. For assembly of the mobile stand, consult booklet entitled Diagnostix TM 972 Assembly Instructions. Operating Instructions Your DIAGNOSTIX 972 Mobile Mercury Sphygmomanometer incorporates a special safety feature that locks the mercury within the reservoir during maintenance. To Operate the DIAGNOSTIX 972: To release the mercury from the reservoir, move the locking lever (located in front of unit below mercury column), left to the position marked ON. Mercury will flow up into the plastic cartridge tube and rest at the 0 mark. If the mercury doesn t rise within the tube, tilt the top of the unit towards you. If the mercury fails to rise, or doesn t reach the 0 level have the unit serviced. To Lock Mercury within the Reservoir: Tilt the top of the unit away from you to permit mercury to flow out of cartridge tube and into reservoir. When cartridge tube is completely emptied of ALL mercury (and while it is still tilted), move locking lever right to the OFF position. PLEASE NOTE: MERCURY SHOULD BE SECURED WITHIN THE RESERVOIR DURING TRANSPORT, OR WHEN SERVICING THE CARTRIDGE TUBE, DI- APHRAGMS, OR FILTERS. NEVER DISASSEMBLE UNIT UNLESS MERCURY IS FIRST LOCKED WITHIN RESERVOIR. CONSULT SEPARATE INSTRUCTIONS FOR ASSEMBLY OF MAIN UNIT (ENCLOSED). 4

MEASUREMENT PROCEDURE 1. Patient Position The patient should sit or lie comfortably. The arm should be fully supported on a flat surface at heart level. (If the arm s position varies, or is not level with the heart, measurement values obtained will not be consistent with the patient s true blood pressure.) Observer should view manometer in a direct line and at eye level to avoid parallax error. 2. Apply the cuff ADCUFF nylon cuffs, with proprietary Size Guide marking system, are specially designed to promote the precise, accurate determination of blood pressure. Index and range markings ensure use of the correct cuff size. The artery mark indicates proper cuff positioning. Select Smaller Cuff THIS SIDE AGAINST LIMB Select Larger Cuff CORRECT Place the cuff over the bare upper arm with artery mark positioned directly over the brachial artery. The bottom edge of the cuff should be positioned approximately one inch above the antecubital fold. Wrap the end of the cuff, not containing the bladder, around arm snugly and smoothly and engage adhesive strips. To verify a correct fit, check that the INDEX line falls between the two RANGE lines. 3. Inflate the cuff Close the ADFLOW valve by turning thumbscrew clockwise. Palpate the radial artery while inflating the cuff. Be sure to inflate cuff quickly by squeezing bulb rapidly. Inflate cuff 20-30mmHg above the point at which the radial pulse disappears. 4. Position the Stethoscope Position the chestpiece in the antecubital space below the cuff, distal to the brachium. Do not place chestpiece underneath the cuff, as this impedes accurate measurement. Use the bell side of a combination stethoscope for clearest detection of the low pitched Korotkoff (pulse) sounds. RANGE RANGE 5

5. Deflate the cuff Open the valve to deflate the cuff gradually at the AHA recommended rate of 2-3mmHg per second. 6. Measurement Record the onset of Korotkoff sounds (Phase 1) as the systolic pressure, and the disappearance of these sounds (Phase 5) diastolic pressure. (Some healthcare professionals prefer recording diastolic 1 and diastolic 2. Diastolic one occurs at phase 4). After measurement is completed, open valve fully to release any remaining air in the cuff. Remove cuff. Care and Maintenance Storage - After measurement, wrap cuff around bulb and store behind main unit in the integral cuff basket. Manometer - Periodically replace the filter plug and diaphragm, and clean the inside of the plastic cartridge tube using alcohol and a lint free pipe brush. Units should be cleaned and serviced periodically depending on frequency of use, or whenever inside of cartridge tube appears dirty. Proper servicing of your mercury unit according to a strict maintenance schedule will assure maximum measurement accuracy and extend the life of your instrument. For directions on disassembling the unit refer to the enclosed schematic diagram and accompanying instructions. PLEASE NOTE: Mercury is toxic. Secure mercury within the reservoir before servicing unit. Always wear protective clothing (gloves and goggles) when cleaning the unit. Clean any spilled mercury promptly using approved mercury containment kits. Consult the enclosed manual entitled Just in Case for procedures on dealing with a mercury spill. 6

Limited Warranty American Diagnostic Corporation (ADC) warrants its products against defects in materials and workmanship under normal use and service as follows: 1. Warranty service extends to the original retail purchaser only and commences with the date of delivery. 2. The entire sphygmomanometer is warranted for one year. 3. The inflation system is warranted for three years. 4. The calibrated plastic cartridge tube is warranted for life an will be replaced FREE OF CHARGE when returned freight prepaid. What Is Covered: Calibration, repair or replacement of parts, and labor. What is Not Covered: Transportation charges to and from ADC. Damages caused by abuse, misuse, accident, or negligence. Incidental, special, or consequential damages. Some states do not allow the exclusion of incidental, special, or consequential damages, so this limitation may not apply to you. To Obtain Warranty Service: Send item(s) postage paid to ADC, Attn: Repair Dept., 55 Commerce Drive, Hauppauge, NY 11788. Please include your name and address, phone no., proof of purchase, a brief note explaining the problem, and $2.00 to cover the cost of return shipping and handling. Implied Warranty: Any implied warranty shall be limited in duration to the terms of this warranty and in no case beyond the original selling price (except where prohibited by law). This warranty gives you specific legal rights and you may have other rights which vary from state to state. 7

To register your product visit us at www.adctoday.com/register FOR QUESTIONS, COMMENTS, OR SUGGESTIONS CALL TOLL FREE: 1-800-ADC-2670 OR VISIT www.adctoday.com/feedback This manual is available online in a variety of languages, follow the links for language options. www.adctoday.com/care ADC 55 Commerce Drive Hauppauge, NY 11788 U.S.A. IB p/n 9357-00 rev 5 Inspected, assembled and packaged in the U.S.A. tel: 631-273-9600 toll free: 1-800-232-2670 fax: 631-273-9659 www.adctoday.com email: info@adctoday.com Printed in the U.S.A. 8