Automated Determination of Dissolved Gases in Water Anne Jurek

Similar documents
Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Rapid and Reliable Detection of Dissolved Gases in Water

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

GCMSD-Headspace Analysis SOP

Retention Time Locking: Concepts and Applications. Application

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Versatility. Accuracy. Compact. Versa. Automated Headspace Vial Sampler

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Agilent 7697A Headspace Sampler

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

NCASI METHOD IM/CAN/WP IMPINGER/CANISTER SOURCE SAMPLING METHOD FOR SELECTED HAPS AND OTHER COMPOUNDS AT WOOD PRODUCTS FACILITIES

Automated Liquid Handling Station

Experiment GC : Analysis of BTEX by GC-FID

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Determination of Dissolved Gases in Ground Waters. By Ray Martrano Laboratory Director Seewald Laboratories Inc.

Standard Test Method for Residual Acrylonitrile Monomer Styrene-Acrylonitrile Copolymers and Nitrile Rubber by Headspace Gas Chromatography 1

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

APPLICATION NOTE. GC Integrated Permeation Device

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Buck Scientific, Inc.

May 1, By John Heim,Doug Staples

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Carrier Gases in Capillary GC

Applied Technology and Best Practices in CEE. Conference

A New GC/MS System Designed for Helium Carrier Gas Conservation

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Determination of Oxygenates in Gasoline by Gas Chromatography and Oxygen Selective Flame Ionization Detector Method

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector

METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1.1 Applicability. This procedure is applicable for

Multiple Gas#5 GC configuration Jan 2016

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

SOP: Buck Scientific BLC-20P HPLC Operation

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Carrier Gases in Capillary GC

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Figure 1: Plug in ion inlet of mass spec

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

This GC has two columns (I think new columns can be ordered from Restek)

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Agilent Inlet EPC Leak Test

* Analysis of TCG ( Total Combustible Gas ) and 6 kinds of Gases * High Sensitivity for C 2

Addressing the World Shortage of Helium

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin

DISSOLVED GAS ANALYSIS OF NATURAL ESTER FLUIDS UNDER ELECTRICAL AND THERMAL STRESS

NOTE: Performance of this method should not be. attempted by persons unfamiliar with the operation of a gas

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

The Tekran Series 2600-IVS Difference!

AKTA pure 25 New Owner s Intro

New improved robot for gas kinetics

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Impact of Air Leaks on the Productivity of GC and GC/MS Systems

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

Agilent G Rebuilt Electron Capture Detector

5890II GC Standard Operating Procedure 9/2/2005

Zebron Gas Management

Setting a New Standard in Flash Chromatography Performance

The Preparation of Low Concentration Hydrogen Sulfide Standards Gulf Coast Conference September 12, 2002 Galveston, TX Paper 050,12:45 pm Moody F

Operation of the Perkin Elmer TGA-GC/MS

Introduction. Experimental

Dissolved Gas Analysis - Past, Present and Future

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs

Your local gas generation partner. Precision series Modular gas generation solution for GC.

The instrument will process the samples one by one till all the samples have been cleaned up.

System Overview. TCD Detector temperature setpoint Regulator

Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

During today's webinar we're going to look at the following topics:

Increasing the productivity of ICP-OES with the Thermo Scientific Qtegra ISDS Software

CHM250 Calibration and Measurement Lab. Balance Calibration

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Bernhard Thomas September EnCal 3000 prochain Energy in Gas with ROI

METHOD: 1024, Issue 2 EVALUATION: FULL Issue 1: 15 August 1987 Issue 2: 15 August 1994

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

CHEM254 #4 The Synthesis of a Tertiary Alcohol Using a Pre-Made Grignard Reagent 1

STANDARD OPERATING PROCEDURES

Your local gas generation partner. Precision series Modular gas generation solution for GC.

FROG-4000 TM Chemical Analysis System

This standard is due for Five Year Review. This process is required by the SEMI Regulations to ensure that the standard is still valid.

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

Transcription:

Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: Tapping the natural gas reservoirs throughout the United States has long been a viable solution for energy independence; however until recently getting to these gas reservoirs was very difficult. Now, through the development of horizontal drilling in conjunction with hydraulic fracturing or fracking, these reservoirs have become much easier to tap for natural gas. However, there are some environmental concerns with the fracking process that have come to light as fracking has gained popularity. One major concern is the potential for natural gas to migrate into drinking water sources. In order to test for dissolved gases, the RSK-175 Standard Operating Procedure (SOP) was developed. Since RSK-175 is an SOP and not an EPA or ASTM method; laboratories have employed different approaches in order to calibrate for the dissolved gases. This study will evaluate a standard calibration using serial dilutions of saturated gas samples. The LGX50 autosampler will do the rest of the work by creating headspace in the sample vial while maintaining sample integrity and transferring the headspace to a GC/FID for analysis. This automation provides the capability to treat calibration samples in the same manner as field samples thus reporting field sample results requires no back calculation using the Henry s Constant. Discussion: The LGX50 was designed to automate the time consuming process of sample preparation for dissolved gas analysis. The automation of this analysis (patent pending) involves using two sample trays, one tray to hold empty 40ml vials with a stir bar and one tray to hold the dissolved gas samples. The LGX50 evacuates the empty vial. Next, a prescribed amount of sample is transferred from the full vial to the empty vial thus creating a headspace that retains the sample integrity by not exposing the sample or the sample pathway to the atmosphere. The system then heats and stirs the sample for a set time. Finally, the sample is pressurized so as to fill a Figure 1: LGX50 Automation Graphic sample loop and transfer the sample to the GC/FID for analysis. See Figure 1. This application note will describe calibration sample preparation and display experimental results showing the reliability and accuracy of the LGX50.

Experimental: The LGX50 Autosampler was configured with a one milliliter loop and interfaced with a GC/FID. The GC column used for this analysis was a Restek RT Q-bond 30m x 0.53mm x 20µm. The LGX50 Autosampler settings and the GC/FID experimental parameters are listed in Tables 1 and 2 respectively. LGX50 Autosampler Parameter Setting Sample Type DGA Sample Fill Mode Loop Sample Volume 20ml Syringe Prime 3 sec. Syringe Needle Rinse 20ml Rinse Cycles On/1 Sample Temperature 60 C Stirrer On/Medium Sample Equilibration Time 10 min. Vial Pressurization Time 5 sec. Loop Fill Time 5 sec. Loop Equilibration Time 5 sec. Valve Temperature 65 C GC Line Temperature 85 C GC Cycle Time 15 min. Rinse Water Temperature 65 C IS 5µl Table 1: LGX50 Autosampler Settings GC/FID Agilent 5890 Inlet Temperature 250 C Inlet Pressure 9psi Gas Helium Inlet Split/Splitless Column Flow 14.33ml/min Column Restek RT Q-bond 30m x 0.53mm x 20µm Oven Program 45 C hold for 1 minute, ramp 16 C/min to 180 C hold for 1.06 min, 10.5 min total runtime FID Temperature 250 C Table 2: GC/FID Parameters High purity gases were ordered from local gas suppliers. Water was purged with the designated gas for at least one hour in order to saturate the water with the gas. The water was kept in an ice water bath and allowed to cool down to 10 C before being purged with the methane, ethane or ethylene gases. The temperature of the water was monitored throughout the purging cycle in order to ensure a constant 10 C temperature. The saturated gas concentrations were determined using the values found in References 1 and 2. Next, using the serial dilutions listed in Tables 3 through 5; the saturated gas curves were made.

Saturated Methane Gas Solution Curve Preparation Standard Amount Final Concentration Saturated Solution 50ml 16.0ppm Saturated Solution 25ml 8.0ppm Saturated Solution 5ml 1.6ppm Saturated Solution 1ml 320.0ppb Saturated Solution 500ul 160.0ppb Saturated Solution 100ul 32.0ppb Saturated Solution 50ul 16.0ppb *Samples Prepared in a 100ml Volumetric Flask Table 3: Methane Dissolved Gas Standard Preparation Saturated Ethane Gas Solution Curve Preparation Standard Amount Final Concentration Saturated Solution 25ml 27.5ppm Saturated Solution 10ml 11.0ppm Saturated Solution 5ml 5.5ppm Saturated Solution 1ml 1.1ppm Saturated Solution 500ul 550.0ppb Saturated Solution 100ul 110.0ppb Saturated Solution 20ul 22.0ppb Saturated Solution 5ul 5.5ppb *Samples Prepared in a 100ml Volumetric Flask Table 4: Ethane Dissolved Gas Standard Preparation Saturated Ethylene Gas Solution Curve Preparation Standard Amount Final Concentration Saturated Solution 25ml 52.5ppm Saturated Solution 10ml 21.0ppm Saturated Solution 5ml 10.5ppm Saturated Solution 1ml 2.1ppm Saturated Solution 200ul 420.0ppb Saturated Solution 50ul 105.0ppb Saturated Solution 10ul 21.0ppb Saturated Solution 2ul 4.2ppb *Samples Prepared in a 100ml Volumetric Flask Table 5: Ethylene Dissolved Gas Standard Preparation Deuterated methyl tert-butyl ether (MtBE-d3) was chosen as the Internal Standard (IS) and added to the IS vessel on the LGX50. The saturated gas standards were poured into 40ml vials with no headspace and loaded on the second tray of the LGX50 while a series of empty vials with stir bars were placed in the first tray. After the vials and IS were loaded, a sample sequence was set up and the LGX50 did all the work. All of the experimental results are listed in Tables 6 through 8. Figures 2 through 4 display a sample chromatogram of each of the gases tested.

Compound Curve Range Individual Gas Results MDL Curve Spike MDL R² Level (ppb) (ppb) Precision Spike Level (ppm) Accuracy (% Recovery) Precision (%RSD) Methane 16ppb to 16ppm 0.999 16.0 6.20 1.6 98.8 5.9 Ethane 5.5ppb to 27.5ppm 0.999 5.5 2.46 5.5 96.7 6.9 Ethylene 4.2ppb to 52.5ppm 1.000 4.2 3.17 10.5 104.9 6.0 Table 6: Saturated Gas Curves Results Compound IS Reproducibility (%RSD) Methane 2.4 Ethane 4.4 Ethylene 5.9 Table 7: Internal Standard Precision Compound High Level Standard Level %Carryover Methane 16.0ppm 0.09 Ethane 27.5ppm 0.05 Ethylene 52.5ppm 0.04 Table 8: Percent Carryover after High Gas Standard Figure 2: 1.6ppm Methane Standard Chromatogram

Figure 3: 5.5ppm Ethane Standard Chromatogram Figure 4: 10.5ppm Ethylene Standard Chromatogram Conclusion: The new LGX50 is both reliable and accurate for performing dissolved gas analysis. The percent carryover was very low and the linearity for each of the saturated gases tested was excellent. The compound precision and accuracy and the IS precision also proved to be very good throughout the study. However, the best part is that the LGX50 is able to automate much of the sample preparation time involved for dissolved gas analysis. Since sample preparation is extremely labor intensive, this system would save both time and money for any lab performing this analysis. For more information visit estanalytical.com or click below: http://www.estanalytical.com/products/environmental/lgx50_for_rsk_analysis

References: 1. Solubility of Gases in Water. Retrieved November 15, 2011, from http://www.engineeringtoolbox.com/gases-solubility-water-d_1148.html 2. Gas Encyclopaedia. (2009). Retrieved November 15, 2011 from http://encyclopedia.airliquide.com/encyclopedia.asp 3. ConocoPhillips Company, Drilling and Completion, Retrieved January 20, 2012, from http://www.powerincooperation.com/en/pages/drilling-and-completion.html 4. Hudson, Felisa, RSKSOP-175, Revision No. 2, May 2004. 5. EPA New England, Technical Guidance for the Natural Attenuation Indicators: Methane, Ethane, and Ethene. Revision 1, July, 2001. 6. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID), PADEP 3686, Rev. 0, April 2012.