Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts

Similar documents
MAGNETIC FIELD INFLUENCES AND YOUR STRATA VALVE. Strata II Valves

Strata Valve ADJUSTMENT REFERENCE GUIDE

URGENT MEDICAL DEVICE RECALL NOTICE

Alaris / Asena GS, GH, CC, TIVA, PK, Enteral Syringe Pump

POLARIS. The first MRI-stable adjustable valve NEW MR SAFE

Urgent Field Safety Notice (Field Safety Corrective Action) for CORAIL AMT Neck Trials Surgical Instruments

progav 2.0 In touch with you

Product Design Specifications

Kimal plc: URGENT FIELD SAFETY NOTICE Field Safety Corrective Action IMMEDIATE ACTION REQUIRED

Injector system for implantation of 1stQ foldable intraocular lenses

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM

UTMB RESPIRATORY CARE SERVICES GUIDELINES Respiratory Care Services Medical Gas Cylinder Storage, Use and Transport

3 Revising Diagnostic Imaging PARs

Department of Neurosurgery, Tuen Mun Hospital, Tuen Mun, Hong Kong

THIS PAGE INTENTIONALLY BLANK

Instructions for Use

prosa Adjustable MIETHKE Shunt Systems Your Simple Choice for Precise Patient Care Aesculap Neurosurgery MIETHKE

Risk Management Series Article 8: Risk Control

CONCUSSION PROTOCOL When in doubt, sit them out!

Name: Oxygen Care Ltd Address: 2 Holfeld Business Park, Kilmacanogue, Co. Wicklow, Ireland URGENT FIELD SAFETY NOTICE - UPDATE

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM

Service Bulletin

Oxygen Dialflow Meter. Instructions for Use

CELEBRATING YEARS of Twin Cities Heart Walk Target Field, May 18 TwinCitiesHeartWalk.org

3T/7T MRI FACILITY. SOP Number: General Safety. Revision Chronology. Associate Director Signature: Date: Version Number Date Changes

Oregon 4-H Bowl Procedures Revised 10/13

Service Bulletin

G92 Series BASOTROL Automatic Pilot Gas Valve

TITAN FLOW CONTROL, INC.

2016 Annual Associate Safety Modules Section 15 Magnetic Resonance Imaging (MRI) Safety

Pressure Relief Valve Instruction Manual

Scald Protection Three-Way Thermostatic Mixing Valve

WE UNDERSTAND. NEUROSURGERY. progav 2.0 IN TOUCH WITH YOU

Pressure Regulators. Operating Instructions. Instrumentation

Powder Measure. Care & Usage Instructions. Instructions # Product # Revision: A

Emergency Action Plans for Dam Safety

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

INSTALLATION & MAINTENANCE INSTRUCTION

Pressure and Flow Characteristics of Ocular Viscoelastic and Silicone Oil Through Glaucoma Drainage Devices

! WARNING. Model PFC-1-G (direct acting) PFC-1-GR (reverse acting) Modulating Pneumatic Liquid Level Controls INSTRUCTION MANUAL MM-110B

FLANGED TWO-PIECE BALL VALVES

Ref: FOI/CAD/ID May Georgie Bevan Dear Ms Bevan. Freedom of Information Act 2000

Minot State University Institutional Review Board ANNUAL UPDATE/REVISION/PROJECT COMPLETION REPORT

IMPORTANT SAFETY RECALL Date: April 11, 2016

Installation Operation Maintenance. Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow IOM.

The following tables provide detailed MRI information for each product. Table # Referenced products MR Safety

URGENT FIELD SAFETY NOTICE: Medical Device Field Safety Corrective Action / Recall

Spasticity / Dystonia

BRAIN AND BANNISTER'S CLINICAL NEUROLOGY (OXFORD MEDICAL PUBLICATIONS) BY ROGER BANNISTER

Pressure Regulator. Instructions for Use

IMPORTANT SAFETY RECALL NOTICE. Voluntary Recall (Field Retrofit) Program for Scotsman Commercial Cubers

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)

National Standard for Commercial Vessels

WHEATLEY Series 500 Swing Check Valve

INSTALLATION, OPERATION AND MAINTENANCE GUIDE

Cylinder Exchange Training Manual

English. ivolve Full Face Mask. User Manual

User Instructions Series Descent Device

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

WHEATLEY Series 822/820 Swing Check Valve

Scald Protection 3-Way Thermostatic Mixing Valve

Traffic Control & Road Closures for Special Events

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM

Family & Consumer Sciences (Home Economics) Bowl

Heart/Lung Perfusion Packs INSTRUCTIONS FOR USE

Healthcare. Ref: Final results of the quench line inspection at your MAGNETOM system. Dear Siemens MR User,

If you have any questions, please call Michelle Tomilloso or Eric Lamonica, between the hours of 7:30 a.m. and 4:00 p.m at (925)

Intracranial Pressure Monitor. Department of Biomedical Engineering University of Wisconsin-Madison BME 301. Final Report

Study Management SM STANDARD OPERATING PROCEDURE FOR Interactions with the Institutional Review Board (IRB)

BD Cytopeia Fluidic Kit User s Guide

Engineering Data Sheet

Alaris Pump module FAQs

Worker Seriously Injured Servicing a Plunger Lift System

CONSTANT FLOW REGULATOR (CFR)

ASSE International Product (Seal) Listing Program. ASSE Performance Requirements for Air Valve and Vent Inflow Preventer

Combination Air Valve

Fetlock Lameness It s importance

North Avenue Rail Station Accessibility Assessment

Jonathan s Landing Golf Club. Old Trail Project

Valve with Integrated Pressure Regulator for medical OXYgen

Via Federal Express Rockville, MD 20850

FLANGED TWO-PIECE BALL VALVES

0311T. General Installation and Maintenance Instructions

Georgia 4-H Horse Quiz Bowl. Procedure for Play. For the 2016 Season

Ambient Weather WS-03 Thermo-Hygrometer

progav 2.0 CHRISTOPH MIETHKE GMBH & CO. KG Instructions for Use USA

URGENT Medical Device Recall

Veterans High School

Pressure Dump Valve Service Kit for Series 2300 Units

2011 Meter Technician Training School Springfield, MO

12V MINI AIR COMPRESSOR

M45 ISO Ball Valve DN25 to 150 Installation and Maintenance Instructions

Pressure Regulators. Instructions for Use

University of Minnesota Center for Magnetic Resonance Research Standard Operating Procedure Human System Magnet Operator Training

Installation Operation Maintenance

G96 Series BASOTROL Dual Operator Valve

EMERGENCY USE 03/02/2016

Model PRS-07i. Stainless Steel Low Pressure, In-Line Pressure Regulator (Reducing Valve)

Transcription:

Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts 125 Cremona Drive Goleta, CA 93117 Catalog Number 42955, STRATAMR VALVE, SMALL E11656, E12556, E12753, E13482, E13483, E13484, E13641, E14518, E14519, E18555, E18556, E18557, E18558, E19670, E19866 Catalog Number 42965, STRATAMR VALVE, REGULAR E11655, E12041, E12265, E12266, E12950, E12951, E12952, E12953, E13485, E13698, E18559, E18560, E19084, E19540, E19671, E19867, E22208, E22242 Catalog Number 46955, STRATAMR SHUNT ASSEMBLY, SMALL E12752, E12932, E16247 Catalog Number 46960, STRATAMR SNAP SHUNT, SMALL E12780, E12934, E14748 Catalog Number 46965, STRATAMR SHUNT ASSEMBLY REGULAR E12779, E12933, E14270 Catalog Number 46970 STRATAMR SNAP SHUNT, REGULAR E12272, E12620, E12935 Date: February 22, 2017 Dear Chief of Neurosurgery and/or Risk Manager: The purpose of this letter is to advise you that Medtronic Neurosurgery (Brain Therapies) is initiating a voluntary recall of all unused units for the Medtronic StrataMR adjustable valves & shunts. You are receiving this letter because you may have received one or more of the affected product(s). Medtronic is initiating this voluntary recall because the products identified above may experience a condition that may, in limited circumstances, affect the flow resistance in the valve and result in underdrainage of cerebrospinal fluid (CSF). The condition can occur when the valve mechanism is adjusted to a position that causes a higher than intended flow resistance. This higher than intended flow resistance is a result of the MRI Resistance Key on the bottom of the valve s rotor mechanism being placed on top of one of the MRI Resistance Walls, instead of in one of the pressure setting wells (see Figures 1 & 2). This condition only occurs during the valve adjustment process. Page 1 of 9

StrataMR Adjustable Valve Figure 1 Example of an Correct rotor orientation Figure 2 Example of an rotor orientation When the StrataMR valve is set in an orientation as seen in Figure 2, pressure flow performance approximately twice the level above StrataMR s highest Performance Level setting of 2.5, can occur and lead to cerebrospinal fluid (CSF) underdrainage. This condition may result in the following adverse health consequences: headaches, nausea, vomiting and lethargy. If left untreated underdrainage has the potential to lead to coma and death. Our investigation determined an increase in the occurrence rate of StrataMR complaints, related to incorrect flow or level change in the month of January 2017. Due to this increase, the complaint rate is 1.45% of total units distributed since November of 2015; 1.1% of total units distributed resulted in revision surgeries. There have been no reports of death. The incorrect valve rotor orientation (Figure 2) can be caused by misalignment and/or off-center positioning of the StrataMR Locator Tool, relative to the valve mechanism. The images that follow are examples that illustrate both correct and incorrect alignment and centering, and are intended to assist clinicians in their efforts to set the valve pressure level setting as intended. Page 2 of 9

Figure 3 provides an example of proper/correct technique. Figures 4, 5, & 6 provide examples of incorrect alignment and centering techniques that should be avoided. Figure 3 Correct Good Alignment and Centering of Locator Too and Valve Mechanism Figure 4 Poor Alignment, but good centering of Locator Tool & Valve Mechanism Figure 5 Good Alignment, but poor centering of Locator Tool & Valve Mechanism Figure 6 Good Alignment, but poor centering of Locator Tool & Valve Mechanism Page 3 of 9

As part of the adjustment process of the StrataMR Adjustable Valve system, it is important to note that if the valve s pressure flow setting is accurately aligned and centered via proper technique as described in the IFU (also see Figure 3 and instructions in Attachment B), the valve MRI Resistance Key will be set to the correct pressure setting well. We ask that the following actions be initiated immediately: 1. Cease use of the listed product. 2. Account for number of product units implanted or discarded and return any unused units by completing the attached individualized customer product accountability form provided in Attachment A. Please follow the instructions described in Attachment A. 3. If any of the indicated products have been implanted in patients, see the valve adjustment instructions in Attachment B for continued patient care. This letter is intended to document this event for your files. We request this letter be provided to all those who need to be aware of this matter within your organization or to any organization where the potentially affected product may have been transferred. We sincerely regret any inconvenience this situation may have caused. Please feel free to contact me should you have any questions or concerns. Fax your Customer Acknowledgement Response Form to Medtronic Neurosurgery at (805) 571-8787. or take a clear picture or scan of signed form and email it to rs.mnsfca@medtronic.com with StrataMR Recall in the subject line. Sincerely, Bob Shokoohi Sr. Quality Operational Excellence Director Page 4 of 9

ATTACHMENT A Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts February 22, 2017 Dear Customer Representative, The purpose of this letter is to advise you that Medtronic Neurosurgery (Brain Therapies) is initiating a voluntary recall of all unused units for the Medtronic StrataMR adjustable valves & shunts. The following products are affected: Catalog Number 42955, STRATAMR VALVE, SMALL E11656, E12556, E12753, E13482, E13483, E13484, E13641, E14518, E14519, E18555, E18556, E18557, E18558, E19670, E19866 Catalog Number 42965, STRATAMR VALVE, REGULAR E11655, E12041, E12265, E12266, E12950, E12951, E12952, E12953, E13485, E13698, E18559, E18560, E19084, E19540, E19671, E19867, E22208, E22242 Catalog Number 46955, STRATAMR SHUNT ASSEMBLY, SMALL E12752, E12932, E16247 Catalog Number 46960, STRATAMR SNAP SHUNT, SMALL E12780, E12934, E14748 Catalog Number 46965, STRATAMR SHUNT ASSEMBLY REGULAR E12779, E12933, E14270 Catalog Number 46970 STRATAMR SNAP SHUNT, REGULAR E12272, E12620, E12935 The products identified above may experience a condition that may, in limited circumstances, affect the flow resistance in the valve and result in underdrainage of cerebrospinal fluid (CSF). Our records indicate that you have received affected product (as shown below). Customer Information Catalog Number Lot Number Quantity Shipped Purchase Order Number (AS APPLICABLE) Page 5 of 9

ATTACHMENT A Please determine the status of the above listed product and complete the table below. The status of all listed product must be provided. If product has been used or discarded, please note the quantities below. Catalog Number Lot Number Quantity Shipped # of Product Available to be Returned # of Product Used and/or Discarded (AS APPLICABLE) If the product is still in inventory, please e-mail or fax this completed form to: Medtronic Neurosurgery Goleta Attention: STRATAMR Recall Email: rs.mnsfca@medtronic.com FAX No. (805) 571-8787 Once the completed form is received, you will receive an RGA with return instructions via e-mail from Customer Care. Once the product has been received, Customer Care will be issuing a credit for the returned product. If you have any questions regarding the return process, please call Customer Care at 1-800-874-5797. The product should be returned to: Medtronic Neurosurgery Goleta RGA# Attention: Quality Control 125 Cremona Drive Goleta, CA 93117 Hospital Representative Signature: Date: Hospital Representative Name (print): Hospital Representative Telephone#: If you have any questions outside of the return process, please do not hesitate to contact: Bob Shokoohi Sr. Director of Quality Operational Excellence Telephone: (805) 571-8725 Email: bob.shokoohi@medtronic.com Page 6 of 9

ATTACHMENT B The incorrect valve rotor orientation (Figure 2) can be caused by misalignment and/or off-center positioning of the StrataMR Locator Tool, relative to the valve. The images that follow are meant to communicate examples of both correct and incorrect alignment and centering to assist clinicians in their efforts to set the valve pressure level setting as intended. Figure 3 provides an example of proper/correct technique. Figures 4, 5, & 6 provide examples of incorrect alignment and centering techniques that should be avoided. Figure 3 Correct Good Alignment and Centering of Valve Mechanism and Locator Tool Figure 4 Poor Alignment, but good centering of Locator Tool and Valve Mechanism Figure 5 Good Alignment, but poor centering of Locator Tool and Valve Mechanism Figure 6 Good Alignment, but poor centering of Locator Tool and Valve Mechanism Page 7 of 9

ATTACHMENT B To ensure the valve is set at the desired setting, it is critical to properly align and center the Locator Tool over the valve mechanism. The following language is taken directly from the Instructions for Use (IFU) and will aid in proper technique to ensure the clinician can achieve the desired pressure level setting. Warning: The StrataMR valve is designed with five wells separated by raised walls. The valve mechanism must be seated within the wells to achieve the intended pressure flow performance otherwise under drainage may occur and result in patient complications The StrataMR Adjustment Tools are for use by qualified personnel only; ensure users have adequate knowledge to prevent misuse The StrataMR valve is designed with five wells separated by raised walls. The valve mechanism must be seated within the wells to achieve the intended pressure flow performance otherwise under drainage may occur and result in patient complications. It is recommended that the initial performance level of the implanted StrataMR valve be changed no more than one performance level in any 24 hour period. Throughout this post adjustment period, the clinical condition of the patient should be carefully monitored. Positioning the Locator Tool after Implantation Warning: the Indicator Tool may give false performance level setting readings unless it is aligned with direction of CSF flow and centered over the magnetic valve mechanism. 1. Position the patient away from large metal objects, such that the implanted valve can be conveniently accessed. The patient may be upright or supine. 2. Palpate the valve implantation site to determine the location of the valve and valve reservoir. The magnetic valve mechanism is adjacent and downstream of the valve reservoir (figure 5a and 5b). Page 8 of 9

ATTACHMENT B 3. Position the Locator Tool above the valve so that the red flow direction arrow is aligned with the direction of CSF flow through the valve and is centered over the magnetic valve mechanism (figure 6a and 6b). 4. Verify that the alignment zone of the Locator Tool is centered over the (hard plastic) valve mechanism, not the valve reservoir dome (figure 7). It s important to note that the previously mentioned instructions are a subset of the entire IFU dedicated to the StrataMR adjustable valve system. Please reference the official IFU for complete instructions. Page 9 of 9