Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Similar documents
Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING

ali-q TM Gravimetric Verification Procedure

Standard Operating Procedure. Air Displacement Pipette Calibration

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

Transferpette -8/-12

Air Displacement Pipetting Modules. Total Integrated Liquid Handling

CH2250: Techniques in Laboratory Chemistry. Outline Measuring Mass Measuring Volume Significant figures. Mass Measurement

Inquiry Module 1: Checking the calibration of a micropipette

Performance Verification of BioTek s Precision 2000 using Artel s Multichannel Volume Verification (MVV ) System

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs

Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx

Air Displacement Pipetting Module. Total Integrated Liquid Handling

CHM250 Calibration and Measurement Lab. Balance Calibration

Improve pipetting results in pharmaceutical formulation by using MICROMAN E

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

CHM Introductory Laboratory Experiment (r17sd) 1/13

PERFORMANCE TEST INSTRUCTIONS FOR BIOHIT PROLINE PIPETTORS

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

Calibration of Volumetric Glassware

Pipettor. User Manual

VITLAB micropipette. Standard Operating Procedure (SOP)

VITLAB Volumetric instruments

Technical Procedure for General Laboratory Equipment

The pipettes cover a volume range from 0.1µl to 10ml.

Purpose. Introduction

RayPette Plus Autoclavable Pipette. User Manual

Experiment #2. Density and Measurements

CHM 100 / Introductory Laboratory Experiment (r10) 1/11

General Accreditation Guidance. User checks and maintenance of laboratory balances

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

Pipetting Small Volumes

CORESTA RECOMMENDED METHOD N 6

Measuring Mass and Volume

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

Density Determination Kit Instruction Manual

Drilling Efficiency Utilizing Coriolis Flow Technology

QAM-I-117 Volumetric Equipment Calibration Verification

EXPERIMENT 1 BASIC LABORATORY TECHNIQUES AND TREATMENT OF DATA MEASUREMENTS

BASIC LABORATORY TECHNIQUES (Revised )

Ti-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T WP

CORESTA RECOMMENDED METHOD Nº 67

METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1.1 Applicability. This procedure is applicable for

Analysis of Deuterium Enrichment by Fourier Transform Infrared Spectrometry (FTIR): Practice. Christine Slater PhD

BASIC LABORATORY TECHNIQUES (Revised )

Adam Equipment. DENSITY DETERMINATION KIT for AFP, AAA and AAA/LE SERIES BALANCES OPERATOR S MANUAL

The use of the analytical balance, and the buret.

MVS Volume Verification Using Any Microtiter Plate or Small Volume Container, Such as a Tube or Vial

Routine Testing. SOP Temperature Calibration of Halogen Moisture Analyzer. Applicable for HR / HG / HB / MJ Moisture Analyzers

Ensuring pipette & weighing device performance: Verification & Calibration

Adam Equipment. DENSITY DETERMINATION KIT for AAA, ADA and ADP SERIES BALANCES OPERATOR S MANUAL

IN-PLACE DENSITY WAQTC FOP AASHTO T 209 (16)

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

UNCC Biotechnology and Bioinformatics Camp. Dr. Jennifer Weller Summer 2010

CHEM 321 Experiment 1

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007

VOLUMETRIC TECHNIQUES

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts

DENSITY CALIBRATIONS AT MIKES. Heikki Kajastie, Jorma Manninen and Kari Riski Centre for Metrology and Accreditation (MIKES) ABSTRACT

Accuracy and Precision

A Flexible, Multi-Purpose, Multi-Detector HTS Automated Platform

Hydrostatics Physics Lab XI

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

Register your instrument! Eppendorf Research plus. User Adjustment

(AS AT 31 st MARCH, 2002)

SOP No. 14. Recommended Standard Operating Procedure for Gravimetric Calibration of Volumetric Standards Using an Electronic Balance

Register your instrument! Eppendorf Xplorer / Eppendorf Xplorer plus. Adjustment

How to Achieve Optimal Weighing Performance

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

EXPERIMENT 7 THE IDEAL GAS LAW AND DENSITY

Lab Equipment ANALYTICAL BALANCE

Experiment 1 Basic Laboratory Operations

INSTALLATION AND OPERATING INSTRUCTIONS

Traceable calibration of automatic weighing instruments in dynamic operation

DS/PI-14 DIAPHRAGM SEAL TEMPERATURE ERROR

Vehicle-mounted meters, pump supplied

Hands-On Experiment Density and Measurement

MULTICHANNEL. Labnet International, Inc.

Chemical Industry. School experiments. Weighing. Helper Booklet. Tips and Tricks for Weighing. Natural science laws experience live learn easily

Calibrating Piston Pipettes

SOLUBILITY OF A SOLID IN WATER

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

Laboratory Activity Measurement and Density. Average deviation = Sum of absolute values of all deviations Number of trials

Commonwealth of Pennsylvania PA Test Method No. 742 Department of Transportation October Pages LABORATORY TESTING SECTION. Method of Test for

MSL Technical Guide 30. Calibrating Piston Pipettes. Introduction. Piston Pipettes. Calibrating Piston Pipettes overview

Application Note AN-107

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

How to Measure R7.1. Reference. I. Linear dimensions

CALIBRATION AND ADJUSTMENT

Lab 1: Precision and Accuracy in Measurement (Using the right tool for the Job) and Density of Metals

Eliminating Sources of Error in IV Pump Testing

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Buoyancy Does it Matter?

NMISA New Gas Flow Calibration Facility

ELEVATION CORRECTION FACTORS FOR RADON MONITORS*

CHEM 355 EXPERIMENT 7. Viscosity of gases: Estimation of molecular diameter

Transcription:

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600 November 2010

Table of Contents ``Abstract...3 ``Introduction...4 ``Methods & Results...6 ``Data Analysis...9 ``Conclusion...12 ``Appendix A...13 2

Abstract The Hamilton MICROLAB 600 syringe pump is compatible with syringes from 10 µl to 50 ml. Dispense accuracy across the various syringe sizes is dictated by two factors. The first factor is the inner diameter (ID) of the syringe barrel; the second is the positional accuracy of the syringe drive. The ID of a Hamilton syringe is held to the tightest possible specifications but some tolerance is expected and allowed. Since the customer installs the syringes after the instrument has been purchased it is critical that the MICROLAB 600 provide sufficient positional accuracy to compensate for the allowable variation in syringe ID. The following experiment will show that the positional accuracy of the MICROLAB 600 is such that a syringe on the large side and the small side of allowable tolerance will provide an accurate dispense. The result is that direct measurement of the drive stem displacement is validated to predict dispense volume and provides a suitable means for calibrating the instrument. 3

Introduction We have the capability to measure the linear syringe drive stem displacement with an accuracy of ± 0.0025 mm (2.5 microns) traceable to N.I.S.T measures. The linear motion delivered by the syringe pump is one of two components that account for the dispense volume. The inner diameter, and thus the cross-sectional area, of the installed syringe makes up the second component. By calculating the dispense volumes that would result with a syringe manufactured at the minimum diameter tolerance per the manufacturing print, and the maximum diameter tolerance per the manufacturing print, the range of possible dispense volumes can be predicted. If the predicted dispense volume falls within the accuracy specifications for the MICROLAB 600, then the pump is accurate for any syringe that meets manufacturing tolerance specifications. The accuracy specifications for a MICROLAB 600 configured with a 1 ml syringe are listed below. Table 1 MICROLAB 600 accuracy specifications for 1%, 5%, and 30% stroke of a 1 ml syringe ML600 Accuracy Specification Allowable Volume Range for a 1 ml Syringe (µl) 1% Syringe Stroke ± 3.0% 9.7 10.3 5% Syringe Stroke ± 1.2% 49.4 50.6 30% Syringe Stroke ± 1.0% 297 303 All Hamilton syringes are manufactured to specifications with a defined minimum (d1) and maximum (d2) allowable tolerance for the inner diameter. For example, the diameter of a 1 ml (1000 µl) syringe must fall between d1 = 4.600 mm and d2 = 4.620 mm. Each syringe is designed to dispense its nominal volume over a 60 mm stroke length. Assuming a theoretically perfect 60 mm stroke, the dispense volume for a syringe built with the minimum and maximum allowed inner diameter would be calculated using the following equations. Dispense Volume = Stroke Length (mm) x Area (mm 2 ) Area = π x (Diameter/2) 2 Minimum Dispense Volume Maximum Dispense Volume Stroke Length x π x (d1/2) 2 Stroke Length x π x (d2/2) 2 60 mm x π x ((4.600)/2) 2 60 mm x π x ((4.620)/2) 2 997.1 µl 1005.8 µl 4

Introduction Continued... The theoretical extremes above fall within 990 µl and 1010 µl which corresponds to the MICROLAB 600 s ± 1% accuracy specification for a 1 ml dispense from a 1 ml syringe. The next step is to incorporate the actual measured stroke of a MICROLAB 600 into the calculations above to show that the accuracy of the entire system meets specification with extreme scenario syringes. In practice, actual syringes will almost never be at exactly the extreme edges of tolerance, but will fall on average somewhere on a bell curve between the extremes. 5

Methods & Results: 1. Ten 1 ml syringes (P/N 59000-35) pulled randomly from stock were labeled and the barrel inner diameters were measured using the appropriate air gauge. The diameter was measured at the top, middle, and bottom of the barrel and these measurements were averaged to obtain the Measured Inside Diameter. See Table 2 for details. Table 2 Cross sectional areas of ten 1 ml syringes pulled from stock Syringe # Measured Inside Diameter (mm) Cross-Sectional Area (mm 2 ) 1 4.61391 16.71969 2 4.60866 16.68166 3 4.60917 16.68534 4 4.61264 16.71048 5 4.60705 16.67002 6 4.60908 16.68473 7 4.61086 16.69760 8 4.61035 16.69392 9 4.60917 16.68534 10 4.61383 16.71907 2. Using a calibrated Mitutoyo Digimatic Digital Indicator, the linear displacement of the MICROLAB 600 was measured 10 times at 1% stroke (0.6 mm), 10 times at 5% stroke (3 mm), and 10 times at 30% stroke (18 mm), ending each stroke at the zero position. The digital interface was used to enter the pump movement commands. The average of 10 measurements for each range was calculated and recorded in Table 3. Table 3 Average measured stroke for MICROLAB 600 Serial Number ML600AG1026 Stroke To Zero Position Average Measured Stroke (mm) 1% (0.6 mm) 0.6022 5% (3.0 mm) 3.0072 30% (18 mm) 18.0171 6

3. The average measured stroke, and measured syringe diameters were used to calculate the expected dispense volume using the equations below. Predicted gravimetric results for each syringe are shown in Table 4. Dispense Volume = Stroke Length (mm) x Area (mm 2 ) Table 4: Predicted dispense volume from measured linear displacement and measured syringe diameters Predicted Volume Syringe # 1% Stroke (µl) 5% Stroke (µl) 30% Stroke (µl) 1 10.068 50.279 301.240 2 10.046 50.165 300.555 3 10.048 50.176 300.621 4 10.063 50.252 301.074 5 10.039 50.130 300.345 6 10.048 50.174 300.610 7 10.055 50.213 300.842 8 10.053 50.202 300.776 9 10.048 50.176 300.621 10 10.068 50.278 301.229 7

4. Using the same pump and the same commands as in step 2, each syringe was tested gravimetrically at 1% (0.6 mm), 5% (3.0 mm), and 30% (18.0 mm), using de-ionized water. The test apparatus was calibrated to N.I.S.T traceable weights and the procedure followed is outlined in Appendix A. As per normal calibration procedures, each volume measurement was repeated 10 times and the average is reported in Table 5. Table 5: Gravimetrically measured dispense volumes Average Measured Volume Syringe # 1% Stroke (µl) 5% Stroke (µl) 30% Stroke (µl) 1 9.954 50.031 301.170 2 9.888 49.906 300.302 3 9.971 50.031 300.443 4 9.919 50.012 300.770 5 9.901 49.930 300.172 6 9.913 49.974 300.329 7 9.898 49.986 300.456 8 9.956 50.007 300.278 9 9.893 50.012 300.218 10 9.804 49.944 300.710 8

Data Analysis The data collected for Predicted and Gravimetric dispense volumes are displayed in Figures 1, 2, and 3 below. The graphs show the correlation between the predicted dispenses and the gravimetric dispenses. They also show the error bands which indicate the allowed dispense volumes at each stroke level. Figure 1: Comparing the Predicted vs. Gravimetrical Volume for a 300 µl dispense The Predicted vs. Gravimetric graph for the 300 µl dispense shows an almost perfect correlation. Syringes that were predicted to dispense slightly above the mean did dispense above the mean. Syringes that were predicted to dispense below the mean did dispense below the mean. Additionally all dispenses fell well within the minimum and maximum allowable dispenses according to the MICROLAB 600 specifications. 9

Figure 2: Comparing the Predicted vs. Gravimetrical Volume for a 50 µl dispense The Predicted vs. Gravimetric graph for the 50 µl dispense shows a correlation with slightly more deviation than the 300 µl graph. The Gravimetric line is consistently less than the Predicted line. This is related to Z factors like evaporation, humidity, density of water, and compressibility of the system. What is critical to note is that all dispenses continued to fall within the minimum and maximum allowable dispense specifications. 10

Figure 3: Comparing the Predicted vs. Gravimetrical Volume for a 10 µl dispense The Predicted vs. Gravimetric graph for the 10 µl dispense shows that all dispenses fall within the minimum and maximum allowable dispense specifications. The peaks and valleys of the graph do not exactly correlate. This is expected because the cross sectional area used to calculate the predicted volume in Table 4 was an average across the entire length of the syringe barrel. A 10 µl dispense only uses 1% of the barrel, so the predicted dispense could be improved by measuring the exact barrel ID over this short distance. In reviewing Figure 3, it is useful to note that from the aspect of the pump, the maximum and minimum allowable volumes shown represent a linear displacement of ± 0.0178 mm. This distance is smaller than the diameter of the average human hair. 11

Conclusion: The MICROLAB 600 is designed to accurately dispense a variety of different liquids over a wide range of volumes. Since it is not practical to test all possible combinations of syringe size, dispense volume, and liquid type, a test protocol was established to measure the linear accuracy of the syringe drive. This testing was experimentally shown to be predictive of the actual dispense volumes achieved using the same syringes with deionized water. As the data shows, under test conditions, the MICROLAB 600 system is easily capable of achieving the published specification. The specification is intentionally broad in an attempt to take into account unexpected factors that can influence dispense results. Potential factors include compressibility of the solvent, atmospheric pressure, humidity, etc. For these reasons it is good practice to test the accuracy and precision of the MICROLAB 600 under true laboratory conditions using the actual solvents to be dispensed. 12

Appendix A: Gravimetric Procedure for Calibrating a MICROLAB 600 Summary This general procedure is based on determining the weighing result of water samples delivered by the syringe. True volume is calculated based on the density of water at specific temperatures. Limitations This method is not recommended for volumes below 2 µl. There is no upper volume limit. Equipment, Materials, Environment 1. Laboratory balances required for the test method should meet or exceed the following performance specifications. They must be regularly maintained and calibrated with the appropriate N.I.S.T. traceable weights. Test Volume, µl Balance Sensitivity, mg 1 10 µl 0.001 mg 10 100 µl 0.01 mg 100 1000 µl 0.1 mg 2. Use a balance table, or suitable equivalent to minimize vibration. Cover the working surface directly in front of the balance with a dark, smooth, non-glare material. Keep the balance area reasonably free of draft currents and the ambient area free of excessive dust. 3. Use a weighing vessel that has a total volume 12 to 40 times the test volume, or 500 µl, whichever is larger (this is for evaporation control). If possible, use a cover that fits over the outside of the vessel top (do NOT allow the cover to come into contact with the test liquid). The vessel should be plastic, glass, metal, or some other non-porous material. The crosssectional area of the opening should be as small as possible to further control evaporation. 4. Handle the vessel with forceps or tweezers. 5. Use distilled water that has equilibrated to room temperature. 6. Use a calibrated thermometer to measure the temperature of the water. 13

Test Procedure 1. Turn on all equipment and allow all test materials to equilibrate to room temperature. 2. Place a small amount of water in the weighing vessel (between 2 and 30 test volumes). 3. Prime the ML600 to eliminate all air bubbles from the fluid path. 4. Run the method to be validated. 5. Open the door of the balance chamber, place the weighing vessel on the balance pan and close the door of the balance chamber. 6. Tare the balance. Retrieve the weighing vessel from the balance chamber, deliver the sample, and return the vessel to the balance pan, closing the door to the chamber. Observe and record balance readout. 7. Deliver a total of n samples (n=10 is recommended) into the weighing vessel, and weigh each sample after delivery. Replicate all motions and time intervals in each sampling cycle as precisely as possible. Keep the distance between the balance and the diluter/dispenser to a minimum. 8. Measure and record the water temperature. 14

Calculations 1. Calculate the volume of each dispense (Vi) by dividing each mass value by the density of water at the measured temperature. Refer to the table below for density values. Density of Water at Various Temperatures C g/ml C g/ml 17 0.998774 24 0.997296 18 0.998595 25 0.997044 19 0.998405 26 0.996783 20 0.998203 27 0.996512 21 0.997992 28 0.995646 22 0.997770 29 0.995944 23 0.997538 30 0.995646 Taken from CRC Handbook of Chemistry and Physics, 50th edition, 1969, page F-4 2. Calculate the average dispensed volume from the individual dispensed volumes, Vi (where i is 1 to 10): V avg = V1 + V2 + V3 + + V10) / 10 3. Calculate the syringe accuracy: Accuracy (%) = 100 x (Vavg Vo) / Vo Note: Vo is equal to the expected dispense volume 4. Calculate the standard deviation (SDEV) of the calculated volumes, then determine the coefficient of variation: CV (%) = 100 x SDEV / V avg Hamilton Company 4970 Energy Way Reno, Nevada 89502 USA Toll-free: 800.648.5950 Telephone: 1+ 775.858.3000 Fax: 1+ 775.856.7259 sales@hamiltoncompany.com Hamilton Bonaduz AG Via Crusch 8 CH-7402 Bonaduz, GR, Switzerland Tel: +41-81.660.60.60 Fax: +41-81.660.60.70 contact@hamilton.ch 15