How to use Solvent Purification System A brief manual

Similar documents
Solvent System Walkthrough

Manual Solvent Purification System. Innovative Technology, Inc.

Ammonia Treatment Instructions Page 1 of 5

Technical Bulletin AL-134

VACUUM & AIRLESS. Select from Kimble vacuum traps, manifolds, Schlenk tubes, adapters, and bubblers for your vacuum and airless glassware needs.

Working correctly with gas

Standard Operating Procedure Inert Vacuum Line. Ryan Mulvenna, May 2012

Douglas Call 4/11/2008 1/5. Ammonia Treatment

School of Chemistry SOP For Operation Of Glove Boxes

The Safe Use of Pyrophoric Substances

7.9. Flash Column Chromatography Guide

A/C SYSTEM GENERAL DIAGNOSTIC PROCEDURES

Birck Nanotechnology Center

STANDARD OPERATING PROCEDURES

ARC Unofficial Manual to the Solvent Dispensing System. Designed by J. C. Meyer

A/C SYSTEM DIAGNOSIS

Solvent Dispensing System Manual. Designed by J. C. Meyer

OPERATION MANUAL NTF-15

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

OPERATION MANUAL NTF-60 Plus

Hot Compressed Air Lance Operation Manual. LAB Manufacturing 9483 Reading Road Cincinnati, OH (800)

METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1.1 Applicability. This procedure is applicable for

Chemical Class Standard Operating Procedures. Pyrophoric Chemicals (PYR) & Water Reactive Chemicats (WRC) Diethylzinc

Armfield Distillation Column Operation Guidelines

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Setup Guide for your HVO Naked Drone

Hydraulic Reliability & Preventive Maintenance Report

Keofitt Aseptic Sampling Bag (Pat. pend.) Item no , & User Manual

KEOFITT ASEPTIC SAMPLING BAG (PAT. PEND.) Item no , & User Manual

MEMORIAL UNIVERSITY OF NEWFOUNDLAND Faculty of Engineering and Applied Science FLUID MECHANICS LABORATORY PIPE FRICTION

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

Neotecha Tank Sampling Sapro

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Review of the Hall B Gas System Hardware. George Jacobs

LNG Regulator Test Kit

1. The Safe Use of Pyrophoric Reagents

Experiment 18 Properties of Gases

User's Manual. MixRite TF 10. Edition 05.08

Lesson 6: Flow Control Valves

Air Ball! Evaluation copy

Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat. Version 5.1

Oxygen Dialflow Meter. Instructions for Use

Liquid Gas Cylinder (LGC)

SS32L Dissolved O 2 Probe

CITY AND COUNTY OF DENVER CR&CF RISK UNIT Compressed Gas Safety Standard

Standard Operating Procedure

Using the Akta Prime plus October 22, 2012

6800 Maintenance Instruction System Flush Procedure

[USING THE VITROBOT April 9, Using the Vitrobot

The Determination of the Value for Molar Volume

Neotecha Tank Sampling Sapro

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

Inert Air (N2) Systems Manual

Contact Person(s) : Isabel M. Fisenne APPLICATION

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

MS.RAJA ELGADY/PRESSURE PAPER 3

Copyright, 2005 GPM Hydraulic Consulting, Inc.

SOP: Buck Scientific BLC-20P HPLC Operation

6900 Maintenance Instruction System Flush

Chemistry 261 Laboratory Experiment 6: Reduced Pressure Distillation of a Mixture of High Boiling Alcohols

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

SOLVENT PURIFICATION SYSTEMS

Setting up and running a column

Customer Responsibilities

5 Gallon Pressure Pot with HVLP Spray Gun and Hose

Instruction Manual Updated 7/26/2011 Ver. 2.2

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Setting a New Standard in Flash Chromatography Performance

Glove Box Installation Manual

Calcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO

Lab-Specific Standard Operating Procedure (LSOP)- Pyrophoric Chemicals

Basic Flows in a Microfluidic Device

BROCKTON AREA MULTI-SERVICES, INC. MEDICAL PROCEDURE GUIDE. Date(s) Reviewed/Revised:

6890 GC Site Preparation

5890II GC Standard Operating Procedure 9/2/2005

300 Held Drive Northampton Pa

SOP for use of the SX19 Stopped-Flow Instrument

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Before you start sampling, be sure to read

Equipment and Accessories. ROMIL LiQuiSure. Designed for High Purity Containment & Dispensing PURE CHEMISTRY

Cover Page for Lab Report Group Portion. Head Losses in Pipes

COMPRESSOR REPLACEMENT PROCEDURE (R-134A)

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

AKTA pure 25 New Owner s Intro

NTF-515. Operation Manual

Unit 8: Gases and States of Matter

PURPOSE OF THE POLICY

Equipment Operating Procedure Glove Box

Research Teaching Production

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

OPERATING MANUAL LO-CROSS-FLOW CF M

LAB Chip Lid Molding LAB PROCEDURE

Standard Operating Procedure

Welcome to the Specialized Medical Services Respiratory training webinar series!

Figure. Experimental set up of catalyst, reactants and product syringes.

Welker Constant Pressure Cylinders With Welker Solid Indicator, Syringe T-Handle, or Vortex Mixer

MODEL MN-PWV PRESSURIZED WATER VESSEL ASSEMBLY OPERATING MANUAL

Properties of Gases Observing Atom Pressure of a Gas Measuring Gas Products of Chemical Inferring Molecule Reactions

INSTRUCTION MANUAL FOR POLYKEG DISPOSABLE KEG

Transcription:

1 How to use Solvent Purification System A brief manual By V. Azev. SPS (Figure 1) can be used to collect dry oxygen free solvents into two different types of containers. One of them is a syringe. Syringes are filled with a solvent in a usual manner using a septum port on the flask (Figure 2). The second type of container is a special flask equipped with a high vacuum valve (Figure 3). These flasks are used by a small population of the department, so their usage will not be discussed in this manual. 1. Sign in the log book. Data, name, solvent and volume taken are required fields. 2. Check out the pressure on the pressure gauge on the Argon tank. If the pressure is around zero then the system will not work at all. 3. Well, now you have to decide, would you trust the purity of the solvent, which remains in the flask, or you will have to use just the fresh one. It is quite hard to measure the desired amount of a solvent. Usually people collect an excess of it. And some of it remains in the flask unused. Solvent is kept under positive argon pressure, so it should remain quite dry and anaerobic. If your reaction is not so sensitive to these contaminants, use the solvent which remains in the flask and go to 5 (if you want to collect more) or 9. 4. In the case when you want a highly pure solvent, remove that remains. First of all, find a bottle with an appropriate solvent label, where you could pour the remaining material. Some of these bottles are in the SPS hoods and some are under the hood which is in the center of d Alarcao lab. After you have found a bottle, open valve # 7. ATTENTION You should turn it so that an arrow would point toward the flask. Then, carefully open valve # 8 HALFWAY WITH AN ARROW POINTING UP. If you open it with an arrow down, the system will be connected to the air the source of junk like water and oxygen (at least). If you would turn it ALL way, the huge 10 psi pressure of argon will flash out all the silicon oil from the bubbler. It will be a mess, which would have to be cleaned by you. The worst part of this mess is that the solvent could be contaminated with this ugly silicon oil. And as a consequence your product too. So, please, turn it approximately half way. The argon should flow through the Page 1 of 5

2 bubbler quite rapidly (~ 3-4 bubbles per second), but not crazy rapidly. Now put the bottle for the solvent under the flask outlet and open valve # 10 slowly (half of the turn). Allow the solvent to pour inside the bottle. As far as the system is under positive Argon pressure und the valve is open slightly, not so much air enter the system. After the flask has been emptied, close valve # 10 and valve # 8. Now we are ready to fill the flask with a fresh solvent. 5. Open valve # 1. It creates the driving force to transfer a solvent from a storage tank into a flask. Also, open valve #2 (ARROW UP), valve # 3 (one way valve), valve #4 (ARROW UP), valve # 5 (one way valve). There are two options at this point. Valve # 7 may be opened (if You did emptify the flask) or not. If not turn it with an arrow toward the flask!!!! 6. Slowly open metering valve # 6. You should be able to see, how the liquid flowes. Try to take as precise as possible. Actually, collect approximately 10 ml less, then necessary. 7. Now, completely close valve # 6. Then, carefully open valve # 8 HALFWAY WITH AN ARROW POINTING UP. If you open it with an arrow down, the system will be connected to the air the source of junk like water and oxygen (at least). If you would turn it ALL way, the huge 10 psi pressure of argon will flash out all the silicon oil from the bubbler with the outcomes described at 4. Solvent stuck in the tubing will come into the flask. 8. Close valves # 8, 7, 5, 4, 3, 2, 1. You may collect the solvent with your syringe now. 9. Before inserting a syringe needle inside the flask it is advisable to increase the bubbling slightly. To do this open valve # 9 a little bit CCW. A flow-rate of 1 bubble per second will suffice for this step. 10. After you have taken the desired amount of solvent bring the flowrate back to around 1 bubble per 2-3 seconds, just turn valve # 9 back (CW). 11. WHEN YOU LEAVE, MAKE SHURE THAT THE BUBBLING IS NOT CRAZY. 12. Feel absolutely free to ask Slava Azev (M 302, 7-20-42) if you have questions. Page 2 of 5

3 Appendix Figure 1. General view of one of the SPS units. Page 3 of 5

4 Figure 2. Collecting solvent with a syringe. Page 4 of 5

5 Figure 3. Solvent storage flask Page 5 of 5