T004.BioTector TOC Analyzer Pressure and Flow Test failure analysis

Similar documents
200 PSI HIGH-FLOW AIR SOURCE KIT

FLUID POWER FLUID POWER EQUIPMENT TUTORIAL OTHER FLUID POWER VALVES. This work covers part of outcome 2 of the Edexcel standard module:

Reclaim Basic Set Up

Test Report # Rev 0. Adiabatic Compression In Nitrogen Service

200 PSI FAST-FILL AIR SOURCE KIT

Bulletin TCR-104 & 109 Filling and adding to the Glycol pressure system

AIR BRAKE SYSTEM TROUBLESHOOTING

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

ASVAD THE SIMPLE ANSWER TO A SERIOUS PROBLEM. Automatic Safety Valve for Accumulator Depressurization. (p.p.)

SPECIFICATIONS ATTENTION

Hydraulic Reliability & Preventive Maintenance Report

Hazards associated with the gas system and how to mitigate them

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank

42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only)

The Water O 3. zonator. Model WOZ3 Manual CONTENTS

MODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE

Pneumatic Powered - Plunger Pumps

BRAKE SYSTEM BLEEDING

INDIAN INSTITUTE OF TECHNOLOGY KHARAGPUR NPTEL ONLINE CERTIFICATION COURSE. On Industrial Automation and Control

Calcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO

Operating Instructions Model and Hydrostatic Test Pump

R E D I C O N T R O L S

Navigator 600 Silica analyzers

OWNER S TECHNICAL MANUAL

WARNING: READ THE GENERAL INFORMATION MANUAL INCLUDED FOR OPERATING AND SAFETY PRECAUTIONS AND OTHER IMPORTANT INFORMATION.

USER S MANUAL BASIC HORIZONTAL OIL-LUBRICATED MICROCOMPRESSOR Reference: RCMPBH

Introduction to Pneumatics

Suggested Installation & Operating Instructions for Sidewinder Pumps

N2 Blanketing Valve DST100 / DST200 TYPE INSTRUCTION MANUAL CONTENTS K.S.P.C. General Description Operation. Installation Maintenance

Working correctly with gas

Instructions for SOTA Water Ozonator Model WOZ5

Exercise 4-2. Centrifugal Pumps EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Pumps

PRESSURE REDUCING VALVE RP45 (EN)

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website:

EMISSION CHAMBERS. Bulk and Surface Emission Detection for the RAD7 User Manual

Generating Calibration Gas Standards

Pump out insert (condenser and still lines) overnight with turbo pump

PACKSETTER HAND PUMP

SERIES 10. Pilot Operated Tank Blanketing Valve, 2" PROTECTOSEAL.

Preliminary Failure Mode and Effect Analysis for CH HCSB TBM

INSTruCTIONS FOr SOTa WaTEr O ZONaTOr model WOZ5 CONTENTS INTrOduCTION COmplEmENTary units

INSTALLATION and OPERATION INSTRUCTIONS

Installation, Operation, and Maintenance Manual

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

The Nitrogen Threat. The simple answer to a serious problem. 1. Why nitrogen is a risky threat to our reactors? 2. Current strategies to deal with it.

OPERATING INSTRUCTIONS AND PARTS LIST HAND OPERATED LPG PISTON PUMP WITH SCALE 301-DH-SP

Chapter 6 Leak detection

University of Minnesota Nano Center Standard Operating Procedure

Digester Processes. 1. Raw Sludge Pumping System

SKP Series SKP25 U.. Pressure Regulating Gas Valve Actuator with Safety Shut-off Function

120 PSI FAST-FILL AIR SOURCE KIT 25% Duty Compressor on 1.5 Gallon Air Tank

Glove Box Installation Manual

RHPS Series RD(H)F40 User Manual. Read the complete manual before installing and using the regulator.

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

MB-SP Series Solvent Purification System

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER FILTER DRYER

Air Eliminator Series

Operating Instructions Model and Hydrostatic Test Pump

Copyright, 2005 GPM Hydraulic Consulting, Inc.

Air Mortar User Manual

Operating Instructions Models 36200, , 46200, Hydrostatic Test Pumps

OPERATING MANUAL LO-CROSS-FLOW CF M

Temperature Controllers

Moving Fluids Concept and Theory

GBK-50M / GBK-110M. Booster. Instructions. Introduction. How to Contact Fluke Calibration

400C & 450C DUAL PERFORMANCE VALUE PACKS

OPERATING INSTRUCTIONS Sherlock Chamber Testing System

Inert Air (N2) Systems Manual

Pressure Regulator. Instructions for Use

HYDRAULIC CYLINDER TEST

2. Determine how the mass transfer rate is affected by gas flow rate and liquid flow rate.

WW-730. Pressure Sustaining/Relief Control Valve

2 Sentry MCL Installation, Operation & Maintenance

Operation Manual. Carbon Filtration Industrial Drive, Orlinda, TN fax


Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

SPCH/ Blowing block

AcornVac Vacuum Plumbing Systems - Trouble Shooting Guide

WW-720. Pressure Reducing Control Valve

HS Bubbler Technology. HyQuest Solutions New Zealand User Group August 2017

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER COMPOSITE GAS SAMPLER

CHECK-VALVES Automatically Position Themselves by the Natural Forces of the Conveying Air. A PNEUMATIC CONVEYING SYSTEM PATENT #: 5,160,222

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply

RG6000. Refrigerant Recovery Machine. Operating Manual. 99 Washington Street Melrose, MA Phone Toll Free

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

Installation Instructions

Delayed Coker Safe Sampling Tar or Spalled Coke

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

ZHE+ Zero Headspace Extractor Operation & Instruction Manual

Standard: NFPA 86 Process Ovens and Furnaces. NFPA 85 Boiler and Combustion Systems Hazards Code

Tank Blanketing Pressure Regulators RHPS Series

Genie Supreme TM Model 123HP Installation & Operation Instructions

VB85/ Pressure regulating valve (Unloader)

The Discussion of this exercise covers the following points: Pumps Basic operation of a liquid pump Types of liquid pumps The centrifugal pump.

Oxygen Dialflow Meter. Instructions for Use

Heat Pump Connections and Interior Piping

Genie Supreme Model 123 Installation & Operation Instructions

Transcription:

March 2013 PAGE 1 of 5 T004.BioTector TOC Analyzer Pressure and Flow Test failure analysis Section 1: Pressure Test. Theory of operation. The Mass Flow Controller (MFC) in the BioTector is used to confirm that there is no leak in the BioTector. The pressure test operates as follows: 1. The normally open exhaust valve is energized closed. 2. The normally closed sample out valve is de-energized so that it remains closed. 3. The MFC is given a set point typically of 40 LPH. 4. The gas flows through the MFC, but because all the outlets from the reactor are sealed, after about 10 seconds, the flow through the MFC will start to fall. 5. If there are no leaks in the system, the flow through the MFC will eventually fall to < 1 LPH, as the pressure in the BioTector equalizes with the pressure on the oxygen regulator. The sections of the BioTector which are covered by the pressure test typically are: 1. Outlet of MFC. 2. Ozone generator. 3. Reactor. 4. Circulation pump. 5. ARS valve. 6. Sample out valve, MV5. 7. Cooler. 8. CO 2 analyser. 9. Ozone destructor. 10. Exhaust valve. See example system drawing below for details.

PAGE 2 of 5 Pressure Test Simulation. The pressure test can be simulated in the Simulate menu as follows: 1. Confirm that the feed oxygen pressure, from the oxygen cylinder or oxygen concentrator is <1.5bar. 2. Energize the sample out valve so that it is open. 3. Remove the plug from the sample out valve so that it closes. This is essential, as the safety interlocks in Process Simulate 1 will prevent you from simulating the pressure test. 4. Energize the exhaust valve so that it closes. 5. Give the MFC as setting of 40 LPH. CAUTION! The BioTector is now under pressure. Therefore it is possible that if there is a leak in the system some liquid (acid or base) may quickly leak out. Always ware safety glasses when working with acid and base. CAUTION! Do not allow the pressure to rise above 350 mbar. 6. In there are no leaks in the system, the flow through the MFC will drop to <1 LPH after about 30 seconds. 7. When the test is complete, or to vent the system at any time, replace the plug on the sample out valve.

PAGE 3 of 5 Possible causes of a pressure test failure. Fitting not tight on Reactor, ARS valve, Circulation pump, Cooler, CO 2 Analyser or Ozone Destructor. Leak in Reactor, ARS valve, Circulation pump, Cooler, CO 2 Analyser or Ozone Destructor. Exhaust or Sample Out valve passing. Damaged ARS valve. The simplest procedure for checking this is to fill the sample in and sample bypass lines before and after the ARS valve with water. If there is a leak in the ARS valve, bubbles should appear in the sample bypass line, during the simulation of the pressure test (see page 3). Piston on the heavy duty pump worn out. When the ceramic piston is worn, the pressure test will fail as the oxygen gas will flow up through the piston (between the ceramic piston and the ceramic liner) and out the water out port. If the water out tube is immersed in liquid, after a few minutes bubbles will be seen leaving the tube, during the simulation of the pressure test (see page 3). Leak in Viton or Kalrez tube in the Ozone Generator. Crack in the Ozone Generator. Faulty MFC, which permanently displays high readings. To test this, give the MFC a low setting, for example 5 LPH and see if the flow drops to this level. Try tapping the control valve lightly if the reading remains high. Fault with the oxygen regulator. If the outlet pressure on the regulator increases as the flow through the MFC drops, then the flow will fall to a certain level and remain constant as the pressure on the regulator increases above its 350mbar setting. When the pressure stops rising, then the flow on the MFC will fall again. Example: 1. The pressure test starts with a flow of 40 LPH, the regulator pressure being 350 mbar. 2. The flow on the MFC quickly drops to 6 LPH, with the regulator output pressure remaining at 350mbar. 3. The output pressure from the regulator now slowly increases, with the result that the flow on the MFC remains constant at 6 LPH. 4. When the output pressure on the regulator stops rising, the flow on the MFC falls again. However, if the flow on the MFC remains at 6 LPH for over 60 seconds, then the pressure test will fail. Confirm that the feed oxygen pressure, from the oxygen cylinder or oxygen concentrator is < 1.5bar.

PAGE 4 of 5 Section 2: Flow Test. Theory of operation. The Mass Flow Controller (MFC) in the BioTector is used to confirm that there is no blockage in the BioTector. The flow test is divided into two sections, and operates as follows: Section 1, MV1 Exhaust gas path. 1. The normally open exhaust valve is de-energized open 2. The sample out valve is de-energized closed. 3. The MFC is given a set point typically of 80 LPH. 4. The gas takes the flowing path: a. MFC. b. Ozone generator c. Reactor. d. Cooler. e. CO 2 analyser. f. Ozone destructor. g. Exhaust valve and exhaust tubing. 5. If there is no blockage in the system, then the flow will remain at 80 LPH. However, if there is a partial blockage in the system, then the gas flow will drop. If the flow is less that 72 LPH (typical setting), a LOW O2 FLOW - EX (LOW FLOW MV1) fault will be generated. Section 2, MV5 Sample out gas path. 1. The normally open exhaust valve is energized closed 2. The sample out valve is energized open. 3. The MFC is given a set point typically of 80 LPH. 4. The gas takes the flowing path: a. MFC. b. Ozone generator c. Reactor. d. Sample out valve and sample out tubing, including the cleaning cycle catch pot (if fitted). 5. If there is no blockage in the system, then the flow will remain at 80 LPH. However, if there is a partial blockage in the system, then the gas flow will drop. If the flow is less that 72 LPH (typical setting), a LOW O2 FLOW - SO (LOW FLOW MV5) fault will be generated. See the drawing on page 2 for details. Flow Test - Simulation. The flow test can be simulated in the Simulate menu as follows: 1. For the MV1 (exhaust valve) gas path, the sample out valve should be de-energized closed and the exhaust valve de-energized open. 2. For the MV5 (sample out valve) gas path, the sample out valve should be energized open and the exhaust valve energized closed. 3. Give the MFC a set point of 80 LPH, and wait at least 60s. The flow and set point should be the same if there are no blockages in the system.

PAGE 5 of 5 Possible causes of a flow warning. Oxygen inlet pressure low. Exhaust (MV1) or sample out (MV5) valves blocked. Tube in the BioTector or outside the BioTector blocked. Ozone inlet to the reactor blocked. Ozone destructor blocked. Possible material built up in the reactor. There may be a build-up of a sludge like material in the reactor. This can have an effect on the flow. If this is the case, the reactor / glass bead chamber should be cleaned, and if necessary, the glass beads should be replaced. Faulty MFC, which permanently displays low readings. To test this, open the tube at the outlet of the MFC, and give the MFC a high setting, for example 80 LPH, and confirm that the flow increases to this level.