Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

Similar documents
NMISA New Gas Flow Calibration Facility

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

CORESTA RECOMMENDED METHOD N 6

Interlaboratory comparison in the pressure range from 0 to 2 MPa for accredited calibration laboratories

Gas Flow Calibration Basics

Uncertainties and Inter-Laboratory Comparisons of Dry Piston Gas Flow Provers

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

FPG8601 Force Balanced Piston Gauge

Hardware Triaxial and Consolidation Testing Systems Pressure Measurement and Control

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

ali-q TM Gravimetric Verification Procedure

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS

Pneumatic high-pressure controller Model CPC7000

Mass Flowmeters. Mass Flowmeters for Gases

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

Traceable calibration of automatic weighing instruments in dynamic operation

Pneumatic high-pressure controller Model CPC7000

VITLAB micropipette. Standard Operating Procedure (SOP)

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

UNDERSTANDING, ACCELERATED

Technical Report. 5th Round Robin Test for Multi-Capillary Ventilation Calibration Standards (2016/2017)

Technical Report. Comparison of Ventilation Standard Calibration Methods

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

Operational experiences with the EuroLoop Liquid Hydrocarbon Flow Facility

Transferpette -8/-12

Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip.

Improve Process Reliability

Paper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF PRESSURE METROLOGY

Pneumatic High-Speed Pressure Controller Model CPC3000

Organisation Internationale de Métrologie Légale

Truck-mounted Mass Flow Metering for LPG Delivery

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

SPECIFICATIONS PARTICLE SENSOR KS-18F Higashimotomachi, Kokubunji, Tokyo , Japan

800 Series Mass Flow Meters and Controllers. High Performance Mass Flow Meters and Controllers

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS

SPECIFICATIONS PARTICLE SENSOR KS-19F Higashimotomachi, Kokubunji, Tokyo , Japan

CalTrak 800 High Accuracy Modular Primary Standard Gas Volume Flow Calibrator

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

ONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Multiple Tube - Teflon PTFE-Glass Flowmeters

PM105 PRESSURE METER USER S MANUAL

Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

HumiSys HF High Flow RH Generator

Hydrostatics Physics Lab XI

Hydraulic pressure balance Model CPB5800

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CALIBRATION CERTIFICATE # 500

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

Automated Mass Handling for High Performance Pressure Balances

PPC4E Pressure Controller/Calibrator Superior rangeability and reliability

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

P3000 Deadweight Tester Setup, Part 1: Limited Partial Correction Method

KENYA ACCREDITATION SERVICE

Investigating the effects of interchanging components used to perform ripple assessments on calibrated vector network analysers

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

Pneumatic dead-weight tester Model CPB3500

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

Auto-Zero Calibration Technique for Pressure Sensors

General Accreditation Guidance. User checks and maintenance of laboratory balances

MET 335W Fluid Mechanics Laboratory. Lab 1: Bourdon Tube Calibration. Nick Peak

F&J SPECIALTY PRODUCTS The nucleus of quality air monitoring programs

Technical Information Conducal CLY421

Performance tests of two insertion type thermal mass flow meters

Gas Proportioners - Back Pressure Compensated

Instruction Manual. Revision A.2 May Cal=Trak SL-800 Primary Gas Flow Calibrator Instruction Manual

Exercise 4-2. Centrifugal Pumps EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Pumps

Buoyancy Does it Matter?

A New Piston Gauge to Improve the Definition of High Gas Pressure and to Facilitate the Gas to Oil Transition in a Pressure Calibration Chain

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C

Electronic Volume Correctors

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

AMS 6915 Board mount pressure sensor with digital output (I²C)

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

THE IMPLEMENTATION OF TOROIDAL THROAT VENTURI NOZZLES TO MAXIMIZE PRECISION IN GAS FLOW TRANSFER STANDARD APPLICATIONS

PG9607 Piston Gauge. Automated primary pressure reference for absolute and gauge pressures to 500 kpa

Design Features. General Description

Agilent Flowmeter ADM1000

MASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510

Gas Diluter Portable Gas Dilution System User s Manual

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Conversion factors for Thermal Mass flow meters for different operating mediums: An Experimental study

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Inquiry Module 1: Checking the calibration of a micropipette

_ pressure transducers. User Manual

Transcription:

Int. J. Metrol. Qual. Eng. 5, 303 (2014) c EDP Sciences 2014 DOI: 10.1051/ijmqe/2014012 Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA) D. Jonker and E.P. Tarnow National Metrology Institute of South Africa (NMISA), Pretoria, South Africa Received: 24 September 2014 / Accepted: 11 June 2014 Abstract. The National Metrology Institute of South Africa (NMISA) Gas Flow Laboratory provides traceability to the South African Industry for gas flow measurements. A new primary standard for gas flow calibrations was purchased and commissioned. With three flow cells, a flow range of 0.5 ml/min to 50 000 ml/min is covered. The main features of this standard are accuracy, speed and convenience. This paper describes the activities of the NMISA Gas Flow Laboratory a discussion of the primary standard, the validation thereof, calibration methods for client instrumentation, analysis of measurement results and the calculation of measurement uncertainties. Keywords: Traceability, gas flow measurements, primary standard, calibration methods, measurement uncertainties 1 Introduction The National Metrology Institute of South Africa (NMISA) is mandated to provide the South African industry with traceable measurements. Gas flow traceability was previously provided through a Califlow MKS primary gas flow standard. The Califlow MKS has become obsolete since it is no longer supported by the manufacturer and has become unreliable. Therefore a new primary standard for gas flow measurement has been purchased to replace the Califlow MKS instrument [1]. The NMISA gas flow laboratory calibrates a wide variety of flow instrumentation for the industry, especially instruments used for environmental monitoring. 2 The NMISA gas flow laboratory The gas flow laboratory is part of the Flow Section in the Physical Metrology Division. The Flow Section consists of the Pressure, Viscosity and Gas Flow laboratories. The laboratory is staffed by one metrologist who is responsible for all gas flow calibrations. Currently, only gas flow calibration services are offered. The flow range covered is from 0.5 ml/min to 50 000 ml/min. Nitrogen gas is used as the flow medium. Typical instruments received for calibration include mass flow controllers, mass flow meters, bubble flow meters and rotameters. The gas flow laboratory is equipped with the following equipment: Bios Met Lab ML-800 primary flow standard consisting of three flow cells. Correspondence: djonker@nmisa.org Bios Integrator 110 Met Lab Command & Control Module. Bios Gas Flow Bench. Vane anemometer (currently out of service). Bubble flow meter. Environmental monitoring equipment (barometers; temperature- and humidity loggers). The NMISA gas flow laboratory is not yet accredited to ISO/IEC 17025. At the moment measurements are traceable to international standards held at NIST (National Institute of Science and Technology). NMISA is aiming to achieve SANAS accreditation in 2014 and in-house traceability shortly thereafter [1]. 3 NMISA primary standard for gas flow calibrations The Bios Met Lab ML-800 was purchased as a primary standard for gas flow calibrations at NMISA. The ML-800 is a positive displacement primary piston prover and can be used for gas flow measurements in pressure or vacuum applications. The ML-800 complies with the requirements necessary for a primary standard, namely that flow measurements are made in terms of volume per unit time. This is accomplished by means of a piston moving in a cylinder of known cross-sectional area over a measured distance in a measured time. Volumetric or standardised flow readings can be displayed at the push of a button, the standardised values being calculated from the pressure, as measured by the built in barometer in the ML-800 and a reference temperature, as entered by the metrologist. Measurements can Article published by EDP Sciences

303-p2 International Journal of Metrology and Quality Engineering be performed manually (one at a time), or automatically. Up to 100 measurements can be performed in an averaging sequence. The ML-800 measures gas flow rates at the measured atmospheric pressure ±7 mmhg. The ML-800 consists of two primary components a common base and selectable flow cells. The main computer and timing crystal are housed in the base component as well as a precision barometer. The various flow cells plug into the base to create a functional system the two components cannot operate independently. The flow cells consist of a borosilicate glass tube with a precision-machined piston. In addition, they also contain an integrated temperature sensor and barometric pressure transducer in the gas flow stream for instant conversion of the volumetric readings into standardised flow. Four interchangeable flow cell models are available: ML-800 Ultra Low Flow Cell: ML-800-3 (in use at NMISA) Flowrange:0.5to50sccm(mL/min) Accuracy: ±0.25%; ML-800 Low Flow Cell: ML-800-10 (future purchase being considered) Flow range: 5 to 500 sccm (ml/min) Accuracy: ±0.15%; ML-800 Medium Flow Cell: ML-800-24 (in use at NMISA) Flow range: 50 to 5 000 sccm (ml/min) Accuracy: ±0.15%; ML-800 High Flow Cell: ML-800-44 (in use at NMISA) Flow range: 500 to 50 000 sccm (ml/min) Accuracy: ±0.15%. The measurements performed with the ML-800 are traceable to international standards. Before dispatch, the standard was calibrated by the calibration laboratory of the supplier, which is NVLAP (National Voluntary Laboratory Accreditation Programme in the USA) accredited. The Clock Period of the timing crystal and the barometer of the base unit were calibrated. The ultra-low flow cell was gravimetrically calibrated together with its temperature and pressure sensors. The medium and high flow cells were dimensionally calibrated together with their temperature and pressure sensors. For each flow cell, temperature and pressure corrections were applied to the volumetric flow readings to obtain standardised flow readings [2]. 4 Validation of the Bios Met Lab ML-800 primary flow standard 4.1 Purpose of the validation The purpose of the validation measurements was to ensure that the flow standard was not damaged during shipping from the United States of America to South Africa. The validation measurements were also required to provide objective evidence to demonstrate the competence of the new flow metrologist in the operation of the ML-800 and the calibration of flow instrumentation [3]. 4.2 Validation method Due to the very limited resources together with severe budgetary constraints, the ML-800 was used to calibrate two mass flow controllers (S/N 0858300004 & 0778900001) of an API Dynamic Dilution Calibrator (Model: 700; Serial Number: 777). These mass flow controllers had been calibrated against the Califlow MKS primary gas flow standard, Type A200, serial number 95068002N previously used in the NMISA flow laboratory. Nitrogen gas was used as the flow medium. Whilst this was far from an ideal validation exercise, since the mass flow controllers were of a lower accuracy than the ML-800, it was still considered to be valuable to establish confidence and competence of the metrologist, as well as identifying any significant shifts in the ML-800 performance after transit from the USA to South Africa. The ML-800 and the API Calibrator, which controlled the mass flow controllers being calibrated, were operated according to the procedures contained in their respective operating manuals. The two mass flow controllers were calibrated against the ML-800, one at a time. The inlet port of the mass flow controller being calibrated was connected to the nitrogen gas cylinder and the outlet port to the inlet port of the appropriate ML-800 flow cell. Connections to the mass flow controllers were made directly to the inlet and outlet ports and not to the ports at the rear of the calibrator. The reason for this was to ensure there was no gas diffusion/leakage possible via any other gas paths, and that the full gas stream passed through the mass flow controller being calibrated. Before any measurements were performed, the gas path was investigated for leaks by means of pressurising the system to 200 kpa and then using Snoop liquid leak detector fluid. Each mass flow controller was calibrated at twenty points and for each calibration point, ten measurements were performed and an average calculated. The measurement results were recorded manually on a worksheet. Ambient conditions were also recorded [3]. 4.3 Validation results The Ref Val values in the graphs in Figures 1 and 2 were the values as obtained through calibration against the Califlow MKS instrument. The Lab Val values were obtained through calibration against the new ML-800 instrument. The Lab Val values have been intentionally offset on the x-axis for clarity. Figure 1 shows that the laboratory values obtained with the ML-800 are higher than the reference values which were measured with the Califlow MKS. The measurement uncertainties calculated for the laboratory values are smaller than that of the reference values. At the nominal flow rates of 2 L/min to 8 L/min, 19 L/min and

D. Jonker and E.P. Tarnow: Gas flow calibrations performed at the NMISA 303-p3 Flow Rate Dev from Nom (L/min) 2,0 1,5 1,0 0,5 0,0-0,5 MFC-0858300004 0,0 1,0 2,0 3,0 4,0 5,0 6,0 7,0 8,0 9,0 10,0 11,0 12,0 13,0 14,0 15,0 16,0 17,0 18,0 19,0 20,0 21,0 Ref Val Lab Val Fig. 1. Flow measurement deviations from nominal flow rate for MFC-0858300004. Flow rate Dev from Nom (L/min) 0,020 0,018 0,016 0,014 0,012 0,010 0,008 0,006 0,004 0,002 0,000 MFC-0778900001 0,000 0,010 0,020 0,030 0,040 0,050 0,060 0,070 0,080 0,090 0,100 0,110 0,120 0,130 0,140 0,150 0,160 0,170 0,180 0,190 0,200 0,210 0,220 0,230 0,240 0,250 Ref Val Lab Val Fig. 2. Flow measurement deviations from nominal flow rate for MFC-0778900001. 20 L/min, the measurement uncertainties of the reference and laboratory values are not overlapping. This indicates that the calculated normalised error (E n )valuesare greater than one at these points. The measurement results of the two calibrations are therefore not equivalent. In Figure 2, again the the laboratory values obtained with the ML-800 seemed to be higher than the reference values which were measured with the Califlow MKS, except for the nominal flow rates of 0.01 L/min to 0.03 L/min. At 0.01 L/min and 0.02 L/min the reference values are higher and at 0.03 L/min the two values are exactly the same. At nominal flow rates up to 0.08 L/min, the measurement uncertainties of the two calibrations are comparable, while at the higher flow rates the uncertainties of the reference values are notably larger. Measurement uncertainties of the reference and laboratory values are not overlapping for most of the nominal flow rates. The calculated normalised error (E n ) values varied between 1.00 and 2.58 (E n = 0.43 and 0.00 for flow rates of 0.02 L/min and 0.03 L/min respectively). Measurement results of the two calibrations are therefore not equivalent. Possible reasons for the differences in measurement results are: 1) Drift of the mass flow controllers since their calibration against the Califlow MKS; 2) Difference in calibration set-up; slightly different method; 3) The mass flow controllers were calibrated against the Califlow MKS shortly before its removal from service. The calibration accuracy of the the Califlow MKS may have been questionable. 5 Calibration methods Flow instruments are calibrated by connecting them in line with the reference standard. The inlet ports of mass flow controllers are connected to the gas cylinder supplying the flow medium, while the outlet ports supply gas to the ML-800 reference standard. Bubble flow meters are connected to the outlet of the reference standard. The reason for this method is to protect the reference standard from possible soap solution contamination of the flow cells. Furthermore for each calibration performed, the volumetric flow readings are converted to standardised flow readings at standard temperature and pressure as requested by the client, using the following equation: Volumetric flow = Q (T m /T std ) (P std /P m ) (1) where Q is the Standard flow rate; T m is the measured temperature of the gas in the flow tube; T std is the standard temperature as requested by client; P m is the gas pressure measured in the flow tube; P std is the standard pressure as requested by client [1]. 5.1 Effects of unit under test calibration position Flow meters are connected in line with the reference standard or to the outlet of the standard, depending on the design of the flow meter. In order to establish whether or not the position of the unit under test (UUT) in the flow line influences the measurement results, measurements were performed connecting a digital flow meter before and after the ML-800 reference standard. The results appear in Figure 3. The difference between the reference and UUT values are smaller when the UUT is connected in line with the reference standard than to the outlet of it, except for the 0.1 L/min flow rate. The measurement results show that the biggest differences in the calibration position were measured at the middle of the calibration range. At a nominal flow rate of 3 L/min the difference is 0.044 L/min and at 4 L/min the difference is 0.049 L/min. The smallest difference occurs at 0.1 L/min and 1 L/min. At the nominal flow rates of 2 L/min; 3 L/min; 4 L/min and 6 L/min, measurement uncertainties are also not overlapping. The differences between the reference and UUT values for the two calibration positions at the above mentioned flow rates are bigger than the uncertainty values

303-p4 International Journal of Metrology and Quality Engineering Ref Val - UUT Val (L/min) 0,16 0,14 0,12 0,10 0,08 0,06 0,04 0,02 0,00 Effect of calibra on posi on -0,02 0,0 1,0 2,0 3,0 4,0 5,0 6,0 7,0 8,0 Connected to the outlet Connected in line Fig. 3. Flow measurement values illustrating the effect of the calibration position of the UUT. at these points. The uncertainty value at 5 L/min is relatively larger than the other uncertainties which resulted from poor repeatability at this point. The two calibration methods indicate a measurable difference in results. 5.2 Effects of gas supply pressure Currently the calibration method used in the laboratory requires that a gas supply pressure of 200 kpa be used. In order to establish whether or not the gas supply pressure influences the measurement results, measurements were performed in which a digital flow meter was calibrated at a nominal flow rate of 7 L/min against the ML-800 and the supply gas pressure varied between 50 kpa and 300 kpa. The results can be seen in Figure 4. The worst case difference between the flow rate values of the reference standard and the UUT in the pressure range 50 kpa to 300 kpa, was 0.023 L/min. Due to the fact that this difference falls within overlapping uncertainties, the effect of the gas supply pressure between 50 kpa and 300 kpa can be considered insignificant. 6 Calculation of measurement uncertainties Uncertainties are calculated according to the Guide to the Expression of Uncertainty in Measurement. The following sources of uncertainty are considered with each calibration: a) ML-800 accuracy Since the calibration certificates for the Flow Cells provided evidence that their accuracy was well within the manufacturer s accuracy specification, the manufacturer s accuracy specification was accepted as the uncertainty estimate. It was treated as a normal distribution variable, at k = 2 and therefore divided by two. Ref - UUT Flow Rate (L/min) 0,10 0,09 0,08 0,07 0,06 0,05 0,04 Effect of gas supply pressure 0,03 0 50 100 150 200 250 300 350 Inlet Pressure (kpa) Fig. 4. Flow measurement values indicating the effect of gas supply pressure. b) Temperature No uncertainty contribution for temperature is considered since the ML-800 sensitivity to temperature is unknown. However, since the ML-800 is used in a laboratory environment within that specified by the manufacturer for the ML-800 to remain within its accuracy specification, the effect of temperature is assumed to be negligible. c) Pressure No uncertainty contribution for pressure is considered since the ML-800 sensitivity to pressure was unknown. The ML-800 compensated for atmospheric pressure, as measured by its internal barometer. d) Repeatability of measurements (ESDM) standard and unit under test as a combined system The experimental standard deviation of the mean is used as the uncertainty estimate for the variation of repeated measurements, since the measurements are performed under repeatable conditions (conditions remained the same) and the calculated mean value of ten independent flow measurements is reported. The degrees of freedom are the number of measurements minus one, which for ten measurements is nine. e) Resolution of ML-800 The displayed resolution of the ML-800 is dependent on which flow rate is being measured. Therefore the uncertainty estimate is accepted as one least significant digit of the displayed flow. Since this is the full range of a rectangular distributed input variable, it is divided by two and then further divided by the square root of three. It is assigned infinite degrees of freedom. f) Resolution of the unit under test The uncertainty estimate is accepted as one least significant digit of the displayed flow. Since this is the full range of a rectangular distributed input variable, it is divided by two and then further divided by the square root of three. It is assigned infinite degrees of freedom.

D. Jonker and E.P. Tarnow: Gas flow calibrations performed at the NMISA 303-p5 Uncertainty estimates are calculated in absolute values. Measurement uncertainties are calculated for each measurement point. Uncertainty calculations indicate the most significant contributors to be the UUT display resolution and system repeatability [1]. 7 Conclusions and recommendations The validation described in Section 4 was unsatisfactory in achieving the method target measurement uncertainty. However, the results provided confidence that no damage had been sustained during shipment of the ML-800 from the USA. A bilateral comparison has therefore been planned with an overseas National Metrology Institute to validate the calibration method adequately. Mass flow controllers with a dedicated read-out and control unit, with a short calibration interval will be used as transfer standards. Initial measurements to determine the effects of UUT calibration position (Sect. 5.1) indicate possible significant effects. Initial measurements to determine the effects of gas supply pressure (Sect. 5.2) indicate possible insignificant effects. However, due to the fact that only one instrument was used for these measurements, the sample was not large enough to represent all instruments. It is therefore recommended that further measurents be performed on a wider range of instrumentation. References 1. D. Jonker, E.P. Tarnow, NMISA new gas flow calibration facility, Proceedings of Test & Measurement 2013 Conference & Workshop (2013) 2. Bios Met Lab Series ML-800 User Manual, 2011, Bios International Corporation, MK-01-33 Rev. J. 3. D. Jonker, NMISA-12-0218, Validation report of the Bios Met Lab ML-800 primary flow standard, 2013