INSTRUCTIONS FOR HIGH PURITY, CHANGEOVER REGULATOR SYSTEMS (PCS, CRS SERIES)

Similar documents
INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

INSTRUCTIONS FOR MODELS UPH & UPHS HIGH FLOW, HIGH PURITY GAS REGULATORS

INSTRUCTIONS FOR MODEL SSC LOW DELIVERY PRESSURE GAS REGULATORS

INSTRUCTIONS FOR MODEL TSC LOW DELIVERY PRESSURE GAS REGULATORS

INSTRUCTIONS FOR MODEL SG6500 GAS HEATERS

INSTRUCTIONS FOR SINGLE STAGE LECTURE BOTTLE REGULATORS

INSTRUCTIONS FOR MODELS APC & APG HIGH PURITY, CORROSIVE GAS REGULATORS

INSTRUCTIONS FOR FM4460N SERIES HIGH PRESSURE, METAL TUBE FLOWMETERS

INSTRUCTIONS FOR MODELS SG6190 & SG6191 GLASS ENCASED, LOW PRESSURE, HIGH PURITY, MOISTURE PURIFIERS

INSTRUCTIONS FOR SERIES 150 TWO TUBE GAS BLENDER

INSTRUCTIONS FOR SERIES 50 & 50K FLOWMETERS

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

Pressure Differential Automatic Manifold For LPG. PDH1000 Series. LPG Installation & Operations Instructions

The 527 Series Automatic Switchover System with Inlet Pressure Transducers

SPECIFICATIONS. Approximate Weight: 10 oz. Surface Finish: Ra micro inch or less MATERIALS OF CONSTRUCTION

TESCOM 50-4X Series Safety, Installation & Start-Up Procedures

MANIFLEX System. 52B Series 52C Series 52S Series INSTALLATION AND OPERATION INSTRUCTIONS

Types 749B and R130 Changeover Manifolds

632 Series Gas Distribution System

The 527 Series Automatic Switchover System

Protocol Switchover Station

Pressure Regulators. Operating Instructions. Instrumentation

Installation and Operating Instructions For MANUAL MANIFOLDS SD and SDHP SERIES

453 Series Steam Heated Vaporizing Regulator

Installation Operating Instructions for Duplex Manual Manifolds PX-TMD Series

623 Series Dome Load Regulators

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters

Petro-Chemical & Analyzer Sampling Systems

Tri-Tech Medical Inc.

INSTRUCTION MANUAL GAS TRAK

SINGLE STAGE ULTRA-HIGH PRESSURE REDUCING REGULATOR INSTALLATION AND OPERATION INSTRUCTIONS. Certified ISO 9001

629 Series MicroManifold & Vent Kit System

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

Tri-Safem Model I1 Regulator. INSTRUCTIONS and PARTS

Petro-Chemical & Analyzer Sampling Systems

620 Series Switchover System

REGULATORS FOR SPECIFIC APPLICATIONS

SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual

624 Gas Panels INSTALLATION AND OPERATION INSTRUCTIONS. Before Installing or Operating, Read and Comply with These Instructions

RK-IR Sample Draw Aspirator Adapter Operator s Manual

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

MEDICAL EQUIPMENT CATALOG

Operating Manual. R280 Pressure regulator made of brass. 1. Intended Use

CSA Sample Draw Aspirator Adapter Operator s Manual

EXD SERIES PRESSURE DIFFERENTIAL SWITCHOVER SYSTEM

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

601 Series Regulators

RK LEL Sample Draw Aspirator Adapter Operator s Manual

AND OPERATION INSTRUCTIONS

Installation Instructions for NP100 Series Nitrogen Control Panels

MEGR-1912 Instruction Manual

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

LRS(H)4 Pressure-Reducing Regulator User Manual

Installation & Maintenance Manual. Economy Manifold

Operation Manual Piston Sensed Gas Pressure Regulators

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

Operator: Save these instructions for future use!

V-24 OXYGEN LANCE VALVE

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION

INSTALLATION MANUAL Matheson Tri-Gas Cabinet Enclosures

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

Model Series 62 Constant Differential Relay

Installation & Maintenance Manual. C2 & C3 Manifolds

ANDERSON GREENWOOD SERIES 9000 POSRV INSTALLATION AND MAINTENANCE INSTRUCTIONS

Economy Manifold Alert -2. Installation and Maintenance Manual V1.6

TOP VALVE. Pat. #5,857,486 & 5,944,050. Mid-Range Pressure PSIG Back Pressure and Pressure Relief Valves. Instruction Manual

- GAS REGULATOR - COMPACT STYLE

Description Advanced Features Typical Applications

Yoke Block Instruction Manual

INDUSTRIAL VALVES MODELS: C62-A; C62-D. INSTRUCTION MANUAL Installation Operation Parts Service DIAPHRAGM BYPASS PRESSURE REGULATING VALVES

SERIES 30. Spring Operated Tank Blanketing Valve PROTECTOSEAL.

TECHNICAL BULLETIN. Sigma-Aldrich Pure-Pac II System. Technical Bulletin AL-236

TOP VALVE. Pat. #5,857,486 & 5,944,050. High Temperature: max. 300 F (149 C) Back Pressure And Pressure Relief Valves. Instruction Manual

Type 1367 High-Pressure Instrument Supply System with Overpressure Protection

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR

GILMONT ACCUCAL FLOWMETERS

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Yoke Block Instruction Manual

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

MATHESON s LaserPro Gases and Gas Delivery Equipment

GALISO SOAP OPTION FOR PCT-ADW-D

Interface Devices, Inc. Hydraulic Mini Mule

SG3 Series Regulators - Pressure Reducing

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

36E DSI, HSI & Proven Pilot Two-Stage Combination Gas Valve INSTALLATION INSTRUCTIONS

1805 Series Relief Valves

SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

Standard Operating Procedure (SOP) Compressed Gas Cylinders (Original SOP from Eastern Washington University)

Safe Handling - Dimethylamine

LUBRICATOR ASSEMBLY AND OPERATING INSTRUCTIONS

Type S301 & S302 Gas Regulators INTRODUCTION INSTALLATION. Scope of Manual. Description. Specifications. Type S301 and S302. Instruction Manual

TECHNICAL DATA. Q = C v P S

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

Float Operated Level Controllers

Transcription:

INSTRUCTIONS FOR HIGH PURITY, CHANGEOVER REGULATOR SYSTEMS (PCS, CRS SERIES) THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR USE IN CONJUNCTION WITH HIGH PURITY, CHANGEOVER REGULATOR SYSTEMS (PCS, CRS SERIES). CRS Series IMPORTANT These instructions are for experienced operators who know the general principles and safety precautions to be observed in handling specialty gases and operating specialty gas equipment. If you are not certain you fully understand the safety precautions for handling gases, we urge you to obtain and read the Material Safety Data Sheet (MSDS) for each gas being used. Do not permit untrained persons to install, operate, or maintain this equipment. Do not attempt to install or operate this equipment until you have read and fully understand these instructions. If you do not fully understand these instructions, contact your Advanced Specialty Gas Equipment Distributor. Be sure this information reaches the operator. Your supplier has extra copies. ADVANCED Specialty Gas Equipment

SAFETY PRECAUTIONS Protect yourself and others. Read and understand the following instructions before attempting to use this equipment. Failure to understand and follow these instructions could result in serious personal injury and/or damage to equipment. Know and understand the physical and chemical properties of the gas being used. Observe general precautions for the use of gases. Observe safety precautions for the gas being used. Read and follow precautions on cylinder labels. Never use this equipment with gases not compatible with the materials of construction. The use of gases not compatible with the materials of construction may cause damage to equipment or injury to personnel. If flammable gases are used with this equipment do not locate it near open flames or any other source of ignition. If toxic or flammable gases are used with this equipment, emergency equipment applicable to the gases in use should be available in the operating area. Many gases can cause asphyxiation by displacing oxygen in the atmosphere. Make certain the area where this equipment is operated is well ventilated. Provide a device to warn personnel of oxygen depletion in the work area. Do not release toxic or flammable gases in the vicinity of personnel. Use this equipment only in well ventilated areas. Vent gases to the outside atmosphere, and in an area away from personnel. Be sure that venting and disposal methods are in accordance with Federal, State and local requirements. Locate and construct vent lines to prevent condensation or gas accumulation. Be sure the vent outlet cannot be obstructed by rain, snow, ice, insects, birds, etc. Do not interconnect vent lines; if more than one vent is needed, use separate lines. Relief devices should be installed and properly vented in all gas handling systems to protect against regulator failure and overpressurization. Never use oil or grease with this equipment. Oil and grease are easily ignited and may combine violently with some gases under pressure. Never connect this equipment to a supply source having a pressure greater than the maximum rated pressure. Refer to the Product Specifications (page 14) for maximum inlet pressures. 2

MANUFACTURER STATEMENT The information contained in this instruction booklet has been compiled by Advanced Specialty Gas Equipment Corp., (the Company), from what it believes are authoritative sources and is offered solely as a convenience to its customers. While the Company believes that this information is accurate and factual as of the date printed, the information including design specifications is subject to change without prior notice. DESCRIPTION PCS Series Primary Changeover Systems are designed for use in processes incorporating downstream line or station regulators located at the point of use. The PCS Series incorporates two regulators set at slightly different delivery pressures. Gas discharges from the side with the higher setting first (primary side) which is indicated by the In Service arrow located on the hand knob. The side with the lower delivery pressure setting will remain closed until the primary side has been exhausted (approximately 150 psi residual pressure). The changeover will automatically switch to the reserve bank (secondary side). A fluctuation in pressure will occur at this point at the outlet of the changeover. Downstream line regulator(s) (not included) will eliminate pressure variations to the process. CRS Series Changeover Regulator Systems provide constant delivery pressure control to instrumentation by incorporating an outlet line regulator and utilize a compact design housing two changeover regulators in a single body. Downstream line or station regulation is not necessary unless various distribution point pressures are required. Operation is the same as described for the PCS Series with approximately 200 psi residual cylinder pressure maintained. 3

OPTIONAL EQUIPMENT Pressure Switches used to monitor line pressure and activate an external alarm (such as a Model SG6550, or SG6551 Annunciator sold separately) when a certain predetermined pressure is reached. These switches are generally connected to the ends of the single row manifold headers. Model SG6540 General Service Pressure Switch Model SG6541 Explosion-Proof Pressure Switch Annunciators used in conjunction with pressure switches to provide both an audible alarm and visual indication of pressure switch activation. Model SG6550 Nema 4x Single Point Annunciator Model SG6551 Nema 1 Single Point Annunciator Flashback Arrestors (SG6545) for acetylene service only designed to be used on acetylene cylinders to prevent a flash or flame from entering the cylinder through a cylinder lead. Flexible Hoses Double Braided (all metal) three foot stainless steel flexible hoses (available for cylinder leads) extend service life and provide ease of connecting cylinders. Check Valves Prevent discharge of gas from manifold and pigtails when changing cylinders. Purge/Vent Valves allow for complete removal of entrapped air and moisture from the system upon start-up, or after a cylinder change out thus maintaining the high purity nature of the system. SG6680 (Set of 2) For Brass Systems SG6681 (Set of 2) For Stainless Steel Systems INSTALLATION WARNING: Before attempting to install and operate this equipment, read and fully understand the safety precautions on page 2 in this booklet. Failure to follow the safety precautions may result in serious personal injury and/or damage to equipment. The PCS and CRS Series Changeover Regulator Systems can be used either with two cylinders by connecting the inlet leads directly to each of the cylinders or with four or more cylinders by connecting the leads to the ends of two single-row manifolds. Typical layouts for use with two single-row manifolds are shown in Figures 1 and 2. Note: The PCS Series is factory preset to changeover at approximately 150 psig. The CRS Series is factory preset to changeover at approximately 200 psig. Liquefied gases or cryogenic liquids with vapor pressures less than 300 psig should not be used with standard preset units. Contact your Advanced distributor if lower factory settings are required. 4

1. Securely mount the metal panel with attached changeover to a wall. Panel mounting dimensions are shown in Fig. 3 and Fig. 4. The panel should be located such that the regulator gauges are at eye level with the operator (approximately 5 6 ft.). The single row manifold outlets (or cylinder valves, if a two-cylinder system is installed) should be located approximately 10 to 12 inches outward from the changeover regulator isolation valves. 2. Connect the process line (downstream piping) to the 1 4 in. NPT outlet port of the changeover regulator system. Use Teflon tape on pipe threads to prevent galling. 3. Attach the inlet leads to the changeover regulator isolation valves by inserting the O-Rings provided in the grooves located in each valve inlet adaptor. The inlet leads (rigid pigtails) may be carefully bent to align them with the manifold or cylinder valve fittings. Avoid sharp bends that could kink the tubing. 4. Single-row manifolds are shipped without cylinder leads connected to prevent damage to the leads. Manifolds over four stations long are shipped in sections and require assembly. To assemble single row manifold sections (if required), align the manifold sections and insert the O-Rings provided in the groove located in the threaded end of the pipe tee. Connect the sections being careful not to chip or dislodge the O-Ring. To attach cylinder leads to the manifold, insert the small O-Rings in the groove of the station valves. Connect the cylinder leads (pigtails) to the station valves being careful not to chip or dislodge the O-Ring. 5. Securely mount the two manifolds on a wall. Mounting dimensions for single-row manifolds are shown in Figure 5 (page 9). 6. Provide a means of securely supporting the cylinders connected to the system. The clamping brackets, chains, straps, etc. used must be capable of holding the cylinders in place to prevent them from falling. 7. If the system is sized for more than two cylinders, connect the inlet leads of the changeover regulator to the outlets of the two single row manifolds. 8. Close the two inlet isolation valves by turning the hand knobs fully clockwise. Turn the changeover regulator handknob clockwise on the PCS Series and counter-clockwise on the CRS Series until the IN SERVICE arrow points left. Close the line regulator on the CRS Series by turning the pressure adjusting knob counterclockwise until it reaches the stop. Do not turn the adjustment knob past the stop. Damage to the regulator could result. 9. After the changeover regulator system is installed and the downstream piping is connected, the system connections should be leak tested at the maximum working pressure. 5

Changeover Regulators Optional Pressure Switch Station Valves Wall Mounting Brackets Optional Pressure Switch 1/4" NPTM Outlet Inlet Isolation Valves Cylinder Valves Cylinder Leads Manifold Header LEFT BANK PCS SERIES RIGHT BANK Figure 1 Typical Installation of PCS Series System with Two Single Row Manifolds and Optional Pressure Switches. Optional Pressure Switch Line Regulator 1/4" NPTF Outlet Port Optional Pressure Switch Station Valves Changeover Regulator Wall Mounting Brackets Inlet Isolation Valves Cylinder Valves Cylinder Leads Manifold Header LEFT BANK CRS SERIES RIGHT BANK Figure 2 Typical Installation of CRS Series System with Two Single Row Manifolds and Optional Pressure Switches. 6

Inlet Gauge 3.0" Outlet Gauge Recommended Screw Size is.25" (2x) Inlet Gauge 7.50" IN SERVICE 1.95" IN SERVICE Isolation Valve ADVANCED OPEN CLOSE Specialty Gas Equipment Outlet.25" NPT male ADVANCED OPEN CLOSE Specialty Gas Equipment Isolation Valve Inlet.25" NPT (Plugged) 5.5" Inlet Adapters 8.70" 8.90" max FRONT VIEW.073" 1.75" 1.0" 1.25" 4.34" 2.0" 2x 1.0" SIDE VIEW 6.13" Figure 3 Mounting Dimensions for PCS Series Changeover Regulator System and Panel 7

Outlet Gauge 5.25" using.25" screws CRS Line Regulator INCREASE Outlet.25" NPT female (plugged) Inlet Gauge DECREASE Outlet.25" NPT female Inlet Gauge 6.75" 9.63" OPEN ADVANCED CLOSE IN SERVICE OPEN ADVANCED CLOSE Specialty Gas Equipment IN SERVICE Specialty Gas Equipment Inlet Adapter Isolation Valve CRS Changeover Regulator 3.75" 4.75" FRONT VIEW Inlet Adapter Isolation Valve 2.81".094" 8.75" SIDE VIEW 3.89" 4.44" Figure 4 Mounting Dimensions for CRS Series Changeover Regulator System and Panel 8

4" 3" 8 1 /4" Min. TOP VIEW 26 1 /8" (Typ.) 3 3 /8" (Typ.) 2 3 /4" (Typ.) 2 1 /4" Min. 6 3 /4" Max. 10" (Typ.) Station Valve CGA Outlet Adaptor 1 7 /8" (Typ.) Header Section Connecting Nut End Nut and Plug Cylinder Lead Height of Cylinders Plus 8" Min. 5/16" (Typ.) Bracket Optional Check Valve (in CGA) Floor Line FRONT VIEW Figure 5 Mounting Dimensions for a Single Row Manifold used on the Right Side of a Changeover Regulator System 9

LEAK TESTING AND PURGING CAUTION: Care should be exercised when bending cylinder leads to connect them to cylinders. Brass is especially subject to work hardening and may fail if continuously re-bent. 1. Connect cylinders to cylinder leads and connectors provided. CAUTION: Cylinder leads containing check valves can be pressurized with system pressure only up to the check valve seat. The threaded joints on the pigtail, including the CGA connection will not be pressurized with the process gas. 2. Use the process gas to leak test and purge the system. If the process gas is hazardous (flammable, toxic and/or corrosive) or sensitive to atmospheric contaminants, use clean dry nitrogen as a purge gas to leak test and purge the changeover system. 3. Isolate downstream side of the changeover regulator system by closing a downstream process isolation valve. 4. Stand to the side of the changeover regulator and slowly open the purge gas from the left side of the changeover. Open the isolation valve located on the left side of the changeover regulator and check inlet gauge for pressure into the regulator. Repeat the same procedure for the right side of the changeover. Note: By opening either isolation valve, cylinder pressure enters the changeover regulator automatically reducing the pressure to approx 200 psig on the CRS Series and 150 psig on the PCS Series. 5. On the CRS system, open the line regulator by turning the pressure adjusting knob clockwise until the desired pressure is indicated on the outlet gauge. 6. With cylinders connected but with the cylinder valves closed, leak check all connections with either a soap solution, such as Snoop or a gas leak detector. Note: The line regulator on the CRS System does not vent downstream system pressure when the pressure adjusting knob is turned counterclockwise to reduce delivery pressure. For applications where atmospheric constituents could contaminate your gas system, install a vent valve on the downstream side of the changeover regulator to vent pressure. Connect outlet of vent valve to a safe disposal area. 7. Purge both right and left sides of the changeover system if the process gas is hazardous or sensitive to atmospheric contaminants. Turn the changeover handknob clockwise on the PCS Series and counterclockwise on the CRS Series until the IN SERVICE arrow points to the left. This will allow gas to flow from the left side first. Turn the handknob 180º (one half turn) ( IN SERVICE arrow points right) to flow gas from the right side. 8. Vent the system to atmospheric pressure. Close both regulator isolation valves by turning the hand knobs fully clockwise. On the CRS Series, close the line regulator by turning hand knob counterclockwise until it reaches the stop. 10

OPERATION WARNING: Never operate a gas handling system under any circumstances if it is leaking or otherwise malfunctioning. DO NOT repair any leaks while system is under pressure. Damage to equipment and/or injury to personnel may result. 1. Close the two regulator isolation valves by turning the hand knobs fully clockwise. Turn the changeover regulator handknob to choose a side to become the primary side ( IN SERVICE arrow points either left or right). The side chosen will function as the primary side (bank no.1). On the CRS System, close the line regulator by turning the pressure adjusting knob counterclockwise until it reaches the stop. 2. Ensure that any purge or system vents are closed. 3. Isolate downstream side of the changeover regulator system by closing the downstream process isolation valve. 4. Stand to the side of the regulator and slowly open the cylinder gas from the primary side (bank no. 1) of the changeover. Open primary side regulator isolation valve and check inlet gauge for pressure into the regulator. Repeat the same procedure for the reserve side (bank no. 2) of the changeover. Note: By opening either isolation valve, cylinder pressure enters the changeover regulator which automatically reduces the pressure to approximately 200 psig on the CRS Series and 150 psig on the PCS Series. 5. On the CRS Series, open the line regulator by turning the pressure adjusting knob clockwise until the desired pressure is indicated on the outlet gauge. 6. Open the downstream process isolation valve to allow gas to flow from the changeover regulator system to the use point. The system is now in operation with gas being supplied from the primary side (bank no. 1), which is indicated by the IN SERVICE arrow located on the handknob. The reserve side (bank no. 2) remains closed. When the gas from the primary bank has been exhausted (approximately 200 psig on the CRS and 150 psig on the PCS systems), the regulator will automatically switch to the reserve bank. 7. Once a changeover has occurred, the hand knob on the changeover regulator should be turned 180 to indicate that bank no. 2 is now in service. This also resets the changeover regulator with bank no. 2 becoming the primary bank. A fluctuation in outlet pressure will occur at this point at the changeover regulator; however, the builtin line regulator on the CRS Series eliminates this fluctuation and insures a constant delivery pressure to the use point. Installation of a downstream line regulator(s) is required with the PCS Series to eliminate pressure variations to the use point. 11

CYLINDER REPLACEMENT 1. Close the regulator isolation valve for the side containing the exhausted cylinder(s). 2. On the side containing the exhausted cylinders, close all cylinder valves and manifold station valves (if single row manifolds are used in the system). WARNING: Hazardous gases must be discharged into a safety vent. Be sure to use a venting procedure that is environmentally acceptable and complies with Federal, State and local requirements. 3. Nonhazardous gases may be vented by carefully loosening the cylinder lead connector(s). If a hazardous gas is in use, vent inlet leads through a safety vent or the optional purge/vent valve. Purge the leads with clean dry nitrogen gas. Continue purging until the hazardous gas level is below the TLV for the gas. 4. Disconnect exhausted cylinder(s) and replace with full cylinder(s). 5. If the gas connected to the manifold is nonhazardous, use it to leak test and purge the cylinder lead. If the process gas is hazardous (flammable, toxic and/or corrosive) or sensitive to atmospheric contaminants, use clean dry nitrogen as a purge gas to leak test and purge the cylinder lead. 6. Slowly open cylinder valve(s) (and manifold station valves, if applicable). Open the regulator isolation valve and check inlet gauge for pressure into the regulator. This now becomes the reserve cylinder bank. 12

SHUTDOWN WARNING: Hazardous gases must be discharged into a safety vent. Be sure to use a venting procedure that is environmentally acceptable and complies with Federal, State and local requirements. 1. Close the cylinder valves. 2. Vent the system pressure to zero psig. If a hazardous gas was used, purge the entire system with clean dry nitrogen gas. Continue purging until the hazardous gas level in the system is below the TLV for the gas. 3. Close all system valves by turning the hand knobs fully clockwise. 4. On the CRS Series, close the line regulator by turning the hand knob counterclockwise until it reaches the stop. REPAIRS If the any part of the changeover leaks or malfunctions, take it out of service immediately. Repairs should be made by Advanced Specialty Gas Equipment Corp. who have the special tools, test equipment and trained personnel required to make a safe repair. Tampering with the changeover voids the warranty. Contact your Advanced Specialty Gas Equipment Distributor to arrange for repair. If a changeover leaks or malfunctions, take it out of service immediately. Do not attempt to repair these changeovers. Repairs should be made only by Advanced Specialty Gas Equipment Corp. who has the special tools, test equipment and trained personnel required to make a safe repair. Tampering with the changeover voids the warranty. Contact your Advanced Specialty Gas Equipment Distributor to arrange for repair. Repairs to changeovers done after the initial warranty period has expired are chargeable to the customer. Upon receipt at the factory, the changeover will be inspected and you will be contacted with a repair cost estimate. No item will be repaired until approval is received. There will be an evaluation charge assessed for equipment not repaired. All repairs should be arranged through your Advanced Specialty Gas Equipment Distributor. Note: All equipment being returned must be purged of all hazardous materials using a clean, dry inert gas (e.g. Dry Nitrogen) prior to return. 13

SPECIFICATIONS Manifolds Maximum Inlet Pressure 3000 psig Flow Coefficient (each station valve) C V = 0.43 Inlet Connections CGA connection as specified Changeover Regulators Maximum Inlet Pressure 3000 psig Minimum Inlet Pressure 300 psig* Inlet Pressure Gauge 0 4000 psig Gauge Size 2 in. dial Operating Temperature Range -40 F to +165 F Flow Coefficient C V = 0.06 Outlet Connection (PCS Series) 1 4 in. NPT male Line Regulators (CRS Series only) Maximum Inlet Pressure 3000 psig Delivery Pressure Range 10 150 psig Delivery Pressure Gauge 0 200 psig Gauge Size 2 in. dial Operating Temperature Range -40 F to +165 F Flow Coefficient C V = 0.15 Outlet Connection 1 4 in. NPT female * Contact your Advanced distributor if lower factory settings are required. 14

200 180 DELIVERY PRESSURE PSIG 160 140 120 100 80 60 40 20 1 2 3 4 5 6 7 8 9 10 FLOW RATE IN SCFM N 2 SUPPLY PRESSURE 2000 psig 400 psig Figure 6 Typical Performance PCS Series 15

150 120 DELIVERY PRESSURE PSIG 90 60 30 0 1 2 3 4 FLOW RATE IN SCFM N 2 Figure 7 Typical Performance CRS Series 16

MATERIALS OF CONSTRUCTION Manifolds Metal Parts Seals Check Valve Seats Station Valve Seats Brass or Type 316 Stainless Steel (as specified when ordering) Teflon EPDM with Brass Pigtails Viton with Stainless Steel Pigtails PCTFE Regulators Body Brass Systems Stn. Stl. Systems Gauges Brass Systems Stn. Stl. Systems Bonnets Brass Systems Stn. Stl. Systems Internal Metal Parts Exposed to Gas Brass Systems Stn. Stl. Systems Seats Diaphragms Seals Brass Bar Stock Type 316 Stn. Stl. Bar Stock Brass Type 316 Stainless Steel Brass Bar Stock 300 Series SS Bar Stock Brass and Stainless Steel Type 316 Stainless Steel Teflon Type 316 Stainless Steel Teflon 17

REPLACEMENT PARTS Teflon O-Ring Kits (package of 25 O-Rings) for connection between pigtail & station valve for header connection on brass systems stainless steel systems Replacement Pigtails, Rigid Type brass without check valves brass with check valves stainless steel without check valves stainless steel with check valves Replacement Pigtails, Flexible Type without check valves with check valves SG6081 SG6082B SG6082S SG6640-(CGA) SG6641-(CGA) SG6642-(CGA) SG6643-(CGA) SG6638-(CGA) SG6639-(CGA) Replacement Station Valves brass 0202-5083A stainless steel not available Additional Stations (for adding stations to the following existing manifolds)* brass system with rigid pigtails without check valves SG6660-(CGA) brass system with rigid pigtails with check valves SG6661-(CGA) brass system with flexible pigtails without check valves SG6662-(CGA) brass system with flexible pigtails with check valves SG6663-(CGA) stainless steel system with rigid pigtails without check valves SG6664-(CGA) stainless steel system with rigid pigtails with check valves SG6665-(CGA) stainless steel system with flexible pigtails without check valves SG6666-(CGA) stainless steel system with flexible pigtails with check valves SG6667-(CGA) * Specify left bank or right bank when ordering. Note: Insert applicable CGA connection number to complete part number. Example: SG6667 580. 18

WARRANTY Advanced Specialty Gas Equipment Corp.,(the Company), warrants to the initial purchaser of each changeover regulator system described herein, that such equipment will be free from defects in material and workmanship which result in breakdown or failure under normal use during a period of 12 months from date of shipment by the Company if used and maintained according to Advanced Specialty Gas Equipment written instructions. Purchaser is aware that this equipment is designed for specific applications and that using this equipment with the wrong or improperly purged gas or at the wrong pressure may damage or corrode the unit and cause personal injury. This warranty does not cover damage or malfunction due to contamination or corrosion. Purchaser must confirm that this equipment is compatible with the gas being passed through it. If there is any doubt about compatibility, consult your Advanced Specialty Gas Equipment Corp. distributor. The Company s liability under this warranty shall be limited to the repair, or at its option, replacement or refund of the purchase price, of such equipment which proves to be defective, provided; however, that this warranty shall only apply if the purchaser (1) gives the Company written notice within ten (10) days after discovery of such defect, (2) immediately on discovery of the claimed defect, discontinues all use of such equipment, and (3) returns such equipment freight prepaid to plant of manufacture. THERE ARE NO EXPRESS WARRANTIES OTHER THAN THOSE SPECIFIED HEREIN. NO WARRANTIES BY ADVANCED SPECIALTY GAS EQUIPMENT CORP. (OTHER THAN WARRANTY OF TITLE AS PROVIDED IN THE UNIFORM COMMERCIAL CODE) SHALL BE IMPLIED OR OTHERWISE CREATED UNDER ANY APPLICABLE LAW, INCLUDING BUT NOT LIMITED TO WARRANTY OF MERCHANTABILITY AND WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE. No claim against the Company of any kind, whether as to equipment delivery or for nondelivery of equipment and whether or not based on contract, warranty, negligence, strict liability in tort or otherwise, shall be greater in amount than the purchase price of the equipment in respect of which such claim is made. Without limiting the generality of the foregoing, Advanced Specialty Gas Equipment Corp. shall not be liable for any special, indirect, or consequential damage, such as failure of parts resulting from corrosion. If it is determined by Advanced Specialty Gas Equipment Corp. that the equipment is to be repaired or replaced under the terms of this warranty, the cost of returning said equipment to the initial purchaser will be paid by the Company. If, however, equipment returned to the Company in connection with a claim under this warranty is found by the Company not to be defective hereunder, then such equipment will be returned to the initial purchaser, shipping charges collect, and additionally, a service charge will be paid by the purchaser to the Company to cover the cost of handling and testing such equipment. 19

ADVANCED Specialty Gas Equipment 241 Lackland Drive, Middlesex, NJ 08846 Phone: 732-271-9300 Fax: 732-271-1630 www.asge-online.com Printed In U.S.A. AI 2074R Copyright 2004 Advanced Specialty Gas Equipment Corporation 3/04 Teflon and Viton are registered trademark of E.I. Du Pont de Nemours & Co. Snoop is a registered trademark of Nupro Co.