Dräger Evita Infinity V500 ventilator ICU Ventilation and Respiratory Monitoring

Similar documents
Evita V300 ICU Ventilation and Respiratory Monitoring

Dräger Savina 300 Classic ICU Ventilation and Respiratory Monitoring

Dräger Savina 300 Select ICU Ventilation and Respiratory Monitoring

Dräger Babylog VN500 Neonatal Open Care

Dräger Savina 300 Select ICU Ventilation and Respiratory Monitoring

Dräger Savina 300 Select ICU Ventilation and Respiratory Monitoring

Oxylog VE300 Emergency & Transport Ventilation

Dräger Savina Sub-Acute Care Ventilation

Dräger Oxylog 3000 plus Emergency & Transport Ventilation

D Beyond essentials DRÄGER SAVINA 300 SELECT

Dräger Carina Sub-Acute Care Ventilation

Dräger Carina Sub-Acute Care Ventilation

Oxylog 2000 plus Emergency & Transport Ventilation

D Transport ventilation without compromise DRÄGER OXYLOG 3000 PLUS

Focused on essentials DRÄGER SAVINA 300

Oxylog 3000 plus Emergency & Transport Ventilation

D Beyond essentials DRÄGER SAVINA 300 SELECT

EvitaXL. Excellence throughout the ventilation process. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Transport ventilation without compromise DRÄGER OXYLOG 3000 PLUS

Evita 2 dura. High-end ventilation in your patient s best interests. Because you care

MT Excellent performance and reliability EVITA 4 EDITION

Oxylog 2000 plus. Step up your performance. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

D Designed for non-invasive ventilation DRÄGER CARINA

Dräger NITROX 200 Diving Equipment

Dräger coaxial breathing circuits Consumables and Accessories

Focused on essentials DRÄGER SAVINA 300

Dräger RZ 7000 Test Equipment

Setting new trends in ventilation. DRÄGER EVITA INFINITY V500

D Quality and Precision you can Trust in Oxygen Administration CONVENTIONAL FLOWMETER DIGITAL FLOWMETER COMPACT FLOWMETER

Primus Anaesthesia Workstations

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

Dräger Perseus A500 Anaesthesia Workstations

Technical Data and Specifications

MT Your customized anesthesia workstation DRÄGER FABIUS GS PREMIUM. Dräger. Technology for Life

D The plus that makes all the difference DRÄGER FABIUS PLUS

Made to meet your needs DRÄGER SAVINA 300

Your customized anesthesia workstation DRÄGER FABIUS GS PREMIUM. Dräger. Technology for Life

Resuscitaire Neonatal Open Care

D The plus that gives you more DRÄGER FABIUS PLUS

Ventilation modes in intensive care

Better to have one OXYLOG 3000

D Accessories that actively cooperate DRÄGER INFINITY ID-ACCESSORIES

Training Benzene Awareness. Basic Information

Update to RS-232 commands. Changing patient from NIV to INVASIVE Vent Type. SNDF command

V8800 Ventilator System. Your healthcare, we care.

EVE. The mobile Ventilation Concept from Emergency to Intensive Care. easy - light - smart. Ventilation

Fabius Tiro TM. Designed for your environment. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

MT The secret of harmonious breathing BABYLOG 8000 PLUS

Fabius plus. The plus that makes all the difference. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Evita 2 dura. First-class ventilation in your patient s best interests. Because you care. Emergency Care OR/Anesthesia Critical Care Home Care

D Rising up to the challenge OXYLOG 3000 PLUS

Dräger PAS Micro Short Term Breathing Apparatus

Dräger PARAT 3260 Filtering Escape Device

Dräger MSI P7 Combustion Gas Analysis

Dräger Fabius GS premium Anaesthesia Workstations

VentStar Oxylog Breathing Systems Consumables and Accessories

Fabius plus. The plus that makes all the difference. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Dräger Secor 7000 Diving Equipment and Systems

Panther 5 Acute Care Ventilator

D The specialist for every environment DRÄGER FABIUS TIRO

Dräger Fabius plus XL Anaesthesia Workstations

INTENSIVE CARE VENTILATORS

Dräger X-pect 8500 Goggles Protective Eyewear

Dräger OCEANUS 140 Mobile Compressor

Dräger Fabius Tiro Anaesthesia Workstations

OxyLine The customised pressure reducer

Dräger Aerotest Navy Test Equipment

Understanding and comparing modes of ventilation

Dräger CAPITANO 140 Mobile Compressor

D Your personal anesthesia assistant DRÄGER PRIMUS. Dräger. Technology for Life

Vamos / Vamos plus Anaesthetic Gas Monitors

WILAmed INTENSA. Intensive Care Ventilator System

O 2 Star - Oxygen Therapy Accessories Consumables and Accessories

SERVO-air ventilator. Datasheet for system version 2.1

Dräger-Tube pump accuro Dräger Tube Pump

Dräger Quaestor 7000 Test Equipment

DATA SHEET. VENTILATION SERVO-s

fabian HFO Our 4-in-1 device

HAMILTON-MR1. Technical specifications HAMILTON-MR1

Servo UNIVERSAL ADVANCED VENTILATION PLATFORM ALL PATIENT CATEGORIES

Fabius Tiro TM. Designed for your environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Dräger ABIL-L Airline Equipment

D Reliability goes XL

Dräger FGT 5400 Diving Equipment

Dräger Tactical Harness Versatile carrying harness for LAR rebreathers.

SERVO-s v8.0 Data sheet

Dräger Short-term Tubes Dräger-Tubes

Increase patient outcome and therapeutic benefits Dedicated NIV accessories for Carina CARINA ACCESSORIES

Dräger X-plore 9000 and PAS X-plore Airline breathing equipment

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes

Dräger PA91 plus Standard and PA91 plus Marine Self Contained Breathing Apparatus

Oxylog Better to have one. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Dräger Compressed Air Breathing Cylinders Accessories for Breathing Apparatus

Dräger Bodyguard 7000 Warning Device

Mechanical Ventilation

Dräger Bodyguard 1000 Warning Device

Dräger PAS AirPack 2 Compressed Air Breathing Apparatus

Anesthesia Workstation

Dräger Saver PP Compressed Air Escape

Technology Insights for protective ventilation and spontaneous breathing in the OR

Transcription:

Dräger Evita Infinity V500 ventilator ICU Ventilation and Respiratory Monitoring Combine fully-featured, high-performance ventilation with Infinity Acute Care System integration to meet the challenges of today s health care environment MT-6073-2008

02 Dräger Evita Infinity V500 ventilator Benefits Tools for your ventilation therapy I Lung diagnostic tools like the Low Flow maneuver (inflection points) Breath by breath recruitment trends (eg PEEP, EIP, VT, Cdyn) Recruitment tools (eg lnspiration Hold, QuickSet, PressureLink) PC-APRV with AutoRelease Volumetric CO 2 -Monitoring (VCO 2, VTCO 2, Slope Phase 3, Vds/VTe) Weaning parameter (eg RSBi, P01, NIF) Automatic weaning with SmartCare/PS Tools for your ventilation therapy II Variable Pressure Support, Proportional Pressure Support, Automatic Tube Compensation Graphical representation of airway resistance and lung compliance with Smart Pulmonary View Room-to-breathe concept (AutoFlow, BIPAP, VG) Applicable for neonatal ventilation (smallest tidal volume: 2 ml) Functions to support your workflow Non-invasive ventilation in all modes and all patient categories O 2 -Therapy allows flow constant oxygen application Flexible screen configuration: 6 different views for each patient related to the individual therapy Full record of all patient data, alarms and trends Data export via USB interface Context-sensitive help function and online Instruction For Use for ventilation modes, alarms and device functions RFID functionality eg to monitor exchange intervals and to transport ventilation settings

Dräger Evita Infinity V500 ventilator 03 Accessories Infinity ID-Accessories Each and every Infinity ID-accessory has been designed to offer additional functionality, which can help you simplify routine tasks, streamline workflow and increase safety levels D-14586-2009

04 Dräger Evita Infinity V500 ventilator Patient type Adults, pediatric patients, neonates Ventilation settings Ventilation mode Volume controlled ventilation: VC-CMV VC-SIMV VC-AC VC-MMV Pressure controlled ventilation: PC-CMV PC-BIPAP 1 / SIMV+ PC-SIMV PC-AC PC-APRV PC-PSV Support of spontaneous breathing: SPN-CPAP/PS SPN-CPAP/VS SPN-CPAP SPN-PPS Enhancements AutoFlow / Volume Guarantee (VC-AC) Variable Pressure Support Smart Pulmonary View Automatic Tube Compensation (ATC) Mask Ventilation (NIV) SmartCare /PS 20 Automated clinical protocol in SPN-CPAP/PS Low Flow PV Loop O 2 -therapy Special procedures Suction maneuver Manual inspiration/hold Medication nebulization P01 PEEPi NIF Therapy types Invasive ventilation (Tube) Non-invasive ventilation (NIV) O 2 -therapy Ventilation frequency (RR) Inspiration time (Ti) Tidal volume (VT) Inspiratory flow (Flow) Adult 05 to 98/min Pediatric patients, Neonates 05 to 150/min Adults 011 to 10 s Pediatric patients, Neonates 01 to 10 s Adults 01 to 30 L Pediatric patients 002 to 03 L Neonates 0002 to 01 L Adults 2 to 120 L/min Pediatric patients, Neonates 2 to 30 L/min Inspiratory pressure (Pinsp) 1 to 95 mbar (or hpa or cmh 2 O) Inspiratory pressure limit (Pmax) 2 to 100 mbar (or hpa or cmh 2 O) PEEP / intermittent PEEP (ΔintPEEP) 0 to 50 mbar (or hpa or cmh 2 O) Pressure assist (Psupp) 0 to 95 mbar (or hpa or cmh 2 O)

Dräger Evita Infinity V500 ventilator 05 Rise time for pressure assist (Slope) O 2 concentration (FiO 2 ) Trigger sensitivity (Flow trigger) Automatic Tube Compensation (ATC) Airway Pressure Release Ventilation (PS-APRV) Inspiratory time (Thigh) Expiratory time (Tlow) Maximum time of low pressure level in APRV/PEF (Tlow max) Inspiratory pressure (Phigh) Expiratory pressure (Plow) Termination criterion (peak expiratory flow) (Exp term) Adults, Pediatric patients 0 to 2 s Neonates 0 to 15 s 21 to 100 Vol% 02 to 15 L/min Inside tube diameter tube Ø Endotracheal tube ET Adults 5 to 12 mm (02 to 047 inch) Pediatric patients 2 to 8 mm (008 to 031 inch) Neonates 2 to 5 mm (008 to 02 inch) Tracheostoma tube (Trach) Adults 5 to 12 mm (02 to 047 inch) Pediatric patients 25 to 8 mm (01 to 031 inch) Degree of compensation 0 to 100 % 01 to 30 s 005 to 30 s 005 to 30 s 1 to 95 mbar (or hpa or cmh 2 O) 0 to 50 mbar (or hpa or cmh 2 O) 1 to 80 % (PEF) Proportional Pressure Support (SPN-PPS) Flow Assist (Flow Assist) Adults 0 to 30 mbar/l/s (or hpa/l/s or cmh 2 O/L/s) Pediatric patients 0 to 100 mbar/l/s (or hpa/l/s or cmh 2 O/L/s) Neonates 0 to 300 mbar/l/s (or hpa/l/s or cmh 2 O/L/s) Volume Assist (Vol Assist) Adults 0 to 100 mbar/l (or hpa/l or cmh 2 O/L) corresponds to compliance compensation 10,000 to 10 ml/mbar (or ml/hpa or ml/cmh 2 O) Pediatric patients 0 to 1,000 mbar/l (or hpa/l or cmh 2 O/L) corresponds to compliance compensation 10,000 to 1 ml/mbar (or ml/hpa or ml/cmh 2 O) Neonates 0 to 4,000 mbar/l (or hpa/l or cmh 2 O/L) corresponds to compliance compensation 1,000 to 03 ml/mbar (or ml/hpa or ml/cmh 2 O) O 2 -therapy Leakage compensation Displayed measured values Airway pressure measurement Flow Measurement Minute volume measurement Continuous Flow 2 to 50 L/min O 2 concentration FiO 2 21 to 100 Vol% On/Off On: full compensation active Off: only trigger compensation active Plateau pressure (Pplat) Positive end-expiratory pressure (PEEP) Positive end-expiratory pressure (PIP) Mean airway pressure (Pmean) Minimum airway pressure (Pmin) Range -60 to 120 mbar (or hpa or cmh 2 O) Expiratory minute volume (MVe) Inspiratory minute volume (MVi) Total minute volume (MV) Mandatory minute volume (MVmand)

06 Dräger Evita Infinity V500 ventilator Tidal volume measurement Respiratory rate measurement Spontaneous expiratory minute volume (MVspon) Range 0 to 99 L/min BTPS Tidal Volume (VT) Inspiratory tidal volume (not leakagecompensated) of mandatory breaths (VTimand) Expiratory tidal volume (not leakagecompensated) of mandatory breaths (VTemand) Inspiratory tidal volume (not leakagecompensated) of spontaneous breaths (VTispon) Range 0 to 5,500 ml BTPS Breathing frequency (RR) Mandatory respiratory rate (RRmand) Spontaneous breathing frequency (RRspon) Range 0/min to 300/min O 2 measurement (inspiratory side) Inspiratory O 2 concentration (FiO 2 ) CO 2 measurement in main flow (adults and pediatric patients only) Displayed calculated values Range 18 to 100 Vol% End-expiratory CO 2 concentration (etco 2 ) Range 0 to 100 mmhg Compliance (C) Range 0 to 650 ml/mbar (or ml/cmh 2 O) Resistance (R) Leakage minute volume (MVleak) Rapid Shallow Breathing (RSB) Range 0 to 1,000 mbar / (L/s) (or cmh 2 O / (L/s)) Range 0 to 99 L/min Range 0 to 9999 (/min/l) Negative Inspiratory Force (NIF) Range -80 mbar to 0 mbar (or hpa or cmh 2 O) Occlusion pressure P01 Range -60 to 130 mbar (or hpa or cmh 2 O) Curve displays Airway pressure Paw (t) -30 to 100 mbar (or hpa or cmh 2 O) Flow (t) -180 to 180 L/min Volume V (t) 2 to 3,000 ml Exp CO 2 concentration (etco 2 ) 0 to 100 mmhg Alarms / Monitoring Expiratory minute volume (MVe) Airway pressure (Paw) Inspiratory O 2 concentration (FiO 2 ) End-expiratory CO 2 concentration (etco 2 ) Tachypnoea monitoring (RR) Volume monitoring (VT) Apnea alarm time (Tapn) Disconnect alarm delay time (Tdisconnect) Performance data Control principle Intermittent PEEP duration Medicament nebulisation Inspiratory flow Base flow, adults Base flow, pediatric patients Base flow, neonates Safety valve High 5 to 60 seconds 0 to 60 seconds time-cycled, volume-constant, pressure-controlled 1 to 20 expiratory cycles for 5, 10, 15, 30 minutes max 180 L/min 2 L/min 3 L/min 6 L/min Opens if medical compressed air supply fails(supply gas flow is not sufficient to provide the inspiratory flow required), enables spontaneous breathing with ambient air

Dräger Evita Infinity V500 ventilator 07 Endotracheal suction Disconnection detection automatic Reconnection detection automatic Initial Oxygen enrichment max 3 minutes Active suction phase max 2 minutes Final oxygen enrichment max 2 minutes Factor for pediatric patients and neonates 1 to 2 Supply system for spontaneous breathing and Psupp adaptive CPAP system with high initial flow Operating data Mains power supply Mains power connection Current consumption at 230 V at 100 V Inrush current Power consumption maximum during ventilation, without charging the battery 100 V to 240 V, 50/60 Hz max 14 A max 30 A approx 8 to 24 A peak approx 6 to 17 A quasi RMS 300 W approx 100 W ventilation unit with Medical Cockpit approx 180 W with GS500 Digital machine output Gas supply O 2 gauge pressure Air gauge pressure Digital output and input via an RS232 C interface Dräger MEDIBUS and MEDIBUSX 27 to 60 bar (or 270 to 600 kpa or 39 to 87 psi) 27 to 60 bar (or 270 to 600 kpa or 39 to 87 psi) Physical Specifications Dimensions (W x H x D) Ventilation unit with lateral standard rail (without Infinity C500) Ventilation unit and Infinity C500 on the trolley, carrier frame without bar Ventilation unit and Infinity C500 on the trolley, carrier frame with bar 361 mm x 320 mm x 410 mm (143 in x 126 in x 161 in) 577 mm x 1,420 mm x 687 mm (227 in x 553 in x 271 in) 577 mm x 1,405 mm x 700 mm (227 in x 553 in x 276 in) Weight Evita Infinity V500 and Infinity C500 Evita Infinity V500 and Infinity C500 on trolley PS500 approx GS500 approx Mounting: Supporting frame Adapter for 38 mm pole approx 25 kg (551 lbs) approx 59 kg (130 lbs) approx 27 kg (595 lbs) approx 105 kg (23 lbs) 165 kg (364 lbs) 235 kg (518 lbs) Infinity C500 Diagonal screen size Infinity C500 17" TFT color touch screen Input / Output ports (at Infinity C500) 3 external RS232 (9-pin) connectors 6 USB ports for data collection 3 DVI for digital video output 1 LAN port 2 RJ 45 Ethernet connectors

08 Dräger Evita Infinity V500 ventilator 1 BIPAP, trademark used under license ATC, trademarked by Dräger AutoFlow, trademarked by Dräger BTPS Body Temperature Pressure Saturated Measured values relating to the conditions of the patient lung (986 F), steamsaturated gas, ambient pressure 1 mbar = 100 Pa Some functionalities are available as an option Not all products, features, or services are for sale in all countries Mentioned Trademarks are only registered in certain countries and not necessarily in the country in which this material is released Go to wwwdraegercom/trademarks to find the current status 90 51 753 1707-5 HQ PP Subject to modifications 2017 Drägerwerk AG & Co KGaA CORPORATE HEADQUARTERS Drägerwerk AG & Co KGaA Moislinger Allee 53 55 23558 Lübeck, Germany wwwdraegercom Manufacturer: Drägerwerk AG & Co KGaA Moislinger Allee 53-55 23558 Lübeck, Germany REGION DACH Drägerwerk AG & Co KGaA Moislinger Allee 53 55 23558 Lübeck, Germany Tel +49 451 882 0 Fax +49 451 882 2080 info@draegercom REGION EUROPE Drägerwerk AG & Co KGaA Moislinger Allee 53 55 23558 Lübeck, Germany Tel +49 451 882 0 Fax +49 451 882 2080 info@draegercom REGION MIDDLE EAST, AFRICA Drägerwerk AG & Co KGaA Branch Office PO Box 505108 Dubai, United Arab Emirates Tel +971 4 4294 600 Fax +971 4 4294 699 contactuae@draegercom REGION ASIA PACIFIC Draeger Singapore Pte Ltd 25 International Business Park #04-20/21 German Centre Singapore 609916 Tel +65 6308 9400 Fax +65 6308 9401 asiapacific@draegercom REGION NORTH AMERICA Draeger, Inc 3135 Quarry Road Telford, PA 18969-1042, USA Tel +1 800 4DRAGER (+1 800 437 2437) Fax +1 215 723 5935 infousa@draegercom REGION CENTRAL AND SOUTH AMERICA Dräger Panama S de RL Complejo Business Park, V tower, 10th floor Panama City Tel +507 377-9100 Fax +507 377-9130 contactcsa@draegercom Locate your Regional Sales Representative at: wwwdraegercom/contact