FPD212 Dual-chamber orifice fitting

Similar documents
ATS430 turbidity sensor Retractable insertion assembly

Un-Pressurized Orefice Fittings FIO EZ. Parts List and Operation Instructions TECHNICAL MANUAL. Dn 2-6 Class Lbs

FPD175 Compact orifice carrier assembly

Endura AZ40 Sensor assembly High temperature filter assembly kit

WHEATLEY Series 822/820 Swing Check Valve

WHEATLEY Series 500 Swing Check Valve

HOW TO NOT MEASURE GAS - ORIFICE. Dee Hummel. Targa Resources

KBV21i and KBV40i Air Actuated Boiler Blowdown Valves

INSTALLATION, MAINTENANCE & OPERATING INSTRUCTIONS 2-4 REDUCED PORT/ FULL PORT (5700/6700) ANSI CLASS 150/300/600/900/1500/2500 TRUNNION BALL VALVES

Endura AZ40 Sensor assembly Smart electronics PCB

KBV21i and KBV40i Key Operated Boiler Blowdown Valves Installation and Maintenance Instructions

BSA6T and BSA64T Stainless Steel Bellows Sealed Stop Valves Installation and Maintenance Instructions

KTM OM-2 SPLIT BODY FLOATING BALL VALVES INSTALLATION AND MAINTENANCE INSTRUCTIONS

WHEATLEY WHEATLEY SERIES 500 SWING CHECK VALVE. Installation, Operation and Maintenance Manual

Apollo Standard Port, Full Port & One Piece Flanged Ball Valves Installation, Operation, & Maintenance Manual

FLANGED MULTI-PORT BALL VALVES

INSTALLATION COMMISSIONING, OPERATION & MAINTENANCE MANUAL

Serie 06-M6. Swing wafer check valve. made in. Application fields. Check valves E U R O P E WATER CONDITIONING INDUSTRY

Eaton Filtration, LLC

Eaton Filtration, LLC

Contents. 1. General information BROEN BALLOMAX Steel Ball Valves Approvals Quality Management...2

PV4 and PV6 Piston Valves

FLANGED TWO-PIECE BALL VALVES

FLANGED TWO-PIECE BALL VALVES

1200B2 Series Service Regulators. Instruction Manual

1800C and 1800C-HC Series Service Regulators

APCO ARV CLEAN WATER AIR RELEASE VALVES. Model 50A

TITAN FLOW CONTROL, INC.

Bray/ VAAS O-Ported Series Knife Gate Valve 770/780 Series Operation and Maintenance Manual

9408 Series Dissolved Oxygen Systems. Dissolved oxygen systems designed for reliable operation with minimal operating costs

Engineering Data Sheet

USM21 Sealed Bimetallic Steam Trap for use with Pipeline Connectors Installation and Maintenance Instructions

KECKLEY. Installation, Operation, and Maintenance Manual. Style KGV - Knife Gate Valve

MODEL 200 KNIFE GATE VALVES INSTALLATION & MAINTENANCE MANUAL

M10HTi ISO Tobacco Ball Valve Screwed, SW, BW and Flanged versions

ABB MEASUREMENT & ANALYTICS DATA SHEET. AP100 ph/redox (ORP) cartridge sensors with water-wash option

Installation & Operation Manual Proven Quality since 1892

SPECIFICATION // DATA SHEET WATER DRAIN VALVE AM SENSORS LTD AMS-S-084/DS/ (ISSUE 1)

Installation, Operation and Maintenance Manual for Back Pressure Regulator

Spiratec ST14, ST16 and ST17 Sensor Chambers and sensors

3-PIECE BALL VALVE, 3600 PSI/ PN 248, WITH ISO DIRECT MOUNTING PAD 306M SERIES/ PED Category II

FTGS14 Ball Float Steam Trap DN15 (½") to DN25 (1")

Installation, Operating and Maintenance Instructions. V914 Cast Iron Flanged Swing Check Valve V914

M33F ISO, M33S ISO and M33V ISO Ball Valves Installation and Maintenance Instructions

IFTGS14 Ball Float Steam Trap ½" and ¾" with Integral Spiratec Sensor

M45 ISO Ball Valve DN25 to 150 Installation and Maintenance Instructions

Models 461-S, 461-8S and S Regulators. R-1330 Rev. 7

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions. IMPORTANT NOTICE

RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL

WKM Model 320F Floating Ball Valve

GREIG FILTERS, INC. OPERATION MANUAL

1.0 - OPENING AND CLOSING THE DOOR

Regulator from 1800 B series

TriPod Safety Coupler Installation & Operating Instructions for Models TP4, TP5, TP6, and TP7

Combination Air Valve

CA44, CA44S, CA46 and CA46S Air and Gas Traps

APT14, APT14HC and APT14SHC Spares 2

Type BBS-03, BBS-05, BBS-06, BBS-25

AP100 series ph/redox (ORP) cartridge sensors with water-wash option

Wedge Type Flow Meter Installation and Operation Manual

299H Series. Introduction. P.E.D. Categories. Specifications. Installation. Warning. Installation Guide English September 2012

INSTALLATION, OPERATION and MAINTENANCE MANUAL

Model PIV UL/FM Post Indicator Valve. Maintenance and Operation Manual UL/FM Post Indicator Valve Configuration AWWA C515

DelVal Flow Controls Private limited

Endura AZ40 Transmitter Main board spares kit

CA10S, CA14 and CA14S Air and Gas Traps

Telefon (+45) Telefax (+45)

Installation, operation & maintenance manual - original version

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

TP1 and TP2 Temporary Cone Shaped Strainers

ABB MEASUREMENT & ANALYTICS DATA SHEET series Dissolved oxygen systems

Installation and Service Manual

Sempell High Pressure Forged Check Valves Forged steel, swing check and tilting disc configuration

Types S100K and S102K Pressure Regulators

UIB30 and UIB30H Sealed Inverted Bucket Steam Traps for use with Pipeline Connectors

Installation Operation Maintenance. Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow IOM.

INSTALLATION, OPERATION AND MAINTENANCE GUIDE

INSTALLATION, OPERATING & MAINTENANCE MANUAL WATER DRAIN VALVE

Manual Actuated Boiler Blowdown Valves

MSC-P and MSC-N Manifolds for Steam Distribution and Condensate Collection

A3S Bellows Sealed Stop Valve Installation and Maintenance Instructions

Instruction Manual Updated 7/26/2011 Ver. 2.2

SA Control Valve types: BM, BMF, BX, SB, NS, KA, KB, KC (Normally open) BMRA, BMFRA, BXRA, SBRA, NSRA, KX, KY (Normally closed)

TD62LM and TD62M Thermodynamic Steam Traps with Replaceable Seats

Operating Procedures for GripTight 15.5 SDR & IPS Test Plugs

ROTATING DISK VALVES INSTALLATION AND MAINTENANCE 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6

COMBINATION AIR RELEASE DEGASSING (CARD) VALVES INSTALLATION AND MAINTENANCE MANUAL

130400A 1/2 NPT Female A 3/4 NPT Female A 1 NPT Female A 1 1/4 NPT Female A 1 1/2 NPT Female A 2 NPT Female

Spirax SafeBloc TM DBB3 Double Block and Bleed Bellows Sealed Stop Valve. 1. Safety information. 2. General product information. 3.

Differential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN

ANDERSON GREENWOOD SERIES 9000 POSRV INSTALLATION AND MAINTENANCE INSTRUCTIONS

MST21 Stainless Steel Balanced Pressure Thermostatic Steam Trap

Installation / Operation / Maintenance Manual KLINGER Reflex Level Gauge Type R 25

1700ESL Kamvalok Coupler

Constant Pressure Crude Oil Container Model CPCCP

Needle valve. Contents. User s Manual. (1) Be sure to read the following warranty clauses of our product 1. (2) General operating instructions 2

Type 2000 INOX. Quickstart. English Deutsch Français. 2/2-way angle seat valve 2/2-Wege Schrägsitzventil Vanne à siège incliné 2/2 voies

THE PROCESS OF JOINT INTEGRITY

TITAN FLOW CONTROL, INC.

S265 Remote seals with capillary tube. 2600T Series Pressure Transmitters. Engineered solutions for all applications

Transcription:

Operating instructions OI/FPD212 EN FPD212 Dual-chamber orifice fitting Measurement made easy Introduction ABB's FPD212 dual-chamber orifice fitting is a proven system for safely and quickly inspecting and changing orifice plates in conformance with strict Oil & Gas industry guidelines. For more information Further publications for the FPD212 dual-chamber orifice fitting are available for free download from www.abb.com/measurement or by scanning this code: These operating instructions provide installation, operation and maintenance procedures for the FPD212 dual-chamber orifice fitting.

Contents 1 Health & Safety...3 1.1 Document symbols... 3 1.2 Safety precautions... 3 1.3 Potential safety hazards... 3 1.4 Product recycling and disposal (Europe only)... 3 1.5 Lifting and handling... 4 1.5.1 Lifting / moving products in crates and on pallets... 4 1.5.2 Lifting / moving meter runs... 5 1.5.3 Lifting / moving orifice fittings... 6 2 Operation...7 3 Key operating parts...8 4 Orifice plate removal...9 5 Orifice plate inspection...11 6 Orifice fitting quick inspection and assessment...12 7 Scheduled teardown maintenance...12 8 Orifice plate replacement...13 9 Greasing procedure...16 10 Preservation and storage...16 11 Specification...17 Notes...18 2 OI/FPD212 EN FPD212 Dual-chamber orifice fitting

1 Health & Safety 1.1 Document symbols Symbols that appear in this document are explained below: DANGER Serious damage to health / risk to life This symbol in conjunction with the signal word 'DANGER' indicates an imminent danger. Failure to observe this safety information will result in death or severe injury. WARNING Bodily injury This symbol in conjunction with the signal word 'WARNING' indicates a potentially dangerous situation. Failure to observe this safety information may result in death or severe injury. 1.2 Safety precautions Be sure to read, understand and follow the instructions contained within this manual before and during use of the equipment. Failure to do so could result in bodily harm or damage to the equipment. WARNING Bodily injury Pressurized equipment Installation, operation, maintenance and servicing of pressurized equipment must be performed: by suitably trained personnel only in accordance with the information provided in this manual in accordance with relevant local regulations 1.3 Potential safety hazards CAUTION Minor injuries This symbol in conjunction with the signal word 'CAUTION' indicates a potentially dangerous situation. Failure to observe this safety information may result in minor or moderate injury. The symbol may also be used for property damage warnings. NOTICE Property damage This symbol indicates a potentially damaging situation. Failure to observe this safety information may result in damage to or destruction of the product and / or other system components. WARNING Bodily injury To ensure safe use when operating this equipment, the following point must be observed: Orifice fittings and their components present both heavy lifting and tip-over hazards. Operators must wear suitable PPE at all times and have strategies in place for safely lifting, moving and storing orifice fittings and their components. Safety advice concerning the use of the equipment described in this manual or any relevant Material Safety Data Sheets (where applicable) can be obtained from the Company, together with servicing and spares information. IMPORTANT (NOTE) This symbol indicates operator tips, particularly useful information or important information about the product or its further uses. The signal word 'IMPORTANT (NOTE)' does not indicate a dangerous or harmful situation. NOTICE Property damage The soft-seat valve seal design does not require lubrication through the grease port. Attempting to lubricate the FPD212 orifice fitting equipped with a soft-seat inner slide valve causes the inner O-ring seal to be ejected from the dovetail groove resulting in valve failure. 1.4 Product recycling and disposal (Europe only) IMPORTANT (NOTE) For return for recycling, please contact the equipment manufacturer or supplier for instructions on how to return end-of-life equipment for proper disposal. FPD212 Dual-chamber orifice fitting OI/FPD212 EN 3

1.5 Lifting and handling This product requires mechanical lifting devices and techniques. All safety activities associated with lifting techniques and equipment must be in accordance with local laws and the receiving company's HSE policies and procedures, and shall be followed without exception. Extreme care must be exercised throughout the operation to prevent both injury to personnel and damage to the meter. Proper rigging and lifting techniques must be followed. Do not stand under suspended loads. Improper handling may cause injury and / or damage. SAFETY FIRST. This product is delivered to its destination in a wooden crate or strapped to a wooden pallet and is easily moved using a standard forklift. There are scenarios, however, where a meter run or fitting must be moved without the aid of a frame. In both cases, care must be taken to assess the weight of the package and ensure that the machinery and lifting aids are appropriate to the items being handled. Crate, pallet and / or meter run weights are documented on the commercial invoice or packing slip accompanying the shipment. 1.5.1 Lifting / moving products in crates and on pallets Select an appropriate forklift based on the weight of the package being handled. Use fork extensions if necessary. Center the forks so the load is evenly distributed. Ensure the load is balanced and secure prior to lifting. Use extra strapping to secure unstable loads. Drive the forks into the load as far as possible. Avoid contact with any parts of the load that may extend past the edge of the pallet. Tilt the load back slightly, then lift. Keep the load as low as possible while traveling, keep an even pace and avoid fast starts and sudden stops. Fig. 1.1 Product packaging 4 OI/FPD212 EN FPD212 Dual-chamber orifice fitting

1.5.2 Lifting / moving meter runs Select an appropriate forklift and lifting straps based on the weight of the meter run being handled. Position the forks as narrowly as possible to create a single lifting point. Determine the meter run s center of gravity and position the lifting straps as shown in Fig. 1.2, ensuring the meter run is balanced when lifted. Ensure the lifting straps are wrapped in opposing directions to prevent the meter run from rotating and swinging. Slide the ends of the lifting straps over the forks, positioning them as close to the forklift body as possible, while ensuring the meter run does not make contact with the forklift. Tilt the load back slightly, then lift. Keep the forks as low as possible while traveling, keep an even pace and avoid fast starts and sudden stops. LIFT Note. Ensure lifting straps wrap around pipe in opposing directions. 45 Center of gravity Lifting strap Lifting point Fig. 1.2 Lifting meter runs IMPORTANT (NOTE) Fig. 1.2 does not reflect all configurations. Centers of gravity may not be exactly as shown. FPD212 Dual-chamber orifice fitting OI/FPD212 EN 5

1.5.3 Lifting / moving orifice fittings Select an appropriate forklift and lifting strap based on the weight of the fitting being handled. Position the forks as narrowly as possible to create a single lifting point. Wrap the lifting strap around the fitting, under the shoulder of the upper chamber as shown in Fig. 1.3. Slide the end of the lifting strap over the forks, positioning the strap as close to the forklift body as possible while ensuring the fitting does not make contact with the forklift. Adjust as necessary. Tilt the load back slightly, then lift. Keep the forks as low as possible while traveling, keep an even pace and avoid fast starts and sudden stops. LIFT Center of gravity Lifting strap Lifting point Fig. 1.3 Lifting orifice fittings IMPORTANT (NOTE) Fig. 1.3 does not reflect all configurations. Centers of gravity may not be exactly as shown. 6 OI/FPD212 EN FPD212 Dual-chamber orifice fitting

2Operation The FPD212 dual-chamber orifice fitting enables accurate placement of an orifice plate in a pipeline, together with easy removal and replacement, without interrupting the flow or dismantling the pipeline. The flow of a gas or liquid through the orifice plate creates the differential pressure that is measured and used to calculate the flow rate through the pipeline. Referring to Fig. 2.1, the orifice fitting is in normal metering mode when: the plate carrier and orifice plate are in the lower chamber, concentric to the pipeline the inner slide valve is closed Pressure sensors installed in the lower chamber's meter taps read the differential pressure across the orifice plate and transmit the readings to a flow computer or chart recorder. Orifice plate condition is critical to accurate metering. The orifice plate must be removed and inspected for damage, general wear and particulate build-up regularly. Prior to beginning the orifice plate removal procedure, the operator must confirm that the clamping bar is tight and the bleeder valve is fully close. During normal operation, only the lower chamber is pressurized. Atmospheric pressure in the upper chamber is necessary for proper functioning of the inner slide valve. the equalizer valve is closed the bleed valve is closed the upper chamber is at atmospheric pressure the clamping bar is tight Clamping bar tight Bleed valve closed Equalizing valve closed Inner slide valve closed Orifice plate in pipeline Fig. 2.1 Normal metering mode FPD212 Dual-chamber orifice fitting OI/FPD212 EN 7

3 Key operating parts The FPD212 dual-chamber orifice fitting is operated by relatively few parts, all of which are controlled with the use of the included operating wrench. The spatial separation of the operating parts, together with the visual differences between them, reduce the chances of operator error. Clamping bar screws Clamping bar Seal bar Bleed valve Specification plate Equalizing valve Grease gun Plate carrier operator gears Flow direction arrow Inner slide valve operator gear Inner slide valve indicator plate Orifice plate carrier Orifice plate seal Operating wrench Orifice plate Drain Meter taps 8 OI/FPD212 EN FPD212 Dual-chamber orifice fitting

4 Orifice plate removal 2. Open inner slide valve. 3. Raise orifice plate carrier with operator gears. WARNING Bodily injury Ensure the clamping bar is tight and the bleed valve is fully closed before proceeding. 2 To remove the orifice plate: 1. Slowly open the equalizer valve. 1 FPD212 Dual-chamber orifice fitting OI/FPD212 EN 9

4. Close inner slide valve. 6. Open bleed valve. 7. Perform greasing procedure see Section 9, page 16. 8. Loosen clamping bar. WARNING Bodily injury Ensure inner slide valve is fully closed before proceeding. 5. Close equalizer valve. WARNING Bodily injury Ensure equalizer valve is fully closed before proceeding. 10 OI/FPD212 EN FPD212 Dual-chamber orifice fitting

9. Dislodge seal bar with plate carrier. 5 Orifice plate inspection Regular orifice plate inspections are crucial to measurement accuracy. ABB recommends that the orifice plate is removed from the orifice fitting and inspected monthly. AGA-3 / API 14.3 sections 2.4.1 and 2.4.2 outline orifice plate parameters that can be affected by service conditions. Damage, excessive wear and build-up can lead to a reduction in flow meter accuracy. Pay particular attention to dirt, grease or ice accumulation, pitting due to corrosion, nicks, dings and other impact damage, as well as general wear and erosion of the normally sharp bore edges. Replace the orifice plate as necessary. 10. Remove clamping and seal bars, rotate orifice plate carrier up through opening. 1 2 Fig. 5.1 Examples of unacceptable damage to the orifice plate Orifice plate seals must also be inspected to reduce the possibility of bypass. Pay attention to dirt or grease accumulation, cracks, tears or cuts and swelling. Replace seals as necessary. FPD212 Dual-chamber orifice fitting OI/FPD212 EN 11

6 Orifice fitting quick inspection and assessment WARNING Bodily injury DO NOT open the inner slide valve to inspect the lower chamber if the upper chamber is open and the fitting is under process pressure. When the clamping and seal bars have been removed from the orifice fitting and the upper chamber is open, visually inspect the interior of the fitting. Pay attention to the condition of the interior surfaces of the orifice fitting, the pinion gear, seal bar face and O-ring. Excessive corrosion, fluid build-up or other visible damage may indicate the need for more extensive maintenance and replacement of parts. Similarly, excessive resistance when turning gear shafts during operation may indicate the need for the replacement of parts. 7 Scheduled teardown maintenance Both measurement accuracy and orifice fitting performance in general can benefit from periodic maintenance. ABB recommends that maintenance is performed every two years or earlier if deemed necessary. 1. De-pressurize the process line. WARNING Bodily injury Ensure the process line is de-pressurized before removing the upper chamber from the orifice fitting body. Failure to de-pressurize could result in serious injury or death. 2. Remove the upper chamber. 3. Visually inspect all interior surfaces of the orifice fitting body for excessive build-up or corrosion. 4. Inspect gear shafts for broken or bent teeth. Replace as necessary. 5. Inspect valve seat surfaces for abrasion or corrosion. Replace as necessary. IMPORTANT (NOTE) Step 6 is applicable only to fittings fitted with grease-type inner slide valves. 6. Replace the valve seat grease in the grease track. 7. Replace the valve strip and all accessible O-rings (including valve seat O-rings on soft-seat units). 8. Remove the drain plugs and remove all accumulated debris from the lower chamber. Fig. 6.1 Top view of orifice fitting (clamping and seal bars removed) 9. Inspect all differential pressure tap holes for build-up. Clean as necessary. 10. Refit the upper chamber. 12 OI/FPD212 EN FPD212 Dual-chamber orifice fitting

WARNING Bodily injury Steps 11 to 14 MUST be performed before re-pressurizing the process line. Failure to follow this procedure could result in serious injury or death. 11. Ensure the orifice plate is in the upper chamber. 12. Ensure the inner slide valve is closed. 13. Ensure the equalizing and bleed valves are closed. 14. Ensure the seal and clamping bars are in place and tightened down. 8 Orifice plate replacement NOTICE Property damage When refitting the orifice plate, proper positioning of the orifice plate and plate carrier is crucial to both correct operation and metering accuracy. To refit the orifice plate: 1. Align the orifice plate in the orifice plate carrier and insert orifice plate carrier into top opening, ensuring correct positioning as shown. 15. Repressurize the process line. Fig. 7.1 Orifice fitting dismantling Ensure orifice plate bevel and orifice plate carrier gear teeth face downstream Ensure orifice plate carrier is level with top of housing FPD212 Dual-chamber orifice fitting OI/FPD212 EN 13

5. Tighten clamping bar screws. 6. Slowly open equalizer valve. 7. Open inner slide valve. NOTICE Property damage During step 2, do not lower orifice plate carrier onto inner slide valve. 2. 3. Close bleed valve. Lower orifice plate carrier into upper chamber. 4. Install seal and clamping bars. 14 OI/FPD212 EN FPD212 Dual-chamber orifice fitting

8. Lower plate carrier into bottom chamber. WARNING Bodily injury Ensure equalizer valve is fully closed before proceeding. 1 11. Open bleed valve. 2 9. Close inner slide valve. 12. Perform greasing procedure see Section 9, page 16. 13. Close bleed valve. WARNING Bodily injury Ensure inner slide valve is fully closed before proceeding. 10. Close equalizer valve. FPD212 Dual-chamber orifice fitting OI/FPD212 EN 15

9 Greasing procedure NOTICE Property damage This procedure is not applicable to orifice fittings equipped with the soft-seat valve seal design. Attempting to lubricate an FPD212 orifice fitting equipped with a soft-seat inner slide valve causes the inner O-ring seal to be ejected from the dovetail groove resulting in valve failure. The metal-to-metal grease seal design requires lubrication on a regular basis. For the procedure to be successful, a pressure differential across the inner slide valve is required, with the chamber immediately below the valve having a higher pressure and forcing the valve strip up against the seat. The double ball grease check valve prevents process pressure from coming through the grease port. For best results, use only valve seal grease supplied by ABB and recommended for the applicable service. 1. Ensure that the inner slide valve and equalizer valve are fully closed. 2. Open the bleed valve and allow the upper chamber to de-pressurize. A pressure differential now exists between the lower and upper chambers, and the inner slide valve is ready to be greased. 10 Preservation and storage Take the following measures to preserve and store orifice fittings that are not currently in service: Store in dry conditions, preferably indoors to prevent rust and corrosion. Fit the end caps supplied with the fitting. Apply rust inhibitor inside the bore every 3 to 6 months. Remove orifice plate and seals from the seal gap and ensure soft-seat inner slide valves remain in the open position during storage. Soft-seat units require no additional maintenance during storage. Do not apply valve seal grease to soft-seat units. For grease-type units, apply valve seal grease every 90 days to prevent hardening of the sealant medium. Ensure grease-type valves remain in the closed position during storage. Hydrostatic testing is required before entering service if stored for more than 1 year. Replace / re-lubricate seals as required. 3. Remove the threaded stem from the grease gun assembly. Insert one stick of grease. 4. Replace the grease gun stem and turn in slowly so as not to rupture the grease channel seal. 5. Repeat steps 3 and 4 if additional grease is required. Larger fittings require more grease. 6. Close the bleed valve. 16 OI/FPD212 EN FPD212 Dual-chamber orifice fitting

11 Specification Body configuration Flange x flange Flange x weld Weld x weld Flanges to FF, RF and RTJ are available in the following formats: ANSI 16.5 or 16.47 Techlok and SPO Grayloc API 6A or to specific client specification Design codes ASME B31.8 Gas Transmission And Distribution Piping NACE MR-01-75 ASME B31.1 Power Piping ASME B31.3 Liquid Petroleum Transportation Piping Systems PED 97/23/EC Pressure Equipment Directive Measurement standards ISO 5167 API MPMS 14.3.2 (AGA Report No. 3) Standard materials of construction Body casting ASTM A216 WCP, WCC ASTM A352 LCC ASTM A351 CF8M ASTM A995 Gr.4A, Gr.6A Other materials are available on request Internal components Stainless steel 316 Stainless steel A351 CF8M Carbon steel AISI430 Pressure tappings 1 or 2 sets per side; 1 /2 in. NPT female Seals Seal bar HNBR O-ring (standard); gasket (optional) Shafts PTFE packing (standard); HNBR O-ring (optional) Inner valves Grease seal (standard); HNBR O-ring (optional) Orifice plate Type 'K' 2000 edition formed HNBR seal on a 316 SS retainer ring Dual ring HNBR O-rings standard on a 316 SS retainer ring PTFE snap seal two-piece virgin PTFE assembly Orifice plates Industry-standard orifice plate thicknesses as per the guidelines set out within ISO 5167 and API 14.3. Maximizing performance To ensure absolute metering accuracy, it is recommended that industry practice is followed and meter tubes are purchased and manufactured to complement the orifice fitting. This ensures concentricity between the pipe and the fitting and eliminates uncertainties associated with misalignment. ABB can supply the following items for a complete metering solution: orifice fitting with meter tube including straightening vane / profiler DP transmitters and process transmitters flow computer and enclosure spare parts for service and commissioning product service training Temperature rating As ratings of material and flanges fitted Pressure rating Size in. (mm) ANSI pressure class (fully-rated) 2 (50) to 12 (300) 150, 300, 600, 900, 1500, 2500 14 (350) to 24 (600) 150, 300, 600, 900, 1500 Note. For other sizes, contact ABB. External bolting ASTM A194 Gr.4 (Zinc coated) ASTM A320 L7 (Zinc coated) Other materials are available on request, including optional coatings such as PTFE, Xylan and Xylar DS/FPD212-EN Rev. A FPD212 Dual-chamber orifice fitting OI/FPD212 EN 17

Notes 18 OI/FPD212 EN FPD212 Dual-chamber orifice fitting

FPD212 Dual-chamber orifice fitting OI/FPD212 EN 19

Contact us ABB Limited Process Automation Oldends Lane Stonehouse Gloucestershire GL10 3TA UK Tel: +44 (0)1453 826661 Fax: +44 (0)1453 829671 ABB Inc. Process Automation 125 E. County Line Road Warminster PA 18974 USA Tel: +1 215 674 6000 Fax: +1 215 674 7183 Note We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document. We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB. Copyright 2015 ABB All rights reserved Sales Service OI/FPD212 EN 05.2015 www.abb.com