Compressive force applied to a manikin s head during mask ventilation

Similar documents
Changing gas flow during neonatal resuscitation: a manikin study

Breathing during mask ventilation of preterm infants at birth

Neonatal resusitation 1: A model to measure inspired and expired tidal volumes and assess leak at the face mask

I nternational consensus 2and guidelines from

Neonatal resuscitation 2: An evaluation of manual ventilation devices and face masks

Operating & troubleshooting a self inflating bag. Victorian Newborn Resuscitation Project Updated February 2012

Neonatal tidal volume targeted ventilation

Learning objectives. First Response Learning Module 2 Based on ILCOR and ANZCOR 2016

First Response Learning Module 2 Based on ILCOR and ANZCOR Victorian Newborn Resuscitation Project Updated March 2018

Unit 15 Manual Resuscitators

Inspire rpap REVOLUTION FROM THE FIRST BREATH

First Response & Advanced Resuscitation Learning Modules 2 and 3 Based on ILCOR and ANZCOR 2016

Cover Page. The handle holds various files of this Leiden University dissertation

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA

birth: a transition better guidelines better outcomes the birth experience a challenging transition the fountains of life: 2/8/2018

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives

Update on chest compressions

Use of Respiratory Function Monitors during Simulated Neonatal Resuscitation

Introduction to Conventional Ventilation

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

Lung Volumes and Capacities

E C C. American Heart Association. BLS for Healthcare Providers. Written Exams. January 12, 2012

N eonatal respiratory distress syndrome and bronchopulmonary

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)


CATALOGUE. Adult and Infant Respiratory Care Catalogue

Endotracheal Suctioning: In Line ETT

R S A B C CPR. Basic Life Support Flow Chart Check for danger. Check Response. Send for Help. Check Airway. Check for Breathing.

Expired CO 2 Levels Indicate Degree of Lung Aeration at Birth

Collin County Community College. Lung Physiology

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

Volume vs Pressure during Neonatal Ventilation

First and Only Disposable Infant T-Piece Resuscitator. Because they are not miniature adults...

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams)

Principles of Mechanical Ventilation: A Graphics-Based Approach

McHENRY WESTERN LAKE COUNTY EMS SYSTEM FALL 2014 CONTINUING EDUCATION MANDATORY FOR ALL PRIMARY AND PROBATIONARY ALS SYSTEM PROVIDERS.

Trust Guideline for Neonatal Volume Guarantee Ventilation (VGV)

Physiological based management of hypoxaemic respiratory failure

NAVA Neurally Adjusted Ventilatory Assist In Neonates

E C C. American Heart Association. Basic Life Support for Healthcare Providers. Written Examinations. March 2011

RESUSCITATION. If baby very floppy and heart rate slow, assist breathing immediately. If baby not breathing adequately by 90 sec, assist breathing

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

B.L.S احیای پایھ کودکان American Heart Association

PART SEVEN THE HISTORY AND APPLICATION OF HIGH FREQUENCY OSCILLATORY VENTILATION (HFOV)

Mechanical Ventilation

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution

Breathing Devices. Chapter 8 KNOWLEDGE OBJECTIVES SKILL OBJECTIVES. 6. List four precautions to take when using oxygen.

COALINGA STATE HOSPITAL. Effective Date: August 31, 2006

Objectives. Fetal Physiology. Fetal Physiology. Resuscitation and Stabilization of the Newborn. Which Babies Require Resuscitation?

Volume Targeted Ventilation on the Neonatal Unit

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

Device overview. 6 Breathing circuit connection. Note: When the remote control foot pedal is connected, the manual switch is disabled.

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

Neonatal Assisted Ventilation. Haresh Modi, M.D. Aspirus Wausau Hospital, Wausau, WI.

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

Chapter 9 Airway Respirations Metabolism Oxygen Requirements Respiratory Anatomy Respiratory Anatomy Respiratory Anatomy Diaphragm

DOWNLOAD OR READ : VENTILATION BLOOD FLOW AND DIFFUSION PDF EBOOK EPUB MOBI

Stratégie ventilatoire pendant la RCP Pr Jean-Christophe M Richard

CHILD DEFIBRILLATION CHEST SKIN LF03611U INSTRUCTION MANUAL

Save a life CPR Practice. CPR and AED practice for general public

Mechanical ven3la3on. Neonatal Mechanical Ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on 8/25/11. What we need to do"

VENTILATORS PURPOSE OBJECTIVES

Children Small Adults Newborns Small Children

The Concept of Q-CPR. The need for Quality CPR

Neonatal and Pediatric Manual Hyperinflation: Influence of Oxygen Flow on Ventilation Parameters

What is ultrasonic plethysmography and how can we use it in airway disease?

BASIC LIFE SUPPORT - PAEDIATRIC

Technology Insights Paediatric Ventilation

Virginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P

Mechanical Ventilation

Recommendations on Checking Anaesthesia Delivery Systems

HAMILTON-C3 HAMILTON-C3. The compact high-end ventilator

SLE5000 Infant Ventilator with HFO

Physiological Basis of Mechanical Ventilation

Volume Diffusion Respiration (VDR)

Pulmonary Circulation Linda Costanzo Ph.D.

Other Oxygen Delivery Systems

MINI- COURSE on Management of OXYGEN in babies with RESPIRATORY DISTRESS

Disclaimer This material is intended for use by trained family members and caregivers of children with tracheostomies who are patients at the Alberta

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo

RESPIRATORY CARE POLICY AND PROCEDURE MANUAL. a) Persistent hypoxemia despite improved ventilatory pattern and elevated Fl02

AUTOVENT 4000 VENTILATOR

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

Lung Volumes and Capacities

Gas exchange measurement in module M- COVX

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation

LAB 7 HUMAN RESPIRATORY LAB. Complete the charts on pgs. 67 and 68 and read directions for using BIOPAC

Simulation for emergency care

Adult, Child and Infant Exam

Once student shouts for help, instructor says, Here s the barrier device. I am going to get the AED. STOP TEST

Influence of Acyclic Sports on Figures of the Respiratory System of Young Athletes of Years

Neonatal Ventilators. Touch-screen Neonatal Ventilator with High Frequency Oscillation SLE5000

A RANDOMISED CROSSOVER STUDY OF LOW-FLOW AIR OR OXYGEN VIA NASAL CANNULAE TO PREVENT DESATURATION IN PRETERM INFANTS

Basic Life Support. Based on UK Resuscitation Guidelines (2010)

NEONATAL VOLUME VENTILATION CLINICAL GUIDELINE V2.0

Standards and guidelines for care and management of patients requiring oxygen therapy.

B. A clinical emergency exists in which a profound hypoxia is determined to be present.

CHAPTER 3: The cardio-respiratory system

OXYGEN THERAPY. (Non-invasive O2 therapy in patient >8yrs)

Transcription:

ADC-FNN Online First, published on December 5, 2011 as 10.1136/archdischild-2011-300336 Original article Compressive force applied to a manikin s head during mask ventilation Jeroen J van Vonderen, 1 Ted A Kleijn, 1 Kim Schilleman, 1 Frans J Walther, 1 Stuart B Hooper, 2 Arjan B te Pas 1 1 Division of Neonatology, Department of Pediatrics, Leiden University Medical Center, Leiden, The Netherlands 2 Department of Physiology, Monash Institute of Medical Research, Melbourne, Australia Correspondence to Arjan B te Pas, Department of Pediatrics, Leiden University Medical Center, albinusdreef 2, J6-S, PO Box 9600, 2300 RC Leiden, The Netherlands; a.b.te_pas@lumc.nl Received 12 May 2011 Accepted 31 October 2011 ABSTRACT Objective To investigate the compressive force applied to the head during mask ventilation and determine whether this force increases in response to an attempt to correct the mask leak. Methods The authors asked 24 participants (consultants, fellows and nurses) to administer positive pressure ventilation to a modifi ed leak-free, term newborn manikin using a self-infl ating bag (SIB) and a Neopuff T-piece device. Recordings were made before and after the participants were informed about their percentage of mask leak and asked to correct this. Airway pressure and fl ow were measured using a Florian monitor, and the force applied to the head was measured using a concealed custom-made load cell weighing scale. Results There were no differences in the mean (SD) force applied to the head between devices used and before or after the attempt to correct the mask leak (SIB before 2215 (892) and after 2195 (989) g; Neopuff before 1949 (957) and after 2028 (909) g). There was a large variation in force with both devices before and after the attempt (coeffi cient of variation: SIB before 40% and after 45%; Neopuff before 50% and after 45%). There was no correlation between mask leak and the difference in force used before and after the attempt to correct the mask leak using both devices. Conclusion During mask ventilation of a manikin, the authors observed that large forces were exerted on the head with either an SIB or a Neopuff, but these forces did not increase during the attempt to minimise the mask leak. INTRODUCTION Approximately 10% of infants, particularly preterm infants, will need some form of resuscitation at birth at which time adequate ventilation is crucial and is usually applied using a facemask. 1 An appropriate mask technique is essential for establishing effective ventilation. 1 However, the leak between mask and face is common, even for experienced operators, and it reduces the efficacy of resuscitation. 2 4 Airway obstruction is also common in preterm infants receiving mask ventilation immediately after birth, 4 5 especially in very low birthweight infants as they have a relatively large tongue and small mandibles. 5 During mask ventilation of infants, obstruction can be caused by liquid in the oropharynx, positioning of the head, neck and tongue or by pressing down on the mask too hard, leading to obstruction of the nose and mouth. 2 5 While What is known about this topic An appropriate mask technique during neonatal resuscitation is essential for establishing effective ventilation. Leak between mask and face is common, even for experienced operators, and reduces the effi cacy of resuscitation. Airway obstruction occurs frequently when effort is taken to minimise the leak. What this study adds During mask ventilation of a manikin, large forces are exerted on the head. These forces did not increase during the attempt to minimise the mask leak. The force applied was not correlated to the amount of mask leak or whether obstruction occurred. investigating mask ventilation techniques using manikins, we observed that when the resuscitator was notified of mask leak, airway obstruction often accompanied the corrective action. 2 Specifically, we observed that the participants pressed harder on the mask with the intention to reduce mask leak. 2 However, there are no data available on how much force is applied to the head of infants during mask ventilation and whether it depends on the device used. This prompted us to investigate the compressive force applied to the head of a manikin during mask ventilation and determine whether this force increases in response to an attempt to correct mask leak, leading to obstruction. We also investigated whether different devices lead to different compressive forces to the head. METHODS The study was performed in the neonatal intensive care centre of the Leiden University Medical Center (LUMC), Leiden, The Netherlands. Neonatal staff members, registrars and nurses of the unit were asked to administer positive pressure ventilation to a modified, leak-free manikin (Laerdal Resusci Baby; Laerdal, Stavanger; Norway) representing a term newborn. In this manikin, the original lung was replaced with a Copyright van Vonderen Article JJ, Kleijn author TA, Schilleman (or their K, et al. employer) Arch Dis Child Fetal 2011. Neonatal Produced Ed (2011). doi:10.1136/archdischild-2011-300336 by BMJ Publishing Group Ltd (& RCPCH) under F1 licence. of 5

50 ml leak-free test lung (Dräger, Lubeck, Germany) that was positioned so that inflation caused a visible chest rise. Nondistensible tubing connected the lung and mouth, which ensured that all connections were airtight. 2 The mandible of the manikin was adapted to cause obstruction when too much downward pressure was given without an efficient chin lift. Positive pressure ventilation was applied using a size 0/1 Laerdal round mask (Laerdal, Stavanger, Norway) in combination with a T-piece infant resuscitator (Neopuff; Fisher & Paykel Healtcare, Auckland, New-Zealand) or an Ambu mark IV self-inflating bag (SIB) (Ambu, Copenhagen, Denmark) with a positive end expiratory pressure (PEEP) valve and manometer attached. The Neopuff was set to a peak inspiratory pressure (PIP) of 30 cmh 2 O and a PEEP of 5 cmh 2 O; gas flow was set to 8 l/min. The PEEP valve of the Ambu was also set to 5 cmh 2 O. All recordings were made using a Florian respiratory monitor (Acutronic Medical Systems, AG, Switzerland), which uses a hot-wire anemometer as a flow sensor and has a dead space of <1 ml. The flow sensor was placed distally between the T-piece of the Neopuff and the facemask or between the end of the SIB and the facemask, and the output signal was integrated to provide inspired and expired tidal volumes. The difference in inspired and expired volumes equals the facemask leak. 6 This was calculated by expressing the volume of the gas that did not return through the flow sensor during expiration as a percentage of the volume that passed through the flow sensor during inflation. 6 The Florian remained on to minimise the drift and was zeroed by the researchers each time the recordings were made. Measuring obstruction is complicated by the confounding effect of the leak as large leaks make it difficult to determine the presence of an obstruction. Therefore, we only noted obstruction from inflations with less than 30% leak. We defined obstruction as a typical flattening of the flow wave and a measured tidal volume <50% of expected values. The compressive force was measured using a custom-made load cell weighing scale (van Loenen, LUMC, Leiden, The Netherlands). This was composed of a spring-loaded rigid platform, which measured the force in grams independently from the surface area it was applied over. The weighing scale was zeroed before every measurement with the head of the manikin on the scale (figure 1). Before starting the measurements, the device was zeroed and the calibration was rechecked by measuring a known weight. The accuracy and reproducibility of the device was tested by measuring different known weights (250 and 500 g, and 1, 2 and 5 kg), and a combination of the weights, in triplicate. The scale constantly measured the force applied to the head of the manikin and the mean force was calculated. The signals of airway flow, tidal volumes, airway pressure and force applied to the head were digitised and recorded at 200 Hz using a data acquisition program (Spectra, Grove Medical Limited, Hampton, UK). All participants were informed that we were investiga ting the effect of leak reduction and were not aware that they participated in a study in which the compressive force exercised on the head was recorded. The weighing scale was concealed by a blanket under the head of the manikin. Also, the screen of the Florian and the laptop were not visible for the participants. Participants were experienced using T-piece and SIB. All participants were asked to give mask ventilation using both Figure 1 Photograph of the used installation, the manikin is shown being ventilated using the Neopuff. The head of the manikin is resting in the middle of the weighing scale. The scale is showed without cover. devices, but they were randomised to start with either the SIB or the T-piece. After 1 min, they were informed of the average mask leak that occurred and they were asked to improve their technique in order to reduce the mask leak. The average mask leak was calculated and reported to the participant. A total of 2 min of mask ventilation was recorded. Statistical analysis Data were analysed using SPSS for Windows version 17.0.0. The results are given as mean (SD). To quantify the difference between compressive force applied to the head and leak before and after the attempt to correct the mask leak, a paired t test was used. A comparison of all forces applied to the head was made using an analysis of variance. Correlations between pressure and leak were analysed using a Pearson correlation test. Correlations of the difference in the pressure between both measurements in Neopuff and SIB, and leak were also analysed using a Pearson correlation test. The coefficient of variation (CV) was calculated as (SD/mean) 100, and a CV<5% was regarded to reflect good agreement and a CV<10% to reflect acceptable agreement. A P value <0.05 was considered statistically significant. Reported P values are two-sided. RESULTS Participants We obtained all four recordings from 24 participants, of whom 10 were consultants, 5 registrars and 9 neonatal intensive care unit nurses. All participants were trained and experienced in neonatal resuscitation. Compressive force applied to the head There was a large variation in the compressive force applied to the manikin s head, which ranged between 606 and 3717 g (figure 2). The mean force applied during resuscitation was similar for both devices (SIB and T-piece, respectively: 2205 (932) and 1989 (925) g) and the variation in force applied was also similar between devices (SIB and T-piece, respectively, CV of 40% and 50%) (figure 2). The variation in the force did not change after the correction of the leak (SIB and T-piece: CV both 45%; figure 2) and primarily arose due to variation in the force applied between operators, irrespective of the device used. The mean CV in the force applied by each F2 of 5 van Vonderen JJ, Kleijn TA, Schilleman K, et al. Arch Dis Child Fetal Neonatal Ed (2011). doi:10.1136/archdischild-2011-300336

Figure 2 Box plots representing the median (IQR) force in g for both devices and for attempts before and after the correction of leak. Selfinfl ating bag (SIB)/Neopuff before and SIB/Neopuff after represent the attempts before and after the effort to correct the mask leak by all participants using a SIB and Neopuff. Figure 3 The δ force (=force fi rst attempt force second attempt) in g against the percentage of mask leak notifi ed to the participant after the fi rst attempt, ventilating with a self-infl ating bag. (Pearson correlation test: r= 0.10; ns.) operator using each device and before and after leak correction was 18.2% (12.4), with six operators having a CV of <5%. There was no change in force during either the PIP or the PEEP phases of inflation. During mask ventilation, there was no difference in the compressive force on the head between SIB and T-piece before (mean (SD) SIB vs Neopuff: 2215 (892) vs 1949 (957) g; ns) and after (2195 (989) vs 2028 (909) g; ns); the effort to correct the mask leak. van Vonderen JJ, Kleijn TA, Schilleman K, et al. Arch Dis Child Fetal Neonatal Ed (2011). doi:10.1136/archdischild-2011-300336 F3 of 5

Figure 4 The δ force (force fi rst attempt force second attempt) in g against the percentage of mask leak notifi ed to the participant after the fi rst attempt, ventilating with a Neopuff. (Pearson correlation test: r= 0.27; ns.) During ventilation with the SIB, there was no difference in the compressive force before and after the effort to correct mask leak (mean (SD) before vs after: 2215 (892) vs 2195 (989) g; ns). The compressive force increased in 14 participants (by 14.0% (18.2)) and decreased in 10 participants (by 19.5% (17.4)). The median (IQR) difference in the force applied between before and after (before to after) mask placement correction was 19 ( 200 to 177) g (figure 3). During ventilation with the T-piece, there was no difference in the compressive force between before and after the effort to correct the mask leak during ventilation (mean (SD) before vs after: 1949 (957) vs 2028 (909) g; ns). However, the majority (15% (63%)) of participants increased (by 21.5% (23.8)) compressive force, whereas only 9 (37%) participants decreased (by 12.5% (11.8)) the amount of compressive force applied. The median (IQR) difference in the force applied between before and after (before to after) mask placement correction was 79 ( 304 to 95) g (figure 4). Correlation of leak and force applied to the head after correction of facemask position There was no correlation between the amount of facemask leak and the difference in compressive force applied before and after attempting to correct the facemask position using an SIB (r= 0.10; ns) or a T-piece (r= 0.27; ns) (figures 3 and 4). Leak and airway obstruction Using the SIB, the mean (SD) amount of facemask leak was similar before correction (47% (33%)) compared with after correction of facemask leak (36% (35.3%)). Using the T-piece, the mask leak was 42.3% (31%) before correction and 38.0% (33.6%) (ns) after correction of facemask positioning. Table 1 Number of occurrence of obstruction for all participants, no obstruction or obstruction could not be evaluated due to high mask leak (N=24) during the attempts of participants Obstruction No obstruction Obstruction cannot be judged SIB before 2 13 9 SIB after 1 15 8 Neopuff before 1 17 6 Neopuff after 2 15 7 SIB, self-infl ating bag. During inflations that were associated with a minimal leak, occasionally airway obstruction was observed, but the occurrence of airway obstruction was not increased by an attempt to correct the mask leak using both devices (table 1). There was no difference in compressive force applied to the head in the presence (2141 (866) g) or absence (2205 (920) g; ns) of an airway obstruction. DISCUSSION This is the first study evaluating the compressive force that is potentially applied to an infant s head during mask ventilation. Although we observed a large variation in compressive force applied by the participants during mask ventilation, even the minimum force (606 g) applied equates to a very significant pressure on a newborn infant s skull (see below). This variation was not associated with the type of device used nor was it associated with attempts to minimise facemask leak. Instead, the large variation was mostly associated with differences between operators, with 25% of participants displaying a CV of less than 5% across all four test conditions (two devices, before and after leak correction). Although there are no normal values available for comparison, the application of force F4 of 5 van Vonderen JJ, Kleijn TA, Schilleman K, et al. Arch Dis Child Fetal Neonatal Ed (2011). doi:10.1136/archdischild-2011-300336

required to ensure a good seal between the mask and the face should be countered by lifting from under the chin ( squeezing the mask), 2 3 and so the net force on the skull should be minimal. However, the amount of compressive force applied to the skull in these studies was significant, but was not correlated to the amount of mask leak before or after leak correction or whether obstruction occurred or not. It is likely that other factors in the mask technique play a role in how much force is pressed on the head. Airway obstruction can occur when more force is exerted on the mask during an attempt to reduce the leak. 2 5 However, in this manikin study, we could not confirm the compressive force increased during the attempt to correct for facemask leak. In addition, no correlation was found between the compressive force and the occurrence of leak and obstruction. However, as we only observed obstruction occasionally, it is difficult to draw conclusions. Although the participants were blinded for the aim of this study, it is possible that our previous study 2 created more awareness of facemask leak and the possibility of airway obstruction, which may have influenced our results. Furthermore, as the participants were familiar with the use of manikins, it is possible that they applied greater force than they would apply to an infant. Nevertheless, the average force measured was considerable and even if it was markedly reduced, we would expect it to have a major impact on the cerebral physiology (see below). When translating our measurements into approximate pressures (mm Hg) that could be applied to preterm infants during mask ventilation, we first measured the area of the occiput of a preterm infant making contact with the underlying surface. To calculate this, a print of the area was made from a preterm infant of 29 weeks of gestation using washable ink (permission was received from the parents before making this measurement). The measured weight force (in g) was then divided by this surface area (9.1 cm 2 ) and converted into mm Hg using the equation; 1 g/cm 2 =0.735561273 mm Hg. With a surface area of 9.1 cm 2 (see methods), we calculated that a mean (SD) pressure of 170 (75) mm Hg was applied to the head. During ventilation with an SIB, a mean (SD) pressure of 179 (72) mm Hg was applied before and 177 (79) mm Hg after the attempt to correct the mask leak, during Neopuff 157 (77) mm Hg and 164 (73) mm Hg. We found no difference in the compressive force when an SIB or a Neopuff was used. Apparently, the force is predominantly determined by the way the mask is held and not by the device used to administer the inflations. This is supported by our observation that during inflations, the compressive force did not change with either device and that the primary source of variation was between participants. This study has been performed with a manikin and it is difficult to extrapolate our findings to infants. There are no clinical data available about the possible adverse effect that can occur when a large force is pressed on a head of a newborn infant. The manikin s head is considerably less compliant than the head of a newborn infant, and it is possible that the participants were more concerned with reducing the facemask leak rather than the force applied in this study. Thus, we would expect that the force applied to the head of an infant be considerably less, would be applied over a much greater area ( 9.1 cm 2 ) and would be mitigated by any structural rigidity and force dispersement provided by the scalp and skull. Nevertheless, the pressures exerted were considerable (averaging 150 180 mm Hg) and even if the real pressures exerted were a quarter of this, we would expect it to have a very significant physiological impact. Indeed, as the systemic arterial pressures are only 50 60 mm Hg (at best), we would expect cerebral capillary blood flow to effectively cease in regions exposed to the greatest force. If the pressures persist, this may cause hypoxic/ischaemic injury as well as increasing the risk hyperperfusion injury following the release of the force. The associated increase in the intracranial pressure may also cause cerebral tissue damage and contribute to a reflex bradycardia, which is often already present in a newborn infant requiring resuscitation. Although it has been described that when a resuscitation mask is jammed on the face of a newborn, this can mould the back of the head and can bruise the face 1, there are no reports describing the influence of the force on cerebral perfusion. Clearly, the exertion of too great a force on an infant s skull is undesirable and this study is the first to measure and highlight the risks of applying large compressive forces during facemask resuscitation. In conclusion, during mask ventilation of a manikin, we observed that large forces were exerted on the head, but these forces did not increase during the attempt to minimise the mask leak. Using an SIB or a Neopuff made no difference in how much force was applied. More experimental and observational studies are needed to see whether large forces are used during resuscitation and what their effect is on a newborn infant s head and especially the cerebral circulation. Acknowledgements The authors are grateful to our biomedical engineer, Miss R van Loenen, for developing the custom-made load cell weighing scale that the authors used for this study. Competing interests None. Provenance and peer review Not commissioned; externally peer reviewed. REFERENCES 1. Kattwinkel J. Textbook of neonatal resuscitation. Fourth edition. Elk Grove Village, IL: American Academy of Pediatrics and American Heart Association, 2000. 2. Schilleman K, Witlox RS, Lopriore E, et al. Leak and obstruction with mask ventilation during simulated neonatal resuscitation. Arch Dis Child Fetal Neonatal Ed 2010;95:F398 402. 3. Wood FE, Morley CJ, Dawson JA, et al. Improved techniques reduce face mask leak during simulated neonatal resuscitation: study 2. Arch Dis Child Fetal Neonatal Ed 2008;93:F230 4. 4. Schmölzer GM, Dawson JA, Kamlin CO, et al. Airway obstruction and gas leak during mask ventilation of preterm infants in the delivery room. Arch Dis Child Fetal Neonatal Ed 2011;96:F254 7. 5. Finer NN, Rich W, Wang C, et al. Airway obstruction during mask ventilation of very low birth weight infants during neonatal resuscitation. Pediatrics 2009;123:865 9. 6. O Donnell CP, Kamlin CO, Davis PG, et al. Neonatal resuscitation 1: a model to measure inspired and expired tidal volumes and assess leakage at the face mask. Arch Dis Child Fetal Neonatal Ed 2005;90:F388 91. van Vonderen JJ, Kleijn TA, Schilleman K, et al. Arch Dis Child Fetal Neonatal Ed (2011). doi:10.1136/archdischild-2011-300336 F5 of 5