PTG100 Precision Test Gauge

Similar documents
陕西创威科技有限公司. Shaanxi Creat Wit Technology Co., Ltd. Precision pressure calibrator-cwy300 Operating Manual

Additel 761 Automated Pressure Calibrators

Additel 761 Automated Pressure Calibrators Selection Guide

Precision digital pressure gauge Model CPG1000

Additel 761 Automated Pressure Calibrators

Hand-held pressure calibrator with integrated pump Model CPH6600

Additel 761 Automated Pressure Calibrators Selection Guide

Additel 761 Automated Pressure Calibrators

DPC-30 DPC-100. Reference Manual

M104 Series: Digital Manometer: 0.1 % of Full Scale

Additel 780 Series Pressure Controller

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

ADT 761 Automated Pressure Calibrator

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. Type: ETd 04. class 0.05

Digital pressure gauge Model CPG500

Additel 780 Series Pressure Controller

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

Pneumatic High-Speed Pressure Controller Model CPC3000

Pneumatic high-pressure controller Model CPC7000

Pneumatic high-pressure controller Model CPC7000

Datasheet: K-30 ASCII Sensor

Manual of SF6 Comprehensive Tester

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications

Digital pressure gauge Model CPG500

Series Ruska Digital Pressure Indicator. GE Sensing & Inspection Technologies. Features

High-speed pressure sensor Model CPT6140

M2110P SMART PRESSURE GAUGE USER S MANUAL

Automatic Isokinetic Sampler. ST5 Evo

Optical Dissolved Oxygen Meter

Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features

Digital deadweight tester Model CPD8500

Precision pressure sensor Basic version Model CPT6020

SAUTER FK Version /2016 GB

Pneumator. Testo Industrial Services More assurance, better service. Pressure calibrator and measuring instrument.

BI-680 Online Dissolved Oxygen Controller Instruction Manual

AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP

ASDX Series Silicon Pressure Sensors

Operating Instructions Part No

Digital dead-weight tester Model CPD8500

LOCK UP VALVES YT-400/405/430/435 SERIES

Overview. Front Panel: Keypad and Display

SPD Pressure Sensor Families

MODEL 100 NITROGEN INFLATION CART

Mod. M M M063..

Digital Pressure Gauge Main features: Data Sheet: BIT02B.417.R6.EN COVER on request

Optical Dissolved Oxygen Meter

Multifunction Altimeter/Variometer AV1

Ultima. X Series Gas Monitor

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

This multi-functional device allows the determination of up to six quality parameters with only one measurement:

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

628 Differential Pressure Decay Leak Tester

DIGITAL AUTOMATIC TYRE INFLATOR

Datasheet TM40. HUBER INSTRUMENTE AG, Grellingerstrasse 23, CH-4208 Nunningen. All rights reserved Paige 1 / 5 Datasheet_TM40_

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

SANP ACTING RELAYS YT-520/525/530 SERIES

Pressure transmitter PASCAL Ci4

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

WTSeries C R Y S T A L. Digital Pressure Gauge. Operation Manual. engineering corporation

GasSense NDIR User Manual

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

Automatic Isokinetic Sampler. ST5 Evo

A4 Operation Manual. Fig.1-1 Controller Socket Diagram

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

Additel 780 Series Pressure Controller

MP15 Jockey Pump Controller

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

BetaGauge PIR/PIR PRO

Compact Manometer. Pressure measurements can easily be taken any time, anywhere. POWER LIGHT

A4s Operation Manual

Industrial pressure controller Model CPC4000

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

D10S/D20S Wall/Post Mount Inflator Quick Start Manual

How to specify a product. Process Sensors and Mechanical Instruments

Manual Weighingblock VB2 series and Uniscale

MODEL 6600 Rev. B4 USER'S MANUAL LAVERSAB INC., 505 GILLINGHAM LANE. SUGAR LAND TX (281) FAX: (281)

ECHO MANUAL WARNING. L B A ltim e te rs. ECHO is a trademark of LB Altimeters, Denmark

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

DO NOT PRINT THIS PAGE

Digital pressure gauges

Touch Screen Guide. OG-1500 and OG Part # T011

Industrial pressure controller Model CPC4000

2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters. Kent-Taylor

BetaGauge PI & PI PRO

10 switches simultaneously. The settings of the master sensor (source of copy) can be copied to the slave sensors. Reducing setting labor

PM105 PRESSURE METER USER S MANUAL

Compact Manometer. Pressure measurements can easily be taken any time, anywhere.

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

Digitale manometers. TYPE: ETD-01, klasse 0.5. Specificaties HOOFDSTUK 6

Instruction Manual EMA 84 Digital Pressure Gauge

EM151/EM201. Electronic Manometer - + INSTRUCTION MANUAL EM Washington Street Melrose, MA Phone Toll Free

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

WELCOME TO THE REVOLUTION

Control units. Installation and operating instructions Series: DP-Control

MPM280 Piezoresistive OEM Pressure Sensor

MANUAL DIRECT PURGE OPTION. UNION Instruments GmbH CWD2005 PLUS. General information, safety standards and regulations for direct purge option

Operating Instructions Part No

Short manual testo

VOLUME BOOSTER RELAYS YT-300 / 305 SERIES

Transcription:

PTG100 Precision Test Gauge User Manual PD1007 Rev B 03/28/2014 Palmer Instruments Inc. 234 Old Weaverville Road Asheville, NC 28804 Toll Free: 800-421-2853 Phone: 828-658-3131 Fax: 828-658-0728 Email: sales@palmerwahl.com

PTG100 Precision Digital Pressure Gauge - Instruction Manual ATTENTION: This manual is subject to change without prior notice. In the event of any revisions, the latest version shall prevail. Please contact us if you encounter any errors or omissions in this manual. We shall not be held liable for any damages or injuries resulting from misuse by the user. The features described in this manual shall not be taken as the reason to use the product for purposes outside its intended function. WARRANTY & CALIBRATION REGISTRATION at www.palmerwahl.com/register Registration is fast and easy. In about a minute you can have your product automatically registered for Warranty Protection and our Calibration Reminder service. Let Palmer Wahl help you protect your investment, and maintain product accuracy and compliance with ISO and other quality standards. Questions? Call Customer Service at 1-800-421-2853 or 828-658-3131 Or email: register@palmerwahl.com Warning! Please read these Safety Instructions before operation. To avoid possible electric shock, personal injury or damage to the digital pressure gauge and other equipment, please follow all equipment safety procedures. Please read this user manual and strictly conform to the provisions in the use of the gauge. Before you use this precision digital pressure gauge, inspect the case. Do not use it if it is damaged. Make sure the battery door is closed and latched before you operate the precision digital pressure gauge. Pay special attention to the requirements for working conditions; do not operate the digital pressure gauge around explosive gas, vapor, or dust. In the even that the gauge will not be used for a long time, turn off the power to save the battery life. To avoid false readings, please replace the batteries as soon as the battery indicator appears. Stop using the gauge before replacing the batteries. Use only AA, 1.5V batteries, properly installed in the digital pressure gauge. Please use clean and soft cloth to wipe the instrument panel; avoid letting water penetrate the gauge. Do not disassemble the gauge. Any problems caused by disassembly are beyond our service range and cause for the warranty to be void. In the event of an abnormal fault, power off the device and take out 1

PTG100 Precision Digital Pressure Gauge -Instruction Manual Contents I. Description... 3 1.1. Overview... 3 1.2. Features... 3 II. Panel and Configuration Introduction... 3 2.2. Panel Keys Function Introduction... 4 2.3. Keys Function Introduction during Data Input... 4 Ⅲ. Basic Functions... 5 3.1. Power On/Off... 5 3.2. Pressure Measurement... 6 3.3. Zeroing... 6 3.4. Pressure Units... 7 3.5. Backlight... 7 3.6. Peak Detection... 7 3.7. Percent Pressure... 7 3.8. Overpressure Alarm... 8 3.9. Sampling Rate Settings... 8 3.10. Damping Time Settings... 8 3.11. Restore Factory Settings... 8 Ⅳ. Calibration Function Introduction... 9 4.1. Two Point Calibration... 9 4.2. Clear Two Point Calibration Data... 10 4.3. Restore Data of Zeroing Value... 11 4.4. Multipoint Calibration... 11 Ⅴ. Faults and Maintenance... 13 2

PTG100 Precision Digital Pressure Gauge - Instruction Manual I. Description 1.1. Overview The PTG100 is a high precision single-range precision digital pressure gauge. It is composed of high precision measuring chips, a battery circuit, and a high precision pressure sensor. The pressure gauge can be used for real time measuring and calibrating of a general pressure gauge, e.g., pointer pressure gauges, pressure sensors, etc. 1.2. Features Engineer Units: Pa, kpa, MPa, mmh 2O, mmhg, bar, mbar, kgf, or psi; Display: LCD with white backlight, 6-digit display; Overpressure Warning: The screen will display Fail if pressure reaches over 110% of full scale; Measuring Speed: User settable, factory default is 3 times/sec Working Time: up to 10,000 hours (3 times per second); if battery power is very low, the PTG100 will power off automatically. Working Conditions: Temperature: -10-50 Temperature Compensation: 0-50 Storage Temperature: -20-70 Power Source: 3 AA Batteries Communication: RS232 port Dimensions: Φ100mm 37mm, total height is132mm Weight: 0.45kg Pressure Connection: ¼ NPT Re-calibration Interval: One year (recommended) II. Panel and Configuration Introduction 2.1. PTG100 Precision Digital Pressure Gauge Front Panel (See figure-1) Figure-1 PTG100 Precision Digital Pressure Gauge Front Panel 3

PTG100 Precision Digital Pressure Gauge -Instruction Manual 2.2. Panel Keys Function Introduction Press and hold. The PTG100 will show the measuring range, and will show pressure value after releasing it. PTG100 has been equipped with batteries, and is ready for use. PTG100 will retain the previous unit set by the user. The next time the gauge is turned on, it will continue to work in the former mode and pressure unit. Press this key to select the next pressure unit. Continue to press it to switch between the set unit of the gauge (Pa, kpa, MPa, mmh 2 O, mmhg, bar, mbar, kgf, psi) Press this key to turn on the backlight. After the backlight setting time of 5 seconds (selectable) has passed, the backlight will turn off automatically. Press this key to enter the menu operations Press this key for clearing the current displayed pressure value (which is regarded as zero drift value). After zeroing, the gauge will keep the status even if powered down and turned back on. Please allow PTG100 process connection to be open to the atmosphere during zeroing. 2.3. Keys Function Introduction during Data Input A. Confirm to modify the data B. Confirm key for no display menu modification. C. Data circulation key, the data circle from 0~9 D. Enter into data modification When the user uses the keys to carry out the related operations, the LCD will remind the user to input or modify certain parameters. The user can carry out the operation according to the following steps: (1) Under the status of the front LCD in display, press, and the first character of the data will blink. (2) If there is no need to modify the blinking value, please skip to step ⑶. If it is required to modify the blinking value, press and select the desired value. The 4

PTG100 Precision Digital Pressure Gauge - Instruction Manual numbers will increase between 0 and 9. Please skip to step(3)after modifying the blinking character. Note: when the decimal point digit blinks, press to clear the prior decimal point. After adjustment, please skip to step(3). (3)Press again to make the next digit blink. (4) Repeat step ⑵ and ⑶ till the last digit modification completed. (5)Press to save the modified data. 2.4. Key Functions during Menu Operation A. Return to Previous Menu B. Confirm the menu without data modification C. Enter the next menu D. Enter the menu Ⅲ. Basic Functions 3.1. Power On/Off 5

PTG100 Precision Digital Pressure Gauge -Instruction Manual Press to turn on/off the PTG100. The instrument will do a self test and then show the instrument s max range. Figure 3-1 shows the first screen 3.2. Pressure Measurement Press to enter the pressure meauring menu; the displayed content includes: Percentage Icon, Pressure Measuring Value, Pressure Units Note: Initially these register values are set to factory calibration values. If the pressure exceeds 110%FS, the screen will display -----. To prevent damage to the sensor, release the pressure immediately. Figure 3-2 shows pressure measurement interface. 3.3. Zeroing (1)Types of Gauge Pressure: When connecting the PTG100 to the atmosphere via connector, if the current pressure is within the range of (-1% - 1%), press ZERO for zeroing. Figure 3-3 shows the zeroing of pressure gauge 2)The Type of Absolute Pressure: 1 When the PTG100 is vented to the atmosphere, the user should know the current atmosphere pressure value (P standard); 2 Press ZERO and hold to enter the data input state, then input the actual pressure ( P Standard); 6

PTG100 Precision Digital Pressure Gauge - Instruction Manual 3 In the pressure measure menu, the measured pressure will change to ( P Standard). Now, the zeroing process is finished. 3.4. Pressure Units Press UNIT to view and select among the pressure units. The order is Pa, kpa, MPa, mmh 2 O, mmhg, bar, mbar, kgf, and psi. Conversions between pressure units is shown in table 3-4-1. Pa kpa MPa mmh 2O mmhg bar mbar Kgf/cm² psi 1000 1 0.001 102.047 7.50061 0.01 10 0.010197 0.1450377 Table 3-4-1: Pressure units conversion relation Figure 3-4 shows the method to switch the different pressure units. 3.5. Backlight Press LIGHT key to turn on/off the backlight. Selectable time is from 5 seconds to 10 seconds. Figure 3-5 backlight time of display setting is 5 seconds 3.6. Peak Detection Figure 3-6 displays the max. /min. recorded peak value 3.7. Percent Pressure % indicator: shows the current pressure percentage; Scale graph: 0% - 100%, the minimum scale is 10%; Example: For a PTG100 set with a scale of (0 1,000) kpa, if the current pressure value is 100kPa, the percent indicator is 10%, as figure 3-7 7

PTG100 Precision Digital Pressure Gauge -Instruction Manual Figure 3-7 3.8. Overpressure Alarm When the current pressure is beyond the range of high/low limit, the whole screen will show as figure 3-8 to warn the user to adjust the pressure. Figure 3-8 3.9. Sampling Rate Settings Sampling rate is selectable between 1 and 10 seconds; the default is three samples per second; if setting is 5 seconds, then the rate is three samples in five seconds. Figure 3-9 displays that the sampling rate is 1 second. Figure 3-9 3.10. Damping Time Settings The default damping time of PTG100 is 0 second; user can set the time according to their requirements (0~32 seconds selectable). See figure 3-10, where the default damping is 0 second. Figure 3-10 3.11. Restore Factory Settings If there is any incorrect data entered by the user, the gauge factory settings can be restored via following steps: 1) Under measuring status, press CONFIG twice, then press LIGHT four times to enter the status of restore factory setting. 8

PTG100 Precision Digital Pressure Gauge - Instruction Manual 2) Press CONFIG once and input the password 12345.0; 3) Press UNIT to restore the factory backup data; 4) Press UNIT three times to return to the measuring interface. Ⅳ. Calibration Function Introduction Calibration Conditions: Environment: Temperature: 20 ±2. Relative humidity: (45-75)%. Atmosphere pressure: (86 to 106) kpa. Calibration Equipment: Use standard pressure equipment(with an accuracy rating higher than 0.025%FS)to perform the calibration. Warning: User must have knowledge of standard pressure equipment and the PTG100 digital pressure gauge. Normally, two point calibrations can meet the accuracy requirements. If multipoint calibration is required, perform the calibration without modifying the standard pressure value of multipoint calibration. 4.1. Two Point Calibration 4.1.1 Low Point Calibration 1) A pressure source with an accuracy higher than 0.025%FS is necessary for performing pressure module calibration; 2) Under measuring status, press CONFIG once to enter the configuring status: 3) Press LIGHT once to enter the status of two point calibration; 4) Press CONFIG once to enter the low point calibration menu; pressurize the standard 9

PTG100 Precision Digital Pressure Gauge -Instruction Manual device to the standard value according to the display value of the LCD. For example: if it is 0 kpa, then the standard device will be 0 kpa; if it is -100kPa, the standard device will be -100kPa. 5) Press CONFIG once to enter the status of low point calibration; the first digit will blink. 6) Wait until the pressure reading is stable; 7) Press UNIT to save the low point calibration data. 4.1.2 High Point Calibration 1) After low point calibration, press ZERO to enter the status of high point calibration. 2) Press CONFIG once to enter the high point modification status. Pressurize the standard device to the standard value according to the display value of the LCD. For example, if the LCD shows 100 kpa, then the standard device pressure will be 100 kpa. 3) Wait till the pressure is stable; 4) Press UNIT to save the high point calibration data. 4.2. Clear Two Point Calibration Data 1) If user finds that the error of measuring data is too much after two point calibration, they can clear the two point calibration data via the function of Clearing Two Point Calibration Data. 2) Press CONFIG once, press ZERO once to enter the two point calibration menu directly; 3) Press CONFIG once, then press ZERO twice to enter the two point calibration clearing menu. The C_ZS at the lower left corner will blink. 4) Press ZERO once, and C_ZS stops blinking. The two point calibration data will be 10

PTG100 Precision Digital Pressure Gauge - Instruction Manual cleared. 5) Press UNIT twice to return back the measuring interface. 4.3. Restore Data of Zeroing Value 1) If user finds that the error of the displayed reading is too much after two point calibration, they can clear the two point calibration data via the function of Clearing Two Point Calibration Data 2) Press CONFIG once, then press ZERO once to enter the two point calibration menu directly. 3) Press CONFIG once, press ZERO twice to enter the two point calibration clearing menu, and the C_ZS at the low left quarter will blink. 4) Press ZERO once, C_ZS stops blinking, and the two point calibration data will be cleared. 5) Press UNIT twice to return back to the measuring interface. 4.4. Multipoint Calibration 1) Under measuring status, press CONFIG once to enter the configuring status: 2) Press LIGHT twice to enter the interface of multipoint calibration: 3) Press CONFIG once to enter the pressure calibration point interface. As the instrument calibration has been carried out with specific temperature compensation, the calibration point is not allowed to be modified: 4) Press LIGHT to enter the first point pressure calibration interface: 11

PTG100 Precision Digital Pressure Gauge -Instruction Manual 5) Press CONFIG once and the first digit will blink. Pressurize the standard device to standard pressure value according to the LCD display pressure value, and wait until the pressure is stable: 6) Press UNIT once to save the first point calibration data. 7) Press LIGHT once to enter the second pressure calibration interface: 8) Press CONFIG once and the first digit will blink. Pressurize the standard device to the standard pressure value according to the LCD displayed pressure value, and wait till the pressure is stable. For example, the LCD displays 100 kpa, the standard device will be 100 kpa; 9) Press UINT once to save the second point calibration data; 10) Repeat the above-mentioned steps until all pressure calibration points have been calibrated. 11) Press UNIT twice to return to the measuring interface. 12

PTG100 Precision Digital Pressure Gauge - Instruction Manual Attention: During use, the following reasons may cause a discrepancy between the display value of the pressure gauge and actual value: Error of parameter setting by the user User s pressure standard differing from manufacturer s pressure standard The sensor characteristic has been changed, due to overpressure, over-temperature, and time drift. The user may correct this in one of the following two ways: 1.) Restore factory backup data. If there are any user parameter error settings or error settings during calibration, the user can restore the gauge to the factory defaults. 2.) Two Point Calibration: On the condition that the pressure standard equipment is available, carry out two point calibration, including low point calibration and high point calibration. Notes: 1.) In order to acquire the highest accuracy, please make sure that the pressurized value is equal to the lower limit while performing low point calibrations, and it is equal to the upper limit while performing high point calibrations. 2.) In order to acquire the highest accuracy, please make sure to perform the high point calibration after completing the low point calibration. 3.) Multipoint calibration: When performing a two point calibrations, if the product linearity can not meet the accuracy requirement, then perform a multipoint calibration within its measuring range. As the product performs temperature compensation, the calibration point and standard pressure value cannot be changed. Ⅴ. Faults and Maintenance 1.) The operation environment should meet the requirements in this manual. 2.) After turning on the gauge, the screen should display normally. If not, please return the gauge for repair. 3.) Make sure the display is normal and the screen shows the pressure range. If the displayed measuring range disagrees with the labeled range, check if the range needs modification. Modify the range to the measuring range on the product label. 4.) Enter the status of measurement. Pressurize smoothly to the full scale first, then depressurize slowly, observing whether the pressure value is normal. If any abnormal readings are found, please calibrate it according the pressure calibration requirements. 13

PTG100 Precision Digital Pressure Gauge -Instruction Manual 234 Old Weaverville Road, Asheville, NC 28804 Toll Free: 800-421-2853 Phone: 828-658-3131 Fax: 828-658-0728 www.palmerwahl.com info@palmerwahl.com 14