Equipment Operating Procedure Glove Box

Similar documents
Glove Box Installation Manual

Standard Operating Procedure

University of Minnesota Nano Center Standard Operating Procedure

School of Chemistry SOP For Operation Of Glove Boxes

Glove Box with Gas Purification System and Digital Control VGB-6. MTI Corporation

MB-SP Series Solvent Purification System

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Chromat Calibration Updated October 27th, 2017

Operating Instructions

with O 2 Controller Instruction Manual

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

256 Pneumatic Pressure Indicator

GETZ EQUIPMENT INNOVATORS RECOVERY SYSTEM CLEAN AGENT FE-36 / HALOTRON 1 PART NO. 4G59751 OPERATIONS MANUAL

IMPORTANT CO2/ HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES. Tank valves must be installed or removed by qualified personnel.

AX5000 Operational Manual

AcornVac Vacuum Plumbing Systems - Trouble Shooting Guide

IMPORTANT CO2/ HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES. Tank valves must be installed or removed by qualified personnel.

Operations Manual. Internal automatic Pressure control for orbital tube welding. By Exel Orbital Systems, Inc. Page

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

Calcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO

University of MN, Minnesota Nano Center Standard Operating Procedure

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Sedan Range. Evaporative Loss System Check Clarified Procedure

Instruction Manual VCG-6

14500 Coy Drive, Grass Lake, Michigan Vinyl Anaerobic Airlock Chamber Instruction Manual

Installation Instructions

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

APPS-160 Gen 2 User Manual

Basic & Controlled Atmosphere Glove Boxes

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

SPUTTER STATION STANDARD OPERATING PROCEDURE

Revised: June 7, 2017

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

NTF-515. Operation Manual

Manual Solvent Purification System. Innovative Technology, Inc.

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Usage Policies Notebook for Trion RIE / ICP Dry Etch

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Solvent System Walkthrough

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

Series IRV1000/2000/3000

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

Karl Suss MJB4 Mask Aligner

FLUSHMATE III FLUSHMATE FLUSHOMETER - TANK SYSTEM. Owner s Service Manual. 503 Series. 503 Series

Rejuvenation Instructions

INSTRUCTION MANUAL FOR POLYKEG DISPOSABLE KEG

Basic ICP Operating Procedures

! WARNING! IMPORTANT HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES ! WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINES

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

Operating Procedures for Metal Evaporator I

WARNING OLED & BASIC MENU QUICK START

MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE

Standard Operating Procedure Inert Vacuum Line. Ryan Mulvenna, May 2012

Using the Akta Prime plus October 22, 2012

Combination Air Valve

Inert Air (N2) Systems Manual

Birck Nanotechnology Center

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Approved by Principal Investigator Date: Approved by Super User: Date:

OPERATION MANUAL NTF-15

Pressure Dump Valve Service Kit for Series 2300 Units

Installation Instructions

User Manual. Quantos Automated Dosing Powder Module

OXFORD PLASMALAB 80PLUS (CLOEY)

OPERATION MANUAL NTF-60 Plus

UM-010 Invivo User Manual

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

ELECTRONIC PLC SERIES Test Benches

FAULT CODE TROUBLESHOOTING INDEX

user manual .68 CALIBER ELECTRONIC PAINTBALL MARKER Meets or exceeds ASTM standards.

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Ammonia Treatment Instructions Page 1 of 5

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

Smooth Vent Valve. Series XVD XL XL Q D- XVD XGT CYV

Standard Operating Manual

User Instruction Manual

3M Liqui-Cel EXF-10x28 Series Membrane Contactor

MANUAL BE SERIES Test Benches

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

Cover Page for Lab Report Group Portion. Flow Visualization in a Water Channel

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Pressure Dump Valve Service Kit for Series 3000 Units

SINORIX Valve Reconditioning

Norton Group Toepler Pump Operating Instructions. Version 3.1

ACCUMULATOR OPERATING & MAINTENANCE INSTRUCTIONS

B. Assemble the Method-25 sample train with a heated probe/filter.

Performa Cv Twin Alternating and High Flow Systems. Manual Supplement

Safety, Operating and Maintenance Manual. TRAIN Rat DS9-TRAIN

Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat. Version 5.1

Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch

PCS 340 Core Saturator : 115 Volt : 230 Volt. Instruction Manual

Cover Page for Lab Report Group Portion. Head Losses in Pipes

UNIVERSITY OF ROCHESTER ENVIRONMENTAL HEALTH & SAFETY

User's Manual. MixRite TF 10. Edition 05.08

TECHNICAL INFORMATION

NTF-230 OPERATION MANUAL

GUIDELINES FOR CONDUCTING HYDROGENATION REACTIONS

Is there any visible damage to the machine? If yes, describe damage:

Transcription:

Equipment Operating Procedure Glove Box Page 1

0.0 Changing the Compressed Gas Cylinder 1. Complete Compressed Gas Cylinder training from EHS website before manually exchanging gas cylinders. In order to access gas cylinders stored in Scott Lab, request the garage key from the 3 rd floor office. Scott Lab office keys have access to cylinder storage. Cylinder storage is the 2 nd door on the left in the 1 st floor garage in West side of Scott Lab. 2. To remove current cylinder, turn valve on top of tank clockwise to close it. 3. Using a wrench, twist the bolt connecting cylinder to safety valve in an upward motion to loosen it. Continue twisting the bolt until the safety valve is removed. Replace protective valve covering on gas cylinder. 4. Important: Tear off the contents tag on the cylinder to show the tank is Empty. 5. To attach new gas cylinder, remove protective valve covering and insert the safety valve in the port on top of the cylinder. 6. Using a wrench, screw the valve bolt in a downward motion to tighten it completely. 7. Turn the control valve in top of the cylinder counter-clockwise to open flow from cylinder. 8. Tear off the contents tag on the cylinder to show the tank is In Use. 9. The cylinder is now fully attached and ready to use. 1.0 Antechamber Controls These steps assume the atmosphere in the glove box is purified and not open to the outside air. Security Code for glove box controls: 7990 1.1 Large Antechamber 1.1.1 Inserting an Item into the Glovebox 1. Important: Verify the antechamber contains purified gas before opening the inner door. 2. Ensure the inner door and refill valve are fully closed. 3. Open the outside door and insert the object into the Antechamber. 4. Close the outside door and open the vacuum valve and evacuate to -29 in. Hg. 5. Close the vacuum valve and open the refill valve until pressure reaches -15 in. Hg. 6. Repeat evacuation and refill at least 3 times, finally reaching atmospheric pressure. 7. Close the refill valve and open the inner door to remove the object from the chamber. Page 2

1.1.2 Removing an Item from the Glovebox 1. Ensure that the outside door and refill valve are fully closed. 2. Open the inner door. 3. Insert the object into the Antechamber and close the inner door. 4. Open the outside door and remove the object. 5. Purge the antechamber before re-opening the inside door. 1.2 Small Antechamber 1.2.1 Inserting an Item into the Glovebox 1. Ensure the inner door and 3-way valve for mini-chamber are fully closed. 2. Open the outside door and insert the object into the Antechamber. 3. Close the outside door and turn 3-way valve to evacuate and evacuate to -29 in. Hg. 4. Turn the 3-way valve to refill until pressure reaches -15 in. Hg. 5. Repeat evacuation and refill at least 3 times. 6. Close the 3-way valve and open the inner door to remove the object from the chamber. 1.2.2 Removing an Item from the Glovebox 1. Important: Verify the antechamber contains purified gas before opening the inner door. 2. Ensure that the outside door and 3-way valve are fully closed. 3. Open the inner door and insert the object. 4. Close the inner door. 5. Open the outer door and remove the object. 6. Purge the antechamber before re-opening the inside door. 2.0 Removing Front Panel to Expose Glove Box 1. Verify that toxic materials will not be present inside of glove box at the time. 2. If planning to purge glove box and reset inert Argon atmosphere, verify that at least six (6) full tanks of argon are available for the glove box. 3. Close valve on argon tank by turning the valve clockwise. Page 3

4. Turn off power to glove box my pressing switch on front-facing panel located beneath antechamber. 5. Raise light fixtures over the top of the glove box. Light fixtures rotate along hinge. 6. Remove gloves from the front panel by removing the two bands securing the glove and then pulling off the glove itself. 7. While carefully holding onto the front panel, open each safety lock sealing the front panel and the glove box. IMPORTANT: Start at the top, work down the sides, and finally unlock the bottom. 8. Warning: Avoid letting the panel edge contact the sealing on front of glove box. This may lead to a scratch that prevents the seal from forming when system restarts. Be extra careful when removing the front panel from the glove box. Set the panel aside. 3.0 Restarting Glove Box after Exposing Atmosphere 1. Make sure at least 6 full argon tanks are on hand to purge and refill the glove box. 2. Verify that all seals are secured on front panels, each antechamber door is tightly closed, exhaust system is functioning, and cooling system is functioning (if applicable). 3. Note: If a front panel was removed from the glove box at any point, clean the seal with rubbing alcohol and apply high vacuum grease. 4. Turn on argon feed to the glove box by turning the top valve clockwise. 5. Turn on glove box power using the switch located on the front panel beneath the antechambers. 6. Perform a pressure check on the glove box (see section 5.0) 7. (step 5-8 is the purging method for the antechamber) Set the Working Maximum pressure to +10.0 mbar and the Working Minimum Pressure to +5.0 mbar. To do this, press the System Config. button and adjust the numeric values with the blue display. The working gas will enter the box automatically. 8. Turn on the vacuum pump by pressing the Start button on the main screen. 9. Open the evacuate valve on the Large Antechamber and allow the chamber to evacuate to evacuate to -30. 10. Close the evacuate valve and repeat for the mini antechamber. 11. Wait 10 minutes for the glovebox to stabilize. 12. Use the appropriate foot pedal to add/remove gas to/from the glovebox to reach desired pressure. 13. Important: See section 5.0 to perform a pressure test on the glove box and verify that all seals are secured. 14. It is extremely important to purge the glove box to displace as much air in the box as possible with inert gas. Important: Purging the glovebox requires 3 full cylinders of Page 4

working gas (6 cylinders for large glove box). Follow the steps in section 4.0 to purge the system. 4.0 Purging the Glovebox 1. Purging the glove box requires 3 full cylinders of working gas (6 cylinders for large glove box). Verify sufficient cylinders are in inventory (lab and 1 st floor garage). 2. Set Working Maximum pressure to +10.0 mbar and the Working Minimum Pressure to +5.0 mbar (see section 3.0). 3. Open the manual purge valve located on the back panel of the glove box beneath the antechambers. 4. Adjust the flow using the purge valve so that there is a steady flow of incoming gas and the box pressure is maintained at >1 mbar. 5. Create turbulence in the box to increase purge efficiency. This can be done by pressing in the gloves or using an electric fan or vacuum inside the glove box. 6. As the working gas cylinder empties, close the purge valve and change the cylinder. Open the valve on the cylinder and then re-open the purge valve. 7. Purge the system with the required number of cylinders or until Oxygen content is 25 ppm or less. Then close the manual purge valve. 8. The oxygen content might increase once the blower has been turned on (Main Screen -> Blower Control). If oxygen level rises above 100 ppm, stop the blower and continue to purge. 9. The blower system circulates the atmosphere through the purification system. It can be set to low speed during weighing, or set to high speed to recover from operator error. 10. Set the blower to 50% during normal operation. (100% = full blower speed) 5.0 Preforming a Pressure Test (Check Seal Quality) 1. Perform a pressure test anytime the front panel of the glove box is removed and replaced. A pressure test should also be performed periodically for safety and quality assurance. 2. Close purge valve located on back of glove box panel beneath antechamber. 3. Turn compressed cylinder on by rotating the valve on top of the cylinder counterclockwise. 4. Set Maximum Working pressure to 12.0 mbar and the Minimum Working pressure to 10.0 mbar. Page 5

5. Allow 10 minutes for the glovebox to stabilize at 11.0 mbar. Use the foot pedals to finely adjust the pressure. 6. Watch the current pressure in the glove box carefully for 25 minutes. If the pressure fluctuates more than 1 mbar, the system is not secured. Troubleshoot the seals/connections or contact Innovative Technology Inc. 6.0 Regenerating Purification Columns (Single-Column, offline) 1. Make sure that regen gas tank is sufficiently full. Column regeneration consumes approx. one-half of the tank. 2. Unless a second purification column is installed, the atmosphere of the glove box will be jeopardized so all experiments should be stopped and all chemical containers should be closed tightly. 3. Connect a cylinder of regenerating gas (3-7% Hydrogen in Nitrogen or Argon) to the regen gas connections (cylinder in left position if looking at them). Ensure connections are tightly secured. 4. Open the valve on top of the tank, allow the gas line to fill. Set the regulator pressure to 7 psi. 5. Turn off the blower on the main screen. 6. Press the Start button in the Regen Controls menu. The screen will prompt the user with two questions. 7. Are the column valves closed? Make sure that column valves are closed. These are located on top of the purification column inside the purification box. If the yellow valves are visible, the column valves are not closed. 8. Click Yes in the blue box next to the question. A click will be heard; the solenoid valves are opening to allow regen gas to flow. 9. Is Flow OK? At this point, the regen gas should have built up pressure and is ready to flow through the purification chamber. Adjust the needle valve on the front of the purification box to set the flow between 25-30 scfh. The flow indicator should be held steady in between these two value. 10. Click Yes in the blue box next to the question. Now the regen gas will stop flowing and the flow indicator will drop to 0. The flow will automatically restart after the column is sufficiently heated (default = 3 hours). 11. The remainder of the column regeneration is automatically controlled by the glove box. The default time is 14.5 hours. At the end of the cycle, the blower will automatically turn back on. Page 6