OPERATING AND MAINTENANCE MANUAL SBE 23 DISSOLVED OXYGEN SENSOR

Similar documents
Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

12mm Dissolved Oxygen Sensor Care and Use Instructions

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Dissolved Oxygen Guide

DISSOLVED OXYGEN SENSOR BT34i

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

Sea-Bird Scientific Carol Janzen, Ph.D. Oceanography David Murphy, M.S.E.E. Dan Quittman, M.S.E.E.

PRODUCT SHEET. Order probe only as RXPROBE02

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa)

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

MODEL GT820 OXYGEN SENSOR

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

Columbus Instruments

Optical Dissolved Oxygen Meter

OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT

BI-680 Online Dissolved Oxygen Controller Instruction Manual

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

Best Practice for Calibrating LTH Conductivity Instruments

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

Vinmetrica Dissolved Oxygen System User Manual

D-Opto Dissolved Oxygen Sensor Operation Manual

OxyProbe II Instruction Manual

D-Opto. Dissolved Oxygen Sensor. Operation Manual (Software Version 1.0)

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Optical Dissolved Oxygen Meter

Heavy Duty Dissolved Oxygen Meter

OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

Water Samplers, Water Sampling with Internal Recording, Cabling, and Deployment

SDX Submersible Depth Transmitter User Manual

STERILIZABLE DISSOLVED OXYGEN SENSORS

Sensor Stability. Summarized by Breck Owens David Murphy, Sea-Bird Electronics AST-14 Wellington March 18-21

Dissolved Oxygen Sensor

Radnoti Liver Perfusion System

D-OptoLogger Dissolved Oxygen Logger Operation Manual

LogIT Dissolved Oxygen Probe

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017

QUICK START GUIDE TO DRINKING WATER MONITORING

RDO Optical Dissolved Oxygen Sensor

SDX Submersible Depth Transmitter User Manual

DPC-30 DPC-100. Reference Manual

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

Overview. 2 Module 13: Advanced Data Processing

Mass Flow Controllers

Dissolved Oxygen/Temp Waterproof Tester

MODEL GENERA PURGE TURBINE INSTR DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA

Victoreen B. Operators Manual. Image Intensifier Ion Chamber

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Best Practice Guide, Servomex 2700

Aerobic Respiration. Evaluation copy

_ pressure transducers. User Manual

Gas Pressure Sensor (Order Code GPS-BTA)

Brooks TMF-Zone 1 Models 5816-Ex/5864-Ex

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

Operation manual Oxygen sensor

AMS 2710 PCB pressure sensor module with V output

Quick Start Guide LIQ-QSG-499ACL-03, Rev I July Rosemount 499ACL-03. Monochloramine Sensor

Instruction Manual HI Portable Dissolved Oxygen Meter. DO meter HI 8043 STB C OFF O 2 ZERO SLOPE C. ins rum n s

FX134, FX1234, FX3030 Problem and Solutions

Triton DO8 Dissolved Oxygen Sensors ELECTRO-CHEMICAL DEVICES

Membrane covered optical CO 2 sensor

ASTM D86 - Procedural Outline

Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip.

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

Unisense In Situ UniAmp Connector System

DIGIVAC 22W Series INSTRUCTION MANUAL. Wide Range Linearized Vacuum Transmitter and Controller YOU MUST READ THIS MANUAL BEFORE USE

Introduction The Smart Q Oxygen Adaptor enables an Oxygen electrode to be connected to an EasySense unit.

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003

Instruction Manual PN SO/rev.A April Model 1181SO. Dissolved Oxygen Two-Wire Transmitter

Sensoric 4-20 ma Transmitter Board Operation Manual

Engineering Note. Algorithms. Overview. Detailed Algorithm Description. NeoFox Calibration and Measurement. Products Affected: NeoFox

Precision level sensing with low-pressure module MS

NT High-temperature flowmeter model Simultaneous flow and pressure outputs for high-temperature applications

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature

Procedures for correcting in situ CTD data and results obtained during the NEAR-GOOS Cross-Basin Climate Monitoring Section project

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

Overview. Front Panel: Keypad and Display

Selecting the right pressure sensor for your application

311XL SERIES. 0-2 PPM Full Scale Trace Oxygen Analyzer DANGER

Kube Engineering Quinn E. Eddins, B.S.E.E Phone

HydroCAT Manual. Sea-Bird Coastal NE 20 th Street Bellevue, WA

Measurements of k L a. Steady-State Mass Balance Method:

GasSense NDIR User Manual

PRACTICAL TIPS FOR TRACE-LEVEL OXYGEN MONITORING WITH ELECTRO-CHEMICAL SENSORS

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Instruction and Maintenance Manual

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS

Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

AMT-Ex Dewpoint Transmitter

Transcription:

OPERATING AND MAINTENANCE MANUAL SBE 23 DISSOLVED OXYGEN SENSOR Sea-Bird Electronics, Inc. 1808-136th Place NE Bellevue, Washington 98005 USA Phone: (206) 643 9866 Telex: 292915 SBEI vers. 000

L I M I T E D L I A B I L I T Y S T A T E M E N T Extreme care should be execised when using or servicing this equipment. It should be used or serviced only by personnel with knowledge of and training in the use and maintenance of oceanographic electronic equipment. SEA-BIRD ELECTRONICS, INC. disclaims all product liability risks arising from the use or servicing of this system. SEA-BIRD ELECTRONICS, INC. has no way of controlling the use of this equipment or of choosing the personnel to operate it, and therefore cannot take steps to comply with laws pertaining to product liability, including laws which impose a duty to warn the user of any dangers involved in operating this equipment. Therefore, acceptance of this system by the customer shall be conclusively deemed to include a covenant by the customer to defend, indemnify, and hold SEA-BIRD ELECTRONICS, INC. harmless from all product liability claims arising from the use of servicing of this system.

DESCRIPTION Model SBE 23 Dissolved Oxygen Sensor Operating Instructions and Maintenance Manual The SBE 23 Dissolved Oxygen Sensor uses a 'Beckman' type high-pressure polarographic oxygen electrode as now manufactured by Sensor-Medics Corporation. The sensor also contains a thermistor for the purpose of determining internal (compensating) temperature. Model SBE 23-01 is intended for use to 3400 meters (5000 psi), while Model SBE 23-02 may be deployed at up to 6800 meters (10,000 psi). The SBE 23 (schematic dwg. no. 30507, assembly dwg. 40166) contains the electronics necessary to polarize the DO sensor, and convert the resulting oxygen-dependent current to a high-level voltage. A second part of the interface circuitry converts the compensating thermistor resistance to a voltage proportional to sensor internal temperature. The interconnections between the electronics, sensor, and the bulkhead I/O connectors are shown on assembly dwg. no. 40016. Note: drawing 40016 also shows a ph electrode (not installed on SBE 23). The SBE 23 is designed for operation from a single-polarity power supply of 6-16 volts. Current consumption is about 3.5 milliamperes. The sensor provides two output voltages. The first (Vdo) is proportional to the electrode current and is in the range of 0-5 volts, approximately corresponding to 0 to 1.3 microamperes. The second output voltage (Vto) is proportional to the internal temperature of the sensor and has a nominal range of 0 to +5 volts for temperatures of -5 to +35 degrees C. Consult the sensor calibration sheet for specific conversion factors. OPERATION Connect a DC power source to the sensor's bulkhead connector as follows: looking at the 6 pin bulkhead connector on SBE 23 (Impulse type AG306) Ground O Vdo o o DC power, 10-20 volts no connection o o no connection o Vto If the connector pigtail supplied with the SBE 23 (Impulse Enterprise type AG306) is used, align the raised bump on the side of the rubber molded connector with the large pin on the bulkhead connector.

The color code with the supplied pigtail (pigtails supplied by manufacturers other than Impulse may have different color codes) is as follows: Pin 1 Ground white Pin 2 Vdo black Pin 3 no conn blue Pin 4 Vto orange Pin 5 no conn green Pin 6 Power red Connect a voltmeter between Vdo and Ground. Apply power. The voltage at Vdo should begin reading about 5 volts. After several seconds, the voltage will start to decrease until an equilibrium value of about 2.5-3.5 volts is produced. This voltage is proportional to the partial pressure of oxygen in air. Connect the voltmeter between Vto and Ground. The reading will be about 3 volts at room ambient temperature. INTERPRETATION OF SENSOR RESPONSES Compute the sensor current and sensor compensation temperature using the relationships and coefficients given on the accompanying calibration sheet. The new algorithm by Owens and Millard (1985) may be used to convert SBE 23 oxygen sensor data to oxygen concentration. OXYGEN ALGORITHM The algorithm has the following form: OX = [Soc * (oc + tau * doc/dt) + Boc] * OXSAT(T,s) * exp(tcor * [T + wt * (To - T)] + pcor * p) where OX dissolved oxygen value in ml/l T water temperature in deg C s salinity in parts per thousand p pressure in decibars oc oxygen current in microamps Boc oxygen current bias Soc oxygen current slope To oxygen sensor internal temperature in deg C wt weighting fraction of oxygen sensor internal temperature tcor temperature correction factor for membrane permeability pcor pressure correction factor for membrane permeability tau oxygen sensor response time OXSAT(T,s) oxygen saturation value after Weiss (1970). Values for tcor, pcor, tau and wt are taken from the Beckman polarographic oxygen sensor technical memorandum:

tcor = -0.033 pcor = 1.50e-4 tau = 2.0 wt = 0.67 Values for Boc and Soc at the time of factory calibration are shown on the calibration sheet supplied with the sensor. RECALIBRATION For recalibration, the program OXFIT, written for the IBM PC/XT/AT and compatibles, computes the value for Soc, oxygen current slope. Determination of sensor zero (Boc) To measure the zero oxygen response of the oxygen sensor, flush the sensor with an inert gas (e.g. N 2, Ar, He) or place the sensor in a 2%- 5% solution of sodium sulfite (Na 2 SO 3 ). When using an inert gas, make sure that there are no air leaks; when using the sodium sulfite solution, make sure that there are no air bubbles trapped on the oxygen sensor membrane. In our experience, the sodium sulfite method produces better (consistently lower) results. Allow the sensor to stabilize and measure Vdo. Compute the sensor current (oc) which will typically be 0.05 microamperes or less. Since the Owens-Millard algorithm assumes that Boc + oc = 0, Boc must be chosen = -oc. Determination of sensitivity slope (Soc) To determine the sensitivity slope of the sensor, place the oxygen sensor and a temperature sensor (accuracy about +/- 0.1 degrees C) in a bath of air saturated water (use an aquarium bubbler to oxygenate the water). The oxygen sensor must be pumped (approximately 15 ml/second) to avoid formation of a low oxygen boundary layer at the oxygen sensor membrane. Wait until the oxygen sensor has had time to come to thermal equilibrium with the air saturated water (approximately 30 minutes). Read Vdo and Vto, compute the water temperature and oxygen sensor current, and verify that the internal oxygen sensor temperature is within about 0.5 degrees of the water temperature. For most accurate results the temperature of the air saturated water should be as close as possible to the temperature of the water at which the oxygen measurements will be made. However, note that it will be difficult to obtain equilibrium values at temperatures other than room ambient unless a temperature-controlled bath is employed. Determine local atmospheric pressure. Run the program OXFIT to compute Soc. The algorithm used by OXFIT is described in Appendix A. OXFIT will prompt for the following:

local barometric pressure (millibars) this is the pressure that would be read on a barometer (not corrected to sea level) water temperature (deg C) water temperature read by the temperature sensor (bath temperature, not compensating temperature oxygen current (microamps) air-saturated water value boc enter the value of boc determined previously MAINTENANCE and STORAGE The sensor module (the brown cylindrical cartridge in the plastic part of the SBE 23) has a working life of at least 6 months if kept moist. If the sensor is stored in dry air, however, operating life may be appreciably shorter. When possible, store the sensor with a loop of Tygon tubing containing distilled or deionized water. Care must be taken to avoid fouling the oxygen membrane with oil or grease. It is recommended that the oxygen sensor be flushed with distilled water after each use and be kept filled with distilled water between use. An additional important benefit of keeping the sensor closed with Tygon is to keep air born contaminants (of which there are an abundance on most research vessels) from entering the sensor. If it is not practical to keep the sensor filled with distilled water between use (for example, in Arctic environments where freezing is a hazard), at least flush it with clean fresh water (preferably distilled) and close the cell with Tygon. Also, remember to keep the Tygon in a clean place (so that it does not pick up contaminants) while the instrument is in use. For routine cleaning, fill the sensor with a 1% solution of Triton X- 100 and let soak for 30 minutes. This is most easily done by using a length of 7/16" Tygon tubing to form a closed loop including the sensor. The teflon membrane should be pointing up so that an air bubble does not block the cleaning solution. After the soak, drain and flush with warm (not hot) fresh water for 1 minute. Refill the sensor with distilled water until the next usage. SENSOR MODULE REPLACEMENT When satisfactory calibrations are no longer obtained, or there is evidence of calibration drift, follow the procedures in the Disassembly (Reassembly) Instructions supplied with the sensor. Replacement modules (P/N 147737) may be obtained from Sea-Bird Electronics, or directly from the manufacturer:

Sensor-Medics Corporation 1630 South State College Blvd Anaheim, CA 92806 800-231-2466 APPENDIX A Correction factor for non-standard atmosphere: nsa(t,bp) = (bp/p0) * (1 - (ph 2 O/bp) / (1 - ph 2 O/p0) bp = barometric pressure in kilopascals p0 = 101.325 kilopascals ph 2 O = water vapor pressure in kilopascals T = water temperature in deg C ph 2 O = exp[(-216961 * X - 3840.7) * X + 16.4754] X = 1/(T + 273.15) For air saturated water at the surface: ((Soc * oc + Boc) / nsa(t,bp)) * exp(tcor * T) = 1 Soc = {[nsa(t,bp) / exp(tcor * T)] - boc } / oc APPENDIX B Computation of OXSAT OXSAT(T,s) = exp(a1 + A2*(100/T) + A3*ln(T/100) + A4*(T/100) + s * (B1 + B2*(T/100) + B3*(T/100)*(T/100))) The units are ml/l, the oxygen saturation value is the volume of the gas (STP) absorbed from water saturated air at a total pressure of one atmosphere, per unit volume of the liquid at the temperature of measurement. where: s = salinity in parts per 1000 T = deg C + 273.15 (absolute temperature) A1 = -173.4292 A2 = 249.6339 A3 = 143.3438 A4 = -21.8492 B1 = -0.033096 B2 = 0.014259 B3 = -0.00170

REFERENCES Owens, W. B., and R. C. Millard Jr., 1985: A new algorithm for CTD oxygen calibration. J. Physical Oceanography., 15, 621-631. Gnaiger, E., and H. Forstner, Ed., 1983: Polarographic Oxygen Sensors: Aquatic and Physiological Applications. Springer-Verlag, 370 pp. Weiss, R. F., 1970: The solubility of nitrogen, oxygen and argon in water and seawater. Deep-Sea Res., 17, 721-735. Fletcher, R., 1972: Fortran subroutines for minimization by quasi-newton methods. Rep R7125 AERE, Harwell, England, 28 pp. Millard, R, C., Jr., 1982: CTD calibration and data processing techniques at WHOI using the 1978 practical salinity scale. Proc. Int. STD Conference and Workshop, La Jolla, Mar. Tech. Soc., 19 pp.