Dissolved Oxygen/Temp Waterproof Tester

Similar documents
DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Heavy Duty Dissolved Oxygen Meter

12mm Dissolved Oxygen Sensor Care and Use Instructions

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

BI-680 Online Dissolved Oxygen Controller Instruction Manual

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

PRODUCT SHEET. Order probe only as RXPROBE02

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

HI 9143, HI 9143M HI 9145

Oxygen Sensor with Electrode

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

Optical Dissolved Oxygen Meter

Bante904 Benchtop Conductivity/Dissolved Oxygen Meter Instruction Manual

Optical Dissolved Oxygen Meter

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

LogIT Dissolved Oxygen Probe

Aqua-DP. Aqua-DP Waterproof Dissolved Oxygen - ph -mv Temperature Meter. TPS Australia - Quality hand-made instruments since 1968.

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature

Instruction Manual DO 6. Economy Hand-held Dissolved Oxygen Meter. 68X Rev 2 Apr 08. Technology Made Easy... Part of Thermo Fisher Scientific

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

SS32L Dissolved O 2 Probe

Vinmetrica Dissolved Oxygen System User Manual

DISSOLVED OXYGEN SENSOR BT34i

NEULOG OXYGEN LOGGER SENSOR GUIDE

Instruction Manual HI 9146 Dissolved Oxygen and Temperature Meter

Instruction Manual DO 6+ Dissolved Oxygen/Temp

User Manual Pocket Pro + Multi 1

COM280. Technical Information. Oxygen measurement Battery-driven pocket meter for dissolved oxygen measurement

Instruction Manual HI Portable Dissolved Oxygen Meter. DO meter HI 8043 STB C OFF O 2 ZERO SLOPE C. ins rum n s

Dissolved Oxygen Probe (Order Code DO-BTA)

Dissolved Oxygen Probe (Order Code DO-BTA)

TQC Bresle Chloride test 2008

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017

STERILIZABLE DISSOLVED OXYGEN SENSORS

Instruction Manual. DO 700 Dissolved Oxygen/ºC/ºF Bench Meter. Technology Made Easy...

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

Introduction The Smart Q Oxygen Adaptor enables an Oxygen electrode to be connected to an EasySense unit.

Aerobic Respiration. Evaluation copy

Micro600DO_ZI INST /1/17 13:16 Page 1. Palintest. Micro 600 DO Meter. Palintest ZI INST 1240

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

Dissolved Oxygen in Water. Evaluation copy. Table 1. Temperature Range ( C) Trout Smallmouth bass Caddisfly larvae

Visit us at Code

SensoDirect Oxi200 Dissolved Oxygen Meter Operating instructions

MARTINI. instruments. THANK YOU FOR CHOOSING. Sales and Technical Service Contacts:

CCS240 and CCS241. Technical Information. Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly

Experiment C-6 Gas Solubility

Pool and Spa. Water Testing

User Manual DOCUMENT # Pro20 Pro20i USER MANUAL. English

Instruction Manual. Multiparameter Benchtop Meter

PRO20 USER MANUAL. English. Français. Español. Deutsch

Short manual testo

Manufactured by: AAA

D-OptoLogger Dissolved Oxygen Logger Operation Manual

testo 511 Bedienungsanleitung Instruction manual Mode d emploi Manual de instrucciones Manuale di istruzioni Manual de instruções

& Dissolved Oxygen Sensor. Instruction Manual

YSI Model 550a Dissolved Oxygen (DO) Meter

QUICK START GUIDE TO DRINKING WATER MONITORING

SAUTER FK Version /2016 GB

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

YSI-85_maint-cal Lab Version Page 1 of 13

STANDARD OPERATING PROCEDURE MAINTENANCE/CALIBRATION CHECKS OF DISSOLVED OXYGEN PROBES

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

Relative Humidity Calibration Kit

Operating manual for the pressure transducer unit CCS 1000

Three Lakes Association Hydrolab for Dummies. Table of Contents:

T02.0.3X.6C-06 Page 1 of 6. Operating Manual for Dissolved Oxygen-Measuring transducer OXY 3610 MP

DO_maint-cal Page 1 of 13

Altimeter and Compass Watch Instruction Manual

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

Dissolved Oxygen Sensor

HI 2314 HI 2315 HI 23151

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

DT 630 ALTIMETER, BAROMETER AND COMPASS WATCH OPERATING INSTRUSTIONS

D i s s o l v e d O x y g e n M e a s u r e m e n t

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications

Euclid Creek Volunteer Watershed Monitoring Program Directions for Performing Physical/Chemical Tests

YSI Model 52 Dissolved Oxygen Meter

Elcometer 138. Bresle Kit and Patches. Operating Instructions

5 Function Indicator. Outside Air Temperature (C) Outside Air Temperature (F) Pressure Altitude Density Altitude Aircraft Voltage STD TEMP SL 15000

AIR FLOW ANEMOMETER INSTRUCTION MANUAL

COOPERATIVE LAKES MONITORING PROGRAM TRAINING FOR. Dissolved Oxygen and Temperature

Sampling Equipment Catalog. MicroPurge Low-Flow Pump Control

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Compact Manometer. Pressure measurements can easily be taken any time, anywhere. POWER LIGHT

Instruction Manual Series Benchtop Meters. ph 2700 ION 2700 CON 2700 DO 2700 PC Technology Made Easy...

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Elcometer 138. Bresle Kit and Patches. Operating Instructions

Oxygen Measurement Systems High Reliability and Wide Application Coverage

Relative Humidity Calibration Kit

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL

Operating manual. Oxi 3205 O mg/l C. [Sal] Dissolved oxygen (D.O.) meter

Operating manual. TriOxmatic 702 IQ. IQ SENSOR NET D.O. sensor

130 SSP Soluble Salt Profiler

Unisense In Situ UniAmp Connector System

Transcription:

User's Guide AM-DO-01 Battery Compartment Cap Main Display Dissolved Oxygen/Temp Waterproof Tester Lanyard Connection old/max/min Button Function Mode Button Power / Calibration Button Probe Coller Probe Module Protective Cap Probe Cap

Index Introduction Features Specifications Accessories Device Description Probe Description Display Description Preparation Calibration Advanced Setting Membrane Cap Replacement

Introduction : Thank you for purchasing/ buying our product, the microprocessorbased waterproof DO/O2/Temp tester. You can now measure a wide range of Dissolved Oxygen and temperature, with a replaceable probe. Please read the manual carefully before using the tester. Features : A large LCD, which displays simultaneously DO or O2 and temperature A Waterproof IP-57 standard having a rugged design convenient for field use. The tester floats on water. Automatic temperature Compensation, Manual Salinity (MSC) and Altitude (MAC) Compensation. Both Centigrade and Fahrenheit degrees can be read by switching to required mode. The Icons for DO, O2 and for units, mg/l, ppm, %, oc, of are easily recognizable, especially during selecting function mode. Displays Maximum /Minimum value and data hold. There is an indicator for low battery and consumption. The DO probe module is easily replaceable. Specifications : DO O2 Temp. Range 0~20.00 mg/l 0~200.0 % 0~90 C 0~20.00 ppm ±0.2+1 digit Accuracy ±0.2+1 digit ±0.2+1 digit 0.1 C Resolution 0.01 mg/l 0.10 % Compensation ATC:0~50 C MSC: 0~50 ppt MAC: 0-20000 ft Accessories : Meter, Membrane Cap x 2,Electrolyte x 50 ml, Plastic burette, Lanyard, Battery (has been installed), Introduction manual, Carrying case. 1

Probe Description DO Probe Membrane Cap Anode Cathode Display Description Measuring Value mg/l% us ms ppt PPM mv p Unit Calibration Mode Maximum value Battery Consumption Indicator Minimum Value old Reading DO O2 COND SALT TDS ORP P MAX CAL MIN CAL C F ATC Function Mode Temperature Indicator Automatic Temperature Compensation 2

Functions of Keyboard : CAL MODE M Power/calibration 1. Press Button to switch power On or Off. 2. Press and hold button to enter calibration mode. Function Mode 1. Press to switch DO (mg/l), DO (ppm) and O2 (%) 2. Press and hold button to change degree oc or of old/max and Min 1. Press Button to enter hold mode 2. Press and hold button to enter maximum/minimum mode. In this mode press button light to get Maximum and Minimum value. 3. Press old button again to exit this mode and return to measurement mode. Preparation : 1. Remove the protection cap and probe cap from meter. 2. Remove the membrane cap carefully. 3. Fill the membrane cap with the electrolyte solution up to the bottom of the threads on the inside of the cap, (please see step 8~11 in Membrane Cap Replacement below for details). 4. Press CAL button to turn the meter power on. Calibration 1. Remove the probe cap. Press CAL to turn on power and then press MODE to choose O2 mode. Wait for 10 to 30 minutes for probe to polarize. After the probe has completely polarized, the reading should be approx. 101.7% (saturation) 2. While holding the probe in air, press and hold CAL for 3 sec. to enter the calibration mode. The display will appear CAL and flash 101.7%. When the display stops flashing, it should indicate SA, then END. This is when calibration ends. The display now should return to measurement mode. 3. Optional Zero Oxygen calibration. (This calibration improves measurement accuracy for very low or very high DO measurements). Place the probe in Zero oxygen solution, such as 5% sodium sulfite, wait for stability and press and hold CAL to enter calibration. Stability in a zero solution may take many minutes, depending on the probe history. Note : 1. If the tester display is not 0% when the probe is not connected, calibrate it in air without probe to bring it to 0%. 3

Measurement <DO> 1. Remove the probe cap and turn on the power. Wait for 10 minutes to 30 minutes for probe to polarize. The reading should be approx. 101.7% (saturation) after the probe has completely polarized. 2. Select the desired units of measure by pressing MODE until they are shown in the display. 3. Place the probe in the sample to be measured. Stir the probe in the sample to remove any trapper air bubbles from the membrane surface. 4. Allow the meter to stabilize, for the final measurement value. Note : 1. The larger the difference in temperature between probe and the solution the longer it will take for the reading to stabilize. Stabilization time can vary from ten (10) seconds to five (5) minutes. 2. After using, cover the probe with probe cap. The sponge contained in the cap should be moistened (not soaked) with DI (distilled water) or clean tap water. Function mode 1. Press M button to enter old Function Mode. The icon OLD will appear, and the value shown on display, can be locked. Press the button again to return to measurement mode. 2. Now to enter maximum and minimum function mode, press the hold M button until the icon MAX and MIN appears on the display. The maximum and minimum value will show on the display when the light button is pressed. To exit this mode, press and hold button until icon MAX M and MIN disappear, return to measurement mode. 3. Press and hold MODE button to change Degree C or F. Note : 1. The display cannot shut off by itself (auto) under MAX/MIN mode. 2. The battery needs to be changed/replaced,when battery indicator flashes on. Advanced Setting 1. Press MODE to choose DO mode 2. Press and hold CAL to enter menu. Now press M to choose SaLC (Salt Compensation) or ALtC (Altitude Compensation) and then press MODE to confirm. MODE 3. Press M ( up) and (down) to set Salt Compensation from 0 to 50 ppt, or set Altitude Compensation from 0 to 20K ft. 4

Membrane Cap Replacement 1. Do not touch the membrane as skin oils will interfere with the oxygen permeability rate of the membrane. Replace the cap carefully. 2. It is recommended that the probe remains attached to the meter during this replacement process. 3. Unscrew the cap firmly and carefully separate it from the probe. 4. Discard the old electrolyte solution from the Cathode and Anode. 5. Use the supplied polishing strip to clean, polish, shine, and/or remove scratches from the cathode. Be sure to moisten the cloth before polishing the cathode. Do not over-polish the sensitive gold cathode. 6. Set the new replacement membrane cap on the flat surface. Leave the cap in this position during the replacement process. 7. Fill the membrane cap with the electrolyte solution up to bottom of the threads on the inside of the cap. 8. Tap the membrane cap to release and prevent air bubbles in the electrolyte solution. 9. Keeping the cap in a fixed position on a flat surface, carefully insert the probe into the new cap by first dipping and removing the probe several times from the cap. With each dip, push the probe progressively deeper into the bonded cap. Finally, screw the probe slowly onto the cap until fully tightened. The dipping and removal technique minimizes the introduction of air bubbles into the electrolyte solution. Air bubbles in the electrolyte can affect measurements. 10.It is normal that excess electrolyte solution will leak out from the cap during this replacement, as it minimizes the introduction of air pockets. Clean off the excess electrolyte before use. Note : 1. We recommend changing the electrolyte when it becomes yellow. 2. Re-calibrate and re-polarize the meter once the membrane cap has been replaced or reinstalled. 5

Dissolved Oxygen/Temp Waterproof Tester