Indumax CLS54D. Operating Instructions

Similar documents
Operating Instructions Flowfit CUA252

Technical Information Conducal CLY421

Oxymax COS61/COS61D. Technical Information

Oxymax H COS21D. Technical Information

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications

Oxymax W COS51D. Technical Information. Digital sensor for measuring dissolved oxygen Membrane-covered amperometric sensor

Operating Instructions Turbimax CUS52D

Oxymax COS22/22D. Technical Information

Technical Information Oxymax COS22D/COS22

Best Practice for Calibrating LTH Conductivity Instruments

DTG - LCD Digital Temperature Gauge USER MANUAL

OUS31. Technical Information

SITRANS P measuring instruments for pressure

COM280. Technical Information. Oxygen measurement Battery-driven pocket meter for dissolved oxygen measurement

Type BBS-03, BBS-05, BBS-06, BBS-25

Technical Information Cleanfit CUA451

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25

OPTIBAR P 1010 C Technical Datasheet

High-performance submersible pressure transmitter For level measurement Model LH-10

L3 Pressure and Level transmitter

Sanitary RTD Temperature sensor TH17. Measuring System Sanitary RTD assembly with connection head TH17 for food and dairy applications.

CCS240 and CCS241. Technical Information. Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure

Pressure Measurement Single-range transmitters for general applications

Operating Instructions Oxymax COS22D

Improve Process Reliability

Technical Data Sheet MF010-O-LC

How to specify a product. Process Sensors and Mechanical Instruments

ph measuring system for hygienic process applications

DMK 458. Pressure Transmitter for Marine and Offshore. Ceramic Sensor. accuracy according to IEC 60770: standard: 0.25 % FSO option: 0.

Pressure Measurement Single-range transmitters for general applications

Pressure Measurement. Transmitters for basic requirements SITRANS P Z for gauge pressure. 2/4 Siemens FI Overview

EC214 EC215 - EC215R Bench Conductivity Meters

JUMO dtrans p30 Pressure Transmitter

GA24 Technical Datasheet

Pressure Transmitter P41

RTD Temperature sensor TH11

SITRANS. Pressure transmitter SITRANS P, Z series for gauge and absolute pressure. Introduction. Safety instructions 2.

Differential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN

OPERATING MANUAL CHEMTRAC 840M MANUAL RETRACTABLE PROCESS HOLDER

High-performance submersible pressure transmitter For level measurement in hazardous areas Model LH-20

High-performance submersible pressure transmitter For level measurement Model LH-20

Electropneumatic Positioner Type 4763 Pneumatic Positioner Type 4765

High-performance submersible pressure transmitter For level measurement Model LH-10

Mounting and Operating Instructions EB 3017 EN. Self-operated Regulators

9700 Transmitter Submersible hydrostatic level transmitter

Modular Analytical Panel System with Sample preparation for Steam and Power Applications

BSP Pressure Sensor Balluff GmbH, BA Marco Zorcic.

HI 2314 HI 2315 HI 23151

LMP 307. Stainless Steel Probe. Stainless Steel Sensor. accuracy according to IEC 60770: standard: 0.35 % FSO option: 0.25 % / 0.

Flow Measurement SITRANS F C

Safety Instructions Liquiphant FailSafe FTL80, FTL81, FTL85

Pressure Reducing Valve DMV 750 set range: bar

High-performance submersible pressure transmitter For level measurement Model LH-20

Type N-10, N-11 Hazardous Area Non-incendive Transmitters

High-performance submersible pressure transmitter For level measurement Model LH-20

ED 701 Hygenic Pressure Transmitter

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

Armatures for analytical probes

Pressure maintenance valve SDV-P / SDV-P-E

Digital Melting Point Apparatus

sitrans p Operating Instructions (de/en) Edition 03/2004

Suction lance SLP / SLP Operating instructions SLP Manufacturer: Keep the operating instructions for future use!

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

VG5 Supplementary instructions

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

Dimensions. Technical Data General specifications Switching element function Rated operating distance s n 5 mm

Pressure Reducing Valve DMV 755 set range: 1,0-9,0 bar

OSAT Series. Intrinsically Safe Infrared Temperature Sensor Operator's Guide

1 Overview. Pressure Measurement Transmitters for basic requirements. 1/16 Siemens FI SITRANS P220 for gauge pressure

Operating Instructions. Oxymax COS51D. Digital sensor for dissolved oxygen With Memosens technology BA00413C/07/EN/

T EK-SUB 4800C 19 mm Submersible Level Transmitter

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI US Edition

RTD temperature probes

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

Precision pressure sensor Basic version Model CPT6020

L (up to 25 m) Electronic part. working temp.: Ñ IP 66 / IP67. Mounting bracket ÐÑ. Sensor type. working temp.:

Operating Instructions. Oxymax W COS41. Dissolved oxygen sensor. BA284C/07/en/

x act i Precision- Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar

Rosemount 400 and 400VP

Electropneumatic Positioner and Pneumatic Positioner Type 3760 JIS

DK48 - DK800. Quick Start. Variable area flowmeter

Un-Pressurized Orefice Fittings FIO EZ. Parts List and Operation Instructions TECHNICAL MANUAL. Dn 2-6 Class Lbs

Operating Instructions Temperature Measuring System MF420-5T-100

Operating Instructions

A17W49_MASEM_F_EN MANUAL

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI

Technical Data. General specifications Switching element function Rated operating distance s n 2 mm

Technical Data. General specifications. Rated operating distance s n 5 mm

KDV HIGH ACCURACY GLASS TUBE ROTAMETER

Type 0404 / /2-way solenoid valve 2/2-Wege-Magnetventil Électrovanne 2/2 voies Operating Instructions Bedienungsanleitung Manuel d utilisation

Type 2000 INOX. Quickstart. English Deutsch Français. 2/2-way angle seat valve 2/2-Wege Schrägsitzventil Vanne à siège incliné 2/2 voies

C-Flow Coriolis Mass Flow Meters

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1

GA24. GA24 Handbook. Variable-area flowmeter. KROHNE 02/ MA GA24 R01 en

Air Sensor. FCS Ex. Manual. AQ Elteknik AB

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Transcription:

Operating Instructions Indumax CLS54D Inductive conductivity sensor with hygienic design for applications in the food, beverage and pharmaceutical industries and in biotechnology BA00508C/07/EN/01.12 71189125

Document function Safety messages The structure, signal words and safety colors of the signs comply with the specifications of ANSI Z535.6 ("Product safety information in product manuals, instructions and other collateral materials"). Safety message structure! DANGER Cause (/consequences) Consequences if safety message is not heeded Corrective action! WARNING Cause (/consequences) Consequences if safety message is not heeded Corrective action! CAUTION Cause (/consequences) Consequences if safety message is not heeded Corrective action NOTICE Cause/situation Consequences if safety message is not heeded Action/note Meaning This symbol alerts you to a dangerous situation. Failure to avoid the situation will result in a fatal or serious injury. This symbol alerts you to a dangerous situation. Failure to avoid the situation can result in a fatal or serious injury. This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or medium injury. This symbol alerts you to situations that can result in damage to property and equipment. Used symbols È ä 1 This symbol indicates a cross reference to a defined page (e.g. p. 1). È å 2 This symbol indicates a cross reference to a defined figure (e.g. fig. 2).

Indumax CLS54D Table of contents 1 Safety instructions.......... 4 1.1 Requirements for personnel............ 4 1.2 Designated use...................... 4 1.3 Workplace safety.................... 4 1.4 Operational safety................... 5 1.5 Product safety...................... 5 2 Identification.............. 6 2.1 Nameplate......................... 6 2.2 Scope of delivery.................... 6 2.3 Certificates and approvals.............. 7 3 Installation................ 8 3.1 Installation conditions................ 8 3.2 Installation........................ 10 3.3 Post-installation check............... 11 4 Wiring.................. 12 4.1 Connection to the transmitter......... 12 5 Commissioning............ 14 6 Maintenance.............. 15 7 Accessories............... 16 7.1 Measuring cables................... 16 7.2 Calibration solutions................ 16 8 Return.................. 17 9 Technical data............ 18 9.1 Input............................ 18 9.2 Performance characteristics........... 18 9.3 Environment...................... 19 9.4 Process.......................... 19 9.5 Mechanical construction............. 21 Index................... 22 Endress+Hauser

Safety instructions Indumax CLS54D 1 Safety instructions 1.1 Requirements for personnel Installation, commissioning, operation and maintenance of the measuring system must only be carried out by trained technical personnel. The technical personnel must be authorized by the plant operator to carry out the specified activities. The electrical connection may only be performed by an electrical technician. The technical personnel must have read and understood these Operating Instructions and must follow the instructions they contain. Measuring point faults may only be rectified by authorized and specially trained personnel. Repairs not described in the enclosed Operating Instructions may only be carried out directly at the manufacturer's or by the service organization. 1.2 Designated use Indumax CLS54D conductivity sensors are especially suitable for application in the food and beverage industry. The six-decade measuring range and the high chemical resistance of the materials in contact with medium permit to use this sensor in a number of various applications, e.g.: Concentration measurement of acids and bases Phase separation of products The sensors are used with Liquiline CM44x, Liquiline CM42 or Liquiline CM14. Any other use than the one described here compromises the safety of persons and the entire measuring system and is not permitted. The manufacturer is not liable for damage caused by improper or non-designated use. NOTICE Use in not specified applications Measurement errors and failures up to the breakdown of the measurement point possible Only use the product acc. to it s specification. Note the technical data of the nameplate. 1.3 Workplace safety As the user, you are responsible for complying with the following safety conditions: Regulations for explosion protection Installation instructions Local standards and regulations 4 Endress+Hauser

Indumax CLS54D Safety instructions 1.4 Operational safety Before commissioning the entire measuring point, make sure all the connections are correct. Ensure that electrical cables and hose connections are not damaged. Do not operate damaged products, and safeguard them to ensure that they are not operated inadvertently. Mark the damaged product as defective. If faults cannot be rectified, the products must be taken out of service and secured against unintentional commissioning. 1.5 Product safety The product is designed to meet state-of-the-art safety requirements, has been tested and left the factory in a condition in which it is safe to operate. Relevant regulations and European standards have been observed. Endress+Hauser 5

Identification Indumax CLS54D 2 Identification 2.1 Nameplate The nameplate can be found on the sensor. The following information is provided on the nameplate: Order code Extended order code Serial number Cell constant (nominal value) Protection class Pressure specification at 20 C Continuous service temperature Compare the data on the nameplate with your order. To find out what CLS54D version you have, enter the order code indicated on the nameplate in the search screen at the following address: www.products.endress.com/order-ident 2.2 Scope of delivery The scope of delivery comprises: 1 sensor Indumax CLS54D in the ordered version 1 Operating Instructions BA00508C/07/EN 6 Endress+Hauser

Indumax CLS54D Identification 2.3 Certificates and approvals 2.3.1 Hygienic certificates FDA All materials in contact with medium are listet at FDA. EHEDG Certified for cleanability according to EHEDG document 2 The cleanability of a sensor also depends on the way of installation. To install the sensor in a pipe system use the appropriate and EHEDG certified flow assembly for the respective process connection. 3-A Certified according to 3-A Standard 74-04 ("3-A Sanitary Standards for Sensor and Sensor Fittings and Connections Used on Milk and Milk Products Equipment"). Biological reactivity (USP class VI) (optional) Certificate (Certificate of Compliance) on biological reactivity tests according to USP (United States Pharmacopeia) part <87> und part <88> class VI with traceability of the materials in contact with medium. 2.3.2 Declaration of Conformity Declaration of conformity The product meets the requirements of the harmonized European standards. It thus complies with the legal requirements of the EC directives. The manufacturer confirms successful testing of the product by affixing the 4 symbol. Endress+Hauser 7

Installation Indumax CLS54D 3 Installation 3.1 Installation conditions The sensor must be completely immersed in the medium. Avoid bubbles in the area of the sensor. Abb. 1: Installation positions of conductivity sensors a0017691 After elbow pipes, turbulences can occur in the medium. Therefore it is necessary to install the sensor with a minimum distance of 1 m (3.3 ft) after an elbow pipe. The medium should flow through the flow opening of the sensor (see indicator arrow of the sensor body). The symmetrical mesauring channel allows a flow in both directions. 8 Endress+Hauser

Indumax CLS54D Installation In narrow installation conditions, the ion flow in the medium is affected by the pipe walls. This effect is compensated by the so-called installation factor. The installation factor can be entered in the transmitter or the cell constant can be corrected by multiplication with the installation factor to ensure correct measurement. The value of the installation factor depends on the diameter and the conductivity of the pipe as well as the sensor s distance from the wall. If the distance from the wall is sufficient (a > 15 mm, from DN 65), it is not necessary to consider the installation factor (f = 1.00). If the distance from the wall is smaller, the installation factor increases in case of electrically insulating pipes (f > 1) and decreases in case of electrically conductive pipes (f < 1). The installation factor can be measured using calibration solutions or it can be approximately determined from the diagram beside. Fig. 2: Installation of CLS54D a Sensor distance from the pipe wall a0017605 f 1.40 2 1.20 1.00 0.80 1 0 5 10 15 20 25 a [mm] Fig. 3: Dependance of installation factor f on wall distance a 1 Conductive pipe 2 Insulating pipe 0.20 0.39 0.59 0.79 0.98 a [inch] a0005441 Endress+Hauser 9

Installation Indumax CLS54D 3.2 Installation Fig. 4: Installation position of the sensor 1 Flow direction of medium a Sensor distance from the pipe wall a0018174 Install the sensor in such a way that the sensor opening is oriented in the flow direction of the medium. The sensor head must be completely immersed in the medium. Airset The sensor has been adjusted at the factory, an on-site compensation is not necessary. 10 Endress+Hauser

Indumax CLS54D Installation 3.3 Post-installation check Are sensor and cable undamaged? Is the installation position correct? Is the sensor installed via process connection and not suspended from the cable? Endress+Hauser 11

Wiring Indumax CLS54D 4 Wiring! WARNING Device is energized Improper connection can cause injury or death. The electrical connection must only be carried out by a certified electrician. Technical personnel must have read and understood the instructions in this manual and must adhere to them. Prior to beginning any wiring work, make sure voltage is not applied to any of the cables. 4.1 Connection to the transmitter The sensor is equipped with a fixed cable. For the wiring diagram refer to the operating instructions of the used transmitter. To extend the cable, a junction box and a CYK11 extension cable are required. 12 Endress+Hauser

Indumax CLS54D Wiring 75 (2.95) 63 (2.48) 1 2 A B Cable gland - shield clamped in gland Shield CYK11 to transmitter Sensor cable 85 PK 86 GY 87 BN 88 WH 97 GN 98 YE 80 (3.15) 1 2 20 (0.79) 20 (0.79) Fig. 5: Junction box A B a0017996 Endress+Hauser 13

Commissioning Indumax CLS54D To extend fixed cable sensors with M12 plug, use the measuring cable CYK11 with M12 female. 1 2 3 4 Fig. 6: CYK11 for extension with M12 connector pair 1 Transmitter 2 Measuring cable CYK11 with M12 connector pair 3 Measuring cable of CLS54D with M12 plug 4 CLS54D sensor a0017842 5 Commissioning Prior to initial commissioning, make sure that: The sensor has been installed correctly The electrical connection is correct 14 Endress+Hauser

Indumax CLS54D Maintenance 6 Maintenance Inductive sensors are less sensitive to soiling than conventional conductive sensors since there is no galvanic contact with the medium. However, dirt may collect in the flow opening (making it narrower), which changes the cell constant. In this case, an inductive sensor also requires cleaning.! WARNING Burning chemicals Danger of chemicals burns to the eyes and skin. Danger of damage to clothing and equipment. It is absolutely essential to protect the eyes and hands properly when working with acids, bases and organic solvents! Wear protective goggles and safety gloves. Clean away splashes on clothes and other objects to prevent any damage. Pay particular attention to the information provided in the safety data sheets for the chemicals used. Clean away fouling on the sensor as follows depending on the particular type of fouling: Oily and greasy films: Clean with grease remover, e.g. alcohol, acetone, as well as hot water and dishwashing detergent if necessary. Lime and metal hydroxide buildup: Dissolve buildup with diluted hydrochloric acid (3 %) and then rinse thoroughly with plenty of clear water. Sulfidic buildup (from flue gas desulfurising or sewage treatment plants): Use a mixture of hydrochloric acid (3 %) and thiocarbamide (commercially available) and then rinse thoroughly with plenty of clear water. Endress+Hauser 15

Accessories Indumax CLS54D 7 Accessories 7.1 Measuring cables CYK11 Memosens data cable Extension cable for digital sensors with Memosens protocol Ordering as per product structure (--> Online configurator, www.products.endress.com/cyk11) Junction box M12 connector/cable Material: aluminum, painted Cable extension: Memosens sensors, Liquiline Order no. 71145498 Junction box cable/cable Material: aluminum, painted Cable extension: Memosens sensors, Liquiline Order no. 71145499 7.2 Calibration solutions Precision solutions, traceable to SRM (standard reference material) by NIST, for qualified calibration of conductivity measurement systems according to ISO 9000, with temperature table CLY11-B 149.6 μs/cm (reference temperature 25 C / 77 F), 500 ml / 16.9 fl.oz Order no. 50081903 CLY11-C 1.406 ms/cm (reference temperature 25 C / 77 F), 500 ml / 16.9 fl.oz Order no. 50081904 CLY11-D 12.64 ms/cm (reference temperature 25 C/ 77 F), 500 ml / 16.9 fl.oz Order no. 50081905 CLY11-E 107.0 ms/cm (reference temperature 25 C / 77 F), 500 ml / 16.9 fl.oz Order no. 50081906 16 Endress+Hauser

Indumax CLS54D Return 8 Return The device must be returned if repairs or a factory calibration are required, or if the wrong device has been ordered or delivered. According to legal regulations, Endress+Hauser, as an ISO-certified company, is required to follow certain procedures when handling returned products that are in contact with medium. To ensure swift, safe and professional device returns, please read the return procedures and conditions on the internet site: www.services.endress.com/return-material Endress+Hauser 17

Technical data Indumax CLS54D 9 Technical data 9.1 Input 9.1.1 Measured variable Conductivity Temperature 9.1.2 Cell constant k = 6.3 cm 1 9.1.3 Measuring ranges Conductivity Temperature recommended range: 100 μs/cm to 2000 ms/cm (uncompensated) 10 to +150 C (+14 to +302 F) 9.1.4 Temperature measurement Pt 1000 (class A acc. to IEC 60751) 9.2 Performance characteristics 9.2.1 Temperature response time t 90 26 s 9.2.2 Conductivity response time t 95 2 s 9.2.3 Measured error ± (0.5 % of measured value + 10 μs/cm) after calibration, valid for T < 100 C (212 F) ± (0.5 % of measured value + 25 μs/cm) after calibration, valid for T > 100 C (212 F) (plus inaccuracy of the calibration) 9.2.4 Repeatability 0.2 % of the measured value 18 Endress+Hauser

Indumax CLS54D Technical data 9.3 Environment 9.3.1 Ambient temperature range 20 to +60 C (-4 to +140 F) 9.3.2 Storage temperature -25 to +80 C (-13 to +176 F) 9.3.3 Relative humidity 5 to 95 % 9.3.4 Ingress protection IP 68 (10 m water column, 25 C, 168 h) / NEMA Type 6P 9.4 Process 9.4.1 Process temperature -10 to +125 C (14 to 257 F) 9.4.2 Sterilization 150 C / 5 bar (300 F / 72.5 psi) (max. 60 minutes) 9.4.3 Process pressure 12 bar (174 psi) up to 90 C (194 F) 8 bar (116 psi) at 125 C (257 F) Underpressure down to 0.1 bar (1.45 psi) absolute Endress+Hauser 19

Technical data Indumax CLS54D 9.4.4 Pressure-temperature diagram Abb. 7: Pressure / temperature load curve A short-time sterilization (max. 60 min.) a0017602 9.4.5 Flow velocity max. 10 m/s (33 ft/s) for pipe diameters 80 mm for low viscous media max. 5 m/s for pipe diameters 50 mm and < 80 mm for low viscous media 20 Endress+Hauser

Indumax CLS54D Technical data 9.5 Mechanical construction 9.5.1 Weight Depending on version 0.3 to 0.5 kg (0.66 to 1.1 lb.) plus cable 9.5.2 Material In contact with medium Not in contact with medium Virgin PEEK PPS-GF40 Stainless steel 1.4404 (AISI 316L) FKM, EPDM (seal) PEEK (cable gland) TPE (cable) 9.5.3 Chemical durability Medium Concentration PEEK Caustic soda NaOH 0 to 15 % 20 to 90 C (68 to 194 F) Nitric acid HNO 3 0 to 25 % 20 to 90 C (68 to 194 F) Phosphoric acid H 3 PO 4 0 to 15 % 20 to 80 C (68 to 176 F) Sulphuric acid H 2 SO 4 0 to 30 % 20 C (68 F) Peracetic acid H 3 C-CO-OOH 0.2 % 20 C (68 F) No responsibility is taken for the correctness of this information. Endress+Hauser 21

Indumax CLS54D Index A Accessories............................. 16 Approvals............................... 7 C Cable extension......................... 12 Calibration solutions...................... 16 Certificates.............................. 7 Cleaning............................... 15 Commissioning.......................... 14 D Declaration of conformity................... 7 Designated use........................... 4 E Electrical connection...................... 12 I Installation conditions...................... 8 M Maintenance............................ 15 Measuring cables........................ 16 N Nameplate.............................. 6 O Operational safety......................... 5 P Product safety............................ 5 R Return................................ 17 S Scope of delivery.......................... 6 T Technical data.......................... 18 Transmitter............................. 12 U Use................................... 4 W Wiring................................ 12 Workplace safety......................... 4 22

BA00508C/07/EN/01.12 FM6 / DT 71189125