Fused-Core TM Particle Columns provide UHPLC performance on any HPLC

Similar documents
Ascentis Express HPLC Columns with Fused-Core Technology

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

More Speed and Resolution for Current Methods

Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

Carrier Gases in Capillary GC

Carrier Gases in Capillary GC

Fast method for Ginseng Analyses using Agilent Poroshell 120 Columns Scaled from a Traditional Method

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Office VOC Mixture Test Report

Ensuring Best Possible UHPLC Performance Through an Innovative Capillary and Fitting Design

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Retention Time Locking: Concepts and Applications. Application

Experiment GC : Analysis of BTEX by GC-FID

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

A VALIDATED STABILITY-INDICATING RP-HPLC ASSAY METHOD FOR BOLDENONE UNDECYLENATE AND ITS RELATED SUBSTANCES

PicoChip. Nanospray System. Column Care & Use Manual

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

Dionex IonPac AS19-4µm

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

Chemical Analysis for Methyltestosterone

Rev. CO Packed Columns for Normal Phase Chromatography. TSKgel NH m INSTRUCTION MANUAL

Extended Application Note

SOP: Buck Scientific BLC-20P HPLC Operation

CONTACT DETAILS AND ORDER INFORMATION

Supporting information for J. Med. Chem., 1994, 37(5), , DOI: /jm00031a018

Setting a New Standard in Flash Chromatography Performance

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Introduction. Experimental

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

Prep LC Training Manual

Advances in Capillary Column Backflushing

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Running HPLC Methods on an ACQUITY UPLC System

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Appendix A COMPARISON OF DRAINAGE ALGORITHMS UNDER GRAVITY- DRIVEN FLOW DURING GAS INJECTION

Video Notes. LC Troubleshooting Series Retention Time Shifts

IJPAR Vol.6 Issue 1 Jan - Mar Journal Home page:

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

Best Practice for Identifying Leaks in GC and GC/MS Systems

Agilent Dimension Software for ELSD User Manual

May 1, By John Heim,Doug Staples

LC, RSLC, Bio RS and Dual Gradient Systems

Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

UP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule

Prep LC Training Manual S.V.

HPLC systems. Overview

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

LCMSMS Solutions For Steroid Analysis

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

no peaks missing peaks too small peaks increasing retention decreasing retention

BETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump

The instrument should have a mass range from 2 to 2000 m/z or better Interface

AKTA pure 25 New Owner s Intro

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

Packing Procedure Recommendation. For. YMC RP silica. ODS-A, ODS-AQ, Octyl, Butyl, YMCbasic

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think! There Are Many Reasons to Reduce Analysis Time

CHAPTER 3 Development and Validation of a RP-HPLC method for Dutasteride and its impurities in bulk drug

CTB3365x Introduction to Water Treatment

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

Your local gas generation partner. Precision series Modular gas generation solution for GC.

LEVANT WORMSEED FOR HOMOEOPATHIC PREPARATIONS CINA FOR HOMOEOPATHIC PREPARATIONS

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

Increasing the Throughput of Liquid Chromatography

Setting up and running a column

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

Sensitive Analysis of Genotoxins by HPLC-ECD Using Boron-Doped Diamond Electrochemical Cell

Ring-Closing Metathesis Approaches for the Solid- Phase Synthesis of Cyclic Peptoids

1. INTRODUCTION CH 3 CH3

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform

HPLC troubleshooting guide

Meteorology & Air Pollution. Dr. Wesam Al Madhoun

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

Longterm Detection of Anabolic Steroid Metabolites in Urine

FROG-4000 TM Chemical Analysis System

Using the Akta Prime plus October 22, 2012

Dissolution test results showing biorelevant media made from SIF Powder Original is the same as media prepared using methylene chloride

Gradients in ion chromatography. Optimal separation with shorter analysis time

FROG-4000 TM Chemical Analysis System

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

Your local gas generation partner. Precision series Modular gas generation solution for GC.

GCMSD-Headspace Analysis SOP

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Validation of the MRM pesticides from the Working Document SANCO/12745/2013 of November 2017 rev. 9(1)

Methylprednisolone detection in urine following local and oral administrations

OPUS 5-80R Pre-packed Chromatography Columns USER GUIDE

Transcription:

Fused-Core TM Particle Columns provide UPLC performance on any PLC Major Contributors and Co-Authors. Bell, W. Campbell, W. Way, K. Buckendahl Supelco a ivision of Sigma Aldrich

Content Influence on Resolution & Van eemter Curve PLC Evolution of particle design Fused-Core TM particle Technology Column performance - System compatibility (pressure requirements) - Comparison to Sub 2µm and 3µm materials 2 Fused-Core is a trademark of Advanced Materials Technology Inc. Ascentis is a trademark of the Sigma-Aldrich Corp.

Resolution in PLC Key Influence Factors 3 Efficiency N α k R= N 4 Retention k k +1.. Selectivity α-1 α - Increase column length - ecrease particle size (N ~ 1/d p ) - Change stationary phase - Change mobile phase solvent - Change buffer p - Increase (weaker solvent) - ecrease (stronger solvent) Resolution (R) Zhao, J.. and P.W. Carr. Analytical Chemistry, 1999. 71(14): p. 2623-2632 3.0 2.5 2.0 1.5 1.0 0.5 α N k 0.0 1.00 1.05 1.10 1.15 1.20 1.25 0 5000 10000 15000 20000 25000 0 5 10 15 20 25 α N k

van eemter-kurve: = A + B/u + Cu B = 2γ m C = (f 1 d P2 +f 2 d f2 ) A = 2λd P 4 u

Evolution of PLC Column Particle esign Irregular ifficult to pack, easily clogged and not very rugged Pellicular Thin porous layer showed low capacity Totally Porous Current PLC workhorse Fused-Core tm The future of PLC columns? 5 Fused-Core is a trademark of Advanced Materials Technology, Inc.

Pressure & Efficiency are reciprocal to the particle size Influence of d P on P >> N Effiiciency bar 6 30,000 25,000 20,000 15,000 10,000 5,000 0 400 350 300 250 200 150 100 50 0 0 5 10 15 20 0 5 10 15 20 d p (μm) N P 1 2 d p 1 d p 2x efficiency 4x higher pressure Smaller particles are not the simplest solution and have draw backs. d P (µm) psi Bar N 1.8 5889 406 27,500 2.5 3089 213 20,000 3 2118 146 16,500 5 769 53 10,000 10 189 13 5,000 15 87 6 3,750 20 44 3 2,500 10 cm column, 3 mm/s linear flow ow much separation power do you have commonly on your PLC- System?

Fused Core-Technology igher Efficiency 1 2 4 3 Agilent 1200 Ascentis Express C18, 15cm x 4.6mm, 2.7 µm 1.0mL/min, 254nm, RT, 10µL inj 4. Toluol N = 30,738 3. Benzol N = 31,696 2. Acetophenon N = 33,786 1. Uracil (dead time marker) Pressure = 183 bar (2690 psi) 0 1 2 3 4 min 7

Ascentis Express Fused-Core Particle Technology Fused-Core particle technology invented by Jack Kirkland Porous silica layer fused to solid core 0.5 µm Ascentis Express PLC columns use this newly engineered particle 2.7 µm 1.7 µm 2.7 micron Silica particle 1.7 micron solid core 0.5 micron porous shell C18 and C8 phases available at launch (RP-Amide later in 2007) 8

Fused Core Technology Reducing dispersion (increasing efficiency) A solid core at the center of a particle, the potential diffusion path length of an analyte molecule is effectively shortened A larger particle then can theoretically generate similar efficiency of a smaller totally porous particle without generating high backpressures 9

Fused-Core Vs. Totally Porous: Improved Peak Shape 0.5 µm Migration Path In Particle 1.5 µm 2.7 µm 1.7 µm 10 Shorter diffusion path reduces axial dispersion and minimizes peak broadening

System Compatibility Comparison of Pressure for ifferent Particles = A + B/u + Cu ΔP = 1000Fη ΔP = L t o πr 2 d 2 p 16,000 35.00 16,000 14,000 1.7 µm 30.00 12,000 ETP (μm) 25.00 20.00 15.00 Pressure (psi) 10,000 8,000 6,000 2.7 µm FC 10.00 4,000 3 µm 5.00 2.7 µm Ascentis Express 2,000 1 2 3 4 5 Mobile Phase Velocity (mm/sec) 2 4 6 8 10 12 Mobile Phase Velocity (mm/sec) 11

van eemter Ascentis Express 2.7 µm Vs. 3 µm Small Molecule Study: Naphthalene: M.W. = 130 8.00 Naphthalene 7.00 h = / d p h 6.00 5.00 4.00 3.00 Acentis C8 3 µm Ascentis Express C18 2.7 µm Ascentis Express C8 2.7 µm 2.00 1.00 12 Reduced plate hight to compare columns with diff. Particle sizes 0.00 0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 v (mm/s) Columns: 50 x 4.6 mm Mobile phase: 55/45 ACN/Water Sample: Naphthalene (20 µg/ml), dissolved in 55/45 ACN/water Injection: 5 µl etection: 254 nm Temperature: 30 C Instrument: Agilent 1100.

Smaller Particles Exponentially Increase Pressure Pressure rop Vs. Particle Size bar 450 400 350 300 250 200 150 10 0 50 0 1.8 µm ~ 400 bar (6,000 psi) 3 µm ~ 150 bar (2,250 psi) 5 μm ~ 50 bar (750 psi) 0 5 10 15 2 0 2 5 d p (μm) 13 10 cm column, 3 mm/s linear velocity

Frit Requirements for PLC columns Column Work horse igher Efficiency Particle size (μm) Frit size (μm) 5 2 3 0.5 UPLC 1.6-2.0 0.2 Ascentis Express 2.7 2 Ascentis Express 3µm Porus material Ascentis Express narrow particle distribution allows for 2 μm-frits Like 5 μm-columns Ascentis Express-Columns can efficiently be protected by 0.5 μm-inline-filter. 14

igh efinition () Resolution Compared to 5µm porus particle column Sub 2 µm columns can provide up to 1.55 x increase in resolution for an 8 fold increase in pressure Ascentis Express Fused-Core (FC) columns can provide 1.55 x increase in resolution for only a 4 fold increase in pressure dp N Rs factor* Pressure (psi) Pressure (bar) 5μm 10.000 1 750 50 3μm 14.000 1,18 2.250 150 2.7μm FC 24.000 1,55 2.800 190 1.8μm 24.000 1,55 6.000 400 15 10cm columns * RS normalised to 5µm column

Twice the Efficiency at the System Limit 2.7 µm Fused-Core Vs. 3 µm Totally Porous Particles Ascentis Express C18 2.7 µm 65 % ACN N = 237,740 p/m or N = 35,661 p/col Psi (approx 4,000) 1 2 3 4 Efficiency Comparison / Columns: 150 x 4.6 mm Mobile Phase: Acetonitrile / Water Flow: 1.5 ml/min Injection: 2.0 μl etection: 220 nm Agilent 1100 PLC System 1. p-hydroxy ethylbenzene 2. Napthalene 3. p-xylene 4. Biphenyl 0 2 4 Ascentis C18, 3 µm 72.5 % ACN N = 140,631p/m or N = 21,095 p/col Psi (approx 4,000) 1 2 3 4 0 2 4 16 NB: Ascentis C18 and Ascentis Express C18 have different Surface Chemistries

mau 0 20 40 60 80 100 120 140 mau 0 20 40 60 80 100 120 140 17 TO11/IP6A Carbonyl-NP Mix Almost Twice the Sensitivity Ascentis Express C18, 2.7 µm Peak 8 N = 260,720 p/m N = 39,108 p/col 0 2 4 6 8 10 12 Ascentis C18, 3 µm Peak 8 N = 146,587p/m N = 21,988p/col 8 0 2 4 6 8 10 12 8 Sensitivity Gap Column: 150 x 4.6 mm I.. Mobile phase: Ascentis Express C18 2.7 µm: 25:75, water: acetonitrile Ascentis C18 3 µm: 30:70, water: acetonitrile Flow rate: 1.0 ml/min. Temp.: 30 C et.: UV at 365 nm Injection: 1 µl Sample: 47285-U TO11/IP6A Carbonyl-NP Mix as indicated below in 40:60, water: acetonitrile Peak Is 1. Formaldehyde-2,4-NP (105 µg/ml) 2. Acetaldehyde-2,4- NP (76.4 µg/ml) 3. Acrolein-2,4- NP (63.2 µg/ml) 4. Acetone-2,4- NP (61.5 µg/ml) 5. Propionaldehyde-2,4- NP (61.5 µg/ml) 6. Crotonaldehyde-2,4- NP (53.6 µg/ml) 7. Butyraldehyde-2,4- NP (52.5 µg/ml) 8. Benzaldehyde-2,4- NP (40.5 µg/ml) 9. Isovaleraldehyde-2,4- NP (46.4 µg/ml) 10. Valeraldehyde-2,4- NP (46.4 µg/ml) 11. o-tolualdehyde-2,4- NP (37.5 µg/ml) 12. m-tolualdehyde-2,4- NP (37.5 µg/ml) 13. p-tolualdehyde-2,4- NP (37.5 µg/ml) 14. exaldehyde-2,4- NP (42 µg/ml) 15. 2,5-imethylbenzaldehyde-2,4- NP (35 µg/ml) NB: Ascentis C18 and Ascentis Express C18 have different Surface Chemistries

igher Sample Throughput on Traditional PLC Systems* 0 2 4 6 0 2 4 6 0 2 4 6 Conventional C18 25 cm x 4.6 mm I.., 5 μm 1.0 ml/min., N= 22147 Pressure: 128 bar (1880 psi) 0 2 4 6 0 2 4 6 0 2 4 6 0 10 20 30 40 Ascentis Express C18 10 cm x 4.6 mm I.., 2.7 μm 1.0 ml/min., N = 22694 Pressure: 167 bar (2450 psi) Method requirement was N >20.000 0 2 4 6 8 10 Ascentis Express C18 10 cm x 4.6 mm I.., 2.7 μm 1.5 ml/min., N = 21297 Pressure: 248 bar (3645 psi) *Agilent 1100 PLC System 18 0 2 4 6 8 10

igh Speed: Increased Flow Rate Ascentis Express Vs. sub-2 µm at System Pressure Limit Ascentis Express C18 0.3 ml/min 45 % acetone. 2130 psi N = 12,500 Sub-2 µm Column 2 0.3 ml/min 51 % acetone. 9500 psi N = 12,170 1 2 3 1.0 2.0 3.0 4.0 1 2 4 4 5 Ascentis Express C18 0.6 ml/min 45 % acetone. 4450 psi N = 8,500 1 2 4 Mobile Phase: water : acetonitrile; isoelutropic for β-estradiol Columns: 100 x 2.1 mm Flow: variable et: 200 nm Inj: 1µL Elution order: 1. Estriol 2. β-estradiol 3. Contaminant 4. Estrone 5. Estrone degradant 3 5 3 5 19 1.0 2.0 3.0 4.0 1.0 2.0 All column efficiencies have been compromised by system dispersion, however the pressure associated with the Ascentis Express column when run at twice the flow rate. It is still at only half that of the sub 2 µm column.

Twice the Efficiency at the System Limit: 4,000 6,000 psi System Ascentis Express C18 100 mm length 45 % acetonitrile 4500 psi N = 8,500 p/col C18 Sub-2 µm Column 50 mm length 46 % acetonitrile 5200 psi N = 3,790 p/col 1 1 2 2 3 4 1.0 2.0 4 5 Mobile Phase: water : acetonitrile; isoelutropic for β-estradiol Columns: 2.1 mm I Flow: 0.6 ml/min et: 200 nm Inj: 1µL Elution order: 1. Estriol 2. β-estradiol 3. Contaminant 4. Estrone 5. Estrone degradant Both column efficiencies have been compromised by system dispersion however N is double for the longer column as would have been predicted at the same pressure 3 5 20 0.2 0.4 0.6 0.8 1.0 1.2

21

Ascentis Express Achieve 100,000 Plates under 7,000 PSI 1 Column 15cm x 4.6mm Agilent 1200, 1.0mL/min, 254nm, ambient temp, 10uL 4. N Toluene = 30,738 3. N Benzene = 31,696 2. N Acetophenone = 33,786 Pressure = 183 bar (2690 psi) 2 columns 30cm N Toluene = 63,212 N Benzene = 64,785 N Acetophenone = 67,390 Pressure = 350 bar (5145 psi) N Toluene = 91,311 3 columns 45cm 2 3 4 N Benzene = 94,979 N Acetophenone = 101,586 Pressure = 473 bar (6953 psi) 0 2 4 6 8 10 12 14 22

Ascentis Express separates deuterated Isomeres Agilent 1200 55cm x 4.6mm (>100,000 N in 14 min) 60% Acetonitrile 1.1mL/min, 254nm, 50 C, 10uL 4 columns B B B Toluene N = 104,235 Benzene N = 106,096 Acetophenone N = 117,475 Pressure = 473 bar (7,000 psi) B 0 2 4 6 8 10 12 14 Toluol N = 105,657 10 Benzol N = 106,260 6 Benzol N = 105,235 Acetophenon N = 118,567 Pressure = 473 bar (7,000 psi) 10 0 2 4 6 8 10 12 14 23

24 euterated Analogues of Toluene and Benzene C 3 C 3 C 3 C 3

igh-efficiency Separations: euterated Analogues of Toluene and Benzene Length: 45 cm Pressure: 10500 psi N (acetophenone) = 73400 1.8 µm, C18 0 10 20 Length: 60 cm Pressure: 10550 psi N (acetophenone) = 113810 2.7 µm, Fused Core 25 0 10 20

26 Further examples of high resolution using Ascentis Express

Increased resolution over 3 µm particles Ascentis Express C18 (2.7μm) 6 0 2 4 6 Efficiency (N) Peak 9: 13160 2990 psi columns: 10 cm x 2.1 mm I.. mobile phase A: 100 mm ammonium acetate (p 7.0; titrated with ammonium hydroxide) mobile phase B: water mobile phase C: methanol online mixing: Ascentis Express C18: A:B:C = 10:28:62 3 µm, C18: A:B:C = 10:25:65 (isoelutropic for peaks 6 & 9) flow rate: 0.3 ml/min. temp.: 55 C det.: UV at 250 nm injection: 1 µl C18 3 µm 6 Efficiency (N) Peak 9: 8290 1750 psi Peak I 1. Uracil 2. Nordoxepin (50 mg/l) 3. esipramine (50 mg/l) 4. Nortiptyline (50 mg/l) 5. oxepin (50 mg/l) 6. Norclomipramine (50 mg/l) 7. Imipramine (50 mg/l) 8. Amitriptyline (50 mg/l) 9. Clomipramine (50 mg/l) 0 2 4 6 27

Carbonic Anhydrase Tryptic igest on Ascentis Express C18 40μL Injection 1.0mL/min flow @ 35ºC 215nm etection Response time >0.010min Gradient: MP A 5% ACN w/0.1% TFA MP B 80% ACN w/0.095% TFA 2 min hold @ 0% B 58 min gradient from 0% B to 50% B 20 min gradient from 50% B to 85% B Ascentis Express 15x4.6mm 2.7μm C18 15x4.6mm 3μm 0 10 20 30 40 50 60 70 80 0 10 20 30 40 50 60 70 80 28

Greater peak resolution with Ascentis Express Magnified regions: each pair is at the same scale 2 14 16 18 20 22 24 26 28 12 14 16 18 20 22 24 26 28 C18 3μm Ascentis Express C18 8 30 32 34 36 38 40 2829 30 32 34 36 38 40

ILIC on Ascentis Express Erythromycin and Related Compounds by O erythromycins.esp 1 3 3 C O 3 C O C 3 O 3 C C 3 O N C 3 2 4 3 C O O C 3 O O O C 3 C 3 O 0 5 10 15 Retention 3 C O O C 3 C 3 30 Column Name Ascentis Express Si Length 15 cm iameter 0.05 cm Particle Size 2.7 um Mobile Phase A 2 O; 100mM ammonium formate p Mobile Phase A 3 Mobile Phase B acetonitrile Gradient 90B Temperature 35 C Injection Volume 5 ul

Comparison of Ascentis Express ILIC and porous particle based silica phases Ascentis Express ILIC 1 2 3 4 1. Norephedrine 2. Ephedrine 3. Pseudoephedrine 4. Synephrine 1 2 3 4 Ascentis Silica 31 0 2 4 6 8 10 Column Ascentis Express ILIC Plates/Column - Synephrine 18500 Ascentis Silica 12878 Other ILIC 10013 0 2 4 6 8 10 12 14 4 1 Other porous ILIC 2 3 0 2 4 6 8 10 12 14

32 Ascentis Express Stability A Customers Experience

Summary Ascentis Express columns with 2.7 µm Fused-Core Particles offer the chromatographer a real Speed Advantage - Greater resolving power per unit pressure on any system Almost double the efficiency of 3 µm columns with totally porous particles Close to half the back pressure of sub 2 µm particles possible use of 2x longer columns - Increase the speed of analysis while maintaining resolution Use a column of half the length or smaller while maintaining the same or better resolving power than 5 µm & 3 µm columns Get the same speed as sub 2 µm columns but at half the pressure Or increase flow and double the speed (or better) at the same pressure - Benefit from the ruggedness and durability of traditional PLC columns 33

Acknowledgements r. illel Brandes r avid Bell ugh Cramer r. Richard A. enry r. Wayne Way r. Russel Gant r. Paul Ross r. William Campbell r. A. aniel Jones, MSU 34

Thank you! C18 C8 ILIC RP-Amide (coming soon) 35 Seminar Offer 25% off for a Test Column