Shukla Medical. TITLE: Manual Surgical Orthopedic Instruments Recommendations for Care, Cleaning, Maintenance and Sterilization

Similar documents
Prestige Endoscopic Graspers Instructions for Use

AMBLER SURGICAL INSTRUCTIONS FOR USE - OPHTHALMIC DUET-STYLE SCISSORS

Suction Regulators. Recommendations for Cleaning, Disinfection and Reprocessing: Boehringer 3800 &3900 Series Suction Regulators

Hercules 3 Universal Stabilizer Arm. Catalog Number

Urgent Field Safety Notice (Field Safety Corrective Action) for CORAIL AMT Neck Trials Surgical Instruments

McLane / Black Lake Fire Department Policy & Procedural Guideline PPG #

Competency Record for Sonic Cleaning

Chemical Spill Clean-up Protocol

TJF-Q180V Cleaning and Disinfection Checklist

Client Care Items and Equipment

TJF-Q180V Cleaning and Disinfection Checklist

Frigitronics CRYO-PLUS

SUGGESTED POLICY FOR THE MONITORING OF THE CLEANING EFFICIENCY OF AN ULTRASONIC CLEANERS( SONIC & TOSI) WEEKLY

Turnout gear is comprised of boots, coat, pants, suspenders, hood, gloves, helmet, and SCBA Facepiece.

LM-900 Cryosurgery System

The health hazards of HF are dependent upon the concentration, temperature and nature of exposure.

Technical Service Bulletin March 2015 TSB142.01

WELLS JOHNSON HIGH VOLUME CANISTERS

Vertebrate Animal Biosafety Level 1 Criteria

User Manual POINT SETTER MODEL:

ESCONDIDO FIRE DEPT TRAINING MANUAL Section Truck Module Page 1 of 5 Ropes Care and Maintenance Revised

VALVES & MEASUREMENT

Working Safely with Hydrofluoric Acid

UNIVERSITY OF ROCHESTER MEDICAL CENTER

COMMON PRECAUTIONS Data Sheet - Be Sure To Read Before Handling Page 1 of 6

To provide adequate disinfection for reusable semi-critical equipment other than scopes. Processing of scopes is covered in

Biological Safety Cabinet Guidelines

Air Removal from the Sterilization Chamber

INFUSION SYSTEM (110V) (220V) USER S MANUAL

GLIDERITE RIGID STYLET. Operations & Maintenance Manual

Biological materials manipulated in the laboratory: Bacteria Virus Protozoan

AFT25 Facemask with Two-Way Non-Rebreathing Valve

Laboratory Emergency Response & Incident Reporting Guide. Occupational & Environmental Safety Office Duke University Health System.

U.S. CONCRETE, INC. SAFETY POLICY and PROCEDURE MANUAL

O&M Receiving and Inspection. Storage of Water Treatment Media

WHEATLEY Series 822/820 Swing Check Valve

Essential Chemical Safety & Hygiene Training 2015

Standard Operating Procedure (SOP)

Hydrofluoric Acid SOP

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

Autoclear Mix & Matt

Heavy Duty Chlorinated Alkaline Foam Cleaner. #1 Trincity Drive Trincity Industrial Estate Trincity Trinidad and Tobago

USER MANUAL. AirPal Air Supply. (AS-1100 North America) 1488 Limeport Pike, Coopersburg, PA airpal.com fax

Effective Date: 24-April-2017 Version: 6 Page: 1 of 8 TITLE: SAF-370, PERSONAL PROTECTIVE EQUIPMENT GUIDELINES. Table of Contents

Fingertip Pulse Oximeter

ENVIRONMENTAL HEALTH AND SAFETY PERSONAL PROTECTIVE EQUIPMENT

Tools to Enhance Laboratory Biosafety and Biosecurity Performing Risk Assessments

1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique.

PERSONAL PROTECTIVE EQUIPMENT GUIDANCE

Aseptic Techniques. Techniques for Sterile Compounding. Pharmacy Technician Training Systems Passassured, LLC

Pressure Cell, French STANDARD OPERATING PROCEDURE

Hazardous Materials Spill Response

Type BBS-03, BBS-05, BBS-06, BBS-25

Household, Institutional and Industrial Disinfectant Supplier Identification:

University of Delaware Respirable Crystalline Silica Dust Program June 2018

Operating Manual for. SUPPORT jacks

OPERATIONAL GUIDELINE P R O C E D U R E

SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER

Ultima Salt System Overview

WHEATLEY Series 500 Swing Check Valve

Print PERSONAL PROTECTIVE EQUIPMENT

VCH Ebola Virus Disease Health Care Worker PPE Checklist/Sign In-Out Sheet

Hazardous Materials Spill Response

Cabinet. 1.0 POLICY It is the policy of Calvin College to comply with the MIOSHA Respiratory Protection Standard Part 451.

Laboratory Personal Protective Equipment & Hazard Assessment Tool

Laboratory Safety Training. Environmental Health and Safety

Ebola Virus Disease PPE Checklist/Sign In-Out Sheet Higher Transmission Risk Option 2: PAPRS

MATERIAL SAFETY DATA SHEET

Phalanx Alpha Phalanx Beta and Ultra Elite

USA & Canada: Outside USA & Canada: Not a dangerous substance according to GHS. Not a hazardous substance or mixture.

TEMPLE UNIVERSITY ENVIRONMENTAL HEALTH AND RADIATION SAFETY

Right-Angle Tube Roller Model

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

INSTRUCTION MANUAL FOR EBA MINUTE SELF-CONTAINED SELF RESCUER (SCSR) Manual Number: NH13747 Revision: A, ECN 2384 September 7, 2001

Service and Repair Manual

CLASS CYCLE P8000 OWNER'S MANUAL JOHNSON HEALTH TECH. CO., LTD.

WARNING! DO NOT THROW AWAY THESE INSTRUCTIONS! READ AND UNDERSTAND BEFORE USING EQUIPMENT!

E3628 THE RESPIRATORY PROTECTION PROGRAM: EMPLOYEE TRAINING. Leader s Guide ERI Safety Videos EMPLOYEE TRAINING

Animal Biosafety Level 2

32374 R. Jazowski 4/4/ D. W. Mudge G. Sivik 1 of 8 ODN TITLE: Installation & Operation Manual for Nautilus G2 Connectors

MATERIAL SAFETY DATA SHEET. COPPER CARE PRODUCT NAME: Aqua-Nap-5. ATTENTION: SAFETY MANAGER MSDS Number: Date of Last Revision: 04/10/2006

MONT BLANC 10400X XL XLED 10401X XL XLED 10403X XL XLED

INSTALLATION INSTRUCTIONS FOR SSI WIDEBAND DIFFUSER

User Guide. Pall Laboratory Manifold. For laboratory use. Not for use in a manner other than indicated. Introduction. Regulatory References

SEPTIC TANK & DRAINFIELD ACTIVATOR

SAFETY DATA SHEET Moisture absorber black

Standard Operating Procedure

Hydrofluoric Acid. Syracuse University Laboratory Standard Operating Procedure. I. Purpose: Physical and Chemical Properties: Potential Hazards:

Cavitron DualSelect. Introduction ENGLISH. Directions For Use. Cavitron DualSelect Dispensing System ENGLISH 1

Underwater Housing for Canon EOS M

NUS Safety & Health Guidelines Procedure No: OSHE/SOP/GL/08

3M DICHROIC Glass Finishes

Pectoral Machine. User manual E S S E N T I A L S T R E N G T H

Active Chemical Pty Ltd

CALCIUM CHLORIDE. Accelerator, adsobant, anti-freeze, mould inhibitor, food additive.

"SAFETY SHOWERS AND EYE WASHES"

AR-15 Rifle Operator Maintenance

Safety data-sheet according to Regulation (EC) No. 1907/2006 (REACH)

Instructions for Cleaning Equipment: Low Level Disinfection. Beaumont Hospital, Dearborn Infection Prevention & Control 2017

STAINLESS STEEL TABLES

Transcription:

Page 1 of 5 1. PROCESSING INSTRUCTIONS These processing instructions are intended to assist the hospital and central supply management in developing procedures for safe and effective reprocessing of both, hospital owned and loaned Shukla Medical instrument sets. A. WARNINGS AND PRECAUTIONS Universal precautions should be observed by all hospital personnel that work with contaminated or potentially contaminated medical devices. Caution should be exercised when handling devices with sharp points or cutting edges. Personal Protective Equipment (PPE) should be worn when handling or working with contaminated or potentially contaminated materials, devices and equipment. PPE includes gown, mask, goggles or face shield, gloves and shoe covers. Metal brushes or scouring pads must not be used during manual cleaning procedures. These materials will damage the surface and finish of instruments. Soft-bristled, nylon brushes and pipe cleaners should be used. Do not place heavy instruments on top of delicate devices. Do not allow contaminated devices to dry prior to reprocessing. All subsequent cleaning and sterilization steps are facilitated by not allowing blood, body fluid, bone and tissue debris, saline, or disinfectants to dry on used instruments. Remove excess body fluids and tissue with a disposable, non-shedding wipe and cover with a damp cloth. Automated cleaning may not be effective. A thorough manual cleaning process is recommended Instruments should be removed from trays and cleaned separately Cleaning agents with chlorine or chloride as the active ingredient are corrosive to stainless steel and must not be used. Enzymatic and cleaning agents with neutral ph are recommended

Page 2 of 5 B. POINT OF USE PREPARATION, CONTAINMENT AND TRANSPORTATION Remove excess body fluids and tissue from instruments with a disposable, non-shedding wipe. Place instruments in a basin of distilled water or in a tray covered with damp towels. Do not allow saline, blood, body fluids, tissue, bone fragments or other organic debris to dry on instruments prior to cleaning Instruments should be cleaned within 30 minutes of use to minimize the potential for drying prior to cleaning Used instruments must be transported to the central supply in closed or covered containers to prevent unnecessary contamination risk C. PREPARATION OF CLEANING AGENTS Neutral ph enzymatic and low foaming cleaning agents are preferred and recommended by Shukla Medical All cleaning agents should be prepared at the use-dilution and temperature recommended by the manufacturer. Softened tap water may be used to prepare cleaning agents. Use of recommended temperatures is important for optimal performance of cleaning agents. Fresh cleaning solutions should be prepared when existing solutions become grossly contaminated (bloody and/or turbid) D. MANUAL CLEANING PROCESS Use the neutral ph enzyme soaking solution that has been prepared Completely submerge the instrument(s) in enzyme solution and allow it to soak for 20 minutes. Use a soft-bristled brush to gently clean the device (particular attention shall be given to crevices, lumens, mated surfaces and other hard-to-clean areas) until all visible soil has been removed. Lumens should be cleaned with a long, narrow, soft-bristled brush (i.e. pipe cleaner brush) NOTE: The enzyme solution should be changed when it becomes grossly contaminated (bloody and/or turbid)

Page 3 of 5 Remove the device from the enzyme solution and rinse in purified water (from one or any combination of the following processes: ultra-filter, RO, DI and/or distilled) for a minimum of 3 minutes. Thoroughly flush lumens, holes and other difficult to reach areas Rinse instrument in purified water (from one or any combination of the following processes: ultra-filter, RO, DI and/or distilled) thoroughly for at least 3 minutes or until there is no sign of blood or soil in the rinse stream Dry the instrument with a clean, disposable, absorbent, non-shedding wipe E. AUTOMATED CLEANING PROCESS Automated washer/disinfector systems are not recommended as the sole cleaning method for surgical instruments. Orthopedic instruments should be cleaned following the manual cleaning procedure outlined in this document An automated washer/disinfector may be used as follow-up to the manual cleaning procedure above but is not necessary F. INSPECTION, MAINTENANCE AND LUBRICATION Carefully inspect each device to ensure that all visible contamination has been removed. If contamination is noted repeat the cleaning/disinfection process Visually inspect for device integrity, damage and/or excessive wear NOTE: If damage or wear is noted that may compromise the function of the instrument, contact Shukla Medical for a replacement Check the action of moving parts (e.g. hinges, sliding parts, etc.) to ensure smooth operation throughout the intended range of motion Hinged, rotating, sliding or articulating instruments should be lubricated with a water soluble product (e.g. Instrument Milk or equivalent lubricant) intended for surgical instruments that must be sterilized. To remain effective, the expiration date specified by the manufacturer should be adhered to for both stock and use-dilution concentrations NOTE: Mineral oil or silicone lubricants should not be used because they 1) coat microorganisms; 2) prevent direct contact of the surface with steam; and 3) are difficult to remove.

Page 4 of 5 G. STERILIZATION INSTRUCTIONS Steam sterilize using a pre-vacuum cycle for 4 minutes at a minimum temperature of 132 C (270 F). When sterilizing multiple instruments in one steam sterilization cycle, ensure that the sterilizer manufacturer s maximum load is not exceeded. Drying times will vary according to load size and should be increased for larger loads Flash sterilization by exposure at 270 F (132 C), unwrapped, in a gravity displacement sterilizer should only be used as an emergency procedure Steam sterilization is the preferred method for metal instrument sets. Instrument sets should be properly prepared and packaged in a case or tray that will allow steam to penetrate and make direct contact with all surfaces The following charts summarize exposure times and temperatures that are customarily recommended by manufacturers of steam sterilizers for metal instruments sterilized alone or in combination with porous materials. SHUKLA medical systems have been validated to meet these standards. Time and temperature relationships indicate holding time after the specific temperatures have been reached and do not include heating or drying times Cycle Type Temperature 4 Pressure 5 Exposure Time 6 Wrapped 7 Unwrapped 8 Dry Time 1,2 Prevacuum/ Pulsating Vacuum 132 C 270 F 1.86bar 27psi 4 min 4 min 45 mins 2,3 Prevacuum/ Pulsating Vacuum 134 C 273 F 3bar 28.5psi 18 min 18 min 1 validated steam sterilization temperature required to achieve a 10-6 sterility assurance level (SAL). 2 Local or national specifications should be followed where steam sterilization requirements are stricter or more conservative than those listed in this table 3 Disinfection/steam sterilization parameters recommended by the World Health Organization (WHO) for reprocessing Instruments where there is concern regarding TSE/CJD contamination 4 Sea Level 5 AAMI/AORN steam sterilization cycles with longer times than those listed are also acceptable 6 Rigid sterilization container that complies with ANSI/AAMI ST46 7 Flash (unwrapped) sterilization by exposure at 132 C /270 F should only be used as an emergency procedure. Instruments must be cleaned and disassembled 8 Drying times vary according to load size and should be increased for larger loads SHUKLA Medical Validation reports: S9KNE-0455-01-R01 Knee System Report MKS1022-0455-01-R01 Broken Stripped System Report MKS1031-0455-01-R01 Spine System validation

Page 5 of 5 H. STORAGE INSTRUCTIONS Sterile, packaged instruments should be stored in a designated, limited access area that is well ventilated and provides protection from dust, moisture, insects, and temperature/humidity extremes Sterile instrument packages should be carefully examined prior to opening to ensure that package integrity has not been compromised 2. HOSPITAL RESPONSIBILITIES FOR SHUKLA MEDICAL LOANER/RENTAL SETS Loaner/rental sets should undergo all steps of decontamination, cleaning, disinfection, inspection, and terminal sterilization before being returned to Shukla Medical. All rental sets are sent out with a Decontamination Certificate that is to be filled out and returned with the set along with any supplementary evidence of decontamination. Missing or damaged instruments from loaner sets should be brought to the attention of the operating room supervisor, to the director of the central supply department, and to Shukla Medical to ensure that the missing/damaged instrument(s) are backfilled. 3. CUSTOMER SERVICE INFORMATION SHUKLA MEDICAL 151 OLD NEW BRUNSWICK ROAD PISCATAWAY, NEW JERSEY 08854 TEL: +1 (888)-4SHUKLA E-MAIL: CUSTOMER.SERVICE@SHUKLAMEDICAL.COM WEBSITE: WWW.SHUKLAMEDICAL.COM